Astrodyne BSM400S-12 400w high density medical Datasheet

 400W High Density Medical/Industrial Grade Open Frame Power Supplies ASM400S/BSM400S Medical Safety Approval – 60601-­‐1 3rd Edition – 2 MOPP Class 2 for home use medical applications (BSM400) Single output with 5V Auxiliary and 12V fan output 400 Watts in compact 3”x 5”x 1.4” footprint High Efficiency, up to 92% @ 230Vac High Power Density up to 19W/Inch3 Universal 90-­‐264 VAC INPUT, 50/60 Hz 230W rating with natural convection
ORDERING INFORMATION — SEE PAGE 6 Model Output Voltage Output Current, A 5V Aux. A Forced/Natural Convection Forced/Natural Convection 12V Fan Output current, A Forced/Natural Convection Efficiency, typ. 230/115VAC A(B)SM400S-­‐12 12 VDC 33.3/19.17 2A/1A 1A/0.5A 90/86 A(B)SM400S-­‐15 15 VDC 26.67/15.34 2A/1A 1A/0.5A 90/86 A(B)SM400S-­‐20 20 VDC 20/12.5 2A/1A 1A/0.5A 91/88 A(B)SM400S-­‐24 24 VDC 16.7/10.4 2A/1A 1A/0.5A 91/88 A(B)SM400S-­‐28 28 VDC 14.3/9 2A/1A 1A/0.5A 91/88 A(B)SM400S-­‐36 36 VDC 11.11/7 2A/1A 1A/0.5A 91/88 A(B)SM400S-­‐48 48 VDC 8.33/5.2 2A/1A 1A/0.5A 91/88 A(B)SM400S-­‐54 54 VDC 7.4/4.6 2A/1A 1A/0.5A 91/88 PRODUCT DESCRIPTION The Astrodyne ASM and BSM 400 Series are miniature open frame power supplies designed for medical applications. The ASM400 is Class 1 and BSM400 is Class 2 with 2 MOPP (method of patient protection) isolation and BF leakage current and have been certified by Underwriters Laboratories for compliance with the latest edition of the international medical safety standard, IEC 60601-­‐1 3rd Edition using the CB reporting scheme. They are also certified to be compliant with the collateral standard 60601-­‐1-­‐2 for EMC and bear the UL Recognized component marks for North America and the EU and the CE mark. These products operate over the input voltage range of 90 to 264 VAC and 50, 60Hz frequency and produce 400 Watts of regulated DC output power in a standard 3 x 5 (inch) form factor that is 1U height compatible. Page 1 of 7 A(B)SM400S series 08-­‐25-­‐15 REV.J www.astrodynetdi.com ©2015 Astrodyne Corporation 400W High Density Medical/Industrial Grade Open Frame Power Supplies ASM400S/BSM400S INPUT SPECIFICATIONS Input Voltage Range Range of Nominal Input Voltages Input Frequency Input Current Inrush Current 90-­‐264 VAC 100-­‐240 VAC 47-­‐63 Hz (50/60 Hz Nom.) 4.5 A Max at 115VAC 2.5A Max at 230VAC 30A Max at 115VAC, 60 Hz 60A Max at 230VAC, 50 Hz 300uA Max at 264VAC, 50Hz Earth Leakage Current ASM400S Patient Leakage Current 100uA Max at 264VAC, 50Hz BSM400S BF rating Input Fusing 8A fuse in both L and N lines Power Factor 0.95 min., 230VAC 50Hz MAIN OUTPUT SPECIFICATIONS Output Voltage Output Power Minimum Load Set Point Accuracy Load Regulation Line Regulation Temp. Drift Transient Response Excursion Transient Response Recovery Time Ripple and Noise 12V, 15V, 20V, 24V, 28V, 36V, 48V or 54V nominal 400 W Continuous – See temp. & Airflow derating curves No minimum load required ± 1% ± 1% Max, 0 to Full Load ± 0.5% Max, 90 to 264 VAC ± 0.025 %/°C Less than ± 5% 50 to 100% Load Step 1A/us Slew Rate 2ms Max 50 to 100% Load Step 1A/us Slew Rate 1% pk-­‐pk Max. 20MHz BW Measured with 47uF Alum and 0.1uF Ceramic at output OVERALL SPECIFICATIONS Efficiency Standby Power Start-­‐up Delay Start-­‐up Rise Time Refer to Ordering Information table <1W 230 VAC 2s maximum 50ms maximum OVERALL SPECIFICATIONS (cont.) Hold-­‐up Time Power Density Switching Frequency MTBF 16ms typ. Full Load, 115VAC 19 W/in³ 200 KHz typ. 100K hrs. (typ.) per MIL-­‐
HDBK-­‐217F ISOLATION SPECIFICATIONS Input to Output Input to Earth Output to Earth 4000VAC, 2 MOPP 1500VAC, 1 MOPP 500VAC PROTECTION Over Current Inception Short Circuit Over Voltage Protection Over Temperature Protection 105 to 135% Rated Current Hiccup Mode, Automatic recovery 130% Vo max. Latching, Recycle Input to Reset Automatic recovery MECHANICAL SPECIFICATIONS Size Weight Input Connector Input Mating Connector Output Connector Output Mating Connector See Outline Drawings for mechanical options 1lbs. (453.6g) Molex 41791 Housing Molex 2139 Contact 2478 See Outline Drawings for mechanical options See Outline Drawings for mechanical options ENVIRONMENTAL SPECIFICATIONS Operating Temp. Range Storage Temp. Range Humidity -­‐20 to +40°C at Full Load with derating -­‐40 to +85°C 0 to 95%, non-­‐condensing Page 2 of 7 A(B)SM400S series 08-­‐25-­‐15 REV.J www.astrodynetdi.com ©2015 Astrodyne Corporation 400W High Density Medical/Industrial Grade Open Frame Power Supplies ASM400S/BSM400S ENVIRONMENTAL SPECIFICATIONS (cont.) Altitude Shock Vibration 0 to 10,000 ft. 0 to 3048 m 30G pk. Half sine, 6 axis 2 G RMS, 5 Hz to 500 Hz 3 axis, 30 min SAFETY CERTIFICATIONS UL/cUL UL EU AAMI ES60601-­‐1: 2005 A1 2012/CSA 22.2 60601-­‐1 2014 EN60601-­‐1: 2006 3rd Edition A1 2013 CB Scheme IEC 60601-­‐1: 2005 A1 2012 EMC CERTIFICATIONS EMC CERTIFICATIONS (cont.) Radiated Emissions ESD Susceptibility Air Discharge ESD Susceptibility Contact Discharge Radiated Susceptibility EFT/Burst Surge Conducted Susceptibility EN60601-­‐1-­‐2 Class B EN55011/A1 Class B EN61000-­‐4-­‐2 Criteria A Level 3 EN61000-­‐4-­‐2 Criteria A Level 2 EN61000-­‐4-­‐3 Criteria A Level 2 EN61000-­‐4-­‐4 Criteria A Level 3 EN61000-­‐4-­‐5 Criteria A Level 2 EN61000-­‐4-­‐6 Criteria A Level 2 Conducted Emissions EN60601-­‐1-­‐2 Class B EN55011/A1 Class B All Specifications are typical at nominal input, full load, 25°C unless specified otherwise. • For EMC Compliance, electrically bond 4 mounting holes to a conductive surface. 18mm
Air Flow
50mm
Ambient Thermal
Sensor AC IN
PUT
PSU
DC OUT
PUT
Note: Keep 6mm minimum between PCB bottom side to system chassis
18mm
Page 3 of 7 A(B)SM400S series 08-­‐25-­‐15 REV.J www.astrodynetdi.com ©2015 Astrodyne Corporation 400W High Power Density Medical Grade Open Frame Power Supplies ASM400S/BSM400S OUTPUT DERATING Output Power Derating Information
Natural Convection -­‐ Output Power vs. Ambient Temperature and Input Voltage for all models: Forced Convection -­‐ Output Power vs. Ambient Temperature, Airflow and Input Voltage:
Page 4 of 7 A(B)SM400S series 08-­‐25-­‐15 REV.J www.astrodynetdi.com ©2015 Astrodyne Corporation 400W High Power Density Medical Grade Open Frame Power Supplies ASM400S/BSM400S Page 5 of 7 A(B)SM400S series 08-­‐25-­‐15 REV.J www.astrodynetdi.com ©2015 Astrodyne Corporation 400W High Power Density Medical Grade Open Frame Power Supplies ASM400S/BSM400S PART NUMBERING INFORMATION: Input Class/Product Family (A or B) A: ASM Class I Input B: BSM Class II Input SM400S – – DC Output (2 characters) 12: 12V 24: 24V 19: 20V 54: 54V Aux. Output Configuration (0, 1 or 2 characters) Blank: +5V Aux. and +12V Fan Outputs both provided A1: +12V Fan Output provided, +5V Aux. not provided 00: Neither Aux. Output Provided Z: Neither +5V Aux. or +12V Fan Output provided Note:
1. PS-ON: Connect this signal to DC_OUT_RTN to enable the main
and FAN outputs. The 5V_AUX output is on when AC is applied.
2. PWR_OK: Open collector logic goes to high 160ms (typ.) after main output
is in regulation.
Page 6 of 7 A(B)SM400S series 08-­‐25-­‐15 REV.J www.astrodynetdi.com ©2015 Astrodyne Corporation 400W High Power Density Medical Grade Open Frame Power Supplies ASM400S/BSM400S MECHANICAL DRAWING Notes: 1.
2.
All dimensions in mm (in.) Maximum Dimensions: Length and Width Tolerance is ± 0.2 (0.008) Maximum Height: 36.1 (1.42) Page 7 of 7 A(B)SM400S series 08-­‐25-­‐15 REV.J www.astrodynetdi.com ©2015 Astrodyne Corporation 