Arch MQF500O-12S Ac-dc medical switching power supply Datasheet

VER : A_0
update : 2016.07.14
AC-DC Medical Switching Power Supply
MQF500O SERIES
500 Watts
KEY FEATURES
 Open Frame Medical Switching Power Supply
 Remote ON/OFF Function
 240 Watt with Free Air Convection
 500 Watt with 30CFM FAN
 Built-in 12V/0.3A Auxiliary Output
 Standby 5V@1A with Fan, @0.4A without Fan
 High Efficiency up to 93%
 With P.F.C. Function >0.95
 Ultra Compact Size: 5.03 x 3.0 x 1.38 Inches
 3-Year Product Warranty
(Pending)
ELECTRICAL SPECIFICATIONS
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
MQF500O-12S
Max Output Wattage
90-264 VAC or 127-370 VDC
Frequency (Hz)
47-63 Hz
Current (Full load)
<6.3 A max. (115 VAC) / <3.15 A max. (230 VAC)
Inrush Current (<2ms) (Clod Start)
< 40 A max. (115 VAC) / < 80 A max. (230 VAC)
Leakage Current
< 0.1 mA max. (Input-Output)
Power Factor (at 230 VAC)
PF>0.95 at Full Load
Voltage (V.DC.)
12V
Voltage Accuracy
±2%
Voltage Adj. Range (V.DC)
Current (with 30CFM FAN) (A) max
Current
(Free air Convection) (A) max
Output
Protection
Isolation
Environment
MQF500O-48S
230 W (115 VAC) / 240 W (230 VAC)
Max Output Wattage (W)
Voltage
Input
MQF500O-24S
500 W (30CFM FAN)
(W)
24V
48V
11.52~12.48
23.04~24.96
46.08~49.44
41.5
20.8
10.41
at 115 VAC
19.16
9.58
4.8
at 230 VAC
20
10
5
2,500µF
Line Regulation (115-264 VAC)
±0.5%
Load Regulation (10-100%) (typ.)
±1%
Minimum Load
3%
Maximum Capacitive Load
10,000µF
5,000µF
Ripple & Noise (typ.)
160mV
240mV
480mV
Efficiency (at 230 VAC)
90.5%
92%
93%
Hold-up Time (at 115 VAC)
8 ms min.
Over Power Protection
Auto recovery
Over Voltage Protection
Auto recovery
Overt Temperature Protection
Auto recovery
Short Circuit Protection
Auto recovery
Input-Output (V.AC)
4000VAC or 5656VDC
Input-FG
(V.AC)
2000V
Output-FG
.(V.AC)
1500V
Operating Temperature
-30°C…+70°C (with derating)
Storage Temperature
-35°C…+85°C
Temperature Coefficient
±0.03%/°C ( 0~50°C )
±0.06%/°C ( -30~0°C )
Humidity
95% RH
MTBF
>160,000 h @ 25°C (MIL-HDBK-217F)
Vibration
10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes.
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-1-
VER : A_0
update : 2016.07.14
AC-DC Medical Switching Power Supply
MQF500O SERIES
500 Watts
ELECTRICAL SPECIFICATIONS
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
MQF500O-12S
Dimension (L x W x H)
Physical
Safety
EMC
MQF500O-24S
MQF500O-48S
( 127.8 x 76.2 x 35.0 mm ) Tolerance 0.5 mm
5.03 x 3.0 x 1.38 Inches
Weight
480 g
Cooling Method
Conducted and radiated EMI
Free convection / 30 CFM FAN
cUL / UL Standard:
UL 60950-1, CAN/CSA C22.2 No. 60950-1-07
ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10), CAN/CSA-C22.2 No. 60601-1
(2008), 2 x MOPP
EN55011 class B, Radiated Class A (Pending)
ESD
EN61000-4-2 air ± 8kV , Contact ± 4Kv
Radiated Immunity
EN61000-4-3 10V/m
Fast Transient
EN61000-4-4 ± 2kV
(Pending)
Surge
EN61000-4-5 ±1kV
(Pending)
Conducted Immunity
EN61000-4-6 10Vrms
PFMF
EN61000-4-8 30A/m
Dips
EN61000-4-11 30% 10ms
Interruption
EN61000-4-11 >95% 5000ms
Approval
(Pending)
(Pending)
(Pending)
(Pending)
(Pending)
(Pending)
NOTE
1. Ripple & Noise are measured at 20MHz of bandwidth with ceramic 0.1uF & chemi-con KY 47uF parallel capacitor.
A 30cm twisted pair of no.18 AWG copper wire is connected to a
47uF and 0.1uF capacitor of proper polarity and voltage rating.
The oscilloscope probe ground led should connect right to the
ground ring of the probe and be as short as possible.
The oscilloscope bandwidth should be at 20MHz and connected
to AC ground.
2. It's recommended to add Varistor 14S471K at L / N input side in parallel.
3. Hold-up Time measured at 90% Vout.
4. Main Vout >3% Load, 12V (Aux) / 0.3A., 12V (Aux) need 0.1A Minimum Load, Auxiliary voltage output ground 10.2~13.3V
5. Strongly recommend to conduct this test with DC Voltage. If customer wishes to test with AC Voltage,
please disconnect all Y-Capacitors within Arch power supply.
DERATING
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-2-
VER : A_0
update : 2016.07.14
AC-DC Medical Switching Power Supply
MQF500O SERIES
500 Watts
BLOCK DIAGRAM
MECHANICAL DIMENSION
( Top View )
PIN#
Single
1
FG
2
AC IN (N)
3
AC IN (L)
4
+DC OUT
5
-DC OUT
Connector Pin (FAN)
PIN#
Single
F1
+12V
F2
GND
Connector Pin (CN1)
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
PIN#
Single
C1
-5VSB
C2
+5VSB
C3
GND
C4
DC OK
C5
-RC
C6
+RC
C7
-S
C8
+S
-3-
VER : A_0
update : 2016.07.14
AC-DC Medical Switching Power Supply
MQF500O SERIES
500 Watts
FUNCTION DESCRIPITON of CN1
Pin No.
Function
C1
-5VSB
C2
+5VSB
Description
This pin connects to the negative terminal(-V). Return for DC-OK and -RC signal output.
Stand by voltage output ground 4.2~5.5V, referenced to pin C1(-5VSB). The maximum load current is 1A with Fan, 0.4A
without Fan..
This pin connects to the negative terminal(-V). Return for DC-OK and -RC signal output.
C3
GND
C4
DC OK
C5
-RC
This pin connects to the negative terminal(-V). Return for DC-OK and -RC signal output.
C6
+RC
C7
-S
C8
+S
Turns the output on and off by electrical or dry contact between pin C5 (-RC), Short: Power OFF, Open: Power ON.
Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be
twisted in pair to minimize noise pick-up effect.
Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be
twisted in pair to minimize noise pick-up effect.
DC-OK Signal is a DC output, referenced to pin C3(DC-OK GND).
FUNCTION MANUAL & APPLICATION NOTE
1. DC-OK Signal
Between
DC-OK and GND
Output
Status
3.7~6V
0~1V
ON
OFF
2. Remote Control
It can be turned ON/OFF by using the "Remote Control" function.
Between
+RC and -RC
Output
Status
SW ON (Short)
SW OFF (Open)
OFF
ON
http://www.archcorp.com.tw