ME20 20W Single Output External Power Medical Series Features Meets UL/EN/IEC60601-1-2, 4th ed. for EMC* Approved to EN/IEC/UL60601-1, 3rd edition with isolation levels which satisfy the 2 MOPP requirements. Meets DoE Efficiency Level VI Requirements Up to 20W of AC-DC Power Universal Input 90-264Vac Input Range LPS No load input power Average Efficiency Desktop and Wall-Plug versions Meets EN55011/CISPR11, FCC Part 15.109 Class B Conducted & Radiated Emissions, with 6db margin E-cap life of >10 years >1,000,000 Hours MTBF 3 Year Warranty IP22 Rated Enclosure Description A high performance AC to DC external power supply family designed for medical applications. The ME20A Medical Series low power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition with isolation levels * which satisfy the 2 MOPP requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC . The ME20A Series models will operate at universal input range of 90 to 264Vac over the wide temperature range of -20°C to +50°C, delivering full rated output power up to +40°C and applicable output power derating at 50°C. These models are available in desktop and wall-plug versions; include an IP22 rating per IEC60529 for the enclosure, and output cable terminated at a variety of output connectors. These models use only high quality electrolytic capacitors, providing greater than 10 years life operating at rated output conditions. *Consult Factory for Table 9 compliance information. Model Selection Model Number ME20A0503F01 ME20A0703F01 ME20A0903F01 ME20A1203F01 ME20A1503F01 ME20A1803F01 ME20A2403F01 ME20A4803F01 ME20A0503N01 ME20A0703N01 ME20A0903N01 ME20A1203N01 ME20A1503N01 ME20A1803N01 ME20A2403N01 ME20A4803N01 www.slpower.com Volts 5.0V 7.5V 9.0V 12.0V 15.0V 18.0V 24.0V 48.0V 5.0V 7.5V 9.0V 12.0V 15.0V 18.0V 24.0V 48.0V Output Current 3.00A 2.00A 2.00A 1.50A 1.20A 1.00A 0.83A 0.42A 3.00A 2.00A 2.00A 1.50A 1.20A 1.00A 0.83A 0.42A Output Power 15W 15W 18W 18W 18W 18W 20W 20W 15W 15W 18W 18W 18W 18W 20W 20W Ripple & Noise¹ 75mV pk-pk 75mV pk-pk 90mV pk-pk 120mV pk-pk 150mV pk-pk 180mV pk-pk 240mV pk-pk 480mV pk-pk 75mV pk-pk 75mV pk-pk 90mV pk-pk 120mV pk-pk 150mV pk-pk 180mV pk-pk 240mV pk-pk 480mV pk-pk Line Regulation ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ME20 Rev 2.4 Load Regulation ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% 31-JAN-16 Output Connector Input Configuration 2.5 x 5.5 x 9.5mm Straight Barrel Type, center positive Class I Desktop, IEC60320 C14 Receptacle 2.5 x 5.5 x 9.5mm Straight Barrel Type, center positive Class II Desktop, IEC60320 C8 Receptacle 1 ME20 20W Single Output External Power Medical Series Model Selection (continued) Model Number ME20A0503Q01 ME20A0703Q01 ME20A0903Q01 ME20A1203Q01 ME20A1503Q01 ME20A1803Q01 ME20A2403Q01 ME20A4803Q01 ME20A0503B01 ME20A0703B01 ME20A0903B01 ME20A1203B01 ME20A1503B01 ME20A1803B01 ME20A2403B01 ME20A4803B01 ME20A0503C01 ME20A0703C01 ME20A0903C01 ME20A1203C01 ME20A1503C01 ME20A1803C01 ME20A2403C01 ME20A4803C01 Notes: Volts 5.0V 7.5V 9.0V 12.0V 15.0V 18.0V 24.0V 48.0V 5.0V 7.5V 9.0V 12.0V 15.0V 18.0V 24.0V 48.0V 5.0V 7.5V 9.0V 12.0V 15.0V 18.0V 24.0V 48.0V Output Current 3.00A 2.00A 2.00A 1.50A 1.20A 1.00A 0.83A 0.42A 3.00A 2.00A 2.00A 1.50A 1.20A 1.00A 0.83A 0.42A 3.00A 2.00A 2.00A 1.50A 1.20A 1.00A 0.83A 0.42A Output Power 15W 15W 18W 18W 18W 18W 20W 20W 15W 15W 18W 18W 18W 18W 20W 20W 15W 15W 18W 18W 18W 18W 20W 20W Ripple & Noise¹ 75mV pk-pk 75mV pk-pk 90mV pk-pk 120mV pk-pk 150mV pk-pk 180mV pk-pk 240mV pk-pk 480mV pk-pk 75mV pk-pk 75mV pk-pk 90mV pk-pk 120mV pk-pk 150mV pk-pk 180mV pk-pk 240mV pk-pk 480mV pk-pk 75mV pk-pk 75mV pk-pk 90mV pk-pk 120mV pk-pk 150mV pk-pk 180mV pk-pk 240mV pk-pk 480mV pk-pk Line Regulation ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% ±1% Load Regulation ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% Output Connector Input Configuration 2.5 x 5.5 x 9.5mm Straight Barrel Type, center positive Class II Desktop, IEC60320 C18 Receptacle 2.5 x 5.5 x 9.5mm Straight Barrel Type, center positive 2.5 x 5.5 x 9.5mm Straight Barrel Type, center positive Class II Wall-Plug, Interchangeable Blades (North American Blade included)² Class II Wall-Plug, Fixed North American Blades³ 1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and 6V models, values listed are typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point. 2. Order blade kit KT-1027K for other blades (EU. UK, Australia) 3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”. 4. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME20B0503F01). 5. All specifications are typical at nominal input, full load, at 25°C ambient unless noted. General Specifications AC Input Input Current Inrush Current Input Fuses Earth Leakage Current Efficiency 100-240Vac, ±10%, 47-63Hz, 1∅ Turn On Time Less than 700mS @115Vac, full load 115Vac: 0.5A, 230Vac: 0.25A Hold-up Time 20mS min., at full Load, 100Vac input 264Vac, cold start: will not exceed 40A F1, F2: 3.15A, 250Vac fuses (line & neutral lines) provided on all models Input-GND: <500μA@264Vac, 60Hz, NC Output-GND: <4mA@264Vac, 60Hz, NC 15 to 20W continuous – See models chart for specific voltage model ratings. No Load Input Power <0.1W per DoE Efficiency Level VI Requirements www.slpower.com Overload Protection Short Circuit Protection Overvoltage Protection >87%, typical Output Power Ripple and Noise Overtemperature Protection Isolation Safety Standards Operating Temperature See models chart on pg 1. ME20 Rev 2.4 31-JAN-16 Will shutdown upon an overtemperature condition, auto-recovery. 130 to 180% of rating, Hiccup Mode Hiccup Mode, auto recovery. 130 to 150% of output voltage, hiccup mode Input-Output: 2 MOPP Input-Ground: 1 MOPP Output-Ground: 1500Vac ANSI/AAMI ES60601-1:2005/(R)2012, CSA CAN/CSA-C22.2 NO, 60601-1:14; IEC606011:2005+CORR.2:2007+AM:2012; EN 606011:2006/A11:2011, EN60601-1: 2006/A12:2014, EN60601-1:2006/A1:2013, -20°C to +50°C Start Up at -40°C, full load, (warmup period before all parameters are within published specifications). 2 ME20 20W Single Output External Power Medical Series General Specifications (continued) Output Voltage Transient Response Regulation Drop Test Vibration Dimensions Weight Temperature Derating See models chart on pg 1. 500µs response time for return to within 0.5% of final value for any 50% load step over the range of 5% to 100% of rated load, ∆i/∆t< 0.2A/µs. Max. voltage deviation is +/-3.5%. Storage Temperature Altitude See models chart on pg 1. 1.4m from table top to wooden platform, 4 faces. Relative Humidity Operating: 0.003g/Hz, 1.5grms overall, 3 axes, 10 min/axis, 1-500Hz. Non-Oper.: random waveform, 3 minutes per axis, 3 axes and Sine waveform, Vib. frequency/acceleration: 10-500Hz/1g, sweep rate of 1 octave / minutes, Vibration time of 10 sweeps / axes, 3 axes Shock See outline drawings MTBF E-Cap Life TBDg Derate output power above 40°C to TBD at 50°C -40°C to +85°C Operating: to 5000m. Non-operating: -500 to 40,000 ft. 5% to 95%, non-condensing Operating: Half-sine, 20gpk, 10mS, 3 axes, 6 shocks total Non-Operating: Half-sine waveform, impact acceleration of 100G, Pulse duration of 6 mS, Number of shocks: 3 for each of the three axis >1.000,000 hours, full load, 110 & 220Vac input, 25°C amb., per Telcordia 332 Issue 6 (stress method). >7 year life based on calculations at 115Vac/60Hz & 230Vac/50Hz, ambient 25°C at 24 hrs per day, 365 days/year, 6 power up cycles per day. All specifications are typical at nominal input, full load, at 25°C ambient unless noted. EMI/EMC Compliance Conducted Emissions: EN55011/CISPR22 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac Radiated Emissions: EN55022/CISPR22 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac Common Mode Noise: High Frequency (100kHz-20MHz): <40mA pk-pk Electro-Static Discharge (ESD) Immunity on Power ports: EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A IEC60601-1-2, 4th Edition, Table 4 Radiated RF EM Fields Susceptibility EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz IEC60601-1-2, 4th Edition, Table 4 Electrical Fast Transients (EFT) /Bursts: EN55024/IEC61000-4-4, Level 4, +/-4kV, 100Khz rep rate, 40A, Criteria A IEC60601-1-2, 4th Edition, Table 5 Surges, Line to Line (Diff Mode) and Line to GND (CMN Mode) EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A Surpasses IEC60601-1-2, 4th Edition requirements. Conducted Disturbances induced by RF Fields EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz IEC60601-1-2, 4th Edition, Table 5 Rated Power frequency magnetic fields EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz IEC60601-1-2, 4th Edition, Table 4 Voltage Interruptions, Dips, Sags & Surges EN55024/IECEN61000-4-11: --100% dip for 20mS, Criteria A --100% dip for 5000mS (250/300 cycles), Criteria B --60% dip for 100mS, Criteria B --30% dip for 500mS, Criteria A IEC60601-1-2, 4th Edition, Table 5 Harmonic Current Emissions EN55011/EN61000-3-2, Class A Flicker Test EN61000-3-3 All specifications are typical at nominal input, full load, at 25°C ambient unless noted. www.slpower.com ME20 Rev 2.4 31-JAN-16 3 ME20 20W Single Output External Power Medical Series Mechanical Drawings IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector Notes: Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector 1. All dimensions in mm. 2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K. 3. The unit should not be covered or enclosed to protect against excessive case temperature rise. Connector Information Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive. Other standard options are listed below. The “03” in the standard model number is replaced by the applicable digits below. For other options, consult the factory. Data Sheet © 2016 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice. www.slpower.com ME20 Rev 2.4 31-JAN-16 4