HP E1412A 6.5-digit high-accuracy multimeter c-size Datasheet

Agilent E1412A
6.5-Digit High-Accuracy Multimeter
C-Size
Data Sheet
Features
• 1-Slot, C-size, message-based
• DCV, ACV, DCI, ACI, 2/4-wire Ω,
frequency, period
• NULL, MIN/MAX, LIMIT,
dB, dBm
• 1000 reading/s into internal
memory at 4.5 digits
• Fast range/function changes
• Reading storage with internal
memory
Description
The Agilent Technologies E1412A
6.5-Digit Multimeter is a C-size,
1-slot, message-based VXI module.
It is identical in electrical design
to the E1312A, differing only in
size. It delivers the widest functionality in Agilent’s DMM line.
It also delivers high performance
and Agilent high quality at
prices you’d expect to pay for
a 5.5-digit DMM.
This multimeter’s wide product
functionality includes volts,
amps, ohms, and frequency with
advanced tests including limit
checks to drive a TTL output and
dc voltage ratios. Standard measurements include ac/dc voltage,
ac/dc current, 2- and 4-wire Ω,
plus frequency/period. When
measuring dcV, this multimeter
can deliver 65 range changes
per second and 30 function
changes per second.
Refer to the Agilent Technologies
Website for instrument driver
availability and downloading
instructions, as well as for recent
product updates, if applicable.
DC Specifications
Specifications are for 1-hour warm-up at an integration time of 100 PLCs.
DC Summary:
DC voltage:
300 V max.
Voltage accuracy (dc):
± 0.0019%
DC Accuracy ± (% of reading + % of range): Specifications for 1-hour warm-up at 6.5 digits
DC voltage:
Range2
Test Current or
Burden Voltage
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
300.0000 V
24 Hour1
23°C ± 1°C
90 Day
23°C ± 5°C
1 Year
23°C ± 5°C
0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0018
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0030
0.0050 + 0.0035
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0030
Range
Resistance:3
Test Current or
Burden Voltage
24 Hour1
23°C ± 1°C
90 Day
23°C ± 5°C
1 Year
23°C ± 5°C
100.0000 Ω
1.000000 kΩ
10.00000 kΩ
100.0000 kΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ
1 mA
1 mA
100 µA
10 µA
5 µA
500 nA
500 nA || 10 MΩ
0.0030 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0005
0.0020 + 0.0005
0.002 + 0.001
0.015 + 0.001
0.300 + 0.010
0.008 + 0.004
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.035 + 0.001
0.800 + 0.010
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.054 + 0.001
0.800 + 0.010
Range
dc current:
Test Current or
Burden Voltage
24 Hour
23°C ± 1° C
90 Day
23°C ± 5° C
1 Year
23°C ± 5°C
10.00000 mA
100.0000 mA
1.000000 A
3.000000 A
<0.1 V
<0.6 V
<1 V
<2 V
0.005 + 0.010
0.01 + 0.004
0.10 + 0.006
0.70 + 0.020
0.050 + 0.020
0.040 + 0.005
0.130 + 0.010
0.720 + 0.020
0.070 + 0.020
0.070 + 0.005
0.150 + 0.010
0.720 + 0.020
Temperature Coefficient
0°C - 18°C
28°C - 55°C
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0003
Temperature Coefficient
0°C - 18°C
28°C - 55°C
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
Temperature Coefficient
0°C - 18°C
28°C - 55°C
0.005 + 0.0020
0.006 + 0.0005
0.005 + 0.0010
0.005 + 0.0020
DC:DC Ratio:
Range2
100 mV to 300 V: (Input Accuracy) + (Reference Accuracy)
Input Accuracy = accuracy specification for the HI-LO input signal
Reference Accuracy = accuracy specification for HI-LO reference input signal
1 Relative to calibration standards.
2 20% overrange on all ranges, except 300 Vdc and 3 A range.
3 Specifications are for 4-wire Ω function, or 2-wire Ω using Math Null, without Math Null, add 0.2 Ω additional error in 2-wire Ω function.
2
DC Specifications (continued)
Specifications are for 1-hour warm-up at an integration time of 100 PLCs.
DC Voltage Characteristics:
Measurement method:
Continuously integrating, multi-slope III A/D converter
A/D linearity:
0.0002% of reading + 0.0001% of range
2/4-wire Ω:
100 MΩ
Input resistance:
0.1 V, 1 V, 10 V ranges:
100 V, 300 V ranges:
Selectable 10 MΩ or 10 GΩ
10 MΩ ± 1%
Input bias current:
<30 pA at 25°C
Input terminals:
Copper alloy
Input protection:
300 V on all ranges
Resistance:
Measurement method:
Selectable 4-wire or 2-wire Ω
(Current source referenced to low input)
Max. lead resistance:
(4-wire Ω) 10% of range per lead for dc Current
100 Ω and 1 kΩ per lead on all other ranges
Input protection:
300 V on all ranges
Shunt resistor:
0.1 Ω for 1 A and 3 A, 5 Ω for 10 mA and 100 mA
Input protection:
Externally accessible 3 A, 250 V fuse
Measurement Noise Rejection:
60 Hz (50 Hz) (For 1 kΩ unbalance in LO lead.)
DC CMMR:
140 dB
Integration Time
100 PLC/1.67s (2s)
10 PLC/167 ms (200 ms)
1 PLC/16.7 ms (20 ms)
<1 PLC/3 ms (800 µs)
Normal mode rejection1
60 dB2
60 dB2
60 dB2
0 dB
1 For power-line frequency ± 0.1%.
2 For power-line frequency ± 1%, subtract 20 dB; for ± 3%, subtract 30 dB.
System Speed:
(Speeds are for 4.5 digits, Delay 0 and Autozero OFF. Includes measurement and data transfer over VXI backplane.)
Function change:
30/s
Range change:
65/s
Autorange time:
<30 ms
Max. internal trigger rate:
1000/s
Max. external trigger rate to memory:
1000/s
3
DC Specifications (continued)
Specifications are for 1-hour warm-up at an integration time of 100 PLCs.
DC:DC Ratio:
Measurement method:
Input HI-LO/Reference HI-LO
Apply “Reference HI-LO” signal to Ohms 4-Wire Sense terminals.
Input HI to Input LO:
100 mV to 300 V
Reference HI to Input LO:
<12 V on 100 mV to 10 V ranges (autoranged)
Reference LO to Input LO:
<2 V
Additional Error with Autozero OFF:
Following instrument warm-up at calibration temperature ± 1°C and <10 minutes.
100 mV-100 V ranges:
add 0.0002% reading + 5 µV
300 V range:
add 0.0006% reading
DC Operating Characteristics: Readings speeds for 60 Hz and (50 Hz) operation, Autozero OFF.
Function
NPLC
Digits
Reading/s
Additional Noise Error
DCV, DCI and Ω
100
6.5
0.6 (0.5)
0% of range
DCV, DCI and Ω
10
6.5
6 (5)
0% of range
DCV, DCI and Ω
1
5.5
60 (50)
0.001% of range*
DCV, DCI and Ω
0.2
5.5
300
0.001% of range*
DCV, DCI and Ω
0.02
4.5
1,000
0.01% of range*
*For 300 V range: use 0.003% of range for 5.5 digits and 0.030% of range for 4.5 digits.
For all ranges: add 20 µV for dc Volts, 4 µA for dc current, or 20 mΩ for resistance.
Considerations
Settling considerations:
Reading settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes.
Measurement considerations:
Agilent recommends the use of Teflon or other high impedance,
low-dielectric absorption wire insulation for these measurements.
4
AC Specifications
AC Summary
AC voltage:
300 V max.
Voltage accuracy (ac):
± 0.07%
AC Accuracy ± (% of reading + % of range): Specifications for 1-hour warm-up at 6.5 digits. Slow ac filter, sinewave input.
True RMS AC Voltage3:
Range2
Frequency
24 Hour1
23°C ± 1°C
90 Day
23°C ± 5°C
1 Year
23°C ± 5°C
Temperature Coefficient
0°C - 18°C
28°C - 55°C
100.0000 mV
3 Hz-5 Hz
1.00 + 0.03
1.00 + 0.04
1.00 + 0.04
0.100 + 0.004
100.0000 mV
5 Hz-10 Hz
0.35 + 0.03
0.35 + 0.04
0.35 + 0.04
0.035 + 0.004
100.0000 mV
10 Hz-20 kHz
0.04 + 0.03
0.05 + 0.04
0.06 + 0.04
0.005 + 0.004
100.0000 mV
20 kHz-50 kHz
0.10 + 0.05
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
100.0000 mV
50 kHz-100 kHz
0.55 + 0.08
0.60 + 0.08
0.60 + 0.08
0.060 + 0.008
100.0000 mV
100 kHz-300 kHz
5.00 + 0.50
5.00 + 0.50
5.00 + 0.50
0.020 + 0.020
1.000000 V to 300.000 V 4
3 Hz-5 Hz
1.00 + 0.02
1.00 + 0.03
1.00 + 0.03
0.100 + 0.003
1.000000 V to 300.000 V 4
5 Hz-10 Hz
0.35 + 0.02
0.35 + 0.03
0.35 + 0.03
0.035 + 0.003
1.000000 V to 300.000 V 4
10 Hz-20 kHz
0.04 + 0.02
0.05 + 0.03
0.06 + 0.03
0.005 + 0.003
1.000000 V to 300.000 V 4
20 kHz-50 kHz
0.10 + 0.04
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
1.000000 V to 300.000 V 4
50 kHz-100 kHz
0.55 + 0.08
0.60 + 0.08
0.06 + 0.08
0.060 + 0.008
1.000000 V to 300.000 V 4
100 kHz-300 kHz5
5.00 + 0.50
5.00 + 0.50
5.00 + 0.50
0.200 + 0.020
Range
Frequency
24 Hour
23°C ± 1°C
90 Day
23°C ± 5°C
1 Year
23°C ± 5°C
Temperature Coefficient
0°C - 18°C
28°C - 55°C
1.000000 A
3 Hz-5 Hz
1.05 + 0.04
1.05 + 0.04
1.05 + 0.04
0.100 + 0.006
1.000000 A
5 Hz-10 Hz
0.35 + 0.04
0.35 + 0.04
0.35 + 0.04
0.035 + 0.006
1.000000 A
10 Hz-1 kHz
0.15 + 0.04
0.15 + 0.04
0.15 + 0.04
0.015 + 0.006
1.000000 A
1 kHz-50 kHz
0.40 + 0.04
0.40 + 0.04
0.40 + 0.04
0.015 + 0.006
3.000000 A
3 Hz-5 Hz
1.70 + 0.06
1.70 + 0.06
1.70 + 0.06
0.100 + 0.006
3.000000 A
5 Hz-10 Hz
0.95 + 0.06
0.95 + 0.06
0.95 + 0.06
0.035 + 0.006
3.000000 A
10 Hz-1 kHz
0.75 + 0.06
0.75 + 0.06
0.75 + 0.06
0.015 + 0.006
3.000000 A
1 kHz-50 kHz
1.00 + 0.06
1.00 + 0.06
1.00 + 0.06
0.15 + 0.06
True RMS AC Current3:
1 Relative to calibration standards.
2 20% overrange on all ranges, except 300 Vac and 3 A ranges which have 1% overrange.
3 100 mV to 100 V range specifications are for sine wave input >5% of range. For inputs from 1% to 5% of range and <50 kHz, add 0.1% of range additional error.
For 50 kHz to 100 kHz, add 0.13% additional error. 300 V range specifications are for sinewave input >15% of range. For inputs from 3% to 15% of range and >50 kHz,
add 0.30% of kHz, add 0.40% of range additional error.
4 For 300 V range, use (% reading) shown in table and multiply each (% range) x 3.
5 300 Vac range limited to 50 kHz. For frequencies >50 kHz, signals must be ≤1.5 x 107 VHz.
5
AC Specifications (continued)
Additional AC Specifications
Low Frequency Errors (% of reading)
Frequency
Crest Factor Errors (non-sinewave)*
AC Filter
Slow
Medium
Fast
Crest Factor
Error (% of reading)
10 Hz-20 Hz
0
0.74
-
1-2
0.05%
20 Hz-40 Hz
0
0.22
-
2-3
0.15%
40 Hz-100 Hz
0
0.06
0.73
3-4
0.30%
100 Hz-200 Hz
0
0.01
0.22
4-5
0.40%
200 Hz-1 kHz
0
0
0.18
>1 kHz
0
0
0
*For frequencies below 100 Hz, slow ac filter specified for sinewave input only.
Noise Rejection: (For 1 kΩ unbalance in LO lead.)
AC CMMR:
70 dB
True RMS AC Voltage:
Measurement method: AC-coupled True RMS — measures the ac component of the input
with up to 300 Vdc of bias on any range. (Max ac+dc = 300 V rms.)
Crest factor:
Maximum 5:1 at full scale
AC filter bandwidth:
Slow:
3 Hz-300 kHz
Medium:
20 Hz-300 kHz
Fast:
200 Hz-300 kHz
Input impedance:
1 MΩ ±2%, in parallel with 100 pF
Input protection:
300 Vrms all ranges
True RMS AC Current:
Measurement method: Direct couple to the fuse and shunt. ac -coupled True RMS measurement
(measures the ac component only).
Shunt resistor:
0.1 Ω for 1 A and 3 A ranges
Burden voltage:
1 A range:
3 A range:
Input protection:
Externally accessible 3 A, 250 V fuse
<1 Vrms
<2 Vrms
6
AC Specifications (continued)
AC Operating Characteristics:
Maximum reading rates 0.01% of ac step additional error. Additional settling delay required when input dc level varies.
Function
Digits
Reading/s
ACV and ACI
6.5
7 s/reading
Slow
ACV and ACI
6.5
1
Medium
ACV and ACI
6.5
1.61
Fast
ACV and ACI
6.5
10
Fast
6.5
502
Fast
ACV and ACI
AC Filter
1 For External Trigger or remote operation using default settling delay (Delay Auto).
2 Maximum useful limit with default settling delays used.
Systems Speeds:
Maximum useful limit with default settling delays used; Speeds are for 4.5 digits, Delay 0, and Fast ac filter.
Function or range change:
5/s
Autorange time:
<0.8 s
ASCII reading to GPIB:
50/s
Max. internal trigger rate:
50/s
Max. external trigger rate to memory:
50/s
Additional Low-Frequency Errors (% of reading):
Input >100 mV. For mV input, multiply % of reading error x 10.
Frequency
6.5 digits
5.5 digits
4.5 digits
3 Hz-5 Hz
0
0.12
0.12
5 Hz-10 Hz
0
0.17
0.17
10 Hz-40 Hz
0
0.2
0.2
40 Hz-100 Hz
0
0.06
0.21
100 Hz-300 Hz
0
0.03
0.21
300 Hz-1 kHz
0
0.01
0.07
>1 kHz
0
0
0.02
7
Frequency and Period Specifications
Frequency and Period Accuracy (% of reading):
Specifications are for 1-hour warm-up at 6.5 digits.
Function
Range2
Frequency
24 Hour1
90 Day
1 Year
23°C ± 1°C
23°C ± 5°C
23°C ± 5°C
Temperature
Coefficient
0°C - 18°C
28°C - 55°C
Frequency, Period
100 mV to 300 V
3 Hz-5 Hz
0.10
0.10
0.10
0.005
Frequency, Period
100 mV to 300 V
5 Hz-10 Hz
0.05
0.05
0.06
0.005
Frequency, Period
100 mV to 300 V
10 Hz-40 Hz
0.03
0.03
0.03
0.001
Frequency, Period
100 mV to 300 V
40 Hz-300 kHz
0.006
0.01
0.01
0.001
1 Relative to calibration standards.
2 20% overrange on all ranges, except 300 Vac range which has 1% overrange.
Measuring Characteristics:
Measurement method:
Reciprocal-counting technique. AC-coupled input using the ac voltage measurement function.
Voltage ranges:
100 mV rms full scale to 300 V rms. Auto or manual ranges.
Gate time:
10 ms, 100 ms, or 1 s
Settling considerations:
Errors will occur when attempting to measure the frequency or period of an input following a
dc offset voltage change. The input blocking RC time constant must be allowed to fully settle
(up to 1 s) before the most accurate measurements are possible.
Measurement considerations: All frequency counters are susceptible to error when measuring low-voltage, low-frequency
signals. Shielding inputs from external noise pickup is critical for minimizing measurement errors.
Max. reading rate:
1K
8
Frequency and Period Specifications (continued)
Operating Characteristics:
Speeds are for 4.5 digits, Delay 0, and Fast ac filter.
Function
Digits
Reading/s
Frequency, Period
6.5
1
Frequency, Period
5.5
9.8
Frequency, Period
4.5
80
Systems Speeds:
Configuration rates:
14/s
Autorange time:
<0.6 s
Max. internal trigger rate:
80/s
Max. external trigger rate to memory:
80/s
Characteristics:
Warmup time:
1 hour
State storage memory:
Power-off state automatically saved
Functions:
Idc:
3A
Iac:
3A
Frequency:
300 kHz
Period:
3.3 µs
9
Ordering Information
General Specifications
Module Current:
VXI Characteristics:
VXI device type:
IPM
IDM
Message
based
+5 V:
0.2
0.1
Data transfer bus:
A16
+12 V:
0.7
0.06
Size:
C
-12 V:
0
0
Slots:
1
+24 V:
0
0
Connectors:
P1/2
-24 V:
0
0
Shared memory:
n/a
-5.2 V:
0
0
VXI buses:
n/a
-2 V:
0
0
Cooling Slot:
Instrument Drivers:
See the Agilent Technologies Website
http://www.agilent.com/find/inst_drivers
for driver availability and downloading.
Command module firmware:
n/a
Command module firmware rev:
n/a
I-SCPI Win 3.1:
n/a
I-SCPI Series 700:
n/a
C-SCPI LynxOS:
n/a
C-SCPI Series 700:
n/a
Panel Drivers:
No
VXIplug&play Win Framework:
Yes
Watts/slot:
9.40
∆P mm H2O:
0.05
Air Flow liter/s:
0.80
VXIplug&play Win95/NT Framework: Yes
VXIplug&play HP-UX Framework:
No
10
Description
Product No.
6.5-Digit Multimeter,
High Accuracy
E1412A
ANSI Z540
Compliant Calibration
E1412A A6J
3 year return to
Agilent to 1 year
OnSite warranty
E1412A W01
Remove all doubt
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
www.agilent.com/find/open
Agilent Open simplifies the process
of connecting and programming
test systems to help engineers
design, validate and manufacture
electronic products. Agilent offers
open connectivity for a broad range
of system-ready instruments, open
industry software, PC-standard I/O
and global support, which are
combined to more easily integrate
test system development.
www.lxistandard.org
LXI is the LAN-based successor to
GPIB, providing faster, more efficient
connectivity. Agilent is a founding
member of the LXI consortium.
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
www.agilent.com
For more information on Agilent Technologies’
products, applications or services, please
contact your local Agilent office. The complete
list is available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
Belgium
Denmark
Finland
France
0820 87 44 11
32 (0) 2 404 93 40
45 70 13 15 15
358 (0) 10 855 2100
0825 010 700*
Germany
01805 24 6333**
*0.125 € fixed network rates
**0.14 €/minute
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland (French)
41 (21) 8113811(Opt 2)
Switzerland (German) 0800 80 53 53 (Opt 1)
United Kingdom
44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: October 24, 2007
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2008
Printed in USA, January 31, 2008
5965-5563E