Keysight N9343C Handheld spectrum analyzer (hsa) Datasheet

Keysight N9343C
Handheld Spectrum Analyzer (HSA)
13.6 GHz
Data Sheet
02 | Keysight | N9343C Handheld Spectrum Analyzer (HSA) - Data Sheet
Field testing just got easier
If you are making measurements in the field, the Keysight Technologies, Inc. N9343C
handheld spectrum analyzer (HSA) makes your job easier. It's got the features you need
for operating in tough field environments, and its measurement performance gives you
confidence the job's been done right. The N9343C HSA lets you automate routine
tasks to save time and ensure consistent results. Field testing just got easier with the
Keysight N9343C HSA.
Definitions and requirements
This data sheet contains specifications and supplemental information for Keysight
N9343C handheld spectrum analyzer. The differences between specifications, typical
performance, and nominal values are described as follows.
Definitions
"Specifications" describe the performance of parameters covered by the product
warranty and apply to temperatures ranging from –10 to 50 °C, unless otherwise noted.
95th percentile values indicate the breadth of the population (> 2) of performance
tolerances expected to be met in 95% of the cases with a 95% confidence, for
any ambient temperature in the range of 20 to 30 °C. In addition to the statistical
observations of a sample of instruments, these values include the effects of the
uncertainties of external calibration references. These values are not warranted. These
values are updated occasionally if a significant change in the statistically observed
behavior of production instruments occurs.
"Typical" describes additional product performance information that is not covered
by the product warranty. It is performance beyond specification that 80% of the units
exhibit with a 95% confidence level over the temperature range of 20 to 30 °C. Typical
performance does not include measurement uncertainty.
"Nominal values" indicate expected performance, or describe product performance that
is useful in the application of the product, but is not covered by the product warranty.
Conditions required to meet specifications
The following conditions must be met for the analyzer to meet its specifications.
–– The analyzer is within its calibration cycle.
–– Under auto couple control, except when Swp Time Rule is set to Accuracy.
–– Any analyzer that has been stored at a temperature range inside the allowed storage
range but outside the allowed operating range must be stored at an ambient
temperature within the allowed operating range for at least two hours before being
turned on.
–– The analyzer has been turned on at least 30 minutes.
Certification
Keysight Technologies certifies that this product met its published specifications at
the time of shipment from the factory. Keysight Technologies further certifies that
its calibration measurements are traceable to the United States National Institute of
Standards and Technology (NIST), to the extent allowed by the Institute’s calibration
facility, and to the calibration facilities of other International Standards Organization
(ISO) members.
Your job just got easier:
–– Get the features you need in
a field-ready instrument.
–– Gain confidence in your
measurements with
benchtop performance in a
handheld instrument.
–– Innovative task planner
(www.keysight.com/find/
taskplanner) reduces
test setup time by 95%,
delivers test automation and
consistency, and makes it
easy to capture test results,
generate reports, and share
task plans with others.
03 | Keysight | N9343C Handheld Spectrum Analyzer (HSA) - Data Sheet
Specifications
Specification
Supplemental information
Frequency
Frequency range
1 MHz to 13.6 GHz (usable to 9 kHz)
AC coupled
Internal 10 MHz frequency reference accuracy
Aging rate
± 1 ppm/year
Temperature stability
± 1 ppm
Referenced to frequency reading at 25 °C.
Temperature varied at max. of 2 °C per
minute. Control voltage held at voltage
control range midpoint
Frequency readout accuracy with marker (start, stop, center, marker)
Marker resolution
(frequency span)/(sweep points - 1)
Uncertainty
± (frequency indication × frequency reference Frequency reference uncertainty = (aging
uncertainty + 1% × span + 20% × resolution
rate × period of time since adjustment +
bandwidth + marker resolution + 1 Hz)
temperature stability)
Marker frequency counter
Resolution
1 Hz
Accuracy
± (marker frequency × frequency reference
uncertainty + counter resolution)
RBW/span ≥ 0.02; marker level to displayed
noise level > 25 dB; frequency offset 0 Hz
Frequency span
Range
0 Hz (zero span), 100 Hz to 13.6 GHz
Resolution
1 Hz
Accuracy
± (0.22% × span + span/(sweep
points - 1))
Nominal
30 kHz
< –86 dBc/Hz, typical –89 dBc/Hz
20 to 30 °C
100 kHz
< –97 dBc/Hz, typical –99 dBc/Hz
Center frequency 500 MHz
1 MHz
< –115 dBc/Hz, typical –119 dBc/Hz
SSB phase noise
Carrier offset
Resolution bandwidth (RBW)
–3 dB bandwidth
10 Hz to 3 MHz
1-3-10 sequence
Accuracy
± 5%, RBW = 10 Hz to 1 MHz
Nominal
± 10%, RBW = 3 MHz
Resolution filter shape factor
< 5:1
Nominal; 60 dB/3 dB bandwidth ratio;
digital, Gaussian-like
EMI bandwidth (CISPR compliant)
200 Hz, 9 kHz, 120 kHz, 1 MHz
Option EMC required
Accuracy
± 10% nominal
Resolution filter shape factor
< 5:1 nominal
−60 dB/−6 dB bandwidth ratio
–3 dB bandwidth
1 Hz to 3 MHz
1-3-10 sequence
Accuracy
± 10%, VBW = 1 Hz to 1 MHz
Nominal
Video bandwidth (VBW)
04 | Keysight | N9343C Handheld Spectrum Analyzer (HSA) - Data Sheet
Specifications (continued)
Amplitude specifications
Supplemental information
Measurement range
1 to 500 MHz
Displayed average noise level (DANL) to +10 dBm
500 MHz to 13.6 GHz
Displayed average noise level (DANL) to +20 dBm
Input attenuator range
0 to 50 dB, in 5 dB steps
Preamp off
Maximum safe input level
Average continuous
power
+30 dBm, 3 minutes maximum
DC voltage
± 50 VDC maximum
Input attenuator setting
≥ 20 dB, 1 MHz to 13.6 GHz
Displayed average noise level 1
Preamp off
Normalized to 1 Hz
Minimum RBW
1 to 10 MHz
–125 dBm, typical –140 dBm
–115 dBm, typical –130 dBm
10 MHz to 3 GHz
–137 dBm, typical –142 dBm
–127 dBm, typical –132 dBm
3 to 7 GHz
–135 dBm, typical –140 dBm
–125 dBm, typical –130 dBm
7 to 10 GHz
–139 dBm, typical –142 dBm
–129 dBm, typical –132 dBm
10 to 13.6 GHz
–137 dBm, typical –140 dBm
–127 dBm, typical –130 dBm
1 to 10 MHz
–140 dBm, typical –156 dBm
–130 dBm, typical –146 dBm
10 MHz to 3 GHz
–150 dBm, typical –154 dBm
–140 dBm, typical –144 dBm
3 to 6 GHz
–145 dBm, typical –150 dBm
–135 dBm, typical –140 dBm
6 to 13.6 GHz
–151 dBm, typical –155 dBm
–141 dBm, typical –145 dBm
Reference level ≤ –50 dBm
Preamp on
Reference level ≤ –70 dBm
Level display range
Log scale
10 to 100 dB, 10 divisions displayed, 1, 2, 5, 10 dB/division
Linear scale
0 to 100%, 10 divisions displayed
Scale units
dBm, dBmV, dBμV, W, V, dBmV EMF, dBμV EMF, V EMF
Sweep (trace) points
461
Number of markers
6
Marker functions
Normal, frequency counter, noise marker, band power and AM/
FM demod (tune and listen)
and listen)
Log scale
0.01 dB
Linear scale
≤ 1% of signal level
Detectors
Normal, positive peak, sample, negative peak, average
(video, RMS, voltage), quasi-peak (option EMC required)
Number of traces
4
1.
Nominal
RMS detector, trace averaging > 40, 0 dB input attenuation, input terminated 50 Ω, 1 kHz resolution bandwidth, 20 to 30 °C.
05 | Keysight | N9343C Handheld Spectrum Analyzer (HSA) - Data Sheet
Specifications (continued)
Amplitude specifications (continued)
Supplemental information
Level display range (continued)
Trace functions
Level measurement
error
Clear/write, maximum hold, minimum hold, average
1 MHz to 7 GHz
Excluding input VSWR mismatch ± 1.3 dB
7 to 13.6 GHz
± 1.6 dB
–– 20 to 30 °C, 30 to 70%
RH, peak detector,
preamp off, input signal
–50 to 0 dBm, 95%
percentile
–– Swp Time Rule is set to
Accuracy
–– Adds ± 0.3 dB when
Swp Time Rule is set to
Speed
–– Adds ±0.3 dB with
5-minute warm-up
Reference level 2
Setting range
Setting resolution
–100 to +30 dBm
Steps of 1 dB
Log scale
0.01 dB
Linear scale
Same as log (2.236 µV to 7.07 V)
Accuracy
0
RF input VSWR (at tuned frequency)
10 MHz to 3 GHz
1 MHz to 7 GHz
< 1.5:1
7 to 13.6 GHz
< 2:1
Nominal, ≥ 10 dB attenuation
Spurious response
Second harmonic
distortion
Third order intermodulation distortion (third
order intercept)
< –65 dBc, typical < –70 dBc, 50 MHz to 7 GHz
< –80 dBc, typical < –90 dBc, 7 to 13.6 GHz
5-min warm-up
30-min warm-up
50 to 300 MHz
+6.5 dBm, typical +7.5 dBm
+8 dBm, typical +9 dBm
300 MHz to 8 GHz
+7.5 dBm, typical +9.5 dBm
+9 dBm, typical +11 dBm
8 to 13.6 GHz
+8.5 dBm, typical
+10.5 dBm
+10 dBm, typical
+12 dBm
< –59 dBc, typical
< –69 dBc
< –60 dBc, typical
< –70 dBc
1 MHz to 7 GHz
< –93.5 dBm, typical
–108.5 dBm
< –95 dBm, typical
–110 dBm
7 GHz to 13.6 GHz
< –83.5 dBm, typical
–91.5 dBm
< –85 dBm, typical
–93 dBm
Input related spurious
Inherent residual
response
Mixer signal level at
–30 dBm, input attenuation
dB, preamp off, 20 to 30 °C
Two –20 dBm tones at input mixer,
spaced by
100 kHz, input attenuation
0 dB, preamp off, reference level ≥
–30 dBm, 20 to 30 °C
–– –30 dBm signal at input
mixer, span < 2.9 GHz
–– Exception:
–55 dBc (2 x F1 = center
frequency –5,890 MHz,
7 GHz < center frequency
< 10 GHz, with F1 input
frequency)
Input terminated and 0 dB RF
attenuation, preamplifier off
2. Reference level only affects the display not the measurement, so trace data markers do not cause additional errors in measurement results.
06 | Keysight | N9343C Handheld Spectrum Analyzer (HSA) - Data Sheet
Specifications (continued)
Sweep specifications
Supplemental information
Sweep time
Range
2 ms to 1000 s
Span ≥ 100 Hz
600 ns to 200 s
Span = 0 Hz (zero span)
Sweep mode
Continuous, single
Sweep time rule
Accuracy, speed
Trigger source
Free run, video, external, RF burst
Trigger slope
Selectable positive or negative edge
Trigger delay
± 12 ms to ± 12 s
Front panel input/output
Nominal, span = 0 Hz (zero span)
Supplemental information
RF input
Connector and impedance
Type-N female, 50 Ω
Nominal
10 MHz reference/external trigger input
Reference input frequency
10 MHz
Reference input amplitude
0 to +10 dBm
Trigger voltage
5 V TTL level
Nominal
Connector
BNC female, 50 Ω
Nominal
Probe power
Voltage/current
+15 Vdc, ± 7% at 0 to 150 mA (nominal)
–12.6 Vdc, ± 10% at 0 to 150 mA (nominal)
GND
Connectivity
USB host
USB Type-A female, compatible with USB
2.0 full speed
USB device
USB Type-mini AB female, compatible with
USB 2.0 full speed
LAN
RJ-45, 10 Base-T
General specifications
Supplemental information
Display
Resolution
640 pixels x 480 pixels
Size and type
170 mm (6.5 in) TFT color display
Internal memory
System memory
64 MB
For system use. Not user accessible
User memory
64 MB
User accessible. Able to store about 14,000
traces
Languages
On-screen GUI
English, Simplified Chinese, Traditional
Chinese, French, German, Italian, Japanese,
Korean, Russian, Spanish, Portuguese
07 | Keysight | N9343C Handheld Spectrum Analyzer (HSA) - Data Sheet
Specifications (continued)
General specifications (continued)
Supplemental information
Power requirements and calibration
Adaptor voltage
100 to 240 V AC, 50 to 60 Hz
Auto-ranging
15 V DC, 5.3 A, 80 W max
Power consumption
16 W
Typical
Battery operating time (fully charged
battery)
3.5 hours
Tracking generator off, preamplifier on
3 hours
Tracking generator on, preamplifier on
Charging time
3 hours
Life time
300 to 500 charge cycles
Warm-up time
5 minutes
Calibration cycle
One year
Environmental and size
Temperature range
Altitude
–10 to +50 °C
Operating (battery: 0 to 50 °C)
–40 to +70 °C
Storage (battery: –20 to 50 °C)
9,144 meters (30,000 feet)
Operating with battery
3,000 meters (9,840 feet)
Operating with AC to DC adapter
15,240 meters (50,000 feet)
Non-operating
< 95%
Weight
3.2 kg (7 Ibs)
Net (shipping) approximately,
3.6 kg (7.9 Ibs) with battery
Dimensions
318 mm × 207 mm × 69 mm
(12.5 in × 8.15 in × 2.7 in)
Approximately (W × H × D)
Option specifications
Supplemental information
Channel scanner (Option SCN)
Scan modes
Top N, bottom N, and list
Channels displayed
1 to 20
Display orientation
Vertical
Number of channels ≤ 5
Horizontal
Number of channels > 5
Chart
Bar chart, and time chart
Log file
.CSV and .KML
Radio standards
Pre-defined and user-defined. Pre-defined standards include the major wireless
communication standards such as GSM, CDMA, W-CDMA, LTE, WiMAX, etc.
Spectrum monitor (Option SIM)
Display modes
Spectrogram
Spectrum trace
Combination of spectrogram and spectrum
trace in one screen
RF preamplifier (Option P13)
Frequency range
1 MHz to 13.6 GHz
Gain
15 dB
Nominal
Tracking generator (Option TG7)
Frequency range
5 MHz to 7 GHz
Output level
0 to –20 dBm
1 dB steps
VSWR
< 2.0:1
Nominal
Connector and impedance
Type-N female, 50 Ω
08 | Keysight | N9343C Handheld Spectrum Analyzer (HSA) - Data Sheet
Specifications (continued)
Option specifications (continued)
Supplemental information
AM/FM modulation analysis (Option AMA)
Frequency range
10 MHz to 13.6 GHz
Carrier power accuracy
< 7 GHz, ± 1.5 dB
Nominal
7 to 13.6 GHz, ± 1.8 dB
Nominal
–30 to +10 dBm
1 to 500 MHz
–30 to +20 dBm
500 MHz to 13.6 GHz
Carrier power range
Carrier power displayed resolution
0.01 dBm
AM measurement
Modulation rate
20 Hz to 100 kHz
Accuracy
1 Hz
Nominal (modulation rate < 1 kHz)
< 0.1% modulation rate
Nominal (modulation rate > 1 kHz)
Depth
5 to 95%
Accuracy
± 4%
Nominal
FM measurement
Modulation rate
20 Hz to 200 kHz
Accuracy
1 Hz
Nominal (modulation rate < 1 kHz)
< 0.1% modulation rate
Nominal (modulation rate > 1 kHz)
Depth
20 Hz to 400 kHz
Accuracy
± 4%
Nominal
ASK/FSK modulation analysis (Option DMA)
Frequency range
2.5 MHz to 6 GHz
Carrier power accuracy
± 2 dB
Nominal
Carrier power range
–30 to +20 dBm
Nominal
Carrier power displayed resolution
0.01 dBm
ASK measurement
Symbol rate range
100 Hz to 100 kHz
Modulation depth/index
5 to 95%
Accuracy
± 4%
Displayed resolution
0.1%
Nominal
FSK measurement
FSK deviation
100 Hz to 400 kHz
Symbol rate range
100 Hz to 20 kHz
1 ≤ ß*≤ 20
20 to 50 kHz
1≤ß≤8
50 to 100 kHz
1≤ß≤4
Accuracy
± 4%
Nominal
Displayed resolution
0.01 Hz
Time-gated spectrum analysis (Option TMG)
Gated sweep
Span range
Any span
RBW range
> = 1 kHz
VBW is fixed and equal to RBW 3
Gate delay range
12 μs to 10 s
200 ns resolution
Gate length range
84 μs to 10 s
200 ns resolution
* ß is the ratio of frequency deviation to symbol rate (deviation/rate).
09 | Keysight | N9343C Handheld Spectrum Analyzer (HSA) - Data Sheet
Specifications (continued)
Option specifications (continued)
Supplemental information
Time-gated spectrum analysis (Option TMG) (continued)
Gated sweep (continued)
Gate sources
External
RF burst
Periodic timer
–– Sync sources include free, external, and
RF burst
–– Period: 0 to 20.0 s
(It should be greater than gate delay
plus gate length)
–– Offset: –5 to +5 s
RF Burst
Level range
–60 to –20 dBm plus attenuation (nominal)
Bandwidth (–10 dB)
8 MHz (nominal)
Frequency limitations
If the start or center frequency is too close
to zero, LO feedthrough can degrade or
prevent triggering. How close is too close
depends on the bandwidth.
Built-in GPS receiver and GPS antenna (Option GPS)
GPS information tagging
Longitude, latitude, and altitude
GPS antenna
Built-in
Frequency accuracy with GPS on
± 50 ppb
External GPS antenna connector
SMA-F
External GPS antenna, N934xC-GPA, is
offered as an optional accessory
USB peak and average power sensor support (Option PWP)
Power sensor supported
Keysight U2020 X-series USB peak and
average power sensor
Frequency range
50 MHz to 40 GHz
Peak power dynamic range
–30 to +20 dBm
Sensor dependent
USB average power sensor support (Option PWM)
Power sensor supported
Keysight U2000 Series USB power sensor
Frequency range
9 kHz to 24 GHz
Sensor dependent
Dynamic range
–60 to +44 dBm
Sensor dependent
Security erase method
Erase the entire user flash memory by
writing single character “1” over all memory
locations
Non-recoverable
Port control
Disable/enable LAN port or USB port
Security features (Option SEC)
Task planner for test automation (Option TPN)
Task plan execution mode
Auto, manual, and manual if fail
Task plan file
.TPN
Number of tasks
Maximum 20 in a single .TPN file
Measurements supported
Regular spectrum analysis and power suite
(channel power, ACPR, and OBW)
Complementary task plan editor is available
with HSA PC software
Visit www.keysight.com/find/taskplanner for more information.
3. For efficiency and convenience, RBW is restricted to be equal to or greater than 1 kHz and VBW is restricted to be equal to RBW.
10 | Keysight | N9343C Handheld Spectrum Analyzer (HSA) - Data Sheet
From Hewlett-Packard through Agilent to Keysight
For more than 75 years, we‘ve been helping you unlock measurement insights.
Our unique combination of hardware, software and people can help you reach
your next breakthrough. Unlocking measurement insights since 1939.
1939
THE FUTURE
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s commitment to superior product quality and lower total cost
of ownership. The only test and measurement company with three-year
warranty standard on all instruments, worldwide.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to five years of protection and no budgetary surprises to ensure your
instruments are operating to specification so you can rely on accurate
measurements.
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-04-23-15)
www.keysight.com/find/n9343c
www.keysight.com/find/hsa-videos
www.keysight.com/find/hsa
This information is subject to change without notice.
© Keysight Technologies, 2014 - 2015
Published in USA, October 22, 2015
5990-7499EN
www.keysight.com