Delta MDS-200ADB24AA Medical ac-dc enclosed power supply Datasheet

Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
MDS-200ADB24 AA
Highlights & Features
–Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
–Compliant with IEC 60601-1-2 4th Ed. Requirements
–2.5" x 4.5" x 1.8" Package
–Power Good Signal
–up to 800K Hours MTBF
–up to 120W convection, 200W forced air
Safety Standards
CB Certified for worldwide use
Model Number:
MDS-200ADB24 AA
Unit Weight:
345g(0.762lb)
Dimensions (W x L x H): 64.7x113.9x44.7mm
(2.5x4.5x1.8 inch)
General Description
The MDS-200ADB24 AA enclosed power supply comes with universal AC input range from 90Vac to 264Vac. Other
features include low leakage, Type BF Patient Access Leakage Currents, and electric shock protection compliance with 2
x MOPP requirements. The MDS-200ADB24 AA is certified for EMC standards according to EN 55011for industrial,
scientific and medical (ISM) radio-frequency equipment; and, EN 55022 for Industrial Technology Equipment (ITE) radiofrequency equipment. In addition, only recognized Japanese capacitors are used to ensure long product life.
The MDS-200ADB24 AA comes with both medical and ITE safety approvals, including UL/CE/CCC (5000meters), and CB
certification. Designs are compliant with RoHS Directive 2011/65/EU for environmental protection.
Model Information
Model Number
Input Voltage
Output Voltage
Convection Current Output Forced Air Current Output
MDS-200ADB24AA
90-264Vac
24Vdc
0-5A
0-8.33A*
*With 5.5CFM forced air
Model Numbering
MDS
Delta Medical power
Supply
1
200
Max wattage in the
product Series.
Maybe lower at some
voltage.
200→200W
A
D
B
24
Family Code
Product Type
Input Type Code
Output Voltage
A~ Z
D: Enclosed
B: 3pin Class I
24 for 24V
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
AA
Revision code
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
100-240Vac
Input Voltage Range
90-264Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current(max)
2.5A
Input Surge Voltage (max)
300Vac for 100ms
Full load Efficiency (typ.)
92% @ 115Vac/60Hz
93% @ 230Vac/50Hz, Reference Fig.1
Standby Power (max)
0.5W @ 115Vac/60Hz, 230Vac/50Hz
Inrush Current (max)
60A@230Vac, cold start
Input-PE(protective earth) leakage current(max)
0.1mA @ NC, 0.3mA @ SFC1)
Output-PE(protective earth) leakage current for Type BF application (max)
0.1mA @ NC, 0.5mA @ SFC 1)
Power Factor (min)
0.95 @ 115V/50Hz, 230V/50Hz, full load
1) NC: normal condition, SFC: single fault condition
Leakage Current
Input-PE Leakage Current
100Vac/60Hz(Typ)
Normal Condition 19.5uA
Single Fault Condition 44.8uA
264Vac/60Hz(Typ) Delta Limit
IEC60601-1 Limit
45.0uA
100uA max
5000uA max
153.0uA
300uA max
10000uA max
81.5uA
100uA max
100uA max
129.4uA
500uA max
500uA max
Output-PE Leakage Current for Type BF application
Normal Condition 38.3uA
Single Fault Condition 43.4uA
100
90
Efficiency (%)
115Vac
80
230Vac
70
60
50
25%
50%
75%
Output Load
Fig.1 Efficiency versus output load
2
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
100%
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
Output Ratings / Characteristics
Nominal Output Voltage (Vrated)
24V
Output Voltage Tolerance
±3%
Output Power
200W 5.5CFM air, up to 120W convection air
Line Regulation (max)
±0.5%
Load Regulation (max)
±1%
Ripple& Noise(typ.)
1%pk-pkVrated@ Full load, Reference Fig. 2
Start-up Time(max)
2000ms@115Vac
Hold-up Time(min)
12ms@115Vac, tested with 120W load
Dynamic Response(Overshoot & Undershoot O/P Voltage)
±5% @50-100% load
Capacitive load (max)
1000uF
Rise time (max)
100ms
(a) 115V (measured value=153mV)
(b) 230V(measured value=158mV)
Fig.2 Ripple & Noise example, 20MHz BW
Ripple & Noise measurement circuit
PSU
Load
5cm
0.47uF Ceramic
10uF Electrolytic
Test Fixture
3
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
Scope
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
Mechanical
Case Cover
Enclosed cover(SPCC)
Dimensions(W x L x H typ.)
64.7x113.9x44.7mm (2.5x4.5x1.8inch)
Unit Weight
345g(0.762lb)
Indicator
NA
Cooling System
NA
Environment
Surrounding Air Temperature
Operating
Storage
Absolute Max
-10°C to+70°C, supported power linearly de-rate from
50°C to 50% rated up to 70°C
Note: see power de-rating curve
-40°C to+85°C
Operating Humidity
5-95% RH (Non-Condensing)
Operating Altitude
5,000 meters (16402 feet)
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration (Operating)
5-500Hz, 2Grms, 15 minute for each three axis
Power De-rating curve
4
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
Protections
Overvoltage (max)
125% of rated voltage, Latch Mode
Overload / Overcurrent (max)
Main output 160% of rated current
Hiccup Mode(Non-Latching, Auto-Recovery)
Over Temperature
Latch Mode
Short Circuit
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Reliability
MTBF(Minimum) at 115Vac, 120W, 35°C, Convection Air Flow
800 kHrs based on Telecordia SR-332
Operating life at 115Vac, 120W, ambient 25 °C, Convection Air Flow
26,280Hrs
Safety Standards / Directives
nd
rd
rd
Medical Safety
IEC60601-12 , 3 and 3 +A1 edition CB report
TUV EN60601-1:2006
UL60601-1+CAN/CSA 60601-1: (Ed.3.2005)
ITE Safety
IEC60950-1 CB report
TUV60950-1
UL60950-1+CAN/CSA60950-1
GB4943.1-2011, GB9254-2008, GB17625.1-2003
CE
MDD Directive 93/42/EEC
Environmental
RoHS Directive 2011/65/EU Compliant
Galvanic Isolation
1)
5
Input to/Output (2XMOPP) 4000 Vac
1)
Input to/Ground(1XMOPP) 1500Vac
Output to/Ground(1XMOPP) 1500Vac ( Type BF application rated)
PSU can support PoE applications with Primary to FG 2500Vac test.
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements)
EMC / Emissions
Harmonic Current Emissions
EN55011/EN55022,FCC Title 47:Class B
IEC61000-3-2
Meet Class D limit
Immunity to
Voltage Flicker
IEC61000-3-3
Electrostatic Discharge
IEC61000-4-2
Level 4 Criteria A
Air Discharge: 15kV
Contact Discharge: 8kV
Radiated Field
IEC61000-4-3
Electrical Fast Transient / Burst
IEC61000-4-4
Criteria A
80MHz-2700MHz, 10V/m AM modulation
385MHz-5785MHz, 28V/m Pulse mode and other
modulation
1)
Level 3 Criteria A :2kV
Surge
IEC61000-4-5
Level 3 Criteria A
3)
Common Mode : 2kV
4)
Differential Mode : 1kV
Conducted
IEC61000-4-6
Level 2 Criteria A
150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and
1)
1)
1)
1)
Amateur radio bands
Power Frequency Magnetic Fields
Voltage Dips
IEC61000-4-8
IEC61000-4-11
1)
Criteria A
Magnetic field strength 30A/m
1)
Criteria A
0% UT, 0.5 cycle (10ms)
o
o
o
o
o
o
o
o
o
, 0 /45 /90 /135 /180 /225 /270 /315 /360
2)
Criteria B
o
0% UT, 1 cycle (20ms), 0
2)
Criteria B
o
70% UT, 25 cycle (500ms), 0
2)
Criteria B
o
0% UT, 250 cycle (5000ms), 0
1) Criteria A: Normal performance within the specification limits
2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test.
3) Asymmetrical: Common mode (Line to earth)
4) Symmetrical: Differential mode (Line to line)
6
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
Dimensions
W x L x H: 64.7 x 113.9 x 44.7 mm
Mechanical drawing (3Pin input type)
Notes
―Dimensions are in mm(inches)
―There are two locations where assembled power supply is connected to the customer’s product
a. Bottom mounting, use (4X) M3 screws to affix assembled power supply to product’s enclosure. Extruded hole with thread must
be withstand 9Kgf.cm (7.81lb-in) min. Maximum allowed screw penetration is 3.5mm (0.138 inch).
b. Side mounting, use (2X) M3 screws to affix one side of assembled power supply to the product’s enclosure. Extruded hole with
thread must be withstand 9Kgf.cm (7.81lb-in) min. Maximum allowed screw penetration is 3.5mm (0.138 inch).
7
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
Input Connector CN1
Pin 1
Neutral
Pin 2
Line
CN1 mates with Molex housing
26033031 and Molex series 6838 crimp
terminals.
Input Line can also be connected to pin
1, and Input Neutral can be connected
to pin 2
Pin 1
Pin 2
CN102 mates with Molex housing
0874390200 and Molex 874210000 crimp terminals.
CN3: KANG YANG:JP-13T mate with KST :FDFNYD1-187 or other
applicable connectors
Output Connector CN101
Pin1
DC RTN
Pin2
DC RTN
Pin3
Vo
Pin4
Vo
CN101 mates with JST housing VHR-4N
and JST terminal SVH-41T-P1.1.
8
Signal Connector CN102
Power Good -(DC RTN)
Power Good +
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
Functions
Start-up Time
The time required for the output voltage to reach 90% of its set value, after the input voltage is applied.
Rise Time
The time required for the output voltage to change from 10% to 90% of its set value.
Hold-up Time
Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time
required for the output to reach 90% of its set value, after the input voltage is removed.
■Graph illustrating the Start-up Time, Rise Time, and Hold-up Time
Dynamic Response
The power supply output voltage will remain within ±5% of its
steady state value, when subjected to a dynamic load 50 to 100%
of its rated current.
■ 50 to 100% Load
Overvoltage Protection
Inrush Current
Inrush current is the peak, instantaneous, input current measured
and, occurs when the input voltage is first applied. For AC input
voltages, the maximum peak value of inrush current will occur
during the first half cycle of the applied AC voltage. This peak
value decreases exponentially during subsequent cycles of AC
voltage.
9
The power supply’s overvoltage circuit will be activated when its
internal feedback circuit fails. The output voltage shall not exceed
its specifications defined on Page 6 under “Protections”. Power
supply will latch off, and require removal/re-application of input
AC voltage in order to restart.
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
Short Circuit Protection
The power supply’s output OLP/OCP function also provides
protection against short circuits. When a short circuit is applied,
the output current will operate in “Hiccup mode”, as shown in the
illustration in the OLP/OCP section on this page. The power
supply will return to normal operation after the short circuit is
removed.
Overload & Overcurrent Protections
The power supply’s Overload (OLP) and Overcurrent (OCP)
Protections will be activated before output current under 160% of
I O (Max load). Upon such occurrence, V O will start to drop. Once
the power supply has reached its maximum power limit, the
protection will be activated and the power supply will go into
“Hiccup mode” (Auto-Recovery). The power supply will recover
once the fault condition causing the OLP and OCP is removed
and I O is back within the specified limit.
Power Good
Power Good+ pin is an open collector transistor (40V/600mA
rating). A resistor (suggested value 20Kohm, 1/8W) can be added
between output pin (or, other available pull-up voltage that is no
greater than 30V) and the Power Good+ pin (refer to figure
below). Value of pull-up resistor may have to be adjusted,
depending on voltage used, and other end-use conditions of the
Power Good+ pin connection to the product. When AC input is
on, Power Good+ pin will be high. When AC input is off, Power
Good+ pin will be low. There will be a minimum of 5 milliseconds
(at 120W load) between the time the power good goes to low
level, and the time when the output reaches 90% of its rated
value.
Additionally, if the Iout is <160% but >100% for a prolong period
of time (depending on the load), the Over Temperature
Protection (OTP) will be activated due to high temperature on
critical components. Then, the power supply will be latched off,
and require recycling of input voltage to restart it.
Over Temperature Protection
As mentioned above, the power supply also has Over
Temperature Protection (OTP). This is activated when the
overload condition persists for an extended duration and the
output current is below the overload trigger point but >100% load.
In the event of a higher operating temperature condition at 100%
load, the power supply will run into OTP when the surrounding air
temperature is higher than the operating temperature. When
activated, the output voltage will go into latch mode until the input
voltage is removed; then, reapplied, and the surrounding air
temperature drops to its normal operating temperature.
10
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 01)
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density MDS-200ADB24 AA
Certificate
All Delta Medical Power products conform to the European directive 2011/65/EU.
RoHS is the abbreviation for “Restriction of the use of certain hazardous
substances
Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN
ISO 13485:2012 for the design and manufacture of switching power supply and
adaptor for medical device.
In addition to a UL Total Certification Program (TCP) approved client laboratory for
IEC60950 and IEC60065. Delta also has participated UL Client Test Data Program
(CDTP) for IEC 60601
11
All parameters are specified at 25°C ambient unless otherwise noted.