MPS-30 30W Single Output for Medical Type series Features : Universal AC input / Full range Low leakage current 0.3mA Protections: Short circuit / Overload / Over voltage / Over temperature Fixed switching frequency at 100KHz Low cost High reliability 3 years warranty SPECIFICATION MODEL OUTPUT INPUT MPS-30-5 MPS-30-12 MPS-30-15 MPS-30-24 MPS-30-27 DC VOLTAGE 5V 12V 15V 24V 27V 48V RATED CURRENT 5A 2.5A 2A 1.2A 1.1A 0.6A CURRENT RANGE 0 ~ 5A 0 ~ 2.5A 0 ~ 2A 0 ~ 1.2A 0 ~ 1.1A 0 ~ 0.6A RATED POWER 25W 30W 30W 28.8W 29.7W 28.8W 120mVp-p 150mVp-p 240mVp-p 240mVp-p 240mVp-p RIPPLE & NOISE (max.) Note.2 80mVp-p VOLTAGE TOLERANCE Note.3 3.0% 3.0% 3.0% 3.0% 3.0% 2.0% LINE REGULATION 1.0% 1.0% 0.5% 0.5% 0.5% 0.5% LOAD REGULATION 2.0% 1.0% 1.0% 1.0% 1.0% 1.0% SETUP, RISE TIME 500ms, 30ms/230VAC HOLD UP TIME (Typ.) 70ms/230VAC VOLTAGE RANGE 88 ~ 264VAC auto switch FREQUENCY RANGE 47 ~ 63Hz EFFICIENCY (Typ.) 72% AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 0.8A/115VAC 12ms/115VAC at full load 120 ~ 370VDC 75% 76% 77% OVERLOAD Above 105% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed WORKING TEMP. WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY TEMP. COEFFICIENT VIBRATION SAFETY STANDARDS SAFETY & WITHSTAND VOLTAGE ISOLATION RESISTANCE EMC (Note 4) EMC EMISSION EMC IMMUNITY MTBF DIMENSION PACKING 13.2 ~ 16.2V 5.5 ~ 6.75V 78% 78% 29.7 ~ 36.45V 52.8 ~ 64.8V 0.5A/230VAC COLD START 30A/230VAC <0.3mA / 264VAC OVER TEMPERATURE NOTE 500ms, 30ms/115VAC at full load LEAKAGE CURRENT PROTECTION OVER VOLTAGE OTHERS MPS-30-48 16.5 ~ 20.25V 26.4 ~ 32.4V Protection type : Hiccup mode, recovers automatically after fault condition is removed Tj 135 typically (U1) detect on main control IC Protection type : Shut down o/p voltage, re-power on to recover -10 ~ +60 (Refer to "Derating Curve") 20 ~ 90% RH non-condensing -20 ~ +85 , 10 ~ 95% RH 0.03%/ (0 ~ 50 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:SHORT I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25 / 70% RH Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3 Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, medical level, criteria A 547Khrs min. MIL-HDBK-217F (25 101.6*65.8*23.5mm (L*W*H) 0.16Kg; 90pcs/15.8Kg/1.02CUFT ) 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) File Name:MPS-30-SPEC 2011-12-27 MPS-30 30W Single Output for Medical Type Mechanical Specification series Unit:mm 101.6 66.1 95.2 18 3.2 3.2 51.4 62.6 65.8 CN2 CN1 4 3 2 1 M3 L= 6m 5 4 3 2 1 m FUSE*2 T2.5A 250V 3.2 2 max. 23.5 max. 4- AC Input Connector (CN1) : Molex 41791-5 or equivalent Pin No. 1 2,4 3 5 Assignment FG No Pin AC/N AC/L Mating Housing Terminal Molex 2139 or equivalent Molex 2478 or equivalent DC Output Connector (CN2) : Molex 41791-4 or equivalent Pin No. 1,2 3,4 Assignment -V +V Mating Housing Molex 2139 or equivalent Terminal Molex 2478 or equivalent Block Diagram fosc : 100KHz RECTIFIERS & FILTER EMI FILTER I/P RECTIFIERS & FILTER POWER SWITCHING PWM CONTROL +V -V DETECTION CIRCUIT FG CROWBAR OVP Derating Curve OUTPUT VOLTAGE (V) 100 80 LOAD (%) 60 40 20 32 350 28 300 24 250 20 200 16 150 12 100 50 8 -10 0 10 20 30 40 50 AMBIENT TEMPERATURE ( ) 60 (HORIZONTAL) OUTPUT RIPPLE (mVp-p) Static Characteristics 60 80 100 120 140 160 180 200 220 240 264 INPUT VOLTAGE (V) 60Hz File Name:MPS-30-SPEC 2011-12-27