Minmax MIHW3022 Dc/dc converter 6w, reinforced insulation, medical safety Datasheet

®
MIHW3000 SERIES
DC/DC CONVERTER 6W, Reinforced Insulation, Medical Safety
FEATURES
►Reinforced Insulation rated for 300VAC Working Voltage
►I/O-isolation Voltage 4000VACrms
►Industrial & Medical Safety Approval
►Wide 2:1 Input Voltage Range
►Fully regulated Output Voltage
►Low Leakage Current
►Operating Temp. Range –40°C to +75 °C
►Input Filter meets EN55022, class A
►Overload Protection
►3 Years Product Warranty
PRODUCT OVERVIEW
The MINMAX MIHW3000 series is a range of high performance DC/DC converter modules with a reinforced insulation system .The I/O- isolation voltage is
specified for 4000VACrms.The product comes in a small DIP-24 package. All 12 models features wide 2:1 input voltage range and fully regulated output
voltage.
The MIHW3000 DC/DC converters offer an economical solution for demanding applications in industrial and medical instrumentation requesting a certified
high supplementary or reinforced insulation system to comply with relative industrial or medical safety standards.
Model Selection Guide
Model
Number
MIHW3022
MIHW3023
MIHW3026
MIHW3027
MIHW3032
MIHW3033
MIHW3036
MIHW3037
MIHW3042
MIHW3043
MIHW3046
MIHW3047
Input
Voltage
(Range)
Output
Voltage
VDC
VDC
5
12
±12
±15
5
12
±12
±15
5
12
±12
±15
12
(9 ~ 18)
24
(18 ~ 36)
48
(36 ~ 75)
Output Current
Input Current
Max.
Min.
@Max. Load
@No Load
Reflected
Ripple
Current
mA
1000
500
±250
±200
1000
500
±250
±200
1000
500
±250
±200
mA
200
100
±50
±40
200
100
±50
±40
200
100
±50
±40
mA(typ.)
570
641
641
641
278
313
313
313
139
156
156
156
mA(typ.)
mA (typ.)
30
60
20
30
10
15
Max. capacitive
Load
uF
1000
470
220#
220#
1000
470
220#
220#
1000
470
220#
220#
Efficiency
(typ.)
@Max. Load
%
75
78
78
78
77
80
80
80
77
80
80
80
# For each output
Input Specifications
Parameter
Model
12V Input Models
24V Input Models
48V Input Models
12V Input Models
24V Input Models
48V Input Models
12V Input Models
24V Input Models
48V Input Models
Input Surge Voltage (1 sec. max.)
Start-Up Voltage
Under Voltage Shutdown
Short Circuit Input Power
Input Filter
Internal Power Dissipation
Conducted EMI
All Models
E-mail:[email protected]
2011/05/18 REV:8
Tel:886-6-2923150
Page 1 of 4
Min.
-0.7
-0.7
-0.7
7
13
30
---------
Typ.
------8
15
33
---------
Max.
25
50
100
9
18
36
8.5
16
34
3000
Unit
VDC
mW
Pi Filter
----2500
mW
Compliance to EN55022,class A and FCC part 15,class A
®
MIHW3000 SERIES
DC/DC CONVERTER 6W, Reinforced Insulation, Medical Safety
Output Specifications
Parameter
Conditions
Min.
Typ.
Max.
Unit
---
±0.5
±1.0
%
Dual Output, Balanced Loads
---
±0.5
±2.0
%
Vin=Min. to Max.
---
±0.3
±0.5
%
Io=25% to 100%
---
±0.5
±1.0
%
5V Output Models
---
75
100
mV P-P
Other Output Models
---
100
150
mV P-P
Over Line, Load & Temp.
---
---
180
mV P-P
---
---
25
mV rms
---
300
500
uS
---
±3
±6
%
---
±0.02
±0.05
%/℃
120
150
---
%
Output Voltage Accuracy
Output Voltage Balance
Line Regulation
Load Regulation
Ripple & Noise (20MHz)
Ripple & Noise (20MHz)
Ripple & Noise (20MHz)
Transient Recovery Time
25% Load Step Change
Transient Response Deviation
Temperature Coefficient
Over Load Protection
Foldback
Short Circuit Protection
Continuous
Isolation, Safety Standards
Parameter
Conditions
Min.
Typ.
Max.
Unit
60 Seconds
4000
---
---
VACrms
Flash tested for 1 Second
6000
---
---
VPK
240VAC, 60Hz
---
---
2
uA
I/O Isolation Resistance
500 VDC
10
---
---
GΩ
I/O Isolation Capacitance
100KHz, 1V
13
pF
I/O Isolation Voltage (rated)
I/O Isolation Test Voltage
Leakage Current
Safety Standards
Approvals
--7
cUL/UL60950-1, CSA C22.2 No. 60950-1-03
UL60601-1,CSA C22.2 No.601-1,
IEC/EN 60950-1, IEC/EN 60601-1
IEC60950-1 CB report, cUL/UL 60950-1 certificate
UL60601-1 UL certificate
General Specifications
Parameter
Conditions
Min.
Typ.
Max.
Unit
---
150
---
KHz
700,000
-----
-----
Hours
Switching Frequency
MTBF(calculated)
MIL-HDBK-217F@25℃, Ground Benign
Input Fuse
12V Input Models
24V Input Models
48V Input Models
1200mA Slow-Blow Type
600mA Slow-Blow Type
300mA Slow-Blow Type
Environmental Specifications
Parameter
Conditions
Min.
Max.
Unit
Ambient
-40
+75
℃
Case Temperature
---
+95
℃
Storage Temperature Range
-50
+125
℃
Humidity (non condensing)
---
95
% rel. H
260
℃
Operating Temperature Range (with Derating)
Cooling
Free-Air convection
Lead Temperature (1.5mm from case for 10Sec.)
---
E-mail:[email protected]
2011/05/18 REV:8
Tel:886-6-2923150
Page 2 of 4
®
MIHW3000 SERIES
DC/DC CONVERTER 6W, Reinforced Insulation, Medical Safety
Output Power (%)
Power Derating Curve
Notes
1
2
3
4
5
6
7
Specifications typical at Ta=+25℃, resistive load, nominal input voltage and rated output current unless otherwise noted.
Transient recovery time is measured to within 1% error band for a step change in output load of 75% to 100%.
Ripple & Noise measurement bandwidth is 0-20 MHz.
These power converters require a minimum output loading to maintain specified regulation, operation under no-load conditions will not damage these
modules; however, they may not meet all specifications listed.
All DC/DC converters should be externally fused at the front end for protection.
Other input and output voltage may be available, please contact factory.
Specifications subject to change without notice.
Package Specifications
Mechanical Dimensions
Pin Connections
3.8 12.0 [0.47]
[0.15]
0.5 [0.02]
Pin
0.60 [0.024]
23
15
1
+Vin
+Vin
11
No Pin
Common
12
-Vout
On Pin
13
+Vout
-Vout
15
No Pin
+Vout
23
-Vin
--Vin
24
-Vin
-Vin
12
Bottom View
24
13
5.08
[0.20]
31.8 [1.25]
2.0
[0.08]
2.54
[0.10]
►All dimensions in mm
20.32 [0.80]
2.54
[0.10]
Physical Characteristics
Case Size
:
31.8x20.3x12.0mm (1.25x0.8x0.47 Inches)
Case Material
:
Non-Conductive Black Plastic (flammability to UL 94V-0 rated)
Weight
:
18g
E-mail:[email protected]
2011/05/18 REV:8
Tel:886-6-2923150
Page 3 of 4
Dual Output
20.3 [0.80]
11
15.22 [0.60]
1
Single Output
►Tolerance: X.X±0.25
(inches)
(X.XX±0.01)
X.XX±0.13 ( X.XXX±0.005)
►Pin
pitch tolerance: ±0.25 (0.01)
►Pin
tolerance: ±0.05 (0.002)
®
MIHW3000 SERIES
DC/DC CONVERTER 6W, Reinforced Insulation, Medical Safety
Test Configurations
Input Reflected-Ripple Current Test Setup
Input reflected-ripple current is measured with a inductor Lin (4.7uH) and Cin (220uF, ESR < 1.0Ω at 100 KHz) to simulate source impedance. Capacitor Cin, offsets
possible battery impedance. Current ripple is measured at the input terminals of the module, measurement bandwidth is 0-500 KHz.
To Oscilloscope
+
+
Battery
+Vin
Lin
DC / DC
Converter
Current
Probe
Cin
+Out
-Vin
Load
-Out
Peak-to-Peak Output Noise Measurement Test
Use a Cout 0.47uF ceramic capacitor. Scope measurement should be made by using a BNC socket, measurement bandwidth is 0-20 MHz. Position the load
between 50 mm and 75 mm from the DC/DC Converter.
+Vin
+Out
Copper Strip
Single Output
DC / DC
Converter
-Vin
Cout
-Out
Resistive
Load
Scope
Copper Strip
+Vin
+Out
Copper Strip
Dual Output
DC / DC
Converter
Com.
Copper Strip
-Vin
-Out
Scope
Cout
Cout
Resistive
Load
Scope
Copper Strip
Design & Feature Considerations
Input Source Impedance
The power module should be connected to a low ac-impedance input source. Highly inductive source impedances can affect the stability of the power module. In
applications where power is supplied over long lines and output loading is high, it may be necessary to use a capacitor on the input to insure startup. By using a
good quality low Equivalent Series Resistance (ESR < 1.0Ω at 100 kHz) capacitor of a 10uF for the 12V input devices and a 4.7uF for the 24V input devices and a
2.2uF for the 48V devices, capacitor mounted close to the power module helps ensure stability of the unit.
+
DC Power
Source
+Vin
+
+Out
DC / DC
Converter
Load
Cin
-
-Vin
-Out
Output Ripple Reduction
A good quality low ESR capacitor placed as close as practicable across the load will give the best ripple and noise performance. To reduce output ripple, it is
recommended to use 3.3uF capacitors at the output.
+
+Vin
DC Power
Source
-
+Out
Single Output
DC / DC
Converter
-Vin
+
Cout
+Vin
DC Power
Source
Load
-Out
-
+Out
Dual Output
DC / DC
Com.
Converter
-Vin
-Out
Cout
Load
Load
Cout
Maximum Capacitive Load
The MIHW3000 series has limitation of maximum connected capacitance on the output. The power module may operate in current limiting mode during start-up,
affecting the ramp-up and the startup time. Connect capacitors at the point of load for best performance. The maximum capacitance can be found in the data
sheet.
Thermal Considerations
Many conditions affect the thermal performance of the power module, such as orientation, airflow over the module and board spacing. To avoid exceeding the
maximum temperature rating of the components inside the power module, the case temperature must be kept below 95℃. The derating curves are determined
from measurements obtained in a test setup.
Position of air velocity
probe and thermocouple
50mm / 2in
15mm / 0.6in
Air Flow
DUT
Design & Feature Considerations
Conducted and radiated emissions < A
with external coupling capacitor Cio=1 nF < B
18, Sin Sin Road, An-Ping Industrial District, Tainan 702, Taiwan
Tel: 886-6-2923150 Fax: 886-6-2923149 E-mail: [email protected]
Minmax Technology Co., Ltd.
2011/05/18 REV:8