XP Power JHM1012S12 International medical approval Datasheet

DCDC
10 Watts
JHM10 Series
xppower.com
•
International Medical Approvals
•
4000 VAC Reinforced Insulation
•
Meets IEC606011, 3rd Edition
•
2 µA Patient Leakage Current
•
DIP24 Package
•
EN55011 Level A With No External Components
•
3 Year Warranty
Specification
Input
Input Voltage Range
Input Current
Inrush Current
Input Filter
Patient Leakage
Current
Undervoltage Lockout
Input Surge
General
• 5 V (4.5-9 VDC)
12 V (9-18 VDC)
24 V (18-36 VDC)
• See table
• 20 A max at 36 V
• Pi network
• 2 µA max
• 5 V models, on at >4.4 V, off <4.2 V
12 V models, on at >8.8 V, off <8.3 V
24 V models, on at >17.5 V, off <17.0 V
• 5 V models 15 V for 3 s
12 V models 25 V for 3 s
24 V models 50 V for 3 s
Output
Output Voltage
Output Voltage Trim
Minimum Load
Initial Set Accuracy
Start Up Delay
Start Up Rise Time
Line Regulation
Load Regulation
•
•
•
•
•
•
•
•
Cross Regulation
•
Transient Response
•
Ripple & Noise
Overload Protection
Overvoltage Protection
Temperature
Coefficient
Short Circuit Protection
•
•
•
•
See table
±10%
No minimum load required
±1% max on V1, ±2% max on V2
5 ms typical
2 ms typical
±0.3% max
±2% max 0% to 10% load,
±1% max 10% to 100% load
±4% max on dual with one output set to
50% load and the other varied from
10% to 100% load (D05 is 20% to 100%)
4% max deviation, recovery to within 1%
in <500 µs for a 50% load change at
0.25 A/µs rate
1% pk-pk max at 20 MHz bandwidth
120% - 200%, trip and restart
115% - 140%
±0.03/°C max
• Trip and Restart (hiccup mode),
auto recovery
Efficiency
Isolation
Input to Output
Capacitance
Switching Frequency
Power Density
MTBF
• See tables
• 4000 VAC for 1 min. double/reinforced with
a working voltage of 250 VAC. Meets 2 x
MOPP per 3rd edition of IEC60601-1
5000 VAC for 10 ms in accordance with
IEC60664-1
• 20 pF max
• 80 kHz to 1.2 MHz variable
• 20.0 W/in3
• >1 MHrs typical to MIL-STD-217F
at 25 °C, GB
Environmental
Operating Temperature
Case Temperature
Storage Temperature
Operating Humidity
Cooling
•
•
•
•
•
-40 °C to +80 °C, see derating curve
+100 °C max
-40 °C to +100 °C
5-90%, non-condensing
Natural convection
EMC & Safety
Emissions
Immunity
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Magnetic Field
Safety Approvals
• EN55011 & EN55022 level A conducted &
radiated with no external components
• IEC60601-1-2, EN61204-3
• EN61000-4-2, level 2, Perf Criteria A
• EN61000-4-3, 10 V/m Perf Criteria A
• EN61000-4-4, level 2 Perf Criteria A
• EN61000-4-5, level 1 Perf Criteria A
• EN61000-4-6, 10 Vm, Perf Criteria A
• EN61000-4-8, 3 A/m Perf Criteria A
• ANSI/AMMI ES60601-1 3rd Edition
CSA-22.2 No.60601-1:2008
IEC60601-1 3rd Edition
Input Voltage
Output Voltage
4.5-9 V
9-18 V
18-36 V
Input Current
No Load(1)
Full Load(2)
Output Current
Maximum
Capacitive Load
Efficiency(3)
Model Number
5.0 V
2000 mA
100 mA
2690 mA
2200 µF
83.5%
JHM1005S05
12.0 V
833 mA
115 mA
2640 mA
1000 µF
85.0%
JHM1005S12
15.0 V
666 mA
115 mA
2640 mA
680 µF
85.0%
JHM1005S15
±5.0 V
±1000 mA
130 mA
2760 mA
±1000 µF
81.5%
JHM1005D05
±12.0 V
±420 mA
115 mA
2695 mA
±470 µF
84.0%
JHM1005D12
±15.0 V
±333 mA
115 mA
2670 mA
±470 µF
84.0%
JHM1005D15
5.0 V
2000 mA
50 mA
1310 mA
2200 µF
86.0%
JHM1012S05
12.0 V
833 mA
50 mA
1280 mA
1000 µF
88.0%
JHM1012S12
15.0 V
666 mA
50 mA
1265 mA
680 µF
89.0%
JHM1012S15
±5.0 V
±1000 mA
50 mA
1345 mA
±1000 µF
84.0%
JHM1012D05
±12.0 V
±420 mA
50 mA
1290 mA
±470 µF
88.0%
JHM1012D12
±15.0 V
±333 mA
50 mA
1280 mA
±470 µF
88.0%
JHM1012D15
5.0 V
2000 mA
25 mA
645 mA
2200 µF
87.0%
JHM1024S05
12.0 V
833 mA
20 mA
630 mA
1000 µF
89.0%
JHM1024S12
15.0 V
666 mA
20 mA
630 mA
680 µF
89.0%
JHM1024S15
±5.0 V
±1000 mA
20 mA
660 mA
±1000 µF
85.0%
JHM1024D05
±12.0 V
±420 mA
25 mA
640 mA
±470 µF
88.0%
JHM1024D12
±15.0 V
±333 mA
25 mA
635 mA
±470 µF
88.0%
JHM1024D15
Notes
1. Input current measured at nominal input voltage.
2. Input current measured at lowest input voltage.
3. Typical values.
Mechanical Details
1.25
(31.75)
0.10
(2.54)
0.20
(5.08)
14
16
0.80
(20.32)
22
23
0.60
(15.24)
Bottom View
11 10
9
3
2
0.90
(22.86)
0.18
(4.57)
Dual
-Vin
-Vin
Common
Trim
-Vout
+Vout
Common
+Vin
+Vin
Notes
Notes:
1. Pin diameter: 0.02 ±0.002 (0.5 ±0.05)
2. Pin pitch tolerance: ±0.014 (±0.35)
3. Case Tolerance: ±0.02 (±0.5)
0.40 (10.40)
Pin Connections
Single
-Vin
-Vin
No Pin
Trim
No Pin
+Vout
-Vout
+Vin
+Vin
Pin
2
3
9
10
11
14
16
22
23
1.
2.
3.
4.
5.
0.12 (3.05) Min.
All dimensions are in inches (mm)
Weight: 0.04 lbs (20 g) approx.
Pin diameter: 0.02 ±0.002 (0.5 ±0.05)
Pin pitch tolerance: ±0.01 (±0.25)
Case tolerance: ±0.02 (±0.5)
Application Notes
Derating Curve
External Output Trim
10
Trim
1
8
2
Trim Up
3
Load (W)
+ Vout
Trim Down
6
-Vout
Trim
4
2
0
-40 -30
-20
-10
0
10
20
30
40
50
Ambient Temperature (°C)
1
JHM1012/24 S05, S12,S15, D12, D15
2
JHM1012/24 D05, JHM1005S05, S12, S15, D12, D15
3
JHM1005D05
60
70
80
90
For 5 V output:
Trim +10%, R = 3.4 k typical
Trim -10%, R = 1 k typical
For ±5 V output:
Trim +10%, R = 12.0 k typical
Trim -10%, R = 8.0 k typical
For 12 V output:
Trim +10%, R = 5.9 k typical
Trim -10%, R = 11.3 k typical
For ±12 V output:
Trim +10%, R = 12.8 k typical
Trim -10%, R = 9.5 k typical
For 15 V output:
Trim +10%, R = 8.3 k typical
Trim -10%, R = 10 k typical
For ±15 V output:
Trim +10%, R = 18 k typical
Trim -10%, R = 14.8 k typical
13-Jan-14
DCDC
JHM10
Models and Ratings
Similar pages