Delta MDS-300ADB18AA Medical ac-dc enclosed power supply Datasheet

Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
MDS-300ADB18 AA
Highlights & Features
– Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
– Compliant with IEC 60601-1-2 4th Ed. Requirements
– Up to 190 Watt Convection in 3.5" x 5.5" x 1.75"
Package
– Up to 800 K Hrs MTBF
– Suited for Type BF Medical Products.
– 2 x MOPP isolation
Safety Standards
CB Certified for worldwide use
Model Number:
Unit Weight:
Dimensions (W x L x H):
MDS-300ADB18 AA
610g(1.35lb)
88.9x140.0x45.0mm (3.5x5.5x1.75 inch)
General Description
3
Delta’s new MDS-300ADB18 AA enclosed design offers a high energy density (8.8 Watts/in ) 3.5x5.5 inch design for Type BF patient
access medical products. With operating parameters of 90 to 264 Vac universal input voltage, temperatures of -10 to +70 degrees
centigrade, and altitudes of up to 5000 meters (16,400 feet), the design is well suited for a variety of both medical and non-medical
applications. Other features include input surge of 300 Vac, low leakage current, no-load input power < 0.5 Watt, and 800K hour MTBF.
This product is certified for EMC standards EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment, and EN
55022 for Industrial Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are
used to ensure long product life.
An enhanced feature set, includes a 5 Volt/0.5 amp standby output, remote on/off, remote sense, and a power good signal that are
included as part of the standard product.
The design, which has protection against shock compliant with 2XMOPP and Type BF requirements, has both medical (with risk report
available), and ITE safety approvals, including cURus(US&Canada) /TUV/GB(China); plus, CB certificates with all national deviations.
Full compliance with RoHS Directive 2011/65/EU for environmental protection is included.
Model Information (All with 5V/0.5A standby available):
Model Number
Input Voltage Output Voltage
Convection Current Output Forced Air Current Output
MDS-300ADB18 AA
90-264Vac
0-10.55A
MDS
Delta Medical
power Supply
1
300
Max wattage in the
product Series.
300 → 300W
A
Family Code
18Vdc
D
Product Type
D: Enclosed
A~ Z
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
B
Input Type Code
B: 3pin Class I
0-16.66A (with 10 CFM forced air)
18
Output Voltage
18 for 18V
AA
Revision code
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
100-240Vac
Input Voltage Range
90-264Vac
Nominal Input Frequency
50-60Hz
Input Frequency Range
47-63Hz
Input Current (max)
4A
Input Surge Voltage (max)
300Vac for 100ms
Full load Efficiency (typ.)
92.5% @ 115Vac/60Hz
93.5% @ 230Vac/50Hz, Reference Fig.1
Standby Power (max)
0.5W (Inhibit signal high, 18 volt turned off)
@ 115Vac/60Hz, 230Vac/50Hz
Inrush Current (max)
40A @ 230Vac, cold start
Input-PE(protective earth) leakage current (max)
0.1mA @ NC , 0.3mA @ SFC
Output-PE(protective earth) leakage current for Type BF application (max)
0.1mA @ NC , 0.5mA @ SFC
Power Factor (min)
1) NC: normal condition, SFC: single fault condition
1)
1)
1)
1)
0.9 @ 115V/50Hz, 230V/50Hz, full load
Leakage Current
Input-PE Leakage Current
100Vac/60Hz (Typ) 264Vac/60Hz(Typ) Delta Limit
IEC60601-1 Limit
Normal Condition 17.5uA
43.5uA
100uA max
5000uA max
Single Fault Condition 54.5uA
153uA
300uA max
10000uA max
81.5uA
100uA max
100uA max
129.4uA
500uA max
500uA max
Output-PE Leakage Current for Type BF application
Normal Condition 29.5uA
Single Fault Condition 43.7uA
Fig.1 Efficiency versus output load
2
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
Output Ratings / Characteristics—Main Output
Nominal Output Voltage (Vrated)
18V
Total Regulation
± 3%
Output Power
300W 10CFM air, up to 190W convection air
Line Regulation (max)
±0.5%
Load Regulation (max)
±1%
Ripple & Noise (typ.)
1% pk-pk Vrated@300W load, (Refer to Fig. 2 & Fig. 3)
Start-up Time (max)
2000ms @ 115Vac
Hold-up Time (min)
10ms @ 100Vac and 115Vac, with 300W load
20ms @ 100Vac and 115Vac, with 190W load
Dynamic Response (Overshoot & Undershoot O/P Voltage)
± 5% @ with 50-100% load change
Capacitive load (max)
1500uF
Rise time (max)
100ms
Remote Sense
Up to 500mV compensation for voltage drop across
external wire connections to load.
Short and reverse connection protected.
(a) 115V (measured value=160mV)
(b) 230V(measured value=140mV)
Fig.2 Ripple & Noise example, 20MHz BW
Output Ratings / Characteristics—Standby Output
Nominal Output Voltage of standby output (Vrated)
5V
Total Regulation of standby output
± 3%
Ripple & Noise of standby output
100mV max (Refer to Fig. 3)
3
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
Fig. 3 Ripple & Noise measurement circuit
PSU
Load
5cm
Scope
0.47uF Ceramic
10uF Electrolytic
Test Fixture
Mechanical
Case Cover
Enclosed cover(SPCC)
Dimensions(W x L x H)
88.9x140.0x45.0mm (3.5x5.5x1.75 inch)
Unit Weight
610g (1.34lb)
Environment
Surrounding Air Temperature
Operating
Storage
Absolute Maximum/Minimum Rating.
-10°C to +70°C. Linearly derate from 100% load at
50°C, to 50% load at 70°C (2.5%/ degree centigrade)
Note: see power de-rating curves below
-40°C to +85°C
Operating Humidity
5-95% RH (Non-Condensing)
Operating Altitude
5,000 meters (16,400 feet or 50kPa)
Non-operating Altitude
5,000 meters (16,400 feet or 50kPa)
Shock Test (Non-Operating)
50G, 11ms, 3 shocks for each direction
Vibration (Operating)
5-500Hz, 2Grms, 15 minute for each three axis
Power De-rating curves
4
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
Protections (for both 18 Volt and 5Volt Standby outputs, unless otherwise indicated)
Overvoltage (max)
Main output 125% of rated voltage, Latch Mode
Standby Output 125% of rated voltage, Latch Mode
Over load / Over current (max)
Main output 160% of rated current
Standby Output 1.06A max
Hiccup Mode(Non-Latching, Auto-Recovery)
Over Temperature
Latch Mode for main output
Short Circuit
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Reliability Data
MTBF(Minimum) at 115Vac, 190W, 35 °C
Convection Air Flow
800 Khrs based on Telecordia SR-332
Operating life(Minimum) at 115Vac, 190W, 25°C
Convection Air Flow
26,280 hrs
Safety Standards / Directives
nd
rd
rd
Medical Safety
IEC60601-1 2 , 3 and 3.1 edition CB report
rd
IEC60601-1 edition 3.1 (2012), EN60601-1 (2006) +
A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14,
ANSI/AAMI ES60601-1:2005/(R)2012
ITE Safety
IEC60950-1 CB report
TUV60950-1
UL60950-1+CAN/CSA60950-1
GB4943.1-2011, GB9254-2008, GB17625.1-2003
CE
MDD Directive 93/42/EEC
Environmental
RoHS Directive 2011/65/EU Compliant
Galvanic Isolation
1)
5
Input to/Output (2XMOPP) 4000 Vac
1)
Input to/Ground(1XMOPP) 1500 Vac
Output to/Ground(1XMOPP) 1500 Vac ( Type BF application rated)
PSU can support PoE applications with Primary to FG 2500Vac test.
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
th
EMC (Compliant with IEC-60601-1-2 4 Ed. requirements)
EMC / Emissions
Harmonic Current Emissions
EN55011/EN55022,FCC Title 47:Class B
IEC61000-3-2
Meet Class D limit
Immunity to
Voltage Flicker
IEC61000-3-3
Electrostatic Discharge
IEC61000-4-2
Level 4 Criteria A
Air Discharge: 15kV
Contact Discharge: 8kV
Radiated Field
IEC61000-4-3
Electrical Fast Transient / Burst
IEC61000-4-4
Criteria A
80MHz-2700MHz, 10V/m AM modulation
385MHz-5785MHz, 28V/m Pulse mode and other
modulation
1)
Level 3 Criteria A :2kV
Surge
IEC61000-4-5
Level 3 Criteria A
3)
Common Mode : 2kV
4)
Differential Mode : 1kV
Conducted
IEC61000-4-6
Level 2 Criteria A
150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and
1)
1)
1)
1)
Amateur radio bands
Power Frequency Magnetic Fields
Voltage Dips
IEC61000-4-8
IEC61000-4-11
1)
Criteria A
Magnetic field strength 30A/m
1)
Criteria A
0% UT, 0.5 cycle (10ms)
o
o
o
o
o
o
o
o
o
, 0 /45 /90 /135 /180 /225 /270 /315 /360
2)
1)
Criteria B (Criteria A with 190W load)
o
0% UT, 1 cycle (20ms), 0
2)
Criteria B
o
70% UT, 25 cycle (500ms), 0
2)
Criteria B
o
0% UT, 250 cycle (5000ms), 0
1) Criteria A: Normal performance within the specification limits
2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test.
3) Asymmetrical: Common mode (Line to earth)
4) Symmetrical: Differential mode (Line to line)
5) Unless otherwise noted, requirements apply at 300 watt output load
6
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
Dimensions
Mechanical drawing (MDS-300ADB18 AA)
Notes
― Dimensions are in mm(inches)
―There are two locations where assembled power supply is connected to the customer’s product
a. Bottom mounting, use (4X) M3 screws to affix assembled power supply to product’s enclosure. Extruded hole with thread must
be withstand 9Kgf.cm (7.81lb-in) min. Maximum allowed screw penetration is 3.5mm (0.138 inch).
b. Side mounting, use (2X) M3 screws to affix one side of assembled power supply to the product’s enclosure. Extruded hole with
thread must be withstand 9Kgf.cm (7.81lb-in) min. Maximum allowed screw penetration is 3.5mm (0.138 inch).
7
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
c. Input Connector CN1
Pin 1
Neutral
Pin 2
Line
CN1 mates with Molex housing 26033031 and
Molex series 6838 crimp terminals.
Input Line can also be connected to pin 1, and
Input Neutral can be connected to pin 2.
Output Connector
CN102
CN102
CN103
CN103
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6
Pin7
Pin 8
Pin 9
Pin 10
Control and STANDBY connector CN101
Remote sense +
Remote sense –
Power Good +
Power Good -(DC RTN)
Remote ON_OFF/INHIBIT +
Remote ON_OFF/INHIBIT -(DC RTN)
5V Standby
DC RTN
5V Standby
DC RTN
CN101 mates with Molex housing 1041421000 and Molex series 104539
crimp terminals.
CN2 PINGOOD : JP-13T mates with KST:FDFNYD1-187 or other applicable connectors.
System protective earth connection can be connected to the enclosed cover or to CN2.
8
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
Functions
Start-up Time
The time required for the output voltage to reach 90% of its final steady state value, after the input voltage is applied.
Rise Time
The time required for the output voltage to change from 10% to 90% of its final steady state value.
Hold-up Time
Time between the collapse of the AC input voltage, and the output falling to 90% of its steady state value.
■ Graph illustrating the Start-up Time, Rise Time, and Hold-up Time
9
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
Dynamic Response (18 Volt Output)
Overload & Over current Protections
The power supply output voltage will remain within ±5% of its
steady state value, when subjected to a dynamic load 50 to 100%
of its rated current.
■ 50 to 100% Load
The power supply’s Overload (OLP) and Over current (OCP)
Protections will be activated before the 18 volt outputs’ current
exceeds 160% of IO (Max load). Upon such an occurrence, VO
will start to drop. Once the power supply has reached its
maximum power limit, the protection will be activated and the
power supply will go into “Hiccup mode” (Auto-Recovery). The
power supply will recover once the fault condition causing the
OLP and OCP is removed and IO is back within the specified
limit. For the standby output, the performance is the same, with
the exception that the protection will be activated before the
output current exceeds 1.06 amps.
Inrush Current
Inrush current is the input current that occurs when the input
voltage is first applied. For AC input voltages, the maximum peak
value of inrush current will occur during the first half cycle of the
applied AC voltage. This peak value decreases exponentially
during subsequent cycles of AC voltage.
Additionally, if the 18 Volts’ Iout is >100%; but <160% for a
prolonged period of time (depending on the load), the Over
Temperature Protection (OTP) will be activated due to high
temperature on critical components. The power supply will then
go into latch mode.
Short Circuit Protection
The power supply’s output OLP/OCP function also provides
protection against short circuits. When a short circuit is applied,
the output current will operate in “Hiccup mode”, as shown in the
illustration in the OLP/OCP section on this page. The power
supply will return to normal operation after the short circuit is
removed.
Overvoltage Protection
Over Temperature Protection
The power supply’s overvoltage circuit will be activated when its
internal feedback circuit fails. The output voltage shall not exceed
its specifications defined on Page 5 under “Protections”. Power
supply will latch off, and require removal/re-application of input
AC voltage in order to restart.
As mentioned above, the power supply also has Over
Temperature Protection (OTP). This is activated when the
overload condition persists for an extended duration and the
output current is below the overload trigger point but >100% load.
In the event of a higher operating temperature condition at 100%
load, the power supply will run into OTP when the surrounding air
temperature is higher than the operating temperature. When
activated, the output voltage will go into latch mode until the input
voltage is removed; then, reapplied, and the surrounding air
temperature drops to its normal operating temperature.
10
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
Power Good
Power Good+ pin is an open collector transistor (40V/600mA
rating). A resistor (suggested value 10Kohm, 1/8W) can be added
between 5V STANDBY pin (or, other available pull-up voltage
that is no greater than 30V) and the Power Good+ pin (refer to
figure below). Value of pull-up resistor may have to be adjusted,
depending on voltage used, and other end-use conditions of the
Power Good+ pin connection to the product. When AC input is
on, Power Good+ pin will be high. When AC input is off, Power
Good+ pin will be low. There will be a minimum of 5 milliseconds
(at 190W load) between the time the power good goes to low
level, and the time when the output reaches 90% of its rated
value.
11
Remote On_Off/Inhibit
Remote ON_OFF/INHIBIT can be used to enable or disable only
the main output. When the main output is disabled, the +5V
Standby output will continue to operate. This signal can be pulled
down to a low level of 0.3 volts, or shorted to DC-Return, in order
for the main output to be enabled; and, floated (no connection to
the signal), or pulled up to a value greater than or equal to 3 volts,
in order to disable the main output.
Remote Sense
Remote sense feature can be used to compensate for the extra
voltage drop on output wires that are connected from the main
output terminals, to the load. With wires connected from the
remote sense pins, at the same locations as the wires from the
main output, the remote sense function can compensate up to
500mV voltage drop. The power supply will not be damaged if the
remote sense pins are shorted, or if a reverse/inverted polarity
connection is made to the load.
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Technical Datasheet
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18 AA
Certificate
All Delta Medical Power products conform to the European directive 2011/65/EU.
RoHS is the abbreviation for “Restriction of the use of certain hazardous
substances
Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN
ISO 13485:2012 for the design and manufacture of switching power supply and
adaptor for medical device.
In addition to a UL Total Certification Program (TCP) approved client laboratory
for IEC60950 and IEC60065. Delta also has participated UL Client Test Data
Program (CDTP) for IEC 60601
12
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 04)
Similar pages