Sensortechnics DSDX005D4D-A Digital pressure transducer Datasheet

DSDX Series A-grade
Digital pressure transducers
FEATURES
· 0...1 to 0...100 psi,
0...±1 to 0...±15 psi
· Absolute, gage or differential pressure
· Digital readout via I2C-bus
· Precision ASIC conditioning
· Temperature compensated
· Non-ratiometric output
· Total accuracy ±0.5 to ±1.0 %FSS
SPECIFICATIONS
ELECTRICAL CONNECTION
Maximum ratings
N/C
Supply voltage VS
6
Lead temperature (2 - 4 sec.)
Temperature ranges
Compensated
Operating
Storage
Vibration
1
250°C
0 to +85°C
-20 to +105°C
-40 to +125°C
5
DSDX
A-grade
Application of supply voltage above the
maximum can cause electrical failure.
February 2009 / 611
SDA
220 nF
IC
3
GND
N/C
Note:
Incorrect applications of excitation
voltage or ground to the wrong pin can
cause electrical failure.
SCL
15 nF
8 4
100 g for 11 ms
Caution! The sensor is not reverse polarity protected.
7
2
10 g at 20 - 2000 Hz
Shock
+Vs
4.75 V to 5.25 VDC
max. 6.50 VDC
A capacitor of 220 nF is required
between +Vs and GND.
Pin 2 is an internal device connection and
should be connected to ground with a
15 nF capacitor.
It is important to place the capacitors as
close to the pins as possible!
Pins 4, 6 and 8 are internal device
connections and should not be
connected for any reason!
1/7
www.sensortechnics.com
DSDX Series A-grade
Digital pressure transducers
PRESSURE RANGES SPECIFICATIONS
(VS = 5.0 VDC, TA = 25°C)
Part number
Pressure range
Burst pressure1
30
Sensitivity (typ.)
DSDX015A2R-A
0 to 15
DSDX030A2R-A
0 to 30
DSDX100A2R-A
0 to 100
150
33
DSDX001G2R-A / D4R-A
0 to 1
5
3277
DSDX005G2R-A / D4R-A
0 to 5
20
655
DSDX015G2R-A / D4R-A
0 to 15
30
218
DSDX030G2R-A / D4R-A
0 to 30
DSDX100G2R-A / D4R-A
0 to 100
DSDX001D4D-A
0 to ±1
5
1638
DSDX005D4D-A
0 to ±5
20
328
DSDX015D4D-A
0 to ±15
30
109
psi(a)
psi(g,d)
60
60
150
218
psi(a)
psi(g,d)
109
counts/
p si
109
33
Specification notes:
1. If maximum burst pressure is exceeded, even momentarily, the package may leak or burst, or the pressure sensing die
may fracture.
2. Span is the algebraic difference between the output signal for the highest and lowest specified pressure.
3. Total accuracy is the combined error from offset and span calibration, linearity, pressure hysteresis, and temperature
effects. Linearity is the measured deviation based on a straight line. Hysteresis is the maximum output difference at
any point within the operating pressure range for increasing and decreasing pressure. Calibration errors include the
deviation of offset and full scale from nominal values.
4. Delay time between sampling and signal change at the output.
5. The smallest change in the output voltage, given any change in pressure.
February 2009 / 611
2/7
www.sensortechnics.com
DSDX Series A-grade
Digital pressure transducers
PERFORMANCE CHARACTERISTICS
The output signal is not ratiometric to the supply voltage (VS = 5.0 VDC, TA = 25°C)
All DSDX...R-A
Characteristics
Zero pressure offset
Min.
Typ.
Max.
DSDX001...
378
410
442
all others
394
410
426
2
3276
Full scale span (FSS)
Full scale output
3
Total accuracy (0 to 85°C)
counts
DSDX001...
3654
3686
3718
all others
3670
3686
3702
DSDX001...
±1.0
all others
±0.5
Sample rate
100
Response delay4
2.73
Units
%FSS
Hz
14.11
Startup time (power up to 1st result)
40
ms
Quantization step5
2
counts
Current consumption
6
mA
All DSDX...D-A
Characteristics
Zero pressure offset
Min.
Typ.
Max.
DSDX001...
2016
2048
2080
all others
2032
2048
2064
Full scale span (FSS)2
Output
Units
3276
at max. specified pressure
at min. specified pressure
3
Total accuracy (0 to 85°C)
DSDX001...
3654
3686
3718
all others
3670
3686
3702
DSDX001...
378
410
442
all others
394
410
426
DSDX001...
±1.0
all others
±0.5
Sample rate
100
Response delay4
2.73
counts
%FSS
Hz
14.11
Startup time (power up to 1st result)
40
ms
Quantization step5
2
counts
Current consumption
6
mA
February 2009 / 611
3/7
www.sensortechnics.com
DSDX Series A-grade
Digital pressure transducers
INTRODUCTION
The DSDX is capable to generate a digital output signal. It
runs a cyclic program, which will store a corrected 12-bit
sensor value about every 10 ms within the output registers
of the internal ASIC. In order to use the pressure transducer
for digital signal readout, the device should be connected
to a bidirectional I²C-bus.
STOP condition (P): LOW to HIGH transition of SDA line while
clock (SCL) is HIGH determines STOP condition. STOP
conditions are always generated by the master. More than
one request for the current pressure value can be transmitted without generation of intermediate STOP condition.
DATA valid (D): State of data line represents valid data when,
after START condition, data line is stable for duration of
HIGH period of clock signal. Data on line must be
changed during LOW period of clock signal. There is
one clock pulse per bit of data.
According to the I²C-bus communication specification, the
bus is controlled by a master device, which generates the
clock signal, controls the bus access and generates START
and STOP conditions. The DSDX is designed to work as a
slave, hence it will only respond to requests from a master
device.
Acknowledge (A): Data is transferred in pieces of 8 bits
(1 byte) on serial bus, MSB first. After each byte receiving
device – whether master or slave – is obliged to pull
data line LOW as acknowledge for reception of data.
Master must generate an extra clock pulse for this purpose. When acknowledge is missed, slave transmitter
becomes inactive. It is on master either to send last command again or to generate STOP condition in that case.
The I²C-bus master-slave concept requires a unique
address for each device. The DSDX has a hard coded slave
address (1111000xb), therefore it is not possible to access
more than one DSDX on the same I²C-bus line.
DIGITAL I2C INTERFACE
Slave address: Each device connected to the bus has a
unique slave address. After generating a START condition, the master has to transmit the slave address for
the DSDX with a READ command: $F1. The DSDX must
not be accessed with a WRITE command ($F0) as the
correct function of the device can not be guaranteed in
this case (note: a power-down power-up change will
force the sensor to use factory data again).
The DSDX complies with the following protocol (FIGURE I):
Bus not busy: During idle periods both data line (SDA) and
clock line (SCL) remain HIGH.
START condition (S): HIGH to LOW transition of SDA line
while clock (SCL) is HIGH is interpreted as START condition. START conditions are always generated by the
master. Each request for the current pressure value must
be initiated with a START.
DATA operation: The DSDX starts to send 2 data bytes containing
the current pressure value placed in the output registers.
SCL
SDA
START
condition
Data
valid
1
S Slave Address R/W A
Data allowed
to change
Data Byte 1
A
STOP
condition
Data Byte 2
S
Slave Address
Read out of pressure value
generated by master
generated by slave
P
Next command
S = START condition
Data Byte 1 = High Byte (MSB first)
A = Acknowledge
Data Byte 2 = Low Byte (LSB last)
P = STOP condition
FIGURE I: I²C-BUS Protocol of DSDX
February 2009 / 611
4/7
www.sensortechnics.com
DSDX Series A-grade
Digital pressure transducers
I2C INTERFACE PARAMETERS
Parameter
Symbol
Input high level
Input low level
Output low level
Pull up current
Load capacitance SDA
SCL clock frequency
Bus free time between
STOP and START condition
Hold time (repeated)
START condition
LOW period of SCL
HIGH period of SCL
Setup time repeated
START condition
Data hold time
Data setup time
Rise time of both
SDA and SCL
Fall time of both
SDA and SCL
Setup time for
STOP condition
Input filter spike
suppression
Conditions
Min.
Typ.
90
0
(open drain, IOL = -4 mA)
(pins SCL and SDA)
Max.
100
10
10
20
400
100
5
Unit
% of
Vs
µA
pF
kHz
FSC L
---
tBUF
4.7
µs
4.0
µs
tLOW
tHIGH
4.7
4.0
µs
µs
tSU.STA
4.7
µs
tHD.DAT
tSU.DAT
0
250
ns
ns
tR
---
300
ns
tF
---
300
ns
tSU.STO
4
to first clock pulse
tHD.STA
tSP
tSU;DAT
µs
sp i ke s o n S D A o r S C L o f
that length are suppressed
tHD;DAT
50
tSU;STA tHD;STA
tR
tSU;STO tBUF
tHD;STA
ns
tF
SCL
SDA
tHIGH
tLOW
SCL
SDA
FIGURE II: Timing characteristics of the I2C interface
February 2009 / 611
5/7
www.sensortechnics.com
DSDX Series A-grade
Digital pressure transducers
PHYSICAL DIMENSIONS
D4 package
1.27 typ
(0.050 typ)
16.89
(0.665)
7.87 typ
(0.310 typ)
13.97
(0.550)
3.63
(0.143)
High pressure port
2.29
(0.090)
3.38
(0.133)
8.13
8.89
(0.350) (0.320)
8.26 typ
(.325 typ)
Low pressure port
0.51 typ
(0.02 typ)
1.78
(0.070)
2.54 typ
(0.100 typ)
16.64 ±0.64
(0.655 ±0.025)
4.5
(0.177)
6.99
(0.275)
G2 package
3.15
(0.124)
1.27 typ
(0.050 typ)
13.97
(0.550)
16.89
(0.665)
4.9
(0.193)
2 ° typ
1.19 typ
(0.047 typ)
2.29 typ
(0.090 typ)
2.79 typ
(0.11 typ)
8.89
5.08 (0.350)
(0.200)
0.51 typ
(0.02 typ)
16.64 ±0.64
(0.655 ± 0.025)
4.5
(0.177)
2.54 typ
(0.100 typ)
A2 package
16.89
(0.665)
8.46
(0.333)
1.27 typ
(0.050 typ)
13.97
(0.550)
2.29 typ
(0.090 typ)
8.0 typ
(0.315 typ)
8.89
(0.350)
16.64 ±0.64
(0.655 ±0.025)
third angle projection
February 2009 / 611
4.5
(0.177)
0.51 typ
(0.02 typ)
2.54 typ
(0.100 typ)
dimensions in mm (inches)
6/7
www.sensortechnics.com
DSDX Series A-grade
Digital pressure transducers
ELECTRICAL CONNECTIONS
DSDX...D4...
Vs (pin 1)
identifier
+Vs
IC
GND
N/C
SCL
N/C
SDA
N/C
+Vs
IC
GND
N/C
DSDX...A2R-A
SCL
N/C
SDA
N/C
DSDX...G2R-A
Vs (pin 1)
identifier
Vs (pin 1)
identifier
+Vs
IC
GND
N/C
SCL
N/C
SDA
N/C
Note: Pins 4, 6 and 8 are internal device
connections and should not be
connected for any reason.
Pin 2 is an internal device connection
and should be connected to ground
with a 15 nF capacitor.
ORDERING INFORMATION
Pressure range
Absolute
Gage
Differential/Gage
0 to 1 psi
DSDX001G2R-A
DSDX001D4R-A
0 to 5 psi
DSDX005G2R-A
DSDX005D4R-A
0 to 15 psi
DSDX015A2R-A
DSDX015G2R-A
DSDX015D4R-A
0 to 30 psi
DSDX030A2R-A
DSDX030G2R-A
DSDX030D4R-A
0 to 100 psi
DSDX100A2R-A
DSDX100G2R-A
DSDX100D4R-A
0 to ±1 psi
DSDX001D4D-A
0 to ±5 psi
DSDX005D4D-A
0 to ±15 psi
DSDX015D4D-A
Other pressure ranges or calibrations are available on request
Sensortechnics reserves the right to make changes to any products herein. Sensortechnics does not assume any liability arising out of the
application or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
February 2009 / 611
7/7
www.sensortechnics.com
Similar pages