Total Power MSHU114 Reinforced insulation rated for 300vac working voltage Datasheet

FEATURES
►Ultra-compact SMD-Package
►Reinforced Insulation rated for 300VAC Working Voltage
►I/O-isolation 4000VACrms
►Industrial & Medical Safety Approval
►Operating Temp. Range –25°C to +80°C
►Qualified for leadfree Reflow Solder Process according
IPC-J-STD-020D
►Tape & Reel Package available
►3 Years Product Warranty
PRODUCT OVERVIEW
The MSHU100 series is a new range of 2W DC/DC-converter modules providing a very high I/O-isolation voltage of 4000 VAC. The product comes
in a small SMD-package. There are 15 models available with 5V, 12V or 24VDC input and single-or dual-output voltages.
The MSHU100 DC/DC converters offer an economical solution for many applications in instrumentation, industrial controls, medical equipment and
everywhere where a certified supplementary- or reinforced insulation system is required to comply with requested safety standards.
Model Selection Guide
Model
Input
Output
Number
Voltage
Voltage
Output Current
Input Current
Load
Max. capacitive
Regulation
Load
Max.
Min.
@Max. Load
@No Load
VDC
mA
mA
mA(typ.)
mA(typ.)
5
400
8
606
12
165
3
600
15
133
2.5
605
±12
±83
±1.5
553
MSHU109
±15
±66
±1
542
10
MSHU112
5
400
8
253
12
12
165
3
250
15
133
2.5
252
±12
±83
±1.5
224
10
±15
±66
±1
220
10
(Range)
VDC
MSHU102
MSHU104
MSHU105
MSHU108
5
(4.5 ~ 5.5)
MSHU114
MSHU115
MSHU118
12
(10.8 ~ 13.2)
MSHU119
% (max.)
8
126
12
3
125
10
15
133
2.5
126
±12
±83
±1.5
112
10
±15
±66
±1
110
10
30
72
73
66
330
10
400
MSHU129
100#
10
165
66
66
10
5
(21.6 ~ 26.4)
330
10
40
%
66
10
90
12
MSHU128
uF
12
MSHU124
24
(typ.)
@Max. Load
MSHU122
MSHU125
Efficiency
66
66
100#
74
75
66
330
10
66
66
100#
74
75
# For each output
Input Specifications
Parameter
Input Voltage Range
Input Surge Voltage (1 sec. max.)
Model
Min.
Typ.
Max.
5V Input Models
4.5
5
5.5
12V Input Models
10.8
12
13.2
24V Input Models
21.6
24
26.4
5V Input Models
-0.7
---
9
12V Input Models
-0.7
---
18
24V Input Models
-0.7
---
30
---
---
0.3
---
---
Reverse Polarity Input Current
Input Filter
Internal Power Dissipation
In
Total Power International,
Inc
All Models
Unit
VDC
A
Internal Capacitor
Toll Free: 877-646-0900
650
www.total-power.com
mW
MSHU100 SERIES
Total Power International, Inc.DC/DC CONVERTER 2W, Reinforced
Tota
Insulation, Medical Safety
Output Specifications
Parameter
Output Voltage Accuracy
Output Voltage Balance
Line Regulation
Load Regulation
Ripple & Noise (20MHz)
Ripple & Noise (20MHz)
Ripple & Noise (20MHz)
Temperature Coefficient
Short Circuit Protection
Conditions
Min.
-------
Dual Output, Balanced Loads
Vin=Min. to Max.
Io=20% to 100%
Typ.
Max.
±2.0
±4.0
±0.1
±1.0
±1.2
±1.5
See Model Selection Guide
100
150
--200
--15
±0.01
±0.02
0.5 Second Max.
---------
Over Line, Load & Temp.
Unit
%
%
%
mV P-P
mV P-P
mV rms
%/℃
Isolation, Safety Standards
Parameter
Conditions
Min.
Typ.
Max.
Unit
60 Seconds
4000
---
---
VACrms
Flash tested for 1 Second
6000
---
---
VPK
240VAC, 60Hz
---
---
2
uA
I/O Isolation Resistance
500 VDC
10
---
---
GΩ
I/O Isolation Capacitance
100KHz, 1V
---
15
20
pF
I/O Isolation Voltage (rated)
I/O Isolation Test Voltage
Leakage Current
cUL/UL60950-1, CSA C22.2 No. 60950-1-03
Safety Standards
UL60601-1,CSA C22.2 No.601-1
IEC/EN 60950-1, IEC/EN 60601-1
IEC60950-1 CB report, cUL/UL 60950-1 certificate
Approvals
UL60601-1 UL certificate
General Specifications
Parameter
Conditions
Switching Frequency
MTBF (calculated)
MIL-HDBK-217F@25℃, Ground Benign
Moisture Sensitivity Level (MSL)
Min.
Typ.
Max.
Unit
50
80
100
KHz
2,000,000
---
---
Hours
IPC/JEDEC J-STD-020D
Level 2
Input Fuse
5V Input Models
1000mA Slow-Blow Type
12V Input Models
500mA Slow-Blow Type
24V Input Models
200mA Slow-Blow Type
Environmental Specifications
Parameter
Operating Temperature Range (with Derating)
Case Temperature
Storage Temperature Range
Humidity (non condensing)
Cooling
Lead Temperature (1.5mm from case for 10Sec.)
Conditions
Ambient
Min.
-25
---50
--Free-Air convection
---
Max.
+80
+90
+125
95
Unit
℃
℃
℃
% rel. H
260
℃
Power Derating Curve
100
Natural
convection
80
100LFM
200LFM
60
400LFM
40
20
0
~
-25
0
20
40
60
80
100 110
Ambient Temperature ]
In
Total Power International,
Inc
Toll Free: 877-646-0900
www.total-power.com
MSHU100 SERIES
Total Power International, Inc. DC/DC CONVERTER 2W, Reinforced
Tota
Insulation, Medical Safety
Notes
1
Specifications typical at Ta=+25℃, resistive load, nominal input voltage and rated output current unless otherwise noted.
2
Ripple & Noise measurement bandwidth is 0-20MHz.
3
These power converters require a minimum output loading to maintain specified regulation, operation under no-load conditions will not damage these modules;
4
All DC/DC converters should be externally fused at the front end for protection.
however they may not meet all specifications listed.
5
Other input and output voltage may be available, please contact factory.
6
That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
7
Specifications subject to change without notice.
8
It is not recommended to use water-washing process on SMT units.
Package Specifications
Mechanical Dimensions
Connecting Pin Patterns
24.0 [0.94]
15.24 [0.6]
17.78 [0.70]
2.54
[0.1]
10 9
1
7 8
13.7 [0.54]
16
15.24 [0.60]
18.1 [0.71]
3.11
[0.12]
2.54 [0.10]
19.3 [0.76]
1.00 [0.04]
9.0 [0.35]
2.8 [0.11]
1.6 [0.06]
►All dimensions in mm (inches)
0~
4
9.25 0.5
[0.36 0.02]
0.25 [0.01]
S SEATING
PLANE
►Tolerance: X.X±0.25 (X.XX±0.01)
0.15 S
X.XX±0.13 ( X.XXX±0.005)
►Pins ±0.05 (±0.002)
Pin Connections
Pin
Single Output
Dual Output
1
-Vin
-Vin
7
NC
NC
8
NC
Common
9
+Vout
+Vout
10
-Vout
-Vout
16
+Vin
+Vin
Physical Characteristics
Case Size
:
24.0x13.7x9.0mm (0.94x0.54x0.35 Inches)
Case Material
:
Non-Conductive Black Plastic (flammability to UL 94V-0 rated)
Weight
:
3.75g
NC: No Connection
In
Total Power International,
Inc
Toll Free: 877-646-0900
www.total-power.com
MSHU100 SERIES
Tota
Total Power International, Inc. DC/DC CONVERTER 2W, Reinforced
Insulation, Medical Safety
Test Configurations
Peak-to-Peak Output Noise Measurement Test
Use a Cout 0.47uF ceramic capacitor. Scope measurement should be made by using a BNC socket, measurement bandwidth is 0-20 MHz. Position the load between
50 mm and 75 mm from the DC/DC Converter.
+Vin
+Out
Copper Strip
Single Output
DC / DC
Converter
-Vin
Cout
-Out
Scope
Resistive
Load
Copper Strip
+Vin
+Out
Copper Strip
Dual Output
DC / DC
Converter
Com.
Copper Strip
-Vin
-Out
Scope
Cout
Cout
Resistive
Load
Scope
Copper Strip
Design & Feature Considerations
Maximum Capacitive Load
The MSHU100 series has limitation of maximum connected capacitance at the output. The power module may be operated in current limiting mode during start-up,
affecting the ramp-up and the startup time. For optimum performance we recommend 100uF maximum capacitive load for dual outputs and 330uF capacitive load for
single outputs. The maximum capacitance can be found in the data sheet.
Input Source Impedance
The power module should be connected to a low ac-impedance input source. Highly inductive source impedances can affect the stability of the power module. In
applications where power is supplied over long lines and output loading is high, it may be necessary to use a capacitor at the input to ensure startup.
Capacitor mounted close to the power module helps ensure stability of the unit, it is recommended to use a good quality low Equivalent Series Resistance (ESR <
1.0Ω at 100 KHz) capacitor of a 2.2uF for the 5V input devices, a 1.0uF for the 12V input devices and a 0.47uF for the 24V input devices.
+
DC Power
Source
+Vin
+
+Out
DC / DC
Converter
Load
Cin
-
-Vin
-Out
Output Ripple Reduction
A good quality low ESR capacitor placed as close as practicable across the load will give the best ripple and noise performance. To reduce output ripple, it is
recommended to use 3.3uF capacitors at the output.
+
+Vin
-
+Out
Single Output
DC / DC
Converter
DC Power
Source
-Vin
+
Cout
Load
-Out
+Vin
DC Power
Source
-
+Out
Dual Output
Com.
DC / DC
Converter
-Vin
-Out
Cout
Load
Load
Cout
Thermal Considerations
Many conditions affect the thermal performance of the power module, such as orientation, airflow over the module and board spacing. To avoid exceeding the
maximum temperature rating of the components inside the power module, the case temperature must be kept below 90℃. The derating curves are determined from
measurements obtained in a test setup.
Position of air velocity
probe and thermocouple
50mm / 2in
15mm / 0.6in
In
Total Power International,
Inc
Air Flow
DUT
Toll Free: 877-646-0900
www.total-power.com
Similar pages