AIMTEC AMEC30-24SMAZ 30 watt | ac/dc converter Datasheet

Series AMEC30-MAZ
30 Watt | AC-DC / DC-DC Converter
FEATURES:
•
•
•
•
4000VAC I/O isolation
Medical Switching power modules for PCB mounting
Operating temperature: -40 to +85°C
Universal Input: 90-260VAC, 47-440Hz, or 120-370VDC
AMEC30-3.3SMAZ
AMEC30-5SMAZ
AMEC30-12SMAZ
AMEC30-15SMAZ
AMEC30-24SMAZ
Input Voltage
(VAC/Hz)
Input Voltage
(VDC)
90-264/47-440
90-264/47-440
90-264/47-440
90-264/47-440
90-264/47-440
120-370
120-370
120-370
120-370
120-370
Input Voltage
(VAC/Hz)
Input Voltage
(VDC)
Output Voltage
(V)
3.3
5
12
15
24
Models
Dual output
Model
Energy star compliant
RoHS compliant
Low Ripple and Noise
CE, cULus approvals
CB
Models
Single output
Model
•
•
•
•
Output Voltage
(V)
Output Current
max
(A)
Efficiency
(%)
6
6
2.5
2
1.25
80
83
85
85
84
Output Current
max
(A)
Efficiency
(%)
AMEC30-5DMAZ
90-264/47-440
120-370
±5
±3
81
AMEC30-12DMAZ
90-264/47-440
120-370
±12
±1.25
84
AMEC30-15DMAZ
90-264/47-440
120-370
±15
±1
83
AMEC30-24DMAZ
90-264/47-440
120-370
±24
±0.625
82
NOTE: All specifications in this datasheet are measured at an ambient temperature of 25°C, humidity<75%, nominal input voltage and at
rated output load unless otherwise specified.
Input Specifications
Parameters
Conditions
Current
Inrush current <2ms
Leakage current
External fuse
Input dissipation
Start-up time
Output Specifications
Parameters
Voltage accuracy
Line regulation
Load regulation
Cross regulation
Maximum Capacitive load
Transient response deviation
Ripple & Noise*
Hold-up time (min)
Minimum Load Current
Typical
115 VAC
230 VAC
115 VAC
230 VAC
Recommended slow blow type
No load; 115VAC
No load; 264VAC
3.15
<0.5
<0.8
117
Conditions
Typical
(LL-HL)
0-100% load single
0-100% load dual
st
nd
25% load - 1 out, 100% load – 2 out
Depending of the model
25% load Step
20MHz bandwidth
115VAC
*Ripple & Noise measured with 0.1μF M/C and 1μF E/C
www.aimtec.com
F 052e R8.N
Tel: +1 514 620 2722
1 of 4
±2
±0.5
±1
±2
±5
470-23 000
±2
50
29
0
Maximum
740
410
25
50
150
Units
mA
A
µA
A
W
ms
Maximum
Units
%
%
%
%
µF
% of Vout
mV p-p
ms
% of Max
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Series AMEC30-MAZ
30 Watt | AC-DC / DC-DC Converter
Isolation Specifications
Parameters
Conditions
Tested I/O voltage
Isolation Resistance
General Specifications
Parameters
Switching frequency
Over current protection
Over voltage protection
Short circuit protection
Short Circuit restart
Operating temperature
Maximum case temperature
Storage temperature
Temperature coefficient
Cooling
Humidity
Case material
Weight
Dimensions (L X H X W)
MTBF
Shock
Vibration
>1000
Conditions
VAC
MΩ
Maximum
140
110
Metal
Units
KHz
%
%
o
C
C
o
C
% / oC
o
100
-40 to +100
0.02
Free air convection
Non condensing
Converter operation
Units
4000
47
110
Zener diode clamp
Hiccup mode, indefinite
Auto recovery
-40 - +85
With derating above 55ºC
Parameters
Wave form
Acceleration amplitude
Bump duration
Converter operation
Number of bumps
Test mode
Displacement
Acceleration
Rated
Typical
Auto recovery
Environment Approval
Test
Typical
3 sec
95
% RH
270
4.92 x 2.56 x 1.38 inch / 125.00 x 65.00 x 35.00 mm
> 800 000 hrs (MIL-HDBK -217F, t=+25oC)
g
Conditions
Half sine wave
5gn
30 ms
Before and after test, body mounted (on chassis)
18 (3 in each direction for every axis)
Sweep sine, 10-100Hz, speed 0.05Hz/s
1 mm
3g, 3 loops 30min one cycle, 3h total, every axis tested
Before and after test, body mounted (on chassis)
Safety Specifications
Parameters
Agency approvals
Standards
www.aimtec.com
F 052e R8.N
cULus, CE, CB Scheme
Medical Electrical Equipment
Information technology Equipment
EMI - Conducted and radiated emission
Harmonic Current Emissions
Voltage fluctuations and flicker
Electrostatic Discharge Immunity
RF, Electromagnetic Field Immunity
Electrical Fast Transient/Burst Immunity
Surge Immunity
RF, Conducted Disturbance Immunity
Power frequency Magnetic Field Immunity
Voltage dips, Short Interruptions Immunity
IEC\EN\UL 60601-1, CSA-C22.2 No. 601.1-M90
EN 60950-1:2006+A11:2009
EN55011, class B
IEC/EN 61000-3-2, (EN60555-2)
IEC/EN 61000-3-3, (EN60555-3)
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-8
IEC 61000-4-11
Tel: +1 514 620 2722
2 of 4
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Series AMEC30-MAZ
30 Watt | AC-DC / DC-DC Converter
Pin Out Specifications
1
2
3
4
5
6
Single
AC Input (L)
AC Input (N)
+V Output
-V Output
N.C.
N.C.
Dimensions
Dual
AC Input (L)
AC Input (N)
+V Output
Common
Common
-V Output
2.56
(65.00)
2.56
(65.00)
1
Mounting
Holes
(including
DIN rail
bracket)
(M3)
2
AC Input
Connector
4.90
(125.00)
0.81
(20.50)
1.28
(32.50)
0.88
(22.25)
2.10
(53.25)
68.63
(2.700)
Pin
3 4 5 6
DC Output
Connector
Bottom View
Top View
Dimensions: inch (mm)
Case Tolerance: ±0.05 (±1.30)
35.00
(1.38)
Side View
PCB Mounting
Screw
PCB Mounting
Screw
Mylar Film
PCB
Bracket
Mounting
Screw
PCB
Mylar Film
2.6 mm Max
Customer Chassis
www.aimtec.com
F 052e R8.N
Tel: +1 514 620 2722
3 of 4
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Series AMEC30-MAZ
30 Watt | AC-DC / DC-DC Converter
Optional DIN Rail Bracket
Derating
Free Air Convection
55º C
Output Power (%)
100
80
Safe Operating Area
60
40
20
0
- 40
- 20
0
20
40
60
80
85
100
Ambient Temperature º C
NOTE: 1. Datasheets are updated as needed and as such, specifications are subject to change without notice. Once printed or downloaded, datasheets are no longer controlled
by Aimtec; refer to www.aimtec.com for the most current product specifications. 2. Product labels shown, including safety agency certifications on labels, may vary based on the
date manufactured. 3. Mechanical drawings and specifications are for reference only. 4. All specifications are measured at an ambient temperature of 25°C, humidity<75%,
nominal input voltage and at rated output load unless otherwise specified. 5. Aimtec may not have conducted destructive testing or chemical analysis on all internal components
and chemicals at the time of publishing this document. CAS numbers and other limited information are considered proprietary and may not be available for release. 6. This
product is not designed for use in critical life support systems, equipment used in hazardous environments, nuclear control systems or other such applications which necessitate
specific safety and regulatory standards other the ones listed in this datasheet. 7. Warranty is in accordance with Aimtec’s standard Terms of Sale available at www.aimtec.com.
www.aimtec.com
F 052e R8.N
Tel: +1 514 620 2722
4 of 4
Toll free: + 1 888 9 AIMTEC
(924 6832)
North America only
Similar pages