SL Power MINT1150A1206K01 The mint1150 family is a high power density for a power supply in a 2â x 4â size. Datasheet

MINT1150
150W Single Output
Medical Series
Features

2” x 4” x 1.3” Package

For 1U Applications

150W w/air, 100W convection cooled

Universal Input 90-264 VAC

Power Fail/Output Good Signal

Approved to CSA/EN/IEC/UL60601-1,
3rd Edition

2 x MOPP Input to Output Isolation

3 Year Warranty

RoHS Compliant
Description
The MINT1150 family is a high power density for a power supply in a 2” x 4” size. Approved to EN/IEC/UL60601-1, 3rd
edition, with isolation levels which satisfy the 2 MOPP requirements. The MINT1150 family is ideal for Portable Medical
Devices, and many other applications where medical certifications, power density and cost are critical. The MINT1150
operates at universal input range of 90 to 264Vac and wide temperature range -10°C to +70°C, delivering full rated output
power up to +50°C. In addition, these models feature Power Fail and DC OK signals.
Model Selection
Model
Number
Volts
MINT1150A1206K01
12V
12.5A
MINT1150A1506K01
15V
MINT1150A2406K01
Ripple &
Noise**
Total
Regulation
OVP
Threshold**
8.33A
0.5%RMS, 1.2% pk-pk
5%
14.0  1.1V
10.0A
6.67A
0.5%RMS, 1.0% pk-pk
5%
18.0 ± 1.5V
24V
6.25A
4.17A
0.5%RMS, 1.0% pk-pk
5%
28.0 ± 2.5V
MINT1150A4806K01
48V
3.13A
2.08A
0.5%RMS, 1.0% pk-pk
5%
55.0 ± 4.0V
MINT1150A5606K01
56V
2.68A
1.79A
0.5%RMS, 1.0% pk-pk
5%
<59.9V
Notes:
Output Current
w/200LFM air
Convection*
*
Maximum output power is 95 Watts for input voltage of 90-105VAC at 50°C convection. For input voltage of 105Vac or more, the total power is
100 Watts at 50°C convection.
** Measured with noise probe directly across output terminals, and load terminated with 0.1μF ceramic and 10μF low ESR capacitors.
General Specifications
AC Input
Input Current
Inrush Current
www.slpower.com
100-240Vac, ±10%, 47-63Hz, 1
120–370Vdc
Turn On Time
Less than 2 sec. @115Vac (inversely
proportional to input voltage and thermistor
temperature)
115Vac: 2A, 230Vac: 1A
Hold-up Time
>12mS at full load, 120Vac
264Vac, cold start: will not exceed 50A
Signals
MINT1150 REV 1.4, 13-NOV-12
AC Power Fail, DC OK
1
MINT1150
150W Single Output
Medical Series
General Specifications (continued)
Input Fuses
Earth Leakage
Current
Efficiency
Output Power
Transient
Response
F1, F2: 4A, 250Vac fuses provided on all
models
<300μA@264Vac, 60Hz, NC
Overload
Protection
Short Circuit
Protection
Overvoltage
Protection
89% typical at 115Vac
150W continuous with 200 lfm airflow, 100W
convection cooled – See chart for specific
voltage model ratings.
50% load step. i/t: <0.2A/μS
Max Volt Deviation = 3%
Switching
Frequency
Isolation
Hiccup Mode
Provided - no damage will occur if the output is
shorted. Hiccup Mode
OVP firing reduces output voltage to <50% of
nominal in <50mS. See chart for trip range.
PFC: Variable 30-400kHz.
Main Converter: Variable 35-180kHz, 65-70kHz
at full load.
Input-Output: 4000Vac, 2 x MOPP
Input-Ground: 1800Vac, 1 x MOPP
Output-Ground: 500Vac
Ripple and Noise
See chart
Operating
Temperature
-10°C to +70°C
Output Voltage
See chart
Temperature
Derating
Derate output power linearly above 50°C to 50%
at 70°C
+/-5% from nominal
Storage
Temperature
-40°C to +85°C
Voltage
Adjustability
Minimum Load
Total Regulation
Vibration
Dimensions
Weight
Altitude
Not required
Relative Humidity
+/- 5%. See chart
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis
Non-Operating: 0.026g2/Hz, 5.0grms overall,
3 axes, 1 hr/axis
Shock
Safety Standards
W: 2.0” x L: 4.0” x H: 1.3”
MTBF
183g
Operating: -500 to 10,000 ft
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Operating: Half-sine, 20gpk, 10ms, 3 axes, 6
shocks total
Non-Operating: Half-sine, 40 gpk, 10 ms, 3
axes, 6 shocks total
EN/CSA/UL/IEC60601-1, 3rd Edition
640,000 hours at 100W convection, 1,500,000
hours at 150W with 200LFM air
Auxiliary Signals
AC Power Fail:
During normal operations, stays HIGH.
Power Fail:
DC OK:
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
Goes LOW with 5 mS warning before loss
of output power due to AC failure
Open collector logic signal goes and stays HIGH
100mS to 500mS after main output reaches
regulation.
2
MINT1150
150W Single Output
Medical Series
EMI/EMC Compliance
Conducted Emissions
EN55011/22 Class B, FCC Part 15, Subpart B, Class B
Radiated Emissions
EN55011/22 Class A, FCC Part 15, Subpart B, Class A w/6db margin
Static Discharge Immunity
EN61000-4-2, Criteria A, 6kV Contact Discharge, 8kV air discharge
Radiated RF Immunity
EN61000-4-3, 3V/m. Criteria A
EFT/Burst Immunity
EN61000-4-4, 2kV/5kHz, Criteria A
Line Surge Immunity
EN61000-4-5, 1kV differential, 2kV common-mode, Criteria A
Conducted RF Immunity
EN61000-4-6, 3Vrms, Criteria A
Power Frequency Magnetic Field Immunity
EN61000-4-8, 3A/m, Criteria A
Voltage Dip Immunity
EN61000-4-11, 0% Vin, 0.5cycle; 40% Vin, 5 cycles; 70% Vin, 25 cycles;
Criteria A
Line Harmonic Emissions
EN61000-3-2, Class A, B, C, & D
Flicker Test
EN61000-3-3, Complies (dmax<6%)
Mechanical Drawing
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
3
MINT1150
150W Single Output
Medical Series
Connector Information
Input Connector
J100
Ground
J101
DC Output Connector
J200
Signal Connector
J300
PIN 1) AC LINE
PIN 2) EMPTY
PIN 3) AC NEUTRAL
0.187” FASTON TAB
PIN 1) +Vout PIN 4) -Vout
PIN 2) +Vout PIN 5) -Vout
PIN 3) +Vout PIN 6) -Vout
PIN 1) PF/DC OK
PIN 2) Common
Mating Connector:
Molex 09-50-3031
Pins= 08-52-0072
Mating Connector:
Molex 01-90020005
Mating Connector:
AMP 640250-6
Pins = 640252-1
Mating Connector:
Molex 1375820-2
Pins = 1375819
Power Fail/DC OK Signals – J300
J300–Pin 1
J300 – Pin 2
AC POWER FAIL SIGNAL
DC OK SIGNAL
Isolation Specifications
Parameter
Conditions/Description
Insulation Safety Rating
Input/Ground
Input/Output
Output/Ground
Electric Strength Test Voltage
Input/Ground
Input/Output
Output/Ground
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
Min
Nom
Max
Units
-
Vac
Vac
Vac
2 MOPP
1 MOPP
N/A
1800
4000
500
-
4
MINT1150
150W Single Output
Medical Series
Input Specifications
All specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted.
Parameter
Conditions/Description
Input Voltage
Min
Nom
Max
90
115/230
264
Units
Vac
Turn-On Input Voltage
Ramping up
82.7
Vac
Turn-Off Input Voltage
Ramping down
67.0
Vac
47
50/60
63
Hz
Inrush Current Limitation
Input Frequency
264Vac, cold start
-
-
50
A
Power Factor
Vi nom, Io nom
0.9
-
-
Efficiency
Vi nom, Io nom
MINT1150A1206K01
MINT1150A2406K01
MINT1150A4806K01
MINT1150A5606K01
-
89%
-
%
Parameter
Conditions/Description
Min
Nom
Max
Units
Input Fuse
Not user accessible
Input Transient Protection
2KV(CM) and 1KV(DM) surge
2
KV (CM)
No-load
Normal
Output
Short Circuit
Hiccup Mode
Overload
Hiccup Mode
Overvoltage Protection
Latch style
See Models chart for trip ranges
Over temperature Protection
Automatic power shutdown at TC =155°C
Protection
Characteristic Curves
Output vs. Temperature
100W convection cooled and 150W continuous with 200 LFM airflow, Derating output power to 50% at 70C.
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
5
MINT1150
150W Single Output
Medical Series
Efficiency vs. Loading
The high efficiency is achieved by using LLC technology, PFC topology minimizing switching losses. Synchronous MOSFET or SCHOTTKY diode is
used as rectifier in MINT1150 family.
Ripple & Noise
To verify that the output ripple and noise does not exceed the level specified in the product specification, measured using a scope probe socket with
0.1uF ceramic and a 10uF electrolytic capacitor connected in parallel across it, 20MHz BW.
www.slpower.com
24V OUT, NO LOAD, 90VAC, 60HZ
24V OUT, FULL LOAD, 90VAC, 60HZ
24V OUT, NO LOAD, 264VAC, 50HZ
24V OUT, FULL LOAD, 264VAC, 50HZ
MINT1150 REV 1.4, 13-NOV-12
6
MINT1150
150W Single Output
Medical Series
Output Transient Response
50% load step within the regulation limits of minimum and maximum load, dI/dt< 0.2A/μSec. Recovery time not specified as there is no laps in regulation
with a 50% Load Step. Maximum voltage deviation is 3%, This test is performed on the MAIN OUTPUT ONLY.
24V OUT, 120VAC, 25% TO 75% LOAD STEP
24V OUT, 240VAC, 25% TO 75% LOAD STEP
Output Overload Characteristic
Supply shall protect itself against Overload conditions. The Power Supply shall recover from Overload Conditions without operator intervention.
24V OUT, 90VAC
24V OUT, 264VAC
Short Circuit Protection
Supply shall protect itself against Short Circuit conditions. No damage will occur if the output is shorted.
24V OUT, 90VAC
www.slpower.com
24V OUT, 264VAC
MINT1150 REV 1.4, 13-NOV-12
7
MINT1150
150W Single Output
Medical Series
Overvoltage Protection
OVP firing reduces output voltage to <50% of nominal in <50ms. See models chart for trip ranges.
24V OUT, NO LOAD
Turn On Time
90VAC, FULL LOAD
264VAC, FULL LOAD
Hold Up Time
120VAC, FULL LOAD
Data Sheet © 2012 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
8