The MBC60 Series of open-frame medical power supplies, with its wide universal 90-264 VAC input range, is available at 60 W of output power and a variety of single and multiple output voltages. The MBC series was designed to 4th edition medical approvals and provides 2x MOPP (Means of Patient Protection) isolation for Class I and Class II installations. These medical power supplies are ideal for, monitoring, home health equipment as well as surgical devices. • 50 - 65 W Single to Triple Outputs • Ultra High Efficiency > 85% • • Form Factor 4 x 2 x 1.2 inches (101.6 x 50.8 x 30.48 mm) Low leakage current < 250 μA • No Load Power < 0.3 W • Approved to EN/IEC 60601-1 • 2x MOPP • IEC Protection Class Options: o Class I: Earth pin J4 (no suffix) o Class II: No Earth pin (-2 suffix) • Conducted EMI EN 55022-B, FCC Part 15 Level B • Medical Safety Agency Approvals • RoHS Compliant • Cover Kit Accessory Available • Dialysis • Surgical Devices • Monitoring • Home Health • Pumps • Ultrasound MBC60 Series 2 OUTPUT VOLTAGE (VDC) 2 OUTPUT CURRENT MAX (A) MINIMUM LOAD (A) 3 RIPPLE & NOISE 4 TOTAL REGULATION MBC60-1005G 5.2 10.0 0.0 1.25% ± 0.8% MBC60-1012G 12 5.4 0.0 1% ± 0.8% MBC60-1015G 15 4.33 0.0 1% ± 0.8% MBC60-1024G 24 2.7 0.0 1% ± 0.8% MODEL1 MBC60-1048G MBC60-3000G MBC60-3001G MBC60-3002G MBC60-3003G Cover-60-XCB 5 48 1.35 0.0 1% ± 0.8% 5.2 8.0 0.5 1.25% ± 0.8% 12.5 3.0 0.1 1% ± 5.3% -12.5 0.5 0.0 1% ± 5.3% 5.2 8.0 0.5 1.25% ± 0.8% 23.8 1.5 0.1 1% ± 5.3% -12.5 0.5 0.0 1% ± 5.3% 5.2 8.0 0.5 1.25% ± 0.8% 14.6 2.5 0.1 1% ± 5.3% -16.2 0.5 0.0 1% ± 5.3% 3.3 6.0 1.0 1.5% ± 0.8% 5.2 3.0 0.1 1% ± 5.3% -12.8 0.5 0.0 1% ± 5.3% Metal cover kit accessory Specifications are for nominal input voltage, 25°C unless otherwise stated. PARAMETER DESCRIPTION / CONDITION SPECIFICATION Input Voltage Universal 90 - 264 VAC Input Frequency6 Input Current No Load Power Inrush Current Leakage Current Switching Frequency 47 to 400 Hz 1.5 A max. 0.75 A max. < 0.3 W < 0.5 W 30 A max. 60 A max. < 140 µA <250 µA 120 VAC: 230 VAC: Single output models Multi output models 120 VAC: 230 VAC: 120 VAC: 230 VAC: Typical 67 67 Hz 1 Single output models deliver 65 W, except MBC60-1005G (50 W). Triple output models deliver 60 W, except MBC60-3003G (45 W). Maximum outputs for each output. Max power rating should not be exceeded 3 Minimum load specified to meet cross regulation. 4 Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Tantalum capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges. 5 When used in Cover Kit, de-rate output power to 70 % under all operating conditions. 6 Safety Approved: 47 to 63 Hz 2 [email protected] MBC60 Series PARAMETER 3 DESCRIPTION / CONDITION SPECIFICATION Efficiency Typical 85% Hold Up Time @ 120 VAC typical >10 ms Output Power Derate output power linearly to 80% from 90 to 80 VAC input. 50 - 65 W 7 Line Regulation Output Voltage Adjustment V1: V2 & V3: V1 +/-0.3% +/-0.5% +/-5% < 10%, recovery time < 5 ms < 100 ms ± 3% ± 5% ± 10% Over Current Protection Typical above rating 130% Over Voltage Protection Typical for V1 only 130% Short Circuit Protection Short term, autorecovery PARAMETER DESCRIPTION / CONDITION SPECIFICATION Operating Temperature Refer to derating curve, Fig. 1 Start-up is guaranteed -20 to 70°C -20 to 0°C Load Regulation V1: V2 & V3: Transient Response 50% to 100% load change, 50/60 Hz, 50% duty cycle, 0.1 A/µs Rise Time Set Point Tolerance Storage Temperature -40 to +85°C Relative Humidity Non Condensing 95% Altitude Operating: Non-Operating: 10,000 ft. 40,000 ft. Reliability MTBF according to Telcordia -SR332-Issue 3 1.87 million hours Cooling Convection Figure 1. Derating Curve De-rate linearly from 100% at 50°C to 50% at 70°C 7 For MBC60-3003G efficiency is 75% typical. Europe, Middle East +353 61 225 977 North America +1 408 785 5200 © 2017 Bel Power Solutions & Protection BCD.00287_AE Asia-Pacific +86 755 298 85888 MBC60 Series 4 Conducted Emissions EN 55011-B, CISPR22-B, FCC PART15-B Pass Radiated Emissions EN 55011 Pass Input Current Harmonics EN 61000-3-2 Class D Voltage Fluctuation and Flicker EN 61000-3-3 Pass ESD Immunity EN 61000-4-2 Level 4, Criterion A Radiated Field Immunity EN 61000-4-3 Level 3, Criterion A Electrical Fast Transient Immunity EN 61000-4-4 Level 3, Criterion A Surge Immunity EN 61000-4-5 Level 3, Criterion A Conducted Immunity EN 61000-4-6 Level 3, Criterion A Magnetic Field Immunity EN 61000-4-8 Level 4, Criterion A Voltage Dips, Interruptions EN 61000-4-11 Criterion A & B PARAMETER Isolation Voltage Safety Standards DESCRIPTION / CONDITION SPECIFICATION Input to Output, 2M0PP Input to Ground, 1M0PP Output to Ground Approved to the latest edition of the following standards: CSA/UL60601-1, EN60601-1 and IEC60601-1. Min. 5700 VDC 1500 VAC 500 VAC Agency Approvals Nemko, UL, C-UL CE mark Complies with LVD Directive CONNECTOR PIN AC Input Connector J1 DC Output Connector J2 Signal Connector J3 Earth J4 DESCRIPTION / CONDITION Pin 1 Pin 2 Pin 1,2 Pin 3,4 Pin 5 Pin 6 Pin 1 Pin 2 AC Neutral AC Line V1 RTN V3 V2 +V1 Sense -V1 Sense MANUFACTURER / PN Molex: 26-60-4030 or equivalent Mating: 09-50-3031; Pins: 08-50-0106 Tyco: 640445-6 or equivalent Mating: 647402-6; Pins: 3-647409-1 Molex: 22-23-2021 or equivalent Mating: 22-01-2021 Molex: 19705-4301 Mating: 190030001 PARAMETER DESCRIPTION / CONDITION Weight 150 g (0.33 lbs.) Dimensions 101.6 x 50.8 x 30.48 mm (4 x 2 x 1.2 inch) [email protected] MBC60 Series 5 Figure 2. Mechanical Drawing MBC60-1xxxG Figure 3. Mechanical Drawing MBC60-3xxxG NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following: 1 Stand off, used to mount PCB has OD of 5.4 mm max. 2 Screws, used to fix PCB on stand off, have head dia of 6.0 mm max. 3 Washer, if used, to have dia of 6.5 mm max. NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems. TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. Specifications are subject to change without notice. Europe, Middle East +353 61 225 977 North America +1 408 785 5200 © 2017 Bel Power Solutions & Protection BCD.00287_AE Asia-Pacific +86 755 298 85888