Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA MDS-400APB12 AA Highlights & Features – 3" x 5" x 1.4" Package –Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1 –Compliant with IEC 60601-1-2 4th Ed. Requirements –Up to 200W convection, 400W force air –IT & Medical Safety Approvals Safety Standards CB Certified for worldwide use Model Number: MDS-400APB12 AA Unit Weight: 405g Dimensions (W x L x H): 76.2x127x35.6mm (3x5x1.4inch) General Description 3 Delta’s new MDS-400APB12 open frame design offers a high energy density (19 Watts/in ) 3x5 inch design for Type BF patient access medical products. With operating parameters of 90 to 264 Vac universal input voltage, temperatures of -10 to +70 degrees centigrade, and altitudes of up to 5000 meters (16,400 feet), the design is well suited for a variety of both medical and non-medical applications. Other features include input surge of 300 Vac, low leakage current, no-load input power < 0.5 Watt, and 500K hour MTBF. This product is certified for EMC standards EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment, and EN 55022 for Industrial Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used to ensure long product life. An enhanced feature set, includes a 5 Volt/2 amp standby output, remote on/off, remote sense, and a power good signal that are included as part of the standard product. The design, which has protection against shock compliant with 2XMOPP and Type BF requirements, has both medical (with risk report available), and ITE safety approvals, including cURus (US&Canada) /TUV/GB(China); plus, CB certificates with all national deviations. Full compliance with RoHS Directive 2011/65/EU for environmental protection is included. The unit comes in a standard 3x5 inch footprint, with height suitable for 1U applications. Robust terminals are capable of multiple connections of the main output to the end product. Model Information Medical AC-DC Open Frame Model Number Main Output Main Output Standby Voltage Current Output Voltage Standby Output Current Fan Output Voltage MDS-400APB12 AA 12Vdc 0-2A 12Vdc 0-33.33A 5Vdc Total Max Output Power 2) 1) 400W 1) Fan output is direct from 12Vdc main output, 2) With 16CFM force air Model Numbering MDS Delta Medical power Supply 1 400 Max wattage in the product Series. 400 →400W A P B 12 Family Code Product Type Input Type Code Output Voltage A~ Z P: Open Frame B: 3pin ClassI 12 for12V All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) A Revision Code A Revision code Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Specifications Input Ratings / Characteristics Nominal Input Voltage 100-240Vac Input Voltage Range 90-264Vac Nominal Input Frequency 50-60Hz Input Frequency Range 47-63Hz Input Current(max) 5.5A Input Surge Voltage (max) 300Vac for 100ms Full load Efficiency (typ.) 89.5% @ 115Vac/60Hz 91.5% @ 230Vac/50Hz, Reference Fig.1 Standby Power (max) 0.5W (only standby working with Inhibit signal high) @ 115Vac/60Hz, 230Vac/50Hz Inrush Current (max) 40A@230Vac, cold start Input-PE(protective earth) leakage current(max) 0.1mA @ NC, 0.3mA @ SFC1) Output-PE(protective earth) leakage current for Type BF application (max) 0.1mA @ NC, 0.5mA @ SFC 1) Power Factor (min) 0.95 @ 115V/50Hz, 230V/50Hz, full load 1) NC: normal condition, SFC: single fault condition Leakage Current Input-PE Leakage Current 100Vac/60Hz(Typ) 264Vac/60Hz(Typ) Delta Limit IEC60601-1 Limit Normal Condition 17.5uA 43.5uA 100uA max 5000uA max Single Fault Condition 32.9uA 90.7uA 300uA max 10000uA max 86.7uA 100uA max 100uA max 128.6uA 500uA max 500uA max Output-PE Leakage Current for Type BF application Normal Condition 28.5uA Single Fault Condition 42.9uA 100.00 90.00 Efficiency (%) 80.00 115Vac 70.00 230Vac 60.00 50.00 25% 50% Output Load (400W) Fig.1 Efficiency versus output load 2 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) 75% 100% Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Output Ratings / Characteristics Nominal Output Voltage (Vrated) 12V Output Voltage Tolerance ±3% Output Power 400W max with 16CFM air cooling Line Regulation (max) ±0.5% Load Regulation (max) ±1% Ripple& Noise (typ.) 150mVpk-pkVrated@ Full load, Reference Fig. 2 Start-up Time (max) 2000ms@115Vac Hold-up Time (min) 10ms@115Vac, with 400W load Dynamic Response (Overshoot & Undershoot O/P Voltage) ±5% @50-100% load Capacitive load (max) 3000uF Rise time (max) 100ms Remote Sense Compensate up to 500mV lead drop with remote sense Short and reverse connection protected. PSU can work normally with remote sense pins left open. (a) 115V (measured value=97.6mV) (b) 230V(measured value=104mV) Fig.2 Ripple & Noise example, 20MHz BW Ripple & Noise measurement circuit PSU Load 5cm 0.47uF Ceramic 10uF Electrolytic Test Fixture 3 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) Scope Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Mechanical Case Cover NA Dimensions(W x L x H typ.) 76.2x127x35.6mm (3x5x1.4inch) Unit Weight 405g Indicator NA Cooling System TBD Environment Surrounding Air Temperature Operating Storage Absolute Max -10°C to+70°C, supported power linearly de-rate from 50°C to 50% rated up to 70°C for forced air. Convection power de-rate from 200W @ 40°C to 100W @ 60°C Note: see power de-rating curve -40°C to+85°C Operating Humidity 5-95% RH (Non-Condensing) Operating Altitude 5,000 meters (16402 fts) Shock Test (Non-Operating) 50G, 11ms, 3 shocks for each direction Vibration (Operating) 5-500Hz, 2Grms, 15 minute for each three axis 4 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Power De-rating curve Output Power 400W 16 CFM forced air 200W Convection 100W -10°C 5 40°C 50°C All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) 60°C Ambient Temperature 70°C Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Protections Overvoltage (max) 135% of rated voltage, Latch Mode Overload / Overcurrent (max) Main output 160% of rated current Standby 3A max Hiccup Mode(Non-Latching, Auto-Recovery) Over Temperature Latch Mode Short Circuit Hiccup Mode, (Non-Latching, Auto-Recovery) Reliability MTBF(Minimum) at 115Vac, 400W, 35°C, 16CFM Air Flow C i A Operating life at 115Vac, 400W, ambient 25 °C,16CFM Air Flow 500 kHrs based on Telecordia SR-332 26,280Hrs Safety Standards / Directives nd rd rd Medical Safety IEC60601-12 and 3 , and 3 +A1 edition CB report TUV EN60601-1:2006 UL60601-1+CAN/CSA 60601-1: (Ed.3.2005) ITE Safety IEC60950-1 CB report TUV60950-1 UL60950-1+CAN/CSA60950-1 GB4943.1-2011, GB9254-2008, GB17625.1-2003 CE MDD Directive 93/42/EEC Environmental RoHS Directive 2011/65/EU Compliant Galvanic Isolation 1) 6 Input to/Output (2XMOPP) 4000 Vac 1) Input to/Ground(1XMOPP) 1500Vac Output to/Ground(1XMOPP) 1500Vac ( Type BF application rated) PSU can support PoE applications with Primary to FG 2500Vac test. All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements) EMC / Emissions Harmonic Current Emissions EN55011/EN55022,FCC Title 47:Class B IEC61000-3-2 Meet Class D limit Immunity to Voltage Flicker IEC61000-3-3 Electrostatic Discharge IEC61000-4-2 Level 4 Criteria A Air Discharge: 15kV Contact Discharge: 8kV Radiated Field IEC61000-4-3 Electrical Fast Transient / Burst IEC61000-4-4 Criteria A 80MHz-2700MHz, 10V/m AM modulation 385MHz-5785MHz, 28V/m Pulse mode and other modulation 1) Level 3 Criteria A :2kV Surge IEC61000-4-5 Level 3 Criteria A 3) Common Mode : 2kV 4) Differential Mode : 1kV Conducted IEC61000-4-6 Level 2 Criteria A 150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and 1) 1) 1) 1) Amateur radio bands Power Frequency Magnetic Fields Voltage Dips IEC61000-4-8 IEC61000-4-11 1) Criteria A Magnetic field strength 30A/m 1) Criteria A 0% UT, 0.5 cycle (10ms) o o o o o o o o o , 0 /45 /90 /135 /180 /225 /270 /315 /360 2) Criteria B o 0% UT, 1 cycle (20ms), 0 2) Criteria B o 70% UT, 25 cycle (500ms), 0 2) Criteria B o 0% UT, 250 cycle (5000ms), 0 1) Criteria A: Normal performance within the specification limits 2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test. 3) Asymmetrical: Common mode (Line to earth) 4) Symmetrical: Differential mode (Line to line) 7 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Dimensions W x L x H: 76.2 x 127.0 x 35.6 mm Suggested air flow direction, draw air from input side to output side Mechanical drawing (3Pin input type) Notes ―Dimensions are in mm(inches) Input Connector CN1 Pin 1 Neutral Pin 2 Line CN1 mates with Molex housing 26033031 and Molex series 6838 crimp terminals. Output Connector CN102 Vo CN103 DC RTN Fan Connector CN104 Pin1 12V Fan Pin2 DC RTN Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Pin 8 Pin 9 Pin 10 Control and STANDBY connector CN101 Remote sense + Remote sense – Power Good + Power Good -(DC RTN) Remote ON_OFF/INHIBIT + Remote ON_OFF/INHIBIT -(DC RTN) 5V Standby DC RTN 5V Standby DC RTN CN101 mates with Molex housing 1041421000 and Molex series 104539 crimp terminals. CN104 mates with JST housing PHR-2 and JST SPH-002T-P0.5S terminals. Two mounting points in mechanical drawing need to be connected to system earth case together, Protective bonding conductor from the end product protective earth terminal (if any) can be tied to CN2 for open frame model. 8 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Functions Start-up Time The time required for the output voltage to reach 90% of its set value, after the input voltage is applied. Rise Time The time required for the output voltage to change from 10% to 90% of its set value. Hold-up Time Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time required for the output to reach 90% of its set value, after the input voltage is removed. ■Graph illustrating the Start-up Time, Rise Time, and Hold-up Time 9 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Dynamic Response Short Circuit Protection The power supply output voltage will remain within ±5% of its steady state value, when subjected to a dynamic load 50 to 100% of its rated current. ■ 50 to 100% Load The power supply’s output OLP/OCP function also provides protection against short circuits. When a short circuit is applied, the output current will operate in “Hiccup mode”, as shown in the illustration in the OLP/OCP section on this page. The power supply will return to normal operation after the short circuit is removed. Inrush Current Inrush current is the peak, instantaneous, input current measured and, occurs when the input voltage is first applied. For AC input voltages, the maximum peak value of inrush current will occur during the first half cycle of the applied AC voltage. This peak value decreases exponentially during subsequent cycles of AC voltage. Overload & Overcurrent Protections The power supply’s Overload (OLP) and Overcurrent (OCP) Protections will be activated before output current under 160% of I O (Max load). Upon such occurrence, V O will start to drop. Once the power supply has reached its maximum power limit, the protection will be activated and the power supply will go into “Hiccup mode” (Auto-Recovery). The power supply will recover once the fault condition causing the OLP and OCP is removed and I O is back within the specified limit. Additionally, if the Iout is <160% but >100% for a prolong period of time (depending on the load), the Over Temperature Protection (OTP) will be activated due to high temperature on critical components. The power supply will then go into latchmode. Over Temperature Protection Overvoltage Protection The power supply’s overvoltage circuit will be activated when its internal feedback circuit fails. The output voltage shall not exceed its specifications defined on Page 4 under “Protections”. Power supply will latch off, and require removal/re-application of input AC voltage in order to restart. 10 As mentioned above, the power supply also has Over Temperature Protection (OTP). This is activated when the overload condition persists for an extended duration and the output current is below the overload trigger point but >100% load. In the event of a higher operating temperature condition at 100% load, the power supply will run into OTP when the surrounding air temperature is higher than the operating temperature. When activated, the output voltage will go into latch mode until the input voltage is removed; then, reapplied, and the surrounding air temperature drops to its normal operating temperature. All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Power Good Power Good+ pin is an open collector transistor (40V/600mA rating). A resistor (suggested value 10Kohm, 1/8W) can be added between 5V STANDBY pin (or, other available pull-up voltage that is no greater than 30V) and the Power Good+ pin (refer to figure below). Value of pull-up resistor may have to be adjusted, depending on voltage used, and other end-use conditions of the Power Good+ pin connection to the product. When AC input is on, Power Good+ pin will be high. When AC input is off, Power Good+ pin will be low. There will be a minimum of 3 milliseconds between the time the power good goes to low level, and the time when the output reaches 90% of its rated value. 11 Remote On_Off/Inhibit Remote ON_OFF/INHIBIT can be used to enable or disable only the main output. When the main output is disabled, the +5V Standby output will continue to operate. This signal can be pulled down to a low level of 0.3 volts, or shorted to DC-Return, in order for the main output to be enabled; and, floated (no connection to the signal), or pulled up to a value greater than or equal to 3 volts, in order to disable the main output. Remote Sense Remote sense feature can be used to compensate for the extra voltage drop on output wires that are connected from the main output terminals, to the load. With wires connected from the remote sense pins, at the same locations as the wires from the main output, the remote sense function can compensate up to 500mV voltage drop. The power supply will not be damaged if the remote sense pins are shorted, or if a reverse/inverted polarity connection is made to the load. All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08) Medical AC-DC Open Frame Power Supply 12V/400W 3”x5” High Density MDS-400APB12 AA Certificate All Delta Medical Power products conform to the European directive 2011/65/EU. RoHS is the abbreviation for “Restriction of the use of certain hazardous substances Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN ISO 13485:2012 for the design and manufacture of switching power supply and adaptor for medical device. In addition to a UL Total Certification Program (TCP) approved client laboratory for IEC60950 and IEC60065. Delta also has participated UL Client Test Data Program (CDTP) for IEC 60601 12 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2016, Rev. 08)