Specifications
Specifications describe the performance of parameters covered by the product warranty. These values are only valid for the stated operating frequency, and apply over 0°C
to +55°C unless otherwise noted.
95th percentile values indicate the breadth of the population (approx. 2 s) of performance tolerances expected to be met in 95 percent of the cases with a 95 percent
confidence, for any ambient temperature in the range of +20°C to +30°C. In addition to
the statistical observations of a sample of instruments, these values include the effects
of the uncertainties of external calibration references. These values are not warranted.
These values are updated occasionally if a significant change in the statistically observed
behavior of production instruments is observed.
Typical describes additional product performance information that is not covered by the
product warranty. It is performance beyond specifications that 80 percent of the units
exhibit with a 95 percent confidence level over the temperature range of +20°C to +30°C.
Typical performance does not include measurement uncertainty.
Nominal values indicate expected performance, or describe product performance that is
useful in the application of the product, but are not covered by the product warranty.
The analyzer will meet its specifications when:
–– It is within its calibration cycle
–– Under auto couple control, except when Auto Sweep Time Rules = Accy
–– Signals measured <10 MHz have DC coupling applied
–– The analyzer has been stored at an ambient temperature within the allowed operating range for at least two hours before being turned on; if it had previously been
stored at a temperature range inside the allowed storage range, but outside the
allowed operating range
–– The analyzer has been turned on at least 30 minutes with Auto Align set to normal,
or, if Auto Align is set to off or partial, alignments must have been run recently
enough to prevent an Alert message; if the Alert condition is changed from Time and
Temperature to one of the disabled duration choices, the analyzer may fail to meet
specifications without informing the user
For the complete specifications guide, visit:
www.keysight.com/find/NFA_X-Series_specifications
03 | Keysight | N8973B, N8974B, N8975B, N8976B NFA X-Series Noise Figure Analyzer, Multi-touch - Data Sheet
Frequency
Frequency range
N8973B
10 MHz to 3.6 GHz
N8974B
10 MHz to 7.0 GHz
N8975B
N8976B
10 MHz to 26.5 GHz
4
10 MHz to 40.0 GHz
Measurement bandwidth (nominal)
N8973B, N8974B, N8975B, N8976B 5
1 Hz to 3 MHz (in E24 series increments 1), 4 MHz, 5 MHz, 6 MHz, 8 MHz
Frequency reference
Accuracy
± [RΔt + T + C]
Aging rate
± 0.1 ppm 2 /year
± 0.15 ppm/2 years
Temperature stability
+20°C to +30°C
Full temperature range
± 0.015 ppm
± 0.05 ppm
Achievable initial calibration accuracy
± 0.04 ppm
Example frequency reference accuracy,
Residual FM </= (use less than or equal to symbol)
(0.25 Hz x N) p-p in 20 ms nominal
1 year since last adjustment
= ± (1 x 10 -7 + 5 x 10 -8 + 4 x 10 -8)
Frequency readout accuracy (start, stop, center,
marker)
± (marker frequency x frequency reference accuracy + 0.25% x span + 5% x RBW + 2 Hz + 0.5 x
horizontal resolution 3)
1.
2.
3.
4.
5.
R = aging rate
Δt = time since last adjustment
T = temperature stability
C = calibration accuracy
= ± 0 0.19 ppm
The E24 series is defined by international standard IEC 60063. E24 is a preferred series of numbers, with each number being approximately 10% larger than
the previous number. It is commonly used for the labeling of 5% tolerance resistors, capacitors, etc.
Parts per million (10-6)
Horizontal resolution is span/(sweep points – 1).
The N8976B ships with 346CK40. The 346CK40 has superior match above 26 GHz, which leads to better uncertainty.
IQ analyzer (basic) mode has up to 25 MHz analysis BW.
Note: The NFA X-Series noise figure analyzer is more than a dedicated noise figure
analyzer. Each model has full featured spectrum analyzer and IQ analyzer (basic) modes.
The analyzer is specified to 44 GHz when in SA or IQ analyzer mode.
04 | Keysight | N8973B, N8974B, N8975B, N8976B NFA X-Series Noise Figure Analyzer, Multi-touch - Data Sheet
Noise figure, gain, and uncertainty
Example DUT uncertainties 1
Without a Preamp
With a USB Preamp 2
When combined with the U7227A/C/F preamp, the NFA X-Series noise figure analyzer
offers improved uncertainty over the previous NFA-A in all of the above hypothetical
cases.
1.
2.
These uncertainties assume a measurement made with a N8975B at 1 GHz with a N4000A noise
source and a non-frequency-converting DUT. The DUT is assumed to have an input/output match of
1.5 VSWR.
Assuming a U7227A/C/F External USB Preamp
05 | Keysight | N8973B, N8974B, N8975B, N8976B NFA X-Series Noise Figure Analyzer, Multi-touch - Data Sheet
Description
Specifications
Supplemental information
Noise figure
‹ 10 MHz
10 MHz to internal preamplifier’s
frequency limit 1
Uncertainty calculator 1
See note 1
Internal and external preamplification recommended 1
Noise source ENR
Measurement range
Instrument uncertainty
4 to 6.5 dB
0 to 20 dB
± 0.02 dB
12 to 17 dB
0 to 30 dB
± 0.025 dB
20 to 22 dB
0 to 35 dB
± 0.03 dB
Gain
Instrument uncertainty 1
DUT Gain range = –20 to +40 dB
‹ 10 MHz
See note 1
10 MHz to 3.6 GHz
± 0.15 dB
› 3.6 GHz
± 0.11 dB additional 1 95th percentile, 5 minutes after
calibration
Noise Figure uncertainty calculator1, 2
Instrument noise figure uncertainty
Instrument gain uncertainty
Instrument noise figure
See the noise figure table above
See the gain table above
See graphs of “Nominal instrument noise figure”; noise
figure is DANL + 176.24 dB (nominal) 1. Note on DC coupling 1
See graph: nominal VSWR. Note on DC coupling 1
See “Displayed average noise level (DANL) (with noise floor
extension) improvement” in the Option NFE - Noise floor
extension chapter.
Instrument input match
Optional NFE improvement/internal Cal 1
Uncertainty versus calibration options1
User calibration
Uncalibrated
Internal calibration
Best uncertainties; noise figure uncertainties calculator applies
Worst uncertainties; noise of the analyzer input acts as a second stage noise on the DUT
Available with Option NFE. Good uncertainties without the need of reconnecting the DUT and running a
clibration. The uncertainty of the analyzer input noise model adds a second-stage noise power to the DUT
that can be positive or negative. Running the noise figure uncertainty calculator will usually show that internal
calibration achieves 90% of the possible improvement between the uncalibrated and user calibration states.
1. Refer to NFA X-Series specifications guide for footnote details
2. Online uncertainty calculator at www.keysight.com/find/nfuc or use the instrument’s built-in calculator.
Nominal instrument noise figure, N8973B, N8974B, N8975B
06 | Keysight | N8973B, N8974B, N8975B, N8976B NFA X-Series Noise Figure Analyzer, Multi-touch - Data Sheet
Internal Preamp noise figure
Frequency
Noise Figure (nominal)
100 kHz to 3.6 GHz
8 dB + (0.001112 * freq in MHz) nominal
3.6 GHz to 8.4 GHz
9 dB nominal
8.4 GHz to 13.6 GHz
10 dB nominal
> 13.6 GHz
DANL + 176.24 dB nominal
Measurement uncertainty is usually dominated by the uncertainty of the noise source,
meaning that the instrument’s noise figure is negligible for most measurements. For
situations when this noise figure becomes non-negligible (i.e. low-gain, low-noise DUTs),
the included U7227 Series USB preamp provides extra measurement reliability.
DANL (N8973B, N8974B, N8975B) 1
Frequency
Specification
Typical
10.0 MHz to 2.1 GHz
–161 dBm
–163 dBm
2.1 GHz to 7.0 GHz
–160 dBm
–162 dBm
7.0 GHz to 13.6 GHz
–160 dBm
–163 dBm
13.5 GHz to 17.1 GHz
–157 dBm
–160 dBm
17.0 GHz to 20.0 GHz
–155 dBm
–159 dBm
20.0 GHz to 26.5 GHz
–150 dBm
–156 dBm
Frequency
Specification
Typical
10.0 MHz to 1.2 GHz
–164 dBm
–165 dBm
1.2 GHz to 2.1 GHz
–163 dBm
–164 dBm
2.1 GHz to 3.6 GHz
–162 dBm
–163 dBm
3.5 GHz to 20.0 GHz
–160 dBm
–162 dBm
20.0 GHz to 26.5 GHz
–158 dBm
–160 dBm
26.4 GHz to 34.0 GHz
–156 dBm
–159 dBm
33.9 GHz to 40.0 GHz
–153 dBm
–155 dBm
DANL (N8976B) 1
Preamp noise figure and gain 2
Specification
U7227A
U7227C
U7227F
Frequency
10 MHz to 4 GHz
100 MHz to 26.5 GHz
2 GHz to 50 GHz
Noise figure
10 MHz to 100 MHz: < 5.5 dB
100 MHz to 4 GHz: < 5 dB
100 MHz to 4 GHz: < 6 dB
4 GHz to 6 GHz: < 5 dB
6 GHz to 18 GHz: < 4 dB
18 GHz to 26.5 GHz: < 5 dB
Gain
10 to 100 MHz: > 16 dB
100 MHz to 4 GHz: > 17 + 0.5F dB
100 MHz to 26.5 GHz: > 16.1 + 0.26F dB
Averaging
Up to 10,000 measurement results
* “F” signifies frequency in GHz
1.
2.
Preamp on, input terminated, sample or average detector, log averaging, 0 dB input attenuation,
IF Gain = High, +20°C to +30°C.
See U7227A/C/F Data Sheet for list of specifications
2 GHz to 50 GHz: > 16.5 + 0.23F dB
07 | Keysight | N8973B, N8974B, N8975B, N8976B NFA X-Series Noise Figure Analyzer, Multi-touch - Data Sheet
RF input
Connector
N8973B, N8974B, N8975B
Type-N female, 50 Ω nominal
N8976B
2.4 mm male, 50 Ω nominal
Input VSWR
Input VSWR
N8973B, N8974B, N8975B
N8976B
10 MHz to 3.6 GHz
< 1.2:1 nominal
1.2:1 nominal
3.6 GHz to 26.5 GHz
< 1.9:1 nominal
1.5:1 nominal
26.5 GHz to 44.0 GHz
N/A
< 1.8:1 nominal
Measurement
Sweep
Number of points setting
2 to 501, or fixed frequency
Start/stop, center/span, Frequency list of up to 501 points
Sweep trigger
Continuous or single
Measurement speed
Local measurement and display update rate
11 ms (90/s)
Remote measurement and LAN transfer rate
6 ms (167/s)
Marker peak search
5 ms
Center frequency tune and transfer (RF)
22 ms
Center frequency tune and transfer (µW)
49 ms
Measurement/mode switching
75 ms
DUT profiles available
Amplifier
Includes any non-frequency-converting device (e.g. amplifiers, attenuators, filters, etc)
Downconverting DUT
With fixed or variable IF.
Instrument capable of controlling an external LO via GPIB, LAN, or USB
Upconverting DUT
With fixed or variable IF.
Instrument capable of controlling an external LO via GPIB, LAN, or USB
System downconverter
Allows the use of an external downconverting mixer as part of the measurement system.
Instrument capable of controlling an external LO via GPIB, LAN, or USB
Measurement
Display type and N.F. results
Type
4U multitouch
Output format
Graphical, table of values, or meter mode
Display channels
2
Number of markers
4
Limit lines
Upper and lower for each of 2 channels
Noise figure
Noise figure (F dB), or as a ratio (F)
Gain
Gain (G dB)
Y-factor
Y-factor (Y dB)
T effective
Effective noise temperature in Kelvin
P hot
Relative power density in dB
P cold
Relative power density in dB
08 | Keysight | N8973B, N8974B, N8975B, N8976B NFA X-Series Noise Figure Analyzer, Multi-touch - Data Sheet
Front panel
Sweep
Probe power
Voltage/current
+15 Vdc ± 7 % at 150 mA max nominal
-12.6 Vdc ± 10 % at 150 mA max nominal
USB 2.0 ports
Master (2 ports)
Standard
Connector
Output current
Compatible with USB 2.0
USB Type-A female
0.5 A nominal
Master (1 port)
High power
Connector
Output current
Compatible with USB 2.0
USB Type-A Female
1.0A nominal
Rear panel connectivity
10 MHz out
Connector
Output amplitude
Frequency
BNC female, 50 Ω nominal
≥ 0 dBm nominal
10 MHz ± (10 MHz x frequency reference accuracy)
Ext ref in
Connector
Input amplitude range
Input frequency
Frequency lock range
BNC female, 50 Ω nominal
–5 to 10 dBm nominal
10 MHz nominal
± 5 x 10 -6 of specified external reference input frequency
Trigger 1 and 2 inputs
Connector
Impedance
Trigger level range
BNC female
> 10 kΩ nominal
–5 to 5 V
Trigger 1 and 2 inputs
Connector
Impedance
Trigger level range
BNC female
50 Ω nominal
5 V TTL nominal
Monitor output
Connector
Format
Resolution
VGA compatible, 15-pin mini D-SUB
XGA (60 Hz vertical sync rates, non-interlaced) analog RGB
1024 x 768
Rear panel
Noise source drive +28 V (pulsed)
Connector
BNC female
SNS Series noise source connector
For use with Keysight SNS Series noise sources
USB 2.0 ports
Master (4 ports)
Standard
Connector
Oupt current
Compatible with USB 2.0
USB Type-A female
0.5 A nominal
Slave (1 port)
Standard
Connector
Output current
Compatible with USB 2.0
USB Type-B female
0.5 A nominal
GPIB interface
Connector
GPIB codes
GPIB mode
IEEE-488 bus connector
SH1, AH1, T6, SR1, LR1, PP0, DC1, C1, C2, C3, C28,DT1, L4, C0
Controller or device
LAN TCP/IP interface
Standard
Connector
1000 Base-T
RJ45 Ethertwist
09 | Keysight | N8973B, N8974B, N8975B, N8976B NFA X-Series Noise Figure Analyzer, Multi-touch - Data Sheet
General Specifications
Temperature range
Operating
0 to 55°C
Storage
–40 to 70°C
EMC
Complies with European EMC Directive 2004/108/EC
–– IEC/EN 61326-1 or IEC/EN 61326-2-1
–– CISPR Pub 11 Group 1, class A
–– AS/NZS CISPR 11:2002
–– ICES/NMB-001
This ISM device cmplies with Canadian ICES-001
Get appareil ISM est conforme à la norme NMB-001 du Canada
Safety
Complies with European Low Voltage Directive 2006/95/EC
–– IEC/EN 61010-1 3rd Edition
–– Canada: CSA C22.2 No. 61010-1-12
–– U.S.A.: UL 61010-1 3rd Edition
Acoustic statement (European Machinery Directive 2002/42/EC, 1.7.4.2u)
Acoustic noise emission
LpA < 70 dB
Operator position
Normal position
Per ISO 7779
Environmental stress
Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration, altitude,
and power line conditions; test method are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3.
Power requirements
Voltage and frequency
Power consumption
On
Standy
100 to 120 V, 50/60/400 Hz
220 to 240 V, 50/60 Hz
350 W maximum
20 W
Display
Resolution
1280 × 768, WXGA
Size
269 mm (10.6 in.) diagonal (nominal)
Data storage
Internal
External
≥ 160 GB nominal (removable solid-state drive)
Supports USB 2.0 compatible memory devices
Weight (without options)
Net
Shipping
16 kg (35 Ibs) nominal
28 kg (62 Ibs) nominal
Dimensions
Height
Width
Length
177 mm (7.0 in)
426 mm (16.8 in)
368 mm (14.5 in)
Warranty
The NFA noise figure analyzer is supplied with a standard 3-year warranty
Calibration cycle
The recommended calibration cycle is two years: calibration services are available through Keysight service centers
10 | Keysight | N8973B, N8974B, N8975B, N8976B NFA X-Series Noise Figure Analyzer, Multi-touch - Data Sheet
From Hewlett-Packard through Agilent to Keysight
For more than 75 years, we‘ve been helping you unlock measurement insights. Our
unique combination of hardware, software and people can help you reach your next
breakthrough. Unlocking measurement insights since 1939.
1939
THE FUTURE
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s committed to superior product quality and lower total cost of
ownership. Keysight is the only test and measurement company with a
three-year warranty standard on all instruments, worldwide. And, we provide
a one-year warranty on many accessories, calibration devices, systems and
custom products.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure your
instruments are operating to specification, so you can rely on accurate
measurements.
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
Keysight Infoline
www.keysight.com/find/service
Keysight’s insight to best in class information management. Free access to
your Keysight equipment company reports and e-library.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-01-01-16)
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
This information is subject to change without notice.