XP Power AMM90PS06 Worldwide medical approval Datasheet

ACDC
90150 Watts
AMM Series
xppower.com
•
CEC2008 & EISA2007 Compliant
•
Worldwide Medical Approvals
•
Class I and Class II Versions
•
High Power Density
•
High Efficiency
•
Operating Temperature up to +60 °C
•
3 Year Warranty
Specification
Input
General
• 90 W: 80-264 VAC, derate output power
10% <90 VAC & 15% <85 VAC,
120/150 W: 80-264 VAC, derate output
power 15% <90 VAC & 20% <85 VAC
Input Frequency
• 47-63 Hz
Input Current
• 90 W: 1.2 A at 115 VAC, 0.6 A at 230 VAC
120 W: 1.6 A at 115 VAC, 0.8 A at 230 VAC
150 W: 2 A at 115 VAC, 1.0 A at 230 VAC
Inrush Current
• 60 A max at 115 VAC, 120 A max at
230 VAC, cold start +25 °C
Power Factor
• 0.98 typical
Earth Leakage Current • 90 µA max at 115 VAC/60 Hz,
(Class I Versions)
150 µA max at 230 VAC/50 Hz
Input Protection
• 90 W: Internal T2.5 A/250 V fuse in line
and neutral lines,
120 W: Internal T3.15 A/250 V fuse in line
and neutral lines,
150 W: Internal T4 A/250 V fuse in line and
neutral lines
No Load Input Power • <0.5 W
Input Voltage
Output
Output Voltage
Output Voltage Trim
Initial Set Accuracy
Minimum Load
Start Up Delay
•
•
•
•
•
Start Up Rise Time
Hold Up Time
Line Regulation
Load Regulation
Transient Response
•
•
•
•
•
Ripple & Noise
•
Overvoltage Protection •
Overtemperature
•
Protection
Overload Protection
•
Short Circuit Protection •
Temperature
•
Coefficient
See tables
Not user-adjustable
±2%, set at 60% load
No minimum load required
90/120 W: 2 s max at 115 VAC,
150 W: 3 s max at 115 VAC
<80 ms at 115 VAC
15 ms minimum at full load & 115 VAC
0.5% maximum
See tables
4% max. deviation, recovery to within
1% in 500 µs for a 25% load change
2% max pk-pk (see note 1)
110-140% Vnom, recycle input to reset
90/150 W: Unit shuts down, auto recovery
120 W: Unit shuts down, recycle input
to reset
110-180%, auto recovery
Trip and restart (Hiccup mode)
0.04%/°C
Efficiency
Isolation
Switching Frequency
Power Density
MTBF
• 85% typical
• 4000 VAC Input to Output
1500 VAC Input to Ground (Class I only)
500 VAC Output to Ground (Class I only)
• 50-110 kHz variable
• 90 W: 3.4 W/In3, 120 W: 3.7 W/In3
150 W: 2.9 W/In3
• 150 kHrs typical to MIL-HDBK-217F
at 25 şC, GB
Environmental
Operating Temperature • 0 °C to +60 °C, derate linearly from 100%
power at +40 °C to 50% power at +60 °C
Cooling
• Convection-cooled
Operating Humidity
• 5-95% RH, non-condensing
Storage Temperature • -40 şC to +85 şC
Operating Altitude
• 3000 m
Shock
• 30 g, 10 ms on 3 axes
Vibration
• 5-100 Hz, 2.31 m/s2, 20 mins, 3 axes
EMC & Safety
Emissions
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Magnetic Field
Dips & Interruptions
Safety Approvals
• EN55011/FCC/VCCI, Class B conducted
EN55011/FCC/VCCI, Class B radiated
• EN61000-3-2, Class A
• EN61000-3-3
• EN61000-4-2, level 3 Perf Criteria A
• EN61000-4-3, level 2 Perf Criteria A
• EN61000-4-4, level 3 Perf Criteria A
• EN61000-4-5, level 3 Perf Criteria A
• EN61000-4-6, 3 V Perf Criteria A
• EN61000-4-8, 3 A/m Perf Criteria A
• EN61000-4-11
90/120 W: 70% 10 ms, 40% 100 ms,
<5% 5000 ms, Perf Criteria A, B, B (Perf
Criteria A, A, B with 70% load)
150 W: >95% 10 ms, 60% 100 ms,
<30% 500 ms, Perf Criteria A, B, B
• EN60601-1, UL60601-1,
CSA22.2 No. 601-1 per cUL
Output
Power
60 W
60 W
70 W
80 W
85 W
85 W
90 W
90 W
90 W
90 W
90 W
Output
Voltage
5.0 V
6.0 V
9.0 V
12.0 V
13.5 V
15.0 V
18.0 V
19.0 V
24.0 V
30.0 V
48.0 V
Total
Regulation(2)
5%
5%
5%
5%
5%
5%
5%
5%
5%
5%
5%
Output
Current
12.00 A
10.00 A
7.78 A
6.67 A
6.30 A
5.67 A
5.00 A
4.74 A
3.75 A
3.00 A
1.87 A
Notes
1. Ripple and noise measured at 20 MHz bandwidth with a 10 µF tantalum and 0.1
µF ceramic cap connected at the measurement point.
2. Total regulation includes initial set accuracy, line and load regulation.
3. For class II versions, add ‘ C2’ to model number e.g. AMM90PS24C2.
Model
Number(3, 4)
AMM90PS05(5)
AMM90PS06(5)
AMM90PS09(5)
AMM90PS12
AMM90PS13
AMM90PS15
AMM90PS18
AMM90PS19
AMM90PS24
AMM90PS30
AMM90PS48
4. For optional Barrel Jack Connector, add ‘ B1’ to model number e.g.
AMM90PS24C2B1 (not available on 5 V or 6 V models).
5. Not CEC2008 or EISA2007 compliant.
Mechanical Details
1 3
Cable length approx 48” (1219mm)
2
4
1.72
PIN CONNECTIONS
PS05 - PS06
Pin
Function
1
Return
2
Return
3
Output+
4
Return
5
Output+
(43.7)
3
Rubber foot
(4 places.)
5
2
1
1
4
PIN CONNECTIONS
PS09 - PS48
Pin
Function
1
Return
2
Output+
3
Return
4
Output+
3
5.12
(130.0)
2
3.00 (76.2)
4
Optional Barrel Jack Connector
0.02
(0.5)
0.433 (11.0)
0.083 (2.1)
0.217
(5.5)
0.413
(10.5)
AC Input
IEC 320-C14 Inlet - For Class I version
IEC 320-C18 Inlet - For Class II version
Notes
1. Dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum,
except output cable length.
2. Weight: 1.46 lb (660 g).
3. Maximum load per pin on output connector is 7.5 A.
4. Output connector for AMM90PS05 and PS06 is 5 pin 180° DIN.
5. Output connector for AMM90PS09-PS13 is 4 Pin DC Power Plug non locking
type. For mating half, use KYCON KPJ-4S-S or equivalent.
6. Output connector for AMM90PS15-PS48 is 4 Pin DC Power Plug locking type.
For mating half, use KYCON KPJ-4S-S or equivalent.
7. For Class I versions shell is connected to ground. For Class II versions shell is
capacitively coupled to input.
8. For European mains lead order part: EU-MAINS-IEC
9. For UK mains lead order part: UK-MAINS-IEC
10. For US mains lead order part: US-MAINS-IEC
11. Mains lead length is 76“ (1930 mm) approx
ACDC
AMM90
Models and Ratings
Output
Power
96 W
105 W
120 W
120 W
120 W
120 W
120 W
120 W
120 W
Output
Voltage
12 V
15 V
18 V
19 V
20 V
24 V
30 V
36 V
48 V
Output
Current
8.00 A
7.00 A
6.67 A
6.32 A
6.00 A
5.00 A
4.00 A
3.34 A
2.50 A
Total
Regulation(2)
5%
5%
5%
5%
5%
5%
5%
5%
5%
Model
Number(3, 4)
AMM120PS12
AMM120PS15
AMM120PS18
AMM120PS19
AMM120PS20
AMM120PS24
AMM120PS30
AMM120PS36
AMM120PS48
Notes
1.
2.
3.
4.
Ripple and noise measured at 20 MHz bandwidth with a 10 µF tantalum and 0.1 µF ceramic cap connected at the measurement point.
Total regulation includes initial set accuracy, line and load regulation.
For Class II versions, add ‘ C2’ to model number e.g. AMM120PS24C2.
For optional barrel jack connector add suffix ‘ B1’ to model number e.g. AMM120PS24C2B1.
Mechanical Details
1 3
Cable length approx 48” (1220mm)
2
4
1.72
Power On Indicator
(43.7)
Rubber foot
(4 places.)
6.30
(160.0)
1
3
2
4
Pin
1
2
3
4
Function
Return
Output+
Return
Output+
Optional Barrel Jack Connector
0.433 (11.0)
0.217
(5.5)
3.00 (76.2)
0.083 (2.1)
0.413
(10.5)
0.02
(0.5)
ACDC
AMM120
Models and Ratings
AC Input
IEC 320-C14 Inlet - For Class I version
IEC 320-C18 Inlet - For Class II version
Notes
1. Dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum,
except output cable length.
2. Weight 1.72 lbs (780 g).
3. Maximum load per pin on output connector is 5 A.
4. Output connector for AMM120PS12 & PS15 is 4 Pin DC Power Plug,
non-locking type. For mating half, use KYCON KPJ-4S-S or equivalent.
5. Output connector for AMM120PS18-PS48 is 4 Pin DC Power Plug, locking
type. For mating half, use KYCON KPJ-4S-S or equivalent.
6. For Class I versions shell is connected to ground. For Class II versions shell is
capacitively coupled to input.
7. For European mains lead order part: EU-MAINS-IEC
8. For UK mains lead order part: UK-MAINS-IEC
9. For US mains lead order part: US-MAINS-IEC
10. Mains lead length is 76“ (1930 mm) approx
Output
Power
132 W
144 W
150 W
150 W
150 W
150 W
Output
Voltage
12 V
15 V
19 V
24 V
27 V
48 V
Total
Regulation(2)
5%
5%
5%
5%
5%
5%
Output
Current
11.00 A
9.60 A
7.90 A
6.25 A
5.56 A
3.13 A
Model
Number(4,5)
AMM150PS12
AMM150PS15
AMM150PS19(6)
AMM150PS24(3,6)
AMM150PS27(3,6)
AMM150PS48(3,6)
Notes
1. Ripple and noise measured at 20 MHz bandwidth with a 10 µF tantalum and 0.1 µF ceramic capacitor connected at the measurement point.
2. Total regulation includes set accuracy, line and load regulation.
3. For optional barrel jack connector, add suffix ‘ B1’ to the model number e.g. AMM150PS24B1 (not available on 12 V models)
4. For optional AC cable restraint on the Class I Version, add suffix ‘ A’ to the model number e.g. AMM150PS24A or AMM150PS24B1A. AC mains lead must be
Interpower Corporation, part number: 70006020300. Optional AC cable restraint is not available on the Class II Version.
5. For Class II Versions, and ‘ C2’ to model number, eg. AMM150PS24C2.
6. Energy Star Level V.
Mechanical Details
1 3
Cable length approx 48” (1219 mm)
2 4
1.96
Power On Indicator
(49.7)
Rubber foot
(4 places.)
8.17
(207.6)
1
3
2
4
Pin
1
2
3
4
Function
Return
Output+
Return
Output+
Optional Barrel Jack Connector
0.433 (11.0)
3.23 (82.0)
0.413
(10.5)
0.02
(0.5)
0.083 (2.1)
0.217
(5.5)
Notes
AC Input
IEC 320/C14 Inlet - For Class I Version
IEC 320/C18 Inlet - For Class II Version
1. Dimensions shown in inches (mm).
Tolerance is 0.02 (0.5) maximum, except output cable length.
2. Weight: 2.16 lbs (980 g).
3. Output connector is 4 Pin DC Power Plug, locking type.
For mating half, use KYCON KPJX-4S-S or equivalent.
4. For Class I versions shell is connected to ground. For Class II Versions shell is
capacitively coupled to input.
5.
6.
7.
8.
For European mains lead order part: EU-MAINS-IEC
For UK mains lead order part: UK-MAINS-IEC
For US mains lead order part: US-MAINS-IEC
Mains lead length is 78“ (2000 mm) approx
31-Jan-12
ACDC
AMM150
Models and Ratings
Similar pages