Minmax AYM-60S12 Ac/dc power module 60w, industrial & medical safety Datasheet

AYM-60 SERIES
AC/DC Power Module 60W, Industrial & Medical Safety
FEATURES
►Fully Encapsulated Plastic Case for
PCB, Chassis and DIN-Rail Mounting Version
►Universal Input 85~264VAC, 47~440Hz
►Protection Class II as per IEC/EN 60536
►I/O Isolation 4000VAC with Reinforced Insulation
►Operating Ambient Temp. Range -40℃ to +80℃
►Overload/Voltage and Short Circuit Protection
►Designed-in EMI Emission meets EN55011/22 Class B & FCC Level B
►Designed-in EMC Immunity meets EN61000-4-2,3,4,5,6,8,11
►Medical EMC Standard meets EMI EN55011 & EMS EN60601-1-2
►Medical Safety meets 2xMOPP per 3rd Edition of IEC/EN 60601-1
& ANSI/AAMI ES60601-1
►UL508 Safety Approval Specifically for Industrial Application
►UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking
PRODUCT OVERVIEW
The new MINMAX AYM-60 series is a range of fully encapsulated AC/DC power modules. These high performance products feature an extended operating
temperature range of -40°C to +80°C. Universal input voltage 85-264VAC and UL/IEC/EN safety approvals including medical safety and UL508 listing
qualify these power supplies modules for applications in products with worldwide markets.EMI-filter meets EN55022, class B and FCC,part15, class B.
The AYM-60 series power modules provide an economical solution for many space critical applications in commercial, medical and industrial electronic
equipment.
Model Selection Guide
Model
Number
Output
Voltage
Output Current
Input Current
115VAC, 60Hz
230VAC, 50Hz
Max.
@Max. Load
VDC
mA
mA(typ.)
AYM-60S051
5.1
10000
880
AYM-60S12
12
5000
AYM-60S15
15
AYM-60S24
24
AYM-60S48
48
Max. capacitive
Load
Efficiency
(typ.)
@Max. Load, 115VAC
μF
%
528
8000
84
1000
600
3900
87
4000
1000
600
3300
87
2500
1000
600
1500
87
1250
988
593
680
88
Input Specifications
Parameter
Model
AC Voltage Input Range
Input Frequency Range
All Models
DC Voltage Input Range
No-Load Power Consumption
Inrush Current (Cold Start at 25℃)
E-mail:[email protected]
2015/12/28
REV:5
Typ.
Max.
Unit
85
---
264
VAC
47
---
440
Hz
120
---
370
VDC
---
---
0.5
W
115VAC
---
---
30
A
230VAC
---
---
60
A
Tel:886-6-2923150
Page 1 of 5
Min.
AYM-60 SERIES
AC/DC Power Module 60W, Industrial & Medical Safety
Output Specifications
Parameter
Conditions / Model
Output Voltage Setting Accuracy
Line Regulation
Load Regulation
Minimum Load
Min.
------No minimum Load Requirement
5.1VDC Output Models
--0-20 MHz Bandwidth
Other Output Models
--Zener diode clamp
------85VAC, Hiccup Mode, auto-recovery
105
(long term overload condition may cause damage)
Hiccup mode, Automatic Recovery
Vin=Min. to Max. @Full Load
Io=0% to 100%
Ripple & Noise(3)
Over Voltage Protection
Temperature Coefficient
Overshoot
Current Limitation
Short Circuit Protection
Typ.
Max.
Unit
±1.0
±0.2
±0.5
±2.0
±1.0
±1.0
%Vnom.
%
%
2.0
1.0
120
±0.02
---
3.0
1.5
----5
%VPP of Vo
%VPP of Vo
% of Vo
%/℃
%
---
---
%Inom.
General Specifications
Parameter
I/O Isolation Voltage
Leakage Current
I/O Isolation Resistance
Switching Frequency
Conditions
Min.
Typ.
Max.
Unit
Reinforced Insulation, Rated For 60 Seconds
4000
--1000
-------
--80
--65
20
80
125,000
-------------
VACrms
μA
MΩ
KHz
ms
ms
Hours
500 VDC
115VAC, 60Hz
230VAC, 50Hz
MIL-HDBK-217F@25℃, Ground Benign
Hold-up Time
MTBF (calculated)
Protection Class II
According IEC/EN 60536
UL/cUL 60950-1, CSA C22.2 No 60950-1
ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1
IEC/EN 60950-1, IEC/EN 60601-1 3rd Edition 2xMOPP
UL508, CSA C22.2 No.107.1-01
UL/cUL 60950-1 recognition (UL certificate), IEC/EN 60950-1 (CB-report), UL/cUL 508 listed certificate
ANSI/AAMI ES60601-1 2xMOPP recognition (UL certificate), IEC/EN 60601-1 3rd Edition (CB-report)
Safety Standards
Safety Approvals
EMC Specifications
Parameter
Standards & Level
EMI
Conduction and Radiation
Performance
EN55011, EN55022, EN61000-6-4, EN61000-6-3, FCC part 15
EN60601-1-2, EN55024, EN61000-6-2, EN61000-6-1
ESD
EN61000-4-2 Air ± 8kV , Contact ± 4kV
Radiated immunity
EN61000-4-3 10V/m
Fast transient
EN61000-4-4 ±2kV
Surge
EN61000-4-5 ±1kV
EN61000-4-6 10Vrms
Conducted immunity
EN61000-4-8 30A/m
PFMF
EN61000-4-11 30% 10ms
Dips
EN61000-4-11 >95% 5000ms
Interruptions
EMS
Class B
A
A
A
A
A
A
A
B
Environmental Specifications
Parameter
Operating Ambient Temperature Range
(See Power Derating Curve)
Storage Temperature Range
Humidity (non condensing)
Power Derating
Cooling
Lead Temperature
(1.5mm from case for 10Sec.)
E-mail:[email protected]
2015/12/28
REV:5
Conditions
Min.
Max.
Unit
Natural Convection
-40
+80
℃
-40
---
+95
95
℃
% rel. H
W/℃
260
℃
Above +60℃
2.3
Natural Convection
---
Tel:886-6-2923150
Page 2 of 5
AYM-60 SERIES
AC/DC Power Module 60W, Industrial & Medical Safety
Notes
1
2
3
4
5
6
7
8
This product is not designed for use in critical life support systems, equipment used in hazardous environment, nuclear control systems or other
such applications which necessitate specific safety and regulatory standards other the ones listed in this datasheet.
Specifications typical at Ta=+25℃, resistive load, 115VAC, 60Hz input voltage, after warm-up time rated output current unless otherwise noted.
Ripple & Noise of PCB mounting type measured with a 0.1μF/50V MLCC and a 1μF/50V Aluminum electrolytic.
Safety approvals cover frequency 47-63 Hz.
We recommend to protect the converter by a slow blow fuse in the input supply line.
Other input and output voltage may be available, please contact factory.
That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
Specifications are subject to change without notice.
Package Specifications PCB Mounting
Mechanical Dimensions
Pin Connections
25.4 [1.00]
6
1.0 [ 0.04]
4
Pin
Function
1
AC (N)
2
AC (L)
4
+Vout
6
-Vout
88.9 [3.50]
17.00
[0.67]
Bottom View
81.3 [3.20]
Mounting M4 Thread
36.45
[1.44]
17.75
[0.70]
1
2
►All dimensions in mm
55.88 [2.20]
0.7 [0.03]
67.5 [2.66]
6.0 [0.24]
Physical Characteristics
Case Size
:
88.9x67.5x34.2mm (3.50x2.66x1.35 inches)
Case Material
:
Plastic resin (flammability to UL 94V-0 rated)
Pin Material
:
Copper Alloy with Gold Plate Over Nickel Subplate
Weight
:
360g
E-mail:[email protected]
2015/12/28
REV:5
Tel:886-6-2923150
Page 3 of 5
34.2 [1.35]
►Tolerance:
►Pin
(inches)
±1.0 (±0.04)
diameter  1.0 ±0.1 (0.04±0.004)
AYM-60 SERIES
AC/DC Power Module 60W, Industrial & Medical Safety
Package Specifications Chassis Mounting (order code suffix C)
Mechanical Dimensions
Connections
4x 3.5
2
5
Top View
1
4
Indication LED
67.8 [2.67]
6
50.0 [1.97]
7
3
DCON
POWER"GOOD"INDICATOR
Pin
Function
1
AC (N)
2
AC (L)
3
NC
4
+Vout
5
NC
6
-Vout
7
NC
100.0 [3.94]
NC: No Connection
112.0 [4.41]
Physical Characteristics
Case Size
:
112.0x67.8x38.0mm (4.41x2.67x1.50 inches)
Case Material
:
Plastic resin (flammability to UL 94V-0 rated)
Weight
:
380g
Package Specifications with DIN Rail Mounting Bracket
Mechanical Dimensions
4x
3.5
Top View
4
1
67.8 [2.67]
2
5
50.0 [1.97]
7
6
3
Indication LED
DCON
POWER"GOOD"INDICATOR
100.0 [3.94]
112.0 [4.41]
Physical Characteristics
Case Size
:
112.0x67.8x38.0mm (4.41x2.67x1.50 inches)
Case Material
:
Plastic resin (flammability to UL 94V-0 rated)
Weight
:
433g
E-mail:[email protected]
2015/12/28
REV:5
Tel:886-6-2923150
Page 4 of 5
38.0 [1.50]
11.0 [0.43]
3.0 [0.12]
5.0[0.20]
11.0 [0.43]
16.8 [0.66]
90.0 [3.54]
38.0 [1.50]
5.0[0.20]
11.0 [0.43]
16.8 [0.66]
90.0 [3.54]
►All dimensions in mm
►Tolerance:
(inches)
±1.0 (±0.04)
AYM-60 SERIES
AC/DC Power Module 60W, Industrial & Medical Safety
DIN-Rail Mounting Bracket (Order Code for Kit : AC-DIN-02)
Order Code Table
PCB Mounting
AYM-60S051
Chassis Mounting
AYM-60S051C
With DIN Rail Mounting by two Order Code
AYM-60S051C
AC-DIN-02
AYM-60S12
AYM-60S12C
AYM-60S12C
AC-DIN-02
AYM-60S15
AYM-60S15C
AYM-60S15C
AC-DIN-02
AYM-60S24
AYM-60S24C
AYM-60S24C
AC-DIN-02
AYM-60S48
AYM-60S48C
AYM-60S48C
AC-DIN-02
Minmax Technology Co., Ltd.
18, Sin Sin Road, An-Ping Industrial District, Tainan 702, Taiwan
Tel: 886-6-2923150 Fax: 886-6-2923149 E-mail: [email protected]
2015/12/28
REV:5
Page 5 of 5