1~4 Output Medical Type 200W Medical series Features Universal AC input / Full range Low leakage current <180 A Protections: Short circuit / Overload / Over voltage / Over temperature Free air convection for 140W and forced air convection for 200W UL60601-1 medical safety approved With power good and fail signal output Built-in remote ON-OFF control Built-in remote sense function Fixed switching frequency at 100KHz 3 years warranty EN60601-1 IEC60601-1 Specification AC INPUT VOLTAGE 90~264 VAC, 47~440Hz / 127~370VDC. POWER FACTOR (Typ.) PF>0.95/230VAC PF>0.98/115VAC at full load AC INPUT CURRENT (Typ.) Maximum input current 3.5A at 115VAC, 60Hz or 1.6A at 230VAC, 60Hz with 100% output load. INRUSH CURRENT (Typ.) Inrush current is less than 25A at 115VAC or less than 40A at 230VAC under cold start conditions. Limiting provided by internal thermistors. SETUP, RISE TIME 1000ms, 20ms / 230VAC at full load 3000ms, 20ms / 115VAC at full load HOLD-UP TIME (Typ.) 16ms / 230VAC at full load 16ms / 115VAC at full load LEAKAGE CURRENT Leakage current is less than 180 A at 264VAC for earth leakage current Leakage current is less than 100 A at 264VAC for patient leakage current DC OUTPUT ADJ. RANGE DC output voltage (or CH1 of multiple output models) can be adjusted between -5%~+10% rated output voltage by potential meter. OVERLOAD PROTECTION Fully protected against short circuit and output overload. The hiccup type protection will be activated at 120~160% rated load and recovers automatically after fault condition is removed. OVER VOLTAGE PROTECTION Provided on output channel 1 only at 115%~135% rated output voltage. Output will be shut down when this protection is activated. OVER TEMPERATURE PROTECTION When the temperature of TSW1 which detect on heat sink of power transistor reaches 95 This protection is activated. Then output will be shut down and recovers automatically after temperature goes down. POWER GOOD / FAIL SIGNAL TTL logic high for power good and TTL low for power fail. When the output voltage reaches 90% of rated value, a +5V TTL signal will be sent out with a 10~500ms delay; At least 1ms before the output voltage goes below 90% of the rated value, the TTL signal will be turned off. * MPS-200-3.3 does not have this function. REMOTE CONTROL RC+/RC-:0 ~ 0.8V=power on; 4 ~ 10V=power off sink current<4~10mA WORKING TEMP. Whole series can operate from -20~70 . Please refer to the derating curves. WORKING HUMIDITY 20~90% RH non-condensing. STORAGE TEMP., HUMIDITY -40~+85 , 10~90% RH TEMP. COEFFICIENT 0.04%/ on all outputs at full load between 0~50 of ambient temperature. VIBRATION 2G of acceleration, vibrating frequency adjust from 10Hz ~500Hz within a 10-minute cycle. 6 testing cycles (60 minutes) each along X, Y, Z axes. SAFETY STANDARDS Medical : UL60601-1, TUV EN60601-1, IEC60601-1 approved Commercial : Also design refer to UL60950-1, TUV EN60950-1 WITHSTAND VOLTAGE 4000VAC between input and output 1500VAC between input and F.G. 1500VAC between output and F.G. ISOLATION RESISTANCE >100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage. EMI COMPLIANCE EMI Specifications Conducted & Radiation Harmonic distortion Voltage flicker EMS COMPLIANCE EMS Specification ESD air ESD contact RF field susceptibility EFT(Electrical Fast Transient)/Burst Lightning/Surge Conducted RF susceptibility Magnetic field immunity Voltage dip, interruption Digital phone carrier immunity MTBF 262,100 hours min. at full load and 25 calculated per MIL-HDBK-217F. Compliance Level EN55011, Class B EN55022, Class B EN61000-3-2 EN61000-3-3 Compliance Level EN61000-4-2, Level 3, 8KV EN61000-4-2, Level 2, 4KV EN61000-4-3, Level 2, 3V/m Level 3, 10V/m EN61000-4-4, Level 2, 1KV/5KHz Level 3, 2KV/5KHz EN61000-4-5, Level 4, 2KV/Line-Line 4KV/Line-Earth EN61000-4-6, Level 2, 3Vrms/m Level 3, 10Vrms/m EN61000-4-8, Level 2, 3A/m Level 3, 10A/m EN61000-4-11,Compliance ENV50204, Level 2, 3V/m, 900MHz Level 3, 10A/m, 900MHz of ambient temperature, DIMENSION (L*W*H) 177.8x107.2x35.5mm or 7"x4.22"x1.4" PACKING 0.66Kg; 24pcs/16.8Kg/0.99CUFT File Name:200W MEDICAL-SPEC 2010-10-18 200W Medical 1~4 Output Medical Type series Output Chart MODEL OUTPUT VOLTAGE RATED CURRENT MPS-200-3.3 3.3V 40A OUTPUT CURRENT RIPPLE & NOISE VOLTAGE PEAK LOAD (Max.) TOLERANCE WITH 25CFM FAN (Note 2) (Note 3) LINE LOAD EFFICIENCY REGULATION REGULATION (typ.) MINIMUM LOAD CONVECTION (max.) WITH FAN (25CFM) (Note 4) 0A 28A 40A 48A 80mVp-p 2.0% 0.5% 1.0% 77% MPS-200-5 5V 40A 0A 28A 40A 48A 80mVp-p 2.0% 0.5% 1.0% 81% MPS-200-12 12V 16.7A 0A 11.7A 16.7A 20A 100mVp-p 2.0% 0.5% 1.0% 84% MPS-200-15 15V 13.4A 0A 9.4A 13.4A 16A 100mVp-p 2.0% 0.5% 1.0% 85% MPS-200-24 24V 8.4A 0A 5.9A 8.4A 10A 150mVp-p 1.0% 0.5% 1.0% 86% MPS-200-48 48V 4.2A 0A 3A 4.2A 5A 200mVp-p 1.0% 0.5% 1.0% 87% MPD-200A 5V 20A 4A 15A 20A 24A 80mVp-p 2.0% 0.5% 1.0% 12V 8A 0.8A 5.4A 8A 9.6A 120mVp-p +8,-5% 1.0% 4.0% 5V 20A 4A 15A 20A 24A 80mVp-p 2.0% 0.5% 1.0% 24V 4A 0.4A 2.7A 4A 4.8A 180mVp-p 6.5% 1.0% +4,-6% 5V 20A 4A 15A 20A 24A 80mVp-p 2.0% 0.5% 1.0% 12V 7.5A 0.8A 5A 7.5A 9A 120mVp-p 8.0% 1.0% 5.0% -5V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% 5V 20A 4A 15A 20A 24A 80mVp-p 2.0% 0.5% 1.0% 12V 6A 0.6A 4.4A 6A 7.2A 120mVp-p 8.0% 1.0% 5.0% -12V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% 5V 20A 4A 15A 20A 24A 80mVp-p 2.0% 0.5% 1.0% 15V 4.7A 0.5A 3.3A 4.7A 5.6A 150mVp-p 8.0% 1.0% 5.0% -15V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% 5V 20A 4A 15A 20A 24A 80mVp-p 2.0% 0.5% 1.0% 24V 3A 0.3A 2.2A 3A 3.6A 180mVp-p 8.0% 1.0% 5.0% 12V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% 5V 15A 3A 12A 15A 18A 80mVp-p 2.0% 0.5% 1.0% 12V 7A 0.7A 5.3A 7A 8.4A 120mVp-p 8.0% 1.0% 5.0% -5V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% -12V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% 5V 15A 3A 12A 15A 18A 80mVp-p 2.0% 0.5% 1.0% 15V 5A 0.5A 4A 5A 6A 150mVp-p 6.0% 1.0% 5.0% -5V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% -15V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% 5V 15A 3A 12A 15A 18A 80mVp-p 2.0% 0.5% 1.0% 24V 3A 0.3A 2.3A 3A 3.6A 180mVp-p 8.0% 1.0% 5.0% 12V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% -12V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% 5V 15A 3A 12A 15A 18A 80mVp-p 2.0% 0.5% 1.0% 24V 2.7A 0.3A 2.1A 2.7A 3.3A 180mVp-p 8.0% 1.0% 5.0% 15V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% -15V 2A 0A 1A 2A 2.4A 80mVp-p 5.0% 0.5% 1.0% MPD-200B MPT-200A MPT-200B MPT-200C MPT-200D MPQ-200B MPQ-200C MPQ-200D MPQ-200F Notes : 82% 83% 80% 80% 80% 81% 78% 78% 79% 81% 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power. 5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) 6. Derating may be needed under low input voltages. Please check the derating curve for more details. File Name:200W MEDICAL-SPEC 2010-10-18 200W Medical 1~4 Output Medical Type Mechanical Specification(MPS-200) series Unit:mm 6.1[0.24"] 177.8[7"] 165.1[6.5"] 6.35[0.25"] SVR1 LED 1 4 5 6 10 9 8 7 6 5 4 3 2 1 95.25[3.75"] 107.2[4.22"] 2 3 CN1 3 2 1 CN4 20 19 18 17 16 15 14 13 12 11 CN2 FG 15cm 4- 3.96[0.156"] 25CFM min. 35.5[1.4"] FAN 3[0.118"] max. DC Output Connector (CN4) : JS-2008-03*2 or equivalent AC Input Connector (CN1) : JST B3P-VH or equivalent Pin No. 1 2 3 Assignment AC/N No Pin AC/L Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent Pin No. 1 2 3 4 5 6 DC Output Connector (CN2) : MOLEX 5566-20 or equivalent Pin No. 1~5,11~15 6~10,16~20 Assignment DC OUTPUT -V DC OUTPUT +V Mating Housing Terminal MOLEX 5557 or equivalent MOLEX 5556 or equivalent Derating Curve (MPS-200) 25CFM forced air cooling JS-2007-03*2 or equivalent JS-2007-T or equivalent 90 80 Convection cooling 70 LOAD(%) LOAD (%) Terminal 100 80 40 20 -20 Mating Housing Static Characteristics (MPS-200) 100 60 Assignment PG RSGND RC+ RS+ RC- 0 10 20 30 40 50 AMBIENT TEMPERATURE ( ) 60 70 (HORIZONTAL) 60 50 40 90 100 115 140 160 180 200 220 240 264 INPUT VOLTAGE (V) 60Hz File Name:200W MEDICAL-SPEC 2010-10-18 200W Medical 1~4 Output Medical Type series Mechanical Specification(MPD/T/Q-200) 6.1[0.24"] 177.8[7"] 165.1[6.5"] 6.35[0.25"] SVR1 LED 1 2 3 4 5 6 CN4 95.25[3.75"] 107.2[4.22"] 1 CN2 CN1 3 2 1 CN3 16 FG 15cm 4- 3.96[0.156"] 25CFM min. 35.5[1.4"] FAN 3[0.118"] max. DC Output Connector (CN4) : JS-2008-03*2 or equivalent AC Input Connector (CN1) : JST B3P-VH or equivalent Pin No. 1 2 3 Assignment AC/N No Pin AC/L Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent Pin No. 1 2 3 4 5 6 DC Output Connector (CN2,3) : JST B8P-VH*2 or equivalent Pin No. 1,2,3,4 5~11 12,13 14 15 16 Assignment V1 COM V2 V3 No pin V4 Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent Derating Curve (MPD/T/Q-200) 25CFM forced air cooling MPD-200 JS-2007-03*2 or equivalent JS-2007-T or equivalent 90 MPT/Q-200 80 Convection cooling 70 MPT/Q-200 LOAD(%) LOAD (%) Terminal 100 80 MPD-200 40 20 -20 Mating Housing Static Characteristics (MPD,T,Q-200) 100 60 Assignment PG RSGND RC+ RS+ RC- 0 10 20 30 40 50 AMBIENT TEMPERATURE ( ) 60 70 (HORIZONTAL) 60 50 40 90 100 115 140 160 180 200 220 240 264 INPUT VOLTAGE (V) 60Hz File Name:200W MEDICAL-SPEC 2010-10-18