Powerbox OBP02016 200-300 watts single & multiple output ac/dc medical Datasheet

MEDICAL & MILITARY
Medline 300 Series
2 0 0 - 3 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Features
• EN61000-3-2 class A and D compliant.
• Power factor 0.98 typical.
• Remote sense on output 1 and output 2.
• Overvoltage protection.
• Overcurrent protection.
• Power fail detection (PFD) signal.
• 100% burn-in at full rated load.
• Remote inhibit - TTL high disables output.
• Short curcuit protection
Specifications
OPERATING
INPUT
Input Voltage
90-264Vac
Frequency
47Hz–63Hz
Input Current
4.7A (rms) for 115Vac
2.2A (rms) for 230Vac
Leakage current
Isolation
Inrush Current
MTBF
350,000 hrs @ 25°C
Efficiency
70% minimum.
ENVIRONMENTAL
Operating Temperature
0°C to 70°C
90µA max. at 115Vac. 60Hz
150µA max. at 230Vac. 50Hz
Derating
Derate from 100% at +50°C linearly to 50%
at 70°C
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
Storage Temperature
-40°C to 85°C
Humidity
5%–95% non-condensing
30A at 115V or 80A at 230V at 25°C and cold
start
Cooling
35 CFM forced air, provided on “C” version, to be
provided for “B” version by user
STANDARDS AND APPROVALS
OUTPUT
Power
See table
Safety standards
UL60601-1, TUV EN60601-1
Voltage
See table
C-Tick
AS/NZS CISPR11 Group 1, Class A
Ripple & Noise
2% peak to peak max
EMC standards
EN55011 Class B, FCC Class B, EN61000-3-2, -3,
EN61000-4-2, -3, -4, -5, -6, -8, -11
Overvoltage Protection
Provided on output #1 only. Set at 115%–140%
of nominal output voltage
MECHANICAL
Overcurrent Protection
All outputs protected for short circuit conditions
Dimensions
177.8x101.6x38.1mm
Weight
0.82Kgs for “B” version.
0.96Kgs for “C” version.
Temperature Coefficient
All outputs ±0.04% per °C
Transient Response
Max. excursion of 4%or better on all models,
recovering to 1% of final value within 500µs after
a 25% step load change
PFD Signal
TTL logic high for normal operation and TTL logic low
upon loss of input power. This signal appears at least
1ms prior to +5V output dropping 5% below its
nominal value. This signal also provides a minimum
delay of 100ms after +5V is within regulation
Remote inhibit
Requires an external TTL high level signal to
inhibit outputs for standard models
Line Regulation
±0.2% maximum at full load
Holdup Time
Typically 12ms at 110VAC
Your dependable power partner – www.powerbox.com.au
347
MEDICAL & MILITARY
Medline 300 Series
2 0 0 - 3 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Selection Table
MAX.
MODEL
OUTPUT #1
OUTPUT #2
VNOM IMIN
IMAX
OUTPUT #3
TOL
VNOM
OUTPUT #4
IMIN IMAX TOL
VNOM
IMIN
OUTPUT
NUMBER
VNOM
IMIN
IMAX
TOL
OBP02013
3.3V
0A
60A
3%
IMAX
TOL
POWER
200W
OBP02016
5.1V
0A
60A
2%
300W
OBP02031
12V
0A
25A
2%
300W
OBP02037
15V
0A
20A
2%
300W
OBP02046
24V
0A
12.5A
2%
300W
OBP02055
30V
0A
10A
2%
300W
OBP02170
48V
0A
6.3A
2%
OBP02403
+3.3V
3.0A
35A
3%
5.1V
2.0A
22A
2%
12V
0A
4A
4%
12V
0A
4A
4%
300W
250W
OBP02404
+5.1V
2.0A
35A
2%
12V
1.0A
10A
2%
12V
0A
4A
4%
5.1V
0A
4A
4%
300W
OBP02407
+5.1V
2.0A
35A
2%
15V
0.8A
8A
2%
15V
0A
4A
4%
24V
0A
2.5A
4%
300W
OBP02410
+5.1V
2.0A
35A
2%
12V
1.0A
10A
2%
12V
0A
4A
4%
12V
0A
4A
4%
300W
OBP02413
+5.1V
2.0A
35A
2%
12V
1.0A
10A
2%
12V
0A
4A
4%
24V
0A
2.5A
4%
300W
OBP02418
+5.1V
2.0A
35A
2%
12V
1.0A
10A
2%
12V
0A
4A
4%
15V
0A
4A
4%
300W
OBP02419
+5.1V
2.0A
35A
3%
24V
0.5A
5A
2%
12V
0A
4A
4%
12V
0A
4A
4%
300W
OBP02424
+5.1V
2.0A
35A
2%
24V
0.5A
5A
2%
5.1V
0A
4A
4%
15V
0A
4A
4%
300W
OBP02427
+5.1V
2.0A
35A
2%
12V
1.0A
10A
2%
5.1V
0A
4A
4%
24V
0A
2.5A
4%
300W
OBP02430
24V
0.5A
6.3A
2%
12V
1.0A
10A
2%
5.1V
0A
4A
4%
12V
0A
4A
4%
300W
OBP02433
24V
0.5A
6.3A
2%
12V
1.0A
10A
2%
5.1V
0A
4A
4%
24V
0A
2.5A
4%
300W
OBP02436
24V
0.5A
6.3A
2%
12V
1.0A
10A
2%
12V
0A
4A
4%
12V
0A
4A
4%
300W
OBP02439
24V
0.5A
6.3A
3%
24V
0.5A
5A
2%
5.1V
0A
4A
4%
15V
0A
4A
4%
300W
OBP02442
24V
0.5A
6.3A
2%
24V
0.5A
5A
2%
12V
0A
4A
4%
12V
0A
4A
4%
300W
NOTES: (1) Add suffix “B” for U-bracket format and “C” for enclosed format.
(2) All outputs are floating. It can be connected externally for postive or negative output.
(3) Output 1 and 2 can be adjusted within ±5% of their nominal voltage.
(4) Output 3 and 4 can be adjusted within ±15% of their nominal voltage.
(5) 300 watts for “C” version with a cover and fan assembly 150 watts for “B” version without moving air (maximum current of ouput #1 & #2 derated to 50%),
or 300 watts with 25 CFM forced air provided by user.
(6) PM300-10-3 is rated 200 watts with 35 CFM forced air cooling or 100 watts convectional cooled. PM300-40-3 is rated 250 watts with 35 CFM forced air
cooling (maximum current of output 1 & 2 derated to 50%) or 125 watts convection cooled.
(7) Single output models may be operated at no-load. At no-load, output voltage tolerance increases to 10%.
Pin Connection
CONNECTOR P1(AC)
MODEL/PIN
OBP02013
OBP02031
OBP02046
OBP02070
OBP02016
OBP02037
OBP02055
1
2
3
Live
Neutral
Ground
P2
P3
P4
OUTPUT OUTPUT #1 OUTPUT #1
P5
P6
OUTPUT #1
OUTPUT #1
#1 RET
CONNECTOR P4
MODEL/PIN
OBP02013
OBP02031
OBP02046
OBP02070
348
OBP02016
OBP02037
OBP02055
1
2
3
4
5
6
7
8
9
10
SIGNAL
GROUND
+S
(Vo1)
-S
(Vo1)
PFD
INHIBIT
N.C.
N.C.
N.C.
OV
(FAN)
12V
(FAN)
MEDICAL & MILITARY
Medline 300 Series
2 0 0 - 3 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Pin Connection ctnd
P1(AC)
MODEL/PIN
OBP02403
OBP02407
OBP02413
OBP02419
OBP02427
OBP02433
OBP02439
OBP02404
OBP02410
OBP02418
OBP02424
OBP02430
OBP02436
OBP02442
P2
1
2
3
Live
Neutral
Ground
P3
OUTPUT
#1
P5
OUTPUT
#1 RET
P7
P6
OUTPUT
#2
OUTPUT
#2 RET
1
2
3
4
OUTPUT
#3
OUTPUT
#3 RET
OUTPUT
#4
OUTPUT
#4 RET
CONNECTOR P4
MODEL/PIN
OBP02403
OBP02407
OBP02413
OBP02419
OBP02427
OBP02433
OBP02439
OBP02404
OBP02410
OBP02418
OBP02424
OBP02430
OBP02436
OBP02442
1
2
3
4
5
6
7
8
9
10
SIGNAL
GROUND
+S
(Vo1)
-S
(Vo1)
PFD
INHIBIT
N.C.
N.C.
N.C.
OV
(FAN)
12V
(FAN)
Technical Illustration
MULTIPLE
SINGLE
Note:
1. Dimensions shown in inches (mm)
2. Tolerance 0.02 (0.5) maximum
Input connector P1 is DT-35-B01W-03 screws are M3, Nickel plated
3. Connector P4 mates with Molex housing 50-37-5103 and pins 5263
4. Connectors P2, P3, P5 and P6: M3*0.5 screw connections
5. 12V (P4) is for DC fan rated 12V/0.1A
6. Weight 1.20 Kgs approx
Note:
1. Dimensions shown in inches (mm)
2. Tolerance 0.02 (0.5) maximum
3. Input connector P1 Dinkle is DT-35-B01W-03 screws are M3, Nickel plated
4. Connector P4 mates with Molex housing 50-37-5103 and pins 5263
5. Connectors P2, P3, P5 and P6: M3*0.5 screw connections
6. Output connector P7 mates Molex housing 09-50-3041 and Molex 2878 series
crimp terminal
7. 12V (P4) is for DC fan rated 12V/0.1A
8. Weight 1.20 Kgs approx
Your dependable power partner – www.powerbox.com.au
349
Similar pages