MPS HPU31B-104 30w desk top type medical power supplies for health care application Datasheet

HPU31B SERIES
30W Desk top type medical power supplies for Home health care applications
Description:
The HPU31B series of AC/DC switching mode power supplies
provide 30 Watts of continuous output power . All supplies are
UL94V-1 min compliant, include IEC-320-C18 input for worldwide
applications. They are suited for use in Home health care
applications. All models meet FCC Part-18 class B and CISPR-11
EN55011 class B emission Limits and are designed to comply with
UL/c-UL(UL 60601-1) ,TUV/T-mark (EN 60601-1) and new CE
requirements. All units are 100% burned in and tested.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C8 Input Inlet
Single Output
Output Voltage Available From 5VDC Thru 48VDC
Optional Output Connector (See appendix)
Over Voltage Protection (Crowbar Design)
Input Surge Current, Over Voltage And Over Load protection
CEC level V and Engergy Star 2.0 Compliance
Approved as Limited Power Source (LPS).
Class II
2 year warranty
Safety Approvals :
PS
E
Electrical Characteristics:
Sym.
Parameter
Test Conditions
Vin
fin
Po
Vo
Io
Input Voltage
Input Frequency
Output Power Range
Output Voltage Range
Output Current Range
Vin=90 to 264VAC
Iil
Iih
Input Current (Low Line)
Input Current ( High Line)
Io=Full load, Vin=115 VAC
Io=Full load, Vin=230 VAC
I rl
Low Line Inrush Current
I rh
High Line Inrush Current
Io=Full load, 25 C , Cool start,
Vin= 115VAC
Io=Full load, 25 C , Cool start,
Vin= 230VAC
Io=Full Load, Vin=230VAC
Operating Voltage
Min. Typ. Max. Unit
264
63
90
47
30
0
See rating chart
See rating chart
A
A
16
24
A
32
48
A
85
88
%
%
%
132
150
%
%
mS
mS
S
Efficiency
OCP
Over Current Protection
Time of Transient Response Io=Full Load to Half Load, Vin=100VAC
4
Hold-Up Time
12
Io=Full Load, Vin=110VAC
0.3 1.5
2
Start Up Time
Io=Full Load, Vi n= 10 0V AC
Ripple & Noise (Peak to Peak) Full Load, Vin=90VAC
1
Temperature Coefficient
-0.04
0.04
All output
Ttr
Thold
Ts
Vrn
TC
78
Io=Full Load
Vin=230VAC
V
A
0.9
0.34
Eff
REG-i
REG -o
OVP
Line Regulation
Load Regulation
Over Voltage Protection
VAC
Hz
W
0.5
5
112
110
%
%/ C
Environmental :
Sym.
Toper
Tstg
Hr
MTBF
Pd
Parameter
Test Conditions
Operating Temperature
Storage Temperature
Relative Humidity
Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Derate linearly from 100% load at 40 C to 50% load at 70 C
Min. Typ. Max.
Unit
40
C
C
%
Hrs
0
-40
5
0.1M
70
85
95
Safety Specifications:
Sym.
Parameter
Vps Dielectric Withstanding Voltage
for Primary to secondary
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
Test Conditions
Primary to secondary
Vin=220VAC
Vin=110VAC
Min. Typ. Max.
Unit
5600
VDC
B
B
CLASS
CLASS
HPU31B SERIES
30W Desk top type medical power supplies for Home health care applications
Output Voltage And Current Rating Chart (Single Output) :
Total
Regulation
Maximum
Output Power
4.00 ~ 3.33 A
5%
20W
6 ~ 8 VDC
3.83 ~ 2.87 A
5%
23W
8 ~ 11 VDC
3.38 ~ 2.45 A
5%
27W
11 ~ 13 VDC
2.74 ~ 2.30 A
5%
30W
HPU31B-106
13 ~ 16 VDC
2.30 ~ 1.88 A
5%
30W
HPU31B-107
16 ~ 21 VDC
1.88 ~ 1.43 A
5%
30W
HPU31B-108
21 ~ 27 VDC
1.43 ~ 1.11 A
5%
30W
HPU31B-109
27 ~ 33 VDC
1.11 ~ 0.91 A
3%
30W
HPU31B-110
33 ~ 40 VDC
0.91 ~ 0.75 A
3%
30W
HPU31B-111
40 ~ 48 VDC
0.75 ~ 0.63 A
3%
30W
Model Number
Output Voltage
HPU31B-102
5 ~ 6 VDC
HPU31B-103
HPU31B-104
HPU31B-105
Mechanical Specifications:
Output Current
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 265-280gs approx.
3. Optional output connector:
See page Appendix.