SynQor AQ0300MU36TQCFD Medical grade ac/dc power supply with pfc Datasheet

Technical
Specification
300 W Single Vout Series
Medical Grade AC/DC Power Supply With PFC
85-264 Vrms
12/24/36/48 V
300 W
400 W
Up to 91 %
Input Voltage
Semi-Regulated Output
Output Continuous
Output Transient
Full Load Efficiency
300 W Open Frame
Configurations
• 300 W Encased or Open Frame
• 600 W Encased Flat or Stacked
• 900 W Encased Flat or Stacked
300 W Encased
900 W Encased Stacked
600 W Encased Flat
300 W Single Vout Series Features
•
•
•
•
•
•
•
•
High efficiency (91% for 48 VOUT Model at 300 W)
Universal input voltage range
Semi-regulated output for bus stability
Parallel operation supported
Integral fan cooling with speed control
Active PFC; EN61000-3-2 compliant
Low leakage; EN60601-1 compliant
Low noise; EN55011 / EN55022 Class B compliant
Product # AQ0300MUxx Series
Phone 1-888-567-9596
• Over–current, over–voltage, and over–temp
protection
•
•
•
•
•
•
DC Power Good / AC Power Good signals
Remote enable input
Fan status output / Fan enable input
Small size: 3" x 5" x 1.45" (open frame)
RoHS 6/6 compliant
5 V (250 mW) standby output
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 1
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
ACuQor 300W Series ELECTRICAL CHARACTERISTICS
All specifications typical with TA = 25 °C, unless otherwise specified.
MAIN OUTPUT SPECIFICATIONS
GENERAL SPECIFICATIONS
Output power (continuous) 85-132/170-264 Vrms
(5 s transient)
85-132/170-264 Vrms
132-170 Vrms
Nominal DC output
12 Vout
voltage (at 250W)
24 Vout
(Semi-regulated)
36 Vout
48 Vout
Efficiency (see figs. 3 - 10) 12
48
12
48
Vout,
Vout,
Vout,
Vout,
Hold-up time (to -20%)
Vout
/ 36 / 48 Vout
Vout
Vout
Vout
Vout
12
24
Maximum load capacitance 12
24
36
48
Output ripple voltage
115
115
230
230
Vrms,
Vrms,
Vrms,
Vrms,
300
300
300
300
300 W
400 W
See Figure 10
12.4 V
25 V
37.5 V
50 V
89%
90%
90%
91%
W
W
W
W
16 ms @ 300 W
20 ms @ 300 W
16,000 µF
8,000 µF
4,000 µF
2,000 µF
Switching frequency (20 MHz BW)
Twice line frequency (at 300W)
0.5% p-p
5.0% p-p
Turn-on delay
2 s max.
Transient response
Iout steps from 50-75%
At 0.2 A/µs
Overvoltage protection
Cyclic restart
Short circuit protection
Cyclic operation
Total regulation
Over line, load and temperature
Auxillary Output
typ.
typ.
typ.
typ.
3% typ / 6% max.
dev.
100 ms recovery
110-120%
115% rated Iout
±6.0%
5 V @ 50 mA
Always on (See Note 1)
Thermal protection
Automatic recovery
+125 ºC (PCB Temp)
REMOTE_ENABLE
Input Low Voltage
Input High Voltage
0.45 V (max)
4.15 V (min)
Fundamental ripple freq.
Audible noise
Input
Output
Fan speed varies with temp.
Weight (EA \ EC)
(SC \ RC)
(UC \ TC)
MTBF
MIL-217
Demonstrated
ISOLATION SPECIFICATIONS
Isolation voltage
Insulation resistance
500 kHz
250 kHz
39 dBA @ 1 m max.
343 g (12.1 oz) \ 446 g (15.7 oz)
846 g (29.8 oz) \ 879 g (31 oz)
1224 g (43.2 oz) \ 1298 g (45.8 oz)
343.6 kHours
TBD kHours
Input to output
Input to ground
Output to ground (BF & CF )
Output to ground (CFD)
4000 Vrms
1500 Vrms
1500 Vrms
5000 Vpulse
Output to ground
10 MΩ min.
Leakage currents
See Note 2
ENVIRONMENTAL CHARACTERISTICS
Thermal performance
Relative humidity
Altitude
Operating ambient (see Figure 9)
Non-operating ambient
Non-condensing
Random vibration
Operating
Non-operating
5-500 Hz
Shock
Half-sine, 10 ms, 3 axes
0 ºC to +70 ºC
-40 ºC to +85 ºC
5-95% RH
10,000 ft max.
30,000 ft max.
0.03 g2/Hz
20 g peak
EMC CHARACTERISTICS
Conducted emissions
Line frequency harmonics
EN55011 and EN55022, FCC
part15
EN61000-3-2
Level B
Voltage fluctuations
EN61000-3-3
Clause 5b
Class A
ESD air
EN61000-4-2
Level 3
INPUT SPECIFICATIONS
ESD contact
EN61000-4-2
Level 3
AC input voltage
Radiated immunity
EN61000-4-3
Level 3
Fast transients
EN61000-4-4
Level 3
Line surge immunity
EN61000-4-5
Level 3
Conducted immunity
EN61000-4-6
Level 3
Power freq. mag. field
EN61000-4-8
Voltage dip immunity
EN61000-4-11
Universal range
Input frequency
Input current
85-264 Vrms
47-63 Hz
115 Vrms @ 400 W
230 Vrms @ 400 W
Power factor
Input surge current
264 Vrms (cold start)
Internal input fuses
Both AC lines
3 Arms
1.5 Arms
>0.98
40 A max.
3 A/m
Perf Criteria A, A, B
<5% UT 10 ms,
70% UT 500 ms,
40% UT 100 ms
6.3 A
SAFETY AGENCY CERTIFICATIONS (pending)
All certification marks appear on individual unit labels.
UL60601-1 and UL60950
IEC/EN 60601-1 and IEC/EN 60950
CE Marked
cUL CSA Standards
Meets NFPA 99 2005 300 µA earth leakage
NOTES:
1. Derate 1 mA per °C above 50 °C ambient temperature.
2. Leakage currents see page 4.
Product # AQ0300MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 2
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
EFFICIENCY, DERATING, AND VOUT DROOP CURVES
100
14.0
95
13.5
13.0
Output Voltage (V)
Efficiency (%)
90
85
80
75
12.5
12.0
11.5
11.0
115 Vrms
70
10.5
230 Vrms
65
10.0
0
100
200
300
400
0
4
8
12
20
24
28
Figure 2: 12 VOUT droop characteristic.
Figure 1: 12 VOUT efficiency curves.
100
28
95
27
26
Output Voltage (V)
90
Efficiency (%)
16
Output Load (A)
Output Power (W)
85
80
75
25
24
23
22
115 Vrms
70
21
230 Vrms
65
20
0
100
200
300
400
0
2
4
6
10
12
14
16
Figure 4: 24 VOUT droop characteristic.
Figure 3: 24 VOUT efficiency curves.
42
100
95
40
Output Voltage (V)
90
Efficiency (%)
8
Output Load (A)
Output Power (W)
85
80
75
115 Vrms
38
36
34
32
70
230 Vrms
65
30
0
100
200
300
400
0
Figure 5: 36 VOUT efficiency curves.
Product # AQ0300MUxx Series
Phone 1-888-567-9596
1
2
3
4
5
6
7
8
9
10
11
Output Load (A)
Output Power (W)
Figure 6: 36 VOUT droop characteristic.
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 3
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
EFFICIENCY, DERATING, AND VOUT DROOP CURVES
100
54
95
52
Output Voltage (V)
Efficiency (%)
90
85
80
75
115 Vrms
50
48
46
44
70
230 Vrms
65
42
0
100
200
300
400
0
1
2
3
Output Power (W)
4
5
6
7
Output Load (A)
Figure 8: 48 VOUT droop characteristic.
Figure 7: 48 VOUT efficiency curves.
500
400
400
Power (W)
Power (W)
300
200
100
300
200
100
Fan Enabled
Transient
Continuous
Fan Disabled
0
0
0
10
20
30
40
50
60
70
132
170
264
85
80
Input AC Voltage (VRMS)
Ambient Temperature (ºC)
Figure 9: Continuous power derating curve in natural convection.
Figure 10: Rated output power vs Input AC Voltage.
Leakage Currents
AC Leakage Current
from Input to Earth
AC Line Connection
240 V L-N, 1 phase
ACuQor Typical at
110% nominal input 208 V L-L, 120 V L-N, 1 of 3 phases
voltage 60 Hz
240 V L-N-L, 120 V L-N, split phase
Normal Open Neutral
Condition
Fault
125 µA
250 µA
65 µA
130 µA
65 µA
130 µA
AC Leakage Current
from Output to Earth
ACuQor Typical at
264 Vac 60 Hz input
Model
Normal
Condition
Open Earth
Fault
AC Backdrive
Fault
AQ . . BF . .
2 µA
36 µA
125 µA
AQ . . CF . .
2 µA
6 µA
18 µA
Contact Type
Normal
Condition
Open Earth
Fault
AC Backdrive
Fault
BF
100 µA
500 µA
5000 µA
CF
10 µA
50 µA
50 µA
For convenience, the following tables show limits allowed by various standards:
AC Leakage Current
from Input to Earth
Maximum Allowed
per Standard
Standard
Normal Open Neutral
Condition
Fault
IEC60601-1
500 µA
1000 µA
NFPA 99 2005
300 µA
—
IEC60950
3500 µA
—
Product # AQ0300MUxx Series
Phone 1-888-567-9596
AC Leakage Current
from Output to Earth
Maximum Allowed per
IEC60601-1
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 4
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
MECHANICAL DRAWINGS (1 Module Open Frame Version — E Package Type)
Front
Back
Top
Bottom
Right
MECHANICAL DRAWINGS (1 Module Encased Version — E Package Type)
Top
Right
Bottom
Front
Back
NOTES (applies to all mechanicals)
1) Recommended screw tightening torque of 6 in.lbs
2) Undimensioned components are shown for visual reference only
3) All dimensions in inches [mm]
Tolerances: x.xx in ± 0.02
x.xxx in ± 0.010
Product # AQ0300MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 5
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
MECHANICAL DRAWINGS (2 Modules Flat Version — R Package Type)
Product # AQ0300MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 6
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
MECHANICAL DRAWINGS (2 Modules Stacked Version — S Package Type)
Product # AQ0300MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 7
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
MECHANICAL DRAWINGS (3 Modules Flat Version — T Package Type)
Product # AQ0300MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 8
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
MECHANICAL DRAWINGS (3 Modules Stacked Version — U Package Type)
Product # AQ0300MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 9
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
CONNECTOR DETAILS
OUTPUT CONNECTOR PINOUT (top side)
Pin 1
FAN_GOOD
Open collector with internal 5V pullup. See Figure A.
Pulsed low on fan failure, 100ms, 50% duty.
Short to VOUT(-) to disable fan.
Pin 2
AC_POWER_GOOD
Open collector with internal 5V pullup. See Figure B.
Pulled low on AC power dropout.
Pin 3
DC_POWER_GOOD
Open collector with internal 5V pullup. See Figure B.
Pulled low during startup ramp and within 5 ºC of
temperature shutdown threshold.
Pin 4
5V_STANDBY
5 V @ 50 mA available whenever AC power is applied.
Pin 5
VOUT(+)
Positive Output Voltage.
Pin 6
VOUT(+)
Positive Output Voltage.
Pin 7
VOUT(+)
Positive Output Voltage.
Pin 8
VOUT(+)
Positive Output Voltage.
Pin 9
Reserved
Reserved for future use.
Pin 10 Reserved
Reserved for future use.
Pin 11 REMOTE_ENABLE
Figure A: Fan status output / Fan enable input interface circuitry.
5V_STANDBY
1N4148
5k
10 Ohms
AC_POWER_GOOD or
DC_POWER_GOOD
Logic input. See Figure C.
Pull high to enable main output.
Pin 12 VOUT(–)
Negative Output Voltage.
Pin 13 VOUT(–)
Negative Output Voltage.
Pin 14 VOUT(–)
Negative Output Voltage.
Pin 15 VOUT(–)
Negative Output Voltage.
Pin 16 VOUT(–)
Negative Output Voltage.
2N3904
From AC/DC
Monitor
Circuitry
0.1 µF
VOUT(-)
Figure B: Power good interface circuitry.
12 V OUTPUT CONNECTOR PINOUT (bottom side)
Pin 1
VOUT(+)
Positive Output Voltage.
Pin 2
VOUT(+)
Positive Output Voltage.
Pin 3
VOUT(+)
Positive Output Voltage.
Pin 4
VOUT(+)
Positive Output Voltage.
Pin 5
VOUT(–)
Negative Output Voltage.
Pin 6
VOUT(–)
Negative Output Voltage.
Pin 7
VOUT(–)
Negative Output Voltage.
Pin 8
VOUT(–)
Negative Output Voltage
5V_STANDBY
5V_STANDBY
10 Ohms
5k
To Enable
Circuitry
REMOTE_ENABLE
0.1 µF
10k
VOUT(-)
Figure C: Remote enable interface circuitry.
MATING CONNECTORS
Connector
Type
Contact
INDIVIDUAL INPUT CONNECTOR PINOUT
OUTPUT (16 pins)
Molex 430251600
Molex 430300008*
Pin 1
Ground
12V_OUTPUT (8 pins)
Molex 436450800
Molex 430300008*
Pin 3
AC Neutral
INPUT
JST VHR-5N
JST SVH-41T-P1.1
Pin 5
* Each contact rated for a maximum of 5.5 A.
AC Line
Product # AQ0300MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 10
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
PARALLEL OPERATION - MULTIPLE UNITS
Chassis configurations for Parallel Units
ACuQor units are available either open-frame or pre-mounted
at the factory. Up to 3 units can be mounted into a chassis, in
either side-by-side or stacked configurations. For a complete list of
options, see the “Part Numbering System” table on the last page,
under “Package Type”, along with the Mechanical Drawings pages.
Only side-by-side configurations can be populated with 500W units,
since each includes a thermal pad underneath.
Interconnection of Parallel Units
ACuQor units mounted in 2 and 3 unit chassis are not connected
together. This allows the physical routing and connectivity of the
external wiring to be customized to each application. The following
table summarizes the recommended wiring to operate multiple
units in parallel:
Specifications of Parallel Units
As a rule, units wired in parallel behave the same as single units.
Any specification will remain unchanged that is expressed in units
of voltage, time, frequency, or efficiency. Specifications expressed
in terms of power, current, or capacitance, should be scaled by the
number of units wired in parallel.
ACuQor units are individually calibrated at the factory, so that the
output voltage vs. output current characteristic is always consistent
(see Vout droop characteristic figures). As such, multiple units will
share output current accurately. Full current is guaranteed from a
bank of multiple units wired in parallel.
Output Connector Signal
Suggested Connection
Behavior with Multiple Units
REMOTE_ENABLE
Wire in parallel
Inputs activated simultaneously
FAN_GOOD
“
Wired-OR outputs – can be pulled low by any unit during an abnormal condition.
AC_POWER_GOOD
“
“
DC_POWER_GOOD
“
“
VOUT(+), VOUT(-)
“
Built-in droop characteristic ensures graceful current sharing.
12V_STANDBY*
“
“
5V_STANDBY
Do not wire in parallel
Fully regulated characteristic does not support current sharing. If placed in
parallel, only the output with the highest set-point will drive current.
*Note: Triple output models only.
Product # AQ0300MUxx Series
Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 11
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
INSTALLATION INSTRUCTIONS
General: ACuQor AC/DC power supplies are intended for use
as components in medical and industrial equipment. ACuQor
units must be properly installed within end use equipment before
they can be safely applied as described in this document. The
suitability of the ACuQor/equipment combination must be verified
through end product investigation.
Mounting: Refer to the Mechanical Drawings section. ACuQor
units are provided with threaded stainless-steel stand-offs or
inserts for mounting. This mounting hardware is internally connected to the input connector protective-earth terminal for functional-earth EMC control. Any orientation (vertical, horizontal,
etc.) may be used. Adequate air space should be provided over
the fan intake (top) and exhaust (sides) to allow for exchange of
cooling air. ACuQor is designed for a pollution degree 2 environment. The suitability of the enclosed ACuQor mechanical assemblies must be verified through end product investigation.
Encased models: A minimum of 5 mm electrical clearance
should be allowed from the connector ends of encased models.
Input: Refer to the Connector Details section for input connector
wiring. ACuQor products require a single phase AC power source
of 100-240V 50/60Hz nominal. Refer to nameplate label for input
current ratings. A protective-earth connection is also required.
Minimum wire size of 18 AWG (0.8mm2) is recommended. Both
sides of the AC line are internally fused (see table for specific
models). These fuses are not user replaceable.
MODEL
AQ0300
AQ0400
AQ0500
Input Fuses (in Both AC Lines)
Littelfuse 6.3A 250V 21606.3XEP
Littelfuse 6.3A 250V 21606.3XEP
Littelfuse 10.0A 250V 216010XEP
Product # AQ0300MUxx Series
Phone 1-888-567-9596
Fuses Total
2
2
2
Output: Refer to the Connector Details section for output con-
nector wiring and signal I/O functionality. Refer to nameplate
label for output current ratings. Main DC output (Vout+, Vout-)
pins should use 20 AWG (0.5mm2) wire size. Individual main
output pins should not be loaded to more than 5.5 A. For currents greater than 5.5 A, multiple main output pins/wires must be
used in parallel. All signal I/O pins are referenced to Vout-.
EMC: ACuQor products have been tested to the EMC specifications listed in the Electrical Characteristics section. However, end
use equipment must be tested to verify EMC compliance..
Hipot Testing: ACuQor products are rated for Hipot testing
levels of 1500 Vac input to protective-earth, 1500 Vac output to
protective-earth, and 4000 Vac input to output. When performing
the 4000 Vac input to output test, the test voltage must be balanced evenly 2000 Vac input and output to protective-earth. Two
oppositely phased test voltage sources or a single test voltage
source with external balancing impedances (capacitors) may be
used to prevent overstressing input or output to protective-earth
insulation per IEC60601-1 2005 sub clause 8.8.1 and IEC60601-1
1990 sub clause 20.4g.
Patient Contact: ACuQor models include versions designed
for BF and CF patient contact application per IEC60601-1. These
ACuQor models provide reinforced insulation at the DC output
voltage level and basic insulation at the 240 Vac level from output
to protective-earth. Note that equipment and wiring may add to
system leakage currents so that the end product must be tested
for compliance. Refer to the Electrical Characteristics section for
typical ACuQor input and output leakage currents. In addition,
ACuQor defibrillation rated models comply with the minimum
output to protective-earth creepage/clearance requirement and
defibrillator pulse test of IEC60601-1.
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 12
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
Technical Specification
PART NUMBERING SYSTEM
The part numbering system for SynQor’s ACuQor AC/DC power supplies follows the format shown in the table below.
Not all combinations make valid part numbers, please contact SynQor for availability.
Family
AQ
ACuQor series
of ac-dc semi-regulated
output power supplies
Output Power
0300: 300 W
0400: 400 W
0500: 500 W
0600: 600 W (2 x 300 W)
0800: 800 W (2 x 400 W)
0900: 900 W (3 x 300 W)
1000: 1000 W (2 x 500 W)
1200: 1200 W (3 x 400 W)
1500: 1500 W (3 x 500 W)
Grade
M: medical
Range
U: universal
(85-264 VRMS)
Output Voltage
Package Type
(xx=Standard; xT=Triple)
(Correlates to Output Power)
12: 12 V
1T: 12 V / 5 & 12 V STBY
24: 24 V
2T: 24 V / 5 & 12 V STBY
36: 36 V
3T: 36 V / 5 & 12 V STBY
48: 48 V
4T: 48 V / 5 & 12 V STBY
E: 1 unit (3" x 5")
R: 2 units; flat
S: 2 units; stacked
T: 3 units; flat
U: 3 units; stacked
Thermal
Design
Options
A: open frame
C: encased
Q: encased only
Medical Grade:
BF: BF isolation rating
CF: CF isolation rating
CFD: CF isolation rating
defibrillator proof
Example: AQ0300MU12EACFD
ACCESSORIES
SynQor offers a series of assemblies that can be ordered according to the table below. Mechanical drawings for these accessories are
available for download in pdf format from the SynQor website.
Part Number
Description
AQ-CBL-INPUT1C
Input mating cable with pre-stripped wire ends (36" long).
AQ-CBL-OUT1C
Output mating cable with pre-stripped wire ends (18" long).
AQ-CBL-OUT1CD
Same as AQ-CBL-OUT1C with an additional 8-pin connector.
AQ-CBL-OUT2C
AQ-CBL-OUT2CD
AQ-INSUL1M
AQ-EVAL-PRL3
Output mating cable with connectors on both ends (18" long).
Same as AQ-CBL-OUT2C with an additional 8-pin connector.
Single module bottom-side Mylar insulator for open frame mounting.
Evaluation board for up to three paralleled modules.
APPLICATION NOTES
A variety of application notes and technical white papers can be downloaded in pdf format from the SynQor website.
PATENTS
SynQor holds the following patents, one or more of which might apply to this product:
5,999,417
6,222,742
6,545,890
6,594,159
6,731,520
6,894,468
6,927,987
7,050,309
7,072,190
7,119,524
7,269,034
7,272,021
Contact SynQor for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
978-849-0600
888-567-9596
978-849-0602
[email protected]
www.synqor.com
155 Swanson Road
Boxborough, MA 01719, USA
Product # AQ0300MUxx Series
Phone 1-888-567-9596
6,577,109
6,896,526
7,085,146
7,272,023
Warranty: SynQor offers a two (2) year limited warranty. Complete warranty information is listed on our website or is available upon request from SynQor.
Limitation on use: SynQor’s products are not authorized for use and should not be
used, without the specific prior written approval of an authorized officer of SynQor
making specific reference hereto, in any life support system, nuclear facility or application, aircraft control application or any other application in which failure of the product
could reasonably result in harm to life, property or the environment.
Information furnished by SynQor is believed to be accurate and reliable. However,
no responsibility is assumed by SynQor for its use, nor for any infringements of patents
or other rights of third parties which may result from its use. No license is granted by
implication or otherwise under any patent or patent rights of SynQor.
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 13