ASC DS-0041 Barometrici pressure sensor Datasheet

barometric pressure sensors
Barometrici Pressure Sensors
600 to 1,100 mbar Barometric Pressure
Options:
all sensors
• Ratiometric Output
• Fixed 4V Output with internal Voltage Reference
• Temperature Compensated (-25C to 85C, -40C to 125C, 5 C
to 50C)
• Calibrated Zero and Span
Applications
• Medical Instrumentation
• Environmental Controls
• Chemical Analysis
e www.allsensors.com
General Description
These Barometric pressure sensors are based upon a proprietary package technology to reduce errors. This
model provides a fixed 4-volt output with superior output characteristics. The sensor housing has been designed
specifically to reduce package induced parasitic stress and strain. In addition the sensor utilizes a silicon,
micromachined (MEMS) structure to provide a very linear output to measured pressure. The sensor embodies a
f 408 225 2079
vacuum reference.
These calibrated and temperature compensated sensors give an accurate and stable output over a wide
temperature range. Each sensor is internally compensated using an ASIC compensation technique. This series is
a 16035 Vineyard Blvd. Morgan HIll, CA 95037 p 408 225 4314
intended for use with non-corrosive, non-ionic working fluids such as air, dry gases and the like.
Physical Dimensions
all sensors
DS-0041 Rev A
01
Pressure Sensor Ratings
Environmental Specifications
Lead Temperature, max
(soldering 2-4 sec.)
250°C
Burst Pressure
60 psi
Temperature Ranges
Compensated
see below
Operating
-40 to 125° C
Storage
-40 to 125° C
Humidity Limits
0 to 95% RH
(non condensing)
Standard Pressure Ranges
Part Number
Operating Pressure
Supply Voltage
Compensated Temperature Range
BARO-A-4V
600 - 1100 mbar
4.5 to 5.5 V
5C to 50C
BARO-A-4V-PRIME
600 - 1100 mbar
4.5 to 5.5 V
-25C to 85C
BARO-A-4V-MIL
600 - 1100 mbar
4.5 to 5.5 V
-40C to 125C
BARO-A-4V-PRIME-REF
600 - 1100 mbar
5.5 to 16 V
-25C to 85C
Performance Characteristics for BARO-A-4V
Parameter, note 1
Minimum
Nominal
Maximum
Units
Operating Range, absolute pressure
600
1100
mBar
Output Voltage @1,100 mbar
4.20
4.25
4.30
volt
Output Voltage @ 600 mbar
0.20
0.25
0.30
volt
±20
mvolt
0.05
0.25
%fs
±1
%span
Maximum
Units
1100
mBar
Offset Temperature Shift (5°C to 50°C), note 2
Linearity, hysteresis error, note 4
Span Shift (5°C to 50°C), note 2
Performance Characteristics for BARO-A-4V-PRIME
Parameter, note 1
Minimum
Nominal
Operating Range, absolute pressure
600
Output Voltage @ 1,100 mbar
4.20
4.25
4.30
volt
Output Voltage @ 600 mbar
0.20
0.25
0.30
volt
±20
mvolt
0.25
%fs
±1
%span
Offset Temperature Shift (-25°C to 85°C), note 2
Linearity, hysteresis error, note 4
Span Shift (-25°C to 85°C), note 2
02
0.05
barometric pressure sensors amplified output
Nominal
Maximum
Units
1100
mBar
Operating Range, absolute pressure
600
Output Voltage @ 1,100 mbar
4.20
4.25
4.30
volt
Output Voltage @ 600 mbar
0.20
0.25
0.30
volt
±20
mvolt
0.25
%fs
±1
%span
Maximum
Units
1100
mBar
Offset Temperature Shift (-40°C to 125°C), note 2
Linearity, hysteresis error, note 4
0.05
Span Shift (-40°C to 125°C), note 2
Performance Characteristics for BARO-A-4V-PRIME-REF
Parameter, note 1
Minimum
Nominal
600
Output Voltage @ 1,100 mbar
4.20
4.25
4.30
volt
Output Voltage @ 600 mbar
0.20
0.25
0.30
volt
±20
mvolt
0.25
%fs
±1
%span
Offset Temperature Shift (-25°C to 85°C), note 2
0.05
Linearity, hysteresis error, note 4
Span Shift (-25°C to 85°C), note 2
a 16035 Vineyard Blvd. Morgan HIll, CA 95037 p 408 225 4314
Operating Range, absolute pressure
Equivalent Circuit with
Internal Voltage Reference
Equivalent Circuit
Specification Notes
NOTE 1:
ALL PARAMETERS ARE MEASURED AT 5.0 VOLT EXCITATION FOR RATIOMETRIC PARTS AND 12.0 FOR PARTS WITH INTERNAL REFERENCE, FOR
THE NOMINAL FULL SCALE PRESSURE AND ROOM TEMPERATURE UNLESS OTHERWISE SPECIFIED.
2: SHIFT IS RELATIVE TO 25°C.
3: SHIFT IS WITHIN THE FIRST HOUR OF EXCITATION APPLIED TO THE DEVICE.
NOTE 4: MEASURED AT ONE-HALF FULL SCALE RATED PRESSURE USING BEST STRAIGHT LINE CURVE FIT.
NOTE
NOTE
All Sensors reserves the right to make changes to any products herein. All Sensors does not assume any liability arising out of the application or
use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
all sensors
DS-0041 Rev A
e www.allsensors.com
Minimum
f 408 225 2079
Parameter, note 1
all sensors
Performance Characteristics for BARO-A-4V-MIL
03
Similar pages