TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply MDS-350AD701 Highlights & Features – ATX 12V standard function & form factor – IEC60601-1 3.1 ed.(Medical) & IEC 60950-1(ITE) approvals – Suitable for type BF patient access applications – Built-in active PFC – Maximum output power of 350 watts th – Compliant with IEC 60601-1-2 4 Ed. requirements Safety Standards CB Certified for worldwide use Model Number: MDS-350AD701 AA Unit Weight: 1.2Kg (2.6 lb) Dimensions (W x L x H): 100x194x40.5 mm 3.94x7.64x1.59 inch General Description The MDS-350AD701 series of internal ATX power supplies comes with universal AC input from 90Vac to 264Vac. Other features include low leakage, availability of risk management report, and electric shock protection compliant with 2 x MOPP requirements. The MDS series is certified for EMC standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment; and, EN 55022 for Industrial Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used to ensure long product life. The MDS-350AD701 series comes with both Medical and ITE safety approvals, including UL/CE/CQC (5000m) and CB certification and are fully compliant with RoHS Directive 2011/65/EU for environmental protection. Model Information Table 1 (For Maximum Output Power Limits, see “Output Power” requirements on Page 2) Output Voltage Minimum Load (A) Normal Load (A) Maximum Load (A) Load Regulation Ripple & Noise +3.3V +5V +5Vsb +12V1 +12V2 -12V -5V 0.5 0.5 0.0 1.0 1.0 0.0 0.0 8 9 1.25 8 8 0.25 0.1 16 18 2.5 16 16 0.5 0.2 +/- 5% +/- 5% +/- 5% +/- 5% +/- 5% +/- 10% +/- 10% 50mVpp 50mVpp 150mVpp 120mVpp 120mVpp 120mVpp 50mVpp Model Numbering MDS 350 Delta Medical Power Supply 1 A Family Code A~ Z D 701 Output Voltage 701: +3.3V, +5V, +5Vsb, +12V1, +12V2, -12V, -5V Max Wattage in Product Type D: Enclosure the Product Series. Maybe lower at some voltage. 060 → 60W All parameters are specified at 25°C ambient unless 150 →150W 1K2 → 1,200W www.DeltaPSU.com (January 2016, Rev. 1.1) otherwise noted. A Revision Code A Revision code TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply Specifications Input Ratings / Characteristics Nominal Input Voltage 100-240Vac Input Voltage Range 90-264Vac Nominal Input Frequency 50-60Hz Input Frequency Range 47-63Hz Input Current (max) 6A Input Surge Voltage (max) 300Vac for 100ms Full load Efficiency (typ.) 82% @ 115Vac/60Hz 85% @ 230Vac/50Hz Inrush Current (max) No damage, cold start Leakage Current (max) 0.1mA @ NC, 0.3mA @ SFC Output-PE (protective earth) Leakage current for BF application (max) 0.1mA @ NC, 0.5mA @ SFC Power Factor (min) 0.90 @ 115V/50Hz, 230V/50Hz, full load 1) 1) NC: normal condition, SFC: single fault condition Output Ratings / Characteristics Nominal Output Voltage (V rated) +12V1, +12V2, +5V, +3.3V, +5Vsb, -5V, -12V Output Voltage Tolerance See Table 1 on page 1 Output Power Maximum combined output power of +3.3V & +5V outputs, shall not exceed 130W. Maximum combined output power of +12V1, +12V2, +3.3V & +5V outputs shall not exceed 331W Maximum combined output power from all the outputs shall not exceed 350W ±0.5% Line Regulation (max) Load Regulation (max) See Table 1 Ripple & Noise (typ.) Start-up Time (max) (Remote on_off @ TTL Low, then AC applied) 1000ms @ 115 Vac Hold-up Time (min) 17ms @ 115 Vac / 60Hz Dynamic Response (Overshoot & Undershoot O/P Voltage) ± 10% @ 50-100% load Rise time (max) 20ms Remote On_Off/INHIBIT TTL Active Low: Output Enable TTL Active High: Output Disable Power Good (PG) Signal(See page 9 for details) 1. 2. PG changes to TTL High 100-500 milliseconds after all outputs reach their final steady state value. After input AC is removed, PG changes to TTL Low at least 1 millisecond before any output falls out of regulation range. 1) When the dynamic load is applied to one of the outputs, other loads is at minimum load Ripple & Noise measurement circuit for each output 2 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (January 2016, Rev. 1.1) 1) TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply Mechanical Case Chassis SECC (Steel, Electrogalvanized, Cold Rolled, Coil) Case Cover SECC (Steel, Electrogalvanized, Cold Rolled, Coil) Dimensions (W x L x H) 100 x 194 x 40.5 mm (3.94x7.64x1.59 inch) Unit Weight (typ.) 1.2Kg (2.6 pounds) Indicator None Cooling System Forced Cooling Environment Surrounding Air Temperature Operating Storage 0°C to +50°C -40°C to +85°C Operating Humidity 5-95% RH (Non-Condensing) Operating Altitude 5,000 meters (16,400 feet) Shock Test (Non-Operating) 50G, 11ms, 3 shocks for each direction Vibration (Operating) 5-500Hz, 2Grms, 15 minute for each three axis 3 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (January 2016, Rev. 1.1) TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply Protections (Latch Mode protection requires removal of fault, plus removal and reapplication of input AC voltage, in order to restart the power supply. See page 9 for additional details on the different types of protection.) Overvoltage (max) For +3.3V output, Maximum voltage < 4.8V For +5V output, Maximum voltage < 7V For +12V1 & +12V2 outputs, Maximum voltage < 16V Protection Method: Latch Mode For +5Vsb output, Maximum voltage < 10V Protection Method: Latch Mode Over load / Over current (max) For +3.3V output, Maximum current < 25A For +5V output, Maximum current < 25A For +12V1/+12V2 output, Maximum current < 25A Protection Method: Latch Mode For +5Vsb output, Maximum current < 5A Protection Method: Auto-recovery Mode Over Temperature Protection Method: Latch Mode Short Circuit on any output. +3.3V, +5V, +12V1 and +12V2, Protection Method: Latch Mode +5Vsb, Protection Method: Auto-recovery Reliability Data MTBF (Minimum) at 115Vac, 35 °C with 350W Forced Air Flow 500 K hrs based on Telecordia SR-332 Operating Life (Minimum) at 115Vac, 25°C with 350W Forced Air Flow 26,280 hrs Safety Standards / Directives Medical Safety IEC 60601-1 Edition 3.1 (2012), EN60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012 ITE Safety IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB92542008, GB17625.1-2003 CE MDD Directive 93/42/EEC Environmental RoHS Directive 2011/65/EU Compliant Galvanic Isolation Input to/Output(2XMOPP) 4000 Vac 2) Input to/Ground(1XMOPP) 1500 Vac Output to/Ground(1XMOPP) 1500 Vac (Type BF application rated) 2) PSU can support PoE applications with Primary to FG 2500Vac test. EMC EMC / Emissions Harmonic Current Emissions EN55011/EN55022,FCC Title 47:Class B IEC61000-3-2 Meet Class D limit Immunity to Voltage Flicker 4 IEC61000-3-3 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (January 2016, Rev. 1.1) 5 0 0 k H r s b a s e d o n T e l e c o r d i a S R 3 3 2 ( T B D ) TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply 1)5) Electrostatic Discharge IEC61000-4-2 Level 4 Criteria A Air Discharge: 15kV Contact Discharge: 8kV Radiated Field IEC61000-4-3 Level 3 Criteria A 80MHz-1000MHz, 10V/m AM modulation IEC60601-1-2 Criteria A 80MHz-2700MHz, 10V/m AM modulation 385MHz-5785MHz, 28V/m Pulse mode and other modulation Electrical Fast Transient / Burst IEC61000-4-4 Level 3 Criteria A Surge IEC61000-4-5 Level 3 Criteria A 3) Common Mode : 2kV 4) Differential Mode : 1kV Conducted IEC61000-4-6 Level 2 Criteria A 150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and Amateur radio bands Power Frequency Magnetic Fields IEC61000-4-8 Criteria A Magnetic field strength 30A/m IEC60601-1-2 Criteria B C 30% 500ms 1) Criteria A 100% 10ms at step: 45° degree step 45 2) Criteria B C 100% 20ms Criteria B C2) 100% 5000ms Voltage Dips 5) 1) 1)5) 1) :2kV 1)5) 1)5) 1)5) 2) 1) Criteria A: Normal performance within the specification limits 2) Criteria C: Output out of regulation, or shuts down during test. (Need to recycle AC power cord or reset enable signal to normal operation after test) 3) Asymmetrical: Common mode (Line to earth) 4) Symmetrical: Differential mode (Line to line) 5) Compliant with IEC-60601-1-2 4th edition requirements. 5 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (January 2016, Rev. 1.1) IE TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply Dimensions W x L x H: 100 x 194 x 40.5 mm Mechanical drawing Notes ― Dimensions are in mm 6 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (January 2016, Rev. 1.1) TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply DC Output Cable 7 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (January 2016, Rev. 1.1) TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply Functions Start-up Time The time required for the output voltage to reach 90% of its final steady state value, after the input voltage is applied. Rise Time The time required for the output voltage to change from 10% to 90% of its final steady state value. Hold-up Time Hold-up Time is the time between removal of input AC voltage, and the subsequently falling output to reach 90% of its steady state value. 8 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (January 2016, Rev. 1.1) TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply Dynamic Response The power supply output voltage will remain within ±10% of its steady state value, when subjected to a dynamic load 50 to 100% of its rated current. 50 to 100% Load their maximum power limit, the protection will be activated and the power supply will latch off. Over Temperature Protection As mentioned above, the power supply also has Over Temperature Protection (OTP). This is activated when the overload condition persists for an extended duration and the output current is below the overload trigger point but >100% load. In the event of a higher operating temperature condition at 100% load, the power supply will run into OTP when the surrounding air temperature is higher than the operating temperature. When activated, the output voltage will go into latch mode until the input voltage is removed; then, reapplied, and the surrounding air temperature drops to its normal operating temperature. Power Good Inrush Current Inrush current is the peak, instantaneous, input current measured and, occurs when the input voltage is first applied. For AC input voltages, the maximum peak value of inrush current will occur during the first half cycle of the applied AC voltage. This peak value decreases exponentially during subsequent cycles of AC voltage. The Power Good signal prevents a computer from attempting to operate on improper voltages and damaging itself by alerting it to improper power supply. The ATX specification defines the PowerGood signal as a +5 volt (V) signal generated in the power supply when it has passed its internal self-tests and the outputs have stabilized. This normally takes between 0.1 and 0.5 seconds after the power supply is switched on. The signal is then sent to the motherboard, where it is received by the processor timer chip that controls the reset line to the processor. When plug in AC power cord, +5Vsb will start first, then Remote On/Off will be pulled low by computer, +12V1/+12V2/+5V/+3.3V will power on and Power good signal will pull high after 100ms~500ms. After input AC is removed, Power Good signal changes to TTL Low at least 1 millisecond before any output falls out of regulation range. Graph illustrating the AC input, +5Vsb, Remote On/Off, +12V1/12V2/5/3.3V, Power good signal Overvoltage Protection The power supply’s overvoltage circuit will be activated when its internal feedback circuit fails. The output voltage shall not exceed its specifications defined on Page 4 under “Protections”. Power supply will latch off, and require removal/re-application of input AC voltage in order to restart. Short Circuit Protection The power supply’s output OLP/OCP function also provides protection against short circuits. When a short circuit is applied to any of the outputs, the power supply will latch off. Overload & Over current Protections The power supply’s Overload (OLP) and Over current (OCP) Protections will be activated before the output currents reach the limits defined on Page 4. Once the power supply has reached 9 Remote On_Off/INHIBIT Remote ON_OFF/INHIBIT can be used to enable or disable all outputs, except for the +5 Volt Standby. This signal can be pulled down to a low level of 0.3 volts, or shorted to DC-Return, in order for the main output to be enabled; and, floated (no connection to the signal), or pulled up to a value greater than or equal to 3 volts, in order to disable the main output. When the main output is disabled, the +5V Standby output will continue to operate. All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (January 2016, Rev. 1.1) TECHNICAL DATASHEET Medical ATX Power Supply ATX 12V 1U Power Supply Certificate All Delta Medical Power products conform to the European directive 2011/65/EU. RoHS is the abbreviation for “Restriction of the use of certain hazardous substances. Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN ISO 13485:2012 for the design and manufacture of switching power supply and adaptor for medical device. In addition to a UL Total Certification Program (TCP) approved client laboratory for IEC60950 and IEC60065. Delta also has participated UL Client Test Data Program (CDTP) for IEC 60601. 10 All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (January 2016, Rev. 1.1)