Astrodyne ASM-0100 10w ultraminiature medical grade switching power supply Datasheet

1 0 W U l t r a m i n i a t ure Medical Grade Switching Power Supplies
asm10
series
Universal 90-264VAC Input
High Efficiency
Regulated Outputs
4000V Isolation
Single Outputs
UL60601-1, EN60601-1 & IEC60601-1
Single Output from 3.3 to 24 VDC
2”W x 2.28”L x 0.66”H
Model Number
Output Voltage
Output Amps (max)
(pk)
ASM-0100
ASM-0101
ASM-0102
ASM-0103
ASM-0105
3.3 VDC
5 VDC
12 VDC
15VDC
24VDC
2.5A
2A
0.85A
0.67A
0.42A
3.3A
2.6A
1.2A
0.9A
0.6A
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
asm10
1 0 W U l t r a m i n i a t ure Medical Grade Switching Power Supplies
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Inrush Current, typ:
series
PHYSICAL SPECIFICATIONS
90-264 VAC
47-63 Hz
20A @ 100VAC Input *
40A @ 200VAC Input *
Size
Construction
Weight
Connectors: Molex P/N:
2” x 2.28” x 0.66”
Open Frame
1.6oz (44g)
50-37-5033 (Input)
50-37-5023 (Output)
OUTPUT SPECIFICATIONS
ENVIRONMENTAL SPECIFICATIONS
Voltage and Current
Load Regulation
Line Regulation
Preset Accuracy
DC Voltage Adjust (typ)
Temperature Coefficient
Ripple/Noise
Over Voltage Protection
Over Temp. Protection
Short Circuit Protection
Hold Up Time
See Selection Chart
+/- 3%
(20% to 100% Load)
Singles +/- 0.1~1%, typ
1%
+/- 6% of FS (Primary)
+/-0.03%/°C
100mV Pk-Pk, typ
Clamp, 130-150% *
Latching, Cool and Repower
Continuous, self-recovering
20 mS, typ
Oper. Temperature
Storage Temperature
Relative Humidity
EMC
MTBF
0 to +50°C(FL)
-25 to +71°C *
0 to + 95%, non-cond *
FCC/A, VDE 0871/A
180,000 Hrs
Mil Std 217, 25°C
OUTPUT DERATING CURVE
100
90
80
GENERAL SPECIFICATIONS
I/P-O/P: 4000VAC
I/P-Ground: 2500VAC
O/P-Ground: 1000VAC
75%~78%, typ
100Khz, (fixed, typical)
UL60601-1, EN60601-1, IEC60601-1
Load (%)
Isolation
Efficiency
Switching Frequency
Safety
70
60
50
40
30
20
10
0
0
10
20
30
40
50
60
70
Ambient Temperature DegC
All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
WWW.ASTRODYNE.COM
Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 0 W U l t r a m i n i a t ure Medical Grade Switching Power Supplies
asm10
series
MECHANICAL DIMENSIONS
0.1
0.12” Dia. Mtg Holes (4 plcs)
2.0
10 Watt
0.68
0.1
2.28
J1 (AC Input)
Pin 1
Pin 3
AC Neutral
AC Line
NOTE:
For Class II operation (no earth
Gnd), insulating type hardware must
be used at this mounting hole
J2 (DC Output)
Pin 1
Pin 3
+Vout
Return
WWW.ASTRODYNE.COM