4.8 kHz Ultralow Noise 24-Bit Sigma-Delta ADC with PGA AD7190 Data Sheet FEATURES Temperature measurement Chromatography PLC/DCS analog input modules Data acquisition Medical and scientific instrumentation RMS noise: 8.5 nV @ 4.7 Hz (gain = 128) 16 noise free bits @ 2.4 kHz (gain = 128) Up to 22.5 noise free bits (gain = 1) Offset drift: 5 nV/°C Gain drift: 1 ppm/°C Specified drift over time 2 differential/4 pseudo differential input channels Automatic channel sequencer Programmable gain (1 to 128) Output data rate: 4.7 Hz to 4.8 kHz Internal or external clock Simultaneous 50 Hz/60 Hz rejection 4 general-purpose digital outputs Power supply AVDD: 4.75 V to 5.25 V DVDD: 2.7 V to 5.25 V Current: 6 mA Temperature range: –40°C to +105°C Interface 3-wire serial SPI, QSPI™, MICROWIRE™, and DSP compatible Schmitt trigger on SCLK Qualified for automotive applications GENERAL DESCRIPTION The AD7190 is a low noise, complete analog front end for high precision measurement applications. It contains a low noise, 24-bit sigma-delta (∑-Δ) analog-to-digital converter (ADC). The on-chip low noise gain stage means that signals of small amplitude can be interfaced directly to the ADC. The device can be configured to have two differential inputs or four pseudo differential inputs. The on-chip channel sequencer allows several channels to be enabled, and the AD7190 sequentially converts on each enabled channel. This simplifies communication with the part. The on-chip 4.92 MHz clock can be used as the clock source to the ADC or, alternatively, an external clock or crystal can be used. The output data rate from the part can be varied from 4.7 Hz to 4.8 kHz. APPLICATIONS The device has two digital filter options. The choice of filter affects the rms noise/noise-free resolution at the programmed output data rate, the settling time, and the 50 Hz/60 Hz rejection. For applications that require all conversions to be settled, the AD7190 includes a zero latency feature. Weigh scales Strain gauge transducers Pressure measurement The part operates with 5 V analog power supply and a digital power supply from 2.7 V to 5.25 V. It consumes a current of 6 mA. It is housed in a 24-lead TSSOP package. FUNCTIONAL BLOCK DIAGRAM AGND AVDD DVDD DGND REFIN1(+) REFIN1(–) REFERENCE DETECT AD7190 AIN1 AIN2 AIN3 AIN4 AINCOM MUX Σ-∆ ADC PGA SERIAL INTERFACE AND CONTROL LOGIC DOUT/RDY DIN SCLK CS SYNC TEMP SENSOR P3 P2 BPDSW AGND MCLK1 MCLK2 P0/REFIN2(–) P1/REFIN2(+) 07640-001 CLOCK CIRCUITRY Figure 1. Rev. C Document Feedback Information furnished by Analog Devices is believed to be accurate and reliable. However, no responsibility is assumed by Analog Devices for its use, nor for any infringements of patents or other rights of third parties that may result from its use. Specifications subject to change without notice. No license is granted by implication or otherwise under any patent or patent rights of Analog Devices. Trademarks and registered trademarks are the property of their respective owners. One Technology Way, P.O. Box 9106, Norwood, MA 02062-9106, U.S.A. Tel: 781.329.4700 ©2008–2013 Analog Devices, Inc. All rights reserved. Technical Support www.analog.com AD7190* PRODUCT PAGE QUICK LINKS Last Content Update: 02/23/2017 COMPARABLE PARTS REFERENCE DESIGNS View a parametric search of comparable parts. • CN0102 EVALUATION KITS REFERENCE MATERIALS • AD7190 Evaluation Board Solutions Bulletins & Brochures DOCUMENTATION • Test & Instrumentation Solutions Bulletin, Volume 10, Issue 3 Application Notes Tutorials • AN-0979: Digital Filtering Options: AD7190, AD7192 • Tutorial on Technical and Performance Benefits of AD719x Family • AN-1069: Zero Latency for the AD7190, AD7192, AD7193, AD7194, and AD7195 • AN-1084: Channel Switching: AD7190, AD7192, AD7193, AD7194, AD7195 DESIGN RESOURCES • AD7190 Material Declaration • AN-1131: Chopping on the AD7190, AD7192, AD7193, AD7194, and AD7195 • PCN-PDN Information Data Sheet • Symbols and Footprints • AD7190: 4.8 kHz Ultralow Noise 24-Bit Sigma-Delta ADC with PGA Data Sheet User Guides • UG-222: Evaluation Board for the AD7190/AD7192 4.8 kHz Ultralow Noise 24-Bit Sigma-Delta ADCs SOFTWARE AND SYSTEMS REQUIREMENTS • AD7190 - Microcontroller No-OS Driver • AD7192 IIO High Precision ADC Linux Driver TOOLS AND SIMULATIONS • Quality And Reliability DISCUSSIONS View all AD7190 EngineerZone Discussions. SAMPLE AND BUY Visit the product page to see pricing options. TECHNICAL SUPPORT Submit a technical question or find your regional support number. • AD7190/AD7192 Digital Filter Models DOCUMENT FEEDBACK • Download the Active Functional Model to evaluate and debug AD719x Submit feedback for this data sheet. This page is dynamically generated by Analog Devices, Inc., and inserted into this data sheet. A dynamic change to the content on this page will not trigger a change to either the revision number or the content of the product data sheet. This dynamic page may be frequently modified. AD7190 Data Sheet TABLE OF CONTENTS Features .............................................................................................. 1 Full-Scale Register ...................................................................... 25 Applications ....................................................................................... 1 ADC Circuit Information.............................................................. 26 General Description ......................................................................... 1 Overview ..................................................................................... 26 Functional Block Diagram .............................................................. 1 Filter, Output Data Rate, Settling Time ................................... 26 Revision History ............................................................................... 2 Digital Interface .......................................................................... 29 Specifications..................................................................................... 3 Circuit Description......................................................................... 33 Timing Characteristics ................................................................ 7 Analog Input Channel ............................................................... 33 Circuit and Timing Diagrams..................................................... 7 PGA .............................................................................................. 33 Absolute Maximum Ratings............................................................ 9 Bipolar/Unipolar Configuration .............................................. 33 Thermal Resistance ...................................................................... 9 Data Output Coding .................................................................. 33 ESD Caution .................................................................................. 9 Clock ............................................................................................ 33 Pin Configuration and Function Descriptions ........................... 10 Burnout Currents ....................................................................... 34 Typical Performance Characteristics ........................................... 12 Reference ..................................................................................... 34 RMS Noise and Resolution............................................................ 15 Reference Detect ......................................................................... 34 4 Reset ............................................................................................. 34 3 System Synchronization ............................................................ 35 4 Temperature Sensor ................................................................... 35 3 Sinc Chop Enabled .................................................................... 18 Bridge Power-Down Switch ...................................................... 35 On-Chip Registers .......................................................................... 19 Logic Outputs ............................................................................. 35 Communications Register ......................................................... 19 Enable Parity ............................................................................... 36 Status Register ............................................................................. 20 Calibration................................................................................... 36 Mode Register ............................................................................. 20 Grounding and Layout .............................................................. 37 Configuration Register .............................................................. 22 Applications Information .............................................................. 38 Data Register ............................................................................... 24 Weigh Scales ................................................................................ 38 ID Register ................................................................................... 24 Outline Dimensions ....................................................................... 39 GPOCON Register ..................................................................... 24 Ordering Guide .......................................................................... 39 Offset Register ............................................................................. 25 Automotive Products ................................................................. 39 Sinc Chop Disabled ................................................................... 15 Sinc Chop Disabled ................................................................... 16 Sinc Chop Enabled .................................................................... 17 REVISION HISTORY 2/13—Rev. B to Rev. C Added Automotive Information (Throughout) ........................... 1 Added Automotive Specifications, Table 1.................................... 3 Changes to Ordering Guide .......................................................... 39 5/09—Rev. A to Rev. B Changes to Table 3 ............................................................................ 9 5/09—Rev. 0 to Rev. A Changes to Table 1 ............................................................................ 3 Changes to Table 3 and Table 4 ....................................................... 9 Changes to Table 5 .......................................................................... 10 Changes to Table 6 and Table 7 ..................................................... 15 Changes to Status Register Section .............................................. 20 Changes to Table 17 ........................................................................ 21 Changes to Table 19 ........................................................................ 23 Changes to Table 20 ....................................................................... 24 Added ID Register Section ............................................................ 24 Changes to Table 21 ....................................................................... 25 Changes to Filter, Output Data Rate, Settling Time Section .... 26 Changes to Continuous Conversion Mode Section................... 31 Changes to Analog Input Channel and Bipolar/Unipolar Configuration Sections .................................................................. 33 Changes to Burnout Currents, Reference, Reference Detect, and Reset Sections.................................................................................. 34 Changes to Temperature Sensor Section ..................................... 35 Changes to Calibration Section .................................................... 36 Changes to Grounding and Layout Section ................................ 37 Changes to Weigh Scales Section ................................................. 38 Changes to Ordering Guide .......................................................... 39 10/08—Revision 0—Initial Version Rev. C | Page 2 of 40 Data Sheet AD7190 SPECIFICATIONS AVDD = 4.75 V to 5.25 V, DVDD = 2.7 V to 5.25 V, AGND = DGND = 0 V, REFINx(+) = AVDD , REFINx(−) = AGND, MCLK = 4.92 MHz, TA = TMIN to TMAX, unless otherwise noted. Table 1. Parameter ADC Output Data Rate No Missing Codes2 Resolution RMS Noise and Output Data Rates Integral Nonlinearity B Grade WB Grade Offset Error4, 5 Offset Error Drift vs. Temperature5 Offset Error Drift vs. Time Gain Error4 B Grade WB Grade Gain Drift vs. Temperature Gain Drift vs. Time Power Supply Rejection B Grade WB Grade Common-Mode Rejection @ DC @ 50 Hz, 60 Hz2 @ 50 Hz, 60 Hz2 Normal Mode Rejection2 Sinc4 Filter Internal Clock @ 50 Hz, 60 Hz @ 50 Hz @ 60 Hz AD7190B Unit Test Conditions/Comments1 4.7 to 4800 1.17 to 1200 1.56 to 1600 24 24 See the RMS Noise and Resolution section See the RMS Noise and Resolution section Hz nom Hz nom Hz nom Bits min Bits min Chop disabled. Chop enabled, sinc4 filter. Chop enabled, sinc3 filter. FS > 1, sinc4 filter3. FS > 4, sinc3 filter3. ±5 ±15 ±7 ±30 ±75/gain ±0.5 ±100/gain ±5 ±5 25 ppm of FSR max ppm of FSR max ppm of FSR max ppm of FSR max μV typ μV typ nV/°C typ nV/°C typ nV/°C typ nV/1000 hours typ ±1 ppm typical, gain = 1. ±5 ppm typical, gain > 1. ±1 ppm typical, gain = 1. ±5 ppm typical, gain > 1. Chop disabled. Chop enabled. Gain = 1 to 16. chop disabled. Gain = 32 to 128. chop disabled. Chop enabled. Gain ≥ 32. ±0.005 % max ±0.0075 % max ±0.0075 ±1 10 95 100 95 % typ ppm/°C typ ppm/1000 hours typ dB typ dB min dB min ±0.001 % typical, gain = 1, TA = 25°C, AVDD = 5 V6. ±0.001 % typical, gain = 1, TA = 25°C, AVDD = 5 V6. Gain > 1, post internal full-scale calibration. 100 110 120 120 dB min dB min dB min dB min Gain = 1, VIN = 1 V2. Gain > 1, VIN = 1 V/gain. 10 Hz output data rate, 50 ± 1 Hz, 60 ± 1 Hz. 50 ± 1 Hz (50 Hz output data rate), 60 ± 1 Hz (60 Hz output data rate). 100 74 dB min dB min 96 97 dB min dB min 10 Hz output data rate, 50 ± 1 Hz, 60 ± 1 Hz. 50 Hz output data rate, REJ607 = 1, 50 ± 1 Hz, 60 ± 1 Hz. 50 Hz output data rate, 50 ± 1 Hz. 60 Hz output data rate, 60 ± 1 Hz. Rev. C | Page 3 of 40 Gain = 1. Gain = 1, VIN = 1 V. Gain > 1, VIN = 1 V/gain. 110 dB typical. Gain > 1, VIN = 1 V/gain. 110 dB typical. AD7190 Data Sheet AD7190B Unit Test Conditions/Comments1 120 82 dB min dB min @ 50 Hz @ 60 Hz Sinc3 Filter Internal Clock @ 50 Hz, 60 Hz 120 120 dB min dB min 10 Hz output data rate, 50 ± 1 Hz, 60 ± 1 Hz. 50 Hz output data rate, REJ607 = 1, 50 ± 1 Hz, 60 ± 1 Hz. 50 Hz output data rate, 50 ± 1 Hz. 60 Hz output data rate, 60 ± 1 Hz. 75 60 dB min dB min @ 50 Hz @ 60 Hz External Clock @ 50 Hz, 60 Hz 70 70 dB min dB min 100 67 dB min dB min 95 95 dB min dB min ±VREF/gain V nom ±(AVDD – 1.25 V)/gain V min/max AGND − 50 mV AVDD + 50 mV AGND + 250 mV AVDD − 250 mV V min V max V min V max ±2 ±3 ±5 nA max nA max pA/°C typ Gain = 1. Gain > 1. ±5 μA/V typ ±1 ±0.05 ±1.6 μA/V typ nA/V/°C typ nA/V/°C typ Gain = 1, input current varies with input voltage. Gain > 1. External clock. Internal clock. AVDD 1 AVDD V nom V min V max AGND – 50 mV AVDD + 50 mV 7 ±0.03 V min V max μA/V typ nA/V/°C typ External clock. 1.3 nA/V/°C typ Internal clock. Parameter External Clock @ 50 Hz, 60 Hz @ 50 Hz @ 60 Hz ANALOG INPUTS Differential Input Voltage Ranges Absolute AIN Voltage Limits2 Unbuffered Mode Buffered Mode Analog Input Current Buffered Mode Input Current2 Input Current Drift Unbuffered Mode Input Current Input Current Drift REFERENCE INPUT REFIN Voltage Reference Voltage Range2 Absolute REFIN Voltage Limits2 Average Reference Input Current Average Reference Input Current Drift Rev. C | Page 4 of 40 10 Hz output data rate, 50 ± 1 Hz, 60 ± 1 Hz. 50 Hz output data rate, REJ60 = 1, 50 ± 1 Hz, 60 ± 1 Hz. 50 Hz output data rate, 50 ± 1 Hz. 60 Hz output data rate, 60 ± 1 Hz. 10 Hz output data rate, 50 ± 1 Hz, 60 ± 1 Hz. 50 Hz output data rate, REJ607 = 1, 50 ± 1 Hz, 60 ± 1 Hz. 50 Hz output data rate, 50 ± 1 Hz. 60 Hz output data rate, 60 ± 1 Hz. VREF = REFINx(+) − REFINx(−), gain = 1 to 128. Gain > 1. REFIN = REFINx(+) − REFINx(−). The differential input must be limited to ± (AVDD – 1.25 V)/gain when gain > 1. Data Sheet Parameter Normal Mode Rejection2 Common-Mode Rejection Reference Detect Levels TEMPERATURE SENSOR Accuracy Sensitivity BRIDGE POWER-DOWN SWITCH RON Allowable Current2 BURNOUT CURRENTS AIN Current DIGITAL OUTPUTS (P0 to P3) Output High Voltage, VOH2 Output Low Voltage, VOL2 Floating-State Leakage Current Floating-State Output Capacitance INTERNAL/EXTERNAL CLOCK Internal Clock Frequency Duty Cycle External Clock/Crystal2 Frequency Input Low Voltage, VINL Input High Voltage, VINH Input Current LOGIC INPUTS Input High Voltage, VINH2 Input Low Voltage, VINL2 Hysteresis2 Input Currents LOGIC OUTPUT (DOUT/RDY) Output High Voltage, VOH2 Output Low Voltage, VOL2 Output High Voltage, VOH2 Output Low Voltage, VOL2 Floating-State Leakage Current Floating-State Output Capacitance Data Output Coding SYSTEM CALIBRATION2 Full-Scale Calibration Limit Zero-Scale Calibration Limit Input Span AD7190 Unit Test Conditions/Comments1 AD7190B Same as for analog inputs 95 0.3 0.6 dB typ V min V max ±2 2815 °C typ Codes/°C typ Applies after user calibration at 25°C. Bipolar mode. 10 30 Ω max mA max Continuous current. 500 nA nom Analog inputs must be buffered and chop disabled. 4 0.4 ±100 10 V min V max nA max pF typ AVDD = 5V, ISOURCE = 200 μA. AVDD = 5V, ISINK = 800 μA. 4.92 ± 4% 50:50 MHz min/max % typ 4.9152 2.4576/5.12 0.8 0.4 2.5 3.5 ±10 MHz nom MHz min/max V max V max V min V min μA max 2 0.8 0.1/0.25 ±10 V min V max V min/V max μA max DVDD − 0.6 0.4 4 0.4 ±10 10 V min V max V min V max μA max pF typ Offset binary 1.05 × FS −1.05 × FS 0.8 × FS 2.1 × FS V max V min V min V max Rev. C | Page 5 of 40 DVDD = 5 V. DVDD = 3 V. DVDD = 3 V. DVDD = 5 V. DVDD = 3 V, ISOURCE = 100 μA. DVDD = 3 V, ISINK = 100 μA. DVDD = 5 V, ISOURCE = 200 μA. DVDD = 5 V, ISINK = 1.6 mA. AD7190 Parameter POWER REQUIREMENTS8 Power Supply Voltage AVDD − AGND DVDD − DGND Power Supply Currents AIDD Current B Grade WB Grade DIDD Current IDD (Power-Down Mode) B Grade WB Grade Data Sheet AD7190B Unit 4.75/5.25 2.7/5.25 V min/max V min/max 1 1.3 4.5 4.75 6.2 6.75 5 5.3 6.8 7.4 0.4 0.6 1.5 mA max mA max mA max mA max mA max mA max mA max mA max mA max mA max mA max mA max mA typ 2 5 μA max μA max 1 Test Conditions/Comments1 0.85 mA typical, gain = 1, buffer off. 1.1 mA typical, gain = 1, buffer on. 3.5 mA typical, gain = 8, buffer off. 4 mA typical, gain = 8, buffer on. 5 mA typical, gain = 16 to 128, buffer off. 5.5 mA typical, gain = 16 to 128, buffer on. 3.5 mA typical, gain = 8, buffer off. 4 mA typical, gain = 8, buffer on. 5 mA typical, gain = 16 to 128, buffer off. 5.5 mA typical, gain = 16 to 128, buffer on. 0.35 mA typical, DVDD = 3 V. 0.5 mA typical, DVDD = 5 V. External crystal used. Temperature range: TMIN = −40°C, TMAX = +105°C. Specification is not production tested but is supported by characterization data at initial product release. 3 FS = decimal equivalent of Bit FS9 to Bit FS0 in the mode register. 4 Following a system or internal zero-scale calibration, the offset error is in the order of the noise for the programmed gain and output data rate selected. A system fullscale calibration reduces the gain error to the order of the noise for the programmed gain and output data rate. 5 The analog inputs are configured for differential mode. 6 Applies at the factory calibration conditions (AVDD = 5 V, gain = 1, TA = 25°C). 7 REJ60 is a bit in the mode register. When the output data rate is set to 50 Hz, setting REJ60 to 1 places a notch at 60 Hz, allowing simultaneous 50 Hz/60 Hz rejection. 8 Digital inputs equal to DVDD or DGND. 2 Rev. C | Page 6 of 40 Data Sheet AD7190 TIMING CHARACTERISTICS AVDD = 4.75 V to 5.25 V, DVDD = 2.7 V to 5.25 V, AGND = DGND = 0 V, Input Logic 0 = 0 V, Input Logic 1 = DVDD, unless otherwise noted. Table 2. Parameter t3 t4 READ OPERATION t1 t2 3 t5 5, 6 t6 t7 WRITE OPERATION t8 t9 t10 t11 Limit at TMIN, TMAX (B Version) 100 100 Unit ns min ns min Conditions/Comments 1, 2 SCLK high pulse width SCLK low pulse width 0 60 80 0 60 80 10 80 0 10 ns min ns max ns max ns min ns max ns max ns min ns max ns min ns min CS falling edge to DOUT/RDY active time DVDD = 4.75 V to 5.25 V DVDD = 2.7 V to 3.6 V SCLK active edge to data valid delay 4 DVDD = 4.75 V to 5.25 V DVDD = 2.7 V to 3.6 V Bus relinquish time after CS inactive edge 0 30 25 0 ns min ns min ns min ns min CS falling edge to SCLK active edge setup time4 Data valid to SCLK edge setup time Data valid to SCLK edge hold time CS rising edge to SCLK edge hold time SCLK inactive edge to CS inactive edge SCLK inactive edge to DOUT/RDY high Sample tested during initial release to ensure compliance. All input signals are specified with tR = tF = 5 ns (10% to 90% of DVDD) and timed from a voltage level of 1.6 V. See Figure 3 and Figure 4. 3 These numbers are measured with the load circuit shown in Figure 2 and defined as the time required for the output to cross the VOL or VOH limits. 4 The SCLK active edge is the falling edge of SCLK. 5 These numbers are derived from the measured time taken by the data output to change 0.5 V when loaded with the circuit shown in Figure 2. The measured number is then extrapolated back to remove the effects of charging or discharging the 50 pF capacitor. This means that the times quoted in the timing characteristics are the true bus relinquish times of the part and, as such, are independent of external bus loading capacitances. 6 RDY returns high after a read of the data register. In single conversion mode and continuous conversion mode, the same data can be read again, if required, while RDY is high, although care should be taken to ensure that subsequent reads do not occur close to the next output update. If the continuous read feature is enabled, the digital word can be read only once. 1 2 CIRCUIT AND TIMING DIAGRAMS ISINK (1.6mA WITH DVDD = 5V, 100µA WITH DVDD = 3V) TO OUTPUT PIN 1.6V ISOURCE (200µA WITH DVDD = 5V, 100µA WITH DVDD = 3V) Figure 2. Load Circuit for Timing Characterization Rev. C | Page 7 of 40 07640-002 50pF AD7190 Data Sheet CS (I) t6 t1 t5 MSB DOUT/RDY (O) LSB t7 t2 t3 07640-003 SCLK (I) t4 I = INPUT, O = OUTPUT Figure 3. Read Cycle Timing Diagram CS (I) t11 t8 SCLK (I) t9 t10 MSB LSB I = INPUT, O = OUTPUT Figure 4. Write Cycle Timing Diagram Rev. C | Page 8 of 40 07640-004 DIN (I) Data Sheet AD7190 ABSOLUTE MAXIMUM RATINGS TA = 25°C, unless otherwise noted. THERMAL RESISTANCE Table 3. θJA is specified for the worst-case conditions, that is, a device soldered in a circuit board for surface-mount packages. Parameter AVDD to AGND Rating −0.3 V to +6.5 V DVDD to AGND AGND to DGND Analog Input Voltage to AGND −0.3 V to +6.5 V −0.3 V to +0.3 V −0.3 V to AVDD + 0.3 V Package Type 24-Lead TSSOP Reference Input Voltage to AGND Digital Input Voltage to DGND −0.3 V to AVDD + 0.3 V −0.3 V to DVDD + 0.3 V ESD CAUTION Digital Output Voltage to DGND AIN/Digital Input Current Operating Temperature Range Storage Temperature Range −0.3 V to DVDD + 0.3 V 10 mA −40°C to +105°C −65°C to +150°C 150°C Maximum Junction Temperature Lead Temperature, Soldering Reflow Table 4. Thermal Resistance 260°C Stresses above those listed under Absolute Maximum Ratings may cause permanent damage to the device. This is a stress rating only; functional operation of the device at these or any other conditions above those indicated in the operational section of this specification is not implied. Exposure to absolute maximum rating conditions for extended periods may affect device reliability. Rev. C | Page 9 of 40 θJA 128 θJC 42 Unit °C/W AD7190 Data Sheet MCLK1 1 24 DIN MCLK2 2 23 DOUT/RDY SCLK 3 22 SYNC CS 4 AD7190 21 DVDD P3 5 TOP VIEW (Not to Scale) 20 AVDD P2 6 19 DGND P1/REFIN2(+) 7 18 AGND P0/REFIN2(–) 8 17 BPDSW NC 9 16 REFIN1(–) AINCOM 10 15 REFIN1(+) AIN1 11 14 AIN4 AIN2 12 13 AIN3 NC = NO CONNECT 07640-005 PIN CONFIGURATION AND FUNCTION DESCRIPTIONS Figure 5. Pin Configuration Table 5. Pin Function Descriptions Pin No. 1 Mnemonic MCLK1 2 MCLK2 3 SCLK 4 CS 5 6 7 P3 P2 P1/REFIN2(+) 8 P0/REFIN2(−) 9 10 NC AINCOM 11 AIN1 12 AIN2 Description When the master clock for the device is provided externally by a crystal, the crystal is connected between MCLK1 and MCLK2. Master Clock Signal for the Device. The AD7190 has an internal 4.92 MHz clock. This internal clock can be made available on the MCLK2 pin. The clock for the AD7190 can be provided externally also in the form of a crystal or external clock. A crystal can be tied across the MCLK1 and MCLK2 pins. Alternatively, the MCLK2 pin can be driven with a CMOS-compatible clock and the MCLK1 pin left unconnected. Serial Clock Input. This serial clock input is for data transfers to and from the ADC. The SCLK has a Schmitttriggered input, making the interface suitable for opto-isolated applications. The serial clock can be continuous with all data transmitted in a continuous train of pulses. Alternatively, it can be a noncontinuous clock with the information transmitted to or from the ADC in smaller batches of data. Chip Select Input. This is an active low logic input used to select the ADC. CS can be used to select the ADC in systems with more than one device on the serial bus or as a frame synchronization signal in communicating with the device. CS can be hardwired low, allowing the ADC to operate in 3-wire mode with SCLK, DIN, and DOUT used to interface with the device. Digital Output Pin. This pin can function as a general-purpose output bit referenced between AVDD and AGND. Digital Output Pin. This pin can function as a general-purpose output bit referenced between AVDD and AGND. Digital Output Pin/Positive Reference Input. This pin functions as a general-purpose output bit referenced between AVDD and AGND. When REFSEL = 1, this pin functions as REFIN2(+). An external reference can be applied between REFIN2(+) and REFIN2(−). REFIN2(+) can lie anywhere between AVDD and AGND + 1 V. The nominal reference voltage, (REFIN2(+) − REFIN2(−)), is AVDD, but the part functions with a reference from 1 V to AVDD. Digital Output Pin/Negative Reference Input. This pin functions as a general-purpose output bit referenced between AVDD and AGND. When REFSEL = 1, this pin functions as REFIN2(−). This reference input can lie anywhere between AGND and AVDD − 1 V. No Connect. This pin should be tied to AGND. Analog Input AIN1 to Analog Input AIN4 are referenced to this input when configured for pseudo differential operation. Analog Input. It can be configured as the positive input of a fully differential input pair when used with AIN2 or as a pseudo differential input when used with AINCOM. Analog Input. It can be configured as the negative input of a fully differential input pair when used with AIN1 or as a pseudo differential input when used with AINCOM. Rev. C | Page 10 of 40 Data Sheet Pin No. 13 Mnemonic AIN3 14 AIN4 15 REFIN1(+) 16 17 18 19 20 21 22 REFIN1(−) BPDSW AGND DGND AVDD DVDD SYNC 23 DOUT/RDY 24 DIN AD7190 Description Analog Input. It can be configured as the positive input of a fully differential input pair when used with AIN4 or as a pseudo differential input when used with AINCOM. Analog Input. It can be configured as the negative input of a fully differential input pair when used with AIN3 or as a pseudo differential input when used with AINCOM. Positive Reference Input. An external reference can be applied between REFIN1(+) and REFIN1(−). REFIN1(+) can lie anywhere between AVDD and AGND + 1 V. The nominal reference voltage, (REFIN1(+) − REFIN1(−)), is AVDD, but the part functions with a reference from 1 V to AVDD. Negative Reference Input. This reference input can lie anywhere between AGND and AVDD − 1 V. Bridge Power-Down Switch to AGND. Analog Ground Reference Point. Digital Ground Reference Point. Analog Supply Voltage, 4.75 V to 5.25 V. AVDD is independent of DVDD. Digital Supply Voltage, 2.7 V to 5.25 V. DVDD is independent of AVDD. Logic input that allows for synchronization of the digital filters and analog modulators when using multiple AD7190 devices. While SYNC is low, the nodes of the digital filter, the filter control logic, and the calibration control logic are reset and the analog modulator is held in its reset state. SYNC does not affect the digital interface but does reset RDY to a high state if it is low. SYNC has a pull-up resistor internally to DVDD. Serial Data Output/Data Ready Output. DOUT/RDYserves a dual purpose. It functions as a serial data output pin to access the output shift register of the ADC. The output shift register can contain data from any of the on-chip data or control registers. In addition, DOUT/RDY operates as a data ready pin, going low to indicate the completion of a conversion. If the data is not read after the conversion, the pin goes high before the next update occurs. The DOUT/RDY falling edge can be used as an interrupt to a processor, indicating that valid data is available. With an external serial clock, the data can be read using the DOUT/RDY pin. With CS low, the data/control word information is placed on the DOUT/RDY pin on the SCLK falling edge and is valid on the SCLK rising edge. Serial Data Input to the Input Shift Register on the ADC. Data in this shift register is transferred to the control registers in the ADC, with the register selection bits of the communications register identifying the appropriate register. Rev. C | Page 11 of 40 AD7190 Data Sheet TYPICAL PERFORMANCE CHARACTERISTICS 8,388,760 8,388,950 8,388,900 8,388,758 8,388,850 8,388,800 8,388,750 CODE 8,388,754 8,388,752 8,388,700 8,388,650 8,388,750 8,388,600 8,388,550 8,388,748 200 400 600 800 1000 SAMPLE Figure 6. Noise (VREF = 5 V, Output Data Rate = 4.7 Hz, Gain = 128, Chop Disabled, Sinc4 Filter) 07640-106 0 8,388,450 0 100 200 300 400 500 600 700 800 900 1000 SAMPLES Figure 8. Noise (VREF = 5 V, Output Data Rate = 4800 Hz, Gain = 128, Chop Disabled, Sinc4 Filter) 250 30 25 200 FREQUENCY 20 150 100 15 10 50 5 0 Figure 7. Noise Distribution Histogram (VREF = 5 V, Output Data Rate = 4.7 Hz, Gain = 128, Chop Disabled, Sinc4 Filter) 8,388,920 07640-109 CODE 8,388,834 8,388,748 8,388,662 8,388,576 8,388,490 8,388,760 8,388,758 8,388,756 07640-107 CODE 8,388,754 8,388,752 8,388,750 8,388,748 8,388,746 0 Figure 9. Noise Distribution Histogram (VREF = 5 V, Output Data Rate = 4800 Hz, Gain = 128, Chop Disabled, Sinc4 Filter) Rev. C | Page 12 of 40 07640-108 8,388,500 8,388,746 FREQUENCY CODE 8,388,756 Data Sheet AD7190 3.0 8,388,820 8,388,800 2.0 INL (ppm of FSR) 8,388,780 8,388,760 8,388,720 8,388,700 0 –1.0 8,388,680 –2.0 8,388,660 0 100 200 300 400 500 600 700 800 900 1000 SAMPLES –3.0 –2.5 07640-110 8,388,620 –2.0 –1.5 –1.0 –0.5 0 0.5 1.0 1.5 2.0 2.5 VIN (V) 07640-112 8,388,640 Figure 12. INL (Gain = 1) Figure 10. Noise (VREF = 5 V, Output Data Rate = 4800 Hz, Gain = 1, Chop Disabled, Sinc4 Filter) 80 70 6 60 4 INL (ppm of FSR) 50 40 30 20 2 0 –2 10 8,388,820 07640-111 8,388,780 8,388,740 8,388,700 8,388,660 CODE Figure 11. Noise Distribution Histogram (VREF = 5 V, Output Data Rate = 4800 Hz, Gain = 1, Chop Disabled, Sinc4 Filter) –6 –0.020 –0.015 –0.010 –0.005 0 0.005 0.010 VIN (V) Figure 13. INL (Gain = 128) Rev. C | Page 13 of 40 0.015 0.020 07640-113 –4 0 8,388,620 FREQUENCY CODE 8,388,740 1.0 AD7190 Data Sheet 66 1.000008 1.000007 1.000006 62 1.000005 GAIN OUTPUT VOLTAGE (µV) 64 60 1.000004 1.000003 58 1.000002 56 –20 0 20 40 60 80 100 120 140 TEMPERATURE (°C) 1.000000 –60 128.002 –0.2 128.001 –0.3 128.000 GAIN –0.1 –0.4 127.999 –0.5 127.998 –0.6 127.997 0 20 40 60 80 100 TEMERATURE (°C) 120 07640-115 OFFSET (µV) 128.003 –20 0 20 40 60 80 100 120 100 120 Figure 16. Gain Error (Gain = 1, Chop Disabled) 0 –40 –20 TEMPERATURE (°C) Figure 14. Offset Error (Gain = 1, Chop Disabled) –0.7 –60 –40 Figure 15. Offset Error (Gain = 128, Chop Disabled) 127.996 –60 –40 –20 0 20 40 60 80 TEMPERATURE (°C) Figure 17. Gain Error (Gain = 128, Chop Disabled) Rev. C | Page 14 of 40 07640-117 –40 07640-114 54 –60 07640-116 1.000001 Data Sheet AD7190 RMS NOISE AND RESOLUTION on a single channel. The effective resolution is also shown, and the output peak-to-peak (p-p) resolution, or noise-free resolution, is listed in parentheses. It is important to note that the effective resolution is calculated using the rms noise, wheras the p-p resolution is calculated based on peak-to-peak noise. The p-p resolution represents the resolution for which there is no code flicker. These numbers are typical and are rounded to the nearest ½ LSB. The AD7190 has a choice of two filter types: sinc4 and sinc3. In addition, the AD7190 can be operated with chop enabled or chop disabled. The following tables show the rms noise of the AD7190 for some of the output data rates and gain settings with chop disabled and enabled for the sinc4 and sinc3 filters. The numbers given are for the bipolar input range with the external 5 V reference. These numbers are typical and are generated with a differential input voltage of 0 V when the ADC is continuously converting SINC4 CHOP DISABLED Table 6. RMS Noise (nV) vs. Gain and Output Data Rate Filter Word (Decimal) 1023 640 480 96 80 32 16 5 2 1 Output Data Rate (Hz) 4.7 7.5 10 50 60 150 300 960 2400 4800 Settling Time (ms) 852.5 533 400 80 66.7 26.7 13.3 4.17 1.67 0.83 Gain of 1 250 310 330 900 970 1460 1900 3000 5000 14,300 Gain of 8 38 45 50 125 140 215 285 480 780 1920 Gain of 16 21 25 30 78 88 125 170 280 440 1000 Gain of 32 12 16 18 45 52 75 100 175 280 550 Gain of 64 10 12 14 33 36 55 75 140 220 380 Gain of 128 8.5 10.5 11.5 28 31 48 67 121 198 295 Table 7. Effective Resolution (Peak-to-Peak Resolution) vs. Gain and Output Data Rate Filter Word (Decimal) 1023 640 480 96 80 32 16 5 2 1 1 Output Data Rate (Hz) 4.7 7.5 10 50 60 150 300 960 2400 4800 Settling Time (ms) 852.5 533 400 80 66.7 26.7 13.3 4.17 1.67 0.83 Gain of 1 1 24 (22.5) 24 (22) 24 (22) 23.5 (20.5) 23.5 (20.5) 22.5 (20) 22.5 (19.5) 21.5 (19) 21 (18) 19.5 (16.5) Gain of 81 24 (22) 24 (22) 24 (22) 23.5 (20.5) 23 (20.5) 22.5 (19.5) 22 (19.5) 21.5 (18.5) 20.5 (18) 19.5 (16.5) Gain of 161 24 (22) 24 (22) 24 (21.5) 23 (20) 22.5 (20) 22.5 (19.5) 22 (19) 21 (18.5) 20.5 (17.5) 19.5 (16.5) The output peak-to-peak (p-p) resolution is listed in parentheses. Rev. C | Page 15 of 40 Gain of 321 24 (22) 24 (21.5) 24 (21.5) 22.5 (20) 22.5 (20) 22 (19.5) 21.5 (19) 21 (18) 20 (17.5) 19 (16.5) Gain of 641 24 (21) 23.5 (21) 23.5 (20.5) 22 (19.5) 22 (19.5) 21.5 (18.5) 21 (18.5) 20 (17.5) 19.5 (16.5) 18.5 (16) Gain of 1281 23 (20.5) 23 (20) 22.5 (20) 21.5 (18.5) 21.5 (18.5) 20.5 (18) 20 (17.5) 19.5 (16.5) 18.5 (16) 18 (15.5) AD7190 Data Sheet SINC3 CHOP DISABLED Table 8. RMS Noise (nV) vs. Gain and Output Data Rate Filter Word (Decimal) 1023 640 480 96 80 32 16 5 2 1 Output Data Rate (Hz) 4.7 7.5 10 50 60 150 300 960 2400 4800 Settling Time (ms) 639.4 400 300 60 50 20 10 3.125 1.25 0.625 Gain of 1 270 320 350 1000 1050 1500 1950 4000 56,600 442,000 Gain of 8 42 50 60 134 145 225 308 590 7000 55,000 Gain of 16 23 27 35 86 95 130 175 330 3500 28,000 Gain of 32 13.5 17 19 50 55 80 110 200 1800 14,000 Gain of 64 10.5 13 15 35 40 58 83 150 900 7000 Gain of 128 9 11.5 12.5 29 32 50 73 133 490 3450 Gain of 321 24 (21.5) 24 (21.5) 24 (21) 22.5 (20) 22.5 (19.5) 22 (19) 21.5 (18.5) 20.5 (18) 17.5 (14.5) 14.5 (11.5) Gain of 641 24 (21) 23.5 (21) 23.5 (20.5) 22 (19.5) 22 (19) 21.5 (18.5) 21 (18) 20 (17.5) 17.5 (14.5) 14.5 (11.5) Gain of 1281 23 (20.5) 22.5 (20) 22.5 (20) 21.5 (18.5) 21 (18.5) 20.5 (18) 20 (17.5) 19 (16.5) 17.5 (14.5) 14.5 (11.5) Table 9. Effective Resolution (Peak-to-Peak Resolution) vs. Gain and Output Data Rate Filter Word (Decimal) 1023 640 480 96 80 32 16 5 2 1 1 Output Data Rate (Hz) 4.7 7.5 10 50 60 150 300 960 2400 4800 Settling Time (ms) 639.4 400 300 60 50 20 10 3.125 1.25 0.625 Gain of 1 1 24 (22.5) 24 (22) 24 (22) 23.5 (20.5) 23 (20.5) 22.5 (20) 22.5 (19.5) 21.5 (18.5) 17.5 (14.5) 14.5 (11.5) Gain of 81 24 (22) 24 (22) 24 (21.5) 23 (20.5) 23 (20.5) 22.5 (19.5) 22 (19) 21 (18.5) 17.5 (14.5) 14.5 (11.5) Gain of 161 24 (22) 24 (21.5) 24 (21.5) 23 (20) 22.5 (20) 22 (19.5) 22 (19) 21 (18) 17.5 (14.5) 14.5 (11.5) The output peak-to-peak (p-p) resolution is listed in parentheses. Rev. C | Page 16 of 40 Data Sheet AD7190 SINC4 CHOP ENABLED Table 10. RMS Noise (nV) vs. Gain and Output Data Rate Filter Word (Decimal) 1023 640 480 96 80 32 16 5 2 1 Output Data Rate (Hz) 1.175 1.875 2.5 12.5 15 37.5 75 240 600 1200 Settling Time (ms) 1702 1067 800 160 133 53 26.7 8.33 3.33 1.67 Gain of 1 177 219 234 637 686 1033 1343 2121 3536 10,200 Gain of 8 27 32 36 89 99 152 202 340 552 1360 Gain of 16 15 18 21 55 63 89 120 198 311 707 Gain of 32 8.5 11.5 13 32 37 53 71 124 198 389 Gain of 64 7 8.5 10 24 26 39 53 99 156 26 Gain of 128 6 7.5 8.5 20 22 34 48 86 140 209 Table 11. Effective Resolution (Peak-to-Peak Resolution) vs. Gain and Output Data Rate Filter Word (Decimal) 1023 640 480 96 80 32 16 5 2 1 1 Output Data Rate (Hz) 1.175 1.875 2.5 12.5 15 37.5 75 240 600 1200 Settling Time (ms) 1702 1067 800 160 133 53 26.7 8.33 3.33 1.67 Gain of 1 1 24 (23) 24 (22.5) 24 (22.5) 24 (21) 24 (21) 23 (20.5) 23 (20) 22 (19.5) 21.5 (18.5) 20 (17) Gain of 81 24 (22.5) 24 (22.5) 24 (22.5) 24 (21) 23.5 (21) 23 (20) 22.5 (20) 22 (19) 21 (18.5) 20 (17) Gain of 161 24 (22.5) 24 (22.5) 24 (22) 23.5 (20.5) 23.5 (20.5) 23 (20) 22.5 (19.5) 21.5 (19) 21 (18) 20 (17) The output peak-to-peak (p-p) resolution is listed in parentheses. Rev. C | Page 17 of 40 Gain of 321 24 (22.5) 24 (22) 24 (22) 23 (20.5) 23 (20.5) 22.5 (20) 22 (19.5) 21.5 (18.5) 20.5 (18) 19.5 (17) Gain of 641 24 (21.5) 24 (21.5) 24 (21) 22.5 (20) 22.5 (20) 22 (19) 21.5 (19) 20.5 (18) 20 (17) 19 (16.5) Gain of 1281 23.5 (21) 23.5 (20.5) 23 (20.5) 22 (19) 22 (19) 21 (18.5) 20.5 (18) 20 (17) 19 (16.5) 18.5 (16) AD7190 Data Sheet SINC3 CHOP ENABLED Table 12. RMS Noise (nV) vs. Gain and Output Data Rate Filter Word (Decimal) 1023 640 480 96 80 32 16 5 2 1 Output Data Rate (Hz) 1.56 2.5 3.33 16.6 20 50 100 320 800 1600 Settling Time (ms) 1282 800 600 120 100 40 20 6.25 2.5 1.25 Gain of 1 191 226 248 708 743 1061 1380 2829 40,100 312,550 Gain of 8 30 36 43 95 103 159 218 418 4950 38,540 Gain of 16 16.5 19 25 61 68 92 124 234 2475 19,800 Gain of 32 10 12 14 36 39 57 78 142 1273 9900 Gain of 64 8 9 11 25 29 41 59 106 637 4950 Gain of 128 6.5 8.5 9 21 23 36 52 94 347 2440 Gain of 321 24 (22) 24 (22) 24 (21.5) 23 (20.5) 23 (20) 22.5 (19.5) 22 (19) 21 (18.5) 18 (15) 15 (12) Gain of 641 24 (21.5) 24 (21.5) 24 (21) 22.5 (20) 22.5 (19.5) 22 (19) 21.5 (18.5) 20.5 (18) 18 (15) 15 (12) Gain of 1281 23.5 (21) 23 (20.5) 23 (20.5) 22 (19) 21.5 (19) 21 (18.5) 20.5 (18) 19.5 (17) 18 (15) 15 (12) Table 13. Effective Resolution (Peak-to-Peak Resolution) vs. Gain and Output Data Rate Filter Word (Decimal) 1023 640 480 96 80 32 16 5 2 1 1 Output Data Rate (Hz) 1.56 2.5 3.33 16.6 20 50 100 320 800 1600 Settling Time (ms) 1282 800 600 120 100 40 20 6.25 2.5 1.25 Gain of 1 1 24 (23) 24 (22.5) 24 (22.5) 24 (21) 23.5 (21) 23 (20.5) 23 (20) 22 (19) 18 (15) 15 (12) Gain of 81 24 (22.5) 24 (22.5) 24 (22) 23.5 (21) 23.5 (21) 23 (20) 22.5 (19.5) 21.5 (19) 18 (15) 15 (12.5) The output peak-to-peak (p-p) resolution is listed in parentheses. Rev. C | Page 18 of 40 Gain of 161 24 (22.5) 24 (22) 24 (22) 23.5 (20.5) 23 (20.5) 22.5 (20) 22.5 (19.5) 21.5 (18.5) 18 (15) 15 (12) Data Sheet AD7190 ON-CHIP REGISTERS The ADC is controlled and configured via a number of on-chip registers, which are described in the following sections. In the descriptions, set implies a Logic 1 state and cleared implies a Logic 0 state, unless otherwise noted. COMMUNICATIONS REGISTER (RS2, RS1, RS0 = 0, 0, 0) The communications register is an 8-bit write-only register. All communications to the part must start with a write operation to the communications register. The data written to the communications register determines whether the next operation is a read or write operation and in which register this operation takes place. For read or write operations, when the subsequent read or write operation to the selected register is complete, the CR7 WEN(0) CR6 R/W(0) CR5 RS2(0) CR4 RS1(0) interface returns to where it expects a write operation to the communications register. This is the default state of the interface and, on power-up or after a reset, the ADC is in this default state waiting for a write operation to the communications register. In situations where the interface sequence is lost, a write operation of at least 40 serial clock cycles with DIN high returns the ADC to this default state by resetting the entire part. Table 14 outlines the bit designations for the communications register. CR0 through CR7 indicate the bit locations, CR denoting that the bits are in the communications register. CR7 denotes the first bit of the data stream. The number in parentheses indicates the power-on/reset default status of that bit. CR3 RS0(0) CR2 CREAD(0) CR1 0(0) CR0 0(0) Table 14. Communications Register Bit Designations Bit Location CR7 Bit Name WEN CR6 R/W CR5 to CR3 RS2 to RS0 CR2 CREAD CR1 to CR0 Description Write enable bit. A 0 must be written to this bit for a write to the communications register to occur. If a 1 is the first bit written, the part does not clock on to subsequent bits in the register. It stays at this bit location until a 0 is written to this bit. After a 0 is written to the WEN bit, the next seven bits are loaded to the communications register. A 0 in this bit location indicates that the next operation is a write to a specified register. A 1 in this position indicates that the next operation is a read from the designated register. Register address bits. These address bits are used to select which registers of the ADC are selected during the serial interface communication. See Table 15. Continuous read of the data register. When this bit is set to 1 (and the data register is selected), the serial interface is configured so that the data register can be continuously read; that is, the contents of the data register are automatically placed on the DOUT pin when the SCLK pulses are applied after the RDY pin goes low to indicate that a conversion is complete. The communications register does not have to be written to for subsequent data reads. To enable continuous read, the instruction 01011100 must be written to the communications register. To disable continuous read, the instruction 01011000 must be written to the communications register while the RDY pin is low. While continuous read is enabled, the ADC monitors activity on the DIN line so that it can receive the instruction to disable continuous read. Additionally, a reset occurs if 40 consecutive 1s are seen on DIN. Therefore, DIN should be held low until an instruction is to be written to the device. These bits must be programmed to Logic 0 for correct operation. Table 15. Register Selection RS2 0 0 0 0 0 1 1 1 1 RS1 0 0 0 1 1 0 0 1 1 RS0 0 0 1 0 1 0 1 0 1 Register Communications register during a write operation Status register during a read operation Mode register Configuration register Data register/data register plus status information ID register GPOCON register Offset register Full-scale register Rev. C | Page 19 of 40 Register Size 8 bits 8 bits 24 bits 24 bits 24 bits/32 bits 8 bits 8 bits 24 bits 24 bits AD7190 Data Sheet STATUS REGISTER (RS2, RS1, RS0 = 0, 0, 0; Power-On/Reset = 0x80) The status register is an 8-bit, read-only register. To access the ADC status register, the user must write to the communications register, select the next operation to be a read, and load Bit RS2, Bit RS1, and Bit RS0 with 0. Table 16 outlines the bit designations for the status register. SR0 through SR7 indicate the bit locations, SR denoting that the bits are in the status register. SR7 denotes the first bit of the data stream. The number in parentheses indicates the power-on/reset default status of that bit. SR7 RDY(1) SR6 ERR(0) SR5 NOREF(0) SR4 Parity(0) SR3 0(0) SR2 CHD2(0) SR1 CHD1(0) SR0 CHD0(0) Table 16. Status Register Bit Designations Bit Location SR7 Bit Name RDY SR6 ERR SR5 NOREF SR4 Parity SR3 SR2 to SR0 0 CHD2 to CHD0 Description Ready bit for the ADC. Cleared when data is written to the ADC data register. The RDY bit is set automatically after the ADC data register is read, or a period of time before the data register is updated with a new conversion result, to indicate to the user that the conversion data should not be read. It is also set when the part is placed in power-down mode or idle mode or when SYNC is taken low. The end of a conversion is also indicated by the DOUT/RDY pin. This pin can be used as an alternative to the status register for monitoring the ADC for conversion data. ADC error bit. This bit is written to at the same time as the RDY bit. The ERR bit is set to indicate that the result written to the ADC data register is clamped to all 0s or all 1s. Error sources include overrange or underrange or the absence of a reference voltage. The bit is cleared by a write operation to start a conversion. No external reference bit. This bit is set to indicate that the selected reference (REFIN1 or REFIN2) is at a voltage that is below a specified threshold. When set, conversion results are clamped to all 1s. This bit is cleared to indicate that a valid reference is applied to the selected reference pins. The NOREF bit is enabled by setting the REFDET bit in the configuration register to 1. Parity check of the data register. If the ENPAR bit in the mode register is set, the parity bit is set if there is an odd number of 1s in the data register. It is cleared if there is an even number of 1s in the data register. The DAT_STA bit in the mode register should be set when the parity check is used. When the DAT_STA bit is set, the contents of the status register are transmitted along with the data for each data register read. This bit will be set to 0. These bits indicate which channel corresponds to the data register contents. They do not indicate which channel is presently being converted but indicate which channel was selected when the conversion contained in the data register was generated. MODE REGISTER (RS2, RS1, RS0 = 0, 0, 1; Power-On/Reset = 0x080060) The mode register is a 24-bit register from which data can be read or to which data can be written. This register is used to select the operating mode, the output data rate, and the clock source. Table 17 outlines the bit designations for the mode register. MR0 through MR23 indicate the bit locations, MR denoting that the bits are in the mode register. MR23 denotes the first bit of the data stream. The number in parentheses indicates the power-on/reset default status of that bit. Any write to the mode register resets the modulator and filter and sets the RDY bit. MR23 MD2(0) MR15 Sinc3(0) MR7 FS7(0) MR22 MD1(0) MR14 0 MR6 FS6(1) MR21 MD0(0) MR13 ENPAR(0) MR5 FS5(1) MR20 DAT_STA(0) MR12 0 MR4 FS4(0) MR19 CLK1(1) MR11 Single(0) MR3 FS3(0) Rev. C | Page 20 of 40 MR18 CLK0(0) MR10 REJ60(0) MR2 FS2(0) MR17 0 MR9 FS9(0) MR1 FS1(0) MR16 0 MR8 FS8(0) MR0 FS0(0) Data Sheet AD7190 Table 17. Mode Register Bit Designations Bit Location MR23 to MR21 MR20 Bit Name MD2 to MD0 DAT_STA MR19 to MR18 CLK1 to CLK0 MR17 to MR16 MR15 SINC3 MR14 MR13 ENPAR MR12 MR11 Single MR10 REJ60 MR9 to MR0 FS9 to FS0 Description Mode select bits. These bits select the operating mode of the AD7190 (see Table 18). This bit enables the transmission of status register contents after each data register read. When DAT_STA is set, the contents of the status register are transmitted along with each data register read. This function is useful when several channels are selected because the status register identifies the channel to which the data register value corresponds. These bits are used to select the clock source for the AD7190. Either the on-chip 4.92 MHz clock or an external clock can be used. The ability to use an external clock allows several AD7190 devices to be synchronized. Also, 50 Hz/60 Hz rejection is improved when an accurate external clock drives the AD7190. CLK1 CLK0 ADC Clock Source 0 0 External crystal. The external crystal is connected from MCLK1 to MCLK2. 0 1 External clock. The external clock is applied to the MCLK2 pin. 1 0 Internal 4.92 MHz clock. Pin MCLK2 is tristated. 1 1 Internal 4.92 MHz clock. The internal clock is available on MCLK2. These bits must be programmed with a Logic 0 for correct operation. Sinc3 filter select bit. When this bit is cleared, the sinc4 filter is used (default value). When this bit is set, the sinc3 filter is used. The benefit of the sinc3 filter compared to the sinc4 filter is its lower settling time when chop is disabled. For a given output data rate, fADC, the sinc3 filter has a settling time of 3/fADC while the sinc4 filter has a settling time of 4/fADC. The sinc4 filter, due to its deeper notches, gives better 50 Hz/60 Hz rejection. At low output data rates, both filters give similar rms noise and similar no missing codes for a given output data rate. At higher output data rates (FS values less than 5), the sinc4 filter gives better performance than the sinc3 filter for rms noise and no missing codes. This bit must be programmed with a Logic 0 for correct operation. Enable parity bit. When ENPAR is set, parity checking on the data register is enabled. The DAT_STA bit in the mode register should be set when the parity check is used. When the DAT_STA bit is set, the contents of the status register are transmitted along with the data for each data register read. This bit must be programmed with a Logic 0 for correct operation. Single cycle conversion enable bit. When this bit is set, the AD7190 settles in one conversion cycle so that it functions as a zero latency ADC. This bit has no affect when multiple analog input channels are enabled or when the single conversion mode is selected. This bit enables a notch at 60 Hz when the first notch of the sinc filter is at 50 Hz. When REJ60 is set, a filter notch is placed at 60 Hz when the sinc filter first notch is at 50 Hz. This allows simultaneous 50 Hz/60 Hz rejection. Filter output data rate select bits. The 10 bits of data programmed into these bits determine the filter cutoff frequency, the position of the first notch of the filter, and the output data rate for the part. In association with the gain selection, it also determines the output noise (and, therefore, the effective resolution) of the device. (see Table 6 through Table 13) When chop is disabled and continuous conversion mode is selected, the output data rate equals Output Data Rate = (fmod/64)/FS where FS is the decimal equivalent of the code in bits FS0 to FS9 and is in the range 1 to 1023, and fmod is the modulator frequency, which is equal to MCLK/16. With a nominal MCLK of 4.92 MHz, this results in a output data rate from 4.69 Hz to 4.8 kHz. With chop disabled, the filter first notch frequency is equal to the output data rate when converting on a single channel. When chop is enabled, the output data rate equals Output Data Rate = (fmod/64)/(N × FS) where: FS is the decimal equivalent of the code in Bit FS0 to Bit FS9 and is in the range 1 to 1023. fmod is the modulator frequency, which is equal to MCLK/16. With a nominal MCLK of 4.92 MHz, this results in a conversion rate from 4.69/N Hz to 4.8/N kHz, where N is the order of the sinc filter. The sinc filter’s first notch frequency is equal to N × Output Data Rate. The chopping introduces notches at odd integer multiples of (Output Data Rate/2). Rev. C | Page 21 of 40 AD7190 Data Sheet Table 18. Operating Modes MD2 0 MD1 0 MD0 0 0 0 1 0 1 0 0 1 1 1 0 0 1 0 1 1 1 0 1 1 1 Mode Continuous conversion mode (default). In continuous conversion mode, the ADC continuously performs conversions and places the result in the data register. The DOUT/ RDY pin and the RDY bit in the status register go low when a conversion is complete. The user can read these conversions by setting the CREAD bit in the communications register to 1, which enables continuous read. When continuous read is enabled, the conversions are automatically placed on the DOUT line when SCLK pulses are applied. Alternatively, the user can instruct the ADC to output each conversion by writing to the communications register. After power-on, a reset, or a reconfiguration of the ADC, the complete settling time of the filter is required to generate the first valid conversion. Subsequent conversions are available at the selected output data rate, which is dependent on filter choice. Single conversion mode. When single conversion mode is selected, the ADC powers up and performs a single conversion on the selected channel. The internal clock requires up to 1 ms to power up and settle. The ADC then performs the conversion, which requires the complete settling time of the filter. The conversion result is placed in the data register, RDY goes low, and the ADC returns to power-down mode. The conversion remains in the data register and RDY remains active (low) until the data is read or another conversion is performed. Idle mode. In idle mode, the ADC filter and modulator are held in a reset state even though the modulator clocks are still provided. Power-down mode. In power-down mode, all AD7190 circuitry, except the bridge power-down switch, is powered down. The bridge power-down switch remains active because the user may need to power up the sensor prior to powering up the AD7190 for settling reasons. The external crystal, if selected, remains active. Internal zero-scale calibration. An internal short is automatically connected to the input. RDY goes high when the calibration is initiated and returns low when the calibration is complete. The ADC is placed in idle mode following a calibration. The measured offset coefficient is placed in the offset register of the selected channel. Internal full-scale calibration. A full-scale input voltage is automatically connected to the input for this calibration. RDY goes high when the calibration is initiated and returns low when the calibration is complete. The ADC is placed in idle mode following a calibration. The measured full-scale coefficient is placed in the full-scale register of the selected channel. A full-scale calibration is required each time the gain of a channel is changed to minimize the fullscale error. System zero-scale calibration. The user should connect the system zero-scale input to the channel input pins as selected by the CH7 to CH0 bits in the configuration register. RDY goes high when the calibration is initiated and returns low when the calibration is complete. The ADC is placed in idle mode following a calibration. The measured offset coefficient is placed in the offset register of the selected channel. A system zero-scale calibration is required each time the gain of a channel is changed. System full-scale calibration. The user should connect the system full-scale input to the channel input pins as selected by the CH7 to CH0 bits in the configuration register. RDY goes high when the calibration is initiated and returns low when the calibration is complete. The ADC is placed in idle mode following a calibration. The measured full-scale coefficient is placed in the full-scale register of the selected channel. A full-scale calibration is required each time the gain of a channel is changed. CONFIGURATION REGISTER (RS2, RS1, RS0 = 0, 1, 0; Power-On/Reset = 0x000117) The configuration register is a 24-bit register from which data can be read or to which data can be written. This register is used to configure the ADC for unipolar or bipolar mode, to enable or disable the buffer, to enable or disable the burnout currents, to select the gain, and to select the analog input channel. Table 19 outlines the bit designations for the configuration register. CON0 through CON23 indicate the bit locations. CON denotes that the bits are in the configuration register. CON23 denotes the first bit of the data stream. The number in parentheses indicates the poweron/reset default status of that bit. Rev. C | Page 22 of 40 Data Sheet CON23 Chop(0) CON15 CH7(0) CON7 Burn(0) AD7190 CON22 0(0) CON14 CH6(0) CON6 REFDET(0) CON21 0(0) CON13 CH5(0) CON5 0(0) CON20 REFSEL(0) CON12 CH4(0) CON4 BUF(1) CON19 0(0) CON11 CH3(0) CON3 U/B (0) CON18 0(0) CON10 CH2(0) CON2 G2(1) CON17 0(0) CON9 CH1(0) CON1 G1(1) CON16 (0) CON8 CH0(1) CON0 G0(1) Table 19. Configuration Register Bit Designations Bit Location CON23 Bit Name Chop CON22, CON21 CON20 REFSEL CON19 to CON16 CON15 to CON8 CH7 to CH0 CON7 Burn CON6 REFDET CON5 CON4 BUF CON3 U/B CON2 to CON0 G2 to G0 Description Chop enable bit. When the chop bit is cleared, chop is disabled. When the chop bit is set, chop is enabled. When chop is enabled, the offset and offset drift of the ADC are continuously minimized. However, this increases the conversion time and settling time of the ADC. For example, when FS = 96 decimal and the sinc4 filter is selected, the conversion time with chop enabled equals 80 ms and the settling time equals 160 ms. With chop disabled, higher conversion rates are allowed. For an FS word of 96 decimal and the sinc4 filter selected, the conversion time is 20 ms and the settling time is 80 ms. However, at low gains, periodic calibrations may be required to remove the offset and offset drift. These bits must be programmed with a Logic 0 for correct operation. Reference select bits. The reference source for the ADC is selected using these bits. REFSEL Reference Voltage 0 External reference applied between REFIN1(+) and REFIN1(−). 1 External reference applied between the P1/REFIN2(+) and P0/REFIN2(-) pins. These bits must be programmed with a Logic 0 for correct operation. Channel select bits. These bits are used to select which channels are enabled on the AD7190. See Table 20. Several channels can be selected, and the AD7190 automatically sequences them. The conversion on each channel requires the complete settling time. When this bit is set to 1, the 500 nA current sources in the signal path are enabled. When burn = 0, the burnout currents are disabled. The burnout currents can be enabled only when the buffer is active and when chop is disabled. Enables the reference detect function. When set, the NOREF bit in the status register indicates when the external reference being used by the ADC is open circuit or less than 0.6 V maximum. The reference detect circuitry only operates when the ADC is active. This bit must be programmed with a Logic 0 for correct operation. Enables the buffer on the analog inputs. If cleared, the analog inputs are unbuffered, lowering the power consumption of the device. If set, the analog inputs are buffered, allowing the user to place source impedances on the front end without contributing gain errors to the system. With the buffer disabled, the voltage on the analog input pins can be from 50 mV below AGND to 50 mV above AVDD. When the buffer is enabled, it requires some headroom; therefore, the voltage on any input pin must be limited to 250 mV within the power supply rails. Polarity select bit. When this bit is set, unipolar operation is selected. When this bit is cleared, bipolar operation is selected. Gain select bits. Written by the user to select the ADC input range as follows: G2 G1 G0 Gain ADC Input Range (5 V Reference, Bipolar Mode) 0 0 0 1 ±5 V 0 0 1 Reserved 0 1 0 Reserved 0 1 1 8 ±625 mV 1 0 0 16 ±312.5 mV 1 0 1 32 ±156.2 mV 1 1 0 64 ±78.125 mV 1 1 1 128 ±39.06 mV Rev. C | Page 23 of 40 AD7190 Data Sheet Table 20. Channel Selection Channel Enable Bits in the Configuration Register CH7 CH6 CH5 CH4 CH3 CH2 CH1 CH0 1 1 1 1 1 1 1 1 Channel Enabled Positive Input Negative Input AIN(+) AIN(−) AIN1 AIN2 AIN3 AIN4 Temperature sensor AIN2 AIN2 AIN1 AINCOM AIN2 AINCOM AIN3 AINCOM AIN4 AINCOM Status Register Bits CHD[2:0] 000 001 010 011 100 101 110 111 Calibration Register Pair 0 1 None 0 0 1 2 3 ID REGISTER DATA REGISTER RS2, RS1, RS0 = 1, 0, 0; Power-On/Reset = 0xX4 (RS2, RS1, RS0 = 0, 1, 1; Power-On/Reset = 0x000000) The conversion result from the ADC is stored in this data register. This is a read-only, 24-bit register. On completion of a read operation from this register, the RDY pin/bit is set. When the DAT_STA bit in the mode register is set to 1, the contents of the status register are appended to each 24-bit conversion. This is advisable when several analog input channels are enabled because the three LSBs of the status register (CHD2 to CHD0) identify the channel from which the conversion originated. The identification number for the AD7190 is stored in the ID register. This is a read-only register. GPOCON REGISTER (RS2, RS1, RS0 = 1, 0, 1; Power-On/Reset = 0x00) The GPOCON register is an 8-bit register from which data can be read or to which data can be written. This register is used to enable the general-purpose digital outputs. Table 21 outlines the bit designations for the GPOCON register. GP0 through GP7 indicate the bit locations. GP denotes that the bits are in the GPOCON register. GP7 denotes the first bit of the data stream. The number in parentheses indicates the power-on/reset default status of that bit. Rev. C | Page 24 of 40 Data Sheet GP7 0(0) GP6 BPDSW(0) AD7190 GP5 GP32EN(0) GP4 GP10EN(0) GP3 P3DAT(0) GP2 P2DAT(0) GP1 P1DAT(0) GP0 P0DAT(0) Table 21. Register Bit Designations Bit Location GP7 GP 6 Bit Name 0 BPDSW GP5 GP32EN GP4 GP10EN GP3 P3DAT GP2 P2DAT GP1 P1DAT GP0 P0DAT Description This bit must be programmed with a Logic 0 for correct operation. Bridge power-down switch control bit. This bit is set by the user to close the bridge power-down switch BPDSW to AGND. The switch can sink up to 30 mA. The bit is cleared by the user to open the bridge powerdown switch. When the ADC is placed in power-down mode, the bridge power-down switch remains active. Digital Output P3 and Digital Output P2 enable. When GP32EN is set, the digital outputs, P3 and P2, are active. When GP32EN is cleared, the P3 and P2 pins are tristated, and the P3DAT and P2DAT bits are ignored. Digital Output P1 and Digital Output P0 enable. When GP10EN is set, the digital outputs, P1 and P0, are active. When GP10EN is cleared, the P1 and P0 outputs are tristated, and the P1DAT and P0DAT bits are ignored. The P1 and P0 pins can be used as a reference input REFIN2 when the REFSEL bit in the configuration register is set to 1. Digital Output P3. When GP32EN is set, the P3DAT bit sets the value of the P3 general-purpose output pin. When P3DAT is high, the P3 output pin is high. When P3DAT is low, the P3 output pin is low. When the GPOCON register is read, the P3DAT bit reflects the status of the P3 pin if GP32EN is set. Digital Output P2. When GP32EN is set, the P2DAT bit sets the value of the P2 general-purpose output pin. When P2DAT is high, the P2 output pin is high. When P2DAT is low, the P2 output pin is low. When the GPOCON register is read, the P2DAT bit reflects the status of the P2 pin if GP32EN is set. Digital Output P1. When GP10EN is set, the P1DAT bit sets the value of the P1 general-purpose output pin. When P1DAT is high, the P1 output pin is high. When P1DAT is low, the P1 output pin is low. When the GPOCON register is read, the P1DAT bit reflects the status of the P1 pin if GP10EN is set. Digital Output P0. When GP10EN is set, the P0DAT bit sets the value of the P0 general-purpose output pin. When P0DAT is high, the P0 output pin is high. When P0DAT is low, the P0 output pin is low. When the GPOCON register is read, the P0DAT bit reflects the status of the P0 pin if GP10EN is set. OFFSET REGISTER FULL-SCALE REGISTER (RS2, RS1, RS0 = 1, 1, 0; Power-On/Reset = 0x800000) (RS2, RS1, RS0 = 1, 1, 1; Power-On/Reset = 0x5XXXX0) The offset register holds the offset calibration coefficient for the ADC. The power-on reset value of the offset register is 0x800000. The AD7190 has four offset registers; therefore, each channel has a dedicated offset register. Each of these registers is a 24-bit read/write register. This register is used in conjunction with its associated full-scale register to form a register pair. The poweron reset value is automatically overwritten if an internal or system zero-scale calibration is initiated by the user. The AD7190 must be placed in power-down mode or idle mode when writing to the offset register. The full-scale register is a 24-bit register that holds the full-scale calibration coefficient for the ADC. The AD7190 has four fullscale registers; therefore, each channel has a dedicated full-scale register. The full-scale registers are read/write registers. However, when writing to the full-scale registers, the ADC must be placed in power-down mode or idle mode. These registers are configured at power-on with factory-calibrated, full-scale calibration coefficients, the calibration being performed at gain = 1. Therefore, every device has different default coefficients. The default value is automatically overwritten if an internal or system full-scale calibration is initiated by the user or if the full-scale register is written to. Rev. C | Page 25 of 40 AD7190 Data Sheet ADC CIRCUIT INFORMATION 5V REFIN1(+) AGND IN+ OUT+ OUT– IN– AVDD DVDD DGND AVDD AIN1 AIN2 AIN3 AIN4 AINCOM MUX Σ-Δ ADC PGA REFERENCE DETECT SERIAL INTERFACE AND CONTROL LOGIC DOUT/RDY DIN SCLK CS SYNC AGND TEMP SENSOR REFIN1(–) P3 P2 BPDSW AD7190 CLOCK CIRCUITRY MCLK1 MCLK2 07640-012 AGND P0/REFIN2(–) P1/REFIN2(+) Figure 18. Basic Connection Diagram OVERVIEW where: fADC is the output data rate. fCLK = master clock (4.92 MHz nominal). FS[9:0] is the decimal equivalent of Bit FS9 to Bit FS0 in the mode register. The AD7190 is an ultralow noise ADC that incorporates a ∑-∆ modulator, a buffer, PGA, and on-chip digital filtering intended for the measurement of wide dynamic range signals such as those in pressure transducers, weigh scales, and strain gauge applications. The part can be configured to have two differential inputs or four pseudo differential inputs that can be buffered or unbuffered. Figure 18 shows the basic connections required to operate the part. The output data rate can be programmed from 4.7 Hz to 4800 Hz; that is, FS[9:0] can have a value from 1 to 1023. The previous equation is valid for both the sinc3 and sinc4 filters. The settling time for the sinc4 filter is equal to tSETTLE = 4/fADC FILTER, OUTPUT DATA RATE, SETTLING TIME The sinc4 filter also leads to better 50 Hz and 60 Hz rejection. While the notch positions are not affected by the order of the filter, the higher order filter has wider notches, which leads to better rejection in the band (±1 Hz) around the notches. It also gives better stop-band attenuation. The benefit of the sinc3 filter is its lower settling time for the same output data rate. tSETTLE = 3/fADC Figure 19 and Figure 20 show the frequency response of the sinc4 and sinc3 filters, respectively, for an output data rate of 50 Hz. 0 –10 –20 Chop Disabled –30 –40 –50 –60 –70 –80 –90 –100 The output data rate (the rate at which conversions are available on a single channel when the ADC is continuously converting) is equal to fADC = fCLK/(1024 × FS[9:0]) Rev. C | Page 26 of 40 0 25 50 75 100 125 FREQUENCY (Hz) Figure 19. Sinc4 Filter Response (50 Hz Output Data Rate) 150 07640-013 At low output data rates (<1 kHz), the noise-free resolution is comparable for the two filter types. However, at the higher output data rates, the sinc4 filter gives better noise free resolution. Whereas the settling time for the sinc3 filter is equal to FILTER GAIN (dB) A ∑-∆ ADC consists of a modulator followed by a digital filter. The AD7190 has two filter options: a sinc3 filter and a sinc4 filter. The filter is selected using the SINC3 bit in the mode register. When SINC3 is set to 0 (default value), the sinc4 filter is selected. The sinc3 filter is selected when SINC3 is set to 1. Data Sheet AD7190 The value of FS[9:0] can be varied from 1 to 1023. This results in an output data rate of 1.173 Hz to 1200 Hz for the sinc4 filter and 1.56 Hz to 1600 Hz for the sinc3 filter. The settling time for the sinc3 or sinc4 filter is equal to 0 –10 –30 tSETTLE = 2/fADC –40 Therefore, with chop enabled, the settling time is reduced for a given output data rate compared to the chop disabled mode. However, for a given FS[9:0] value, the output data rate is less with chop enabled compared with the chop disabled mode. –50 –60 –70 –80 For either the sinc3 or the sinc4 filter, the cutoff frequency f3dB is equal to 0 25 50 75 100 125 150 FREQUENCY (Hz) Figure 20. Sinc3 Filter Response (50 Hz Output Data Rate) The sinc4 filter provides 50 Hz (±1 Hz) rejection in excess of 120 dB, assuming a stable master clock, while the sinc3 filter gives a rejection of 100 dB. The stop-band attenuation is typically 53 dB for the sinc4 filter but equal to 40 dB for the sinc3 filter. f3dB = 0.24 × fADC Figure 21 and Figure 22 show the filter response for the sinc4 and sinc3 filters, respectively, when chop is enabled. As shown in the plots, the stop-band attenuation is less compared with the chop disabled modes. 0 –10 –20 The 3 dB frequency for the sinc4 filter is equal to FILTER GAIN (dB) f3dB = 0.23 × fADC and for the sinc3 filter is equal to f3dB = 0.272 × fADC Chop Enabled –60 –90 –100 0 25 50 75 100 125 150 FREQUENCY (Hz) Figure 21. Sinc4 Filter Response (Output Data Rate = 12.5 Hz, Chop Enabled) 0 –10 –20 –30 –40 –50 –60 –70 fADC = fCLK/(3 × 1024 × FS[9:0]) where: fADC is the output data rate. fCLK = master clock (4.92 MHz nominal). FS[9:0] is the decimal equivalent of Bit FS9 to Bit FS0 in the mode register. –50 –80 FILTER GAIN (dB) Chopping affects the output data rate and settling time of the ADC. For the sinc4 filter, the output data rate is equal to For sinc3 filter, the output data rate is equal to –40 –70 With chop enabled, the ADC offset and offset drift are minimized. When chop is enabled, the analog input pins are continuously swapped. Therefore, with the analog input pins connected in one direction, the settling time of the sinc filter is allowed to elapse until a valid conversion is available. The analog input pins are then inverted and another valid conversion is obtained. Subsequent conversions are then averaged so that the offset is minimized. This continuous swapping of the analog input pins and the averaging of subsequent conversions means that the offset drift is also minimized. fADC = fCLK/(4 × 1024 × FS[9:0]) –30 07640-015 –100 07640-014 –90 –80 –90 –100 0 25 50 75 100 FREQUENCY (Hz) 125 150 07640-016 FILTER GAIN (dB) –20 Figure 22. Sinc3 Filter Response (Output Data Rate = 16.6 Hz, Chop Enabled) Rev. C | Page 27 of 40 AD7190 Data Sheet 0 –10 –20 –30 –40 –50 –60 –70 Figure 23 and Figure 24 show the frequency response of the sinc4 and sinc3 filters, respectively, when the output data rate is programmed to 50 Hz and REJ60 is set to 1. –80 –90 –100 0 0 25 50 75 100 125 150 FREQUENCY (Hz) –20 Figure 25. Sinc4 Filter Response (12.5 Hz Output Data Rate, Chop Enabled, REJ60 = 1) –30 0 –40 –10 –50 –20 –60 –30 FILTER GAIN (dB) –70 –80 0 25 50 75 100 125 150 FREQUENCY (Hz) –50 –60 –70 07640-017 –90 –100 –40 –80 4 Figure 23. Sinc Filter Response (50 Hz Output Data Rate, REJ60 = 1) –90 –100 0 0 25 –10 75 100 125 150 FREQUENCY (Hz) Figure 26. Sinc3 Filter Response (16.7 Hz Output Data Rate, Chop Enabled, REJ60 = 1) –20 FILTER GAIN (dB) 50 07640-126 FILTER GAIN (dB) –10 07640-125 Normal mode rejection is one of the main functions of the digital filter. With chop disabled, 50 Hz rejection is obtained when the output data rate is set to 50 Hz, whereas 60 Hz rejection is achieved when the output data rate is set to 60 Hz. Simultaneous 50 Hz/60 Hz rejection is obtained when the output data rate is set to 10 Hz. Simultaneous 50 Hz/60 Hz rejection can also be achieved using the REJ60 bit in the mode register. When the output data rate is programmed to 50 Hz and the REJ60 bit is set to 1, notches are placed at both 50 Hz and 60 Hz. filter is used. Figure 25 and Figure 26 show the filter response for both output data rates when REJ60 is set to 1. FILTER GAIN (dB) 50 Hz/60 Hz Rejection –30 Zero Latency –40 Zero latency is enabled by setting the SINGLE bit in the mode register to 1. With zero latency, the complete settling time is allowed for each conversion. Therefore, –50 –60 –70 fADC = 1/tSETTLE –80 –100 0 25 50 75 100 FREQUENCY (Hz) 125 150 07640-018 –90 Figure 24. Sinc3 Filter Response (50 Hz Output Data Rate, REJ60 = 1) Again, the sinc4 filter provides better 50 Hz/60 Hz rejection than the sinc3 filter. In addition, better stop-band attenuation is achieved with the sinc4 filter. When chop is enabled, lower output data rates must be used to achieve 50 Hz and 60 Hz rejection. With REJ60 set to 1, an output data rate of 12.5 Hz gives simultaneous 50 Hz/60 Hz rejection when the sinc4 filter is selected, whereas an output data rate of 16.7 Hz gives simultaneous 50 Hz/ 60 Hz rejection when the sinc3 Zero latency means that the output data rate is constant irrespective of the number of analog input channels enabled; the user does not need to consider the effects of channel changes on the output data rate. The disadvantages of zero latency are the increased noise for a given output data rate compared with the nonzero latency mode. For example, when zero latency is not enabled, the AD7190 has a noise-free resolution of 18.5 bits when the output data rate is 50 Hz and the gain is set to 128. When zero latency is enabled, the ADC has a resolution of 17.5 bits peak-to-peak when the output data rate is 50 Hz. The filter response also changes. Figure 19 shows the filter response for the sinc4 filter when the output data rate is 50 Hz (zero latency disabled). Figure 27 shows the filter response when zero latency is enabled and the output data rate Rev. C | Page 28 of 40 Data Sheet AD7190 is 50 Hz (sinc4 filter); 50 Hz rejection is no longer achieved. The ADC needs to operate with an output data rate of 12.5 Hz to obtain 50 Hz rejection when zero latency is enabled. To obtain simultaneous 50 Hz/60 Hz rejection, the REJ60 bit in the mode register can be set when the output data rate is equal to 12.5 Hz. The stop-band attenuation is considerably reduced also (3 dB compared with 53 dB in the nonzero latency mode). For example, if the sinc4 filter is selected, chop is disabled and zero latency is disabled, conversions are available at 1/fADC when converting on a single channel, where fADC is equal to the output data rate. The settling time is equal to tSETTLE = 4/fADC The time required to sample N channels is 4/(fADC × N) 0 –20 RDY –30 CONVERSIONS CHANNEL A –40 CHANNEL B 1/fADC –50 CHANNEL C 07640-019 FILTER GAIN (dB) –10 Figure 28. Channel Sequencer –60 –70 DIGITAL INTERFACE –80 As indicated in the On-Chip Registers section, the programmable functions of the AD7190 are controlled using a set of on-chip registers. Data is written to these registers via the serial interface of the part. Read access to the on-chip registers is also provided by this interface. All communication with the part must start with a write to the communications register. After power-on or reset, the device expects a write to its communications register. The data written to this register determines whether the next operation is a read operation or a write operation and also determines which register this read or write operation affects. Therefore, write access to any of the other registers on the part begins with a write operation to the communications register, followed by a write to the selected register. A read operation from any other register (except when continuous read mode is selected) starts with a write to the communications register, followed by a read operation from the selected register. –100 0 50 100 150 200 250 300 350 400 450 500 550 600 FREQUENCY (Hz) 07640-020 –90 Figure 27. Sinc4 Filter Response (50 Hz Output Data Rate, Zero Latency) Channel Sequencer The AD7190 includes a channel sequencer, which simplifies communications with the device in multichannel applications. The sequencer also optimizes the channel throughput of the device as the sequencer switches channels at the optimum rate rather than waiting for instructions via the SPI interface. Bits CH0 to Bit CH7 in the configuration register are used to enable the required channels. In continuous conversion mode, the ADC selects each of the enabled channels in sequence and performs a conversion on the channel. The RDY pin goes low when a valid conversion is available on each channel. When several channels are enabled, the contents of the status register should be attached to the 24-bit word so that the user can identify the channel that corresponds to each conversion. To attach the status register value to the conversion, Bit DAT_STA in the mode register should be set to 1. When several channels are enabled, the ADC must allow the complete settling time to generate a valid conversion each time that the channel is changed. The AD7190 takes care of this: when a channel is selected, the modulator and filter are reset and the RDY pin is taken high. The AD7190 then allows the complete settling time to generate the first conversion. RDY only goes low when a valid conversion is available. The AD7190 then selects the next enabled channel and converts on that channel. The user can then read the data register while the ADC is performing the conversion on the next channel. The serial interface of the AD7190 consists of four signals: CS, DIN, SCLK, and DOUT/RDY. The DIN line is used to transfer data into the on-chip registers, whereas DOUT/RDY is used for accessing data from the on-chip registers. SCLK is the serial clock input for the device, and all data transfers (either on DIN or DOUT/RDY) occur with respect to the SCLK signal. The DOUT/ RDY pin functions as a data ready signal also, the line going low when a new data-word is available in the output register. It is reset high when a read operation from the data register is complete. It also goes high prior to the updating of the data register to indicate when not to read from the device, to ensure that a data read is not attempted while the register is being updated. CS is used to select a device. It can be used to decode the AD7190 in systems where several components are connected to the serial bus. The time required to read a valid conversion from all enabled channels is equal to tSETTLE × Number of Enabled Channels Rev. C | Page 29 of 40 AD7190 Data Sheet Figure 3 and Figure 4 show timing diagrams for interfacing to the AD7190, with CS being used to decode the part. Figure 3 shows the timing for a read operation from the output shift register of the AD7190, and Figure 4 shows the timing for a write operation to the input shift register. It is possible to read the same word from the data register several times even though the DOUT/RDY line returns high after the first read operation. However, care must be taken to ensure that the read operations have been completed before the next output update occurs. In continuous read mode, the data register can be read only once. The serial interface can operate in 3-wire mode by tying CS low. In this case, the SCLK, DIN, and DOUT/RDY lines are used to communicate with the AD7190. The end of the conversion can be monitored using the RDY bit or pin. This scheme is suitable for interfacing to microcontrollers. If CS is required as a decoding signal, it can be generated from a port pin. For microcontroller interfaces, it is recommended that SCLK idle high between data transfers. The AD7190 can be operated with CS used as a frame synchronization signal. This scheme is useful for DSP interfaces. In this case, the first bit (MSB) is effectively clocked out by CS because CS normally occurs after the falling edge of SCLK in DSPs. The SCLK can continue to run between data transfers, provided the timing numbers are obeyed. The serial interface can be reset by writing a series of 1s to the DIN input. If a Logic 1 is written to the AD7190 DIN line for at least 40 serial clock cycles, the serial interface is reset. This ensures that the interface can be reset to a known state if the interface gets lost due to a software error or some glitch in the system. Reset returns the interface to the state in which it is expecting a write to the communications register. This operation resets the contents of all registers to their power-on values. Following a reset, the user should allow a period of 500 µs before addressing the serial interface. The AD7190 can be configured to continuously convert or to perform a single conversion. See Figure 29 through Figure 31. Single Conversion Mode In single conversion mode, the AD7190 is placed in powerdown mode after conversions. When a single conversion is initiated by setting MD2, MD1, and MD0 to 0, 0, 1, respectively, in the mode register, the AD7190 powers up, performs a single conversion, and then returns to power-down mode. The on-chip oscillator requires 1 ms approximately to power up. DOUT/RDY goes low to indicate the completion of a conversion. When the data word has been read from the data register, DOUT/RDY goes high. If CS is low, DOUT/RDY remains high until another conversion is initiated and completed. The data register can be read several times, if required, even when DOUT/RDY has gone high. If several channels are enabled, the ADC sequences through the enabled channels and performs a conversion on each channel. When a conversion is started, DOUT/RDY goes high and remains high until a valid conversion is available. As soon as the conversion is available, DOUT/RDY goes low. The ADC then selects the next channel and begins a conversion. The user can read the present conversion while the next conversion is being performed. As soon as the next conversion is complete, the data register is updated; therefore, the user has a limited period in which to read the conversion. When the ADC has performed a single conversion on each of the selected channels, it returns to power-down mode. If the DAT_STA bit in the mode register is set to 1, the contents of the status register are output along with the conversion each time that the data read is performed. The four LSBs of the status register indicate the channel to which the conversion corresponds. CS 0x08 0x280060 0x58 DIN DATA 07640-021 DOUT/RDY SCLK Figure 29. Single Conversion Rev. C | Page 30 of 40 Data Sheet AD7190 Continuous Conversion Mode Continuous conversion is the default power-up mode. The AD7190 converts continuously, the RDY bit in the status register going low each time a conversion is complete. If CS is low, the DOUT/RDY line also goes low when a conversion is completed. To read a conversion, the user writes to the communications register, indicating that the next operation is a read of the data register. When the data word has been read from the data register, DOUT/RDY goes high. The user can read this register additional times, if required. However, the user must ensure that the data register is not being accessed at the completion of the next conversion or else the new conversion word is lost. When several channels are enabled, the ADC continuously loops through the enabled channels, performing one conversion on each channel per loop. The data register is updated as soon as each conversion is available. The DOUT/RDY pin pulses low each time a conversion is available. The user can then read the conversion while the ADC converts on the next enabled channel. If the DAT_STA bit in the mode register is set to 1, the contents of the status register are output along with the conversion each time that the data read is performed. The status register indicates the channel to which the conversion corresponds. CS 0x58 0x58 DIN DATA DATA 07640-022 DOUT/RDY SCLK Figure 30. Continuous Conversion Rev. C | Page 31 of 40 AD7190 Data Sheet Continuous Read conversion is complete and the new conversion is placed in the output serial register. Rather than write to the communications register each time a conversion is complete to access the data, the AD7190 can be configured so that the conversions are placed on the DOUT/ RDY line automatically. By writing 01011100 to the communications register, the user need only apply the appropriate number of SCLK cycles to the ADC, and the conversion word is automatically placed on the DOUT/RDY line when a conversion is complete. The ADC should be configured for continuous conversion mode. To exit the continuous read mode, the instruction 01011000 must be written to the communications register while the RDY pin is low. While in the continuous read mode, the ADC monitors activity on the DIN line so that it can receive the instruction to exit the continuous read mode. Additionally, a reset occurs if 40 consecutive 1s are seen on DIN. Therefore, DIN should be held low in continuous read mode until an instruction is to be written to the device. When several channels are enabled, the ADC continuously steps through the enabled channels and performs one conversion on each channel each time that it is selected. DOUT/ RDY pulses low when a conversion is available. When the user applies sufficient SCLK pulses, the data is automatically placed on the DOUT/RDY pin. If the DAT_STA bit in the mode register is set to 1, the contents of the status register are output along with the conversion. The status register indicates the channel to which the conversion corresponds. When DOUT/RDY goes low to indicate the end of a conversion, sufficient SCLK cycles must be applied to the ADC; the data conversion is then placed on the DOUT/RDY line. When the conversion is read, DOUT/RDY returns high until the next conversion is available. In this mode, the data can be read only once. Also, the user must ensure that the data-word is read before the next conversion is complete. If the user has not read the conversion before the completion of the next conversion, or if insufficient serial clocks are applied to the AD7190 to read the word, the serial output register is reset when the next CS 0x5C DIN DATA DATA DATA 07640-023 DOUT/RDY SCLK Figure 31. Continuous Read Rev. C | Page 32 of 40 Data Sheet AD7190 CIRCUIT DESCRIPTION ANALOG INPUT CHANNEL The AD7190 has two differential/four pseudo differential analog input channels which can be buffered or unbuffered. In buffered mode (the BUF bit in the configuration register is set to 1), the input channel feeds into a high impedance input stage of the buffer amplifier. Therefore, the input can tolerate significant source impedances and is tailored for direct connection to external resistive-type sensors such as strain gauges or resistance temperature detectors (RTDs). When BUF = 0, the part is operated in unbuffered mode. This results in a higher analog input current. Note that this unbuffered input path provides a dynamic load to the driving source. Therefore, resistor/capacitor combinations on the input pins can cause gain errors, depending on the output impedance of the source that is driving the ADC input. Table 22 shows the allowable external resistance/capacitance values for unbuffered mode at a gain of 1 such that no gain error at the 20-bit level is introduced. Table 22. External R-C Combination for No 20-Bit Gain Error C (pF) R (Ω) 50 1.4 k 100 850 500 300 1000 230 5000 30 The absolute input voltage range in buffered mode is restricted to a range between AGND + 250 mV and AVDD – 250 mV. Care must be taken in setting up the common-mode voltage so that these limits are not exceeded. Otherwise, there is degradation in linearity and noise performance. The absolute input voltage in unbuffered mode includes the range between AGND – 50 mV and AVDD + 50 mV. The negative absolute input voltage limit does allow the possibility of monitoring small true bipolar signals with respect to AGND. The analog input range must be limited to (AVDD − 1.25 V)/gain because the PGA requires some headroom. Therefore, if AVDD = 5 V, the maximum analog input that can be applied to the AD7190 is 0 to 3.75 V/gain in unipolar mode or ±3.75 V/ gain in bipolar mode. BIPOLAR/UNIPOLAR CONFIGURATION The analog input to the AD7190 can accept either unipolar or bipolar input voltage ranges. A bipolar input range does not imply that the part can tolerate negative voltages with respect to system AGND. In pseudo-differential mode, signals are referenced to AINCOM while in differential mode, signals are referenced to the negative input of the differential pair. For example, if AINCOM is 2.5 V and the AD7190 AIN1 analog input is configured for unipolar mode with a gain of 2, the input voltage range on the AIN1 pin is 2.5 V to 3.75 V when a 2.5 V reference is used. If AINCOM is 2.5 V and the AD7190 AIN1 analog input is configured for bipolar mode with a gain of 2, the analog input range on AIN1 is 1.25 V to 3.75 V. The bipolar/unipolar option is chosen by programming the U/B bit in the configuration register. DATA OUTPUT CODING When the ADC is configured for unipolar operation, the output code is natural (straight) binary with a zero differential input voltage resulting in a code of 00...00, a midscale voltage resulting in a code of 100...000, and a full-scale input voltage resulting in a code of 111...111. The output code for any analog input voltage can be represented as Code = (2N × AIN × gain)/VREF When the ADC is configured for bipolar operation, the output code is offset binary with a negative full-scale voltage resulting in a code of 000...000, a zero differential input voltage resulting in a code of 100...000, and a positive full-scale input voltage resulting in a code of 111...111. The output code for any analog input voltage can be represented as Code = 2N – 1 × [(AIN × gain/VREF) + 1] PGA When the gain stage is enabled, the output from the buffer is applied to the input of the programmable gain array (PGA). The presence of the PGA means that signals of small amplitude can be gained within the AD7190 while still maintaining excellent noise performance. For example, when the gain is set to 128, the rms noise is 8.5 nV, typically, when the output data rate is 4.7 Hz, which is equivalent to 23 bits of effective resolution or 20.5 bits of noise-free resolution. The AD7190 can be programmed to have a gain of 1, 8, 16, 32, 64, and 128 using Bit G2 to Bit G0 in the configuration register. Therefore, with an external 2.5 V reference, the unipolar ranges are from 0 mV to 19.53 mV to 0 V to 2.5 V, and the bipolar ranges are from ±19.53 mV to ±2.5 V. where: AIN is the analog input voltage. gain is the PGA setting (1 to 128). N = 24. CLOCK The AD7190 includes an internal 4.92 MHz clock on-chip. This internal clock has a tolerance of ±4%. Either the internal clock or an external crystal/clock can be used as the clock source to the AD7190. The clock source is selected using the CLK1 and CLK0 bits in the mode register. When an external crystal is used, it must be connected across the MCLK1 and MCLK2 pins. The crystal manufacturer recommends the load capacitances required for the crystal. The MCLK1 and MCLK2 pins of the AD7190 have a capacitance of 15 pF, typically. If an external Rev. C | Page 33 of 40 AD7190 Data Sheet clock source is used, the clock source must be connected to the MCLK2 pin and the MCLK1 pin must be left floating. The internal clock can also be made available at the MCLK2 pin. This is useful when several ADCs are used in an application and the devices need to be synchronized. The internal clock from one device can be used as the clock source for all ADCs in the system. Using a common clock, the devices can be synchronized by applying a common reset to all devices, or the SYNC pin can be pulsed. BURNOUT CURRENTS The AD7190 contains two 500 nA constant current generators, one sourcing current from AVDD to AIN(+) and one sinking current from AIN(−) to AGND, where AIN(+) is the positive analog input terminal and AIN(−) is the negative analog input terminal in differential mode and AINCOM in pseudodifferential mode. The currents are switched to the selected analog input pair. Both currents are either on or off, depending on the burnout current enable (burn) bit in the configuration register. These currents can be used to verify that an external transducer remains operational before attempting to take measurements on that channel. After the burnout currents are turned on, they flow in the external transducer circuit, and a measurement of the input voltage on the analog input channel can be taken. It will take some time for the burnout currents to detect an open circuit condition as the currents will need to charge any external capacitors There are several reasons why a fault condition can be detected. The front-end sensor may be open circuit. It could also mean that the front-end sensor is overloaded, or the reference may be absent and the NOREF bit in the status register is set, thus clamping the data to all 1s. The user needs to check these three cases before making a judgment. If the voltage measured is 0 V, it may indicate that the transducer has short circuited. The current sources work over the normal absolute input voltage range specifications when the analog inputs are buffered and chop is disabled. REFERENCE The ADC has a fully differential input capability for the reference channel. In addition, the user has the option of selecting one of two external reference options (REFIN1(x) or REFIN2(x)). The reference source for the AD7190 is selected using the REFSEL bit in the configuration register. The REFIN2(x) pins are dual purpose: they can function as two general-purpose output pins or as reference pins. When the REFSEL bit is set to 1, these pins automatically function as reference pins. The common-mode range for these differential inputs is from AGND to AVDD. The reference input is unbuffered; therefore, excessive R-C source impedances introduce gain errors. The reference voltage REFIN (REFINx(+) − REFINx(−)) is AVDD nominal, but the AD7190 is functional with reference voltages from 1 V to AVDD. In applications where the excitation (voltage or current) for the transducer on the analog input also drives the reference voltage for the part, the effect of the low frequency noise in the excitation source is removed because the application is ratiometric. If the AD7190 is used in a nonratiometric application, use a low noise reference. Recommended 2.5 V reference voltage sources for the AD7190 include the ADR421 and ADR431, which are low noise references. Also note that the reference inputs provide a high impedance, dynamic load. Because the input impedance of each reference input is dynamic, resistor/capacitor combinations on these inputs can cause dc gain errors, depending on the output impedance of the source driving the reference inputs. Reference voltage sources like those previously recommended (for example, ADR431) typically have low output impedances and are, therefore, tolerant to having decoupling capacitors on REFINx(+) without introducing gain errors in the system. Deriving the reference input voltage across an external resistor means that the reference input sees a significant external source impedance. External decoupling on the REFINx pins is not recommended in this type of circuit configuration. REFERENCE DETECT The AD7190 includes on-chip circuitry to detect whether the part has a valid reference for conversions or calibrations. This feature is enabled when the REFDET bit in the configuration register is set to 1. If the voltage between the selected REFINx(+) and REFINx(–) pins is between 0.3 V and 0.6 V, the AD7190 detects that it no longer has a valid reference. In this case, the NOREF bit of the status register is set to 1. If the AD7190 is performing normal conversions and the NOREF bit becomes active, the conversion result is all 1s. Therefore, it is not necessary to continuously monitor the status of the NOREF bit when performing conversions. It is only necessary to verify its status if the conversion result read from the ADC data register is all 1s. If the AD7190 is performing either an offset or fullscale calibration and the NOREF bit becomes active, the updating of the respective calibration registers is inhibited to avoid loading incorrect coefficients to these registers, and the ERR bit in the status register is set. If the user is concerned about verifying that a valid reference is in place every time a calibration is performed, check the status of the ERR bit at the end of the calibration cycle. RESET The circuitry and serial interface of the AD7190 can be reset by writing consecutive 1s to the device; 40 consecutive 1s are required to perform the reset. This resets the logic, the digital filter, and the analog modulator, whereas all on-chip registers are reset to their default values. A reset is automatically performed on power-up. When a reset is initiated, the user must allow a period of 500 µs before accessing any of the on-chip registers. A reset is useful if the serial interface loses synchronization due to noise on the SCLK line. Rev. C | Page 34 of 40 Data Sheet AD7190 SYSTEM SYNCHRONIZATION TEMPERATURE SENSOR The SYNC input allows the user to reset the modulator and the digital filter without affecting any of the setup conditions on the part. This allows the user to start gathering samples of the analog input from a known point in time, that is, the rising edge of SYNC. SYNC needs to be taken low for four master clock cycles to implement the synchronization function. Embedded in the AD7190 is a temperature sensor. This is selected using the CH2 bit in the configuration register. When the CH2 bit is set to 1, the temperature sensor is enabled. When the temperature sensor is selected and bipolar mode is selected, the device should return a code of 0x800000 when the temperature is 0 K. A one-point calibration is needed to get the optimum performance from the sensor. Therefore, a conversion at 25°C should be recorded and the sensitivity calculated. The sensitivity is approximately 2815 codes/°C. The equation for the temperature sensor is If multiple AD7190 devices are operated from a common master clock, they can be synchronized so that their data registers are updated simultaneously. A falling edge on the SYNC pin resets the digital filter and the analog modulator and places the AD7190 into a consistent, known state. While the SYNC pin is low, the AD7190 is maintained in this state. On the SYNC rising edge, the modulator and filter are taken out of this reset state and, on the next clock edge, the part starts to gather input samples again. In a system using multiple AD7190 devices, a common signal to their SYNC pins synchronizes their operation. This is normally done after each AD7190 has performed its own calibration or has had calibration coefficients loaded into its calibration registers. The conversions from the AD7190s are then synchronized. The part is taken out of reset on the master clock falling edge following the SYNC low-to-high transition. Therefore, when multiple devices are being synchronized, the SYNC pin should be taken high on the master clock rising edge to ensure that all devices begin sampling on the master clock falling edge. If the SYNC pin is not taken high in sufficient time, it is possible to have a difference of one master clock cycle between the devices; that is, the instant at which conversions are available differs from part to part by a maximum of one master clock cycle. The SYNC pin can also be used as a start conversion command. In this mode, the rising edge of SYNC starts conversion, and the falling edge of RDY indicates when the conversion is complete. The disadvantage of this scheme is that the settling time of the filter has to be allowed for each data register update. This means that the rate at which the data register is updated is reduced. For example, if the ADC is configured to use the sinc4 filter, zero latency is disabled and chop is disabled, the data register update takes four times longer. Temp (K) = (Conversion – 0x800000)/2815 K Temp (°C) = Temp (K) – 273 Following the one point calibration, the internal temperature sensor has an accuracy of ±2 °C, typically. BRIDGE POWER-DOWN SWITCH In bridge applications such as strain gauges and load cells, the bridge itself consumes the majority of the current in the system. For example, a 350 Ω load cell requires 15 mA of current when excited with a 5 V supply. To minimize the current consumption of the system, the bridge can be disconnected (when it is not being used) using the bridge power-down switch. Figure 18 shows how the bridge power-down switch is used. The switch can withstand 30 mA of continuous current, and it has an on resistance of 10 Ω maximum. LOGIC OUTPUTS The AD7190 has four general-purpose digital outputs, P0, P1, P2, and P3. These are enabled using the GP32EN and GP10EN bits in the GPOCON register. The pins can be pulled high or low using the P0DAT to P3DAT bits in the GPOCON register; that is, the value at the pin is determined by the setting of the P0DAT to P3DAT bits. The logic levels for these pins are determined by AVDD rather than by DVDD. When the GPOCON register is read, the P0DAT to P3DAT bits reflect the actual value at the pins. This is useful for short-circuit detection. These pins can be used to drive external circuitry, for example, an external multiplexer. If an external multiplexer is used to increase the channel count, the multiplexer logic pins can be controlled via the AD7190 general-purpose output pins. The general-purpose output pins can be used to select the active multiplexer pin. Because the operation of the multiplexer is independent of the AD7190, the AD7190 modulator and filter should be reset using the SYNC pin each time that the multiplexer channel is changed. Rev. C | Page 35 of 40 AD7190 Data Sheet ENABLE PARITY The AD7190 also has a parity check function on-chip that detects 1-bit errors in the serial communications between the ADC and the microprocessor. When the ENPAR bit in the mode register is set to 1, parity is enabled. The contents of the status register must be transmitted along with each 24-bit conversion when the parity function is enabled. To append the contents of the status register to each conversion read, the DAT_STA bit in the mode register should be set to 1. For each conversion read, the parity bit in the status register is programmed so that the overall number of 1s transmitted in the 24-bit data-word is even. Therefore, for example, if the 24-bit conversion contains eleven 1s (binary format), the parity bit is set to 1 so that the total number of 1s in the serial transmission is even. If the microprocessor receives an odd number of 1s, it knows that the data received has been corrupted. The parity function only detects 1-bit errors. For example, two bits of corrupt data can result in the microprocessor receiving an even number of 1s. Therefore, an error condition is not detected. CALIBRATION The AD7190 provides four calibration modes that can be programmed via the mode bits in the mode register. These modes are internal zero-scale calibration, internal full-scale calibration, system zero-scale calibration, and system full-scale calibration. A calibration can be performed at any time by setting the MD2 to MD0 bits in the mode register appropriately. A calibration should be performed when the gain is changed. After each conversion, the ADC conversion result is scaled using the ADC calibration registers before being written to the data register. The offset calibration coefficient is subtracted from the result prior to multiplication by the full-scale coefficient. To start a calibration, write the relevant value to the MD2 to MD0 bits. The DOUT/RDY pin and the RDY bit in the status register go high when the calibration is initiated. When the calibration is complete, the contents of the corresponding calibration registers are updated, the RDY bit in the status register is reset, the DOUT/ RDY pin returns low (if CS is low), and the AD7190 reverts to idle mode. During an internal zero-scale or full-scale calibration, the respective zero input and full-scale input are automatically connected internally to the ADC input pins. A system calibration, however, expects the system zero-scale and system full-scale voltages to be applied to the ADC pins before initiating the calibration mode. In this way, errors external to the ADC are removed. From an operational point of view, treat a calibration like another ADC conversion. A zero-scale calibration, if required, must always be performed before a full-scale calibration. Set the system software to monitor the RDY bit in the status register or the DOUT/RDY pin to determine the end of calibration via a polling sequence or an interrupt-driven routine. With chop disabled, both an internal zero-scale calibration and a system zero-scale calibration require a time equal to the settling time, tSETTLE, (4/fADC for the sinc4 filter and 3/fADC for the sinc3 filter). With chop enabled, an internal zero-scale calibration is not needed because the ADC itself minimizes the offset continuously. However, if an internal zero-scale calibration is performed, the settling time, tSETTLE, (2/fADC) is required to perform the calibration. Similarly, a system zero-scale calibration requires a time of tSETTLE to complete. To perform an internal full-scale calibration, a full-scale input voltage is automatically connected to the selected analog input for this calibration. For a gain of 1, the time required for an internal full-scale calibration is equal to tSETTLE. For higher gains, the internal full-scale calibration requires a time of 2 × tSETTLE. A full-scale calibration is recommended each time the gain of a channel is changed to minimize the full-scale error. A system full-scale calibration requires a time of tSETTLE. With chop disabled, the zero-scale calibration (internal or system zero-scale) should be performed before the system full-scale calibration is initiated. An internal zero-scale calibration, system zero-scale calibration and system full-scale calibration can be performed at any output data rate. An internal full-scale calibration can be performed at any output data rate for which the filter word FS[9:0] is divisible by 16, FS[9:0] being the decimal equivalent of the 10-bit word written to Bit FS9 to Bit FS0 in the mode register. Therefore, internal full-scale calibrations can be performed at output data rates such as 10 Hz or 50 Hz when chop is disabled. Using these lower output data rates results in better calibration accuracy. The offset error is, typically, 100 µV/gain. If the gain is changed, it is advisable to perform a calibration. A zero-scale calibration (an internal zero-scale calibration or system zero-scale calibration) reduces the offset error to the order of the noise. The gain error of the AD7190 is factory calibrated at a gain of 1 with a 5 V power supply at ambient temperature. Following this calibration, the gain error is 0.001%, typically, at 5 V. Table 23 shows the typical uncalibrated gain error for the different gain settings. An internal full-scale calibration reduces the gain error to 0.001%, typically, when the gain is equal to 1. For higher gains, the gain error post internal full-scale calibration is 0.0075%, typically. A system full-sale calibration reduces the gain error to the order of the noise. Table 23. Typical Precalibration Gain Error vs. Gain Gain 8 16 32 64 128 Rev. C | Page 36 of 40 Precalibration Gain Error (%) −0.11 −0.20 −0.23 −0.29 −0.39 Data Sheet AD7190 The AD7190 gives the user access to the on-chip calibration registers, allowing the microprocessor to read the calibration coefficients of the device and also to write its own calibration coefficients from prestored values in the EEPROM. A read of the registers can be performed at any time. However, the ADC must be placed in power-down or idle mode when writing to the registers. The values in the calibration registers are 24-bits wide. The span and offset of the part can also be manipulated using the registers. GROUNDING AND LAYOUT Because the analog inputs and reference inputs are differential, most of the voltages in the analog modulator are commonmode voltages. The high common-mode rejection of the part removes common-mode noise on these inputs. The analog and digital supplies to the AD7190 are independent and separately pinned out to minimize coupling between the analog and digital sections of the device. The digital filter provides rejection of broadband noise on the power supplies, except at integer multiples of the modulator sampling frequency. Connect an R-C filter to each analog input pin to provide rejection at the modulator sampling frequency. A 100 Ω resistor in series with each analog input, a 0.1 μF capacitor between the analog inputs along with a 0.01 μF capacitor from each analog input to AGND is advised. The digital filter also removes noise from the analog and reference inputs provided these noise sources do not saturate the analog modulator. As a result, the AD7190 is more immune to noise interference than a conventional high resolution converter. However, because the resolution of the AD7190 is so high and the noise levels from the converter so low, care must be taken with regard to grounding and layout. The printed circuit board (PCB) that houses the ADC must be designed so that the analog and digital sections are separated and confined to certain areas of the board. This facilitates the use of ground planes that can be easily separated. A minimum etch technique is generally best for ground planes because it gives the best shielding. Although the AD7190 has separate pins for analog and digital ground, the AGND and DGND pins are tied together internally via the substrate. Therefore, the user must not tie these two pins to separate ground planes unless the ground planes are connected together near the AD7190. In systems in which the AGND and DGND are connected somewhere else in the system (that is, the power supply of the system), they should not be connected again at the AD7190 because a ground loop results. In these situations, it is recommended that ground pins of the AD7190 be tied to the AGND plane. In any layout, the user must keep in mind the flow of currents in the system, ensuring that the paths for all currents are as close as possible to the paths the currents took to reach their destinations. Avoid forcing digital currents to flow through the AGND. Avoid running digital lines under the device because this couples noise onto the die and allow the analog ground plane to run under the AD7190 to prevent noise coupling. The power supply lines to the AD7190 must use as wide a trace as possible to provide low impedance paths and reduce the effects of glitches on the power supply line. Shield fast switching signals like clocks with digital ground to prevent radiating noise to other sections of the board, and never run clock signals near the analog inputs. Avoid crossover of digital and analog signals. Run traces on opposite sides of the board at right angles to each other. This reduces the effects of feedthrough through the board. A microstrip technique is by far the best but is not always possible with a double-sided board. In this technique, the component side of the board is dedicated to ground planes, whereas signals are placed on the solder side. Good decoupling is important when using high resolution ADCs. Decouple all analog supplies with 10 μF tantalum in parallel with 0.1 μF capacitors to AGND. To achieve the best from these decoupling components, place them as close as possible to the device, ideally right up against the device. Decouple all logic chips with 0.1 μF ceramic capacitors to DGND. In systems in which a common supply voltage is used to drive both the AVDD and DVDD of the AD7190, it is recommended that the system AVDD supply be used. For this supply, place the recommended analog supply decoupling capacitors between the AVDD pin of the AD7190 and AGND and the recommended digital supply decoupling capacitor between the DVDD pin of the AD7190 and DGND. Rev. C | Page 37 of 40 AD7190 Data Sheet APPLICATIONS INFORMATION bridge power-down switch is connected in series with the cold side of the bridge. In normal operation, the switch is closed and measurements can be taken. In applications in which current consumption is being minimized, the AD7190 can be placed in standby mode, thus significantly reducing the power consumed in the application. In addition, the bridge power-down switch can be opened while in standby mode, thus avoiding unnecessary power consumption by the front-end transducer. When the part is taken out of standby mode and the bridge power-down switch is closed, the user should ensure that the front-end circuitry is fully settled before attempting a read from the AD7190. The AD7190 provides a low-cost, high resolution analog-todigital function. Because the analog-to-digital function is provided by a ∑-∆ architecture, it makes the part more immune to noisy environments, making it ideal for use in sensor measurement and industrial and process control applications. WEIGH SCALES Figure 32 shows the AD7190 being used in a weigh scale application. The load cell is arranged in a bridge network and gives a differential output voltage between its OUT+ and OUT– terminals. Assuming a 5 V excitation voltage, the full-scale output range from the transducer is 10 mV when the sensitivity is 2 mV/V. The excitation voltage for the bridge can be used to directly provide the reference for the ADC because the reference input range includes the supply voltage. For simplicity, external filters are not included in Figure 32. However, an R-C antialias filter should be included on each analog input. This is required because the on-chip digital filter does not provide any rejection around the modulator sampling frequency or multiples of this frequency. Suitable values are a 100 Ω resistor in series with each analog input, a 0.1 μF capacitor between the analog inputs and 0.01 μF capacitors from each analog input pin to AGND. A second advantage of using the AD7190 in transducer-based applications is that the bridge power-down switch can be fully utilized to minimize the power consumption of the system. The 5V REFIN1(+) AGND OUT+ OUT– IN– AIN1 AIN2 AIN3 AIN4 AINCOM MUX DVDD DGND AVDD REFERENCE DETECT AVDD SERIAL INTERFACE AND CONTROL LOGIC Σ-Δ ADC PGA DOUT/RDY DIN SCLK CS SYNC AGND REFIN1(–) TEMP SENSOR P3 P2 BPDSW AD7190 CLOCK CIRCUITRY AGND MCLK1 MCLK2 P0/REFIN2(–) P1/REFIN2(+) Figure 32. Typical Application (Weigh Scale) Rev. C | Page 38 of 40 07640-024 IN+ Data Sheet AD7190 OUTLINE DIMENSIONS 7.90 7.80 7.70 24 13 4.50 4.40 4.30 6.40 BSC 1 12 PIN 1 0.65 BSC 0.15 0.05 0.30 0.19 1.20 MAX SEATING PLANE 0.20 0.09 8° 0° 0.75 0.60 0.45 0.10 COPLANARITY COMPLIANT TO JEDEC STANDARDS MO-153-AD Figure 33. 24-Lead Thin Shrink Small Outline Package [TSSOP] (RU-24) Dimensions shown in millimeters ORDERING GUIDE Models 1, 2 AD7190BRUZ AD7190BRUZ-REEL AD7190WBRUZ AD7190WBRUZ-RL EVAL-AD7190EBZ 1 2 Temperature Range –40°C to +105°C –40°C to +105°C –40°C to +105°C –40°C to +105°C Package Description 24-Lead TSSOP 24-Lead TSSOP 24-Lead TSSOP 24-Lead TSSOP Evaluation Board Package Option RU-24 RU-24 RU-24 RU-24 Z = RoHS Compliant Part. W = Qualified for Automotive Applications. AUTOMOTIVE PRODUCTS The AD7190W models are available with controlled manufacturing to support the quality and reliability requirements of automotive applications. Note that these automotive models may have specifications that differ from the commercial models; therefore, designers should review the Specifications section of this data sheet carefully. Only the automotive grade products shown are available for use in automotive applications. Contact your local Analog Devices account representative for specific product ordering information and to obtain the specific Automotive Reliability reports for these models. Rev. C | Page 39 of 40 AD7190 Data Sheet NOTES ©2008–2013 Analog Devices, Inc. All rights reserved. Trademarks and registered trademarks are the property of their respective owners. D07640-0-2/13(C) Rev. C | Page 40 of 40