AC-DC 200/250 Watts CCB200/250 Series xppower.com • 200 & 250 W Convection Cooled • Very Low Heat Loss • +70 °C Full Power Operation (CCB200) • Very High Efficiency up to 95% • 5 V Standby, Inhibit & Power Fail Signals • IT & Medical (BF) Safety Approvals • 3 Year Warranty Specification Input • See derating curve • 47-63 Hz • 1.8/2.4 A typical at 115 VAC, full load (CCB200/250) 0.9/1.2 A typical at 230 VAC, full load (CCB200/250) Inrush Current • 30 A typical at 230 VAC, cold start at 25 °C Power Factor • >0.9 Earth Leakage Current • 250 µA max at 264 VAC/60 Hz No Load Input Power • <0.5 W when inhibit used (CCB200) Patient Leakage Current • 100 µA max, 264 VAC 60 Hz Input Protection • Internal T5.0 A/250 V fuse in line and neutral Input Voltage Input Frequency Input Current Output • 12-48 VDC (see tables) • ±3% CCB250, ±5% CCB200 (See Mechanical Details) Initial Set Accuracy • ±0.5% V1, ±5% V2 Minimum Load • No minimum load required Start Up Delay • 1 s typical Start Up Rise Time • 0.5 s typical Hold Up Time • 16/20 ms min at 90 VAC (CCB250/200) Drift • ±0.2% after 20 min warm up Line Regulation • ±0.5% max Load Regulation • ±1% V1, ±5% V2 Over/Undershoot • 5% max Transient Response • 4% max. deviation, recovery to within 1% in 500 µs for a 50-75-50% load change Ripple & Noise • 1% pk-pk, 20 MHz bandwidth Overvoltage Protection • 115-140% Vnom, recycle input to reset (V1 only) Overload Protection • 125-165% Inom V1 (CCB250), 110 - 150% Inom V1 (CCB200) Short Circuit Protection • Continuous trip and restart (hiccup mode) Temperature • 0.05%/°C Coefficient Remote On/Off • Uncommitted isolated optocoupler diode, (Inhibit/Enable) powered diode inhibits V1 Output Voltage Output Voltage Trim General Efficiency Isolation Switching Frequency Signals MTBF • Up to 95% • 4000 VAC Input to Output 2 x MOPP, 1500 VAC Input to Ground 1 x MOPP, 1500 VAC Output to Ground 1 x MOPP • See longform datasheet • Power Fail Inhibit, 5 V Standby • 369/230 kHrs to MIL-HDBK-217F at 25 °C, GB (CCB250/200) Environmental Operating Temperature Cooling Operating Humidity Storage Temperature Operating Altitude Shock Vibration EMC & Safety • • • • • • See derating curve & longform datasheet Convection cooled 95% RH, non-condensing -40 °C to +85 °C 3048 m for CCB250, 5000 m for CCB200 ±3 x 30g shocks in each plane, total 18 shocks. 30g = 11ms (±0.5 ms), half sine, conforms to EN60068-2-27 & EN60068-2-47 • Single axis 10 - 500 Hz at 2g sweep and endurance at resonance in all 3 planes. Conforms to EN60068-2-6 Low Voltage PSU EMC • EN61204-3, high severity level Emissions • EN55011/22 level B conducted EN55011/22 level A radiated, See longform datasheet for level A Harmonic Currents • EN61000-3-2, class A, >140w class C Voltage Flicker • EN61000-3-3 Radiated Immunity • EN61000-4-3, level 3 Perf Criteria A EFT/Burst • EN61000-4-4, level 3 Perf Criteria A Surge • EN61000-4-5, installation class 3 Perf Criteria A Conducted Immunity • EN61000-4-6, level 3 Perf Criteria A Dips & Interruptions • EN61000-4-11, 30% 10 ms, 60% 100 ms, 100% 5000 ms, Perf Criteria A, B, B, EN60601-1-2, 30% 500 ms, 60% 100 ms, 100% 10 ms, 100% 5000 ms, Perf Criteria A, A, A, B - 230 VAC. Consult longform datasheet for 115 V operation. Safety Approvals • IEC60950-1 CB report, UL60950-1, TUV EN60950-1, IEC60601-1 CB report, UL60601-1, TUV 60601-1, Equipment Protection • EN60601-1, ANSI/AAMI ES60601-1, Class CSA22.2 No.60601-1 per cUL BF Rated, Including Risk Management, EN60950, UL60950 Output Power Output Voltage V1 Max Output Current V1 200 W 200 W 200 W 200 W 200 W 200 W 12.0 V 15.0 V 24.0 V 28.0 V 48.0 V 56.0 V 16.7 13.3 8.3 7.1 4.2 3.6 5 V Standby (Optional) A A A A A A 5 V/0.5 5 V/0.5 5 V/0.5 5 V/0.5 5 V/0.5 5 V/0.5 Model Number A A A A A A CCB200PS12 CCB200PS15 CCB200PS24 CCB200PS28 CCB200PS48 CCB200PS56 Notes 1. For covered version add suffix -C to model number e.g. CCB200PS12-C 2. Add suffix -A for 5 V standby option, or -AC for standby and cover options combined. Mechanical Details 0.30 (7.50) 5.00 ±0.02 (127.0 ±0.5) ø0.14 (3.5) thru in 4 positions L3 A C64 C53 2.55 (64.77) 0.92 (23.48) 1 2 T1 0.22 (5.71) 0.22 (5.71) 3.00 ±0.02 (76.2 ±0.5) CCB200PSxx Models Output Voltage Control -VE 1.49 (37.91) 4.55 (115.57) +VE Output Terminals M4 Screws in 2 positions. 0.80 (20.2) 1.43 ±0.04 (36.32 ±1.0) Mounting holes married with must be connected to safety earth. Pin 2 Pin 1 Pin 10 Pin 9 Torque to 8lbs-in (90 cNm) max. Detail A Signals Connector CON4 10 way JST B10B-PHDSS 0.13 (3.5) CCB200PSxx-A Models ø0.14 (3.5) thru in 4 positions 5.00 ±0.02 (127.0 ±0.51) 0.30 (7.5) A L3 A C64 2.55 (64.7) B Standby Board (Option -A) T1 0.22 (5.71) Output Voltage Control C53 under standby board 0.92 (23.4) 0.22 (5.71) 3.00 ±0.02 (76.2 ±0.51) -VE 1.49 37.8) S +VE 0.80 (20.2) 4.55 (115.5) Output Terminals M4 Screws in 2 positions. Torque to 8lbs-in Pin 3 (90 cNm) max. Pin 4 Pin 2 Pin 1 1.43 ±0.04 (36.32 ±1.0) Detail B Standby Board O/P Connector Cons JST S48-PH-K-S 0.13 (3.5) Input Connector CON 1 Line Neutral Mates with Molex housing 09-50-1031 and Molex series 5194 crimp terminals. 1 2 Notes 1. All dimensions in inches (mm). 2. Tolerance .xx = ±0.02 (0.50); .xxx = ±0.01 (0.25) Output Connector CON 3 +V1 1 V1 RTN 2 Signals Connector CON 4 Power Fail (Emitter) 1 Power Fail (Collector) 2 Inhibit (Cathode) 3 Inhibit (Anode) 4 N/C 5 N/C 6 - VE Sense 7 M - VE Sense 8 +VE Sense 9 +VE Sense 10 Mates with JST PHDR-10VS housing and JST SPHD-001TP0.5 crimp terminals 3. Weight: 0.88 lb (400 g) approx 1 2 3 4 Standby Output (Option -A) +5 V +5 V 0V 0V Mates with JST PHDR-4 housing and JST SPH-002T-P0.5L crimp terminals AC-DC CCB200 Models and Ratings AC-DC CCB200 Mechanical Details CCB200PSxx-C Models 0.39 (9.85) M3 Mounting Holes in 2 positions 4.72 (120.0) 0.93 (23.7) Pin 2 Pin 1 Pin 10 Pin 9 Detail A Signals Connector CON4 10 way JST B10B-PHDSS 5.50 ±0.02 (139.7 ±0.5) 1.75 (44.4) 0.59 (15.0) Con 4 Output Voltage Control A Con 3 3.48 ±0.02 (88.5 ±0.5) 1 Con 1 -VE 1.73 (43.9) 2 Output Terminals M4 Screws in 2 positions torque to 8 lbs (90 cNm) max. +VE 1.03 (26.1) Con 2 4.72 (120.0) 0.39 (9.85) M3 mounting holes in 4 positions maximum penetration by screw to be 3.0 mm from bottom face of cover Input Connector CON 1 1 Line 2 Neutral 1.97 (50.0) Mates with Molex housing 09-50-1031 and Molex series 5194 crimp terminals. Output Connector CON 3 1 +V1 2 V1 RTN 0.7 (19.3) CCB200PSxx-AC Models 0.39 (9.85) 4.72 (120.0) 0.93 (23.7) Pin 2 Pin 1 Pin 10 Pin 9 Detail A Signals Connector CON4 10 way JST B10B-PHDSS 5.50 ±0.02 (139.7 ±0.5) 1.75 (44.4) Signals Connector CON 4 1 Power Fail (Emitter) 2 Power Fail (Collector) 3 Inhibit (Cathode) 4 Inhibit (Anode) 5 N/C 6 N/C 7 - VE Sense 8 - VE Sense 9 +VE Sense 10 +VE Sense M3 Mounting Holes in 2 positions 0.59 (15.0) Con 4 Output Voltage Control A Con 3 3.48 ±0.02 (88.5 ±0.5) 1 Con 1 B 1.73 (43.9) 2 -VE +VE 1.03 (26.1) 0.39 (9.85) Output Terminals M4 Screws in 2 positions torque to 8 lbs (90 cNm) max. Con 2 4.72 (120.0) M3 mounting holes in 4 positions maximum penetration by screw to be 3.0 mm from bottom face of cover 1.97 (50.0) 0.7 (19.3) Notes 1. All dimensions in inches (mm). 2. Tolerance .xx = ±0.02 (0.50); .xxx = ±0.01 (0.25) 3. Weight: 1.36 lb (618 g) approx Mates with JST PHDR10VS housing and JST SPHD-001T-P0.5 crimp terminals Standby Output (Option -A) 1 2 3 4 +5 V +5 V 0V 0V Mates with JST PHDR-4 housing and JST SPH-002T-P0.5L crimp terminals Output Power Output Current V1 Output Voltage V1 Standby Supply V2 Model Number 25.00 A 5.0 V/0.5 A CCB250PS12 20.00 A 5.0 V/0.5 A CCB250PS15 10.4 A 12.50 A 5.0 V/0.5 A CCB250PS24 28.0 V 8.9 A 10.70 A 5.0 V/0.5 A CCB250PS28 300 W 36.0 V 6.9 A 8.30 A 5.0 V/0.5 A CCB250PS36 300 W 48.0 V 5.2 A 6.25 A 5.0 V/0.5 A CCB250PS48 Nominal Peak(1) Nominal Peak(1) 250 W 300 W 250 W 300 W 12.0 V 20.8 A 15.0 V 16.7 A 250 W 300 W 24.0 V 250 W 300 W 250 W 250 W Notes 1. Peak duration is 500ms max, average power must not exceed 250 W. Mechanical Details 6.00 (152.4) 0.24 (6.1) Output Voltage Adjustment Signal Connector* ø 3.5 through 4 places L2 C12 3.73 (94.6) 0.74 (19.0) 0.14 (3.5) -VE 38.1 (1.5) Mounting holes married with must be connected to safety earth. Signals Connector CON 1 Output Connector CON 3 1 +V1 2 V1 RTN 1 2 3 Pin 10 Pin 1 Pin 9 Input Connector CON 4 Line Neutral Earth Mates with Molex housing 09-50-1051 and Molex series 5194 crimp terminals. Notes 1. All dimensions in inches (mm). 2. Tolerance .xx = ±0.02 (0.50); .xxx = ±0.01 (0.25) 1.98 (50.4) Output Terminals M4 Screws in 2 positions torque to 8 lbs-in (90 cNm) max C21 5.73 (145.42) Pin 2 1 5 V Standby Return 2 5 V Standby 3 5 V Standby Return 4 5 V Standby 5 5 V Standby Return 6 5 V Standby 7 Power Fail (Collector) 8 Power Fail (Emitter) 9 Remote On / Off (Cathode) 10 Remote On / Off (Anode) +VE 2.60 (66.3) Output connectors allow connections from either top or side L CON 4 N E 0.14 (3.5) CON 3 101.6 (4.0) 3. Weight 1.7 lb (780 g) approx *Signal Connector CON 1 10-way JST-BI0B-PHDSS Mating Half - p/n JST PHDR-10VS, Contact- 26-22 AWG p/n JST SPHD-001T-P0.5 AC-DC CCB250 Models and Ratings Signals - CCB200 Power Fail Inhibit (High) Inhibit (Low) 5V 5V POWER SUPPLY POWER SUPPLY POWER SUPPLY 8V Max 3 mA Pin 2 CON 4 Power Fail Collector 5 mA 8 V Max 1K Inhibit Hi Pin 4 CON 4 Transistor On: Input & Output Not OK Transistor Off: Input & Output OK Power Fail Emitter Pin 1 CON 4 8 V Max Inhibit Hi Pin 4 CON 4 Inhibit Lo Pin 3 CON 4 5 mA VI RTN Pin 7 CON 4 Signal Connector Signals Inhibit Lo Pin 3 CON 4 VI RTN Pin 7 CON 4 Signal Connector Remote On/Off (Inhibit/Enable) Notes & Conditions Uncomitted opto isolated transistor, normally off when AC is good (see fig.5) Provides 5 - 15ms warning of loss of output from AC failure Uncommited isolated optocoupler diode, powered diode inhibits the supply (see fig.6 & 7) 5 V Standby (Optional -A) 5 V/0.5 A Supply, always present when AC applied Power Fail Signal Connector Input Voltage & Temperature Derating - CCB200 Output Power (Watts) 220 Output Power (W) 200 5 Seconds Maximum 180 160 140 120 90 Input Voltage (VAC) 85 80 200 182 Standby Covered Convection (-AC) 150 Covered Convection (-C) 100 0 -20 300 264 Open Frame Convection 160 -10 0 +70 +50 +85 Ambient Temperature (ºC) Signals - CCB250 Power Fail Inhibit (High) POWER SUPPLY Pin 2, 4 & 6 CON 1 5 V Standby 5 V Standby 2, 4 & 6 CON 1 Pin 7 CON 1 Power Fail Collector Power Fail Emitter Pin 8 CON 1 Inhibit (Low) 2K2 5 mA 8V Max 3 mA POWER SUPPLY Transistor On (<0.8 V): Input & Output Not OK Transistor Off (>4.5 V): Input & Output OK 8 V Max 8 V Max Inhibit Hi Pin 10 CON 1 Inhibit Lo Pin 9 CON 1 5 mA 1K Inhibit Lo Pin 9 CH 1 1K Logic GND Pin 1, 3 & 5 CON 1 Logic GND Pin 1, 3 & 5 CON 1 Signal Connector Signal Connector Signal Connector Signals Remote On/Off (Inhibit/Enable) Notes & Conditions Uncommitted opto isolated transistor, normally off when AC is good Provides ≥ 5 ms (typically 20-30 ms) warning of loss of output from AC failure Uncommitted isolated optocoupler diode, powered diode inhibits the supply Standby Supply V2 Isolated 5 V/0.5 A supply, always present when AC supplied. Power Fail POWER SUPPLY 5 V Standby Pin 2, 4 & 6 CON 1 Inhibit Hi Pin 10 CON 1 5 V Standby Rtn Pin 1, 3 & 5 CON 1 Input Voltage & Temperature Derating - CCB250 300 300 Peak Power Availability Peak Power Availability Output Power (W) 250 Output Power (W) AC-DC CCB Mechanical Details 200 150 250 200 150 100 100 50 50 0 80 85 90 Input Voltage (VAC) 264 275 -20 -10 0 10 Ambient Temperature (°C) 40 50 60 70 10 Dec 15