MPS HPU100-111 100w desk top type medical power supplies for health care application Datasheet

HPU100 SERIES
100W Desk top type medical power supplies for Home health care applications.
Description:
The HPU100 series of AC/DC switching mode power supplies
provide 100 Watts of continuous output power . All supplies are
UL94V-1 min compliant, include IEC-320-C14 input for worldwide
applications. They are suited for use in Home health care
applications. All models meet FCC Part-18 Calss B and CISPR-11
EN55011 class B emission Limits and are designed to comply with
UL/c-UL(UL 60601-1) ,TUV/T-mark (EN 60601-1) and new CE
requirements. All units are 100% burned in and tested.
Features:
Wide Input Voltage 90 to 260 VAC,47 to 63 Hz
IEC-320-C14 Input Inlet
Single Output
Output Voltage Available From 11VDC Thru 48VDC
Optional Output Connector (See appendix)
Input Surge Current, Over Voltage And Over Load protection
Output Voltage Protection
Active Power Factor Correction
Class I
2 year warranty
Safety Approvals :
Electrical Characteristics:
Sym.
Parameter
Test Conditions
Vin
fin
PF
Po
Vo
Io
Input Voltage
Input Frequency
Power Factor Correction
Output Power Range
Output Voltage Range
Output Current Range
Io=Full load, Vin=230 VAC
Vin=90 to 260 VAC
Iil
Iih
Input Current (Low Line)
Input Current ( High Line)
Io=Full load, Vin=115 VAC
Io=Full load, Vin=230 VAC
I rl
Low Line Inrush Current
I rh
High Line Inrush Current
Operating Voltage
Eff
REG-i
REG -o
Efficiency
Io=Full load, 25 C , Cool start,
Vin= 115VAC
Io=Full load, 25 C , Cool start,
Vin= 230VAC
Io=Full Load, Vin=230VAC
Line Regulation
Load Regulation
Io=Full Load
Vin=230VAC
OVP
Over Voltage Protection
Over Current Protection
Time of Transient Response
Hold-Up Time
Start Up Time
Ripple & Noise (Peak to Peak)
Safety Ground Leakage C urrent
Temperature Coefficient
OCP
Ttr
Thold
Ts
Vrn
Ilk
TC
Min. Typ. Max. Unit
260
90
47
63
0.95
1
100
0
See rating chart
See rating chart
W
V
A
1.35
0.5
A
A
44
50
A
85
100
A
87
90
%
0.5
1
%
3
5
132
150
4
2
1.5
0.5
1
0.15 0.25
-0.04
0.04
%
%
%
mS
mS
S
%
mA
%/ C
Min. Typ. Max.
Unit
85
112
110
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vi n= 10 0V AC
Full Load, Vin=90VAC
Io = Fu ll Load, V i n = 240 V A C/60Hz
All output
VAC
Hz
12
0.3
Environmental :
Sym.
Toper
Tstg
Hr
MTBF
Pd
Parameter
Test Conditions
Operating Temperature
Storage Temperature
Relative Humidity
Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Derate linearly from 100% load at 40 C to 50% load at 70 C
0
-40
5
0.1M
70
85
95
C
C
%
Hrs
HPU100 SERIES
100W Desk top type medical power supplies for Home health care applications.
Safety Specifications:
Sym.
Parameter
Test Conditions
Dielectric Withstanding Voltage
for Primary to secondary
Dielectric Withstanding Voltage
Vpg for
Primary to Ground
Ris Isolation Resistance
CISPR EMI requirements for CISPR-11
EMI requirements for FCC PART-18
FCC
Vps
Min. Typ. Max.
Unit
Primary to secondary
5600
VDC
Primary to ground
2800
VDC
50
B
B
M
CLASS
CLASS
Test Voltage = 2100VDC
Vin=230VAC
Vin=120VAC
Output Voltage And Current Rating Chart (Single Output) :
Model Number
HPU100-105
HPU100-106
HPU100-107
HPU100-108
HPU100-109
HPU100-111
Output Voltage
11
13
16
21
27
40
~
~
~
~
~
~
13
16
21
27
33
48
VDC
VDC
VDC
VDC
VDC
VDC
Output Current
9.09
7.69
6.25
4.76
3.70
2.50
~
~
~
~
~
~
7.69
6.25
4.76
3.70
3.03
2.08
A
A
A
A
A
A
Total Regulation ①
5%
4%
4%
4%
3%
3%
Maximum
Output Power
100W
100W
100W
100W
100W
100W
The output voltage under 30V had been approved by TUV/PSE.
①
The total regulation on each model is required to use AWG#18×3C+ AWG#16×3C/4FT output cable.
The regulation will be changed by modified output cable.
Mechanical Specifications:
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 460-800gs approx.
3. Optional output connector:
See page Appendix.