VER : PA_3 update : 2016.07.21 AC-DC Medical Switching Power Supply MQF240O SERIES 240 Watts KEY FEATURES Open Frame Medical Switching Power Supply 160 Watt with Free Air Convection 240 Watt with 10CFM FAN 4000VAC Input to Output 2MOPP Insulation High Efficiency up to 94% With P.F.C. Function >0.9 <0.5W No Load Input Power Low Leakage Current under 100uA rd ANSI/AAMI ES60601-1, EN60601-1, & IEC60601-1 3.1 Edition Safety Approvals 3-Year Product Warranty (Pending) ELECTRICAL SPECIFICATIONS All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated. Model No. Max Output Wattage (with 10CFM FAN) (W) Max Output Wattage (Free air Convection) Voltage Input Output MQF240O-12S 240 W (Note 4) 47-63 Hz Current (Full load) < 3.0 A max. (115 VAC) / < 1.5 A max. (230 VAC) Inrush Current (<2ms) < 45 A max. (115 VAC) / < 90 A max. (230 VAC) Leakage Current < 0.1mA / 264 VAC (Touch Current) Power Factor PF>0.9 at Full Load Voltage (V.DC.) 12V Voltage Adj Range (V.DC.) ±5% Output Voltage Voltage Accuracy ±2% Current (with 10CFM FAN) (A) max Current (Free air Convection) (A) max Line Regulation ±1% Load Regulation (0-100%) ±1% Efficiency (at 230VAC) Hold-up Time (at 115 VAC) Environment Physical 24V 48V 20 10 5 13.3 6.66 3.33 Pending Pending Pending 150mV 1% Vout 92% 93% ─ (Note 1) Ripple & Noise max. Isolation 90-264 VAC Frequency (Hz) Maximum Capacitive Load MQF240O-48S 160 W (W) Minimum Load Protection MQF240O-24S (Note 2) 93.5% 10 ms min. Over Power Protection Auto recovery, Hiccup mode Over Voltage Protection Zener diode clamp Short Circuit Protection Auto recovery, Hiccup mode Input-Output 4000VAC or 5656VDC Input-FG 2000VAC or 2828VDC Output-FG 1500VAC or 2121VDC Operating Temperature -30°C…+70°C (with derating) Storage Temperature -30°C…+85°C Temperature Coefficient ±0.05%/°C Humidity 95% RH MTBF >250,000 h @ 25°C (MIL-HDBK-217F, Notice 1) Vibration 10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes. Dimension (L x W x H) 4.1 x 2.05 x 1.023 Inches Weight Pending Cooling Method Free convection http://www.archcorp.com.tw TEL: +886-2-26989508 FAX: +886-2-26981319 ( 103.9 x 52.0 x 26.0 mm ) Tolerance 0.5 mm -1- VER : PA_3 update : 2016.07.21 AC-DC Medical Switching Power Supply MQF240O SERIES 240 Watts ELECTRICAL SPECIFICATIONS All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated. Model No. Safety Approval Conducted and radiated EMI EMC MQF240O-12S MQF240O-24S MQF240O-48S cUL / UL Standard: UL 60950-1, CAN/CSA C22.2 No. 60950-1-07 ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10), CAN/CSA-C22.2 No. 60601-1 (2008), 2 x MOPP (Input to Output) EN55011 class B (Pending) ESD EN61000-4-2 air ± 8kV , Contact ± 4Kv Radiated Immunity EN61000-4-3 10V/m Fast Transient EN61000-4-4 ± 2kV (Pending) Surge EN61000-4-5 ±1kV (Pending) Conducted Immunity EN61000-4-6 10Vrms PFMF EN61000-4-8 30A/m Dips EN61000-4-11 30% 10ms Interruption EN61000-4-11 >95% 5000ms (Pending) (Pending) (Pending) (Pending) (Pending) (Pending) NOTE 1. Ripple & Noise are measured at 20MHz of bandwidth with 0.1uF & 47uF parallel capacitor. 2. Hold-up Time measured at 90% Vout. 3. Main Vout must be >50% Load, 12V (Aux) / 0.5A. 4. Please check the derating curve for more details. 5. Strongly recommend to conduct this test with DC Voltage. If customer wishes to test with AC Voltage, please disconnect all Y-Capacitors within Arch power supply. DERATING http://www.archcorp.com.tw TEL: +886-2-26989508 FAX: +886-2-26981319 -2- VER : PA_3 update : 2016.07.21 AC-DC Medical Switching Power Supply MQF240O SERIES ECHANICAL DIMENSION 240 Watts ( Top View ) PIN# Single 1 AC IN (N) 2 AC IN (L) 3 +DC OUT 4 +DC OUT 5 -DC OUT 6 -DC OUT 7 -AUX OUT 8 +AUX OUT BLOCK DIAGRAM Pending http://www.archcorp.com.tw TEL: +886-2-26989508 FAX: +886-2-26981319 -3-