MOLEX TM-638019100

Auto Grip Pull Tester
PT-1500
Auto Grip Pull Tester
Operation Manual
Order No. 63801-9100
Description
Operation
Maintenance
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 1 of 15
Auto Grip Pull Tester
WARNING
NEVER
OPERATE, SERVICE, INSTALL, OR ADJUST THIS PULL TESTER WITHOUT PROPER
INSTRUCTION AND WITHOUT FIRST READING AND UNDERSTANDING THE INSTRUCTIONS
IN THIS MANUAL.
NEVER
REGARDLESS OF WHETHER THE DISPLAY MODULE (FORCE GAUGE) IS ON OR OFF, DO
NOT EXCEED THE CAPACITY OF THE DISPLAY MODULE. AT 230 LBF (105 kgf), THE DISPLAY
WILL FLASH. NEVER EXCEED THE MAXIMUM CAPACITY OF 440 LBF (200 kgf), OR THE LOAD
CELL WILL BE DAMAGED. AVOID SHOCK LOAD.
WARNING! FAILURE TO ALLOW FOR SUFFICIENT OVERLOAD CAPACITY IN THE LOAD, STRUCTURE
AND MOUNTING ELEMENTS, MAY RESULT IN PROPERTY DAMAGE, SERIOUS INJURY, AND
EVEN DEATH.
CAUTION
When mounting display module, after recalibration or repair, use M4 mounting screws with a
maximum insertion depth of 5mm into the display module. Only use mounting hardware supplied.
CAUTION
Hand tighten wire terminal grip only. Do not use tools.
CAUTION
Do not remove the warranty seal or disassemble the display module. Disassembly will void the
warranty.
CAUTION
Molex crimp specifications are valid only when used with Molex terminals, terminators and tooling.
CAUTION
Specifications subject to change without notice.
CAUTION
Always wear safety glasses.
WORK SAFELY AT ALL TIMES
For Service, Contact Your
Local Molex Sales Office
Molex Application Tooling Group
2200 Wellington Court
Lisle, Illinois 60532, USA
Tel: 630-969-4550
Fax: 630-505-0049
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 2 of 15
Auto Grip Pull Tester
Table of Contents
SECTION
1
General Description
2
Setup and Operation
3
Maintenance
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 3 of 15
Auto Grip Pull Tester
Section 1
General Description
1.1
Description
1.2
Features
1.3
Technical Specifications
1.4
Delivery Check
1.5
Tools
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 4 of 15
Auto Grip Pull Tester
General Description
1.1. Description
The PT-1500 Auto Grip Pull Tester, order no. 63801-9100, is a manual, lever-operated digital pull tester for
checking pull strength of crimped terminals. The tester is designed to automatically grip the wire as the hand
lever is pulled down.
The unit is complete with display module, two multi-terminal grips, pull tester frame, and an AC adapter/charger
for the display module. It has a 3-point, NIST certification.
Principal Mechanical Parts of the Auto Grip Pull Tester
See Figure 1-1
1.
2.
3.
4.
63801-9100
63801-9101
63801-9102
63801-9103
63801-9105
5. 63801-9104
4
PT-1500 Pull Tester assembly
Multi-terminal grip
Clamp type terminal grip
A.C. adapter / charger, 120 VAC
A.C. adapter / charger, 230 VAC
Display module
3
1.2. Features
5
Single-stroke, ergonomic pull-down motion
Automatically grips sample wire as lever is pulled
1
Grips wire diameters up to 0.23" (8-30 AWG)
Lbf, kgf, and Newton units are push-button selectable
Digimatic and analog outputs
Display module is easily removed for recalibration or repair
Runs on internal rechargeable batteries or AC adapter (120 or 230 VAC)
NIST Calibration Certificate included
Figure 1-1
2
1.3. Technical Specification
Dimensions See Figure 1-2
Width:
Depth:
Height:
475.0 to 5.45.0mm (18.70 to 21.45")
(Depending on the handle position)
150.00mm
(5.90")
400.0mm
(15.74")
Weight
10.88kg
Doc No: TM-638019100
Revision: B
(24 lbs)
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 5 of 15
Auto Grip Pull Tester
Figure 1-2
63801-9100 Range
(Resolution) ± 0.2% F.S. ± 1 LSD
Capacity (Resolution)
Pound-force Kilogram-force Newtons
63801-9100 220.0 (0.1 lbf) 100.0 (0.1 kgf) 1000 (1 N)
Model
1.4. Delivery Check
Carefully remove the assembly from its shipping container and determine that the following items are included in the
package.
Description
Quantity
63801-9100 Pull Tester assembly
1
63801-9102 Clamp type terminal grip
1
63801-9103 A.C. Adapter / charger (120 VAC) or
1
63801-9105 A.C. Adapter / charger (230 VAC)
1
TM-638019100 Operation Manual
1
1.5. Tools
The following tools are recommended for set up and maintaining the tester:
Open end wrench, 10mm
Hex wrench, 5mm
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 6 of 15
Auto Grip Pull Tester
Section 2
Set-Up and Operation
2.1
Display Module Operation
2.2
LCD Display
2.3
Installation
2.4
General Operation
2.5
Removing the Display Module
2.6
Settings and Functions
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 7 of 15
Auto Grip Pull Tester
Read the following instructions before attempting to operate the Auto Grip Pull Tester.
2.1. Display Module Operation
See Figure 2-1.
Display
To set the measurement units to pound-force (lbf), kilogram-force (kgf), or Newtons (N), follow the steps below:
1. Turn off the display module.
2. Then set the display units by pressing and holding the
ZERO button until the desired measuring unit is
displayed.
3. Press the ON/OFF button.
4. Then press the ZERO button to cycle the units.
5. Push the SEND button to select.
6. The newly selected units will stay as the default.
MEASURING
SHAFT
DISPLAY
ON / OFF
BUTTON
On / Off Button
Power ON, capacity is displayed. After 10 minutes, unit
powers down if no key is pressed.
ZERO
BUTTON
SEND
BUTTON
Peak Button
Press to activate the Peak mode.
PEAK
BUTTON
Send Button
Transmits display value.
Zero Button
Tares weight of attachment and resets peak value.
CHARGER
PORT
RS-232
PORT
Measuring Shaft
Attach any of the included standard attachments or optional
special attachments to measure tension.
Figure 2-1
Charger Port
Recharge the internal NiMH battery or connect directly to an AC power supply by using one of the included AC
charger / adapters, appropriate for your local electrical requirements. (See Figure 1.1).
RS-232 Port
For data transmission to the external device.
2.2 LCD Display
See figure 2-2.
Compression Icon
Indicates the compression measurement.
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 8 of 15
Auto Grip Pull Tester
Tension Icon
Indicates the tension measurement, (used for crimp pulls).
Unit Indicator
Displays selected measuring unit. Display
pound-force (lbf), kilogram-force (kgf) or
Newtons (N).
Battery Icon
Flashes when gauge needs to be recharged.
TENSION
ICON
COMPRESSION
ICON
NEGATIVE
SIGN
GO/NO GO Indicator
-- Under low set point
Between low and high set point
Over high set point
Negative Sign
Will displayed when measuring tension.
Peak Icon
Will displays continuously when peak mode is
active.
PEAK ICON
GO / NO GO
INDICATOR
UNITS
INDICATOR
BATTERY
ICON
Figure 2-2
2.3 Installation (See Figure 2-3)
1. Place the Auto grip Pull Tester securely on a stable work surface.
2. To adjust clearance between the two wire gripping fixtures if necessary, loosen the cap screw on the displaymodule mounting bracket. Adjust the position and
the re-tighten the cap screw.
DISPLAY MODULE
3. To remove the grip, use a 10mm open-end wrench
HEX NUT
to hold the hex nut while turning the grip counterAC ADAPTER
GRIP
clockwise (CW). Installation is the reverse of
removal but do not over tighten the grip on the
display module.
4. Connect the appropriate AC adapter (120 or 230
VAC included) to the display module and plug it
into an AC outlet or use the internal rechargeable
CAP SCREW
battery.
Figure 2-3
2.4
General Operation (See figure 2-4)
5mm HEX WRENCH
1. Press the ON/OFF button to turn on the display module on, press the PEAK button to enter the peak
measuring mode.
2. If using the ring wire terminal grip, turn the turret to select an appropriate aperture and bring the hand lever
forward. See Figure A.
3. Insert the sample so that the wire terminal is held in place by the grip. If necessary, press the ZERO button to
reset. See Figure B.
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 9 of 15
Auto Grip Pull Tester
4. Thread the wire between the gripper and the cylinder which is attached to the end of the hand lever. See
Figure C.
5. Slowly pull the hand lever back to automatically grip the wire and pull off the terminal. See Figure D.
NOTE: Pull the lever slowly and steadily to insure consistent readings.
6. Read the result from the display module and press the ZERO button to reset for the next test.
Figure B
Figure A
Figure 2-4
Figure C
Figure D
2.5 Removing the Display Module
(See Figure 2-5)
The display module can be easily removed for
recalibration or repair. To do this:
1. Loosen the cap screw on the display
module mounting bracket and slide the
mounting bracket off the base.
Remove the (4) M4 Phillips screws from the
back of the mounting plate.
2. Remove the display module.
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
MOUNTING
PLATE
CAP SCREW
5mm HEX
WRENCH
Figure 2-5
UNCONTROLLED COPY
Page 10 of 15
Auto Grip Pull Tester
2.6 Settings and Functions
RS-232C Bi-directional Interface Functions (See Chart below)
Connect the gauge to an external data receiving device, with the optional RS-232C cable (Order No. 638019151). See Optional Cables and Software below.
All gauge functions can be duplicated from a remote location by using the RS-232C interface.
All commands must be sent in uppercase ASCII character format followed by [Enter].
Signal level: RS-232C, 8 data bits, 1 stop bit, and no parity bit.
Baud rate: 19200 bps.
RS-232C Interface Functions (Upper case ASCII format)
Function
Response*
Select “kgf/gf” units
Select “N” units
Select “lbf/ozf” units
Select peak mode
R [Enter] executed E [Enter] error*
Select real time mode
Tare Display
Turn off power
Set high/low set points(4 digit) HHHH=High, LLLL=Low
EHHHHLLLL [Enter]**
Read high/low set points
set point values (4 digit)
HHHH=High, LLLL=Low
Transmit display data
[value][units][mode][Enter]
*E [Enter] response if the command is not accepted.
**Ignore decimal point for high low set points
Command
K [Enter]
N [Enter]
O [Enter]
P [Enter]
T [Enter]
Z [Enter]
Q [Enter]
EHHHHLLLL [Enter]**
E [Enter]**
D [Enter]
Optional Cables and Software
To capture and analyze the peak data from the force gauge. A running log of all the data is displayed along
with a chart. Calculate max/min, average, and standard deviation.
Order No.
63901-9150
Description
Software, Data Acquisition.
63801-9151
RS-232C cable (10', 9 pin female)
63801-9152
RS-232C to USB Adapter
Type
Recharging NI-MH Battery
1. To maximize the life of the battery, power is shut off after 10 minutes of non-use. Automatic shut off is
bypassed when used with the AC adapter/charger.
2. Battery icon will flash when the gauge needs to be recharged.
3. Turn off power. Only use one of the AC adapter / chargers provided, 63801-9103 for 120VAC, 638019105 for 230VAC. Plug into the correct AC output. It takes 10 hours to charge fully.
4. When the gauge is turned off, make sure the AC adapter/charger is disconnected to avoid overcharging.
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 11 of 15
Auto Grip Pull Tester
Section 3
Maintenance
3.1
Cleaning
3.2
Calibration
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 12 of 15
Auto Grip Pull Tester
3.1
Cleaning
CAUTION: Always disconnect power supply before all maintenance
The Auto Grip Pull Tester maintenance will increase instrument life and produce a consistent, crimp strength
measurement capability. The following procedure should be followed at the end of each operation.
1. Always disconnect the power supply from this tool when not in use.
2. For efficient operation, the Auto Grip Pull Tester should be cleaned daily. Use a soft bristle brush to remove
debris from critical areas such as the terminal grip and clamp.
3.2.
Calibration
The display module should be calibrated on a regular schedule (usually annually). Molex does not offer a
Calibration service for the display module. A local calibration service should be used.
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 13 of 15
Auto Grip Pull Tester
For more information use the Quality Crimping Handbook
And Industrial Crimping Handbook
There is no charge for these books, they can be found on the Molex Website (www.molex.com) or contact you
local Molex sales engineer
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 14 of 15
Auto Grip Pull Tester
2 YEAR WARRANTY
(RESTRICTIONS APPLY)
Molex warrants its products to the original purchaser to be free from defects in workmanship and material under
normal use and proper maintenance for two years (one year for adapters, attachments, batteries, and cables) from
original purchase. This warranty shall not be effective if the product has been subject to overload, shock load,
misuse, negligence, accident, or repairs attempted by others than Molex.
During the warranty period, we will, at our option, either repair or replace defective products. Please call our
customer service department for a return authorization number and return the defective product to us with freight
prepaid.
The foregoing warranty constitutes the SOLE AND EXCLUSIVE WARRANTY, and we hereby disclaim all other
warranties, express, statutory or implied, applicable to the products and/or software, including but not limited to all
implied warranties of merchantability, fitness, non-infringement, results, accuracy, security and freedom from
computer virus. In no event shall Molex be liable for any incidental, consequential or punitive damages in
connection with the use of its products and/or software.
CAUTION: To prevent injury, wear eye protection when using this tool.
http://www.molex.com
Doc No: TM-638019100
Revision: B
Release Date: 08-10-09
Revision Date: 10-01-09
UNCONTROLLED COPY
Page 15 of 15