ETC H2-CA-1000

MEMBRAPOR
SPECIFICATION SHEET FOR FOR H2 SENSOR TYPE H2/CA-1000
PERFORMANCE CHARACTERISTICS
PHYSICAL CHARACTERISTICS
Nominal Range
Maximum Overload
Expected Operation Life
Output Signal
Resolution
Temperature Range
Pressure Range
Pressure Coefficient
T90 Response Time
Relative Humidity Range
Weight
Position Sensitivity
Storage Life
0 – 1’000 ppm
2’000 ppm
2 years in air
17 – 40 nA/ppm
1)
Recommended Storage
Temperature
Warranty Period
2 ppm
0 °C to 45 °C 1)
Atmospheric ± 10%
No data
< 50 sec
15 % to 90 % R.H.
non-condensing
< 10 ppm
Typical Baseline Range (pure
air, 20°C)
Maximum Zero Shift (+20°C
to +40°C)
Long Term Output Drift
Recommended Load Resistor
Bias Voltage
Repeatability
Output Linearity
~ 13 g
None
Six months in
container
5 °C – 20 °C
12 months from date
of dispatch
Compact-Size Outline Dimensions
BOTTOM VIEW
No data yet
< 2% signal loss/month
10 Ohm
Not recommended
< 5% of signal
Linear
45°
Reference
45°
Sensing
Counter
17.0 mm
PCD
Nonconnected
Pin
CO cross sensitivity can be > 5% below 0 °C.
24 mm
CROSS-SENSITIVITY DATA
Interfering Gas
Concentration
CO
2)
100 ppm
Reading
< 2 ppm
2)
for temperature ≥ 20 °C
SIDE VIEW
32 mm
27.1 mm
O-Ring
23.7 mm
Performance data conditions:
20 °C, 50% RH and 1013 mbar
APPLICATIONS
H2 Detection in H2/CO-Mixtures
Medical Applications
Safety and Environmental Control
3.5 mm
Pin Projection
1 mm Pins
on 17.0 mm PCD
Note:
The sensor must be operated by using a suitable gas ket
for gas flow exposure
The data contained in this document is for guidance only. Membrapor AG accepts no liability for any consequential losses, injury or damage resulting
from the use of this document or the information contained within it. The data is given for guidance only. Customers should test under their own
conditions, to ensure that the sensors are suitable for their own requirements.
REV.: 6/2004
Phone: +41 43 311 72 00
Fax : +41 43 311 72 01
Email: [email protected]
www.membrapor.ch
Page 1 of 1