NI NIPXI-2533

256-Crosspoint SSR Matrix Modules
NI PXI-2533, NI PXI-2534 NEW!
• PXI-2533: 4x64 (1-wire)
• PXI-2534: 8x32 (1-wire)
• Switch capacity
• ±55 VDC/30 VAC
• 1 A switching/carry
• 55 W
• 400 crosspoints/s
• Unlimited mechanical lifetime
• Unlimited simultaneous
crosspoint connections
• 32,000-step scan list for
determining scanning
• Fully software programmable
• Onboard relay count tracking
Operating Systems
• Windows Vista/XP/2000
Recommended Software
•
•
•
•
•
•
LabVIEW
LabVIEW Real-Time
LabWindows™/CVI
Measurement Studio
NI Switch Executive
NI TestStand
Other Compatible Software
• Microsoft Visual Basic
• C/C++
Driver Software (included)
• NI-SWITCH
Overview
Automatic Scanning
The National Instruments PXI-2533 and PXI-2534 high-density solidstate relay (SSR) matrix modules feature 256 crosspoints in a compact,
single-slot, 3U PXI form factor. The NI PXI-2533 is configured as a
4x64 one-wire matrix and the NI PXI-2534 is configured as an 8x32
one-wire matrix. Because the modules use SSR technology, they offer
unique benefits such as unlimited mechanical lifetime and the ability
to route 55 V at 1 A on all channels simultaneously (unlimited
simultaneous crosspoint connections). These features make the
PXI-2533 and PXI-2534 ideal for routing general-purpose signals in highdensity matrix applications. Use these modules with the NI PXI-4110
programmable power supply and the NI PXI-4071 71⁄2-digit FlexDMM to
build high-density automated test systems up to ±55 VDC/30 VAC and 1 A.
The PXI-2533 and PXI-2534 maximize throughput in automated test
applications through scanning. Scanning improves throughput by
downloading a list of up to 32,000 connections to the switch and cycling
through the list using an event (trigger) without any interruption from the
host processor. Scanning is most efficiently accomplished by mating the
PXI-2533 and PXI-2534 modules with an instrument, such as the PXI4071 71⁄2-digit FlexDMM, which issues a trigger after each measurement.
Software
All PXI switch modules from NI are shipped with NI-SWITCH, an
IVI-compliant driver offering complete functionality for all switch
modules. When programming high-channel-count switch modules such
as the PXI-2533 and PXI-2534, switch management software like
NI Switch Executive can help simplify programming. With NI Switch
Executive, you gain increased development productivity by interactively
configuring and naming switch modules, external connections, and
signal routes. You also increase test code reuse and system performance
using switch programming with NI TestStand test management
software, the NI LabVIEW and LabWindows/CVI development
environments, and NI Measurement Studio for Microsoft Visual Basic 6.0.
Connectivity
The PXI-2533 and PXI-2534 modules use a single 68-pin SCSI connector.
You can connect signals to these modules using the NI SH68-68-S cable
and the NI TBX-68 external terminal block.
Ordering Information
NI PXI-2533............................................................................778572-33
NI PXI-2534............................................................................778572-34
Includes NI-SWITCH and NI-DAQmx driver software.
Accessories
NI TBX-68 connector block....................................................777141-01
NI SH68-68-S 68-pin shielded cable ....................................185262-01
NI Switch Executive
Development System ........................................................778546-01
Deployment Engine ..........................................................778548-00
BUY NOW!
For complete product specifications, pricing, and accessory
information, call 800 813 3693 (U.S.) or go to ni.com/switches.
256-Crosspoint SSR Matrix Modules
Solid-State Relays
Topology Diagrams
A solid-state relay (SSR) is constructed using a photo-sensitive metaloxide semiconductor field-effect transistor (MOSFET) that is controlled
using a light emitting diode (LED). Light from the encapsulated LED
actuates the photo-sensitive MOSFET and allows current to flow through
it. LED actuation provides a galvanic isolation barrier (no charge can
flow) between the control circuitry and the MOSFET. This is analogous to
the isolation barrier formed between the magnetic coil and the
mechanical armature in electromechanical relays.
LED
Control
Circuitry
C0
C1
C2
C3
C63
R0
R1
R2
R3
MOSFET
High Voltage
Load
Mixed Case Isolation Barrier
R2
Diagram 1. PXI-2533 4x64 1-Wire
Figure 1. Solid-State Relay Construction
There are two main advantages to using SSR technology in PXI:
extended lifetime and unlimited simultaneous crosspoint connections.
Because SSRs use solid-state devices (MOSFETS) to switch states
instead of mechanical contacts, they have infinite mechanical lifetime.
PXI switch modules that are based on SSR technology last longer and
help reduce overhead in many applications by eliminating the need for
frequent switching component replacement.
Compared to electromechanical relays, SSRs require very little power to
transition between ON and OFF states.This feature is especially useful
when building PXI switch systems. The PXI backplane supplies limited
power to each slot in the chassis. Most high-density electromechanical
relay-based matrix modules require more power than that available from
the backplane to close all crosspoints simultaneously. SSR-based matrix
modules are not limited in the same way due to the low power consumption
of SSR technology.
The PXI-2533 and PXI-2534 256-crosspoint matrix modules are
high-density switch products that incorporate SSR technology. Both
modules offer unlimited mechanical lifetime and unlimited
simultaneous crosspoint connections while providing the ability to
switch up to 55 V at 1 A on all channels.
C0
C1
C2
C3
C31
R0
R1
R6
R7
R6
Diagram 2. PXI-2534 8x32 1-Wire
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)
256-Crosspoint SSR Matrix Modules
Safety and Compliance
Specifications
All specifications are subject to change without notice.
Visit ni.com/manuals for the most current specifications.
Safety
Topologies
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
PXI-2533 .............................................. 4x64 (1-wire) matrix
PXI-2534 .............................................. 8x32 (1-wire) matrix
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Input Characteristics
Note: For UL and other safety certifications, refer to the product label or
visit ni.com/certification, search by model number or product line, and
click the appropriate link in the Certification column.
All input characteristics are DC, ACrms, or a combination unless
otherwise specified. Typical values are representative of an average unit
operating at room temperature (25 ± 3 °C) unless otherwise specified.
Maximum switching voltage ............. ±55 VDC, 30 VACrms
(channel-to-ground and
channel-to-channel)
Maximum switching current............... 1 A
Maximum switching power ................ 55 W
DC isolation resistance....................... >2 GΩ, typical
Offset voltage ..................................... 2 µV, typical
Total path resistance row-to-column
Typical ............................................. 1 Ω
Maximum........................................ 1.4 Ω
Dynamic Characteristics
FET operate time
Typical ............................................
Maximum........................................
Maximum scan rate ............................
Simultaneous drive limit.....................
Expected relay life ..............................
724 µs
2.5 ms
400 crosspoints/s
256 relays
Unlimited, when operated
within specified limits
Trigger Characteristics
Input trigger
Sources ...........................................
Minimum pulse width.........................
Output trigger
Destinations....................................
Pulse width .....................................
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
• EN 61326 EMC requirements; Minimum Immunity
• EN 55011 Emissions; Group 1, Class A
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Note: For EMC compliance, operate this device according to
product documentation.
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
• 2006/95/EC; Low-Voltage Directive (safety)
• 2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Note: Refer to the Declaration of Conformity (DoC) for this product for
any additional regulatory compliance information. To obtain the DoC for
this product, visit ni.com/certification, search by model number or
product line, and click the appropriate link in the Certification column.
PXI trigger lines 0-7
70 ns
PXI trigger lines 0-7
Programmable (1 to 62 µs)
Physical Characteristics
Relay type ........................................... Solid-state relay (SSR)
I/O connectors..................................... 68-pin male SCSI
Power requirement ............................ 1 W at 3.3 V, typical
8 W at 5 V, typical
(all crosspoints closed)
Dimensions.......................................... 3U, single-slot PXI
Weight................................................. 300 g (12 oz)
Environment
Operating temperature .......................
Storage temperature...........................
Relative humidity ................................
Pollution degree..................................
Approved at altitudes up to 2,000 m.
0 to 55 °C
-20 to 70 °C
5 to 85%, noncondensing
2
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)
IN
TA
PL
AN
MA
IN
NI Services and Support
P
DE
LO
Y
EL
OP
SERVICE
NEEDS
V
DE
NI has the services and support to meet
your needs around the globe and through
the application life cycle – from planning
and development through deployment
and ongoing maintenance. We offer
services and service levels to meet
customer requirements in research,
design, validation, and manufacturing.
Visit ni.com/services.
Training and Certification
NI training is the fastest, most certain route to productivity with our
products. NI training can shorten your learning curve, save development
time, and reduce maintenance costs over the application life cycle. We
schedule instructor-led courses in cities worldwide, or we can hold a
course at your facility. We also offer a professional certification program
that identifies individuals who have high levels of skill and knowledge on
using NI products. Visit ni.com/training.
Professional Services
Our NI Professional Services team is composed of NI applications
and systems engineers and a worldwide National Instruments Alliance
Partner program of more than 600 independent consultants and
integrators. Services range
from start-up assistance to
turnkey system integration.
Visit ni.com/alliance.
OEM Support
We offer design-in consulting and product integration assistance if you
want to use our products for OEM applications. For information about
special pricing and services for OEM customers, visit ni.com/oem.
ni.com • 800 813 3693
National Instruments • [email protected]
Local Sales and Technical Support
In offices worldwide, our staff is local to the country, giving you access
to engineers who speak your language. NI delivers industry-leading
technical support through online knowledge bases, our applications
engineers, and access to 14,000 measurement and automation
professionals within NI Developer Exchange forums. Find immediate
answers to your questions at ni.com/support.
We also offer service programs that provide automatic upgrades to
your application development environment and higher levels of technical
support. Visit ni.com/ssp.
Hardware Services
NI Factory Installation Services
NI Factory Installation Services (FIS) is the fastest and easiest way to
use your PXI or PXI/SCXI combination systems right out of the box.
Trained NI technicians install the software and hardware and configure
the system to your specifications. NI extends the standard warranty by
one year on hardware components (controllers, chassis, modules)
purchased with FIS. To use FIS, simply configure your system online
with ni.com/pxiadvisor.
Calibration Services
NI recognizes the need to maintain properly calibrated devices for
high-accuracy measurements. We provide manual calibration
procedures, services to recalibrate your products, and automated
calibration software specifically designed for use by metrology
laboratories. Visit ni.com/calibration.
Repair and Extended Warranty
NI provides complete repair services for our products. Express repair
and advance replacement services are also available. We offer
extended warranties to help you meet project life-cycle requirements.
Visit ni.com/services.
*351559A-01*
351559A-01
2008-9241-101-D
© 2008 National Instruments Corporation. All rights reserved. CVI, FlexDMM, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, ni.com,
NI TestStand, and SCXI are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Other product and company names listed
are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or
joint-venture relationship with NI.