KNOWLES 2403-263

8x15x1.5 HAC Receiver
Order No.2403-263-00092
©2011 Knowles Electronics
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
Change History
Status Version
Date
ECR
Comment / Changes
Initials of
owner
Draft
A1.0
09.30.10
Draft specification for first samples
Paul. Song
Draft
A1.1
10.20.10
Paul. Song
Draft
A1.2
11.05.10
Draft
A1.3
11.23.10
Draft
A1.3.1
01.21.11
Release
A 1.4
03.11.11
Release
A 1.4.1
03.22.11
Release
A 1.5
04.07.11
Release
A 1.5.1
04.29.11
Updated Chapter 2.1.
Updated the mechanical drawing of receiver[1.1].
Updated the part marking/labelling[1.3].
Updated the mass, material of basket and cover [1.4].
Updated the test results on baffle[2.1].
Updated the test results on BK 3.2 high leak[2.2].
Updated the mechanical drawing of baffle adapter[2.6].
Updated Environmental Tests[4].
Added Recommend PAD areas[1.1].
Added the limit for spring force[1.2].
Updated the part marking/labelling[1.3].
Added forces on component[1.4].
Updated the material of cover [1.5].
Changed the limit of Baffle test[2.1].
Added Maximum diaphragm excursion above the top of
the surface[2.3].
Added Linear Displacement Volume[2.3].
Added Maximum Displacement Volume[2.3].
Removed Maximum usable excursion Xmax[2.3].
Added measure setup[2.8.1].
Updated Environmental Tests[4].
Removed Linear Displacement Volume[2.3].
Removed Maximum Displacement Volume[2.3].
Add weighting detail[2.4.1].
Updated the test results on baffle[2.1].
Updated the test results on BK 3.2 high leak[2.2].
Updated the characteristic field strength sensitivity and
frequency response[2.5].
Updated test sweep range, steps and add microphone
details[2.8].
Updated the schematic diagrammeasurement setup for
Hearing Aid[2.9].
Updated the Packing Information[2.10].
Updated the Main Dimensions [1.1].
Updated the material of basket [1.5].
Updated the material of POT/ TOPPLATE[1.5].
Add tolerance to Nominal characteristic sensitivity[2.3]
Knowles. 2011 All rights reserved.
Paul. Song
Paul. Song
Paul. Song
Paul. Song
Paul. Song
Paul. Song
Paul. Song
page 2 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
Contents
1.
Mechanical Characteristics ................................................................................................. 4
1.1. Main Dimensions ......................................................................................................... 4
1.2. Spring Force ................................................................................................................ 5
1.3. Part Marking/Labeling .................................................................................................. 5
1.4. Forces on Component ................................................................................................. 6
1.5. Material List ................................................................................................................. 6
2.
Electro-Acoustic Characteristics ......................................................................................... 7
2.1. Frequency response .................................................................................................... 7
2.2. Frequency response .................................................................................................... 8
2.3. Electro-Acoustic Parameters ....................................................................................... 9
2.4. Power Handling ........................................................................................................... 9
2.5. Magnetic field for Hearing Aid characteristics ............................................................ 10
2.6. Measurement setup on baffle .................................................................................... 11
2.7. Measurement setup on BK 3.2 high leak ................................................................... 13
2.8. Measured Parameters ............................................................................................... 15
2.9. Measurement setup for Hearing Aid .......................................................................... 15
2.10.
Packing information................................................................................................ 16
2.11.
Technical Delivery Terms ....................................................................................... 16
3.
Environmental Conditions ................................................................................................. 17
3.1. Storage ...................................................................................................................... 17
3.2. Transportation ........................................................................................................... 17
3.3. Functionality .............................................................................................................. 17
4.
Environmental Tests ......................................................................................................... 18
4.1. Qualification Tests ..................................................................................................... 18
4.2. Requalification Tests ................................................................................................. 18
4.3. Sample Size, Sequence ............................................................................................ 18
4.4. Period of Shelf-Life .................................................................................................... 18
4.5. Testing Procedures.................................................................................................... 18
5.
Related Documents .......................................................................................................... 25
Knowles. 2011 All rights reserved.
page 3 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
1. Mechanical Characteristics
1.1. Main Dimensions
_
+
Positive voltage on pin ‘+’ moves membrane in direction of red arrow!
Recommend PAD layout
Knowles. 2011 All rights reserved.
page 4 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
1.2. Spring Force
1
Force (N)
0.8
0.6
0.4
0.2
0
0
0.2
0.4
0.6
0.8
Height (mm)
1.3. Part Marking/Labeling
(Numbers are generic example)
Knowles. 2011 All rights reserved.
page 5 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
1.4. Forces on Component
State
From Back To
Front(cover)
Membrane
Maximum Permanent
Force[N]
Maximum Handling
Force[N]
10
0
15
0
5
10
From Side to Side
1.5. Material List
1. MATERIAL of BASKET:
PC
2. MATERIAL of MEMBRANE:
POLYARYLATE - COMPOUND
3. MATERIAL of POT/ TOPPLATE:
SOFT MAGNETIC IRON, Ni Plating
4. MATERIAL of MAGNET:
Nd Fe B
5. MATERIAL of SPRING CONTACT:
STAINLESS STEEL
6. PLATING of SPRING CONTACT:
2.5-3.5 m Ni, 0.5-0.7 m Au (CONTACT POINT)
7. MATERIAL of COVER:
Brass CuZn, Ni Plating
8. DIMENSION:
8x15x1.5mm
9. MASS:
0.554 g
10.CONNECTIONS:
SPRING CONTACT
11. DIRECTION of CONNECTOR:
VERTICAL TO PCB (ORIENTATED)
Knowles. 2011 All rights reserved.
page 6 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2. Electro-Acoustic Characteristics
2.1. Frequency response
TYPICAL FREQUENCY RESPONSE
measured on baffle according to chapter 2.6
(D=1cm; P=10mW @32)
20
Relative Level (dB re 1KHz)
15
10
5
0
-5
-10
-15
-20
100
1000
Frequency[Hz]
up limit
Typical SPL Curve
low limit
10000
Tolerance window
f
[Hz]
lower limit
[dB re1KHz]
f
[Hz]
upper limit
[dB re1KHz]
160
350
450
999
1000
1001
2000
4000
6500
7000
10000
-14
-3
-3
-3
0
-3
-3
-1
4
-16
-16
160
270
450
999
1000
1001
2000
4000
6500
0
4
4
3
0
3
3
6
14
10000
14
Knowles. 2011 All rights reserved.
page 7 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2.2. Frequency response
TYPICAL FREQUENCY RESPONSE
20
120
10
100
0
80
-10
60
-20
40
-30
20
-40
0
100
SPL Up limit
1000
Frequency (Hz)
SPL Low limit
Typical THD Curve
THD limit
THD (%)
Relative Level (dB re 1 KHz)
measured on BK 3.2 high leak according to chapter 2.7
(P=10mW @32, ERP)
10000
Typical SPL Curve
Tolerance window
f
[Hz]
200
250
350
999
1000
1001
1250
1500
1600
1700
2000
2500
3000
4000
Knowles. 2011 All rights reserved.
lower limit
[dB re1KHz]
-12
-7.3
-2.5
-2.5
0
-2.5
-0.5
4.5
5.5
3.5
-4.5
-11.5
-14
-14
upper limit
[dB re1KHz]
-2
2.5
2.5
2.5
0
2.5
5.5
12.5
13
12.5
6.5
-2.5
-7
-7
f
[Hz]
200
300
500
600
700
800
1000
1250
6000
upper limit
[%]
70
45
14
8
8
10
4
3
3
page 8 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2.3. Electro-Acoustic Parameters
Receiver mounted in baffle acc. to chapter 2.6 and 3.2 high leak ear acc. to chapter 2.7
1. Rated impedance
Z:
32
2. Voice coil resistance
R:
28.8±2.88
3. Resonance frequency
f0:
4. Maximum diaphragm excursion above the top of the surface:
5. Nominal characteristic sensitivity(@3.2 high leak; 1KHz, 10mW)
6. Nominal characteristic sensitivity
(@baffle, 10mW, 1cm; average from 1KHz to 3KHz)
7. THD
300Hz±45Hz
0.2mm
25.5±2.5 dBPa/V
89 dBPa re 20uPa
according to chapter 2.2
2.4. Power Handling
Receiver mounted in lifetime test device (open rear/open front).
1. RATED NOISE POWER
(500h, pink noise with the weighting, crest factor is 12dB)
Type
High Pass
High Pass
Low Pass
Order
1
10
10
10mW (RMS)
F0[Hz]
250
100
3500
2. OVERLOAD POWER
(50h, sinusoidal, 200Hz~6 KHz, allow increase R&B)
30mW (RMS)
3. RATED SINUSOIDAL POWER
(50h, sinusoidal, 200Hz~6 KHz, no damage)
10mW (RMS)
Knowles. 2011 All rights reserved.
page 9 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2.5. Magnetic field for Hearing Aid characteristics
RECEIVER MOUNTED IN MEASUREMENT SETUP ACC. TO Chapter 2.9
1. CHARACT. FIELD STRENGTH SENSITIVITY
at 566mV (10mW)1kHz, measurement plane15mm
Axial : 12.13 dB A/m per V
Radial (Average of 4 Position): 4.50 dB A/m per V
2. FREQUENCY RESPONSE (AXIAL relative to value at 1kHz)
Magnet field strength frequency
response
Level [dB re 1KHz]
4
2
0
-2
Typical Curve
-4
Low limit
-6
Up limit
-8
100
1000
10000
Frequency [Hz]
Knowles. 2011 All rights reserved.
page 10 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2.6. Measurement setup on baffle
2.6.1.
Measured setup
D=1.0cm.
Knowles. 2011 All rights reserved.
page 11 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2.6.2.
Measured adapter
Knowles. 2011 All rights reserved.
page 12 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2.7. Measurement setup on BK 3.2 high leak
2.7.1.
Measured setup
Knowles. 2011 All rights reserved.
page 13 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2.7.2.
Measured adapter
Knowles. 2011 All rights reserved.
page 14 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2.8. Measured Parameters
1.Measure Setup
For the test on baffle, the test signal is cosine log sweep signal, from 10KHz to 100 Hz, 160
steps, used BK 4939 1/4” free field microphone, sn is 2624669.
For the test on BK3.2 high leak, the test signal is cosine log sweep signal, from 10KHz to 100
Hz, 160 steps, used BK 4195 Ear Simulator, sn is 2538591.
2. Sensitivity
Sensitivity is expressed in dB rel 1V/1Pa measured at input voltage of 0.566V using the
measurement setup according chapter 2.7.
3. Frequency response
Frequency response on baffle is measured according to the test setup in chapter 2.6 at input
voltage of 0.566V and checked against the tolerance window defined in chapter 2.1.
Frequency response on BK3.2 high leak is measured according to the test setup in chapter 2.7
at input voltage of 0.566V and checked against the tolerance window defined in chapter 2.2.
4. Total harmonic distortion (THD)
Is measured according to IEC 268-5 (2nd to 5th harmonics) and the test setup in chapter 2.7
at input voltage of 0.566V and checked against the tolerance window defined in chapter 2.2.
2.9. Measurement setup for Hearing Aid
Tests are conducted at Z (Axial) direction, Y and X (Radial) directions
Measurement Positions acc. to ANSI C63.19[4]
Knowles. 2011 All rights reserved.
page 15 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
*Receiver must be mounted on a plastic surface,magnetic ambient levels be low as to not
significantly affect the measurement, magnetic shielded chamber would be better.
2.10. Packing information
See document: Packaging Instruction for 8x15x1.5 HAC Receiver 2403-263-00092-299.
2.11. Technical Delivery Terms
See Chapter 4.
Knowles. 2011 All rights reserved.
page 16 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
3. Environmental Conditions
3.1. Storage
The transducer fulfils the specified data after treatment according to the conditions of
ETS 300 019-2-1
Specification of environmental test: Storage
Test spec. T 1.2: Weather protected, not temperature controlled storage
locations.
3.2. Transportation
The transducer fulfils the specified data after treatment according to the conditions of
ETS 300 019-2-2
Specification of environmental test: Transportation
Test Spec. T 2.3: Public Transportation
3.3. Functionality
The transducer fulfils the specified data after treatment according to the conditions of
ETS 300 019-2-5
ETS 300 019-2-7
Specification of environmental test: Ground vehicle installations
Test spec. T 5.1: Protected installation
Specification of environmental test: Portable and non-stationary use
Test spec. T 7.3E: Partly weather protected and non-weather protected
locations.
Knowles. 2011 All rights reserved.
page 17 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
4. Environmental Tests
4.1. Qualification Tests
According to our milestone plan (Product Creation Process), a complete qualification test will
be done at design validation on products, manufactured under serial conditions and 1x per
year and product family. The qualification process covers all tests described under 4.5 and a
complete inspection takes place.
4.2. Requalification Tests
1x per month and product family samples are taken and submitted to tests described under
4.5.2
4.3. Sample Size, Sequence
Unless otherwise stated 20 arbitrary new samples will be used to perform each test for both,
qualification and requalification test as described under 4.1 and 4.2.
4.4. Period of Shelf-Life
The period of shelf-life is 5 years.
4.5. Testing Procedures
4.5.1.
Storage Tests
Low Temperature Storage Test
Parameter
Low Temperature
Storage
(Ref.
Test Method and
Conditions
-40°C
rel. humidity not
controlled
BSI BS EN-60068-21)
Knowles. 2011 All rights reserved.
Duration
Evaluation Standard
168h
Measurements of acoustic
parameters shall be still within
specification.
Allowed deviations: Sensitivity
is allowed to be within +/-3dB
from initial value;
The distortion shall not
increase by more than a factor
of 1.4 of the initial value
measured at a single
frequency bin.
page 18 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
E.g.: distortion at frequency bin
400Hz which is initially 10% is
allowed to be 14% after the
test.
High Temperature Storage Test
Parameter
Dry Heat Storage
(Ref.
BSI BS EN60068-2-2 )
Test Method and
Conditions
+85°C
rel. humidity not
controlled
Duration
Evaluation Standard
168h
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test.
Duration
Evaluation Standard
20 cycles
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test.
Thermal shock Sequence
Parameter
Change of
Temperature
(Ref. IEC 600682-14)
Test Method and
Conditions
-40°C for 30min and
then move to +85°C
for 30min; Transition
time < 3 min; 20
cycles (40 shocks);
Knowles. 2011 All rights reserved.
page 19 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
Humidity Stress Test
Parameter
Damp heat,
cyclic
(Ref.
IRef. IEC 600682-3)
Test Method and
Conditions
Duration
85°C and 95% RH;
168hrs
Evaluation Standard
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test.
4.5.2. Operating Tests
Cold Operation Test
Parameter
Cold Operation
Test
(Ref.
Test Method and
Conditions
-20°C
rel. humidity not
controlled
BSI BS EN-60068-22,
Signal used to be
agreed upon with
RIM )
Knowles. 2011 All rights reserved.
Duration
Evaluation Standard
72h
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test.
page 20 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
Dry Heat Operation Test
Test Method and
Conditions
+70°C
rel. humidity not
controlled
Parameter
Dry Heat
Operation
(Ref.
Duration
Evaluation Standard
500h
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test.
Duration
Evaluation Standard
3 cycles
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test.
BSI BS EN-60068-21
Signal used to be
agreed upon with
RIM )
4.5.3. Salt Mist Test
Parameter
Salt Mist
(Ref. IEC 600682-52)
Test Method and
Conditions
The part must be
subjected to 2 hours
spray of 5% NaCl salt
mist, at 35°C then be
left at 40°C and 95%
RH for 22 hrs.
4.5.4. Shock Resistance Test (Free Fall Test) - protected product
Parameter
Mechanical shock
(Ref. IEC 60068-
Test Method and
Conditions
Drop in test box or test
adapter; Dropped onto
Knowles. 2011 All rights reserved.
Duration
40 drops
Evaluation Standard
Measurements of acoustic
parameters shall be still
page 21 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
2-32)
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test there shall be
no loose particles dissipated
from the components during
the test
concrete from 1.5m
height; Two drops on
each side (2x6); One
drop on each edge
(1x12); Two drops on
each corner (2x8);
In total of 40 drops
4.5.5. Impact Durability Test (Tumble Test) – protected product
Test Method and
Conditions
Parameter
Impact durability
(in a
Tumble Tester)
(Ref. IEC60068-232Ed)
Impact durability (in a
Tumble Tester)
DUT in drop Test
adapter/Jig 170grams;
Jig size:
120mm*60mm*18mm;
100drops, 1m
Random drops on
steel base.
Duration
Evaluation Standard
100 drops,
1m
DUT power
off
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test there shall be
no loose particles dissipated
from the components during
the test
Duration
Evaluation Standard
4.5.6. G-Force Test
Parameter
G-Force Test
Test Method and
Conditions
Test adapter
Knowles. 2011 All rights reserved.
/
Measurements of acoustic
page 22 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test
There shall be no loose
particles dissipated from the
components during the test
170grams; Jig size:
120mm*60mm*18mm;
5000g, 0.2msec;
4.5.7. Resistance to Electrostatic Discharge
Parameter
Resistance to
ESD
IEC61000-4-2
Level 4
(SPR c 2.5.1)
Test Method and
Duration
Conditions
One pole is grounded 10 exposures
and.the ESD pulse is
on
applied to the other
each polarity
pole. The speaker
/5
must be stressed first
units
with one polarisation
DUT Power
and then with the
off
other polarisation.DUT
must be discharged
between each ESD
exposure. Level 4:
contact +/- 8kV,
air +/- 15kV
Evaluation Standard
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test.
4.5.8. Sine Vibration Test
Parameter
Sine Vibration
Test
(Ref. IEC 600682-6)
Test Method and
Conditions
10–150 Hz, 50 m/sec2
acceleration; 90
minute in each axis;
270 minutes (3 axes)
Knowles. 2011 All rights reserved.
Duration
270minutes
Evaluation Standard
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
page 23 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
Sensitivity is allowed to be
within +/-3dB from initial
value;
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test.
There shall be no loose
particles dissipated from the
components during the test
in total;
4.5.9. Random Vibration Test
Test Method and
Conditions
Random Vibration 5 – 500 Hz, 0.01g2/Hz;
Test
100 minutes at each
Ref. IEC 60068-2- axis, total 300 minutes
64
(3 axes)
Parameter
4.5.10.
Duration
Evaluation Standard
300minutes
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value; The distortion shall
not increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test. There shall be
no loose particles dissipated
from the components during
the test.
Duration
Evaluation Standard
Mechanical Shock
Test Method and
Conditions
Mechanical Shock Peak acceleration
Test
30G, 18 ms duration,
Ref. IEC 60068-2- 3.4m/s (half-sine);
27
Six successive
shocks shall be
applied in each
Parameter
Knowles. 2011 All rights reserved.
18 shocks
Measurements of acoustic
parameters shall be still
within specification.
Allowed deviations:
Sensitivity is allowed to be
within +/-3dB from initial
value;
page 24 of 25
Knowles Electronics
Specification: 8x15x1.5 HAC Receiver
Order No. 2403 263 00092
Revision: A1.5.1 / 04.29.11
direction of three
mutually perpendicular
axes; In total of 18
shocks
The distortion shall not
increase by more than a
factor of 1.4 of the initial
value measured at a single
frequency bin.
E.g.: distortion at frequency
bin 400Hz which is initially
10% is allowed to be 14%
after the test there shall be
no loose particles dissipated
from the components during
the test.
5. Related Documents
EN-60068-2
IPC-SM-785
IEC68-2-14
IEC68-2-3,1984
IEC60068-2-52
IEC60068-2-32
IEC61000-4-2
IEC68-2-27
Knowles. 2011 All rights reserved.
Environmental testing
Guidelines for Accelerated Reliability Testing of Surface
Mount Solder Attachment
Tempreture Change Testing Guidline
Humidity and Tempreture Test guidline
Low air pressure environmentaltesting procedure
Free fall Test Procedure
ESD Test Guidline
Mechanical Shock Test Guidline
page 25 of 25