ETC BM101/4

INSULATION & CONTINUITY TESTERS
MEGGER®
BM100/4
AND BM101/4
• Suitable for testing to
the IEE Wiring
Regulations
• Complies with the
requirements of
BS 7671, HD 384 and
IEC 364
• Combined insulation
and continuity testing
• Insulation testing at
500 V d.c. to the
VDE 0413 Part 1
Analogue Insulation and
Continuity Testers
DESCRIPTION

The MEGGER BM100/4 and BM101/4
Insulation and Continuity Testers
provide an insulation test voltage of
500 V d.c.
The instruments have two continuity
measuring ranges: 0 to 2 Ω and 0 to
200 Ω. The BM100/4 has an automatic
voltage indicating facility that warns,
before a test is executed, if the
instrument has been inadvertently
connected to an energised supply.
This warning is given by a reading on
a 0 to 600 V a.c. scale on the meter;
this scale also monitors the
discharging voltage following tests
on a capacitive circuit. The BM101/4
does not have the voltage indicating
facility, but instead has an additional
resistance range of 0 to 1 MΩ.
A single rotary switch selects each
range, and has a position for
checking the condition of the internal
9 V replaceable battery. A pushbutton is pressed to execute a test,
and upon release, the automatic
discharge circuit is connected across
the instrument’s terminals and
therefore across the item under test.
Test leads must only be removed
when the capacitive circuits have
discharged.
Readings are shown on an analogue
meter; this has a fast response. It has
a white-on-black scale and an orange
“dayglow” pointer for clear, easy
reading in any lighting conditions.
The tough ABS plastic case
incorporates
shrouded
safety
terminal sockets and test leads with
right-angled connectors to prevent
them being accidentally pulled out of
the instrument. There is a shatterproof polycarbonate window for
protection of the meter. A fold away
support stand allows the user to
stand the instrument up rather than
holding it in the hand.
Also
available
as
optional
accessories are 4 mm right-angled
adaptors (black and red) enabling
use of leads with straight connectors.
A third optional accessory is a
synthetic test and carry case.
All instruments are tough and robust,
well able to withstand the treatment
they are likely to receive in everyday
use or while being carried in an
electrician’s toolbag.
• Suitable for testing to the IEE
wiring regulations
APPLICATIONS
The BM100/4 and BM101/4 are
intended
for
insulation
and
continuity testing during installation,
servicing or maintenance work. In
particular, they are suited to
domestic and industrial wiring
systems being tested to the IEE
wiring regulations. They may be used
for testing other items of electrical
equipment
(e.g.,
transformers,
motors, generators, etc.).
The instruments have been designed
to conform to the IEC 1010-1 safety
specification, and to performance
standard VDE 0413 Part 1.
Optional test leads with fused prods
are available for the BM100/4. It is
essential (to comply with HSE
Guidance Note GS 38) that these be
used when checking (by performing a
voltage test) that equipment has
been isolated from the supply,
especially in high energy situations.
FEATURES AND BENEFITS
• Complies with BS 7671, HD 384 and
IEC 364
• Combined insulation and
continuity testing
• Insulation testing at 500 V d.c. to
the VDE 0413 Part 1 specification
• Designed under the IEC 1010-1
safety specification
• Shrouded safety terminals with
right-angled test lead connectors
• Automatic discharge of capacitive
circuits after test
• Lightweight, tough and robust
INSULATION & CONTINUITY TESTERS
SPECIFICATION
General
BM100/4
BM101/4
Insulation Test Voltage Ranges
500 V d.c.
500 V d.c.
Accuracy (at 20° C)
±2,5% of scale length on all ranges
[i.e., ±1,9 mm (0,075 in.) on insulation
resistance range]
Insulation Resistance Range
0 to 200 MW
0 to 200 MW
Resistance Range
—
0 to 1 MW
Continuity Ranges
(i) 0 to 200 W
0 to 200 W
(ii) 0 to 2 W
0 to 2 W
Voltage Range
0 to 600 V a.c.
—
Terminal Voltage D.C.
(nominal on open circuit)
Insulation Resistance Range
<600 V
<600 V
(>500V at 0,5 MΩ)
(>500V at 0,5 MΩ)
Resistance Range
—
53 V
Continuity Ranges
(i) 0,76 V
0,76 V
(ii) 4,7 V
4,7 V
Terminal Current
(nominal on short circuit)
Insulation Resistance Range
1,6 mA
1,6 mA
Resistance Range
—
325 µA
Movement
250 µA full scale deflection
Discharge
Automatic discharge of capacitive
circuits via an internal resistor
(<500 kΩ) when TEST pushbutton is
released following an insulation test
Voltmeter Input Impedance
BM100/4 Only: 330 kΩ
2 ampere, 500 V ceramic HBC type F,
32 x 6 mm, 50 kA breaking capacity
Power Supply
Single 9 volt battery, IEC 6 LR61 type
Current Consumption: 110 mA max.
on insulation range, 40 mA max. on
resistance range, 220 mA max. on
2Ω range
Battery Life
More than 1800 5 second operations
on 2Ω range
Safety
The instruments will, in general, meet
the requirements of the IEC 1010-1
(1990) safety specification. Safety
Class II
Flash test to 4 kV peak impulse
Temperature Coefficient
±0,1% per °C
The instruments have been designed
for use on high energy systems with
phase-to-earth
voltages
not
exceeding 300 V a.c. and with phaseto-phase voltages not exceeding
500 V a.c.
Humidity
Operation: 90% RH max. at 20° C,
80% RH max. at 35° C
Dimensions
175 H x 95 W x 57 D mm
(6,9 H x 3,75 W x 2,25 D in. approx)
Temperature Range
Operation: –5 to +40° C
Storage:
95% RH max. at 35° C
Weight
485 g (1 lb approx)
BM100/4 and BM101/4
Terminal Voltage (V d.c.)
Continuity Ranges
(i) 12,3 mA
12,3 mA
(ii) 210 mA
210 mA
Class 1,5 (as VDE 0413 Part 1) for
insulation resistance range marked
by a white band, Class 3 for other
ranges
Fuses
500 mA, 250 V ceramic HBC type F,
20 x 5 mm, IEC 127/1
600
500
400
300
200
100
0
0 0,2 0,4
0,1 0,3
0,5
1
2
5
10
∞
20 50 200
100
Resistance Under Test (MΩ)
Typical scale (not full size)
BM100/4 (BM101/4 similar)
Typical terminal voltage characteristic
ORDERING INFORMATION
Item (Qty)
Order Code
Insulation and Continuity Tester . . . . . . . . . BM100/4
Insulation and Continuity Tester . . . . . . . . . BM101/4
Included Accessories
Zip-up carry case. . . . . . . . . . . . . . . . . . . . . . . 6420-075
Test leads including prods and clips (1 set). 6220-434
Operating instruction book . . . . . . . . . . . . . . 6171-549
Note: When measuring or detecting voltage, the
advice given in Health and Safety Executive
Guidance Note GS 38 must be followed, particularly
in regard to the use of test leads with fused prods.
Optional Accessories
Order Code
Synthetic test and carry case . . . . . . . . . . . 6420-030
Right-angled adaptors, enabling use of
leads with straight connectors, 4 mm
Black . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6320-176
Red . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6320-177
Test leads with fused prods, FPK4 —
unsuitable for continuity measurements:
comply with Health and Safety Executive
Guidance Note GS 38 (1 set) . . . . . . . . . . . . 6110-920