ETC TDS7104

Digital Phosphor Oscilloscopes
TDS7000B Series
Features & Benefits
Up to 7.25 GHz True
Analog Bandwidth and
Down to 43 ps Rise Time
(20% to 80%)
>400,000 wfms/s Waveform
Capture Rate, Powered by
Exclusive DPX® Acquisition
Technology
20 GS/s Maximum Real-time
Sample Rate
Exceptional Delta-time
Accuracy for High-confidence
in Critical Timing Measurements
Powerful Triggering Features
for Fast Detection of
Relevant Faults
Communications Mask
Testing Up to 4.25 Gb/s Rates
Clock Recovery from Serial
Data Streams up to 3.125 Gb/s
Up to 64 MB Record Length
with MultiView ZoomTM for
Quick Navigation of Long
Records
Superior Measurement Fidelity, Unrivaled Analysis,
Uncompromised Usability
The TDS7000B Series’ unique combination
of superior measurement fidelity, unrivaled
analysis, and uncompromised usability
makes it the ultimate test machine to simplify and speed the design of high-speed,
complex systems. This family offers the
industry’s best solution to the challenging
signal integrity issues faced by designers
verifying, characterizing and debugging
sophisticated electronic designs.
They deliver up to 7.25 GHz (typical) true
analog bandwidth, down to 43 ps rise time
(20% to 80%) and 20 GS/s maximum realtime sample rate to capture critical events
with fine detail. Exclusive DPX® acquisition
technology enables waveform capture rates
up to 400,000 wfms/s to quickly find rare
glitches in seconds or minutes, instead of
hours or days. Innovative software solutions
deliver domain expertise for advanced
analysis and compliance testing, while the
OpenChoice® architecture enables users to
integrate their expertise through the ability
to easily write custom programs or utilize
popular commercial software. The intuitive
graphical user interface delivers sophisticated capability to advanced users without
intimidating occasional users.
TekConnect® Interface for
High Fidelity Connection
Classic Direct Controls,
Touch-sensitive Display, or
Mouse Navigation
OpenChoice® with Windows
2000 Delivers Built-in
Networking and Analysis
XGA 1024x768 display
Applications
Signal Integrity, Jitter and
Timing Analysis
Verification, Debug and
Characterization of
Sophisticated Designs
Design Development and
Compliance Testing of Serial
Data Streams Up to
4.25 Gb/s Rates
Debug of Telecom, Datacom
and Storage Area Network
Equipment Designs and
High-speed Backplanes
Spectral Analysis
Investigation of Transient
Phenomena
1
Oscilloscopes • www.tektronix.com/tds7000B
Digital Phosphor Oscilloscopes
TDS7000B Series
Characteristics
Ve r t i c a l S y s t e m
Input Channels
Hardware Analog Bandwidth (–3 dB)
Rise Time 10% to 90% (typical)
TDS7054
TDS7104
TDS7154B
TDS7254B
TDS7404B
4
4
4
4
4
4
500 MHz
1 GHz
1.5 GHz*1
2.5 GHz*1
4 GHz*1
7.25 GHz*1
(typical), 7 GHz*1
guaranteed
800 ps
400 ps
200 ps
130 ps
100 ps
62 ps
135 ps
83 ps
72 ps
43 ps
Rise Time 20% to 80% (typical)
DC Gain Accuracy
Hardware Bandwidth Limits
Input Coupling
Input Impedance
1%
250 MHz or 20 MHz
±(2% +
(2% x offset))
TDS7704B
±(2.5% +
(2% x offset))
Requires TCA-1 MEG
AC, DC, GND
DC, GND
1 MΩ ±0.5% or 50 Ω ±1%
50 Ω ±2.5%
Input Sensitivity, 1 MΩ
1 mV/div to 10 V/div
—
Input Sensitivity, 50 Ω
1 mV/div to 1 V/div
2 mV/div to 1 V/div
8-Bit (>11-Bit with averaging)
8-Bit (>11-Bit with averaging)
Max Input Voltage, 1 MΩ
±150 V CAT I Derate at
20 dB/decade to
9 VRMS above 200 kHz
—
Max Input Voltage, 50 Ω
5 VRMS with peaks less than
±30 Volts
<1 VRMS for <100 mV/div,
<5 VRMS for ≥100 mV/div settings
Also determined by TekConnect® accessory
Offset Range
1 mV/div to 100 mV/div: ±1 V
101 mV/div to 1 V/div: ±10 V
1.01 V/div to 10 V/div: ±100 V
2 mV to 50 mV/div: ±0.5 V
50.5 mV to 99.5 mV: ±0.25 V
100 mV to 500 mV: ±5 V
505 mV to 1 V/div: ±2.5 V
Channel-to-Channel Isolation
Any Two Channels at Equal
Vertical Scale Settings
≥100:1 at 100 MHz and
≥30:1 at the rated bandwidth
≥80:1 at 1.5 GHz and
≥15:1 at rated bandwidth
Vertical Resolution
*1
At ≥10 mV/div.
Note: Typical system bandwidth of TDS7404B with P7240: 4 GHz.
Note: Typical system bandwidth of TDS7404B with P7330: 3.5 GHz.
Note: Typical system bandwidth of TDS7704B with P7260: 6 GHz
Note: Typical system bandwidth of TDS7704B with P7350: 5 GHz.
2
Oscilloscopes • www.tektronix.com/tds7000B
Digital Phosphor Oscilloscopes
TDS7000B Series
Ti m e B a s e S y s t e m
TDS7054/TDS7104
TDS7154B/TDS7254B/TDS7404B/TDS7704B
200 ps/div to 40 s/div
50 ps to 10 s/div
16 ns to 250 s
5 ns to 250 s
Channel-to-channel Deskew Range
±25 ns
±75 ns
Delta Time Measurement Accuracy
±((0.06/sample rate) +
(15 ppm x reading)) RMS
±((0.06/sample rate) +
(2.5 ppm x reading)) RMS
8 psRMS (typical)
2 psRMS (typical) (7254B/7154B)
1.5 psRMS (typical) (7404B)
1.2 psRMS (typical) (7704B)
±15 ppm over ≥1 ms interval
2.5 ppm over any ≥100 ms interval
Time Base Range
Time Base Delay Time Range
Trigger Jitter (RMS)
Long Term Sample Rate and Delay
Time Accuracy
Acquisition System
TDS7054
TDS7104
TDS7154B/ TDS7254B/ TDS7404B/ TDS7704B
1 channel (max)
5 GS/s
10 GS/s
20 GS/s
2 channels (max)
5 GS/s
5 GS/s
10 GS/s
3-4 channels (max)
2.5 GS/s
2.5 GS/s
5 GS/s
Equivalent Time Sample Rate (max)
250 GS/s
250 GS/s
Real-time Sample Rates
Maximum Record Length per Channel
with Standard Memory
1 TS/s
2 Mb (1-CH.), 1 Mb (2-CH.),
500 Kb (4-CH.)
4 Mb (1-CH.), 2 Mb (2-CH.), 1 Mb (4-CH.)
With Memory Opt. 2M
8 Mb (1-CH.), 4 Mb (2-CH.), 2 Mb (4-CH.)
With Memory Opt. 3M
16 Mb (1-CH.), 8 Mb (2-CH.), 4 Mb (4-CH.)
With Memory Opt. 4M
32 Mb (1-CH.), 16 Mb (2-CH.), 8 Mb (4-CH.)
With Memory Opt. 5M
64 Mb (1-CH.), 32 Mb (2-CH.), 16 Mb (4-CH.)
Oscilloscopes • www.tektronix.com/tds7000B
3
Digital Phosphor Oscilloscopes
TDS7000B Series
Maximum Duration at Highest Real-time Resolution (1-CH)
Time Resolution (Single-shot)
TDS7054
TDS7104
TDS7154B/TDS7254B/TDS7404B/TDS7704B
200 ps (5 GS/s)
100 ps (10 GS/s)
50 ps (20 GS/s)
Max Duration with Standard Memory
400 µs
200 µs
200 µs
Max Duration with Opt. 2M
1.6 ms
800 µs
400 µs
Max Duration with Opt. 3M
3.2 ms
1.6 ms
800 µs
Max Duration with Opt. 4M
1.6 ms
Max Duration with Opt. 5M
3.2 ms
Acquisition Modes
TDS7054/TDS7104
FastAcq Acquisition
Maximum FastAcq Waveform
Capture Rate
Waveform Database
Powered by exclusive DPX acquisition technology, FastAcq optimizes the instrument
for analysis of dynamic signals and capture of infrequent events
>200,000 wfms/sec
>400,000 wfms/sec
Accumulate Waveform Database providing three-dimensional array of amplitude, time and counts
Sample
Acquire sampled values
Peak Detect
Minimum Peak Detect Pulse Width
TDS7154B/TDS7254B/TDS7404B/TDS7704B
®
Captures narrow glitches at all real-time sampling rates
≤1 ns
400 ps
Averaging
From 2 to 10,000 waveforms included in average
Envelope
From 2 to 2x10 9 waveforms included in min-max envelope
Hi-res
FastFrame™ Acquisition
4
Oscilloscopes • www.tektronix.com/tds7000B
Real-time boxcar averaging reduces random noise and increases resolution
Acquisition memory divided into segments; maximum trigger rate >265,000 waveforms per second.
Time of arrival recorded with each event
Digital Phosphor Oscilloscopes
TDS7000B Series
P i n p o i n t ™ Tr i g g e r S y s t e m
TDS7054
TDS7104
TDS7154B/TDS7254B/TDS7404B/TDS7704B
0.35 div DC
to 50 MHz
increasing to
1 div at
500 MHz
0.35 div DC
to 50 MHz
increasing to
1 div at
1 GHz
0.5 div DC to 50 MHz increasing to 1.5 div at 3 GHz
TDS7404B/TDS7704B: 2.7 div at 4 GHz (typical)
400 mV from DC
to 50 MHz
increasing to
750 mV at
100 MHz
250 mV from DC
to 50 MHz
increasing to
500 mV at
100 MHz
150 mV from DC to 50 MHz increasing
to 500 mV at 2.5 GHz
Sensitivity
Internal DC Coupled
External (Auxiliary Input)
Main Trigger Modes
Auto, Normal and Single
Trigger Sequences
Main, Delayed by Time, Delayed by Events. All sequences can include separate horizontal delay
after the trigger event to position the acquisition window in time
Trigger Characteristics
Standard Trigger Types
A Event and Delayed B Event
Trigger Types
Communications-related Triggers
(requires Option SM)
Edge, Glitch, Runt, Width, Transition
Time, Timeout, Pattern, State,
Setup/Hold
Edge, Glitch, Runt, Width, Transition Time, Timeout, Pattern,
State, Setup/Hold, Window – all except Edge,
Pattern and State can be logic qualified by up to two channels
A Event: All above types
A Event and Delayed B Event: All above types
Delayed B Event: Edge
Reset: Reset if B Event does not occur within specified time, before
specified number of events, or before specified State changes
Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ encoded communications signals.
Select among isolated positive or negative one, zero pulse form or eye patterns as applicable to standard
Serial Pattern Trigger
(requires Option ST)
64-Bit serial word recognizer, bits specified in binary
(high, low, don’t care) or hex format.
Trigger on NRZ-encoded data up to 1.25 GBaud
Trigger Level Range
Internal
External (Auxiliary In)
Line
Trigger Coupling
Trigger Holdoff Range
±12 divisions from center of screen
±8 V
±5 V
fixed at 0 V
DC, AC (attenuates <60 Hz), HF Rej (attenuates >30 kHz), LF Rej (attenuates <80 kHz),
Noise Reject (reduces sensitivity)
250 ns minimum to 12 s maximum
Oscilloscopes • www.tektronix.com/tds7000B
5
Digital Phosphor Oscilloscopes
TDS7000B Series
Clock Recovery System (Option SM, ST)
TDS7154B/TDS7254B/TDS7404B/TDS7704B
Clock Recovery Phase Locked Loop Bandwidth
Tracking/Acquisition Range
Clock Recovery Jitter (typical)
Input Sensitivity for Clock Recovery
Input Data Rates
Trigger Modes
Edge – Positive and/or negative slope on any
channel or front panel auxiliary input. Coupling
includes DC, AC, noise reject, HF reject and
LF reject.
Fbaud/1600 typical
±2% of requested baud
0.25% period +5 psRMS for PRBS data pattern or 4 psRMS for repeating “011” data patterns
1 division peak-to-peak displayed signal
1.5 Mbaud to 3.125 Gbaud
Timeout – Trigger on an event which remains high,
low or either, for a specified time period, selectable
from 1 ns (TDS7104/TDS7054) or 340 ps (B models)
to 1 s with 200 ps (TDS7104/TDS7054) or 100 ps
(B models).
Glitch – Trigger on or reject glitches of positive,
negative or either polarity. Minimum glitch width is
1.0 ns with 200 ps resolution (TDS7104/TDS7054).
Minimum glitch width is 170 ps (TDS7704B) or
225 ps (all other B models) with rearm time of
250 ps (B models).
Transition – Trigger on pulse edge rates that are
faster or slower than specified. Slope may be
positive, negative or either.
Width – Trigger on width of positive or negative
pulse (down to 170 ps on B models) either within
or out of selectable time limits – 1 ns (TDS7104/
TDS7054) or 340 ps (B models) to 1 s.
Pattern – Trigger when pattern goes false or stays
true for specified period of time. Pattern (AND, OR,
NAND, NOR) specified for four input channels
defined as HIGH, LOW or Don’t Care.
Runt – Trigger on a pulse that crosses one threshold
but fails to cross a second threshold before crossing
the first again. Optional time qualification.
State – Any logical pattern of channels (1, 2, 3)
clocked by edge on channel 4. Trigger on rising or
falling clock edge.
6
Oscilloscopes • www.tektronix.com/tds7000B
Setup/Hold – Trigger on violations of both setup
time and hold time between clock and data present
on any two input channels.
Window – Trigger on an event that enters or exits
a window defined by two user-adjustable thresholds.
Event can be time or logic qualified (B models only).
Logic Qualified Trigger applicable to Glitch, Width,
Runt, Timeout, Transition, Setup/Hold, Window
triggers – Trigger on the specified event only if the
logic state defined with the remaining unused channels occurs (B models only).
Trigger Delay by Time – 16 ns (5 ns for B models)
to 250 seconds.
Trigger Delay by Events – 1 to 10,000,000 Events.
Digital Phosphor Oscilloscopes
TDS7000B Series
Waveform Measurements
Waveform Processing/Math
Display Characteristics
Amplitude – Amplitude, High, Low, Maximum,
Minimum, Peak-to-Peak, Mean, Cycle Mean, RMS,
Cycle RMS, Positive Overshoot, Negative Overshoot.
Algebraic Expressions – Define extensive algebraic
expressions including waveforms, scalars and
results of parametric measurements e.g. (Integral
(CH.1-Mean(CH.1)) x 1.414).
Display Type – Liquid crystal active-matrix
color display.
Time – Rise Time, Fall Time, Positive Width,
Negative Width, Positive Duty Cycle, Negative Duty
Cycle, Period, Frequency, Delay.
Combination – Area, Cycle Area, Phase, Burst Width.
Histogram-related – Waveform Count, Hits in Box,
Peak Hits, Median, Maximum, Minimum, Peak-toPeak, Mean (µ), Standard Deviation (σ), µ + 1σ,
µ + 2σ, µ + 3σ.
Eye Pattern-related (B models only) – Extinction
Ratio (absolute, %, and dB), Eye Height, Eye Top,
Eye Base, Eye Width, Crossing %, Jitter (Peak-toPeak, RMS and 6σ), Noise (Peak-to-Peak and
RMS), S/N ratio, Cycle Distortion, Q-factor.
Arithmetic – Add, subtract, multiply, divide
waveforms and scalars.
Relational – Boolean result of comparison >, <,
>=, <=, ==, !=.
Calculus – Integrate, differentiate.
Frequency Domain Functions – Spectral magnitude
and phase, real and imaginary spectra.
Display Size – Diagonal: 264 mm (10.4 in.).
Display Resolution –
TDS7104/TDS7054: 640 horizontal x 480
vertical pixels.
TDS7154B/TDS7254B/TDS7404B/TDS7704B:
1024 horizontal x 768 vertical pixels.
Waveform Styles – Vectors, Dots, Variable
Persistence, Infinite Persistence.
Vertical Units – Magnitude: Linear, dB, dBm;
Phase: Degrees, radians.
Window Functions – Rectangular, Hamming,
Hanning, Kaiser-Bessel, Blackman-Harris,
Gaussian, Flattop2, Tek Exponential.
Waveform Definition – As arbitrary math expression.
Oscilloscopes • www.tektronix.com/tds7000B
7
Digital Phosphor Oscilloscopes
TDS7000B Series
Computer System and
Peripherals
CPU –
TDS7104/TDS7054: Intel Celeron processor,
850 MHz.
TDS7154B/TDS7254B/TDS7404B/TDS7704B:
Intel Pentium 4 processor, 2.0 GHz.
PC System Memory – 512 MB.
Hard Disk Drive – 40 GB removable hard disk
drive: rear-panel or (B models only) front-panel,
(option FHD).
Floppy Disk Drive – 1.44 MB 3.5 in floppy disk
drive: front-panel, or (B models only) rear-panel
(option FHD).
CD-R/W Drive – Rear-panel CD-R/W drive with CD
creation software application.
Mouse – Thumb wheel model included, USB
interface.
Keyboard – Order 118-9402-00 for small
keyboard (fits in pouch); PS-2 interface. Order
119-6633-00 for full-size keyboard; USB
interface and hub.
Input/Output Ports
Probe Compensator Output – Front panel BNC
connector, requires Probe Cal-Deskew Fixture
(included) for probe attachment. Amplitude 200 mV
(TDS7104/TDS7054) or 500mV (B models) ±20%
into a 50 Ω load, frequency 1 kHz ±5%.
Oscilloscopes • www.tektronix.com/tds7000B
USB Port – Allows connection or disconnection of
USB keyboard and/or mouse while oscilloscope
power is on. B models have 2 USB ports.
Keyboard Port – PS-2 compatible.
Analog Signal Output Amplitude – Front-panel
BNC connector, provides a buffered version of the
signal that is attached to the Ch 3 input when Ch 3
is selected as trigger source. 20 mV/div ±20% into
a 1.8 MΩ load, 10 mV/div ±20% into a 50 Ω load.
Mouse Port – PS-2 compatible.
Analog Signal Output Bandwidth, Typical –
TDS7054, TDS7104: 100 MHz into a 50 Ω load.
TDS7154B/TDS7254B/TDS7404B/TDS7704B:
1 GHz into a 50 Ω load.
Windows Video Port – 15 pin D-Sub connector
on the rear panel; connect a second monitor to
use dual-monitor display mode. Video is DDC2B
compliant.
External Time Base Reference In – Rear-panel
BNC connector, time base system can phase-lock
to external 10 MHz reference.
GPIB Port – IEEE 488.2 standard.
Time Base Reference Out – Rear-panel BNC connector, accepts TTL-compatible output of internal
10 MHz reference oscillator.
Auxiliary Output Levels – Front-panel BNC
connector, provides a TTL-compatible, polarity
switchable pulse when the oscilloscope triggers.
Parallel Port – IEEE 1284, DB-25 connector.
8
Audio Ports – Miniature phone jacks for stereo
microphone input and stereo line output.
LAN Port – RJ-45 connector, supports 10Base-T
and 100Base-T.
Serial Port – DB-9 COM1 port.
Oscilloscope Video Port – 15 pin D-Sub connector
on the rear panel, video is IBM XGA compatible on
B models. Connect to show the oscilloscope
display, including live waveforms on an external
monitor or projector. The primary Windows desktop
can also be displayed on an external monitor using
this port.
Power Source
Power – 100 to 240 VRMS ±10%, 50/60 Hz;
115 VRMS ±10%, 400 Hz; CAT II, <300 W (450 VA).
Digital Phosphor Oscilloscopes
TDS7000B Series
Physical Characteristics
Environmental
BENCHTOP CONFIGURATION
Dimensions
mm
Temperature –
Operating: 0 ºC (TDS7104/TDS7054) or +5 ºC
(B models) to +50 ºC, excluding floppy disk and
CD-R/W drives; +10 ºC to +45 ºC, including floppy
disk and CD-ROM drives.
Nonoperating: -22 ºC to +60 ºC.
Height
Width
Depth
277
455
425
in.
10.9
17.9
16.75
Weight
kg
lbs.
Net
Shipping
19.1
37
42.0
85
RACKMOUNT CONFIGURATION
Dimensions
mm
Height
Width
Depth
Weight
Net
Kit
10.5
19.75
19.125
kg
lbs.
18.6
5.6
41
12.25
MECHANICAL
Cooling – Required mm
Clearance
Top
Bottom
Left side
Right side
Front
Rear
in.
277
502
486
0 or >76
0
76
76
0
0
Humidity –
Operating: 20% to 90% relative humidity with a
maximum wet bulb temperature of +29 ºC at or
below +50 ºC, noncondensing. Upper limit derated
to 25% relative humidity at +50 ºC.
Nonoperating: With no diskette in floppy disk drive.
20% to 90% relative humidity with a maximum wet
bulb temperature of +29 ºC at or below +60 ºC,
noncondensing. Upper limit derated to 20% relative
humidity at +60 ºC.
Altitude –
Operating: 10,000 ft. (3,048 m).
Nonoperating: 40,000 ft. (12,190 m).
Random Vibration –
Operating: 0.000125 G2/Hz from 5 to 350 Hz,
–3 dB/octave from 350 to 500 Hz,
0.0000876 G2/Hz at 500 Hz. Overall level of
0.24 GRMS.
Nonoperating: 0.0175 G2/Hz from 5 to 100 Hz,
–3 dB/octave from 100 to 200 Hz, 0.00875 G2/Hz
from 200 to 350 Hz, –3 dB/octave from 350 to
500 Hz, 0.006132 G2/Hz at 500 Hz. Overall level
of 2.28 GRMS.
Electromagnetic Compatibility – 89/336/EEC.
Safety – UL 3111-1, CSA 1010.1, EN61010-1.2,
IEC 61010-1.
in.
0 or >3
0
3
3
0
0
Oscilloscopes • www.tektronix.com/tds7000B
9
Digital Phosphor Oscilloscopes
TDS7000B Series
Options (available on models indicated by “x”)
TDS7054
TDS7104
TDS7154B
TDS7254B
TDS7404B
TDS7704B
Acquisition Memory Options
2M
8 Msamples max,
2 Msamples/ch
x
x
x
x
x
x
3M
16 Msamples max,
4 Msamples/ch
x
x
x
x
x
x
4M
32 Msamples max,
8 Msamples/ch
x
x
x
x
5M
64 Msamples max,
16 Msamples/ch
x
x
x
x
x
x
x
Software Options
DVI
TDSDVI DVI compliance
test solution
DVD
TDSDVD Optical
storage analysis
x
x
x
x
x
x
ET3
TDSET3 Ethernet compliance
test software
x
x
x
x
x
x
JT3
TDSJIT3 Advanced jitter
analysis software
x
x
x
x
x
x
J3E
TDSJIT3 Essentials
x
x
x
x
x
x
J2
TDSDDM2 Disk drive
analysis software
x
x
x
x
x
x
CP2*1
TDSCPM2 ANSI/ITU
Telecom pulse compliance
testing software
x
x
x
x
x
x
USB*2
TDSUSBS USB2.0
Compliance test S/W only
x
x
x
x
x
x
Serial communications
mask testing
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
SM
ST
Serial pattern trigger
PW3
TDSPWR3 Power measurement
and analysis software
RTE
Serial Data Compliance
and Analysis Software
PCE*3
PCI Express™ Compliance
Module for option RTE
x
x
IBA*3
InfiniBand® Compliance
Module for option RTE
x
x
*1 Requires
Option SM.
*2 Requires
Option TDSUSBF (USB Test Fixture).
*3 Requires
Option RTE.
To view instrument upgrades for the TDS7000 and TDS7000B Series, please go to www.tektronix.com/tds7000b_upgrades.
10 Oscilloscopes • www.tektronix.com/tds7000B
Digital Phosphor Oscilloscopes
TDS7000B Series
Ordering Information
Instrument Options
TDS7054
Power Plug Options
500 MHz Digital Phosphor Oscilloscope.
Opt. A0 – North America power.
Opt. A1 – Universal EURO power.
Opt. A2 – United Kingdom power.
Opt. A3 – Australia power.
Opt. A5 – Switzerland power.
Opt. A6 – Japan power.
Opt. A99 – No power cord or AC adapter.
Opt. A10 – China power.
TDS7104
1 GHz Digital Phosphor Oscilloscope.
TDS7104 and TDS7054 include: (4) P6139A
500 MHz, 10x Passive Probes, accessory pouch,
front cover, mouse, probe calibration and deskew
fixture (067-0405-xx), quick reference kit
(020-2404-xx), user manual (071-1035-xx),
TDS7000 Series product software CD-ROM,
TDS7000 Series operating system restoration
CD-ROM, TDS7000 Series optional applications
software CD-ROM, performance verification procedure
PDF file, GPIB programmer’s reference (on product
software CD-ROM), calibration certificate documenting NIST traceability, 2 540-1 compliance and
ISO9000, power cord, one year warranty.
TDS7154B
1.5 GHz Digital Phosphor Oscilloscope.
TDS7254B
2.5 GHz Digital Phosphor Oscilloscope.
TDS7404B
4 GHz Digital Phosphor Oscilloscope.
TDS7704B
7 GHz Digital Phosphor Oscilloscope.
TDS7154B, TDS7254B, TDS7404B, and
TDS7704B Include: Accessory pouch, front cover,
mouse, user manual (071-1226-xx), quick
reference kit (020-2519-xx), TDS7000B Series
product software CD-ROM 063-3632-00
(TDS7154/7254/7404B) 063-3633-00
(TDS7704B), TDS7000B Series operating system
restoration CD-ROM, TDS7000B Series optional
applications software CD-ROM, performance
verification procedure PDF file, GPIB programmer’s
reference (on product software CD-ROM),
calibration certificate documenting NIST traceability,
2 540-1 compliance and ISO9000, power cord,
one year warranty.
For TDS7154B and TDS7254B Only, Also Includes:
(4) TekConnect® to BNC adapters (TCA-BNC),
(2) TekConnect high-impedance buffer amplifiers
(TCA-1MEG), and deskew fixture (067-0405-xx).
For TDS7404B and TDS7704B Only, Also Includes:
(4) TekConnect to SMA adapters (TCA-SMA).
For TDS7404B Only, Also Includes:
Deskew fixture (067-0405-xx).
TDS7704B Only, Also Includes:
Deskew fixture (067-0484-xx).
Please specify disk drive and power cord option
when ordering.
Mounting Options
1K – K4000 Oscilloscope Cart.
1R – Rackmount Kit.
Disk Drive Options
(for TDS7154B,
TDS7254B, TDS7404B,
and TDS7704B only)
Opt. FHD– Front-panel 40 GB removable hard
disk drive, replaces floppy disk drive that goes
on the rear panel.
Service Options
Opt. C3 – Calibration Service 3 Years.
Opt. C5 – Calibration Service 5 Years.
Opt. D1 – Calibration Data Report.
Opt. D3 – Calibration Data Report 3 Years
(with Option C3).
Opt. D5 – Calibration Data Report 5 Years
(with Option C5).
Opt. R3 – Repair Service 3 Years.
Opt. R5 – Repair Service 5 Years.
Adapters
TCA-1MEG – TekConnect® High-impedance
Buffer Amplifier.
TCA-SMA – TekConnect-to-SMA Adapter.
TCA-N – TekConnect-to-N Adapter.
TCA-BNC – TekConnect-to-BNC Adapter.
TCA75 – 4 GHz Precision TekConnect 75 Ω to
50 Ω Adapter with 75 Ω BNC Input Connector.
AFTDS – Telecom Differential Electrical Interface
Adapter (for line rates <8 MB/s; requires
TCA-BNC adapter for TDS7404B).
Cables
GPIB Cable (1 m) – Order 012-0991-01.
GPIB Cable (2 m) – Order 012-0991-00.
RS-232 Cable – Order 012-1298-00.
Centronics Cable – Order 012-1214-00.
Test Fixtures
TDSUSBF – TDSUSB test fixture for use
with Opt. USB.
Software
WSTRO – WaveStarTM waveform capture
and documentation software.
Miscellaneous
Keyboard (USB interface) – Order 119-6633-00.
Service Manual – TDS7054/TDS7104:
Order 071-0711-xx.
TDS7154B/TDS7254B/TDS7404B/TDS7704B:
Order 071-0898-xx.
Transit Case – Order 016-1522-00.
Recommended
Accessories
Probes
P7350 – 5 GHz Differential Probe recommended
for TDS7404B/TDS7704B.
P7350SMA – 5 GHz SMA input Differential Probe
P7260 – 6 GHz Active Probe recommended
for TDS7404B/TDS7704B.
P6158 – 3 GHz, 20x Low C Probe.
P6247 – 1.0 GHz Differential Probe.
P6139A – 500 MHz, 10x Passive Probe.
P6245 – 1.5 GHz Active Probe.
P6248 – 1.7 GHz Differential Probe.
P7240 – 4 GHz Active Probe.
P7330 – 3.5 GHz Differential Probe.
Oscilloscopes • www.tektronix.com/tds7000B
11
Digital Phosphor Oscilloscopes
TDS7000B Series
Contact Tektronix:
ASEAN / Australasia / Pakistan (65) 6356 3900
Austria +41 52 675 3777
Balkan, Israel, South Africa and other
ISE Countries +41 52 675 3777
Belgium 07 81 60166
Brazil & South America 55 (11) 3741-8360
Canada 1 (800) 661-5625
Central East Europe, Ukraine and Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Denmark 80 88 1401
Finland +41 52 675 3777
France & North Africa +33 (0) 1 69 81 81
Germany +49 (221) 94 77 400
Hong Kong (852) 2585-6688
Tektronix (India) Private Limited (91) 80-22275577
Italy +39 (02) 25086 1
Japan 81(3)6714-3010
Luxembourg +44 (0) 1344 392400
Mexico, Central America & Caribbean 52 (55) 56666-333
Middle East, Asia and North Africa +41 52 675 3777
The Netherlands 090 02 02797
Norway 800 16098
People’s Republic of China 86 (10) 6235 1230
Poland +41 52 675 3777
Portugal 80 08 12370
Republic of Korea 82 (2) 528-5299
Russia, CIS & The Baltics 7 095 775 1064
South Africa +27 11 254 8360
Spain (+34) 901 988 054
Sweden 020 08 80371
Switzerland +41 52 675 3777
Taiwan 886 (2) 2722-9622
United Kingdom & Eire +44 (0) 1344 392400
USA 1 (800) 426-2200
USA (Export Sales) 1 (503) 627-1916
For other areas contact Tektronix, Inc. at: 1 (503) 627-7111
Last Updated 1 November, 2004
Our most up-to-date product information is available at:
www.tektronix.com
Product(s) are manufactured
in ISO registered facilities.
Copyright © 2005, Tektronix, Inc. All rights reserved. Tektronix products are covered
by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specification and price change
privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix,
Inc. All other trade names referenced are the service marks, trademarks or registered trademarks of their respective companies.
01/05
12 Oscilloscopes • www.tektronix.com/tds7000B
HB/WOW
55W-13766-9