EXTECH 407736

User's Guide
Digital Sound Level Meter
MODEL 407736
•
1.5 dB accuracy meets ANSI and IEC 651
Type II standards
•
A/C weighting & Fast/Slow response
•
Built-in calibration check
•
0.1dB resolution
Introduction
Congratulations on your purchase of the Extech Model 407736 Digital Sound Level Meter.
This meter is shipped fully tested and calibrated and, with proper use, will provide years of
reliable service.
Meter Description
1.
2.
3.
4.
5.
6.
7.
8.
9.
Microphone
LCD display
POWER OFF & RANGE select switch
RESPONSE & HOLD select switch
A/C weighting and Calibration switch
Calibration adjust screw for 94dB
Reset key (resets max hold reading)
AC/DC analog output 3.5mm phone jack
Battery compartment and Tripod mount are located on the
back of meter
1
2
4
5
3
7
6
Lo = 35 ~ 100dB
HI = 75 ~ 130dB
8
9
Measurement Considerations
1.
Wind blowing across the microphone adds extraneous noise to the measurement.
Use the supplied windscreen to cover the microphone when applicable.
2.
Calibrate the instrument before each use if possible. Especially if the meter has not
been used for a long period of time.
3.
Do not store or operate the instrument in areas of high temperature or humidity.
4.
Keep meter and microphone dry.
5.
Avoid severe vibration when using the meter.
6.
Remove the battery when the meter is to be stored for long periods of time.
2
407736 V2.0 5/08
Meter Operation
Quick Start
1. Power the meter by moving the RANGE switch to the LO or HI position. The meter will
begin displaying sound levels. If the LCD does not display, check that the battery is
good.
2. Place the meter on a tripod or hold the meter in hand facing the microphone toward the
source of the sound to be measured. Note that the tripod mount is on the back of meter.
3. View the measurement on the meter’s LCD. An indication of ‘OVER’ means that the
measurement is out of range. Select another measurement range in the event of an
over-range condition.
‘A’ and ‘C’ Frequency Weighting
Use the FUNCT (function) switch to select ‘A’ or ‘C’ frequency weighting. Use ‘A’ weighting
to have the meter simulate the response of the human ear (the human ear boosts and cuts
sound levels at specific points over the frequency spectrum). ‘A’ weighting is used for
environmental measurements, OSHA regulatory testing, law enforcement, and workplace
design. Select ‘C’ weighting for flat response measurements (no boosting or dampening
across the frequency spectrum). ‘C’ weighting is suitable for the operational maintenance
and analysis of machinery, motors, pumps, engines, etc.
FAST/SLOW Response Time
Use the RESPONSE switch to select FAST (125 ms) or SLOW (1 second) response time.
The application at hand (and any directives or standards) will dictate which response to
select. For example, most hearing conservation or OSHA related testing is done using
SLOW mode and ‘A’ weighting.
MAX HOLD
In this mode, the meter only updates the LCD when a higher reading than the one
presently on the display is detected. Select MAX HOLD using the RESPONSE switch. The
LCD shows ‘MAX HOLD’ in this mode. Press RESET to reset the MAX HOLD reading.
Analog Outputs
The meter includes an AC and a DC analog output for use with chart recorders,
dataloggers, etc. The AC output is 0.65V rms full scale and the DC output is 10mV per dB.
The 3.5mm output jack is located on the right side of the instrument. Use a stereo miniplug as shown in the diagram below to connect the meter to a datalogger or other
recorder.
1.
2.
3.
1
Sleeve: ground
Ring: DC out
Tip: AC out
2 3
Calibration Check
This meter provides a built-in calibration adjustment. The calibration adjustment
potentiometer is located on the front panel beneath the FUNCT switch. Put the FUNCT
switch to the CAL 94dB position and adjust the potentiometer for a display of exactly
94dB. For calibration with an external sound level calibrator, refer to the instructions
accompanying the calibrator.
3
407736 V2.0 5/08
Battery Replacement
When the low battery message appears on the LCD replace the 9V battery as soon as
possible. The battery compartment cover resides at the bottom, rear of the meter. Slide
the battery compartment cover off, change the battery, and replace the compartment
cover.
Typical ‘A’ weighted Sound Pressure Levels
dBA
140
50HP Siren (100')
130
Jet takeoff (200')
120
110
Riveting machine
100
Subway (20')
90
Pneumatic drill (50')
80
Vacuum cleaner (10')
70
Large store
60
Small office
50
Night residential area
40
Whisper (5'0)
30
20
Chain saw
Boiler room
Freight train (100')
Speech (1')
Large Office
Residence
Sound studio
10
Threshold of hearing
0
Warranty
EXTECH INSTRUMENTS CORPORATION (A FLIR COMPANY) warrants this instrument to
be free of defects in parts and workmanship for one year from date of shipment (a six
month limited warranty applies to sensors and cables). If it should become necessary to
return the instrument for service during or beyond the warranty period, contact the
Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our
website www.extech.com for contact information. A Return Authorization (RA) number
must be issued before any product is returned to Extech. The sender is responsible for
shipping charges, freight, insurance and proper packaging to prevent damage in transit.
This warranty does not apply to defects resulting from action of the user such as misuse,
improper wiring, operation outside of specification, improper maintenance or repair, or
unauthorized modification. Extech specifically disclaims any implied warranties or
merchantability or fitness for a specific purpose and will not be liable for any direct, indirect,
incidental or consequential damages. Extech's total liability is limited to repair or
replacement of the product. The warranty set forth above is inclusive and no other
warranty, whether written or oral, is expressed or implied.
4
407736 V2.0 5/08
Specifications
Display
Display update rate
Microphone
Measurement Bandwidth
Dynamic range
Measurement Range
Frequency weighting
Applicable standards
Accuracy / Resolution
Maximum Hold decay time
Response time
Built-in calibration check
AC Analog output
3-1/2 digit (2000 count) LCD
0.5 seconds
0.5” Electret condensor
31.5Hz to 8KHz
55dB
35 to 130dB (Low: 35 to 100; High: 65 to 130dB)
‘A’ and ‘C’ (selectable)
IEC-651 & ANSI S1.4 Type 2
± 1.5dB / 0.1dB
<1dB / 3min
Fast: 125 milliseconds / Slow: 1 second
1KHz internal sine wave @ 94dB
DC Analog output
10mVDC / dB; 100Ω output impedance (approx).
9V Battery
50 hours (typical); Low battery indicator alerts user
o
o
32 to 104 F (0 to 40 C)
10 to 90% RH
9.45 x 2.68 x 1” (240 x 68 x 25mm) / 6.75oz (210g)
Power
Battery life
Operating temperature
Operating humidity
Dimensions/weight
0.65VAC rms (full scale); 600Ω output impedance
Support line (781) 890-7440
Technical Support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Road, Waltham, MA 02451
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also
provides NIST certification for most products. Call the Customer Care Department for
information on calibration services available for this product. Extech recommends that
annual calibrations be performed to verify meter performance and accuracy.