EXTECH 407766

User’s Manual
Model 407744 and 407766
Sound Level Calibrators
Introduction
Congratulations on your purchase of Extech’s Sound Level Calibrator. Extech Calibrators
accommodate Sound Level Meters with 0.5” and 1” microphones. The Calibrator’s die-cast
aluminum housing is built for ruggedness and durability. The 407744 provides a 94dB
calibration output signal while the 407766 provides both 94dB and 114dB outputs. This
professional calibrator, with proper care, will provide years of safe reliable service.
Specifications
Output signals
407744: 94dB
407766: 94dB and 114dB
Output signal accuracy
407744: ± 0.5dB;
407766: ± 0.5dB (94dB), ±0.8dB (114dB)
Output signal frequency
1KHz sine wave
Accuracy of output signal frequency
± 5%
Compatible microphones sizes
0.5” and 1” microphones
Total Harmonic Distortion (THD)
<2%@94dB, <5%@114dB (407766)
Operating temperature
32 to 122 F (0 to 50 C)
o
o
Power supply
Two (2) 9V batteries
Power consumption
Approx. 10mA DC
Battery check
Status LED
Dimensions
2.2” diameter x 5.7” length (50 x 145mm)
Weight
0.75 lbs. (340g)
Page 2
Models 407744 & 407766 - Version 5.0 - May 2005
Description
Note that only the 407766 is pictured.
1. ½” Microphone adaptor *
4
2. 407766: OFF-94dB-114dB switch
407744: OFF-BATTERY TEST-94dB switch
3. 407766: POWER-ON / LOW-BATTERY status LED
407744: BATTERY OK LED
4. Battery compartment cap
* To remove the ½” adaptor, pull straight down
as shown in diagram below.
3
2
1
Operation
Theory of Operation
The 407744 sound level calibrator generates a 94dB sine wave at 1 KHz. The 407766
generates a 94dB and a 114dB sine wave at 1kHz. When a microphone from a sound level
meter is inserted into the calibrator, the sound level meter picks up the signal and displays
the signal in dB (decibel) units. The sound level meter should be adjusted (if necessary) to
match the calibrator’s output signal as closely as possible (see diagram).
Procedure
1. Insert the microphone into the calibrator’s microphone cavity as shown; ensure that it is
fully inserted. Use the ½” microphone adaptor (supplied attached to
the calibrator) for ½” microphones. Remove the adaptor (remove as
shown in the Description section) for 1” microphones.
2. On the 407744, slide the function switch to the ‘I’ (Power On) position.
As soon as the unit is powered, the calibration tone is generated. On
the 407766, slide the function switch to the 94dB or the 114dB
position as desired. The calibrator generates its signal as soon as the
switch is set.
3. Read the sound level meter’s display. The sound level meter should
read 94dB or 114dB depending on the calibrator’s setting. If the sound
level meter does not match the calibrator’s signal (within
specification), adjust it as explained in the Operation section (typically
a front or side-panel pot adjustment). Refer to diagram at right.
4. Slide function switch to the OFF position, and remove the microphone.
Page 3
Models 407744 & 407766 - Version 5.0 - May 2005
Battery Check
On the 407744, check the condition of the battery by moving the function switch to the
‘Battery test’ position. The Battery status LED should illuminate if the battery is OK. If it
does not light, replace the 9V battery. On the 407766, the POWER-ON / LOW-BATTERY
LED glows brightly when the battery is fresh and dimly when the battery is weak.
Battery Replacement
Lift off battery compartment cap (shown in the Description section) to access the two (2)
9V batteries. Remove the batteries and replace with new alkaline 9V batteries. Replace the
cap.
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects
in parts and workmanship for one year from date of shipment (a six month limited warranty
applies to sensors and cables). If it should become necessary to return the instrument for
service during or beyond the warranty period, contact the Customer Service Department at
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact
information. A Return Authorization (RA) number must be issued before any product is
returned to Extech. The sender is responsible for shipping charges, freight, insurance and
proper packaging to prevent damage in transit. This warranty does not apply to defects
resulting from action of the user such as misuse, improper wiring, operation outside of
specification, improper maintenance or repair, or unauthorized modification. Extech
specifically disclaims any implied warranties or merchantability or fitness for a specific
purpose and will not be liable for any direct, indirect, incidental or consequential damages.
Extech's total liability is limited to repair or replacement of the product. The warranty set
forth above is inclusive and no other warranty, whether written or oral, is expressed or
implied.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also
provides NIST certification for most products. Call the Customer Service Department for
information on calibration services available for this product. Extech recommends that
annual calibrations be performed to verify meter performance and accuracy.
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Copyright © 2005 Extech Instruments Corporation
All rights reserved including the right of reproduction in whole or in part in any form.
Page 4
Models 407744 & 407766 - Version 5.0 - May 2005