HP E5071C-245 Configurable multiport test set Datasheet

[ Obsolete product ]
Agilent has a new name
Keysight Technologies.
Keysight Technologies Inc. is the world's leading electronic measurement
company, transforming today's measurement experience through innovations in
wireless, modular, and software solutions. With its HP and Agilent legacy,
Keysight delivers solutions in wireless communications, aerospace and defense
and semiconductor markets with world-class platforms, software and
consistent measurement science.
Alldatasheet.com
Agilent
E5071C ENA Network Analyzer
• 9 kHz to 4.5/6.5/8.5 GHz
• 100 kHz to 4.5/6.5/8.5 GHz (with bias tees)
• 300 kHz to 14/20 GHz (with bias tees)
E5091A Multiport Test Set
E5092A Configurable Multiport Test Set
Data Sheet
Table of Contents
Options ........................................................................................................................ 3
Definitions ................................................................................................................. 4
Boundary Conditions ............................................................................................. 4
Corrected System Performance........................................................................ 5
System dynamic range ..................................................................................................... 5
Corrected system performance with calibration kit.................................................... 7
Uncorrected System Performance ................................................................13
Test Port Output (Source) .................................................................................14
Test port output frequency ............................................................................................ 14
Test port output power .................................................................................................. 15
Test port output signal purity ........................................................................................ 17
Test Port Input ........................................................................................................18
Test port input levels ...................................................................................................... 18
Test port input (trace noise) .......................................................................................... 20
Test port input (stability) ............................................................................................... 22
Test port input (dynamic accuracy) .............................................................................. 23
Test port input (group delay) ......................................................................................... 25
General Information.............................................................................................27
System bandwidths ......................................................................................................... 27
Front panel information .................................................................................................. 27
Rear panel information ................................................................................................... 28
LXI compliance................................................................................................................. 30
EMC, safety and environment ....................................................................................... 31
Analyzer environmental specifications and dimensions .......................................... 32
Measurement Throughput Summary ............................................................36
Typical cycle time for measurement completion (ms).............................................. 36
Cycle time (ms) vs. number of points.......................................................................... 48
Cycle time (ms) vs. IF bandwidth ................................................................................. 49
Data transfer time (ms) .................................................................................................. 50
E5091A Multiport Test Set.................................................................................51
Test set input/output performance .............................................................................. 51
Option E5091A-009 port performance ......................................................................... 51
Option E5091A-016 port performance ......................................................................... 52
Front panel information .................................................................................................. 52
Rear panel information ................................................................................................... 52
Test set dimensions and block diagram ...................................................................... 53
E5092A Configurable Multiport Test Set .....................................................56
Test set input/output performance .............................................................................. 56
Option E5092A-020 port performance ......................................................................... 56
Control line ....................................................................................................................... 58
DC source.......................................................................................................................... 58
Operating storage environment .................................................................................... 58
Non-operating storage environment ............................................................................ 58
Front panel information .................................................................................................. 59
Rear panel information ................................................................................................... 59
Test set dimensions and block diagram ...................................................................... 59
Corrected System Performance for 75 Ω Measurements with
11852B 50 to 75 Ω Minimum-Loss Pads ................................................ 63
2
Options
This document provides technical specifications for the E5071C ENA network
analyzer and the E5091A multiport test set.
E5071C-230
E5071C-235
E5071C-240
E5071C-245
E5071C-260
E5071C-265
E5071C-280
E5071C-285
E5071C-430
E5071C-435
E5071C-440
E5071C-445
E5071C-460
E5071C-465
E5071C-480
E5071C-485
E5071C-2D5
E5071C-4D5
E5071C-2K5
E5071C-4K5
E5091A
E5092A
2-port test set, 9 kHz to 3 GHz without bias tees (Discontinued)
2-port test set, 100 kHz to 3 GHz with bias tees (Discontinued)
2-port test set, 9 kHz to 4.5 GHz without bias tees
2-port test set, 100 kHz to 4.5 GHz with bias tees
2-port test set, 9 kHz to 6.5 GHz without bias tees
2-port test set, 100 kHz to 6.5 GHz with bias tees
2-port test set, 9 kHz to 8.5 GHz without bias tees
2-port test set, 100 kHz to 8.5 GHz with bias tees
4-port test set, 9 kHz to 3 GHz without bias tees (Discontinued)
4-port test set, 100 kHz to 3 GHz with bias tees (Discontinued)
4-port test set, 9 kHz to 4.5 GHz without bias tees
4-port test set, 100 kHz to 4.5 GHz with bias tees
4-port test set, 9 kHz to 6.5 GHz without bias tees
4-port test set, 100 kHz to 6.5 GHz with bias tees
4-port test set, 9 kHz to 8.5 GHz without bias tees
4-port test set, 100 kHz to 8.5 GHz with bias tees
2-port test set, 300 kHz to 14 GHz with bias tees
4-port test set, 300 kHz to 14 GHz with bias tees
2-port test set, 300 kHz to 20 GHz with bias tees
4-port test set, 300 kHz to 20 GHz with bias tees
Multiport test set
Configurable multiport test set
Calibration kits and ECal modules
This E5071C data sheet also provides technical specifications for the
following calibration kits and ECal modules. For models not listed in
this data sheet, please download the free Uncertainty Calculator from
www.agilent.com/find/na_calculator to generate the curves for your
calibration kit and enable ENA setup.
85032F
85033E
85052D
85092C
85093C
N4691B
Calibration kit
Calibration kit
Calibration kit
Electronic calibration (ECal) module
Electronic calibration (ECal) module
Electronic calibration (ECal) module
3
Definitions
Specification (spec.):
Warranted performance. All specifications apply at 23 ºC (± 5 ºC), unless
otherwise stated, and 90 minutes after the instrument has been turned on.
Specifications include guard bands to account for the expected statistical
performance distribution, measurement uncertainties, and changes in
performance due to environmental conditions.
Supplemental information is intended to provide information that is helpful for
using the instrument but that is not guaranteed by the product warranty.
Typical (typ.):
Describes performance that will be met by a minimum of 80% of all products.
It is not guaranteed by the product warranty.
Supplemental performance data (SPD):
Supplemental performance data represents the value of a parameter that is
most likely to occur; the expected mean or average. It is not guaranteed by the
product warranty.
General characteristics:
A general, descriptive term that does not imply a level of performance.
Boundary Conditions
In this data sheet, boundary conditions are given for the specifications. For
example, system dynamic range is 98 dB with the following boundary conditions.
Option: 485
Frequency: 10 MHz
IF bandwidth: 3 kHz
If the same boundary conditions fall under more than one category in a table,
apply the best value.
4
Corrected System Performance
The specifications in this section apply to measurements made with the
Agilent E5071C network analyzer under the following conditions:
• No averaging applied to data
• Environmental temperature of 23 °C (±5 °C) with less than 1 °C deviation
from the calibration temperature
• Response and isolation calibration performed
System dynamic range
Table 1. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Description
Specification
SPD
System dynamic range1, 2
9 kHz to 300 kHz
300 kHz to 10 MHz
72 dB
IF bandwidth = 3 kHz
82 dB
10 MHz to 6 GHz
98 dB
6 GHz to 8.5 GHz
92 dB
9 kHz to 300 kHz
97 dB
115 dB
300 kHz to 10 MHz
107 dB
115 dB
123 dB
130 dB
6 GHz to 7 GHz
117 dB
128 dB
7 GHz to 8 GHz
117 dB
126 dB
8 GHz to 8.5 GHz
117 dB
124 dB
10 MHz to 6 GHz
IF bandwidth = 10 Hz
System dynamic range [dB]
160
150
140
130
120
110
E5071C specification
100
90
80
1.E+05
1.E+09
2.E+09
3.E+09
4.E+09
5.E+09
6.E+09
7.E+09
8.E+09
Test frequency [Hz]
Figure 1. System dynamic range (specification and actual measurement data example,
IF bandwidth 10 Hz)
1. The test port dynamic range is calculated as the difference between the test port rms noise floor
and the source maximum output power. The effective dynamic range must take measurement
uncertainty and interfering signals into account.
2. The specification might not be met at 5 MHz or 50 MHz.
5
System dynamic range (continued)
Table 2. Option 2D5/2K5/4D5/4K5
Description
Specification
SPD
System dynamic range1, 2
300 kHz to 1 MHz
70 dB
1 MHz to 10 MHz
82 dB
10 MHz to 100 MHz
95 dB
100 MHz to 6 GHz
IF bandwidth = 3 kHz
98 dB
6 GHz to 8.5 GHz
92 dB
8.5 GHz to 10.5 GHz
80 dB
10.5 GHz to 15 GHz
75 dB
15 GHz to 20 GHz
71 dB
300 kHz to 1 MHz
95 dB
105 dB
1 MHz to 10 MHz
107 dB
115 dB
10 MHz to 100 MHz
120 dB
129 dB
123 dB
130 dB
6 GHz to 8 GHz
117 dB
129 dB
8 GHz to 8.5 GHz
117 dB
127 dB
8.5 GHz to 10.5 GHz
105 dB
115 dB
10.5 GHz to 15 GHz
100 dB
111 dB
15 GHz to 20 GHz
96 dB
105 dB
100 MHz to 6 GHz
IF bandwidth = 10 Hz
160
System dynamic range[dB]
150
140
130
120
110
100
E5071C-2D5/2K5/4D5/4K5 specification
90
80
0.0E+00
5.0E+09
1.0E+10
1.5E+10
2.0E+10
Frequency [Hz]
Figure 2. System dynamic range (specification and actual measurement data example,
IF bandwidth 10 Hz)
1. The test port dynamic range is calculated as the difference between the test port’s rms noise
floor and the source’s maximum output power. Effective dynamic range must take measurement
uncertainty and interfering signals into account.
2. The specification might not be met at 5 MHz or 50 MHz.
6
Corrected system performance with calibration kit
Option 230/235/240/245/260/265/280/285/430/435/440/445/460/465/480/485
Table 3. Corrected system performance with type-N device connectors,
85032F calibration kit
Network analyzer:
Calibration kit:
Calibration:
E5071C
85032F (Type-N, 50 Ω)
full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed
Specification (dB)
Description
9 kHz to
300 kHz
300 kHz to
10 MHz
10 MHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
8.5 GHz
Directivity
49
49
46
40
38
Source match
41
41
40
36
35
Load match
49
49
46
40
37
Reflection tracking
±0.011
±0.011
±0.021
±0.032
±0.054
Transmission tracking
±0.027
±0.015
±0.018
±0.056
±0.088
Transmission uncertainty (specification)
Magnitude
9 KHz to 300 KHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
1
0.1
0.01
10
Phase
100
Uncertainty (degrees)
Uncertainty (dB)
10
9 KHz to 300 KHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
10
1
S 1 1 = S 2 2 = 0 ; C a l p o w e r = -1 0 d B m ; M e a s p o w e r = -1 0 d B m
0
–10
–20 –30
–40
–50
–60
S 1 1 = S 2 2 = 0 ; C a l p o w e r = -1 0 d B m ; M e a s p o w e r = -1 0 d B m
0.1
–70 –80 –90
10
Transmission coefficient (dB)
0
-10
-20
-30
-40
-50
-60
-70
-80
-90
Transmission coefficient (dB)
Reflection uncertainty (specification)
Magnitude
10
9 KHz to 300 KHz, 300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
0.04
Uncertainty (degrees)
Uncertainty (linear)
0.05
0.03
0.02
0.01
Phase
9 KHz to 300 KHz, 300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
8
6
4
2
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
0
0
0
0.2
0.4
0.6
0.8
1
0
0.2
Reflection coefficient (linear)
0.4
0.6
Reflection coefficient (linear)
7
0.8
1
Table 4. Corrected system performance with type-N device connectors,
85092C electronic calibration (ECal) module
Network analyzer:
Calibration module:
E5071C
85092C (Type-N, 50 Ω) Electronic calibration (ECal)
module
full 2-port
Calibration:
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed
Specification (dB)
Description
300 kHz to
10 MHz
10 MHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
8.5 GHz
Directivity
45
52
52
47
Source match
36
44
41
36
Load match
41
47
44
39
Reflection tracking
±0.100
±0.040
±0.060
±0.070
Transmission tracking
±0.055
±0.039
±0.068
±0.136
Transmission uncertainty (specification)
Magnitude
Phase
10
100
3 0 0 K H z to 1 0 M H z
3 0 0 kH z to 1 0 M H z
1 0 M H z to 3 G H z
1 0 M H z to 3 G H z
Uncertainty (degrees)
3 G H z to 6 G H z
Uncertainty (dB)
6 G H z to 8 .5 G H z
1
0 .1
3 G H z to 6 G H z
6 G H z to 8 .5 G H z
10
1
S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
0 .1
0 .0 1
10
0
-1 0
-2 0
-3 0
-4 0
-5 0
-6 0
-7 0
-8 0
10
-9 0
0
-1 0
-2 0
-3 0
-4 0
-5 0
-6 0
-7 0
-8 0
-9 0
Transmission coefficient (dB)
Transmission coefficient (dB)
Reflection uncertainty (specification)
Magnitude
Phase
0 .0 5
10
3 0 0 kH z to 1 0 M H z
3 0 0 kH z to 1 0 M H z
1 0 M H z to 3 G H z
3 G H z to 6 G H z
6 G H z to 8 .5 G H z
1 0 M H z to 3 G H z
8
3 G H z to 6 G H z
6 G H z to 8 .5 G H z
Uncertainty (linear)
Uncertainty (linear)
0 .0 4
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
0 .0 3
0 .0 2
0 .0 1
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
6
4
2
0
0
0
0 .2
0 .4
0 .6
0 .8
1
0
0 .2
Reflection coefficient (linear)
0 .4
0 .6
Reflection coefficient (linear)
8
0 .8
1
Table 5. Corrected system performance with 3.5 mm device connector type,
85033E calibration kit
Network analyzer:
Calibration kit:
Calibration:
E5071C
85033E (3.5 mm, 50 Ω)
full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed
Specification (dB)
9 kHz to
300 kHz
Description
300 kHz to
10 MHz
10 MHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
8.5 GHz
Directivity
46
46
44
38
38
Source match
43
43
40
37
36
Load match
46
46
44
38
38
Reflection tracking
±0.006
±0.006
±0.007
±0.009
±0.010
Transmission tracking
±0.026
±0.015
±0.020
±0.058
±0.079
Transmission uncertainty (specification)
Magnitude
9 KHz to 300 KHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
1
0.1
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
0
-10
-20
-30
-40
-50
-60
-70
-80
9 KHz to 300 KHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
10
1
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
0.1
0.01
10
Phase
100
Uncertainty (degrees)
Uncertainty (dB)
10
10
-90
Transmission coefficient (dB)
0
-10
-20
-30
-40
-50
-60
-70
-80
-90
Transmission coefficient (dB)
Reflection uncertainty (specification)
Magnitude
10
Phase
9 KHz to 300 KHz, 300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
0.04
Uncertainty (degrees)
Uncertainty (linear)
0.05
0.03
0.02
0.01
0
0.2
0.4
0.6
Reflection coefficient (linear)
0.8
8
6
4
2
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
S21 = S12 = 0; Cal power = -10 dBm; Meas power = -10 dBm
0
9 KHz to 300 KHz, 300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
0
1
0
0.2
0.4
0.6
Reflection coefficient (linear)
9
0.8
1
Table 6. Corrected system performance with 3.5 mm device connector type,
85093C electronic calibration (ECal) module
Network analyzer:
Calibration module:
E5071C
85093C (3.5 mm, 50 Ω) electronic calibration (ECal)
module
full 2-port
Calibration:
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed
Specification (dB)
Description
300 kHz to
10 MHz
10 MHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
8.5 GHz
Directivity
45
52
51
47
Source match
36
44
39
34
Load match
41
47
44
40
Reflection tracking
±0.100
±0.040
±0.050
±0.070
Transmission tracking
±0.055
±0.049
±0.068
±0.117
Transmission uncertainty (specification)
Phase
Magnitude
10
100
Uncertainty (degrees)
Uncertainty (dB)
3 0 0 K Hz to 1 0 MHz
1 0 MHz to 3 G Hz
3 G Hz to 6 G Hz
6 G Hz to 8 .5 G Hz
1
0 .1
3 0 0 kH z to 1 0 M H z
1 0 M H z to 3 G H z
3 G H z to 6 G H z
6 G H z to 8 .5 G H z
10
1
S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
0 .1
0 .0 1
10
0
-1 0
-2 0
-3 0
-4 0
-5 0
-6 0
-7 0
-8 0
10
-9 0
0
-1 0
-2 0
-3 0
-4 0
-5 0
-6 0
-7 0
-8 0
-9 0
Transmission coefficient (dB)
Transmission coefficient (dB)
Reflection uncertainty (specification)
Phase
Magnitude
10
0 .0 5
3 0 0 kH z to 1 0 M H z
1 0 M H z to 3 G H z
3 G H z to 6 G H z
6 G H z to 8 .5 G H z
3 0 0 kH z to 1 0 M H z
1 0 M H z to 3 G H z
Uncertainty (degrees)
Uncertainty (linear)
8
3 G H z to 6 G H z
0 .0 4
6 G H z to 8 .5 G H z
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
0 .0 3
0 .0 2
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
6
4
2
0 .0 1
0
0
0
0 .2
0 .4
0 .6
0 .8
1
0
0 .2
0 .4
0 .6
Reflection coefficient (linear)
Reflection coefficient (linear)
10
0 .8
1
Option 2D5/2K5/4D5/4K5
Table 7. Corrected system performance with 3.5 mm device connectors,
85052D calibration kit
Network analyzer:
Calibration kit:
Calibration:
E5071C
85052D (3.5 mm, 50 Ω)
full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed
Specification (dB)
Description
300 kHz to
500 MHz
500 MHz to
2 GHz
2 GHz to
6 GHz
6 GHz to
20 GHz
Directivity
42
42
38
36
Source match
37
37
31
28
Load match
42
42
38
36
Reflection tracking
±0.003
±0.003
±0.004
±0.008
Transmission tracking
±0.068
±0.034
±0.100
±0.208
Transmission uncertainty (specification)
Phase
Magnitude
100
3 0 0 K H z to 5 0 0 M H z
3 0 0 K H z to 5 0 0 M H z
5 0 0 M H z to 2 G H z
5 0 0 M H z to 2 G H z
Uncertainty (Degrees)
Uncertainty (dB)
10
2 G H z to 6 G H z
6 G H z to 2 0 G H z
1
0 .1
2 G H z to 6 G H z
6 G H z to 2 0 G H z
10
1
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
S11 = S22 = 0; Cal power = -10 dBm; Meas power = -10 dBm
0 .1
0 .0 1
10
0
-1 0
-2 0
-3 0
-4 0
-5 0
-6 0
-7 0
-8 0
10
-9 0
0
-1 0
-2 0
-3 0
-4 0
-5 0
-6 0
-7 0
-8 0
-9 0
T rans mis s ion C oefficient (dB )
Transmission Coefficient (dB)
Reflection uncertainty (specification)
Phase
Magnitude
10
0 .0 5
3 0 0 K H z to 5 0 0 M H z, 5 0 0 M H z to 2 G H z
3 0 0 K H z to 5 0 0 M H z, 5 0 0 M H z to 2 G H z
2 G H z to 6 G H z
2 G H z to 6 G H z
8
Uncertainty (degrees)
Uncertainty (linear)
0 .0 4
6 G H z to 2 0 G H z
0 .0 3
0 .0 2
6 G H z to 2 0 G H z
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
6
4
2
0 .0 1
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
0
0
0
0 .2
0 .4
0 .6
0 .8
1
0
0 .2
0 .4
0 .6
Reflection coefficient (linear)
Reflection coefficient (linear)
11
0 .8
1
Table 8. Corrected system performance with 3.5 mm device connectors,
N4691B electronic calibration (ECal) module
Network analyzer:
Calibration module:
E5071C
N4691B (3.5 mm, 50 Ω) electronic calibration (ECal)
module
full 2-port
Calibration:
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
(±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed
Specification (dB)
Description
10 MHz to
500 MHz
500 MHz to
2 GHz
2 GHz to
6 GHz
6 GHz to
20 GHz
Directivity
46
56
54
48
Source match
41
47
45
44
Load match
38
41
39
36
Reflection tracking
±0.050
±0.020
±0.030
±0.040
Transmission tracking
±0.087
±0.086
±0.097
±0.130
Transmission uncertainty (specification)
Phase
Magnitude
100
10
1 0 M H z to 5 0 0 M H z
1 0 M H z to 5 0 0 M H z
5 0 0 M H z to 2 G H z
Uncertainty (Degrees)
5 0 0 M H z to 2 G H z
Uncertainty (dB)
2 G H z to 6 G H z
6 G H z to 2 0 G H z
1
0 .1
S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10dBm
2 G H z to 6 G H z
6 G H z to 2 0 G H z
10
1
S 11 = S 22 = 0 ; Cal power = -10 dBm; Meas power = -10dBm
0 .1
0 .0 1
10
0
-1 0
-2 0
-3 0
-4 0
-5 0
-6 0
-7 0
-8 0
10
-9 0
0
-1 0
-2 0
-3 0
-4 0
-5 0
-6 0
-7 0
-8 0
-9 0
Transmission Coefficient (dB)
Transmission Coefficient (dB)
Reflection uncertainty (specification)
Phase
Magnitude
10
0 .0 5
1 0 M H z to 5 0 0 M H z
1 0 M H z to 5 0 0 M H z
5 0 0 M H z to 2 G H z
5 0 0 M H z to 2 G H z
2 G H z to 6 G H z
0 .0 4
Uncertainty (degrees)
6 G H z to 2 0 G H z
Uncertainty (linear)
2 G H z to 6 G H z
8
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
0 .0 3
0 .0 2
6 G H z to 2 0 G H z
S 2 1 = S 1 2 = 0 ; Cal power = -10 dBm; Meas power = -10 dBm
6
4
2
0 .0 1
0
0
0
0 .2
0 .4
0 .6
0 .8
1
0
0 .2
0 .4
0 .6
Reflection coefficient (linear)
Reflection coefficient (linear)
12
0 .8
1
Uncorrected System Performance
Table 9. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
User correction: OFF, system correction: ON
Specification (dB)
Description
9 kHz to
300 kHz
300 kHz to
3 GHz
3 GHz to
6 GHz
6 GHz
to 8.5 GHz
Directivity
20 dB
25 dB
20 dB
15 dB
Source match
20 dB
25 dB
20 dB
15 dB
Load match
12 dB
17 dB
12 dB
10 dB
Transmission tracking
±1.5 dB
±1.0 dB
±1.0 dB
±1.0 dB
Reflection tracking
±1.5 dB
±1.0 dB
±1.0 dB
±1.0 dB
Table 10. Option 2D5/2K5/4D5/4K5
User correction: OFF, system correction: ON
Specification (dB)
Description
300 kHz 1 MHz to 1 GHz to 3 GHz to 6 GHz to
to 1 MHz 1 GHz
3 GHz
6 GHz
8.5 GHz
8.5 GHz to 11 GHz to
11 GHz
20 GHz
Directivity
20 dB
25 dB
25 dB
20 dB
15 dB
15 dB
15 dB
Source match 20 dB
25 dB
25 dB
20 dB
15 dB
15 dB
15 dB
Load match
17 dB
15 dB
11 dB
9 dB
8 dB
7 dB
Transmission ±1.0 dB
tracking
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
Reflection
tracking
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
±1.0 dB
9 dB
±1.0 dB
13
Test Port Output (Source)
Test port output frequency
Table 11. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485/2D5/2K5/4D5/4K5
Description
Specification
Frequency range
Option 230/430
Option 240/440
Option 260/460
Option 280/480
Option 235/435
Option 245/445
Option 265/465
Option 285/485
Option 2D5/4D5
Option 2K5/4K5
9 kHz to 3 GHz
9 kHz to 4.5 GHz
9 kHz to 6.5 GHz
9 kHz to 8.5 GHz
100 kHz to 3 GHz
100 kHz to 4.5 GHz
100 kHz to 6.5 GHz
100 kHz to 8.5 GHz
300 kHz to 14 GHz
300 kHz to 20 GHz
Resolution
1 Hz
Source stability
standard
Option 1E5
CW accuracy
standard
Option 1E5
Typical
±5 ppm (5 to 40 ºC)
±0.05 ppm (5 to 40 ºC),
±0.5 ppm/year
±5 ppm
±1 ppm
14
Test port output power1
Table 12. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Description
Specification
Nominal power
(preset power)
0 dBm
Level accuracy2, 6
(stepped sweep mode)
±0.650 dB (at 0 dBm,
50 MHz absolute)
±1.0 dB
Level accuracy2
(swept sweep mode)
Level linearity3, 5, 6
(stepped sweep mode)
9 kHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
Typical
±2.5 dB
±0.75 dB (–20 to 10 dBm)
±0.75 dB (–20 to 9 dBm )
±0.75 dB (–20 to 8 dBm )
±0.75 dB (–20 to 7 dBm )
Level linearity5
(swept sweep mode)4
9 kHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
±1.5 dB (at –20 to 10 dBm)
±1.5 dB (at –20 to 9 dBm)
±1.5 dB (at –20 to 8 dBm)
±1.5 dB (at –20 to 7 dBm)
Range5, 6
9 kHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
–55 to 10 dBm
–55 to 9 dBm
–55 to 8 dBm
–55 to 7 dBm
Sweep range5, 6
9 kHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
–55 to 10 dBm
–55 to 9 dBm
–55 to 8 dBm
–55 to 7 dBm
Level resolution
0.05 dB
Description
Specification
SPD
power5, 6
Max leveled
(Option 230/235/240/245/260/
265/280/285)
9 kHz to 8.5 GHz
10 dBm
(Option 430/435/440/445/460/
465/480/485)
9 kHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
10 dBm
9 dBm
8 dBm
1.
2.
3.
4.
5.
6.
Source output performance on port 1 only. Other port output performance is typical.
Level accuracy is taken at 0 dBm, relative to 50 MHz reference unless otherwise stated.
Level linearity given is relative to 0 dBm.
The specification might not be met at 5 MHz or 50 MHz.
The level accuracy specification needs to be taken into account for test port output power level.
Power calibration using an external power meter improves level accuracy of the test port output power.
Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration.
15
Test port output power7 (continued)
Table 13. Option 2D5/2K5/4D5/4K5
Description
Specification
Nominal power (preset power)
–5 dBm
Level accuracy6
(stepped sweep mode)1
300 kHz to 1 MHz
1 MHz to 5 MHz
5 MHz to 8.5 GHz
8.5 GHz to 20 GHz
±0.650 dB
+2.0 dB, –6.0 dB
±2.0 dB
±1.0 dB
±2.5 dB
Level accuracy
(swept sweep mode)2
300 kHz to 1 GHz
1 GHz to 8.5 GHz
8.5 GHz to 20 GHz
Level linearity5, 6
(stepped sweep mode)3
300 kHz to 1 MHz
1 MHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10.5 GHz
10.5 GHz to 15 GHz
15 GHz to 20 GHz
±5.0 dB
±2.5 dB
+5.0 dB, –7.0 dB
±0.75 dB (–25 to 8 dBm)
±0.75 dB (–25 to 10 dBm)
±0.75 dB (–25 to 9 dBm )
±0.75 dB (–25 to 7 dBm )
±0.75 dB (–25 to 3 dBm )
±0.75 dB (–25 to 0 dBm )
Level linearity5
(swept sweep mode)3
300 kHz to 1 MHz
1 MHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10.5 GHz
10.5 GHz to 15 GHz
15 GHz to 20 GHz
Range5, 6
300 kHz to 1 MHz
1 MHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10.5 GHz
10.5 GHz to 15 GHz
15 GHz to 20 GHz
Typical
±1.5 dB (–25 to 8 dBm)
±1.5 dB (–25 to 10 dBm)
±1.5 dB (–25 to 9 dBm )
±1.5 dB (–25 to 7 dBm )
±1.5 dB (–25 to 3 dBm )
±1.5 dB (–25 to 0 dBm )
–85 to 8 dBm
–85 to 10 dBm
–85 to 9 dBm
–85 to 7 dBm
–85 to 3 dBm
–85 to 0 dBm
Sweep range4, 5, 6
300 kHz to 1 MHz
1 MHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10.5 GHz
10.5 GHz to 15 GHz
15 GHz to 20 GHz
(Source attenuator = 0 dB)
–25 to 8 dBm
–25 to 10 dBm
–25 to 9 dBm
–25 to 7 dBm
–25 to 3 dBm
–25 to 0 dBm
Level resolution
0.05 dB
1.
2.
3.
4.
5.
6.
Level accuracy is taken at –5 dBm, relative to 50 MHz reference unless otherwise stated.
Level accuracy is taken at –5 dBm, relative to 50 MHz reference.
Level linearity given is relative to –5 dBm.
The sweep range shifts based on the selected source attenuator value (0 dB to 60 dB, 10 dB step).
The level accuracy specification needs to be taken into account for test port output power level.
Power calibration using an external power meter improves level accuracy of the test port output power.
Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration.
7. Source output performance on port 1 only. Other port output performance is typical.
16
Test port output power6 (continued)
Table 13. Option 2D5/2K5/4D5/4K5
Description
Specification
SPD
power3, 4
Max leveled
300 kHz to 1 MHz
1 MHz to 10 GHz
10 GHz to 13 GHz
13 GHz to 15 GHz
15 GHz to 18 GHz
18 GHz to 20 GHz
9 dBm
10 dBm
9 dBm
7 dBm
5 dBm
4 dBm
Test port output signal purity
Table 14. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Description
Specification
Typical
Harmonics (2nd or 3rd)
9 kHz to 2 GHz
2 GHz to 8.5 GHz
< –25 dBc (at 5 dBm)
< –20 dBc (at 5 dBm)
Non-harmonic spurious
9 kHz to 8.5 GHz
< –30 dBc (at 5 dBm)
Table 15. Option 2D5/2K5/4D5/4K5
Description
Harmonics (2nd to
300 kHz to 1 GHz
Specification
Typical
5th)1
1 GHz to 20 GHz
Non-harmonic spurious2
300 kHz to 20 GHz
< –12 dBc (at maximum
output power)5
< –15 dBc (at maximum
output power)5
< –30 dBc (at –5 dBm)
1.
2.
3.
4.
This includes 6th and 7th harmonics when the test frequency is from 1 MHz to 1 GHz.
The carrier ±1/8th spurious is excluded from 8.76 GHz to 17.52 GHz.
The level accuracy specification needs to be taken into account for test port output power level.
Power calibration using an external power meter improves level accuracy of the test port output power.
Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration.
5. Maximum output power is the maximum power of "Range" in Table 13 Test port output power.
6. Source output performance on port 1 only. Other port output performance is typical.
17
Test Port Input
Test port input levels
Table 16. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Description
Specification
Typical
Maximum test port input level
9 kHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
+10 dBm
+9 dBm
+8 dBm
+7 dBm
Damage level
9 kHz to 8.5 GHz
+26 dBm
±35 VDC
Crosstalk1
9 kHz to 300 kHz
300 kHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
–100 dB
–110 dB
–120 dB
–110 dB
–100 dB
Description
Specification
SPD
Test port noise floor
9 kHz to 10 MHz
10 MHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8 GHz
8 GHz to 8.5 GHz
–115 dBm/Hz
–130 dBm/Hz
–131 dBm/Hz
–130 dBm/Hz
–129 dBm/Hz
–127 dBm/Hz
1. The specification might not be met at 5 MHz or 50 MHz.
18
Test port input levels (continued)
Table 17. Option 2D5/2K5/4D5/4K5
Description
Specification
Typical
Maximum test port input level
300 kHz to 20 GHz
+10 dBm
Damage level
300 kHz to 20 GHz
+26 dBm
or ±35 VDC
Crosstalk1
300 kHz to 1 MHz
1 MHz to 5 MHz
5 MHz to 10 MHz
10 MHz to 45 MHz
45 MHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 8.5 GHz
8.5 GHz to 15 GHz
15 GHz to 20 GHz
–68 dB
–70 dB
–100 dB
–110 dB
–118 dB
–123 dB
–120 dB
–112 dB
–106 dB
Description
Specification
SPD
Test port noise floor
300 kHz to 1 MHz
1 MHz to 10 MHz
10 MHz to 100 MHz
100 MHz to 8.5 GHz
8.5 GHz to 15 GHz
15 GHz to 20 GHz
–110 dBm/Hz
–115 dBm/Hz
–129 dBm/Hz
–130 dBm/Hz
–118 dBm/Hz
–115 dBm/Hz
1. The specification might not be met at 5 MHz or 50 MHz.
19
Test port input (trace noise)
Table 18. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Description
Specification
SPD
Compression level
Magnitude
9 kHz to 5 GHz
5 GHz to 8.5 GHz
(maximum test port input level = +10 dBm)
0.08 dB
0.1 dB
Phase
9 kHz to 3 GHz
3 GHz to 5 GHz
5 GHz to 8.5 GHz
(maximum test port input level = +10 dBm)
0.3 deg
0.6 deg
1.0 deg
Trace noise magnitude1
9 kHz to 30 kHz (IFBW = 3 kHz)
(maximum test port input level = +10 dBm)
0.004 dBrms
0.001 dBm
30 kHz to 100 kHz (IFBW = 3 kHz)
(maximum test port input level = +10 dBm)
0.003 dBm
0.001 dBm
100 kHz to 10 MHz (IFBW = 3 kHz)
(maximum test port input level = +10 dBm)
0.003 dBrms
0.0005 dBm
10 MHz to 4.38 GHz (IFBW = 70 kHz)
(maximum test port input level = +10 dBm)
0.004 dBrms
0.001 dBrms
4.38 GHz to 5 GHz (IFBW = 70 kHz)
(maximum test port input level = +10 dBm)
0.006 dBrms
0.0012 dBm
5 GHz to 6 GHz (IFBW = 70 kHz)
(maximum test port input level = +9 dBm)
0.006 dBrms
0.0012 dBm
6 GHz to 7 GHz (IFBW = 70 kHz)
(maximum test port input level = +8 dBm)
0.006 dBrms
0.0012 dBm
7 GHz to 8.5 GHz (IFBW = 70 kHz)
(maximum test port input level = +7 dBm)
0.006 dBrms
0.0012 dBm
Trace noise phase2
9 kHz to 30 kHz (IFBW = 3 kHz)
(maximum test port input level = +10 dBm)
0.035 deg rms
30 kHz to 10 MHz (IFBW = 3 kHz)
(maximum test port input level = +10 dBm)
0.020 deg rms
10 MHz to 4.38 GHz (IFBW = 70 kHz)
(maximum test port input level = +10 dBm)
0.035 deg rms
4.38 GHz to 5 GHz (IFBW = 70 kHz)
(maximum test port input level = +10 dBm)
0.050 deg rms
5 GHz to 6 GHz (IFBW = 70 kHz)
(maximum test port input level = +9 dBm)
0.050 deg rms
6 GHz to 7 GHz (IFBW = 70 kHz)
(maximum test port input level = +8 dBm)
0.050 deg rms
7 GHz to 8.5 GHz (IFBW = 70 kHz)
(maximum test port input level = +7 dBm)
0.050 deg rms
1. The specification might not be met at the following frequencies: 333.333 kHz, 406.25 kHz,
857.143 kHz, 928.571 kHz, 1.3 MHz, 2.4 MHz and 4.333333 MHz.
2. The specification might not be met at 5 MHz or 50 MHz.
20
Test port input (trace noise) (continued)
Table 19. Option 2D5/2K5/4D5/4K5
Description
Specification
SPD
Compression level
Magnitude
300 kHz to 10 MHz
0.3 dB
10 MHz to 10 GHz
0.182 dB
10 GHz to 15 GHz
0.712 dB
15 GHz to 20 GHz
1.87 dB
(maximum test port input
level = +10 dBm)
Phase
300 kHz to 5 GHz
2.3 deg
5 GHz to 10 GHz
4.3 deg
10 GHz to 15 GHz
17.3 deg
15 GHz to 20 GHz
20.3 deg
(maximum test port input
level = +10 dBm)
Trace noise magnitude1, 3
300 kHz to 1 MHz (IFBW = 3 kHz)
0.006 dBrms
0.0009 dBrms
1 MHz to 10 MHz (IFBW = 3 kHz)
0.003 dBrms
0.0005 dBrms
10 MHz to 4.38 GHz (IFBW = 70 kHz)
0.004 dBrms
0.0010 dBrms
4.38 GHz to 8.5 GHz (IFBW = 70 kHz)
0.006 dBrms
0.0012 dBrms
8.5 GHz to 13.137 GHz (IFBW = 70 kHz)
0.009 dBrms
0.0024 dBrms
13.137 GHz to 17 GHz (IFBW = 70 kHz)
0.013 dBrms
0.0040 dBrms
17 GHz to 20 GHz (IFBW = 70 kHz)
0.023 dBrms
0.0065 dBrms
300 kHz to 1 MHz (IFBW = 3 kHz)
0.040 deg rms
0.0120 deg rms
1 MHz to 10 MHz (IFBW = 3 kHz)
0.020 deg rms
0.0025 deg rms
10 MHz to 4.38 GHz (IFBW = 70 kHz)
0.035 deg rms
0.0075 deg rms
4.38 GHz to 8.5 GHz (IFBW = 70 kHz)
0.050 deg rms
0.0150 deg rms
8.5 GHz to 13.137 GHz (IFBW = 70 kHz)
0.064 deg rms
0.0250 deg rms
13.137 GHz to 17 GHz (IFBW = 70 kHz)
0.095 deg rms
0.0320 deg rms
17 GHz to 20 GHz (IFBW = 70 kHz)
0.165 deg rms
0.0520 deg rms
(at maximum output power level
of sweep range)
Trace noise phase2, 3
(at maximum output power level
of sweep range)
1. The specification might not be met at the following frequencies: 406.25 kHz, 666.667 kHz,
722.222 kHz, 857.143 kHz, 928.571 kHz, 1.444444 MHz, 1.714286 MHz, 1.8 MHz, 1.857143 MHz,
1.95 MHz, 2.4375 MHz, 2.571429 MHz, 3.714286 MHz, 4.8 MHz, 5 MHz, 5.416667 MHz,
7.583333 MHz, 10 MHz, 10.833333 MHz, 12.5 MHz.
2. The specification might not be met at 5 MHz or 50 MHz.
3. Trace noise SPD is defined with transmission measurements only.
21
Test port input (stability)1
Table 20. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Description
Specification
Typical
Stability magnitude
9 kHz to 3 GHz
±0.005 dB/°C
3 GHz to 6 GHz
±0.01 dB/°C
6 GHz to 8.5 GHz
±0.04 dB/°C
Stability phase
9 kHz to 3 GHz
±0.1 deg/°C
3 GHz to 6 GHz
±0.2 deg/°C
6 GHz to 8.5 GHz
±0.8 deg/°C
Table 21. Option 2D5/2K5/4D5/4K5
Description
Specification
Typical
Stability magnitude
300 kHz to 3 GHz
±0.005 dB/°C
3 GHz to 6 GHz
±0.01 dB/°C
6 GHz to 20 GHz
±0.04 dB/°C
Stability phase
300 kHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 20 GHz
±0.1 deg/°C
±0.2 deg/°C
±0.8 deg/°C
1. Stability is defined as a ratio measurement at the test port.
22
Test port input (dynamic accuracy)
Table 22. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Accuracy of the test port input power reading is relative to –10 dBm reference input
power level.
Description
Specification
Typical
Dynamic accuracy magnitude
10 dBm
±0.21 dB
–30 dBm
±0.05 dB
–100 dBm
±2.01 dB
Dynamic accuracy phase
10 dBm
±5 deg
–30 dBm
±0.3 deg
–100 dBm
±15.1 deg
Specification
Typical
Magnitude
Phase
100
Accuracy [degrees]
Accuracy [dB]
10
1
0.1
0.01
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90
10
1
0.1
-100
Test port power [dBm]
10
0
-10
-20
-30
-40
-50
-60
Test port power [dBm]
±3.0 dB (at –110 dBm, Ref.= –10 dBm, typical)
23
-70
-80
-90
-100
Test port input (dynamic accuracy) (continued)
Table 23. Option 2D5/2K5/4D5/4K5
Accuracy of the test port input power reading is relative to –10 dBm reference input
power level.
Description
Specification
Typical
Dynamic accuracy magnitude
10 dBm
±2.4972 dB
–30 dBm
±0.05 dB
–100 dBm
±2.01 dB
Dynamic accuracy phase
10 dBm
±20.66 deg
–30 dBm
±0.3 deg
–100 dBm
±15.1 deg
Specification
Typical
Phase
Magnitude
100
Accuracy [degrees]
Accuracy [dB]
10
1
0.1
0.01
10
0
10
1
0.1
-10 -20 -30 -40 -50 -60 -70 -80 -90 -100
10
0
-10 -20 -30 -40 -50 -60 -70 -80 -90 -100
Test port power [dBm]
Test port power [dBm]
±3.0 dB (at –110 dBm, Ref.= –10 dBm, typical)
24
Test port input (group delay)1
Table 24. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Description
Specification
Aperture (selectable)
(frequency span)/(number
of points - 1)
Supplemental information
Maximum aperture
25% of frequency span
Minimum delay
Limited to measuring no more
than 180 °C of phase change
within the minimum aperture.
Accuracy
See graph below, typical
The following graph shows group delay accuracy with type-N connectors, full 2-port
calibration and a 10 Hz IF bandwidth.
Calibration kit (85032F). Insertion loss is assumed to be < 2 dB.
100
Accuracy (nSec)
10
1
0.1
0.01
0.001
0.01
0.1
1
10
100
Aperture (MHz)
In general, the following formula can be used to determine the accuracy, in seconds, of a
specific group delay measurement:
±phase accuracy (degrees) / [360 x aperture (Hz)]
1. Group delay is computed by measuring the phase change within a specified step (determined by
the frequency span and the number of points per sweep).
25
Test port input (group delay)1 (continued)
Table 25. Option 2D5/2K5/4D5/4K5
Description
Specification
Aperture (selectable)
(frequency span)/(number
of points - 1)
Supplemental information
Maximum aperture
25% of frequency span
Minimum delay
Limited to measuring no more
than 180 °C of phase change
within the minimum aperture.
Accuracy
See graph below, typical
The following graph shows group delay accuracy with 3.5 mm (male) connectors, full
2-port calibration and a 10 Hz IF bandwidth.
Calibration kit (85052D). Insertion loss is assumed to be < 2 dB.
100
Accuracy (nSec)
10
1
0 .1
0 .0 1
Frequency = 1 G H z
C al pow er = -10 dBm ; M eas pow er = -10 dBm ; Electrical Length = 10 m
0 .0 0 1
0 .0 1
0 .1
1
10
100
Aperture (MHz)
In general, the following formula can be used to determine the accuracy, in seconds, of a
specific group delay measurement:
±phase accuracy (degrees) / [360 x aperture (Hz)]
1. Group delay is computed by measuring the phase change within a specified step (determined by
the frequency span and the number of points per sweep).
26
General Information
Table 26. System bandwidth
Description
General characteristics
IF bandwidth settings
Range
10 Hz to 500 kHz
Nominal settings are:
10, 15, 20, 30, 40, 50, 70, 100, 150, 200, 300, 400, 500, 700, 1 k,
1.5 k, 2 k, 3 k, 4 k, 5 k, 7 k, 10 k, 15 k, 20 k, 30 k, 40 k, 50 k, 70 k,
100 k, 150 k, 200 k, 300 k, 400 k, 500 kHz
Table 27. Front panel information
Description
Typical
General characteristics
RF connectors
Type-N, female, 50 Ω
Option 230/235/240/
245/260/265/280/285/
430/435/440/445/460/
465/480/485
Option 2D5/2K5/4D5/4K5
3.5 mm (male), 50 Ω nominal
Probe power
Connector
Voltage & maximum
current
3 terminal connector x 2
+15 V ±2% (400 mA)
–12.6 V ±5% (300 mA)
(Combined load for both
probe connections)
Display
Type
10.4 in TFT color LCD
with touch screen
Resolution
XGA (1024 x 768)1
1. Valid pixels are 99.99 % and more. Below 0.01 % (approx. 30 points) of fixed points of black, blue,
green or red are not regarded as failure.
27
Table 28. Rear panel information
Description
Typical
General characteristics
External trigger input connector
Type
BNC, female
Input level
Low threshold voltage: 0.5 V
High threshold voltage: 2.1 V
Input level range: 0 to + 5 V
Pulse width
≥ 2 μsec
Polarity
Positive or negative
External trigger output connector
Type
BNC, female
Maximum output current
50 mA
Output level
Low level voltage: 0 V
High level voltage: 5 V
Pulse width
1 μsec
Polarity
Positive or negative
External reference signal input connector
Type
BNC, female
Input frequency
10 MHz ±10 ppm
Input level
–3 to + 10 dBm
Internal reference signal output connector
Type
BNC, female
Output frequency
10 MHz ±5 ppm
Signal type
Sinewave
Output level
0 dBm ±3 dB into 50 Ω
Output impedance
50 Ω
Internal reference signal oven connector
Type
BNC, female
Output frequency
10 MHz ±1 ppm
Output level
0 dBm minimum
28
Table 28. Rear panel information (continued)
Description
Typical
General characteristics
Bias tee input connector
Type
BNC, female (for each port)
Maximum voltage
±35 VDC
Maximum current (no degradation in RF specifications)
Option 230/235/240/245/260/265/280/
285/430/435/440/445/460/465/480/485
±50 mA (1 MHz < test
frequency < 500 MHz)
±200 mA (test frequency
≤ 1 MHz, ≥ 500 MHz)
Option 2D5/2K5/4D5/4K5
±200 mA
Maximum current (damage level)
±500 mA
Fuse
500 mA, bi-pin style
Video output
15-pin mini D-Sub; female;
drives XGA compatible
monitors
GPIB
24-pin D-Sub (Type D-24),
female; compatible with
IEEE-488
USB-host port
Universal serial bus jack, Type A configuration (4 contacts
inline, contact 1 on left); female; provides connection to
printer, ECal module, USB/GPIB interface or multiport test
set
Contact 1
Vcc: 4.75 to 5.25 VDC, 500
mA, maximum
Contact 2
– Data
Contact 3
+ Data
Contact 4
Ground
29
Table 28. Rear panel information (continued)
Description
Typical
General characteristics
USB (USBTMC1)
interface port
Universal serial bus jack,
Type B configuration
(4 contacts inline); female;
provides connection to an
external PC; compatible
with USBTMC-USB488 and
USB 2.0.
LAN
10/100BaseT Ethernet, 8-pin
configuration; auto selects
between the two data rates
Handler I/O port
36-pin centronics, female;
provides connection to
handler system
Line power2
Frequency
47 Hz to 63 Hz
Voltage
90 to 132 VAC, or 198 to
264 VAC (automatically
switched)
VA max
350 VA max.
Description
Specification
General characteristics
AUX input connector
Type
BNC, female x 2
Input range
Accuracy
±1 V or ±10 V selectable
1 % + 1 mV for ±1 V input
1 % + 10 mV for ±10 V input
Table 29. LXI compliance
Description
General characteristics
Class C (only applies to units that are shipped with
firmware revision A.08.00 or later)
1. USB Test and Measurement Class (TMC) interface that communicates over USB, complying with
the IEEE 488.1 and IEEE 488.2 standards.
2. A third-wire ground is required.
30
Table 30. EMC, safety and environment
Description
General characteristics
EMC
European Council Directive 2004/108/EC
IEC 61326-1:2005
EN 61326-1:2006
CISPR 11:2003+A1:2004
EN 55011:2007
Group 1, Class A
IEC 61000-4-2:1995 +A2:2000
EN 61000-4-2:1995 +A2:2001
4 kV CD / 8 kV AD
IEC 61000-4-3:2006
EN 61000-4-3:2006
1-3 V/m, 80-1000 MHz / 1.4 GHz - 2.7 GHz, 80% AM
IEC 61000-4-4:2004
EN 61000-4-4:2004
1 kV power lines / 0.5 kV signal lines
IEC 61000-4-5:2005
EN 61000-4-5:2006
0.5 kV line-line / 1 kV line-ground
IEC 61000-4-6:2003 + A1:2004+ A2:2006
EN 61000-4-6:2007
3 V, 0.15-80 MHz, 80% AM
IEC 61000-4-11:2004
EN 61000-4-11:2004
0.5-300 cycle, 0% / 70%
ICES-001:2006 Group 1, Class A
AS/NZS CISPR11:2004
Group 1, Class A
Safety
European Council Directive 2006/95/EC
IEC 61010-1:2001 / EN 61010-1:2001
Measurement Category I
Pollution Degree 2
Indoor Use
CAN/CSA C22.2 No. 61010-1-04
Measurement Category I
Pollution Degree 2
Indoor Use
Environment
This product complies with the WEEE Directive (2002/96/EC) marking
requirements. The affixed label indicates that you must not discard
this electrical/electronic product in domestic household waste.
Product Category: With reference to the equipment types in the
WEEE Directive Annex I, this product is classed as a “Monitoring
and Control instrumentation” product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Agilent office, or
see www.agilent.com/environment/product/ for more information.
31
Table 31. Analyzer environmental specifications and dimensions
Description
General characteristics
Operating environment
Temperature
+5 °C to +40 °C
Error-corrected temperature range
23 °C (±5 °C) with < 1 °C deviation from calibration temperature
Humidity
20% to 80% at wet bulb temperature < +29 °C (non-condensation)
Altitude
0 to 2,000 m (0 to 6561 feet)
Vibration
0.21 G maximum, 5 Hz to 500 Hz
Non-Operating storage environment
Temperature
–10 °C to +60 °C
Humidity
20% to 90% at wet bulb temperature < +40 °C (non-condensation)
Altitude
0 to 4572 m (0 to 15,000 feet)
Vibration
0.5 G maximum, 5 Hz to 500 Hz
Dimensions
See next page.
Weight (net)
Option 230/240/260/280 (2-port)
Option 235/245/265/285 (2-port)
Option 430/480/440/460 (4-port)
Option 435/445/465/485 (4-port)
Option 2D5/2K5 (2-port)
Option 4D5/4K5 (4-port)
18.2 kg
18.3 kg
19.9 kg
20.0 kg
19.8 kg
21.8 kg
32
22
19
22
222
426
27
58
20
51
122
37
Figure 3. Dimensions (front view, E5071C with Option 230/235/240/245/260/265/280/285,
in millimeters)
22
19
22
222
426
27
58
20
51
41
41
41
Figure 4. Dimensions (front view, E5071C with Option 430/435/440/445/460/465/480/485,
in millimeters)
33
37
26
19
22
222
426
27
58
20
51
122
Figure 5. Dimensions (front view, E5071C with Option 2D5/2K5, in millimeters)
Figure 6. Dimensions (front view, E5071C with Option 4D5/4K5, in millimeters)
34
37
424
422
128
1
96
21 18
27
25
25
290
25
58
109
148
25
144
146
159
179
190
215
1
25
25
25
165
Figure 7. Dimensions (rear view with Option 1E5, in millimeters)
472
23.4
417
13
18
235
215
222
32
15
Figure 8. Dimensions (side view, E5071C Option 230/235/240/245/260/265/280/285/430/435/440/445/460/465/480/485,
in millimeters)
472
417
23
13
18
215
222
235
32
14
Figure 9. Dimensions (side view, E5071C with Option 2D5/2K5/4D5/4K5, in millimeters)
35
Measurement Throughput Summary
Cycle time for measurement completion1, 2 (ms)
Table 32. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: OFF
Number of points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth
Uncorrected
3.4
4.9
7.1
20
2-port cal
6.3
9.3
14
41
4-port cal
13
19
28
92
Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
3.7
5.2
7.3
20
2-port cal
6.6
10
15
42
4-port cal
13
21
32
94
Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth
Uncorrected
54
202
397
1563
2-port cal
107
401
792
3125
4-port cal
213
800
1581
6247
Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth
Uncorrected
8.7
11
12
25
2-port cal
17
21
24
48
4-port cal
34
42
49
102
Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
8.8
11
13
26
2-port cal
17
22
24
54
4-port cal
34
42
49
118
Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth
Uncorrected
56
206
404
1583
2-port cal
112
410
805
3163
4-port cal
222
818
1608
6324
1. Supplemental performance data.
2. Measured with firmware version A.09.30.
36
Table 32. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485 (continued)
Number of points
51
201
401
1601
Start 100 kHz, stop 8.5 GHz, 500 kHz IF bandwidth
Uncorrected
15
18
19
24
2-port cal
28
35
38
48
4-port cal
55
69
74
104
Start 100 kHz, stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected
15
18
20
26
2-port cal
28
36
38
52
4-port cal
56
70
75
117
Start 100 kHz, stop 8.5 GHz, 1 kHz IF bandwidth
Uncorrected
57
207
404
1583
2-port cal
114
412
806
3163
4-port cal
225
821
1611
6324
37
Cycle time for measurement completion1, 2 (ms)
Table 33. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: ON
Number of Points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth
Uncorrected
3.5
5.5
8.0
24
2-port cal
6.4
9.8
15
49
4-port cal
13
20
29
108
Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
3.6
5.4
8.0
23
2-port cal
6.6
11
16
48
4-port cal
13
22
33
107
Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth
Uncorrected
54
202
397
1563
2-port cal
107
401
792
3125
4-port cal
213
800
1581
6247
Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth
Uncorrected
8.7
11
13
28
2-port cal
18
21
24
53
4-port cal
34
42
49
112
Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
8.8
11
13
28
2-port cal
17
22
25
57
4-port cal
34
42
50
122
Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth
Uncorrected
56
206
404
1583
2-port cal
112
410
805
3163
4-port cal
222
818
1608
6324
1. Supplemental performance data.
2. Measured with firmware version A.09.30.
38
Table 33. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485 (continued)
Number of points
51
201
401
1601
Start 100 kHz, stop 8.5 GHz, 500 kHz IF bandwidth
Uncorrected
15
18
20
27
2-port cal
28
35
38
52
4-port cal
55
69
74
111
Start 100 kHz, stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected
15
18
20
27
2-port cal
28
36
38
56
4-port cal
56
70
75
120
Start 100 kHz, stop 8.5 GHz, 1 kHz IF bandwidth
Uncorrected
57
207
404
1583
2-port cal
114
412
806
3163
4-port cal
225
821
1611
6324
39
Cycle time for measurement completion1, 2 (ms)
Table 34. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Sweep mode: Stepped, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: ON
Number of Points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth
Uncorrected
4.4
7.7
12
29
2-port cal
8.2
15
22
56
4-port cal
16
29
45
126
Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
4.8
9.2
15
41
2-port cal
9.0
18
28
80
4-port cal
18
35
56
163
Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth
Uncorrected
54
202
397
1563
2-port cal
107
401
792
3125
4-port cal
213
800
1581
6247
Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth
Uncorrected
6.7
13
18
48
2-port cal
13
24
36
94
4-port cal
26
47
70
190
Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
7.1
14
22
60
2-port cal
14
27
42
119
4-port cal
27
53
82
237
Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth
Uncorrected
56
206
404
1583
2-port cal
112
410
805
3163
4-port cal
222
818
1608
6324
1. Supplemental performance data.
2. Measured with firmware version A.09.30.
40
Table 34. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485 (continued)
Number of points
51
201
401
1601
Start 100 kHz, stop 8.5 GHz, 500 kHz IF bandwidth
Uncorrected
7.5
13
19
48
2-port cal
15
26
37
94
4-port cal
29
50
73
189
Start 100 kHz, stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected
7.9
15
22
60
2-port cal
16
29
43
120
4-port cal
30
56
85
237
Start 100 kHz, stop 8.5 GHz, 1 kHz IF bandwidth
Uncorrected
57
207
404
1583
2-port cal
114
412
806
3163
4-port cal
225
821
1611
6324
41
Cycle time for measurement completion1, 2 (ms)
Table 35. Option 2D5/2K5/4D5/4K5
Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: OFF
Number of points
51
201
401
1601
Start 300 kHz, stop 20 GHz, 500 kHz IF bandwidth
Uncorrected
23
36
39
43
2-port cal
45
71
77
85
4-port cal
88
142
155
169
Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth
Uncorrected
23
37
40
43
2-port cal
45
72
78
86
4-port cal
89
144
156
171
Start 300 kHz, stop 20 GHz, 1 kHz IF bandwidth
Uncorrected
60
212
410
1591
2-port cal
120
421
817
3180
4-port cal
237
840
1632
6358
Start 8 GHz, stop 18 GHz, 500 kHz IF bandwidth
Uncorrected
17
22
22
23
2-port cal
34
42
44
45
4-port cal
67
83
86
102
Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth
Uncorrected
18
22
23
24
2-port cal
34
42
44
47
4-port cal
67
84
87
106
Start 8 GHz, stop 18 GHz, 1 kHz IF bandwidth
Uncorrected
58
207
405
1583
2-port cal
115
413
807
3163
4-port cal
229
823
1612
6324
1. Supplemental performance data.
2. Measured with firmware version A.09.30.
42
Table 35. Option 2D5/2K5/4D5/4K5 (continued)
Number of points
51
201
401
1601
Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth
Uncorrected
3.0
3.2
4.9
17
2-port cal
5.6
6.1
9.9
37
4-port cal
11
14
23
84
Start 11 GHz, stop 12 GHz, 100 kHz IF bandwidth
Uncorrected
3.0
3.4
5.5
18
2-port cal
5.6
6.4
11
38
4-port cal
11
14
24
85
Start 11 GHz, stop 12 GHz, 1 kHz IF bandwidth
Uncorrected
52
200
396
1566
2-port cal
104
399
790
3130
4-port cal
206
795
1578
6258
43
Cycle time for measurement completion1, 2 (ms)
Table 36. Option 2D5/2K5/4D5/4K5
Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: ON
Number of points
51
201
401
1601
Start 300 kHz, stop 20 GHz, 500 kHz IF bandwidth
Uncorrected
23
36
39
43
2-port cal
45
71
77
85
4-port cal
88
142
155
170
Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth
Uncorrected
23
37
40
43
2-port cal
45
72
78
86
4-port cal
89
144
156
171
Start 300 kHz, stop 20 GHz, 1 kHz IF bandwidth
Uncorrected
60
212
410
1591
2-port cal
120
421
817
3180
4-port cal
237
840
1632
6358
Start 8 GHz, stop 18 GHz, 500 kHz IF bandwidth
Uncorrected
17
22
22
23
2-port cal
34
42
44
48
4-port cal
67
83
86
107
Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth
Uncorrected
18
22
23
24
2-port cal
34
42
44
49
4-port cal
67
83
87
110
Start 8 GHz, stop 18 GHz, 1 kHz IF bandwidth
Uncorrected
58
207
405
1583
2-port cal
115
413
807
3163
4-port cal
229
823
1612
6324
1. Supplemental performance data.
2. Measured with firmware version A.09.30.
44
Table 36. Option 2D5/2K5/4D5/4K5 (continued)
Number of points
51
201
401
1601
Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth
Uncorrected
3.0
3.4
5.9
20
2-port cal
5.6
6.6
12
43
4-port cal
11
14
26
98
Start 11 GHz, stop 12 GHz, 100 kHz IF bandwidth
Uncorrected
3.0
3.5
5.8
20
2-port cal
5.6
6.8
12
43
4-port cal
11
15
26
98
Start 11 GHz, stop 12 GHz, 1 kHz IF bandwidth
Uncorrected
52
200
396
1566
2-port cal
104
399
790
3130
4-port cal
206
795
1578
6258
45
Cycle time for measurement completion1, 2 (ms)
Table 37. Option 2D5/2K5/4D5/4K5
Sweep mode: Stepped, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: ON
Number of points
51
201
401
1601
Start 300 kHz, stop 20 GHz, 500 kHz IF bandwidth
Uncorrected
11
18
24
56
2-port cal
21
35
48
112
4-port cal
41
69
94
222
Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth
Uncorrected
11
20
28
69
2-port cal
22
38
54
137
4-port cal
42
75
107
272
Start 300 kHz, stop 20 GHz, 1 kHz IF bandwidth
Uncorrected
60
212
410
1591
2-port cal
120
421
817
3180
4-port cal
237
840
1632
6358
Start 8 GHz, stop 18 GHz, 500 kHz IF bandwidth
Uncorrected
8.4
14
19
48
2-port cal
17
27
38
94
4-port cal
32
52
74
188
Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth
Uncorrected
8.8
15
22
60
2-port cal
17
30
44
120
4-port cal
34
58
86
237
Start 8 GHz, stop 18 GHz, 1 kHz IF bandwidth
Uncorrected
58
207
405
1583
2-port cal
115
413
807
3163
4-port cal
229
823
1612
6324
1. Supplemental performance data.
2. Measured with firmware version A.09.30.
46
Table 37. Option 2D5/2K5/4D5/4K5 (continued)
Number of points
51
201
401
1601
Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth
Uncorrected
2.8
6.4
11
31
2-port cal
5.1
13
21
61
4-port cal
9.6
24
41
131
Start 11 GHz, stop 12 GHz, 100 kHz IF bandwidth
Uncorrected
3.4
8.0
14
43
2-port cal
5.9
16
27
86
4-port cal
12
30
53
171
Start 11 GHz, stop 12 GHz, 1 kHz IF bandwidth
Uncorrected
52
200
396
1566
2-port cal
104
399
790
3130
4-port cal
206
795
1578
6258
47
Cycle time (ms)1, 2 vs. number of points
Table 38. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth, error correction: OFF,
display update: OFF, number of traces = 1
Number of
points
Sweep mode:
Fast swept,
system error
correction: OFF
Sweep mode:
Fast swept,
system error
correction: ON
Sweep mode:
Std stepped,
system error
correction: ON
3
3.1
3.1
3.4
11
3.3
3.3
3.3
51
3.5
3.5
4.4
101
3.9
4.1
5.6
201
4.9
5.5
7.7
401
7.1
8.0
12
801
12
14
18
1601
20
24
29
Cycle time (ms)1, 2 vs. number of points
Table 39. Option 2D5/2K5/4D5/4K5
Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth, error correction: OFF,
display update: OFF, number of traces = 1
Number of
points
Sweep mode:
Fast swept,
system error
correction: OFF
Sweep mode:
Fast swept,
system error
correction: ON
Sweep mode:
Std stepped,
system error
correction: ON
51
3.0
3.0
2.8
201
3.2
3.4
6.5
401
4.9
5.9
11
1601
17
20
31
1. Supplemental performance data.
2. Measured with firmware version A.09.30.
48
Cycle time (ms)1, 2 vs. IF bandwidth
Table 40. All options
Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: OFF, NOP = 201
Option 230/235/240/245/260/265/280/
285/430/435/440/445/460/465/480/485
Option 2D5/2K5/4D5/4K5
Frequency
[Hz]
IFBW
[Hz]
Cycle time
[ms]
Frequency
[Hz]
IFBW
[Hz]
Cycle time
[ms]
4G
500000
5.0
10 G
500000
2.8
4G
400000
5.0
10 G
400000
2.8
4G
300000
4.9
10 G
300000
2.8
4G
200000
4.9
10 G
200000
2.8
4G
150000
4.9
10 G
150000
2.8
4G
100000
5.0
10 G
100000
3.0
4G
70000
5.7
10 G
70000
3.8
4G
50000
6.9
10 G
50000
4.9
4G
40000
7.8
10 G
40000
5.9
4G
30000
9.4
10 G
30000
7.5
4G
20000
13
10 G
20000
11
4G
15000
16
10 G
15000
14
4G
10000
23
10 G
10000
21
4G
7000
31
10 G
7000
29
4G
5000
42
10 G
5000
40
4G
4000
52
10 G
4000
50
4G
3000
68
10 G
3000
66
4G
2000
100
10 G
2000
98
4G
1500
133
10 G
1500
132
4G
1000
198
10 G
1000
196
4G
700
278
10 G
700
276
4G
500
391
10 G
500
389
4G
400
487
10 G
400
485
4G
300
648
10 G
300
646
4G
200
970
10 G
200
968
4G
150
1291
10 G
150
1289
4G
100
1935
10 G
100
1933
4G
70
2739
10 G
70
2739
4G
50
3864
10 G
50
3869
4G
40
4829
10 G
40
4837
4G
30
6437
10 G
30
6449
4G
20
9653
10 G
20
9671
4G
15
12870
10 G
15
12892
4G
10
19302
10 G
10
19330
1. Supplemental performance data.
2. Measured with firmware version A.09.30.
49
Data transfer time1, 2 (ms)
Table 41. All options
Number of points
51
SCPI over
201
401
1601
GPIB3
64-bit floating point
4
12
25
88
32-bit floating point
3
7
13
45
ASCII
30
108
223
841
REAL 64
1
1
2
4
REAL 32
1
1
1
2
ASCII
15
59
117
457
SCPI over 100 Mbps LAN (Telnet)3
SCPI over 100 Mbps LAN (SICL-LAN)3
REAL 64
3
3
4
6
REAL 32
3
3
3
5
4
8
14
47
2
2
2
3
ASCII
SCPI over USB
(SICL-USB)3
REAL 64
REAL 32
2
2
2
3
ASCII
4
13
24
91
REAL 64
8
15
21
68
REAL 32
7
10
13
37
ASCII
75
290
579
2313
1
1
1
1
SCPI over GPIB/USB (82357B)
COM4
Variant type
1. Supplemental performance data.
2. Measured with firmware version A.09.30.
3. Measured using a VEE Pro 7.0 program running on a 3.2 GHz Pentium 4 DELL Precision 370,
Transferred complex S11 data, using :CALC{1-36}:DATA:FDAT?.
4. Measured using an E5071C VBA macro running inside the analyzer. Transferred complex S11 data.
50
E5091A Multiport Test Set
The section provides port performance of test set without calibration.
Table 42. Test set input/output performance
Description
Specification
Range
Damage level
50 MHz to 8.5 GHz
Typical
20 dBm, ±25 VDC (typical)
Table 43. Option E5091A-009 port performance
Specification
50 MHz to
300 MHz
300 MHz to
1.3 GHz
1.3 GHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
8.5 GHz
A, T2, R1+, R1–
19 dB
20 dB
18 dB
12 dB
10 dB
T1, R2+, R2–, R3+, R3–
15 dB
17 dB
15 dB
11 dB
8 dB
Description
Load match
Test port selected
Test port unselected
A, T2, R1+, R1–, R3+, R3– 23 dB
25 dB
19 dB
12 dB
11 dB
T1, R2+, R2–
18 dB
20 dB
16 dB
12 dB
9 dB
19 dB
19 dB
17 dB
13 dB
9 dB
A, T2, R1+, R1–
3 dB
3 dB
4 dB
5 dB
6 dB
T1, R2+, R2–, R3+, R3–
5 dB
5 dB
7 dB
8 dB
9.5 dB
Stability, typical
0.005 dB/° C
0.005 dB/° C
0.005 dB/° C
0.01 dB/° C
0.015 dB/° C
–100 dB
–100 dB
–100 dB
–100 dB
–90 dB
Interconnect port, typical
P1, P2, P3, P4
Insertion loss
Test port
Isolation
Over arbitrarily test ports
51
Table 44. Option E5091A-016 port performance
Description
Specification
50 MHz to
300 MHz
300 MHz
to 1.3 GHz
1.3 GHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
8.5 GHz
17 dB
15 dB
9 dB
8 dB
Load match
Test port selected
A, T4, R1+, R1–, R2+, R2–, 15 dB
R3+, R3–, R4+, R4–
T1, T2, T3
12 dB
14 dB
14 dB
8 dB
6 dB
Test port unselected
A, T4, T2, R1+, R1–, R2+, R2– 18 dB
R3+, R3–, R4+, R4–, R4–
20 dB
16 dB
10 dB
9 dB
T1, T2, T3
13 dB
15 dB
14 dB
8 dB
6 dB
Interconnect port, typical
P1, P2, P3, P4
12 dB
12 dB
12 dB
9 dB
7 dB
Test port
A, T4, R1+, R1–, R2+, R2–, 6 dB
R3+, R3–, R4+, R4–
6 dB
7 dB
8 dB
9.5 dB
T1, T2, T3
6 dB
Stability per switch, typical 0.005 dB/° C
9 dB
0.005 dB/° C
10.5 dB
12 dB
0.005 dB/° C 0.01 dB/° C
14.5 dB
0.015 dB/° C
–100 dB
–100 dB
–80 dB
Insertion loss
Isolation
Over arbitrarily test ports
–100 dB
Table 45. Front panel information
Description
RF connectors
Interconnect ports
Test ports (Option E5091A-009)
Test ports (Option E5091A-016)
Control line
General characteristics
Type-N, female, 50 Ω, nominal
Number of ports 4 ports
Type-N, female, 50 Ω, nominal
Number of ports 9 ports
SMA, female, 50 Ω, nominal
Number of ports 25 ports
(includes configurable switch port)
15 pin D-sub, female
Table 46. Rear panel information
Description
USB port
Line power 1
Frequency
Voltage
VA max
General characteristics
Type B-receptacle, provide
connection to the E5071C
47 to 63 Hz
90 to 132 VAC, or 198 to 264 VAC
(automatically switched)
150 VA max.
1. A third-wire ground is required.
52
–100 dB
Table 47. Test set dimensions and block diagram
Description
Dimensions
Option E5091A-009
Option E5091A-016
Weight
Option E5091A-009
Option E5091A-016
Block diagram
Option E5091A-009/016
General characteristics
See Figure 10, 12, and 13
See Figure 11, 12, and 15
6 kg
7 kg
See Figure 15
Figure 10. Dimensions (front view, with Option E5091A-009, in millimeters, nominal)
Figure 11. Dimensions (front view, with Option E5091A-016, in millimeters, nominal)
53
Figure 12. Dimensions (rear view, in millimeters, nominal)
Figure 13 Dimensions (side view, with Option E5091A-009, in millimeters, nominal)
Figure 14. Dimensions (side view, with Option E5091A-016, in millimeters, nominal)
54
Figure 15. Block diagram
55
E5092A Configurable multiport test set
The section provides test input/output performance without calibration by the E5071C.
Table 48. Test set input/output performance
Description
Specification
Frequency range
50 MHz to 20 GHz
Damage level
Typical
20 dBm, ±35 VDC
Table 49. Option E5092A-020 port performance
Description
Specification
Typical
Load match (selected port)
SPDT switch1
50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
17 dB
11 dB
8 dB
7 dB
4 dB
4 dB
SP4T switch2
50 MHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
17 dB
11 dB
8 dB
7 dB
4 dB
4 dB
Load match (unselected port)
SPDT switch1
50 MHz to 3 GHz
3 GHz to 10 GHz
10 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 20 GHz
17 dB
11 dB
8 dB
6 dB
4 dB
SP4T switch2
50 MHz to 3 GHz
3 GHz to 10 GHz
10 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 20 GHz
17 dB
11 dB
8 dB
6 dB
4 dB
1. SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.)
2. SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.)
56
Table 49. Option E5092A-020 port performance (continued)
Description
Specification
Typical
Load match (common port)
SPDT switch1
50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 20 GHz
16 dB
11 dB
8 dB
7 dB
4 dB
SP4T switch2
50 MHz to 1.3 GHz
1.3 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 20 GHz
16 dB
11 dB
8 dB
7 dB
4 dB
Insertion loss
SPDT switch1
50 MHz to 100 MHz
100 MHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 14 GHz
14 GHz to 18 GHz
18 GHz to 20 GHz
4 dB
3.5 dB
4.5 dB
5 dB
5.5 dB
7 dB
8 dB
8.5 dB
10 dB
11.5 dB
SP4T switch2
50 to 100 MHz
100 MHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 14 GHz
14 GHz to 18 GHz
18 GHz to 20 GHz
4 dB
3.5 dB
4.5 dB
5.5 dB
6 dB
7.5 dB
8.5 dB
9.5 dB
10.5 dB
12 dB
Stability
50 MHz to 6 GHz
6 GHz to 20 GHz
0.003 dB/°C
0.01 dB/°C
(Stability per switch)
Isolation3
50 MHz to 500 MHz
500 MHz to 1 GHz
1 GHz to 2 GHz
2 GHz to 6 GHz
6 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
65 dB
80 dB
85 dB
90 dB
85 dB
75 dB
65 dB
(Over arbitrarily test ports)
1. SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.)
2. SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.)
3. This specification is defined when all ports are terminated with a 50 ohm load.
57
Table 50. Control line
Description
Specification
Number of groups
4
Group A: 8 bits
Group B,C,D: 4 bits
Input voltage range1
0 V to +5 V (positive input)
–5 V to 0 V (negative input)
Maximum current
Group A, B: 50 mA in total of each group
Group C, D: 500 uA in total of each group
Typical
Impedance
Group A, B: < 10 ohm
Group C, D: < 200 ohm
Table 51. DC source
Description
Specification
Number of sources
4
Typical
0 V to +5.2 V (nominal)2
Output voltage range
Output voltage accuracy
±3 % of setting (+1 V to +5 V) at
1 M ohm load impedance
10 mV (nominal)3
Voltage resolution
Maximum current
150 mA for each source
Output impedance
< 5 ohm
Table 52. Operating storage environment
Description
General characteristics
Temperature
+5 °C to +40 °C
Humidity
20 to 80% at wet bulb temperature
< +29 °C (non-condensing)
Altitude
0 to 2,000m (0 to 6,561 feet)
Vibration
0.21 G max., 5 to 500 Hz
Table 53. Non-operating storage environment
Description
General characteristics
Temperature
–10 °C to +60 °C
Humidity
20 to 90 % at wet bulb temperature
< +40 °C (non-condensing)
Altitude
0 to 4,572 m (0 to 15,000 feet)
Vibration
0.5 G max., 5 Hz to 500 Hz
1. Input voltage will be clipped at about ±5.2 V when over this range.
2. The output voltage can be set in this range.
3. The output voltage resolution becomes effective between 0 V to 5.2 V.
58
Table 54. Front panel information
Description
General characteristics
RF connectors
SMA
Test ports
38 ports
Control line
15-pin D-sub, female
25-pin D-sub, female
Table 55. Rear panel information
Description
General characteristics
USB port
Type B-receptacle, provide
connection to the E5071C
Line power1
Frequency
Voltage
VA max
47 to 63 Hz
90 to 132 VAC, or 198 to 264 VAC
(automatically switched)
300 VA max.
Table 56. Test set dimensions and block diagram
Description
General characteristics
Dimensions
E5092A Option 020
See Figures 16, 17, 18 and 19
Weight
E5092A Option 020
9 kg
426
192
133
18
129
Figure 16. Dimensions (front view, with Option E5092A-020, in millimeters, nominal)
1. 1A third-wire ground is required.
59
40.6
40.6
36.8
17
40.6
27
23.38
132.72
27
27
27
27.8
13.5
27
65.34
27
18.57 23.38
27
69.8
Figure 17. Dimensions (pitch between switches, with Option E5092A-020, in millimeters, nominal)
37
51
83
133
129
Figure 18. Dimensions (rear view, with Option E5092A-020, in millimeters, nominal)
472
417
22
133
146
32
13
18
144
126
23
Figure 19. Dimensions (side view, with Option E5092A-020, in millimeters, nominal)
60
SW 1
PORT 1
1A 1B 1C 1D
SW 5
5A 5B
PORT 2
SW 2
SW 3
6A 6B
6C O M
5C O M
7A 7B
SW 7
SW 4
3A 3B 3C 3D
2A 2B 2C 2D
SW 6
PORT 3
SW 8
7C O M
8A 8B
SW 9
4A 4B 4C 4D
9A 9B
9C O M
8C O M
PORT 4
S W 10
10A 10B
10C O M
Figure 20. Switch configuration (E5092A-020)
Control Line 1 (Group A)
DC Source (Group A)
Max Current
DC Source Output
Control Line 1 (Group A)
Control Line 2 (Group A)
Positive Input
(High)
Control Line 3 (Group A)
Output
Impedance
Output
Voltage
Control Line 4 (Group A)
:
Negative Input
(Low)
DC Source (Group B)
DC Source Output
Impedance
DC Source Output
Control Line 1 (Group B)
Control Line 2 (Group B)
Control Line 3 (Group B)
Control Line 4 (Group B)
Control Line 2 (Group C)
Control Line 1 (Group C)
Control Line 2 (Group C)
Control Line 3 (Group C)
Control Line 4 (Group C)
Positive Input (High)
Negative Input (Low)
DC Source (Group D)
DC Source Output
:
Control Line 8 (Group A)
Control Line 2 (Group B)
Positive Input (High)
Negative Input (Low)
DC Source (Group C)
:
Control Line 2 (Group D)
Control Line 1 (Group D)
Control Line 2 (Group D)
Control Line 3 (Group D)
Control Line 4 (Group D)
Positive Input (High)
Negative Input (Low)
Figure 21. DC control line (E5092A-020)
61
23
22
21
20
19
18
17
16
Positive input
DC source
GND
Negative input
Positive input
DC source
GND
Negative input
12
11
GROUP B
10
9
8
7
6
5
GROUP C
12
11
10
9
GND
Negative input
Positive input
DC source
Reserved
Reserved
Reserved
Line 8
Line 7
Line 6
Line 5
Line 4
Line 3
Line 2
Line1
8
7
6
5
4
3
2
1
GROUP C
4
3
Line 1
13
Line 2
Line 3
Line 4
Line 1
GROUP B
Line 2
Line 3
Line 4
Line 1
Line 2
Line 3
Line 4
GND
14
15
14
DC source
24
Negative input
13
15
Positive input
25
GND
D-SUB 15 pin
assignment
GROUP A
GROUP A
D-SUB 25pin
assignment
GROUP D
2
1
GROUP D
Figure 22. Control line pin assignment (E5092A-020)
62
Corrected System Performance for 75 Ω Measurements with 11852B
50 to 75 Ω Minimum-Loss Pads (Supplemental Information)
Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Table 55. Corrected system performance with type-N 75 Ω device connectors,
85036E calibration kit
Network analyzer:
calibration kit:
50 to 75 Ω adapters:
calibration:
E5071C
85036E (type-N 75 Ω)
11852B
full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23° C
±5° C with < 1° C deviation from calibration temperature, isolation calibration performed
Description
Typical (dB)
10 MHz to 3 GHz
Directivity
Source match
Load match
Reflection tracking
Transmission tracking
37
33
39
±0.015
±0.019
Transmission uncertainty 10 MHz to 3 GHz (typical)
Phase
Magnitude
Transmission using 85036E and 11852B
100
1
10
Phase (deg)
Mag (dB)
Transmission using 85036E and 11852B
10
0.1
1
10 MHz to 3 GHz
10 MHz to 3 GHz
S11 = S22 = 0; Source power = –10 dBm
0.01
10
0
S11 = S22 = 0; Source power = –10 dBm
0.1
10
–10 –20 –30 –40 –50 –60 –70 –80 –90
0
–10 –20 –30 –40 –50 –60 –70 –80 –90
Transmission coefficient (dB)
Transmission coefficient (dB)
Reflection uncertainty 10 MHz to 3 GHz (typical)
Phase
Magnitude
Reflection using 85036E and 11852B
Reflection using 85036E and 11852B
10
0.1
Phase (deg)
Mag (linear)
10 MHz to 3 GHz
8
10 MHz to 3 GHz
0.08
S21 = S12 = 0; Source power = –10 dBm
0.06
0.04
S21 = S12 = 0; Source power = –10 dBm
6
4
2
0.02
0
0
0
0.2
0.4
0.6
0.8
1
0
0.2
Reflection coefficient (linear)
0.4
0.6
0.8
Reflection coefficient (linear)
63
1
www.agilent.com
www.agilent.com/find/ena
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
www.lxistandard.org
LXI is the LAN-based successor to
GPIB, providing faster, more efficient
connectivity. Agilent is a founding
member of the LXI consortium.
Remove all doubt
Our repair and calibration services
will get your equipment back to
you, performing like new, when
promised. You will get full value out
of your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
For information regarding self maintenance of this product, please
contact your Agilent office.
Agilent offers a wide range of additional
expert test and measurement services
for your equipment, including initial
start-up assistance, onsite education
and training, as well as design, system
integration, and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
For more information on Agilent
Technologies’ products, applications
or services, please contact your local Agilent office. The complete list is
available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
01 36027 71571
Belgium
32 (0) 2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
07031 464 6333
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland
0800 80 53 53
United Kingdom 44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: March 24, 2009
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2007-2009
Printed in USA, July 30, 2009
5989-5479EN
Similar pages