Wall ME12D12-250W 5600vdc i/o isolation single and dual outputs 5~6 watt medical dc/dc converter Datasheet

Rev B
ME Series
5600VDC I/O Isolation
Single and Dual Outputs
5~6 Watt Medical DC/DC Converter
FEATURES
● Efficiency up to 80%
● Low Leakage Current
● Single & Dual Outputs
● 5600VDC I/O Isolation
● Short Circuit Protection
● MTBF > 700,000 Hours
● Low Isolation Capacitance
● 2:1 Wide Input Voltage Range
● Operating Temperature: -40°C to +55°C
● Medical Safety Approval UL/cUL 60601-1
● All I/O Clearance and Creepage Distance 2.0mm min.
● Complies with EN55022 Class A with External Coupling Capacitor Cio=1nF < Class B
DESCRIPTION
The ME series of medical approved (UL/cUL 60601-1) dc/dc converters are specially designed to provide ultra high levels of isolation (5600VDC) in
a 24-pin DIP package. These converters operate over input voltage ranges of 9-18VDC, 18-36VDC, and 36-75VDC. This series also provides
regulated single and dual output voltages of 5, 12, ±12, and ±15VDC. The -40°C to +55°C operating temperature makes this series of dc/dc
converters ideal for data communication equipment, mobile battery driven equipment, distributed power systems, process/machine control
equipment, telecommunications equipment, mixed analog/digital subsystems, computer peripheral systems, and industrial robot systems. These
converters have a power rating of 6W and a typical full load efficiency of 80%. This series complies with EN55022 Class A conducted noise and
radiated emissions with an external coupling capacitor Cio=1nF < Class B. The EN55022 Class A conducted noise compliance minimizes design
time, cost, and eliminates the need for external filter components. All models are over load and short circuit protected.
SPECIFICATIONS: ME Series
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Nom
Max
Unit
INPUT (Vin)
12V nominal input models
9
12
18
Input Voltage Range
24V nominal input models
18
24
36
VDC
48V nominal input models
36
48
75
12V nominal input models
7
8
9
Start Voltage
24V nominal input models
13
15
18
VDC
48V nominal input models
30
33
36
12V nominal input models
8.5
Under Voltage Shutdown
24V nominal input models
16
VDC
48V nominal input models
34
Short Circuit Input Power
All models
3000
mW
12V nominal input models
-0.7
25
Input Surge Voltage (1000ms)
24V nominal input models
-0.7
50
VDC
48V nominal input models
-0.7
100
Input Filter
Pi Filter
Leakage Current
240VAC, 60Hz
2
μA
OUTPUT (Vo)
Output Voltage
See Table
Output Voltage Accuracy
±0.5
±1.0
%
Output Voltage Balance
Dual Output, Balanced Loads
±0.5
±2.0
%
Load Regulation
Io = 25% to 100%
±0.5
±1.0
%
Line Regulation
Vin = min. to max.
±0.3
±0.5
%
Output Power
See Table
Output Current Range
See Table
5V output models
75
100
Ripple & Noise (20MHz)
mVpk-pk
Other output models
100
150
Ripple & Noise (20MHz)
Over Line, Over Load, and Over Temperature
180
mVpk-pk
Ripple & Noise (20MHz)
25
mVrms
Transient Recovery Time
25% load step change
300
500
μs
Transient Response Deviation
25% load step change
±3
±6
%
PROTECTION
Over Load Protection
120
%
Short Circuit Protection
Continuous
12V nominal input models
1200mA slow-blow type
Input Fuse Recommendation
24V nominal input models
600mA slow-blow type
48V nominal input models
300mA slow-blow type
GENERAL
Efficiency
See Table
Switching Frequency
150
KHz
Isolation Voltage Rated
60 seconds
5600
VDC
Isolation Voltage Test
Flash Test for 1 second
6000
VDC
Isolation Resistance
500VDC
1000
MΩ
Isolation Capacitance
100KHz, 1V
7
13
pF
Internal Power Dissipation
2500
mW
Max. Capacitive Load
See Table
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
ME Series
5600VDC I/O Isolation
Single and Dual Outputs
5~6 Watt Medical DC/DC Converter
Rev B
SPECIFICATIONS (CONTINUED)
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Nom
Max
Unit
ENVIRONMENTAL
Operating Temperature (Ambient)
Ambient
-40
+55
°C
Operating Temperature (Case)
Case
-40
+95
°C
Storage Temperature
-40
+125
°C
Over Temperature Protection
Case Temperature, automatic
107
112
117
°C
Lead Temperature
1.5mm from case for 10 seconds
260
°C
Humidity
95
%
Cooling
Free air convection
Temperature Coefficient
±0.02
±0.05
%/°C
MTBF
MIL-HDBK-217F @ 25°C, Ground Benign
700,000 hours
Conducted EMI
EN55022 Class A
PHYSICAL
Weight
18 grams
1.25 x 0.8 x 0.47 inches
Dimensions (L x W x H)
31.8 x 20.3 x 12.0 mm
Case Material
non-conductive black plastic
Flammability
UL94V-0
MODEL SELECTION GUIDES
SINGLE OUTPUT MODELS
Model Number
Input Voltage
ME12S5-1000W
12 VDC
(9 ~ 18 VDC)
ME12S12-500W
ME24S5-1000W
ME24S12-500W
ME48S5-1000W
ME48S12-500W
24 VDC
(18 ~ 36 VDC)
48 VDC
(36 ~ 75 VDC)
Output
Voltage
Output Current
5 VDC
200 mA 1000 mA
Min
Max
Input Current
Reflected
Output Power
No Load Max Load Ripple Current
570 mA
30 mA
12 VDC
100 mA
500 mA
5 VDC
200 mA 1000 mA
Efficiency
Maximum
(Typ)
Capacitive Load
5W
75%
1000 μF
6W
78%
470 μF
5W
77%
1000 μF
60 mA
641 mA
278 mA
20 mA
30 mA
12 VDC
100 mA
500 mA
313 mA
6W
80%
470 μF
5 VDC
200 mA 1000 mA
139 mA
5W
77%
1000 μF
12 VDC
100 mA
6W
80%
470 μF
10 mA
500 mA
15 mA
156 mA
DUAL OUTPUT MODELS
Model Number
ME12D12-250W
ME12D15-200W
ME24D12-250W
ME24D15-200W
ME48D12-250W
ME48D15-200W
Input Voltage
12 VDC
(9 ~ 18 VDC)
Output
Voltage
Output Current
Min
Max
±12 VDC ±50 mA ±250 mA
Input Current
Reflected
Output Power
No Load Max Load Ripple Current
30 mA
641 mA
Efficiency
Maximum
(Typ)
Capacitive Load
6W
78%
220 μF #
100mA
±15 VDC ±40 mA ±200 mA
30 mA
641 mA
6W
78%
220 μF #
±12 VDC ±50 mA ±250 mA
24 VDC
(18 ~ 36 VDC) ±15 VDC ±40 mA ±200 mA
20 mA
313 mA
6W
80%
220 μF #
20 mA
313 mA
6W
80%
220 μF #
±12 VDC ±50 mA ±250 mA
48 VDC
(36 ~ 75 VDC) ±15 VDC ±40 mA ±200 mA
10 mA
156 mA
6W
80%
220 μF #
10 mA
156 mA
6W
80%
220 μF #
50mA
25mA
# For each output
NOTES
1. Transient recovery time is measured to within 1% error band for a step change in output load of 75% to 100%.
2. The ME series requires a minimum output loading to maintain specified regulations. Operation under no-load conditions will not damage
these devices, however they may not meet all listed specifications.
3. All DC/DC converters should be externally fused at the front end for protection.
4. Other input and output voltages may be available, please contact factory.
*Due to advances in technology, specifications subject to change without notice.
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
ME Series
5600VDC I/O Isolation
Single and Dual Outputs
5~6 Watt Medical DC/DC Converter
Rev B
BLOCK DIAGRAMS
Single Output
Dual Output
INPUT VOLTAGE TRANSIENT RATING
150
140
130
48VDC Input Models
120
110
Vin ( VDC )
100
90
80
24VDC Input Models
70
60
50
12VDC Input Models
40
30
20
10
0
10uS
100uS
1mS
10mS
100mS
DERATING CURVES
Derating Curve (5V Output Models Only)
Derating Curve (For All Other Models)
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
ME Series
5600VDC I/O Isolation
Single and Dual Outputs
5~6 Watt Medical DC/DC Converter
Rev B
MECHANICAL DRAWING
Unit: inches (mm)
PIN CONNECTIONS
Pin
Single Output
Dual Output
1
+Vin
+Vin
11
No Pin
Common
12
-Vout
No Pin
13
+Vout
-Vout
15
No Pin
+Vout
23
-Vin
-Vin
24
-Vin
-Vin
1. Tolerance: X.X±0.25 (X.XX±0.01)
X.XX±0.13 (X.XXX±0.005)
2. Pin: ±0.05 (±0.002)
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
ME Series
5600VDC I/O Isolation
Single and Dual Outputs
5~6 Watt Medical DC/DC Converter
Rev B
DESIGN & FEATURE CONSIDERATONS
Input Source Impedance
The power module should be connected to a low ac-impedance input source. Highly inductive source impedances can affect
the stability of the power module. In applications where power is supplied over long lines and output loading is high, it may
be necessary to use a capacitor at the input to ensure startup. A capacitor mounted close to the power module helps ensure
stability of the unit. It is recommended to use a good quality low Equivalent Series Resistance (ESR < 1.0Ω at 100KHz)
capacitor of 10μF for the 12V input models, 4.7μF for the 24V input models, and a 2.2μF for the 48V input models.
+
+Vin
+
DC Power
Source
+Out
DC / DC
Converter
Load
Cin
-
-Vin
-Out
Maximum Capacitive Load
The ME Series has a limit of connected capacitance at the output. The power module may operate in current limiting mode
during start-up, which affects the ramp-up and the startup time. Connect capacitors at the point of load for best performance.
The maximum capacitive load can be found in the “Output Voltage / Current Rating Chart” on page 2.
TEST CONFIGURATIONS
Input Reflected-Ripple Current Test Setup
Input reflected-ripple current is measured with an inductor Lin (4.7μH) and Cin (220μF, ESR < 1.0Ω at 100KHz) to simulate
source impedance.
To Oscilloscope
+
+
Battery
+Vin
Lin
Cin
Current
Probe
+Out
DC / DC
Converter
-Vin
Load
-Out
Capacitor Cin offsets possible battery impedance.
Current ripple is measured at the input terminals of the module. Measurement bandwidth is 0 ~ 500KHz.
Peak-to-Peak Output Noise Measurement Test
Use a Cout 0.47μF ceramic capacitor.
Scope measurement should be made by using a BNC socket; measurement bandwidth is 0 ~ 20MHz. Position the load
between 50mm and 75mm from the DC/DC Converter.
+Vin
+Out
Single Output
DC / DC
Converter
-Vin
Copper Strip
Cout
-Out
+Vin
Resistive
Scope
Load
+Out
Copper Strip
Cout
Dual Output
DC / DC
Converter
Com.
-Vin
-Out
Resistive
Load
Cout
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Scope
Scope
ME Series
5600VDC I/O Isolation
Single and Dual Outputs
5~6 Watt Medical DC/DC Converter
Rev B
Output Ripple Reduction
A good quality low ESR capacitor placed as close as possible across the load will give the best ripple and noise performance.
To reduce output ripple it is recommended to use a 3.3μF capacitor at the output.
+
+Vin
Single Output
DC / DC
Converter
DC Power
Source
-
+
+Out
-Vin
+Vin
Dual Output
DC / DC
Com.
Converter
DC Power
Source
Cout
Load
-
-Out
+Out
-Vin
-Out
Cout
Load
Thermal Considerations
Many conditions affect the thermal performance of the power module, such as orientation, airflow over the module, and board
spacing. To avoid exceeding the maximum temperature rating of the components inside the power module, the case
temperature must be kept below 95°C. The derating curves are determined from measurements obtained in an experimental
apparatus.
Position of air velocity
probe and thermocouple
50mm / 2in
15mm / 0.6in
Air Flow
DUT
Electromagnetic Emission EN 55022 < A
Conducted and radiated emissions < A with external coupling capacitor Cio=1nF < B
COMPANY INFORMATION:
Wall Industries, Inc. has created custom and modified units for over 40 years. Our in-house research and development
engineers will provide a solution that exceeds your performance requirements on time and on budget. Our ISO9001-2000
certification is just one example of our commitment to producing a high quality, well documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working
closely with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to
provide ongoing support for your project above and beyond the design and production phases. Give us a call today to
discuss your future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
(603)778-2300
(888)587-9255
(603)778-9797
[email protected]
www.wallindustries.com
5 Watson Brook Rd.
Exeter, NH 03833
Wall Industries, Inc. 37 Industrial Drive Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Similar pages