Keysight N9912A N9912a fieldfox rf handheld analyzer Datasheet

Keysight Technologies
N9912A FieldFox RF
Handheld Analyzer
4/6 GHz
Data Sheet
Documentation Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED,
WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW,
KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE
USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL
CONTROL.
U.S. Government Rights
U.S. Government Rights. The Software is “commercial computer software,” as defined by Federal Acquisition
Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement
(“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which
the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government
customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a
copy of which can be found at http://www.keysight.com/find/sweula. The license set forth in the EULA represents the
exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA
and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical
information related to commercial computer software or commercial computer software documentation that is not
customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of
these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose
commercial computer software or commercial computer software documentation. No additional government
requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or
licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the
DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to
update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101,
pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited
Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
2
U.S. Government Rights .................................................................................................................. 2
Definitions ....................................................................................................................................... 4
Cable and Antenna Analyzer........................................................................................................... 5
Cable and Antenna Analyzer (continued) ....................................................................................... 7
Network Analyzer .......................................................................................................................... 10
Time Domain ................................................................................................................................. 14
Spectrum Analyzer ........................................................................................................................ 15
Tracking Generator or Independent Signal Source ...................................................................... 22
Preamplifier ................................................................................................................................... 23
Interference Analyzer .................................................................................................................... 23
Channel Scanner ........................................................................................................................... 24
Channel Power Meter ................................................................................................................... 24
External USB Power Sensor Support............................................................................................ 24
Power Sensor Measurements vs. Frequency................................................................................ 25
Pulse Measurements ..................................................................................................................... 27
Remote Control Capability............................................................................................................ 27
General Information ...................................................................................................................... 28
FieldFox Data Link Software ......................................................................................................... 31
Supported Cal Kits ........................................................................................................................ 32
3
Definitions
All specifications and characteristics apply over a 25 ±5 °C range (unless otherwise stated) and 90 minutes
after the instrument has been turned on.
Specification (spec.)
Warranted performance. Specifications include guardbands to account for the expected statistical
performance distribution, measurement uncertainties, and changes in performance due to environmental
conditions. The following conditions must be met:
-
FieldFox has been turned on at least 90 minutes
-
FieldFox is within its calibration cycle
-
Storage or operation at 25 ±5 °C range (unless otherwise stated)
Typical (typ.)
Expected performance of an average unit over a 20 °C to 30 °C temperature range after being at ambient
temperature for two hours, unless otherwise indicated; does not include guardbands. It is not covered by the
product warranty. The FieldFox must be within its calibration cycle.
Nominal (nom.)
A general, descriptive term or design parameter. It is not tested, and not covered by the product warranty.
Calibration
The process of measuring known standards to characterize an instruments systematic (repeatable) errors.
Corrected (residual)
Indicates performance after error correction (calibration). It is determined by the quality of calibration
standards and how well "known" they are, plus system repeatability, stability, and noise.
Uncorrected (raw)
Indicates instrument performance without error correction. The uncorrected performance affects the stability of
a calibration.
4
Cable and Antenna Analyzer
Specification
Typical
10 minute
warm up
Supplemental Information
90 minute
warm up
Frequency Range
Option 104
2 MHz to 4 GHz
Option 106
2 MHz to 6 GHz
Frequency Reference (0 to 55 ºC)
Accuracy
±2 ppm
±2 ppm
Aging Rate
±1 ppm/yr
±1 ppm/yr
Temperature Stability
±1 ppm
±1 ppm
Frequency Resolution
2 MHz to 1.6 GHz
2.5 kHz
> 1.6 to 3.2 GHz
5 kHz
> 3.2 to 6 GHz
10 kHz
Resolution (Number of data points)
101, 201, 401, 601, 801, 1001, 1601, 4001, 10001
Custom number of points can be set using SCPI
Measurement Speed (Sweep time)
Return Loss 1, 1.75 to 3.85 GHz, 1001 points, Cal ON
0.4 ms/point (nominal)
DTF , 0 to 500 ft, 601 points, Cal ON
0.5 ms/point (nominal)
2
Output Power (RF Out Port)
High
2 MHz to 4 GHz
< +8 dBm, +6 dBm (nominal)
> 4 to 6 GHz
< +7 dBm, +2 dBm (nominal)
Low (Typically 31 dB below high power)
2 MHz to 4 GHz
< –23 dBm, –25 dBm (nominal)
> 4 to 6 GHz
< –24 dBm, –25 dBm (nominal)
Immunity to interference signals
+16 dBm (nominal)
1.
1.5 ms/pt; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500
2.
2.4 ms/pt; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500
5
Cable and Antenna Analyzer (continued)
Specification
Typical
10 minute warm up
90 minute warm up
Directivity
Corrected with OSL calibration 1
Corrected with QuickCal (Option 111)
> 42 dB
> 42 dB
2
≥ 42 dB
Raw
2 MHz to 3.5 GHz
> 20 dB
> 3.5 to 6 GHz
> 14 dB
Source Match
Corrected with OSL calibration 1
Corrected with QuickCal (Option 111)
> 36 dB
> 36 dB
2
≥ 35 dB
Raw
2 MHz to 3 GHz
> 25 dB
> 3 to 6 GHz
> 16 dB
Reflection Tracking
Corrected with OSL calibration 1
Corrected with QuickCal (Option 111)
± 0.06 dB
± 0.06 dB
2
± 0.15 dB
Reflection Dynamic Range
Reflection (RF Out port) (High power out)
2 MHz to 4 GHz
60 dB
> 4 to 6 GHz
55 dB
Maximum Measurable Cable Loss Using 1–Port CAT Measurement Model 3
Refl Dyn Range /2
1.
Using recommended calibration kits
2.
QuickCal is performed with the connect LOAD step
3.
Higher cable losses can be measured using transmission or S21 measurements. Cable losses measured in transmission mode limited
by transmission dynamic range
6
Cable and Antenna Analyzer (continued)
Specification
Typical
10 minute warm up
90 minute warm up
Transmission Dynamic Range (Option 110), 300 Hz IF Bandwidth
2 MHz to 2 GHz
72 dB
> 2 to 3 GHz
67 dB
> 3 to 5 GHz
58 dB
> 5 to 6 GHz
49 dB
Return Loss
Display Range
0 to 100 dB
Resolution
0.01 dB
VSWR
Display Range
1 to 500
Resolution
0.01
Cable Loss
Display Range
0 to 100 dB
Resolution
0.01 dB
Distance–to–Fault
Horizontal Range
Range = [(number of points – 1) / frequency
span * 2] * velocity factor * speed of light
Number of points auto coupled according to
start and stop distance entered
Horizontal Resolution
Resolution = Range / (number of points – 1)
Number of points settable by user
Bandpass Mode
Window Types
Maximum, medium, and minimum windows
7
Cable and Antenna Analyzer (continued)
Figure 1: CAT Mode, Type–N Calibration Kit – Magnitude (Specification)
8
Cable and Antenna Analyzer (continued)
Figure 2: CAT Mode, QuickCal – Magnitude (Typical)
Figure 3: CAT Mode, Preset Cal – Magnitude (Typical)
9
Network Analyzer
The following CAT mode performance parameters apply to NA mode: frequency accuracy, frequency
resolution, output power, directivity, source match, reflection tracking, and reflection and transmission
dynamic range. NA mode performance that is in addition to CAT mode is listed in the table below.
Specification
Supplemental Information
Frequency Range
Option 104
2 MHz to 4 GHz
Option 106
2 MHz to 6 GHz
Measurement Speed (Sweep time)
S11: 1.75 GHz – 3.85 GHz,1001 Points, Cal ON 1
0.4 ms/point (nominal)
S21: 1.78 GHz – 2.06 GHz, 201 Points, Cal ON
1.8 ms/point (nominal)
2
S11 Phase Uncertainty 3
See Figure 5 on page 12
Display Range
–180º to +180º
System Impedance
50Ω (nominal)
75Ω with appropriate adapter and Cal Kit
1.
1.5 ms/pt; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500
2.
1.9 ms/pt; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500
3.
Using recommended calibration kits
10
Network Analyzer (continued)
Information
Measurements
S11 magnitude and phase
S21 magnitude (Option 110)
A receiver magnitude
R receiver magnitude
Formats
Log magnitude, Linear magnitude
Available ONLY for S11: VSWR, Phase, Smith Chart, Polar, Group delay, Unwrapped phase
Resolution
(Number of data points)
101, 201, 401, 601, 801, 1001, 1601, 4001, 10001
Custom number of points can be set using SCPI
Averaging
Sweep and point averaging; 2 to 999 points.
Number of traces
Four traces available. Tr1, Tr2, Tr3, Tr4
Data markers
Each trace has six independent markers that can be displayed simultaneously. Delta
markers are available for each marker.
Marker formats
Default marker format is the trace format. In Smith chart or polar format, [Real +Imag] or
[Mag and Phase] formats are also available.
Marker functions
Peak, Next Peak, Peak Left, Peak Right, Mkr→ Center, Min Search, Peak Excursion, Peak
Threshold, Target, Bandwidth, Tracking
Display formats
Single-trace
Dual-trace overlay (both traces on one graticule)
Dual-trace split (each trace on separate graticules)
Three-trace overlay (all three traces on one graticule)
Three-trace split (each trace on separate graticules)
Quad-trace split (each trace on separate graticules)
Display data
Display data, memory, data and memory, or data math
Trace math
Vector division or subtraction of current linear measurement values and memory data.
Scale
Autoscale, scale, reference level, reference position
Autoscale: Automatically selects scale resolution and reference value to center the trace.
Autoscale all scales all visible traces.
Title
Add custom titles to the display.
Limit lines
Define test limit lines that appear on the display for go/no go testing. Lines may be any
combination of horizontal, sloping lines, or discrete data points. Each trace can have its
own limit line.
Limit Lines can be Fixed, Relative to center frequency and reference level, and can be built
from existing traces.
11
Network Analyzer (continued)
Figure 4: NA Mode, Type–N Calibration Kit – Magnitude (Specification)
Figure 5: NA Mode, Type–N Calibration Kit – Phase (Specification)
12
Network Analyzer (continued)
Figure 6: NA Mode, QuickCal – Magnitude (Typical)
Figure 7: NA Mode, Preset Cal – Magnitude (Typical)
13
Network Analyzer (continued)
Figure 8: NA Mode, Preset Cal – Phase (Typical)
Time Domain
Using time domain, data from transmission or reflection measurements in the frequency domain are converted to the time
domain. The time-domain response shows the measured parameter value versus time.
Time stimulus modes
Low-pass step
Similar to a traditional time domain reflectometer (TDR) stimulus waveform, Low-pass step
is used to measure low-pass devices. The frequency-domain data should extend from DC
(extrapolated value) to a higher value.
Low-pass impulse
Also used to measure low-pass devices
Bandpass impulse
Stimulates a pulsed RF signal and is used to measure the time-domain response of bandlimited devices
Windowing
The windowing function is used to filter the frequency-domain data and thereby reduce overshoot and ringing in the timedomain response.
Gating
The gating function is used to selectively remove reflection or transmission time-domain responses. When converted back to
the frequency domain, the effects of the responses outside the gate are removed.
14
Spectrum Analyzer
Specification
Supplemental Information
Option 230
100 kHz to 4 GHz
Usable to 5 kHz 1
Option 231
100 kHz to 6 GHz
Usable to 5 kHz 1 (Tunable to 6.1 GHz)
Frequency Range
Frequency Reference (–10 to 55 °C)
Accuracy
±2 ppm
Aging Rate
± 1 ppm/yr
Temperature Stability
± 1 ppm
Frequency Readout Accuracy (start, stop, center, marker)
± (readout frequency x frequency
reference accuracy + RBW centering +
0.5 x horizontal resolution)
Horizontal resolution = span/(trace points – 1)
RBW centering :
5% x RBW, FFT mode (nominal)
16% x RBW, Step mode (nominal)
Frequency Span
Range
0 Hz (zero span), 10 Hz to max freq
Accuracy
±(2 x RBW centering + horizontal
resolution)
Resolution
1 Hz
±(2 x RBW centering +2 x horizontal resolution)
for detector = Normal
Sweep Time, Span = 0 Hz
Range
Minimum
1.0 us
Maximum
RBW = 2 MHz
2.18 ms
RBW = 1 MHz
3.28 ms
RBW = 300 kHz
5.46 ms
RBW = 100 kHz
16.38 ms
RBW = 30 kHz
54.60 ms
RBW = 10 kHz
163.84 ms
RBW = 3 kHz
546.00 ms
RBW = 1 kHz
1.64 s
RBW = 300 Hz
2.54 s
Resolution
100.0 ns
Readout
Entered value representing trace horizontal scale range.
1.
With signal at center frequency.
15
Spectrum Analyzer (continued)
Description
Specification
Supplemental Information
Sweep Acquisition, Span > 0 Hz
Range
1 to 5000. Number of data acquisitions per
trace point. Value is normalized to the
minimum required to achieve amplitude
accuracy with CW signals.
Resolution
1
Readout
Measured value representing time required to tune receiver, acquire data, and process
trace.
Auto coupled. For pulsed RF signals,
manually increase the sweep acquisition
value to maximize the pulse spectrum
envelope.
Trigger
Type
Free Run, Video, External
Slope
Positive, Negative edge
Delay
Range: 0 to 10 sec
Range
Resolution: 100 nsec
Auto Trigger
Forces a periodic acquisition in the absence of a trigger event
Range: 0 sec (OFF) to 10 sec
Time Gating
Gate Method
Triggered FFT
Gate Delay Range
Same as Trigger Delay
Trace Update
Span = 20 MHz, RBW = 3 kHz 1
5.9 updates/s (nominal)
Span = 100 MHz, RBW auto coupled 2
16.7 updates/s (nominal)
Span = 6 GHz, RBW auto coupled 3
1.7 update/s (nominal)
Trace Points
101, 201, 401, 601, 801, 1001 (Defaults to 401)
1.
1.5 updates/s; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500.
2.
7 updates/s; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500.
3.
1 update/s; applicable for N9912A with serial number prefix <MY5607/SG5607/US5607 and N9912A not upgraded with Option
N9910HU-500.
16
Spectrum Analyzer (continued)
Description
Specification
Supplemental Information
Resolution Bandwidth (RBW)
Range (–3 dB bandwidth)
Zero Span
300 Hz to 1 MHz in 1, 3, 10 sequence; 2
MHz
Non–Zero Span
10 Hz to 300 kHz in 1/1.5/2/3/5/7.5/10
sequence; 1 MHz, 2 MHz
Step keys change RBW in 1, 3, 10 sequence
Bandwidth Accuracy
1 kHz to 1 MHz
± 5% (nominal)
10 Hz to 100 kHz non–zero span
± 1% (nominal)
2 MHz
± 10% (nominal)
300 Hz zero span
± 10% (nominal)
Selectivity (-60 dB/ -3 dB)
4:1 (nominal)
Video Bandwidth (VBW)
Range
1 Hz to 2 MHz in 1/1.5/2/3/5/7/10
sequence
Stability
Specification
VBW ≥ RBW in zero span
Typical
Noise Sidebands, CF = 1 GHz
10 minute warm up
< – 85 dBc/Hz
– 88 dBc/Hz
– 88 dBc/Hz
– 89 dBc/Hz
– 89 dBc/Hz
– 95 dBc/Hz
– 95 dBc/Hz
– 115 dBc/Hz
– 115 dBc/Hz
Amplitude Range
Measurement Range
Displayed average noise level (DANL) to + 20 dBm
Input Attenuator Range
0 to 31 dB (1 dB steps)
Maximum Safe Input Level
Average Continuous Power
+27 dBm (0.5 W)
DC
±50 VDC
17
90 minute warm up
Spectrum Analyzer (continued)
Displayed Average Noise Level (DANL) 10 Hz RBW, 10 Hz VBW, 50 ohm input termination, 0 dB attenuation, average detector
Specification
Typical
10 minute warm up
90 minute warm up
–10 to 55 °C
20 to 30 °C
20 to 30 °C
-
-
-
–130 dBm
> 2.4 to 5.0 GHz
-
-
-
–125 dBm
> 5.0 to 6.0 GHz
-
-
-
–119 dBm
Preamp on
20 to 30 °C
–10 to 55 °C
20 to 30 °C
20 to 30 °C
10 MHz to 2.4 GHz
< –143 dBm
< –141 dBm
-
–148 dBm
> 2.4 to 5.0 GHz
< –140 dBm
< –138 dBm
-
–145 dBm
> 5.0 to 6.0 GHz
< –132 dBm
< –130 dBm
-
–138 dBm
Preamp off
20 to 30 °C
10 MHz to 2.4 GHz
Display Range
Log Scale
Ten divisions displayed; 0.1 to 1.0 dB/division in 0.1 dB steps and 1 to 20 dB/division in 1
dB steps
Traces
Detectors
Normal, Positive Peak, Negative Peak, Sample, Average
States
Clear/Write, Max Hold, Min Hold, Average, View, Blank
Number of averages: 1 to 10,000
Number of traces
4
Reference Level
Range
–170 to +30 dBm
Resolution
0.1 dB
Accuracy
0 dB
18
Spectrum Analyzer (continued)
Absolute Amplitude Accuracy at 50 MHz
Peak detector, 10 dB attenuation, preamplifier off, RBW < 2 MHz, input signal –5 dBm to –50 dBm, all settings auto–coupled
Specification
Typical
10 minute warm up
90 minute warm up
20 to 30 ºC
± 0.8 dB
± 0.8 dB
± 0.4 dB
–10 to 55 ºC
± 1.1 dB
-
± 0.8 dB
Frequency Response
Relative to 50 MHz, Peak detector, 10 dB attenuation, preamplifier off, RBW = 30 kHz, input signal 0 dBm to –50 dBm, all
settings auto–coupled
Preamp off
20 to 30 ºC
2 to 10 MHz
± 1.1 dB
± 1.0 dB
± 0.5 dB
> 10 MHz to 3 GHz
± 0.9 dB
± 0.6 dB
± 0.3 dB
> 3 to 5 GHz
± 1.3 dB
± 1.1 dB
± 0.5 dB
> 5 to 6 GHz
± 1.5 dB
± 1.5 dB
± 0.5 dB
± 2.0 dB
-
± 1.0 dB
> 10 MHz to 3 GHz
± 1.5 dB
-
± 0.6 dB
> 3 to 5 GHz
± 2.0 dB
-
± 1.1 dB
> 5 to 6 GHz
± 2.6 dB
-
± 1.5 dB
-
-
± 0.7 dB
> 10 MHz to 3 GHz
-
-
± 0.5 dB
> 3 to 5 GHz
-
-
± 0.7 dB
> 5 to 6 GHz
-
-
± 0.7 dB
-
-
± 1.2 dB
> 10 MHz to 3 GHz
-
-
± 0.8 dB
> 3 to 5 GHz
-
-
± 1.3 dB
> 5 to 6 GHz
-
-
± 1.7 dB
–10 to 55 ºC
2 to 10 MHz
Preamp on
20 to 30 ºC
2 to 10 MHz
–10 to 55 ºC
2 to 10 MHz
19
Spectrum Analyzer (continued)
Specification
Typical
10 minute
warm up
Supplemental Information
90 minute
warm up
Resolution Bandwidth Switching Uncertainty
RBW < 2 MHz
0.0 dB
0.7 dB peak–to–peak 1
Total Absolute Amplitude Accuracy 2
Peak detector, 10 dB attenuation, preamplifier off, RBW < 2 MHz, input signal 0 dBm to –50 dBm, all settings auto coupled
Absolute Amplitude at 50 MHz + Frequency Response 3
20 to 30 ºC:
2 to 10 MHz
± 1.8 dB
± 1.28 dB
± 0.60 dB
> 10 MHz to 3 GHz
± 1.5 dB
± 1.0 dB
± 0.50 dB
> 3 to 5 GHz
± 1.9 dB
± 1.36 dB
± 0.60 dB
> 5 to 6 GHz
± 2.1 dB
± 1.7 dB
± 0.60 dB
RF Input VSWR
At all attenuation settings
1.5:1 (nominal)
Second harmonic distortion (SHI)
–30 dBm signal at input mixer 4
2 MHz to 1.35 GHz
< –70 dBc, +40 dBm (nominal)
1.35 to 3 GHz
< –80 dBc, +50 dBm (nominal)
Third order intermodulation distortion (TOI)
Two –30 dBm tones at input mixer
< –96 dBc, +18 dBm (nominal)
1.
For signals not at center frequency
2.
With signal at center frequency
3.
The specification for Total Absolute Amplitude Accuracy is less than the sum of the Absolute Amplitude Accuracy and Frequency
Response specifications because redundant uncertainty is removed
4.
Mixer level = RF input level – input attenuation
20
Spectrum Analyzer (continued)
Residual Responses
Input terminated, 0 dB attenuation, preamplifier off, RBW ≤ 1 kHz, VBW auto coupled
20 MHz to 3 GHz
–90 dBm (nominal)
> 3 to 6 GHz
–85 dBm (nominal)
Spurious Responses
Input Mixer level –30 dBm
RFsig = RFtune + 417 MHz
–70 dBc (nominal)
RFsig = RFtune + 1.716 GHz
–80 dBc (nominal)
Input Mixer level –10 dBm; First IF Image Response
Rfsig = Rftune – 2 x 0.8346 GHz
(for Rftune 5.7 to 6 GHz)
–50 dBc (nominal)
Sidebands
–80 dBc (nominal)
–60 dBc (nominal) when battery charging, 260 kHz offset
Figure 10
21
Spectrum Analyzer (continued)
Figure 11
Tracking Generator or Independent Signal Source
The independent source or tracking generator is included with either spectrum analyzer option. The source can be used in
continuous wave (CW) or stimulus/response (S/R) mode. In CW mode, the source frequency is independent of the receiver
frequency. The source can be tuned to a frequency that is different from the receiver. In stimulus/response mode, the
source operates the same as a traditional tracking generator - the receiver tracks the source.
Frequency range
2 MHz to 4 GHz (Option 230) or 2 MHz to 6 GHz (Option 231)
Amplitude
High power
2 MHz to 4 GHz < +8 dBm, +6 dBm (nominal)
>4 GHz to 6 GHz <+7 dBm, +2 dBm (nominal)
Low power
2 MHz to 4 GHz <-23 dBm, -25 dBm (nominal)
>4 GHz to 6 GHz < -24 dBm, -29 dBm (nominal)
Attenuation
0 to 31 dB
Functions
Continuous wave, stimulus/response
22
AM/FM Tune and Listen
Description
Audio demodulation types
AM, FM Narrow, FM Wide
Audio Bandwidth
16 kHz
Receiver IF Bandwidth
AM
35 kHz
FM Narrow
12 kHz
FM Wide
150 kHz
Listen Time Range
0 to 100 seconds
Audio Signal Strength Indicator
Audio Signal Strength Indicator helps locate signals. The tone and frequency of the beep varies with signal strength.
Radio Standards
With a Radio Standard applied, pre-defined frequency bands, channel numbers or Uplink / Downlink selections can be used
instead of manual frequency entry. The pre-defined FieldFox Radio Standards include bands such as W-CDMA, LTE, and GSM.
Custom Radio Standards can also be defined, imported, and applied to the FieldFox.
FieldFox Power Suite Measurement types
Channel Power, Occupied Bandwidth, and Adjacent Channel Power Ratio
Preamplifier
Specification
Typical
10 minute warm up
Frequency Range
100 kHz to 4/6 GHz
Gain
22 dB
Interference Analyzer
Description
Spectrogram display
Overlay, full screen, top, or bottom with active trace
Waterfall angle
Moderate, steep, gradual, wide angle
Markers
Time, delta time
Trace playback and recording
Record all spectrum analyzer measurements
Store data internally/USB/SD card
Playback recorded data using FieldFox
Frequency mask trigger allows recording to occur upon trigger
23
Channel Scanner
Description
Scan Mode
Range or custom list
Display Type
Bar chart vertical, bar chart horizontal, channel power, strip chart, chart overlay, scan &
listen
Data logging mode
Time with geo tagging
Trace playback and recording
Record channel power measurement
Store data internally or USB or SD card in .csv or .kml format
Playback recorded data using FieldFox
Data in .kml format can be exported to Google Earth
Channel Power Meter
Channel power meter is a built-in power measurement that application does not require an external power sensor. Set the
center frequency and channel bandwidth. The results are shown on a large analog display.
Specification
Frequency range
Typical
100 kHz to 4/6 GHz
Power accuracy
2 to 10 MHz
± 1.8 dB
± 0.6 dB
> 10 MHz to 3 GHz
± 1.5 dB
± 0.5 dB
> 3 to 5 GHz
± 1.9 dB
± 0.6 dB
> 5 to 6 GHz
± 2.1 dB
± 0.6 dB
External USB Power Sensor Support
The external USB power sensor option supports various Keysight USB Power Sensors. Supported power sensors:
www.keysight.com/find/fieldfoxsupport
24
Power Sensor Measurements vs. Frequency
This feature allows the FieldFox source frequency to be set independently from the power sensor (receiver) frequency. With
frequency-offset using power sensor (FOPS), the frequency of both the source and receiver are swept, and the two track
each other. The offset frequency can be negative, zero, or positive.
FOPS can be used to characterize the scalar transmission response of devices such as mixers and converters. This
frequency-offset capability is necessary for conversion loss/gain measurements on frequency-translating devices, since by
definition, the input and output frequencies of the DUT are different. The FieldFox source stimulates the DUT and the power
sensor is used as the measurement receiver.
Since power sensors are inherently broadband devices (not frequency-selective), the user should ensure that only the signal
of interest is present at the power sensor input and that all others signals are filtered appropriately.
Setup parameter
Source frequency
Center/span or start/stop. Range determined by FieldFox
Receiver frequency
Range determined by power sensor range
Frequency offset
0, > 0, < 0
Frequency step size
30 kHz minimum
Number of points
2 to 1601
Combination of number of points and frequency step size limited by span.
Dwell time/point:
0 to 1.0 sec
Source frequency span must be equal to receiver frequency span.
Receiver sweep direction: forward (default setting) or reverse.
For some DUTs, the output frequency may sweep in a reverse direction, as compared to the source frequency. The basic
relationships between the source, receiver and offset frequencies are shown in the table below. The FieldFox analyzer
includes an offset calculator that ensures a fast measurement setup.
Src sweep direction
Rx sweep direction
Frequency calculations
Forward f2rc > f1src
Forward f2rx > f1rx
Receiver frequency = Source frequency ± Offset
Forward f2src > f1src
Reverse f2rx < f1rx
Receiver frequency = Offset – Source frequency Offset > Source
frequency
Description
Measurements
Source power, gain/loss and receiver (Rx) power
Gain = Rx power / source power (memory). Source power (memory) is measured during
setup.
Output power
Refer to the test port output power typical data on page 5
Dynamic range
The dynamic range with FOPS is dependent on FieldFox’s output power and the power
sensor’s dynamic range. Supported USB power sensors:
www.keysight.com/find/fieldfoxsupport
25
Power Sensor Measurements vs. Frequency (continued)
Description
Measurements
Source power, gain/loss and receiver (Rx) power
Gain = Rx power / source power (memory). Source power (memory) is measured during
setup.
Output power
Refer to the test port output power typical data on page xx
Dynamic range
The dynamic range with FOPS is dependent on FieldFox’s output power and the power
sensor’s dynamic range. Supported USB power sensors:
www.keysight.com/find/fieldfoxsupport
The graph below shows a filter measurement using two different power sensors, the U2002A (-60 to +20 dBm) and the
U2021XA (-45 to +20 dBm).
While a filter is not commonly measured using FOPS, it is a useful device for demonstrating dynamic range.
For both measurements, the FieldFox source power was set to 0 dBm, the maximum available in the selected frequency
range of 150 to 450 MHz.
26
Pulse Measurements
The FieldFox pulse measurement option can be used to characterize RF pulses such as those used in radar and electronic
warfare systems. Measurements are made using FieldFox and Keysight's UBS peak power sensors.
Performance specifications such as frequency, dynamic range and minimum pulse width depend on the peak power sensor.
Supported peak power sensors: www.keysight.com/find/fieldfoxsupport
Description
Setup parameters
Frequency, time (center), time/division, gating, triggering, video bandwidth, resolution
averaging
Functions
Average power, peak power, and peak to average ratio, standard and gated
Analog gauge display and digital display, dBm and watts
Relative/absolute measurements, dB or %, minimum and maximum limits
Trace graph for pulse profiling with gating
Rise time, fall time, pulse width, pulse period, pulse repetition frequency
Remote Control Capability
Option 030 adds remote control capability to FieldFox analyzers, so that FieldFox can be controlled via an iOS device. The
FieldFox app, running on the iOS device, combined with Option 030 on the FieldFox analyzer provides full control of the
instrument from a remote location. The app emulates the front panel of FieldFox, so users can press the FieldFox hardkeys or
softkeys using their iPhone or iPad, and make measurements remotely.
-
iOS device requirements
iPhone, iPad, or iPod Touch
iOS of 6.1 or higher
A WiFi or 3G/4G connection
The FieldFox app communicates with FieldFox via a network connection, so both the iOS device and FieldFox need to be on a
network where both devices can reach the other. For example, a company intranet or a site installation using a wireless
router. FieldFox can directly be connected to a LAN cable, or if wired LAN is not available, a user supplied wireless router can
be configured to work with FieldFox.
FieldFox app without Option 030
The FieldFox app can be installed on an iOS device independent of the presence of Option 030 on the analyzer. Without
Option 030, users can view the live display screen of their FieldFox remotely, but cannot control the instrument. With 030
purchased and installed on their FieldFox, users can both view and control their FieldFox.
Option 030 and the FieldFox app are not applicable to Android, BlackBerry, or Windows phone/tablet devices.
27
General Information
Specification
Typical
Calibration Cycle
1 Year
Weight
2.8 kg or 6.2 lb. including
battery
Dimension H x W x D
292 x 188 x 72 mm (11.5” x 7.4” x 2.8”)
Supplemental Information
Environmental
MIL–PRF–28800F class 2
Operating temperature
Storage temperature
Operating humidity
Random vibration
Functional shock
Bench drop
Altitude – Operating
9,144 m (30,000 ft) using battery
Altitude – Non–Operating
15,240 m (50,000 ft)
Altitude – AC to Dc adapter
3, 000 m (9, 840 ft)
IP Class
30
Temperature Range
Operating, AC power
–10 to 55 °C
Operating, battery
–10 to 50 °C
Storage 1
–51 to 71 °C
1.
–10 to 55 °C
With the battery pack removed. The battery packs should be stored in an environment with low humidity. Extended exposure to
temperature above 45 ºC could degrade battery performance and life
28
General Information (continued)
EMC: Complies with the essential requirements of the European EMC Directive as well as current editions of the following
standards (dates and editions are cited in the Declaration of Conformity).
IEC/EN 61326–1
CISPR Pub 11 Group 1, class A
AS/NZS CISPR 11
ICES/NMB–001
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la
norme NMB-001 du Canada.
When subjected to continuously present radiated electromagnetic phenomena, some degradation of performance may occur.
Safety: Complies with the essential requirements of the European EMC Directive as well as current editions of the following
standards (dates and editions are cited in the Declaration of Conformity).
IEC/EN 61010–1
Canada: CSA C22.2 No. 61010–1
USA: UL std no. 61010–1
Power Supply
External DC Input
15 to 19 VDC (40 W maximum when battery charging)
External AC Power Adapter
Efficiency Level IV, 115 VAC
Input
100 to 250 VAC, 50 to 60 Hz, 1.25 – 0.56 A
Output
15 VDC, 4 A
Power Consumption
12 W (On)
Battery
Lithium ion
10.8 V, 4.6 A-h
Operating time
4 hours (typical)
Charge time
A full discharged battery takes about 1.5 hours to recharge to 80%. 4 hours to 100%
Discharge temperature limits 1
–10 to 60 ºC, ≤ 85% RH
Charge temperature limits
0 to 45 ºC, ≤ 85% RH
1
Storage temperature limits
1
–20 to 50 ºC, ≤ 85% RH
The battery packs should be stored in an environment with low humidity. Extended
exposure to temperature above 45 ºC could degrade battery performance and life
1.
Charge and discharge temperatures are internal temperatures of the battery as measured by a sensor embedded in the battery. The
Battery screen displays temperature information. To access the screen, select System, Service, Diagnostics, and Battery
29
General Information (continued)
Test port connectors
RF out port
Type-N, female, 50 Ω (nominal)
Damage level: > + 23 dBm, > ± 50 VDC
RF In port
Type-N, female, 50 Ω (nominal)
Damage level: > + 27 dBm, > ± 50 VDC
LO emission
(0 dB atten, preamp off)
- 65 dBm (nominal)
Display
6.5” transflective color VGA LED–backlit (640 x 480 with anti–glare coating)
Headphone jack connector
3.5mm (1/8 inch) miniature audio jack
USB
USB-A (2 ports)
Hi-speed USB 2.0
Mini USB (1 port) 1
Hi-speed USB 2.0 used for SCPI programming; USBTMC (USB IEEE488)
Keyboard
USB keyboard are supported (user supply own keyboard)
LAN
Connector
RJ-45 (100 base-T only) 10 base-T not supported
Programming
SCPI, using built-in LAN and mini USB interface
Languages
English, Spanish, German, Italian, French, Russian, Japanese, Chinese, Turkish, Korean,
Portuguese
Preset
User preset for both mode preset and complete system preset
Data storage
1.
Internal
Minimum 4 GB (Up to 1000 instrument states and trace)
External
Supports USB 2.0 compatible memory devices and SD/SDHC memory cards
Data types
Trace, trace+state, picture(png), data (csv), S2P
SCPI over USB is only available for N9912A with serial number prefix starting with MY5607/SG5607/US5607 or N9912A analyzer
upgraded with N9910HU-500
30
General Information (continued)
External reference/trigger in
Connector
BNC (f)
Input frequency
10 MHz
Input amplitude range
–5 dBm to +10 dBm, 50 Ω (nominal)
Lock range
±10 ppm of external reference frequency (nominal)
Trigger input
Impedance
10 kΩ (nominal)
Level Range
Rising edge: 1.7 V (nominal)
Falling edge: 1 V (nominal)
FieldFox Data Link Software
FieldFox Data Link software, installed on a PC, provides the following capabilities:
-
Capture of current trace and settings
-
Opening of data files (s1p, s2p, csv, sta, and png) residing
on the instrument
-
Editing cal kit and cable files on the instrument, or creating
new cal kits and cables
-
Transferring files to/from the instrument
-
Annotating plots for documentation purposes
-
Marker, limit line, and format changes on the PC
-
Report generation
-
Printing function
FieldFox Data Link software is available from the following website:
http://www.keysight.com/find/fieldfoxsupport
31
Supported Cal Kits
The following list of calibration kits are loaded in the FieldFox. You can add additional calibration kits to the FieldFox using
FieldFox Data Link Software.
The basic 50-ohm QuickCal does not require cal standards. However, for higher accuracy, perform QuickCal with a load. 75ohm QuickCal does require a 75-ohm load.
Model number
Description
N9910X-800
3-in-1 OSL calibration kit, DC to 6 GHz, Type-N (m) 50 ohm
N9910X-801
3-in-1 OSL calibration kit, DC to 6 GHz, Type-N (f) 50 ohm
N9910X-802
3-in-1 OSL calibration kit, DC to 6 GHz, 7/16 DIN (m)
N9910X-803
3-in-1 OSL calibration kit, DC to 6 GHz, 7/16 DIN (f)
85031B
Economy calibration kit, DC to 6 GHz, 7 mm
85032E
Economy calibration kit, DC to 6 GHz, Type-N, 50-ohm
85032F
Standard calibration kit, DC to 9 GHz, Type-N, 50-ohm
85033E
Standard calibration kit, DC to 9 GHz, 3.5 mm
85036B
Standard calibration kit, DC to 3 GHz, Type-N 75-ohm
85036E
Economy calibration kit, DC to 3 GHz, Type-N 75-ohm
85038A
Standard calibration kit, DC to 7.5 GHz, 7-16
85039B
Economy calibration kit, DC to 3 GHz, Type-F, 75-ohm
85052D
Economy calibration kit, DC to 26.5 GHz, 3.5 mm
85054B
Standard calibration kit, DC to 18 GHz, Type-N, 50-ohm
85054D
Economy calibration kit, DC to 18 GHz, Type-N, 50-ohm
85514A
Calibration kit, 4-in-1, open, short, load and through, DC to 9 GHz, Type-N(m), 50
85515A
Calibration kit, 4-in-1, open, short, load and through, DC to 9 GHz, Type-N(f), 50
85516A
Calibration kit, 4-in-1, open, short, load and through, DC to 3 GHz, Type-N(m), 75 ohm
85517A
Calibration kit, 4-in-1, open, short, load and through, DC to 3 GHz, Type-N(f), 75 ohm
32
This information is subject to
change without notice.
© Keysight Technologies 2016
Print Date: June 2016
N9912-90006
www.keysight.com
Similar pages