Arch MQF240U-48S U bracket medical switching power supply Datasheet

VER : PA_3
update : 2016.07.21
AC-DC Medical Switching Power Supply
MQF240U SERIES
240 Watts
KEY FEATURES
 U Bracket Medical Switching Power Supply
 160 Watt with Free Air Convection
 240 Watt with 10CFM FAN
 4000VAC Input to Output 2MOPP Insulation
 High Efficiency up to 94%
 With P.F.C. Function >0.9
 <0.5W No Load Input Power
 Low Leakage Current under 100uA
rd
 ANSI/AAMI ES60601-1, EN60601-1, & IEC60601-1 3.1 Edition Safety Approvals
 3-Year Product Warranty
(Pending)
ELECTRICAL SPECIFICATIONS
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
Max Output Wattage (with 10CFM FAN)
(W)
Max Output Wattage (Free air Convection)
Voltage
Input
Output
MQF240U-12S
240 W
(Note 4)
47-63 Hz
Current (Full load)
< 3.0 A max. (115 VAC) / < 1.5 A max. (230 VAC)
Inrush Current (<2ms)
< 45 A max. (115 VAC) / < 90 A max. (230 VAC)
Leakage Current
< 0.1mA / 264 VAC (Touch Current)
Power Factor
PF>0.9 at Full Load
Voltage (V.DC.)
12V
Voltage Adj Range (V.DC.)
±5% Output Voltage
Voltage Accuracy
±2%
Current (with 10CFM FAN) (A) max
Current (Free air Convection) (A) max
Line Regulation
±1%
Load Regulation (0-100%)
±1%
Efficiency (at 230VAC)
Hold-up Time (at 115 VAC)
Environment
Physical
24V
48V
20
10
5
13.3
6.66
3.33
Pending
Pending
Pending
150mV
1% Vout
92%
93%
─
(Note 1)
Ripple & Noise max.
Isolation
90-264 VAC
Frequency (Hz)
Maximum Capacitive Load
MQF240U-48S
160 W
(W)
Minimum Load
Protection
MQF240U-24S
(Note 2)
94%
10 ms min.
Over Power Protection
Auto recovery, Hiccup mode
Over Voltage Protection
Zener diode clamp
Short Circuit Protection
Auto recovery, Hiccup mode
Input-Output
4000VAC or 5656VDC
Input-FG
2000VAC or 2828VDC
Output-FG
1500VAC or 2121VDC
Operating Temperature
-30°C…+70°C (with derating)
Storage Temperature
-30°C…+85°C
Temperature Coefficient
±0.05%/°C
Humidity
95% RH
MTBF
>250,000 h @ 25°C (MIL-HDBK-217F, Notice 1)
Vibration
10~500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes.
Dimension (L x W x H)
4.29 x 2.44 x 1.5 Inches
Weight
Pending
Cooling Method
Free convection
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
( 109.0 x 62.0 x 38.1 mm ) Tolerance 0.5 mm
-1-
VER : PA_3
update : 2016.07.21
AC-DC Medical Switching Power Supply
MQF240U SERIES
240 Watts
ELECTRICAL SPECIFICATIONS
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
Safety
Approval
Conducted and radiated EMI
EMC
MQF240U-12S
MQF240U-24S
MQF240U-48S
cUL / UL Standard:
UL 60950-1, CAN/CSA C22.2 No. 60950-1-07
ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10), CAN/CSA-C22.2 No. 60601-1
(2008), 2 x MOPP (Input to Output)
EN55011 class B (Pending)
ESD
EN61000-4-2 air ± 8kV , Contact ± 4Kv
Radiated Immunity
EN61000-4-3 10V/m
Fast Transient
EN61000-4-4 ± 2kV
(Pending)
Surge
EN61000-4-5 ±1kV
(Pending)
Conducted Immunity
EN61000-4-6 10Vrms
PFMF
EN61000-4-8 30A/m
Dips
EN61000-4-11 30% 10ms
Interruption
EN61000-4-11 >95% 5000ms
(Pending)
(Pending)
(Pending)
(Pending)
(Pending)
(Pending)
NOTE
1. Ripple & Noise are measured at 20MHz of bandwidth with 0.1uF & 47uF parallel capacitor.
2. Hold-up Time measured at 90% Vout.
3. Main Vout must be >50% Load, 12V (Aux) / 0.5A.
4. Please check the derating curve for more details.
5. Strongly recommend to conduct this test with DC Voltage. If customer wishes to test with AC Voltage,
please disconnect all Y-Capacitors within Arch power supply.
DERATING
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-2-
VER : PA_3
update : 2016.07.21
AC-DC Medical Switching Power Supply
MQF240U SERIES
ECHANICAL DIMENSION
240 Watts
( Top View )
PIN#
Single
1
AC IN (N)
2
AC IN (L)
3
+DC OUT
4
+DC OUT
5
-DC OUT
6
-DC OUT
7
-AUX OUT
8
+AUX OUT
BLOCK DIAGRAM
Pending
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-3-
Similar pages