MSP-450 450W Single Output Medical Type series ■ Features : Universal AC input / Full range Built-in active PFC function, PF>0.95 High efficiency up to 89.5% Withstand 300VAC surge input for 5 seconds Protections: Short circuit / Overload / Over voltage / Over temperature Built-in constant current limiting circuit Medical safety approved (MOOP level) Built-in cooling Fan ON-OFF control Built-in DC OK signal Built-in remote ON-OFF control Stand by [email protected] Built-in remote sense function No load power consumption<0.6W (Note.7) 5 years warranty SPECIFICATION MODEL OUTPUT MSP-450-3.3 MSP-450-5 MSP-450-7.5 MSP-450-12 MSP-450-15 MSP-450-24 MSP-450-36 MSP-450-48 DC VOLTAGE 3.3V 5V 7.5V 12V 15V 24V 36V 48V RATED CURRENT 90A 90A 60A 37.5A 30A 18.8A 12.5A 9.5A CURRENT RANGE 0 ~ 90A 0 ~ 90A 0 ~ 60A 0 ~ 37.5A 0 ~ 30A 0 ~ 18.8A 0 ~ 12.5A 0 ~ 9.5A RATED POWER 297W 450W 450W 450W 450W 451.2W 450W 456W RIPPLE & NOISE (max.) Note.2 80mVp-p 80mVp-p 100mVp-p 120mVp-p 150mVp-p 150mVp-p 240mVp-p 240mVp-p VOLTAGE ADJ. RANGE 4.3 ~ 5.8V ±2.0% 6.8 ~ 9V ±2.0% 10.2 ~ 13.8V ±1.0% 13.5 ~ 18V ±1.0% 21.6 ~ 28.8V ±1.0% 28.8 ~ 39.6V ±1.0% 40.8 ~ 55.2V ±1.0% 2.8 ~ 3.8V VOLTAGE TOLERANCE Note.3 ±2.0% LINE REGULATION ±0.5% ±0.5% ±0.5% ±0.3% ±0.3% ±0.2% ±0.2% ±0.2% LOAD REGULATION ±1.0% ±1.0% ±1.0% ±0.5% ±0.5% ±0.5% ±0.5% ±0.5% SETUP, RISE TIME 1000ms, 100ms/230VAC HOLD UP TIME (Typ.) 16ms/230VAC 88% 89% 89.5% 41.4 ~ 48.6V 57.6 ~ 67.2V VOLTAGE RANGE Note.5 85 ~ 264VAC FREQUENCY RANGE 47 ~ 63Hz INPUT PROTECTION 120 ~ 370VDC POWER FACTOR (Typ.) PF>0.95/230VAC EFFICIENCY (Typ.) 80% 83% AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 5A/115VAC 35A/115VAC 2.4A/230VAC 70A/230VAC PF>0.99/115VAC at full load 86.5% 88% 89% LEAKAGE CURRENT Earth leakage current < 300μA/264VAC , Touch leakage current < 100μA/264VAC OVERLOAD 105 ~ 135% rated output power Protection type : Constant current limiting, recovers automatically after fault condition is removed 6 ~ 7V 9.4 ~ 10.9V 14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V 3.96 ~ 4.62V OVER VOLTAGE 5V STANDBY Protection type : Shut down o/p voltage, re-power on to recover Shut down o/p voltage, recovers automatically after temperature goes down 5VSB : [email protected] ; tolerance±5%, ripple : 50mVp-p(max.) DC OK SIGNAL PSU turn on : 3.3 ~ 5.6V ; PSU turn off : 0 ~ 1V OVER TEMPERATURE FUNCTION 2500ms, 100ms/115VAC at full load 16ms/115VAC at full load REMOTE CONTROL RC+ / RC-: 4 ~ 10V or open = power on ; 0 ~ 0.8V or short = power off FAN CONTROL (Typ.) Load 20±10% or RTH2≧50℃ Fan on WORKING TEMP. -40 ~ +70℃ (Refer to "Derating Curve") 20 ~ 90% RH non-condensing WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85℃, 10 ~ 95% RH TEMP. COEFFICIENT ± 0.03%/℃ (0 ~ 50℃ ) VIBRATION 10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes SAFETY & EMC (Note 4) OTHERS SAFETY STANDARDS ISOLATION LEVEL ANSI/AAMI ES60601-1, IEC60601-1 approved Primary-Secondary: 2×MOOP, Primary-Earth: 1×MOOP, Secondary-Earth: 1×MOOP WITHSTAND VOLTAGE I/P-O/P:4KVAC ISOLATION RESISTANCE EMC EMISSION I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3 EMC IMMUNITY Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2 MTBF 159.3K hrs min. DIMENSION 218*105*41mm (L*W*H) 1.19Kg; 12pcs/15.3Kg/0.82CUFT PACKING NOTE I/P-FG:2KVAC O/P-FG:0.5KVAC MIL-HDBK-217F (25℃) 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) 5. Derating may be needed under low input voltages. Please check the derating curve for more details. 6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time. 7. No load power consumption<0.5W when RC- & RC+ (CN100 pin1,2) 0 ~ 0.8V or short. 8. When the input voltage is less than 40VAC, the SPS may exhibit degradation of performance. The final product manufacturers must re-confirm this deviation that does not affect basic safety or essential performance. File Name:MSP-450-SPEC 2016-09-14 MSP-450 450W Single Output Medical Type Mechanical Specification Case No. 995A Unit:mm 32.8 20.75 152.4 series O/P 11 1 2 4 I/P 105 3 63.5 FAN 9.2 Air flow direction 8.2 5 6 1 1 7 2 8 218 152.4 32.8 18max. LED (Vo ADJ.)SVR1 4-M4 L=5 3 CN100 10 2 12.5 41 18 12max. AC Input Terminal Pin No. Assignment Pin No. 1 2 3 8-M4 L=4 (Both Sides) Connector Pin No. Assignment(CN100) : HRS DF11-8DP-2DS or equivalent DC Output Terminal Pin No. Assignment Assignment AC/L AC/N FG Pin No. 1~3 4~6 Assignment RC+ RCAUX AUXG Pin No. 1 2 3 4 Assignment -V +V Pin No. 5 6 7 8 Assignment DC-OK GND +S -S Mating Housing HRS DF11-8DS HRS DF11-**SC or equivalent or equivalent Block Diagram PWM fosc : 70KHz ACTIVE INRUSH CURRENT LIMITING EMI FILTER I/P Terminal RECTIFIERS & PFC AUX POWER RECTIFIERS & FILTER AUX AUXG +S POWER SWITCHING RECTIFIERS & FILTER +V -V -S DETECTION CIRCUIT O.L.P. O.T.P. FG DC OK O.T.P. PFC CONTROL PWM CONTROL Derating Curve O.V.P. RC+ REMOTE CONTROL RC- Output Derating VS Input Voltage 100 125 90 100 80 70 LOAD(%) LOAD (%) 80 50 20 -40 0 10 20 30 40 50 AMBIENT TEMPERATURE (℃) 60 70 (HORIZONTAL) 60 50 40 85 100 125 135 155 264 INPUT VOLTAGE (V) 60Hz File Name:MSP-450-SPEC 2016-09-14 MSP-450 450W Single Output Medical Type series Function Description of CN100 Pin No. Function Description 1 RC+ Turns the output on and off by electrical or dry contact between pin 2 (RC-), Short: Power OFF, Open: Power ON. RC- Remote control ground. 2 Auxiliary voltage output, 4.75~5.25V, referenced to pin 4(AUXG). The maximum load current is 0.3A. This output is not controlled by the 3 AUX "remote ON/OFF control". 4 5 6 7 8 AUXG Auxiliary voltage output ground. The signal return is isolated from the output terminals (+V & -V). DC-OK DC-OK Signal is a TTL level signal, referenced to pin6(DC-OK GND). High when PSU turns on. GND This pin connects to the negative terminal(-V). Return for DC-OK signal output. Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to +S minimize noise pick-up effect. The maximum line drop compensation is 0.5V. -S Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V. Function Manual 1.Remote Sense The remote sensing compensates voltage drop on the load wiring up to 0.5V. CN100 CN100 2 1 8 7 2 RC- RC+ 1 AUXG AUX GND DC-OK 8 +S (Pin7) - + LOAD Fig 1.1 -S +S 7 -S (Pin8) Sense lines should be twisted in pairs 2.DC-OK Signal DC-OK signal is a TTL level signal. High when PSU turns on. Between DC-OK(pin5) and GND(pin6) CN100 Output Status CN100 3.3 ~ 5.6V ON 0 ~ 1V OFF 2 1 8 7 2 RC- RC+ 1 AUXG AUX GND DC-OK 8 -S +S 7 Fig 2.1 3.Remote Control The PSU can be turned ON/OFF by using the "Remote Control" function. Between RC+(pin1) and RC-(pin2) Output Status SW ON (Short) OFF SW OFF (Open) ON CN100 2 1 8 7 1 2 CN100 RC- RC+ AUXG AUX GND DC-OK 8 -S +S 7 SW Fig 3.1 File Name:MSP-450-SPEC 2016-09-14