Bel MBC75-1024L 3 x 2 x 1 inch form factor Datasheet

The MBC75 Series of open frame medical
power supplies feature a wide universal AC
input range of 85 V – 264 VAC, offering 75 W
of output power in a compact 3 x 2 x 1 inch
footprint, with a variety of isolated single
output voltages.

3 x 2 x 1 Inch Form Factor

75 Watts with Convection Cooling

Approval to EN60601 3rd Edition

Efficiencies 93%

-40 To 70 Degree Operating Temperature

Dual Fusing

Thermal Shut-Down Feature

>3.00 Million Hours, Telcordia -SR332-Issue 3

Standby Power < 0.3W

Class II Option Available

RoHS Compliant

CE Marked
The MBC series was designed to 3rd edition
medical approvals and provides 2x MOPP
(Means of Patient Protection) isolation for
Class I and Class II installations.
These power supplies are ideal for medical,
telecom, Datacom, industrial equipment and
other applications.






Diagnostic
Drug Pump
Dialysis
Home Health Care
Monitoring
Portable Equipment
North America
+1-866.513.2839
Asia-Pacific
+86.755.29885888
Europe, Middle East
+353 61 225 977
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00665_001
MBC75 Series
MODEL NUMBER 2
DESCRIPTION
MBC75-1T12L
MBC75-1012L
MBC75-1T15L
MBC75-1015L
MBC75-1T24L
MBC75-1024L
MBC75-1T30L
MBC75-1030L
MBC75-1T48L
MBC75-1048L
MBC75-1T58L
MBC75-1058L
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
VOLTAGE
MAX. LOAD
(CONVECTION)
MIN. LOAD
RIPPLE & NOISE 1
12 V
6.25 A
0.0 A
1%
15 V
5.00 A
0.0 A
1%
24 V
3.12 A
0.0 A
1%
30 V
2.50 A
0.0 A
1%
48 V
1.56 A
0.0 A
1%
58 V
1.29 A
0.0 A
1%
COVER-75-XBC
1
2
metal cover kit accessory
Ripple is peak to peak with 20 MHz bandwidth and 10 μF (Tantalum capacitor) in parallel with a 0.1 μF capacitor at rated line
voltage and load ranges.
For Class II (without input Earth pin) add suffix -2 (e.g.: MBC75-1012L-2).
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00665_001
MBC75 Series
Specifications are for nominal input voltage, 25°C unless otherwise stated.
PARAMETER
Input Voltage
DESCRIPTION / CONDITION
Universal
(Derate from 75 W at 10 0VAC to 65 W at 85 VAC)
Input Frequency
SPECIFICATION
85-264 VAC / 390 VDC3
Power Factor
115 VAC:
230 VAC:
Typical
115 VAC:
230 VAC:
264 VAC:
Typical (N.A. For Class II Option)
Touch current
@ Full Load
47-63 Hz
1 A max.
0.5 A max.
< 0.3 W
25 A
45 A
75 A
300 uA
<100 uA
> 0.95
Switching Frequency
Typical
60 kHz
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Output Power
Convection:
48 V, 58 V:
24 V, 30 V:
12 V, 15 V:
Typical
75 W
93%
91%
90%
>16 ms
Input Current
No Load Power
Inrush Current
Leakage Current
3
Functional, not approved.
Efficiency
Hold-up Time
Line Regulation
+/-0.5%
Load Regulation
Transient Response
+/-1%
25% step load change, at 0.1A/uS slew rate,
50% duty cycle, 50 Hz = 4%
Voltage Adjustment
Rise Time
Typical
55 ms
Set Point Tolerance
+/-1%
Over Current Protection
> 110%
Over Voltage Protection
Latch type (AC recycling required)
Short Circuit Protection
Hiccup mode
PARAMETER
DESCRIPTION / CONDITION
Operating Temperature
recovery time < 5 ms
+/-3%
110 to 140%
Startup is guaranteed with spec deviation, see Fig 1.
Storage Temperature
SPECIFICATION
-40 to +70°C
-40 to 0°C
-40 to +85°C
Cooling
With natural convection cooling at 100 to 264 VAC
75 W
Relative Humidity
Noncondensing
Operating:
Nonoperating:
MTBF according to Telcordia -SR332-Issue 3
5% to 95%
16,000 ft
40,000 ft.
3.00 million hours
Altitude
Reliability
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00665_001
MBC75 Series
PARAMETER
DESCRIPTION / CONDITION
Conducted Emissions
EN55022-B, CISPR22-B, FCC PART15 - B
Static Discharge
EN61000-4-2:
Level-3
RF Field Susceptibility
EN61000-4-3:
Level-3
Fast Transients/Bursts
Level-3
Level A
Level B
Surge Susceptibility
EN61000-4-4:
Radiated:
Radiated with external core:
(King core K5B RC 25x12x15-M in input cable with 5 Turns)
EN61000-4-5:
Harmonic Current
EN61000-3-2:
Class D
PARAMETER
SPECIFICATION
4000 VAC
1500 VAC
1500 VAC
500 VAC
Agency Approvals
DESCRIPTION / CONDITION
Input to Output: (For medical applications)
Input to GND: (Not Applicable For Class II Option)
Output to GND: for type BF
for type B (N/A For Class II Option)
Approved to the latest edition of the following standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1.
Nemko, UL, C-UL
CE mark
Complies with LVD Directive
Radiated Emissions
Isolation Voltage
Safety Standard(s)
SPECIFICATION
Level-3
Figure 1 – Derating Curve
CONNECTOR
PIN
AC Input Connector
J1
DC Output Connector
J2
DESCRIPTION / CONDITION
Pin 1
Pin 2
Pin 3
AC Line
Not Fitted
AC Neutral
Pin 1, 2
Pin 3, 4
V1 -VE
V1 + VE
Screw Terminal (Option 1)
Molex Connector (Option 2)
Screw Terminal (Option 1)
Molex Connector (Option 2)
MANUFACTURER / PN
Molex: 39357-0003
Tyco: 2-1776112-3
Molex: 1722861103
(Mating conn: Molex 1722561003)
Molex: 39357 -0004
Tyco: 2-1776112-4
Molex: 1722861104
(Mating conn: Molex 1722561004)
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00665_001
MBC75 Series
PARAMETER
DESCRIPTION / CONDITION
Weight
150 g
Dimensions
76.2 x 50.8 x 25.4 mm (3 x 2 x 1 inch)
Figure 2 - Mechanical Drawing - Screw Terminal (Option 1)
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00665_001
MBC75 Series
Figure 3 - Mechanical Drawing - Molex Connector (Option 2)
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support
systems, equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending
on the date manufactured. Specifications are subject to change without notice.
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00665_001
Similar pages