ETC2 LD20-2600B Low flow sensing up to 1200 ml/h (20 ml/min) Datasheet

LD20 Liquid Flow Sensor Series – PROTOTYPE
Single-Use Flow Sensor for Biomedical Applications
 Low flow sensing up to 1200 ml/h (20 ml/min)
 Luer lock or barbed fluidic connectors
 Millisecond-fast response time
Product Summary
The LD20 liquid flow sensor series enables precise and
reliable measurements of dynamic liquid flow rates
from a few hundred µl/h up to 1200 ml/h bidirectionally. This sensor has been designed for singleuse applications in the biomedical field. The LD20
series is still being developed. Thus any specification
can be subject to change in the future (performance,
fittings, communication, etc.).
1
Electrical Interface:
Digital
- I2C-Bus
For more information on communication, please refer
to section 3 of this document (preliminary).
Sensor Performance (Preliminary data, subject to change)
The following Table 1 lists preliminary specifications of the current prototypes. Please contact Sensirion if different
specifications are required.
3D-printed samples of the small (LD20-0600) and medium (LD20-1300) LD20 versions are available on request.
Please contact Sensirion.
Parameter
H2O Full scale flow rate
H2O Sensor output limita
Accuracy below full scale
(whichever error is larger)
Repeatability below full scale
(whichever error is larger)
Temperature coefficientc
(additional error / °C; whichever is larger)
Mounting orientation sensitivityd
Flow detection response time 63
a
b
c
d
LD20-0600x
LD20-1300x
LD20-2600B
Unit
20 (0.3)
25 (0.4)
tbd
tbd
tbd
tbd
tbd
tbd
tbd
tbd
250 (4.2)
300 (5)
tbd
tbd
tbd
tbd
tbd
tbd
tbd
tbd
1000 (16.6)
1300 (21.67)
5
0.25
0.5
0.025
0.2
0.01
1.0
tbd
ml/h (ml/min)
ml/h (ml/min)
% of m.v.b
% of full scale
% of m.v.
% of full scale
% m.v. / °C
% full scale / °C
% of full scale
ms
Flow rate at which the sensor output saturates. See section 2 for performance between full scale and saturation point.
Measured value
Additional accuracy error for temperatures deviating from 23 °C.
Maximum additional offset when flow channel is vertical.
www.sensirion.com
Preliminary version 0.6 – December 2016
1/6
LD20 Liquid Flow Sensor Series
Parameter
Response time on power-up
Operating temperature
Operating humidity
Maximum recommended operating pressure
Burst pressure
Recommended maximum use duration
Short term storage temperaturee
Short term storage humiditye
Shelf lifef
LD20-0600x
LD20-1300x
LD20-2600B
Unit
tbd
tbd
+15 … +45 (+59 … +113)
0…95 %, non-condensing
3 (45)
10 (145)
7
-10 … +60 (+14 … +140)
0…95 %, non-condensing
3
tbd
ms
°C (°F)
% RH
bar (psi)
bar (psi)
days
°C (°F)
% RH
years
Table 1: Performance of LD20 series (all data for medium H2O, at 23°C)
2
Specifications Charts
15%
LD20-2600B Absolute Accuracy with H 2 O
At 6°C/40°C
At 6°C/40°C
10%
Repeatability
5%
0%
-1300 -1000
At 23°C
200
At 23°C
Accuracy (ml/h)
Accuracy (% of measured value)
LD20-2600B Relative Accuracy with H2O
20
Repeatability
2
0.2
0.02
-500
0
Flow rate (ml/h H2O)
500
1000 1300
1
10
100
1000
Flow rate (ml/h H2O)
Figure 1: Liquid flow sensor accuracy and repeatability across the flow range of the LD20-2600B. Relative error in %
of measured value (left) and absolute error in ml/h (right) for H2O.
e Flow path empty. Short term storage refers to temporary
f When stored at 15°C - 35°C and 20 - 80 % r.H.
www.sensirion.com
conditions during e.g. transport.
Preliminary version 0.6 – December 2016
2/6
LD20 Liquid Flow Sensor Series
3
Communication with the Sensor (Preliminary data, subject to change)
The LD20-2600B flow sensor shows bi-directional,
linear transfer characteristics. The product comes fully
calibrated for water.
Digital sampling time, 16 bit
Digital sampling time, 9 bit
74 ms
1 ms
3.1 Electrical Specifications
Parameter
Power supply
DC, VDD
Operating
current
Conditions
Min. Typ. Max. Units
Sensor only
3.3
VDD = 3.5 V
3.5
3.7
5.1
V
mA
Table 2: DC characteristics
The liquid flow sensor is equipped with metal contact
pads for electrical connection.g
Description
SDA (data)
SCL (clock)
VDDD
VDDA
GND
Digital communication between a master and the
LD20 sensor runs via the standard I²C-interface. The
physical interface consists of two bus lines, a data line
(SDA) and a clock line (SCL) which need to be
connected via pull-up resistors to the bus voltage of
the system. By default, the I2C address is set to 64
(hexadecimal: 40, binary: 1000000).
These lines can be used on the 3.3 V level with a
clock frequency of 100 kHz. For the detailed
specifications of this I²C communication, please refer
to the specific I²C Application Notes from Sensirion.
Parameter
3.2 Electrical Connector and Pad Assignment
Pad
1
2
3
4
5
3.3 Digital Communication via I²C-Bus
1 … 5
bus clock frequency
Output voltage low
(SDA/SCL), Isink = 6 mA
Low level output current
(SDA/SCL)
High level input voltage
(SDA/SCL)
Low level input voltage
(SDA/SCL)
Min.
I2 C
Typ.
Max. Units
100
0.1
kHz
0.5
V
6
mA
2.0
V
1.0
V
Table 4: I²C output characteristics
Table 3: Electrical pad layout
g Please
note, that the electrical connector provided as part of the
evaluation kit for testing purposes will not be commercially
available. Its design is in the scope of the customer.
www.sensirion.com
Preliminary version 0.6 – December 2016
3/6
LD20 Liquid Flow Sensor Series
4
Fluidic Specifications and Connections
Parameter
LD20-0600x
LD20-1300x
Wetted materials
3D-printed samples are available on
request
Barbed fittings for 3D printed samples
(Final fitting design to be defined)
Fluidic connector ports (fittings)
OD of barbed fittings
Recommended tubing ID for barbed fittings
Pressure drop (at full scale flow rate, H2O, 23 °C)
LD20-2600B
Polyetherimide (PEI),
Liquid-Crystal-Polymer (LCP),
medical-grade adhesive
Barbed fittings
1.75 mm
tbd
4.0 mm
1.6 mm (1/16”)
<1.0 mbar
tbd
tbd
3.2 mm (1/8”)
< 0.2 mbar
Table 5: Fluidic specifications and connections
5
Mechanical Specifications
Parameter
Largest dimensions
Total mass
Inner diameter
LD20-0600x
LD20-1300x
LD20-2600B
~ 35.4 x 15 x 12 mm
tbd
~ 0.8 mm
tbd
tbd
tbd
~ 48 x 15 x 12 mm
1.7 g
~ 2.6 mm
Table 6: Mechanical specifications
Figure 2: Dimensions of LD20-2600B (in mm)
www.sensirion.com
Preliminary version 0.6 – December 2016
4/6
LD20 Liquid Flow Sensor Series
6
Order Information
The LD20 Evaluation Kit contains:


3 pcs disposable liquid flow sensors of type
LD20-2600B
Base station used as a reusable electrical
interface connector (prototype)


PC Software (Viewer & Data Export Tool)
SCC1-USB Sensor Cable with USB connector
for plug-and-play connection to a PC.
Product
LD20-2600B Liquid Flow Sensor with barbed fittings
Base station for LD20 series (prototype)
SCC1-USB sensor cable
Evaluation Kit LD20-2600B
Article Number
1-101564-01
1-101481-01
1-101007-01
1-101431-02
Table 7: Order information
3D-printed samples of the small (LD20-0600x) and medium (LD20-1300x) LD20 versions are available on request also
as part of an Evaluation Kit. Please contact Sensirion.
www.sensirion.com
Preliminary version 0.6 – December 2016
5/6
LD20 Liquid Flow Sensor Series
Important Notices

Warning, personal injury
Do not use this product as safety or emergency stop devices
or in any other application where failure of the product could
result in personal injury (including death). Do not use this
product for applications other than its intended and authorized
use. Before installing, handling, using or servicing this
product, please consult the datasheet and application notes.
Failure to comply with these instructions could result in death
or serious injury.
If the Buyer shall purchase or use SENSIRION products for any
unintended or unauthorized application, Buyer shall defend,
indemnify and hold harmless SENSIRION and its officers,
employees, subsidiaries, affiliates and distributors against all
claims, costs, damages and expenses, and reasonable attorney
fees arising out of, directly or indirectly, any claim of personal injury
or death associated with such unintended or unauthorized use,
even if SENSIRION shall be allegedly negligent with respect to the
design or the manufacture of the product.
ESD Precautions
The inherent design of this component causes it to be sensitive to
electrostatic discharge (ESD). To prevent ESD-induced damage
and/or degradation, take customary and statutory ESD precautions
when handling this product.
Warranty
SENSIRION warrants solely to the original purchaser of this
product for a period of 12 months (one year) from the date of
delivery that this product shall be of the quality, material and
workmanship defined in SENSIRION’s published specifications of
the product. Within such period, if proven to be defective,
SENSIRION shall repair and/or replace this product, in
SENSIRION’s discretion, free of charge to the Buyer, provided that:



notice in writing describing the defects shall be given to
SENSIRION within fourteen (14) days after their appearance;
such defects shall be found, to SENSIRION’s reasonable
satisfaction, to have arisen from SENSIRION’s faulty design,
material, or workmanship;
the defective product shall be returned to SENSIRION’s factory
at the Buyer’s expense; and
the warranty period for any repaired or replaced product shall be
limited to the unexpired portion of the original period.
This warranty does not apply to any equipment which has not been
installed and used within the specifications recommended by
SENSIRION for the intended and proper use of the equipment. EXCEPT
FOR THE WARRANTIES EXPRESSLY SET FORTH HEREIN,
SENSIRION MAKES NO WARRANTIES, EITHER EXPRESS OR
IMPLIED, WITH RESPECT TO THE PRODUCT. ANY AND ALL
WARRANTIES, INCLUDING WITHOUT LIMITATION, WARRANTIES
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSLY EXCLUDED AND DECLINED.
SENSIRION is only liable for defects of this product arising under the
conditions of operation provided for in the datasheet and proper use of
the goods. SENSIRION explicitly disclaims all warranties, express or
implied, for any period during which the goods are operated or stored
not in accordance with the technical specifications.
SENSIRION does not assume any liability arising out of any application
or use of any product or circuit and specifically disclaims any and all
liability, including without limitation consequential or incidental damages.
All operating parameters, including without limitation recommended
parameters, must be validated for each customer’s applications by
customer’s technical experts. Recommended parameters can and do
vary in different applications.
SENSIRION reserves the right, without further notice, (i) to change the
product specifications and/or the information in this document and (ii) to
improve reliability, functions and design of this product.
Copyright © 2001-2016, SENSIRION.
CMOSens® is a trademark of SENSIRION.
All rights reserved.
RoHS, REACH and WEEE Statement
The flow sensors of the LD20 series comply with requirements of the
following directives and regulations:



EU Directive 1907/2006/EC concerning Registration, Evaluation,
Authorization and Restriction of Chemicals (REACH)
EU Directive 2002/96/EC on waste electrical and electronic
equipment (WEEE), OJ13.02.2003; esp. its Article 6 (1) with Annex II.
EU Directive 2002/65/EC on the restriction of certain hazardous
substances in electric and electronic equipment (RoHS),
OJ01.01.2011
Headquarters and Subsidiaries
SENSIRION AG
Laubisruetistr. 50
CH-8712 Staefa ZH
Switzerland
Sensirion Inc., USA
phone: +1 805 409 4900
[email protected]
www.sensirion.com
Sensirion Korea Co. Ltd.
phone: +82 31 337 7700-3
[email protected]
www.sensirion.co.kr
phone: +41 44 306 40 00
fax:
+41 44 306 40 30
[email protected]
www.sensirion.com
Sensirion Japan Co. Ltd.
phone: +81 3 3444 4940
[email protected]
www.sensirion.co.jp
Sensirion China Co. Ltd.
phone: +86 755 8252 1501
[email protected]
www.sensirion.com.cn
Sensirion Taiwan Co. Ltd.
[email protected]
www.sensirion.com
To find your local representative, please visit www.sensirion.com/contact
www.sensirion.com
Preliminary version 0.6 – December 2016
6/6
Similar pages