Bel MBC40-1012G Cover kit accessory available Datasheet

The MBC40 Series of open-frame medical power supplies, with its
wide universal 90-264 VAC input range, is available at 40 W of output
power and a variety of single and multiple output voltages.
The MBC series is designed and approved to the latest Medical
standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I
& Class II applications.
These medical power supplies are ideal for, monitoring, home health
equipment as well as surgical devices.
•
4 x 2 x 1.2 Inch Form Factor
•
Single to Triple Outputs
•
Ultra High Efficiency > 85%
•
Approved to EN/IEC 60601-1
•
Low leakage current < 250 µA
•
No Load Power < 0.3 W
•
IEC Protection Class Options:
o Class I: Earth pin J4 (no suffix)
o Class II: No Earth pin (-2 suffix)
•
RoHS Compliant
•
Cover Kit Accessory Available
•
Dialysis
•
Surgical Devices
•
Monitoring
•
Home Health
•
Pumps
•
Ultrasound
MBC40 Series
2
MODEL1
OUTPUT VOLTAGE (VDC)
MAX LOAD (A)2
MINIMUM LOAD (A)3
RIPPLE & NOISE 4
Specifications are for nominal input voltage, 25°C unless otherwise stated.
PARAMETER
Input Voltage
Input Frequency
DESCRIPTION / CONDITION
SPECIFICATION
Universal
90 - 264 VAC
6
Input Current
No Load Power
Inrush Current
Leakage Current
Switching Frequency
47 to 400 Hz
0.85 A max.
0.45 A max.
< 0.3 W
< 0.5 W
30 A max.
60 A max.
< 140 µA
<250 µA
120 VAC:
230 VAC:
Single output models
Multi output models
120 VAC:
230 VAC:
120 VAC:
230 VAC:
Typical
67
67 Hz
1
For Class II add suffix -2 (e.g.: MBC40-1012G-2).
Maximum current per output channel. Do not exceed total output power rating.
3
Minimum load specified to meet cross regulation.
4
Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Tantalum capacitor) in parallel with a 0.1 µF capacitor at rated line voltage
and load ranges.
5
For MBC40-3003G efficiency is 75% typical.
6
Safety Approved: 47 to 63 Hz
2
[email protected]
MBC40 Series
3
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Efficiency
Typical
85%
Hold Up Time
@ 120 VAC typical
>10 ms
Output Power
Derate output power linearly to 80% from 90 VAC to 80 VAC input.
40 W
Line Regulation
+/-0.3%
Load Regulation
V1:
V2 & V3:
Transient Response
50% to 100% load change, 50/60 Hz, 50% duty cycle, 0.1 A/μs
Rise Time
+/-0.5%
+/-5%
< 10%,
recovery time < 5 ms
< 100 ms
Set Point Tolerance
V1:
V2 & V3:
± 3%
± 5%
Output Voltage Adjustment
V1
± 10%
Over Current Protection
Typical above rating
130%
Over Voltage Protection
Typical for V1 only
130%
Short Circuit Protection
Short term, autorecovery
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Operating Temperature
Refer to derating curve, Fig. 1
Start-up is guaranteed
-20 to 70°C
-20 to 0°C
Storage Temperature
-40 to +85°C
Relative Humidity
Non Condensing
95%
Altitude
Operating:
Non-Operating:
10,000 ft.
40,000 ft.
Reliability
MTBF according to Telcordia -SR332-Issue 3
1.87 million hours
Cooling
Convection
Figure 1. Derating Curve
Europe, Middle East
+353 61 225 977
North America
+1 408 785 5200
© 2017 Bel Power Solutions & Protection
BCD.00286_AE
Asia-Pacific
+86 755 298 85888
MBC40 Series
4
Conducted Emissions
EN 55011-B, CISPR22-B, FCC PART15-B
Pass
Radiated Emissions
EN 55011 B
Pass
Input Current Harmonics
EN 61000-3-2
Class D
Voltage Fluctuation and Flicker
EN 61000-3-3
Pass
ESD Immunity
EN 61000-4-2
Level 4, Criterion A
Radiated Field Immunity
EN 61000-4-3
Level 3, Criterion A
Electrical Fast Transient Immunity
EN 61000-4-4
Level 3, Criterion A
Surge Immunity
EN 61000-4-5
Level 3, Criterion A
Conducted Immunity
EN 61000-4-6
Level 3, Criterion A
Magnetic Field Immunity
EN 61000-4-8
Level 4, Criterion A
Voltage Dips, Interruptions
EN 61000-4-11
Criterion A & B
PARAMETER
Isolation Voltage
Safety Standards
DESCRIPTION / CONDITION
SPECIFICATION
Input to Output, 2M0PP
Input to Ground, 1M0PP
Output to Ground
Approved to the latest edition of the following standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1.
Min. 5700 VDC
1500 VAC
500 VAC
Agency Approvals
Nemko, UL, C-UL
CE mark
Complies with LVD Directive
CONNECTOR
PIN
AC Input Connector
J1
DC Output Connector
J2
Signal Connector
J3
Earth
J4
DESCRIPTION / CONDITION
Pin 1
Pin 2
Pin 1,2
Pin 3,4
Pin 5
Pin 6
Pin 1
Pin 2
AC Neutral
AC Line
V1
RTN
V3
V2
+V1 Sense
-V1 Sense
MANUFACTURER / PN
Molex: 26-60-4030 or equivalent
Mating: 09-50-3031; Pins: 08-50-0106
Tyco: 640445-6 or equivalent
Mating: 647402-6; Pins: 3-647409-1
Molex: 22-23-2021 or equivalent
Mating: 22-01-2021
Molex: 19705-4301
Mating: 190030001
PARAMETER
DESCRIPTION / CONDITION
Weight
150 g (0.33 lbs.)
Dimensions
101.6 x 50.8 x 30.48 mm (4 x 2 x 1.2 inch)
[email protected]
MBC40 Series
5
Figure 2. Mechanical Drawing MBC40-1xxxG
Figure 3. Mechanical Drawing MBC40-3xxxG
NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following:
1
Stand off, used to mount PCB has OD of 5.4 mm max.
2
Screws, used to fix PCB on stand off, have head dia of 6.0 mm max.
3
Washer, if used, to have dia of 6.5 mm max.
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
Europe, Middle East
+353 61 225 977
North America
+1 408 785 5200
© 2017 Bel Power Solutions & Protection
BCD.00286_AE
Asia-Pacific
+86 755 298 85888
Similar pages