MPD MPU300MS24YZ Mpu-300m sery Datasheet

MPU-300M Series
Single Output, 300W
Medical Approved
AC/DC Power Supplies
Key Features:
• Small 300W Supply
• EN 60601-1 3RD Edition
• PFC to EN 61000-3-2 “D”
• Low Profile <1U Height
• 600W Peak Power
• Universal 90-264 VAC Input
• Three Mechanical Options
Electrical Specifications
Specifications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.
Input
Parameter
Input Voltage Range
Input Frequency
Input Current, Full Load
Conditions
Universal
Inrush Current, Cold Start
Leakage Current
Power Factor Correction
Input Protection
Min.
90
47
Typ.
90 VAC
5
110 VAC
220 VAC
240 VAC
Meets EN 61000-3-2 Class D
Two F5A/250V Fuse
Max.
264
63
35
70
300
Units
VAC
Hz
A
A
µA
Output
Parameter
Output Voltage Adjustment
Output Regulation, See Note 1
Hold Time
Ripple & Noise (20 MHz) See Note 2
Overload Protection
Over Voltage Protection
Over Temperature Protection
Temperature Coefficient
Transient Recovery Time, See Note 4
Transient Response Deviation
Overshoot/Undershoot
Turn On Delay
Output Short Circuit
Conditions
Min.
By Trim Pot
Typ.
±5.0
±1.0
Max.
120 VAC, 80% Load
16
See Model Selection Guide
Foldback Circuit, Autorecovery
110
140
>130% of Rated Output Voltage. Recycle AC Input.
Autorecovery
+110
±0.04
2.5
50% Load Change
5
At Turn On/Off
±5.0
120 VAC
1
Continuous With Autorecovery
Units
%
%
mSec
%
ºC
%/ºC
mS
%
%
S
General
Parameter
Isolation Voltage, See Note 5
Switching Frequency
Conditions
Input - Output
Input - FG (Frame Ground)
Primary - Core
Fixed
Min.
4,000
1,500
1,500
Typ.
Max.
Units
VAC
70
kHz
Interface Signals
Power Supply On
Green LED on the PCB
An open collector output rated for 28V/5 mA sink current maximum.
Goes high if a fan failure is detected
Goes TTL high 100 to 500 mS after regulation. Goes low at least 1 mS
before the loss of regulation. Will sink 100 mA.
A TTL low signal inhibits the output. Hiccup mode.
Fan Fail
Power Good Signal
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
T: (781) 344-8226
F: (781) 344-8481
E: [email protected]
W: www.micropowerdirect.com
Remote On/Off
Environmental
Parameter
Operating Temperature Range
Output Derating
Storage Temperature Range
Cooling
Operating Humidity
Conditions
Min.
Typ.
Max.
Units
0
+25
+70
ºC
2.5%/ ºC from +50 ºC to + 70 ºC
-20
+85
ºC
See Model Selection Guide
RH, Non-condensing
90
%
Ambient
Reliability Specifications
Parameter
MTBF
Safety Standards
EMI Compliance
EMS Immunity Compliance
Vibration
Conditions
Min.
Typ.
Max.
Units
MIL HDBK 217F, 30ºC, Gnd Benign
100
kHours
EN 60601 3RD Edition (UL)
Compliance to EN 55022 (CISPR22) Class B; EN 61000-3-2,3
EN 6100-4-2,3,4,5,6, 8, 11; EN 55024; CE Marked (LVD)
Sinusoidal 5~50 Hz, Acceleration ±7.35 m/s2 on X, Y & Z Axis
www.micropowerdirect.com
www.micropowerdirect.com
Model Selection Guide
Output Voltage (VDC) Output Current (Max A) Ripple
& Noise Efficiency
(%)
PreSet
Range
Convection 25With
CFM (% p-p)
Model
Number
MPU-300MS-12YZ 12 VDC 12.0 - 13.8
12.50
25.00
±1%
80%
MPU-300MS-15YZ 15 VDC 14.0 - 16.0
10.00
20.00
±1%
80%
MPU-300MS-24YZ 24 VDC 23.0 - 28.0
6.25
12.50
±1%
80%
MPU-300MS-48YZ 48 VDC 44.0 - 52.0
3.12
6.25
±1%
80%
Model Number
Notes:
1. Output regulation includes line & load.
2. Ripple & noise is measured from 10 Hz to 20 MHz. Connection to the unit is made with a
0.1 µF ceramic capacitor & a 22 µF electrolytic capacitor connected in parallel.
3. A 1% minimum load is required to maintain regulation & ripple specifications.
4. Transient recovery is measured to within a 1% error band for a load step of 50% to 100%.
5. Isolation specifications are production HI-Pot tested for 3 seconds.
6. The full output range (see table) is covered in the safety agency certification. Standard
models are factory set to the Preset voltage, but may be set to other levels within the range
without affecting the agency certification. For more information, contact the factory.
7. Output power is given for the factory preset voltage. The maximum continuous output
power level is 150W without providing 25 CFM airflow. All models provide a peak power
level of 600W for a maximum duration of 500 µS. For more information, contact the factory.
8. Each unit includes two 250V/5A fuses. Since these fuses are not field replaceable, it is
recommended that an external fuse of the same size be used on the input of the power
supply for protection.
Models with other output voltage levels
are available (i.e. 36 VDC, etc)
Contact the factory for details at:
[email protected]
Mechanical Dimensions
Connections
Input Connector (CN1):
• Terminal Block:
Howder HD-601-3P
3 pins, 6.35 mm Centers
• Molex Mating Part No:
Molex 035977-0500 or equivalent
(two pins removed)
Pin
Function
1
2
3
AC-Line
AC-Neutral
Field Ground
Logic Signal Connector (CN3):
• Mating Part No:
JST XHP-4 or equivalent
(CHYAO SHIUNN JS-2001-04)
• Mating Pins: JST SXH-002T-PO.6
For AWG 30 to 26
Output Connector (CN2):
• Terminal Block (Must be used on 12 VOUT):
Dinkle P-830N
M5 Screws
• Molex Mating Part No:
Molex 035977-0600 or equivalent
Pin
Molex
Pin
Dinkle
Pin
Function
1
2
3
4
5
6
-VOUT
-VOUT
-VOUT
+VOUT
+VOUT
+VOUT
1
2
-VOUT
+VOUT
1
2
3
4
Remote On/Off
Fan Fail
Common
Power Good
MicroPower Direct
Fan Driver Connector (CN4):
• Mating Part No:
JST XHP-3 or equivalent
(CHYAO SHIUNN JS-2001-03)
3 Pins 0.98 Pitch
Pin
Function
1
2
3
Fan Fail
Minus
Plus
Notes:
• All dimensions are typical in inches (mm)
• Tolerance x.xx = ±0.02 (±0.50)
We Power Your Success - For Less!
292 Page Street Ste D Stoughton, MA 02072
•
TEL: (781) 344-8226
•
FAX: (781) 344-8481
•
E-Mail: [email protected]