AD ADAS1000-4

Data Sheet
Low Power, Five Electrode
Electrocardiogram (ECG) Analog Front End
ADAS1000/ADAS1000-1/ADAS1000-2
FEATURES
Biopotential signals in; digitized signals out
5 acquisition (ECG) channels and one driven lead
Parallel ICs for up to 10+ electrode measurements
Master ADAS1000 or ADAS1000-1 used with Slave
ADAS1000-2
AC and DC lead-off detection
Internal pace detection algorithm on 3 leads
Support for user’s own pace
Thoracic impedance measurement (internal/external path)
Selectable reference lead
Scalable noise vs. power control, power-down modes
Low power operation from
11 mW (1 lead), 15 mW (3 leads), 21 mW (all electrodes)
Lead or electrode data available
Supports AAMI EC11:1991/(R)2001/(R)2007, AAMI EC38
R2007, EC13:2002/(R)2007, IEC60601-1 ed. 3.0 b:2005,
IEC60601-2-25 ed. 2.0 :2011, IEC60601-2-27 ed. 2.0
b:2005, IEC60601-2-51 ed. 1.0 b: 2005
Fast overload recovery
Low or high speed data output rates
Serial interface SPI-/QSPI™-/DSP-compatible
56-lead LFCSP package (9 mm × 9 mm)
64-lead LQFP package (10 mm × 10 mm body size)
APPLICATIONS
ECG: monitor and diagnostic
Bedside patient monitoring, portable telemetry, Holter,
AED, cardiac defibrillators, ambulatory monitors, pace
maker programmer, patient transport, stress testing
GENERAL DESCRIPTION
The ADAS1000/ADAS1000-1/ADAS1000-2 measure electro
cardiac (ECG) signals, thoracic impedance, pacing artifacts,
and lead-on/off status and output this information in the form
of a data frame supplying either lead/vector or electrode data at
programmable data rates. Its low power and small size make it
suitable for portable, battery-powered applications. The high
performance also makes it suitable for higher end diagnostic
machines.
Rev. A
The ADAS1000 is a full-featured, 5-channel ECG including
respiration and pace detection, while the ADAS1000-1 offers
only ECG channels with no respiration or pace features.
Similarly, the ADAS1000-2 is a subset of the main device and
is configured for gang purposes with only the ECG channels
enabled (no respiration, pace, or right leg drive).
The ADAS1000/ADAS1000-1/ADAS1000-2 are designed to
simplify the task of acquiring and ensuring quality ECG signals.
They provide a low power, small data acquisition system for
biopotential applications. Auxiliary features that aid in better
quality ECG signal acquisition include: multichannel averaged
driven lead, selectable reference drive, fast overload recovery,
flexible respiration circuitry returning magnitude and phase
information, internal pace detection algorithm operating on
three leads, and the option of ac or dc lead-off detection.
Several digital output options ensure flexibility when
monitoring and analyzing signals. Value-added cardiac post
processing is executed externally on a DSP, microprocessor, or
FPGA.
Because ECG systems span different applications, the
ADAS1000/ADAS1000-1/ADAS1000-2 feature a power/
noise scaling architecture where the noise can be reduced
at the expense of increasing power consumption. Signal
acquisition channels may be shut down to save power. Data
rates can be reduced to save power.
To ease manufacturing tests and development as well as offer
holistic power-up testing, the ADAS1000/ADAS1000-1/
ADAS1000-2 offer a suite of features, such as dc and ac test
excitation via the calibration DAC and CRC redundancy testing
in addition to readback of all relevant register address space.
The input structure is a differential amplifier input thereby
allowing users a variety of configuration options to best suit
their application.
The ADAS1000/ADAS1000-1/ADAS1000-2 are available in
two package options: either a 56-lead LFCSP or a 64-lead
LQFP package; they are specified over −40°C to +85°C
temperature range.
Document Feedback
Information furnished by Analog Devices is believed to be accurate and reliable. However, no
responsibility is assumed by Analog Devices for its use, nor for any infringements of patents or other
rights of third parties that may result from its use. Specifications subject to change without notice. No
license is granted by implication or otherwise under any patent or patent rights of Analog Devices.
Trademarks and registered trademarks are the property of their respective owners.
One Technology Way, P.O. Box 9106, Norwood, MA 02062-9106, U.S.A.
Tel: 781.329.4700 ©2012–2013 Analog Devices, Inc. All rights reserved.
Technical Support
www.analog.com
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
TABLE OF CONTENTS
Features .............................................................................................. 1
Extend Switch On Respiration Paths ....................................... 39
Applications ....................................................................................... 1
Pacing Artifact Detection Function (ADAS1000 Only) ....... 40
General Description ......................................................................... 1
Biventricular Pacers ................................................................... 42
Revision History ............................................................................... 2
Pace Detection Measurements ................................................. 42
Functional Block Diagram .............................................................. 3
Evaluating Pace Detection Performance ................................. 43
Specifications..................................................................................... 4
Pace Latency ................................................................................ 43
Noise Performance ....................................................................... 9
Timing Characteristics .............................................................. 10
Pace Detection via Secondary Serial Interface (ADAS1000
and ADAS1000-1 Only) ............................................................ 43
Absolute Maximum Ratings .......................................................... 13
Filtering ....................................................................................... 43
Thermal Resistance .................................................................... 13
Voltage Reference ....................................................................... 45
ESD Caution ................................................................................ 13
Gang Mode Operation ............................................................... 45
Pin Configurations and Function Descriptions ......................... 14
Interfacing in Gang Mode ......................................................... 47
Typical Performance Characteristics ........................................... 18
Serial Interfaces............................................................................... 48
Applications Information .............................................................. 25
Standard Serial Interface ........................................................... 48
Overview...................................................................................... 25
Secondary Serial Interface......................................................... 52
ECG Inputs—Electrodes/Leads ................................................ 28
RESET .......................................................................................... 52
ECG Channel .............................................................................. 29
PD Function ................................................................................ 52
Electrode/Lead Formation and Input Stage Configuration .. 30
SPI Output Frame Structure (ECG and Status Data) ................ 53
Defibrillator Protection ............................................................. 33
SPI Register Definitions and Memory Map ................................ 54
ESIS Filtering............................................................................... 33
Control Registers Details ............................................................... 55
ECG Path Input Multiplexing ................................................... 33
Examples of Interfacing to the ADAS1000 ............................. 72
Common-Mode Selection and Averaging .............................. 34
Software Flowchart .................................................................... 75
Wilson Central Terminal (WCT) ............................................. 35
Power Supply, Grounding, and Decoupling Strategy ............ 76
Right Leg Drive/Reference Drive ............................................. 35
AVDD .......................................................................................... 76
Calibration DAC ......................................................................... 36
ADCVDD and DVDD Supplies ............................................... 76
Gain Calibration ......................................................................... 36
Unused Pins/Paths ..................................................................... 76
Lead-Off Detection .................................................................... 36
Layout Recommendations ........................................................ 76
Shield Driver ............................................................................... 36
Outline Dimensions ....................................................................... 77
Respiration (ADAS1000 Model Only)..................................... 37
Ordering Guide .......................................................................... 78
Evaluating Respiration Performance ....................................... 39
REVISION HISTORY
1/13—Rev. 0 to Rev. A
Changes to Features Section............................................................ 1
Changes to Table 1 ............................................................................ 3
Changes to Excitation Current, Test Conditions/Comments,
Table 2 ................................................................................................ 5
Added Table 3; Renumbered Sequentially .................................... 9
Changes to Respiration (ADAS1000 Model Only) Section,
Figure 66, and Internal Respiration Capacitors Section............ 37
Changes to Figure 67 ...................................................................... 38
Changes to Figure 68...................................................................... 39
Added Evaluating Pace Detection Performance Section .......... 43
Added Table 15 ............................................................................... 47
Changes to Clocks Section ............................................................ 51
Changes to RESPAMP Name, Function, Table 28 ..................... 57
Changes to Bits[14:9], Function, Table 30 .................................. 59
Changes to Ordering Guide .......................................................... 78
8/12—Revision 0: Initial Version
Rev. A | Page 2 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
FUNCTIONAL BLOCK DIAGRAM
REFIN REFOUT
CAL_DAC_IO
RLD_OUT CM_IN
RLD_SJ
CM_OUT/WCT
DRIVEN
LEAD
AMP
–
VREF
CALIBRATION
DAC
SHIELD
SHIELD
DRIVE
AMP
+
AVDD
IOVDD
ADCVDD
ADCVDD, DVDD
1.8V
REGULATORS
DVDD
VCM_REF
(1.3V)
RESPIRATION
DAC
COMMONMODE AMP
AC
LEAD-OFF
DAC
AC
LEAD-OFF
DETECTION
10kΩ
BUFFER
PACE
DETECTION
MUXES
CS
SCLK
5× ECG PATH
AMP
EXT_RESP_LA
EXT_RESP_LL
AMP
ADC
SDO
DRDY
GPIO0/MCS
GPIO1/MSCLK
GPIO2/MSDO
GPIO3
ADC
EXT_RESP_RA
RESPIRATION PATH
ADAS1000
SDI
FILTERS,
CONTROL,
AND
INTERFACE
LOGIC
CLOCK GEN/OSC/
EXTERNAL CLK
SOURCE
XTAL1
CLK_IO
09660-001
ELECTRODES
×5
XTAL2
Figure 1. ADAS1000 Full Featured Model
Table 1. Overview of Features Available from ADAS1000 Generics
Generic
ADAS1000
ADAS1000-1
ADAS1000-2 2
ADAS1000-3
ADAS1000-4
1
2
ECG
5 ECG channels
5 ECG channels
5 ECG channels
3 ECG channels
3 ECG channels
Operation
Master/slave
Master/slave
Slave
Master/slave
Master/slave
Right Leg Drive
Yes
Yes
Yes
Yes
Respiration
Yes
Pace
Detection
Yes
Yes
Yes
Shield
Driver
Yes
Yes
Master
Interface 1
Yes
Yes
Yes
Yes
Yes
Yes
Package
Option
LFCSP, LQFP
LFCSP
LFCSP, LQFP
LFCSP, LQFP
LFCSP, LQFP
Master interface is provided for users wishing to utilize their own digital pace algorithm; see the Secondary Serial Interface section.
This is a companion device for increased channel count purposes. It has a subset of features and is not intended for standalone use. It may be used in conjunction with
any master device.
Rev. A | Page 3 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
SPECIFICATIONS
AVDD = 3.3 V ± 5%, IOVDD = 1.65 V to 3.6 V, AGND = DGND = 0 V, REFIN tied to REFOUT, externally supplied crystal/clock =
8.192 MHz. Decoupling for reference and supplies as noted in the Power Supply, Grounding, and Decoupling Strategy section. TA =
−40°C to +85°C, unless otherwise noted. Typical specifications are mean values at TA = 25°C.
For specified performance, internal ADCVDD and DVDD linear regulators have been used. They may be supplied from external
regulators. ADCVDD = 1.8 V ± 5%, DVDD = 1.8 V ± 5%.
Front-end gain settings: GAIN 0 = ×1.4, GAIN 1 = ×2.1, GAIN 2 = ×2.8, GAIN 3 = ×4.2.
Table 2.
Parameter
ECG CHANNEL
Min
Typ
Max
Unit
0.3
0.63
0.8
0.97
−40
1.3
1.3
1.3
1.3
±1
2.3
1.97
1.8
1.63
+40
V
V
V
V
nA
+200
−7
nA
mV
−7
−15
−22
±2
1||10
mV
mV
mV
μV/°C
GΩ||pF
110
dB
80
19
18
dB
Bits
Bits
16
Bits
30
5
ppm
ppm
4.9
9.81
39.24
3.27
6.54
26.15
2.45
4.9
19.62
1.63
µV/LSB
μV/LSB
μV/LSB
μV/LSB
μV/LSB
μV/LSB
μV/LSB
μV/LSB
μV/LSB
μV/LSB
3.27
13.08
μV/LSB
μV/LSB
Electrode Input Range
Input Bias Current
−200
Input Offset
Input Offset Tempco 1
Input Amplifier Input
Impedance 2
CMRR2
Crosstalk1
Resolution2
Integral Nonlinearity Error
Differential Nonlinearity Error
Gain2
GAIN 0 (×1.4)
GAIN 1 (×2.1)
GAIN 2 (×2.8)
GAIN 3 (×4.2)
105
Rev. A | Page 4 of 80
Test Conditions/Comments
These specifications apply to the following pins:
ECG1_LA, ECG2_LL, ECG3_RA, ECG4_V1, ECG5_V2,
CM_IN (CE mode), EXT_RESP_xx pins when used in
extend switch mode
Independent of supply
GAIN 0 (gain setting ×1.4)
GAIN 1 (gain setting ×2.1)
GAIN 2 (gain setting ×2.8)
GAIN 3 (gain setting ×4.2)
Relates to each electrode input; over operating range;
dc and ac lead-off are disabled
AGND to AVDD
Electrode/vector mode with VCM = VCM_REF
GAIN 3
GAIN 2
GAIN 1
GAIN 0
At 10 Hz
51 kΩ imbalance, 60 Hz with ±300 mV differential dc
offset; per AAMI/IEC standards; with driven leg loop
closed
Between channels
Electrode/vector mode, 2 kHz data rate, 24-bit data-word
Electrode/vector mode, 16 kHz data rate, 24-bit dataword
Electrode/analog lead mode, 128 kHz data rate, 16-bit
data-word
GAIN 0; all data rates
GAIN 0
Referred to input. (2 × VREF)/Gain/(2N − 1). Applies
after factory calibration. User calibration adjusts this
number.
At 19-bit level in 2 kHz data rate
At 18-bit level in 16 kHz data rate
At 16-bit level in 128 kHz data rate
At 19-bit level in 2 kHz data rate
At 18-bit level in 16 kHz data rate
At 16-bit level in 128 kHz data rate
At 19-bit level in 2 kHz data rate
At 18-bit level in 16 kHz data rate
At 16-bit level in 128 kHz data rate
No factory calibration for this gain setting
At 19-bit level in 2 kHz data rate
At 18-bit level in 16 kHz data rate
At 16-bit level in 128 kHz data rate
Data Sheet
Parameter
Gain Error
Gain Matching
ADAS1000/ADAS1000-1/ADAS1000-2
Min
−1
Typ
+0.01
Max
+1
Unit
%
−2
−0.1
−0.5
+0.1
+0.02
+0.1
25
+2
+0.1
+0.5
%
%
%
ppm/°C
Gain Tempco1
Input Referred Noise1
Analog Lead Mode
6
10
12
11
12
14
16
100
65
104
100
Electrode Mode
Digital Lead Mode
Power Supply Sensitivity2
Analog Channel Bandwidth1
Dynamic Range1
Signal-to-Noise Ratio1
COMMON-MODE INPUT
Input Voltage Range
Input Impedance2
Input Bias Current
COMMON-MODE OUTPUT
VCM_REF
Output Voltage, VCM
Output Impedance1
Short Circuit Current1
Electrode Summation
Weighting Error2
RESPIRATION FUNCTION
(ADAS1000 ONLY)
Input Voltage Range
Input Voltage Range (Linear
Operation)
Input Bias Current
Input Referred Noise1
Frequency2
Excitation Current
Resolution2
Measurement Resolution1
In-Amp Gain1
Gain Error
Gain Tempco1
0.3
−40
−200
1.28
0.3
μV p-p
μV p-p
μV p-p
μV p-p
μV p-p
μV p-p
μV p-p
dB
kHz
dB
dB
2.3
1||10
±1
1.3
1.3
0.75
4
1
0.3
+40
+200
1.32
2.3
±1
0.85
46.5 to 64
V
V
kΩ
mA
%
2.3
V
V p-p
+10
nA
μV rms
kHz
1.8/gain
−10
V
GΩ||pF
nA
nA
64
32
16
8
24
0.2
μA p-p
μA p-p
μA p-p
μA p-p
bits
Ω
0.02
1 to 10
Ω
1
25
%
ppm/C
Rev. A | Page 5 of 80
Test Conditions/Comments
GAIN 0 to GAIN 2, factory calibrated. Programmable
user or factory calibration option enables. Factory gain
calibration applies only to standard ECG interface.
GAIN 3 setting, no factory calibration for this gain
GAIN 0 to GAIN 2
GAIN 3
GAIN 2, 2 kHz data rate, see Table 4
0.5 Hz to 40 Hz; high performance mode
0.05 Hz to 150 Hz; high performance mode
0.05 Hz to 150 Hz; low power mode
0.05 Hz to 150 Hz; high performance mode
0.05 Hz to 150 Hz; low power mode
0.05 Hz to 150 Hz; high performance mode
0.05 Hz to 150 Hz; low power mode
At 120 Hz
GAIN 0, 2 kHz data rate, −0.5 dBFS input signal, 10 Hz
−0.5 dB FS input signal
CM_IN pin
Over operating range; dc and ac lead-off disabled
AGND to AVDD
CM_OUT pin
Internal voltage; independent of supply
No dc load
Not intended to drive current
Resistor matching error
These specifications apply to the following pins:
EXT_RESP_LA, EXT_RESP_LL, EXT_RESP_RA and selected
internal respiration paths (Lead I, Lead II, Lead III)
AC-coupled, independent of supply
Programmable gain (10 states)
Applies to EXT_RESP_xx pins over AGND to AVDD
Programmable frequency, see Table 28
Respiration drive current corresponding to differential
voltage programmed by RESPAMP bits in RESPCTL
register. Internal respiration mode, cable 5 kΩ/200 pF,
1.2 kΩ chest impedance
Drive Range A
Drive Range B2
Drive Range C2
Drive Range D2
Update rate 125 Hz
Cable <5 kΩ/200 pF per electrode, body resistance
modeled as 1.2 kΩ
No cable impedance, body resistance modeled as 1.2 kΩ
Digitally programmable in steps of 1
LSB weight for GAIN 0 setting
ADAS1000/ADAS1000-1/ADAS1000-2
Parameter
RIGHT LEG DRIVE/DRIVEN LEAD
(ADAS1000/ADAS1000-1 ONLY)
Output Voltage Range
RLD_OUT Short Circuit Current
Closed-Loop Gain Range2
Slew Rate2
Input Referred Noise1
Amplifier GBP2
DC LEAD-OFF
Data Sheet
Min
Typ
Max
Unit
0.2
−5
±2
AVDD − 0.2
+5
V
mA
25
V/V
mV/ms
μV p-p
MHz
200
8
1.5
Lead-Off Current Accuracy
High Threshold Level1
±10
2.4
%
V
Low Threshold Level1
Threshold Accuracy
AC LEAD-OFF
0.2
25
V
mV
Frequency Range
Lead-Off Current Accuracy
REFIN
Input Range2
Input Current
2.031
±10
1.76
450
1.8
113
675
kHz
%
1.84
950
V
μA
μA
REFOUT
Output Voltage, VREF
Reference Tempco1
Output Impedance2
Short Circuit Current1
Voltage Noise1
1.785
1.8
±10
0.1
4.5
33
17
1.815
V
ppm/°C
Ω
mA
μV p-p
μV p-p
CALIBRATION DAC
DAC Resolution
Full-Scale Output Voltage
Zero-Scale Output Voltage
DNL
Output Series Resistance2
Input Current
CALIBRATION DAC TEST TONE
Output Voltage
Square Wave
Low Frequency Sine Wave
High Frequency Sine Wave
SHIELD DRIVER (ADAS1000/
ADAS1000-1 ONLY)
Output Voltage Range
Gain
Offset Voltage
Short Circuit Current
Stable Capacitive Load2
CRYSTAL OSCILLATOR
Frequency2
Start-Up Time2
2.64
0.24
−1
0.9
10
2.7
0.3
10
Bits
V
V
LSB
kΩ
±5
nA
1
1
10
150
0.3
2.76
0.36
+1
15
External protection resistor required to meet
regulatory patient current limits; output shorted to
AVDD/AGND
0.05 Hz to 150 Hz
Internal current source, pulls up open ECG pins;
programmable in 10 nA steps: 10 nA to 70 nA
Of programmed value
Inputs are compared to threshold levels; if inputs
exceed levels, lead-off flag is raised
Programmable in 4 steps: 12.5 nA rms, 25 nA rms,
50 nA rms, 100 nA rms
Fixed frequency
Of programmed value, measured into low impedance
Channel gain scales directly with REFIN
Per active ADC
5 ECG channels and respiration enabled
On-chip reference voltage for ADC; not intended to
drive other components reference inputs directly,
must be buffered externally
Short circuit to ground
0.05 Hz to 150 Hz (ECG band)
0.05 Hz to 5 Hz (respiration)
Available on CAL_DAC_IO (output for master, input for
slave)
No load, nominal FS output is 1.5 × REFOUT
No load
Not intended to drive low impedance load, used for
slave CAL_DAC_IO configured as an input
When used as input
1.1
mV p-p
Hz
Hz
Hz
Rides on common-mode voltage, VCM_REF = 1.3 V
2.3
V
V/V
mV
μA
nF
Rides on common-mode voltage, VCM
1
−20
Test Conditions/Comments
+20
25
10
Output current limited by internal series resistance
Applied to XTAL1 and XTAL2
8.192
15
MHz
ms
Rev. A | Page 6 of 80
Internal startup
Data Sheet
Parameter
CLOCK_IO
Operating Frequency2
Input Duty Cycle2
Output Duty Cycle2
DIGITAL INPUTS
Input Low Voltage, VIL
Input High Voltage, VIH
Input Current, IIH, IIL
Pin Capacitance2
DIGITAL OUTPUTS
Output Low Voltage, VOL
Output High Voltage, VOH
Output Rise/Fall Time
DVDD REGULATOR
Output Voltage
Available Current1
Short Circuit Current limit
ADCVDD REGULATOR
Output Voltage
Short Circuit Current Limit
POWER SUPPLY RANGES2
AVDD
IOVDD
ADCVDD
DVDD
POWER SUPPLY CURRENTS
AVDD Standby Current
IOVDD Standby Current
EXTERNALLY SUPPLIED ADCVDD
AND DVDD
AVDD Current
ADCVDD Current
DVDD Current
INTERNALLY SUPPLIED ADCVDD
AND DVDD
AVDD Current
POWER DISSIPATION
Externally Supplied ADCVDD
and DVDD 3
All 5 Input Channels and RLD
Internally Supplied ADCVDD
and DVDD
All 5 Input Channels and RLD
ADAS1000/ADAS1000-1/ADAS1000-2
Min
Typ
Max
Unit
80
MHz
%
%
8.192
20
50
Test Conditions/Comments
External clock source supplied to CLK_IO; this pin
is configured as an input when the device is
programmed as a slave
Applies to all digital inputs
0.3 × IOVDD
0.7 × IOVDD
−1
−20
+1
+20
3
0.4
V
V
ns
1.85
V
mA
mA
IOVDD − 0.4
4
1.75
1.8
1
40
V
V
μA
μA
pF
RESET has an internal pull-up
ISINK = 1 mA
ISOURCE = −1 mA
Capacitive load = 15 pF, 20% to 80%
Internal 1.8 V regulator for DVDD
Droop < 10 mV; for external device loading purposes
Internal 1.8 V regulator for ADCVDD; not
recommended as a supply for other circuitry
1.75
1.8
40
1.85
V
mA
3.15
1.65
1.71
1.71
3.3
1.8
1.8
5.5
3.6
1.89
1.89
V
V
V
V
785
1
975
60
μA
μA
If applied by external 1.8 V regulator
If applied by external 1.8 V regulator
All 5 channels enabled, RLD enabled, pace enabled
3.4
3.1
4.25
6.2
4.7
7
2.7
1.4
3.4
6.25
5.3
6.3
9
6.5
9
5
3.5
5.5
mA
mA
mA
mA
mA
mA
mA
mA
mA
High performance mode
Low performance mode
High performance mode, respiration enabled
High performance mode
Low performance mode
High performance mode, respiration enabled
High performance mode
Low performance mode
High performance mode, respiration enabled
All 5 channels enabled, RLD enabled, pace enabled
12.5
9.4
14.8
15.3
12.4
17.3
mA
mA
mA
High performance mode
Low performance mode
High performance mode, respiration enabled
All 5 channels enabled, RLD enabled, pace enabled
27
21
mW
mW
High performance (low noise)
Low power mode
All 5 channels enabled, RLD enabled, pace enabled
41
31
mW
mW
High performance (low noise)
Low power mode
Rev. A | Page 7 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Parameter
OTHER FUNCTIONS 4
Power Dissipation
Respiration
Shield Driver
Min
Typ
7.6
150
Data Sheet
Max
Unit
Test Conditions/Comments
mW
μW
Guaranteed by characterization, not production tested.
Guaranteed by design, not production tested.
3
ADCVDD and DVDD can be powered from an internal LDO or, alternatively, can be powered from external 1.8 V rail, which may result in a lower power solution.
4
Pace is a digital function and incurs no power penalty.
1
2
Rev. A | Page 8 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
NOISE PERFORMANCE
Table 3. Typical Input Referred Noise over 0.5 Second Window (µV p-p) 1
Mode
Analog Lead Mode 3
High Performance Mode
Data Rate 2
GAIN 0 (×1.4)
±1 VCM
GAIN 1 (×2.1)
±0.67 VCM
GAIN 2 (×2.8)
±0.5 VCM
GAIN 3 (×4.2)
±0.3 VCM
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
8
14
6
11
5
9
4
7.5
Typical values measured at 25°C, not subject to production test.
Data gathered using the 2 kHz packet/frame rate is measured over 0.5 seconds. The ADAS1000 internal programmable low-pass filter is configured for either 40 Hz or
150 Hz bandwidth. The data is gathered and post processed using a digital filter of either 0.05 Hz or 0.5 Hz to provide data over noted frequency bands.
3
Analog lead mode as shown in Figure 58 and Figure 57.
1
2
Table 4. Typical Input Referred Noise (μV p-p) 1
Mode
Analog Lead Mode 3
High Performance Mode
Low Power Mode
Electrode Mode 4
High Performance Mode
Low Power Mode
Digital Lead Mode 5, 6
High Performance Mode
Low Power Mode
Data Rate 2
GAIN 0 (×1.4)
±1 VCM
GAIN 1 (×2.1)
±0.67 VCM
GAIN 2 (×2.8)
±0.5 VCM
GAIN 3 (×4.2)
±0.3 VCM
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
128 kHz
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
128 kHz
12
20
95
180
13
22
110
215
8.5
14.5
65
130
9.5
15.5
75
145
6
10
50
105
7.5
12
59
116
5
8.5
39
80
5.5
9
45
85
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
128 kHz
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
128 kHz
13
21
100
190
14
22
110
218
9.5
15
70
139
9.5
15.5
75
145
8
11
57
110
7.5
12
60
120
5.5
9
41
85
5.5
9.5
45
88
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
16
25
130
18
30
145
11
19
90
12.5
21
100
9
15
70
10
16
80
6.5
10
50
7
11
58
Typical values measured at 25°C, not subject to production test.
Data gathered using the 2 kHz packet/frame rate is measured over 20 seconds. The ADAS1000 internal programmable low-pass filter is configured for either 40 Hz or
150 Hz bandwidth. The data is gathered and post processed using a digital filter of either 0.05 Hz or 0.5 Hz to provide data over noted frequency bands.
3
Analog lead mode as shown in Figure 58 and Figure 57.
4
Single-ended input electrode mode as shown in Figure 59 and Figure 57.
5
Digital lead mode as shown in Figure 59 and Figure 57.
6
Digital lead mode is available in 2 kHz and 16 kHz data rates.
1
2
Rev. A | Page 9 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
TIMING CHARACTERISTICS
Standard Serial Interface
AVDD = 3.3 V ± 5%, IOVDD = 1.65 V to 3.6 V, AGND = DGND = 0 V, REFIN tied to REFOUT, externally supplied crystal/clock =
8.192 MHz. TA = −40°C to +85°C, unless otherwise noted. Typical specifications are mean values at TA = 25°C.
Table 5.
Parameter 1
Output Rate 2
3.3 V
2
SCLK Cycle Time
tCSSA
tCSHA
tCH
tCL
tDO
tDS
tDH
tCSSD
tCSHD
tCSW
25
8.5
3
8
8
8.5
11
2
2
2
2
25
tDRDY_CS2
tCSO
RESET Low Time2
0
6
20
2
1.8 V
128
Unit
kHz
40
9.5
3
8
8
11.5
19
2
2
2
2
40
50
12
3
8
8
20
24
2
2
2
2
50
ns min
ns min
ns min
ns min
ns min
ns typ
ns max
ns min
ns min
ns min
ns min
ns min
0
7
20
0
9
20
ns min
ns typ
ns min
Description
Across specified IOVDD supply range; three programmable output data
rates available as configured in FRMCTL register (see Table 35) 2 kHz,
16 kHz, 128 kHz; use skip mode for slower rates
See Table 19 for details on SCLK vs. packet data rates
CS valid setup time to rising SCLK
CS valid hold time to rising SCLK
SCLK high time
SCLK low time
SCLK falling edge to SDO valid delay; SDO capacitance of 15 pF
SDI valid setup time from SCLK rising edge
SDI valid hold time from SCLK rising edge
CS valid setup time from SCLK rising edge
CS valid hold time from SCLK rising edge
CS high time between writes (if used). Note that CS is an optional input,
it may be tied permanently low. See a full description in the Serial
Interfaces section.
DRDY to CS setup time
Delay from CS assert to SDO active
Minimum pulse width; RESET is edge triggered
Guaranteed by characterization, not production tested.
Guaranteed by design, not production tested.
SCLK
tCSSA
tCH
tCSSD
tCL
tCSHA
tCSHD
CS
tCSW
tDS
tDH
MSB
LSB
DB[31]
SDI
DB[30]
R/W
DB[29]
DB[25]
DB[24]
ADDRESS
tCSO
DB[23]
DB[1]
DATA
LSB
MSB
SDO
DRDV
DB[0]
DO_31LAST
DO_30LAST
DO_29LAST
DO_25LAST
DO_1LAST
tDO
Figure 2. Data Read and Write Timing Diagram (CPHA = 1, CPOL = 1)
Rev. A | Page 10 of 80
DO_0LAST
09660-002
1
IOVDD
2.5 V
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
tDRDY_CS
DRDY
tCH
SCLK
tCL
tCSSA
tCSSD
tCSHD
tCSHA
CS
tDS
tCSW
tDH
MSB
SDI
LSB
DB[29]
N
DB[25]
N
DB[24]
N
DB[31]
N
DB[30]
N
R/W
ADDRESS = 0x40 (FRAMES)
DB[23]
DB[1]
LSB
MSB
DB[0]
N
DB[1]
N+1
DB[30]
N+1
DB[31]
N+1
DB[0]
N+1
DATA = NOP or 0x40
DATA
MSB
SDO
DRDY
DB[30]
N–1
DB[24] DB[23]
N–1
N–1
DB[25]
N–1
DB[1]
N–1
LSB
MSB
LSB
DB[31]
N–1
DB[0]
N–1
DB[31]
N
tDO
PREVIOUS DATA
DB[30]
N
DB[1]
N
DB[0]
N
HEADER (FIRST WORD OF FRAME)
09660-003
tCSO
Figure 3. Starting Read Frame Data (CPHA = 1, CPOL = 1)
tCH
SCLK
tCSSD
tCL
tCSSA
tCSHA
tCSHD
CS
tCSW
tDH
DB[31]
SDI
tDO
SDO
tDS
LSB
DB[30]
R/W
DB[29]
DB[28]
DB[24]
ADDRESS
DB[2]
DB[1]
DATA
MSB
DO_31LAST
DB[0]
LSB
DO_30LAST
DO_29LAST
DO_28LAST
DO_1LAST
tDO
Figure 4. Data Read and Write Timing Diagram (CPHA = 0, CPOL = 0)
Rev. A | Page 11 of 80
DO_0LAST
09660-004
MSB
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Secondary Serial Interface (Master Interface for Customer-Based Digital Pace Algorithm) ADAS1000/ ADAS1000-1 Only
AVDD = 3.3 V ± 5%, IOVDD = 1.65 V to 3.6 V, AGND = DGND = 0 V, REFIN tied to REFOUT, externally supplied crystal/clock =
8.192 MHz. TA = −40°C to +85°C, unless otherwise noted. Typical specifications are mean values at TA = 25°C. The following timing
specifications apply for the master interface when ECGCTL register is configured for high performance mode (ECGCTL[3] = 1), see
Table 26.
Table 6.
Parameter 1
Output Frame Rate 2
fSCLK2
Min
tMCSSA
tMDO
tMCSHD
tMCSW
Typ
128
2.5 × crystal
frequency
24.4
0
48.8
2173
Max
2026
1
2
Unit
kHz
MHz
Description
All five 16-bit ECG data-words are available at frame rate of 128 kHz only
Crystal frequency = 8.192 MHz
ns
ns
ns
ns
MCS valid setup time
MSCLK rising edge to MSDO valid delay
MCS valid hold time from MSCLK falling edge
MCS high time, SPIFW = 0, MCS asserted for entire frame as shown in
Figure 5, and configured in Table 31
MCS high time, SPIFW = 1, MCS asserted for each word in frame as shown in
Figure 6 and configured in Table 31
ns
Guaranteed by characterization, not production tested.
Guaranteed by design, not production tested.
tMSCLK
2
tMSCLK
MSCLK
tMCSSA
tMCSHD
MCS
SPIFW = 0*
tMCSW
MSB
MSDO
D0_15
D0_14
D0_1
LSB
MSB
D0_0
D1_15
D1_14
LSB
MSB
D5_0
D6_15
LSB
D6_14
D6_0
tMDO
16-BIT CRC WORD
5 × 16-BIT ECG DATA
09660-105
HEADER: 0xF AND 12-BIT COUNTER
*SPIFW = 0 PROVIDES MCS FOR EACH FRAME, SCLK STAYS HIGH FOR 1/2 MSCLK CYCLE BETWEEN EACH WORD.
Figure 5. Data Read and Write Timing Diagram for SPIFW = 0, Showing Entire Packet Of Data (Header, 5 ECG Words, and CRC Word)
tMSCLK
MSCLK
tMCSSA
tMSCLK
tMCSHD
MCS
SPIFW = 1*
tMCSW
LSB
MSB
MSDO
D0_15
D0_14
D0_1
D0_0
MSB
D1_15
LSB
D1_14
D5_0
MSB
D6_15
LSB
D6_14
D6_0
HEADER: 0xF AND 12-BIT COUNTER
5 × 16-BIT ECG DATA
16-BIT CRC WORD
*SPIFW = 1 PROVIDES MCS FOR EACH FRAME, SCLK STAYS HIGH FOR 1 MSCLK CYCLE BETWEEN EACH WORD.
Figure 6. Data Read and Write Timing Diagram for SPIFW = 1, Showing Entire Packet Of Data (Header, 5 ECG Words, and CRC Word)
Rev. A | Page 12 of 80
09660-005
tMDO
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
ABSOLUTE MAXIMUM RATINGS
THERMAL RESISTANCE
Table 7.
Parameter
AVDD to AGND
IOVDD to DGND
ADCVDD to AGND
DVDD to DGND
REFIN/REFOUT to REFGND
ECG and Analog Inputs to AGND
Digital Inputs to DGND
REFIN to ADCVDD
AGND to DGND
REFGND to AGND
ECG Input Continuous Current
Storage Temperature Range
Operating Junction Temperature Range
Reflow Profile
Junction Temperature
ESD
HBM
FICDM
Rating
−0.3 V to +6 V
−0.3 V to +6 V
−0.3 V to +2.5 V
−0.3 V to +2.5 V
−0.3 V to +2.1 V
−0.3 V to AVDD + 0.3 V
−0.3 V to IOVDD + 0.3 V
ADCVDD + 0.3 V
−0.3 V to + 0.3 V
−0.3 V to + 0.3 V
±10 mA
−65°C to +125°C
−40°C to +85°C
J-STD 20 (JEDEC)
150°C max
θJA is specified for the worst-case conditions, that is, a device
soldered in a circuit board for surface-mount packages.
Table 8. Thermal Resistance1
Package Type
56-Lead LFCSP
64-Lead LQFP
1
θJA
35
42.5
Unit
°C/W
°C/W
Based on JEDEC standard 4-layer (2S2P) high effective thermal conductivity
test board (JESD51-7) and natural convection.
ESD CAUTION
2500 V
1000 V
Stresses above those listed under Absolute Maximum Ratings
may cause permanent damage to the device. This is a stress
rating only; functional operation of the device at these or any
other conditions above those indicated in the operational
section of this specification is not implied. Exposure to absolute
maximum rating conditions for extended periods may affect
device reliability.
Rev. A | Page 13 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
AVDD
CM_IN
RLD_OUT
RLD_SJ
CM_OUT/WCT
AVDD
AGND
AGND
ADCVDD
XTAL1
XTAL2
CLK_IO
DVDD
DGND
56
55
54
53
52
51
50
49
48
47
46
45
44
43
NC
AGND 1
ECG5_V2 2
ECG4_V1 3
ECG3_RA 4
ECG2_LL 5
ECG1_LA 6
REFIN 7
REFOUT 8
REFGND 9
EXT_RESP_LA 10
EXT_RESP_LL 11
EXT_RESP_RA 12
RESPDAC_RA 13
AGND 14
64 63 62 61 60 59 58 57 56 55 54 53 52 51 50 49
NC
47
DGND
RESPDAC_RA 3
46
IOVDD
EXT_RESP_RA 4
45
SDO
EXT_RESP_LL
5
44
SCLK
EXT_RESP_LA
6
43
SDI
NC 1
PIN 1
2
7
ADAS1000
42
DRDY
REFOUT
8
64-LEAD LQFP
41
CS
TOP VIEW
(Not to Scale)
40
DGND
39
GPIO3
ECG2_LL 11
38
GPIO2/MSDO
ECG3_RA 12
37
GPIO1/MSCLK
ECG4_V1 13
36
GPIO0/MCS
ECG5_V2 14
35
IOVDD
AGND 15
34
DGND
NC 16
33
NC
REFIN 9
ECG1_LA 10
NC
DGND
IOVDD
GPIO0/MCS
GPIO1/MSCLK
GPIO2/MSDO
GPIO3
DGND
CS
DRDY
SDI
SCLK
SDO
IOVDD
DGND
NOTES
1. THE EXPOSED PADDLE IS ON THE TOP OF THE PACKAGE;
IT IS CONNECTED TO THE MOST NEGATIVE POTENTIAL, AGND.
09660-007
DGND
DVDD
CLK_IO
XTAL2
XTAL1
ADCVDD
AGND
AGND
AVDD
CM_OUT/WCT
RLD_SJ
RLD_OUT
CM_IN
NC
17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32
AVDD
TOP VIEW
(Not to Scale)
42
41
40
39
38
37
36
35
34
33
32
31
30
29
15
16
17
18
19
20
21
22
23
24
25
26
27
28
REFGND
ADAS1000
56-LEAD LFCSP
NOTES
1. NC = NO CONNECT. DO NOT CONNECT TO THIS PIN.
Figure 8. ADAS1000 56-Lead LFCSP Pin Configuration
56
55
54
53
52
51
50
49
48
47
46
45
44
43
AVDD
CM_IN
RLD_OUT
RLD_SJ
CM_OUT/WCT
AVDD
AGND
AGND
ADCVDD
XTAL1
XTAL2
CLK_IO
DVDD
DGND
Figure 7. ADAS1000 64-Lead LQFP Pin Configuration
PIN 1
INDICATOR
ADAS1000-1
56-LEAD LFCSP
TOP VIEW
(Not to Scale)
42
41
40
39
38
37
36
35
34
33
32
31
30
29
DGND
IOVDD
GPIO0/MCS
GPIO1/MSCLK
GPIO2/MSDO
GPIO3
DGND
CS
DRDY
SDI
SCLK
SDO
IOVDD
DGND
AVDD
NC
CAL_DAC_IO
SHIELD
VREG_EN
AVDD
AGND
AGND
ADCVDD
RESET
PD
SYNC_GANG
DVDD
DGND
15
16
17
18
19
20
21
22
23
24
25
26
27
28
AGND 1
ECG5_V2 2
ECG4_V1 3
ECG3_RA 4
ECG2_LL 5
ECG1_LA 6
REFIN 7
REFOUT 8
REFGND 9
NC 10
NC 11
NC 12
NC 13
AGND 14
NOTES
1. THE EXPOSED PADDLE IS ON THE TOP OF THE PACKAGE;
IT IS CONNECTED TO THE MOST NEGATIVE POTENTIAL, AGND.
Figure 9. ADAS1000-1 56-Lead LFCSP Pin Configuration
Rev. A | Page 14 of 80
09660-008
AGND
PIN 1
INDICATOR
AVDD
RESPDAC_LL
CAL_DAC_IO
SHIELD/RESPDAC_LA
VREG_EN
AVDD
AGND
AGND
ADCVDD
RESET
PD
SYNC_GANG
DVDD
DGND
48
09660-006
DGND
DVDD
SYNC_GANG
PD
RESET
ADCVDD
AGND
AGND
AVDD
VREG_EN
SHIELD/RESPDAC_LA
CAL_DAC_IO
RESPDAC_LL
AVDD
NC
PIN CONFIGURATIONS AND FUNCTION DESCRIPTIONS
AVDD
CM_IN
NC
RLD_SJ
NC
AVDD
AGND
AGND
ADCVDD
NC
NC
CLK_IN
DVDD
DGND
NC
DGND
DVDD
SYNC_GANG
PD
RESET
ADCVDD
AGND
AGND
AVDD
VREG_EN
NC
CAL_DAC_IN
ADAS1000/ADAS1000-1/ADAS1000-2
NC
AVDD
NC
Data Sheet
56
55
54
53
52
51
50
49
48
47
46
45
44
43
64 63 62 61 60 59 58 57 56 55 54 53 52 51 50 49
48 NC
PIN 1
2
NC
3
NC
4
45 SDO
NC
5
44 SCLK
47 DGND
AGND 1
ECG5 2
ECG4 3
ECG3 4
ECG2 5
ECG1 6
REFIN 7
REFOUT 8
REFGND 9
NC 10
NC 11
NC 12
NC 13
AGND 14
46 IOVDD
43 SDI
NC
6
REFGND
7
ADAS1000-2
REFOUT
8
64-LEAD LQFP
REFIN 9
42 DRDY
41 CS
40 DGND
TOP VIEW
(Not to Scale)
ECG1 10
39 GPIO3
ECG2 11
37 GPIO1
ECG4 13
36 GPIO0
ECG5 14
35 IOVDD
AGND 15
34 DGND
NC 16
33 NC
NC
TOP VIEW
(Not to Scale)
NOTES
1. THE EXPOSED PADDLE IS ON THE TOP OF THE PACKAGE;
IT IS CONNECTED TO THE MOST NEGATIVE POTENTIAL, AGND.
2. NC = NO CONNECT. DO NOT CONNECT TO THIS PIN.
09660-010
DGND
DVDD
CLK_IN
NC
NC
ADCVDD
AGND
AGND
AVDD
NC
RLD_SJ
NC
CM_IN
NC
17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32
AVDD
56-LEAD LFCSP
DGND
IOVDD
GPIO0
GPIO1
GPIO2
GPIO3
DGND
CS
DRDY
SDI
SCLK
SDO
IOVDD
DGND
AVDD
NC
CAL_DAC_IN
NC
VREG_EN
AVDD
AGND
AGND
ADCVDD
RESET
PD
SYNC_GANG
DVDD
DGND
ECG3 12
ADAS1000-2
42
41
40
39
38
37
36
35
34
33
32
31
30
29
15
16
17
18
19
20
21
22
23
24
25
26
27
28
38 GPIO2
PIN 1
INDICATOR
NOTES
1. NC = NO CONNECT. DO NOT CONNECT TO THIS PIN.
Figure 10. ADAS1000-2 Companion 64-Lead LQFP Pin Configuration
09660-009
NC 1
AGND
Figure 11. ADAS1000-2 Companion 56-Lead LFCSP Pin Configuration
Table 9. Pin Function Descriptions
ADAS1000
LQFP
LFCSP
18, 23,
15, 20,
58, 63
51, 56
35, 46
30, 41
ADAS1000-1
LFCSP
15, 20, 51, 56
30, 41
ADAS1000-2
LQFP
LFCSP
18, 23,
15, 20,
58, 63
51, 56
35, 46
30, 41
26, 55
23, 48
23, 48
26, 55
23, 48
ADCVDD
30, 51
27, 44
27, 44
30, 51
27, 44
DVDD
2, 15,
24, 25,
56, 57
31, 34,
40, 47,
50
59
1, 14,
21, 22,
49, 50
28, 29,
36, 42,
43
19
1, 14, 21, 22,
49, 50
2, 15,
24, 25,
56, 57
31, 34,
40, 47,
50
59
1, 14, 21,
22, 49,
50
28, 29,
36, 42,
43
19
AGND
Description
Analog Supply. See recommendations for bypass capacitors in the Power
Supply, Grounding, and Decoupling Strategy section.
Digital Supply for Digital Input/Output Voltage Levels. See
recommendations for bypass capacitors in the Power Supply, Grounding,
and Decoupling Strategy section.
Analog Supply for ADC. There is an on-chip linear regulator providing the
supply voltage for the ADCs. This pin is primarily provided for decoupling
purposes; however, the pin may also be supplied by an external 1.8 V
supply should the user wish to use a more efficient supply to minimize
power dissipation. In this case, use the VREG_EN pin tied to ground to
disable the ADCVDD and DVDD regulators. The ADCVDD pin should not
be used to supply other functions. See recommendations for bypass
capacitors in the Power Supply, Grounding, and Decoupling Strategy
section.
Digital Supply. There is an on-chip linear regulator providing the supply
voltage for the digital core. This pin is primarily provided for decoupling
purposes; however, the pin may also be overdriven supplied by an
external 1.8 V supply should the user wish to use a more efficient supply
to minimize power dissipation. In this case, use the VREG_EN pin tied to
ground to disable the ADCVDD and DVDD regulators. See
recommendations for bypass capacitors in the Power Supply, Grounding,
and Decoupling Strategy section.
Analog Ground.
DGND
Digital Ground.
VREG_EN
10
11
12
13
6
5
4
3
Enables or disables the internal voltage regulators used for ADCVDD and
DVDD. Tie this pin to AVDD to enable or tie this pin to ground to disable
the internal voltage regulators.
Analog Input, Left Arm (LA).
Analog Input, Left Leg (LL).
Analog Input, Right Arm (RA).
Analog Input, Chest Electrode 1 or Auxiliary Biopotential Input (V1).
28, 29, 36, 42,
43
19
6
5
4
3
Mnemonic
AVDD
IOVDD
ECG1_LA
ECG2_LL
ECG3_RA
ECG4_V1
Rev. A | Page 15 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
ADAS1000
LQFP
LFCSP
14
2
ADAS1000-1
LFCSP
2
ADAS1000-2
LQFP
LFCSP
10
11
12
13
14
6
5
4
3
2
Data Sheet
Mnemonic
ECG5_V2
ECG1
ECG2
ECG3
ECG4
ECG5
EXT_RESP_RA
EXT_RESP_LL
EXT_RESP_LA
RESPDAC_LL
Description
Analog Input, Chest Electrode 2 or Auxiliary Biopotential Input (V2).
Analog Input 1.
Analog Input 2.
Analog Input 3.
Analog Input 4.
Analog Input 5.
Optional External Respiration Input.
Optional External Respiration Input.
Optional External Respiration Input.
Optional path for higher performance respiration resolution, respiration
DAC drive, Negative Side 0.
Shared Pin (User-Configured).
Output of Shield Driver (SHIELD).
Optional Path for Higher Performance Respiration Resolution, Respiration
DAC Drive, Negative Side 1 (RESPDAC_LA).
Optional Path for Higher Performance Respiration Resolution, Respiration
DAC Drive, Positive Side.
Common-Mode Output Voltage (Average of Selected Electrodes). Not
intended to drive current.
Common-Mode Input.
Summing Junction for Right Leg Drive Amplifier.
Output and Feedback Junction for Right Leg Drive Amplifier.
Calibration DAC Input/Output. Output for a master device, input for a
slave. Not intended to drive current.
Reference Input. For standalone mode, use REFOUT connected to REFIN.
External 10 μF with ESR < 0.2 Ω in parallel with 0.1 μF bypass capacitors to
GND are required and should be placed as close to the pin as possible. An
external reference can be connected to REFIN.
Reference Output.
Reference Ground. Connect to a clean ground.
External crystal connects between these two pins; external clock drive
should be applied to CLK_IO. Each XTAL pin requires 15 pF to ground.
Buffered Clock Input/Output. Output for a master device; input for a slave.
Powers up in high impedance.
Chip Select and Frame Sync, Active Low. CS can be used to frame each
word or to frame the entire suite of data in framing mode.
Clock Input. Data is clocked into the shift register on a rising edge and
clocked out on a falling edge.
Serial Data Input.
Power-Down, Active Low.
4
5
6
62
12
11
10
16
60
18
SHIELD/
RESPDAC_LA
3
13
RESPDAC_RA
22
52
52
19
21
20
61
55
53
54
17
55
53
54
17
19
21
55
53
CM_IN
RLD_SJ
RLD_OUT
CAL_DAC_IO
9
7
7
9
7
REFIN
8
7
27, 28
8
9
47, 46
8
9
47, 46
8
7
8
9
REFOUT
REFGND
XTAL1, XTAL2
29
45
45
41
35
35
41
35
CS
44
32
32
44
32
SCLK
43
53
33
25
33
25
43
53
33
25
SDI
PD
45
31
31
45
31
SDO
42
34
34
42
34
DRDY
54
24
24
54
24
RESET
52
26
26
52
26
SYNC_GANG
36
40
40
GPIO0/MCS
Digital Input. This pin has an internal pull-up. This pin resets all internal
nodes to their power-on reset values.
Digital Input/Output (Output on Master, Input on Slave). Used for
synchronization control where multiple devices are connected together.
Powers up in high impedance.
General-Purpose I/O or Master 128 kHz SPI CS.
37
38
39
39
38
37
39
38
37
GPIO1/MSCLK
GPIO2/MSDO
GPIO3
General-Purpose I/O or Master 128 kHz SPI SCLK.
General-Purpose I/O or Master 128 kHz SPI SDO.
General-Purpose I/O.
CM_OUT/WCT
CLK_IO
Serial Data Output. This pin is used for reading back register configuration
data and for the data frames.
Digital Output. This pin indicates that conversion data is ready to be read
back when low, busy when high. When reading packet data, the entire
packet must be read to allow DRDY to return high.
Rev. A | Page 16 of 80
Data Sheet
ADAS1000
LQFP
LFCSP
1, 16,
17, 32,
33, 48,
49, 64
ADAS1000/ADAS1000-1/ADAS1000-2
ADAS1000-1
LFCSP
10, 11, 12, 13,
16
ADAS1000-2
LQFP
LFCSP
10, 11,
1, 3, 4,
12, 13,
5, 6,
16, 18,
16, 17,
46, 47,
20, 22,
52, 54
27, 28,
32, 33,
48, 49,
60, 62,
64
36
40
37
39
38
38
39
37
57
Description
No connect. Do not connect to these pins. (see Figure 7, Figure 9,
Figure 10, Figure 11).
General-Purpose I/O.
General-Purpose I/O.
General-Purpose I/O.
General-Purpose I/O.
Output of Shield Driver.
Calibration DAC Input. Input for companion device. Calibration signal
should come from the master.
Buffered Clock Input. This pin should be driven from the master CLK_IO pin.
Exposed Pad. The exposed paddle is on the top of the package; it is
connected to the most negative potential, AGND.
61
17
GPIO0
GPIO1
GPIO2
GPIO3
SHIELD
CAL_DAC_IN
29
45
57
CLK_IN
EPAD
18
57
Mnemonic
NC
Rev. A | Page 17 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
TYPICAL PERFORMANCE CHARACTERISTICS
8
15
10
INPUT REFERRED NOISE (µV)
4
2
0
–2
1
2
3
4
5
6
7
8
9
10
–5
Figure 12. Input Referred Noise for 0.5 Hz to 40 Hz Bandwidth, 2 kHz Data
Rate, GAIN 0 (1.4)
8
–15
09660-039
0
TIME (Seconds)
0.5Hz TO 150Hz
GAIN SETTING 3 = 4.2
DATA RATE = 2kHz
10 SECONDS OF DATA
0
1
2
3
4
5
6
7
8
9
10
TIME (Seconds)
Figure 15. Input Referred Noise for 0.5 Hz to 150 Hz Bandwidth, 2 kHz Data
Rate, GAIN 3 (4.2)
25
0.5Hz TO 40Hz
GAIN SETTING 3 = 4.2
DATA RATE = 2kHz
10 SECONDS OF DATA
LA 150Hz
LA 40Hz
INPUT REFERRED NOISE (µV)
6
INPUT REFERRED NOISE (µV)
0
–10
–4
–6
5
09660-042
INPUT REFERRED NOISE (µV)
0.5Hz TO 40Hz
GAIN SETTING 0 = 1.4
DATA RATE = 2kHz
6 10 SECONDS OF DATA
4
2
0
–2
20
15
10
5
0
1
2
3
4
5
6
7
8
9
10
TIME (Seconds)
0
09660-040
–6
Figure 13. Input Referred noise for 0.5 Hz to 40 Hz Bandwidth, 2 kHz Data
Rate, GAIN 3 (4.2)
GAIN 0
GAIN 1
GAIN 2
GAIN 3
GAIN SETTING
09660-043
–4
Figure 16. ECG Channel Noise Performance over a 0.5 Hz to 40 Hz or 0.5 Hz to
150 Hz Bandwidth vs. Gain Setting
15
0.0020
AVDD = 3.3V
GAIN SETTING 0 = 1.4
0.0018
GAIN ERROR (%)
5
0
–5
0.0016
0.0014
–15
0.5Hz TO 150Hz
GAIN SETTING 0 = 1.4
DATA RATE = 2kHz
10 SECONDS OF DATA
0
1
2
3
4
5
6
TIME (Seconds)
7
8
9
10
Figure 14. Input Referred Noise for 0.5 Hz to 150 Hz Bandwidth, 2 kHz Data
Rate, GAIN 0 (1.4)
Rev. A | Page 18 of 80
0.0010
LA
LL
RA
V1
ELECTRODE INPUT
Figure 17. Typical Gain Error Across Channels
V2
09660-044
0.0012
–10
09660-041
INPUT REFERRED NOISE (µV)
10
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
0.121
0.215
AVDD = 3.3V
AVDD = 3.3V
0.210
0.101
THRESHOLD (V)
GAIN ERROR (%)
0.205
0.081
0.061
0.041
0.200
0.195
0.190
0.021
GAIN 0
GAIN 1
GAIN 2
GAIN 3
0.180
–40
ECG DC LEAD-OFF THRESHOLD
RLD DC LEAD-OFF THRESHOLD
–20
0
GAIN SETTING
2.420
0 = 1.4
1 = 2.1
2 = 2.8
3 = 4.2
HIGH THRESHOLD (V)
–0.15
–0.20
2.405
2.400
2.395
2.390
2.385
GAIN ERROR G0
GAIN ERROR G1
GAIN ERROR G2
GAIN ERROR G3
–0.35
–40
–20
0
20
40
60
2.380
80
TEMPERATURE (°C)
Figure 19. Typical Gain Error for All Gain Settings Across Temperature
–20
0
20
40
60
80
Figure 22. DC Lead-Off Comparator High Threshold vs. Temperature
0
AVDD = 3.3V
GAIN SETTING 0 = 1.4
+85°C
+55°C
+25°C
–5°C
–40°C
ECG DC LEAD-OFF THRESHOLD
RLD DC LEAD-OFF THRESHOLD
TEMPERATURE (°C)
AVDD = 3.3V
–1
–2
2
–3
1
–4
GAIN (dB)
LEAKAGE (nA)
2.375
–40
09660-046
–0.30
09660-049
–0.25
0
–1
–5
–6
–2
–7
–3
–8
–4
–9
1.3
VOLTAGE (V)
1.8
2.3
09660-047
0.8
–10
Figure 20. Typical ECG Channel Leakage Current over Input Voltage Range vs.
Temperature
1
10
100
FREQUENCY (Hz)
1k
09660-050
GAIN ERROR (%)
–0.10
–5
0.3
AVDD = 3.3V
2.410
–0.05
3
80
2.415
0
4
60
Figure 21. DC Lead-Off Comparator Low Threshold vs. Temperature
0.15
5
40
TEMPERATURE (°C)
Figure 18. Typical Gain Error vs. Gain
AVDD = 3.3V
GAIN SETTING
0.10 GAIN SETTING
GAIN SETTING
0.05 GAIN SETTING
20
09660-048
0.001
09660-045
0.185
Figure 23. Filter Response with 40 Hz Filter Enabled, 2 kHz Data Rate; See
Figure 72 for Digital Filter Overview
Rev. A | Page 19 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
0
0
AVDD = 3.3V
–1
–1
–2
–2
–4
GAIN (dB)
GAIN (dB)
–3
–5
–6
–3
–4
–7
–8
–5
–9
10
100
1k
FREQUENCY (Hz)
AVDD = 3.3V
–6
1
0
0
–1
–2
–2
10k
100k
AVDD = 3.3V
–3
–4
GAIN (dB)
–5
–6
–4
–5
–6
–7
–7
–8
10
100
1k
FREQUENCY (Hz)
Figure 25. Filter Response with 250 Hz Filter Enabled, 2 kHz Data Rate; See
Figure 72 for Digital Filter Overview
0
–9
09660-052
1
1
10
100
1k
10k
100k
FREQUENCY (Hz)
09660-055
–8
–9
Figure 28. Filter Response Running at 128 kHz Data Rate; See Figure 72 for
Digital Filter Overview
1.8010
AVDD = 3.3V
–1
1.8005
–2
1.8000
–3
VOLTAGE (V)
1.7995
–4
–5
–6
1.7990
1.7985
1.7980
–7
1.7975
–8
1.7970
1
10
100
FREQUENCY (Hz)
1k
09660-053
–9
Figure 26. Filter Response with 450 Hz Filter Enabled, 2 kHz Data Rate; See
Figure 72 for Digital Filter Overview
Rev. A | Page 20 of 80
1.7965
–40
–20
0
20
40
60
TEMPERATURE (°C)
Figure 29. Typical Internal VREF vs. Temperature
80
09660-056
GAIN (dB)
–3
GAIN (dB)
1k
Figure 27. Analog Channel Bandwidth
–1
–10
100
FREQUENCY (Hz)
Figure 24. Filter Response with 150 Hz Filter Enabled, 2 kHz Data Rate; See
Figure 72 for Digital Filter Overview
–10
10
09660-054
1
09660-051
–10
Data Sheet
805
AVDD = 3.3V
AVDD = 3.3V
800
AVDD SUPPLY CURRENT (µA)
1.3005
1.2995
1.2990
1.2985
1.2980
785
780
775
0
20
40
60
80
TEMPERATURE (°C)
765
–40
0
20
40
60
80
TEMPERATURE (°C)
Figure 33. Typical AVDD Supply Current vs. Temperature in Standby Mode
Figure 30. VCM_REF vs. Temperature
12.65
AVDD = 3.3V
5 ECG CHANNELS ENABLED
INTERNAL LDO UTILIZED
12.45 HIGH PERFORMANCE/LOW NOISE MODE
12.60
12.40
12.55
CURRENT (mA)
12.50
12.35
12.30
LOW NOISE/HIGH
PERFORMANCE MODE
12.50
12.45
12.40
12.25
–20
0
20
40
60
80
TEMPERATURE (°C)
12.35
3.0
09660-060
12.20
–40
3.5
4.0
4.5
5.0
5.5
6.0
VOLTAGE (V)
Figure 34. Typical AVDD Supply Current vs. AVDD Supply Voltage
Figure 31. Typical AVDD Supply Current vs. Temperature, Using Internal
ADVCDD/DVDD Supplies
3.430
0.142955
AVDD = 3.3V
5 ECG CHANNELS ENABLED
ADCVDD AND DVDD SUPPLIED EXTERNALLY
HIGH PERFORMANCE/LOW NOISE MODE
RESPIRATION MAGNITUDE (V)
3.425
–20
09660-069
–20
09660-057
1.2970
–40
AVDD SUPPLY CURRENT (mA)
790
770
1.2975
AVDD SUPPLY CURRENT (mA)
795
09660-059
VOLTAGE (V)
1.3000
3.420
3.415
3.410
3.405
0.142945
0.142940
0.142935
0.142930
3.400
–20
0
20
40
TEMPERATURE (°C)
60
80
0.142925
09660-058
3.395
–40
AVDD = 3.3V
ECG PATH/DEFIB/CABLE IMPEDANCE = 0 Ω
PATIENT IMPEDANCE = 1k Ω
RESPIRATION RATE = 10RESPPM
0.142950 RESPAMP = 11 = 60µA p-p
RESPGAIN = 0011 = 4
Figure 32. Typical AVDD Supply Current vs. Temperature, Using Externally
Supplied ADVCDD/DVDD
0
5
10
15
TIME (Seconds)
20
25
30
09660-062
1.3010
ADAS1000/ADAS1000-1/ADAS1000-2
Figure 35. Respiration with 200 mΩ Impedance Variation, Using Internal
Respiration Paths and Measured with a 0 Ω Patient Cable
Rev. A | Page 21 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
0.517390
AVDD = 3.3V
ECG PATH/DEFIB/CABLE IMPEDANCE = 0Ω
PATIENT IMPEDANCE = 1kΩ
0.121140 RESPIRATION RATE = 10RESPPM
RESPAMP = 11 = 60µA p-p
RESPGAIN = 0011 = 4
0.121135
0.121130
0.121125
0.517375
0.517370
0.121120
0.517365
0.121115
0.517360
0
5
10
15
20
25
30
TIME (Seconds)
Figure 36. Respiration with 100 mΩ Impedance Variation, Using Internal
Respiration Paths and Measured with a 0 Ω Patient Cable
0.663160
0.663155
0.159770
0.663150
0.663145
0.663140
15
20
25
30
AVDD = 3.3V
TIME (seconds)
ECG PATH/DEFIB/CABLE IMPEDANCE = 1.5k Ω/600pF
PATIENT IMPEDANCE = 1k Ω
EXTCAP = 1nF
RESPIRATION RATE = 10RESPPM
RESPAMP = 11 = 60µA p-p
RESPGAIN = 0011 = 4
0.159765
0.159760
0.159755
15
20
25
30
0.159745
0.159125
RESPIRATION MAGNITUDE (V)
0.062350
0.062345
AVDD = 3.3V
ECG PATH/DEFIB/CABLE IMPEDANCE = 0Ω
PATIENT IMPEDANCE = 1kΩ
EXTCAP = 100pF
RESPIRATION RATE = 10RESPPM
RESPAMP = 11 = 60µA p-p
RESPGAIN = 0011 = 4
0
5
10
15
TIME (ms)
15
20
25
30
AVDD = 3.3V
ECG PATH/DEFIB/CABLE IMPEDANCE = 1.5kΩ/600pF
PATIENT IMPEDANCE = 1k Ω
0.159124
0.159123
0.159122
0.159121
0.159120
0.159119 EXTCAP= 1nF
20
25
30
0.159118
09660-065
0.062340
10
Figure 40. Respiration with 200 mΩ Impedance Variation, Using External
Respiration DAC Driving 1 nF External Capacitor and Measured with a 1.5 kΩ
Patient Cable
0.159126
0.062355
5
TIME (ms)
0.062365
0.062360
0
09660-066
10
Figure 37. Respiration with 200 mΩ Impedance Variation, Using Internal
Respiration Paths and Measured with a 5 kΩ Patient Cable
RESPIRATION MAGNITUDE (V)
10
Figure 38. Respiration with 200 mΩ Impedance Variation, Using External
Respiration DAC Driving 100 pF External Capacitor and Measured with a 0 Ω
Patient Cable
RESPIRATION RATE = 10RESPPM
RESPAMP = 11 = 60µA p-p
RESPGAIN = 0011 = 4
0
5
10
15
TIME (Seconds)
20
25
30
09660-068
5
09660-064
0
TIME (ms)
0.062335
5
0.159750
0.663135
0.663130
0
Figure 39. Respiration with 200 mΩ Impedance Variation, Using External
Respiration DAC Driving 100 pF External Capacitor and Measured with a
5 kΩ Patient Cable
0.159775
AVDD = 3.3V
ECG PATH/DEFIB/CABLE IMPEDANCE = 5k Ω
PATIENT IMPEDANCE = 1k Ω
RESPIRATION RATE = 10RESPPM
RESPAMP = 11 = 60µA p-p
RESPGAIN = 0011 = 4
RESPAMP = 11 = 60µA p-p
RESPGAIN = 0011 = 4
TIME (Seconds)
RESPIRATION MAGNITUDE (V)
RESPIRATION MAGNITUDE (V)
0.517380
09660-067
RESPIRATION MAGNITUDE (V)
AVDD = 3.3V
ECG PATH/DEFIB/CABLE IMPEDANCE = 5k Ω/250pF
PATIENTIMPEDANCE = 1k Ω
EXTCAP = 100pF
0.517385 RESPIRATION RATE = 10RESPPM
09660-063
RESPIRATION MAGNITUDE (V)
0.121145
Figure 41. Respiration with 100 mΩ Impedance Variation, Using External
Respiration DAC Driving 1 nF External Capacitor and Measured with a 1.5 kΩ
Patient Cable
Rev. A | Page 22 of 80
Data Sheet
50
ADAS1000/ADAS1000-1/ADAS1000-2
40
30
LA
LL
RA
V1
V2
AVDD = 3.3V
100
20
50
10
INL (µV/RTI)
DNL ERROR (µV RTI)
150
LA
LL
RA
V1
V2
AVDD = 3.3V
0
–10
0
–50
–20
–30
–100
0.5
0.7
0.9
1.1
1.3
1.5
1.7
1.9
2.1
2.3
INPUT VOLTAGE (V)
Figure 42. DNL vs. Input Voltage Range Across Electrodes at 25°C
50
30
0.5
0.7
0.9
1.1
1.3
1.5
1.7
1.9
2.1
2.3
INPUT VOLTAGE (V)
Figure 45. INL vs. Input Voltage Across Electrode Channel for 2 kHz Data Rate
150
–40°C
–5°C
+25°C
+55°C
+85°C
AVDD = 3.3V
40
AVDD = 3.3V
GAIN0
GAIN1
GAIN2
GAIN3
100
20
INL (µV/RTI)
DNL ERROR (µV RTI)
–150
0.3
09660-070
–50
0.3
09660-074
–40
10
0
–10
50
0
–20
–50
–30
0.5
0.7
0.9
1.1
1.3
1.5
1.7
1.9
2.1
2.3
INPUT VOLTAGE (V)
Figure 43. DNL vs. Input Voltage Range Across Temperature
AVDD = 3.3V
150
1.1
1.3
1.5
1.7
1.9
AVDD = 3.3V
2.1
2.3
GAIN0
GAIN1
GAIN2
GAIN3
100
50
INL (µV/RTI)
0
0
–50
–50
–100
–100
0.5
0.7
0.9
1.1
1.3
1.5
INPUT VOLTAGE (V)
1.7
1.9
2.1
2.3
–150
0.3
09660-073
INL (µV/RTI)
0.9
Figure 46. INL vs. Input Voltage Across Gain Setting for 16 kHz Data Rate
50
–150
0.3
0.7
INPUT VOLTAGE (V)
GAIN0
GAIN1
GAIN2
GAIN3
100
0.5
Figure 44. INL vs. Input Voltage Across Gain Setting for 2 kHz Data Rate
0.5
0.7
0.9
1.1
1.3
1.5
INPUT VOLTAGE (V)
1.7
1.9
2.1
2.3
09660-076
150
–100
0.3
09660-071
–50
0.3
09660-075
–40
Figure 47. INL vs. Input Voltage Across Gain Setting for 128 kHz Data Rate
Rev. A | Page 23 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
0
120
AVDD = 3.3V
GAIN 0
DATA RATE = 2kHz
FILTER SETTING = 150Hz
–20
80
60
LOOP GAIN (dB)
–60
–80
–100
–120
40
20
0
–140
–40
–160
–60
–180
0
50
100
150
200
250
300
350
400
450
500
FREQUENCY (Hz)
1
10
100
1k
10k
100k
1M
10M 100M
1G
FREQUENCY (Hz)
Figure 51. Open-Loop Gain Response of ADAS1000 Right Leg Drive Amplifier
Without Loading
Figure 48. FFT with 60 Hz Input Signal
150
–80
100m
09660-080
–20
09660-077
AMPLITUDE (dBFS)
–40
100
0
AVDD = 3.3V
–0.5dBFS
10Hz INPUT SIGNAL
–50
100
LOOP GAIN (Phase)
AMPLITUDE (dB)
SNR
50
0
–100
–150
–200
–250
–50
–300
GAIN 0
GAIN 1
GAIN 2
–350
100m
09660-078
–100
GAIN 3
GAIN SETTING
DRDY
2
AVDD
2.48V
09660-079
1
A CH1
100
1k
10k
100k
FREQUENCY (Hz)
AVDD = 3.3V
M1.00ms
T 22.1%
10
1M
10M 100M
1G
Figure 52. Open-Loop Phase Response of ADAS1000 Right Leg Drive Amplifier
Without Loading
Figure 49. SNR and THD Across Gain Settings
CH1 2.00V CH2 1.00V
1
09660-081
THD
Figure 50. Power Up AVDD Line to DRDY Going Low (Ready)
Rev. A | Page 24 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
APPLICATIONS INFORMATION
OVERVIEW
The ADAS1000/ADAS1000-1/ADAS1000-2 are electro cardiac
(ECG) front-end solutions targeted at a variety of medical
applications. In addition to ECG measurements, the ADAS1000
version also measures thoracic impedance (respiration) and
detects pacing artifacts, providing all the measured information
to the host controller in the form of a data frame supplying
either lead/vector or electrode data at programmable data rates.
The ADAS1000/ADAS1000-1/ADAS1000-2 are designed to
simplify the task of acquiring ECG signals for use in both
REFIN REFOUT
CAL_DAC_IO
RLD_SJ
RLD_OUT CM_IN
monitor and diagnostic applications. Value-added cardiac
post processing may be executed externally on a DSP,
microprocessor, or FPGA. The ADAS1000/ADAS1000-1/
ADAS1000-2 are designed for operation in both low power,
portable telemetry applications and line powered systems;
therefore, the parts offer power/noise scaling to ensure
suitability to these varying requirements.
The devices also offer a suite of dc and ac test excitation via
a calibration DAC feature and CRC redundancy checks in
addition to read-back of all relevant register address space.
DRIVEN
LEAD
AMP
–
VREF
CALIBRATION
DAC
SHIELD
CM_OUT/WCT
SHIELD
DRIVE
AMP
+
VCM_REF
(1.3V)
ADCVDD, DVDD
1.8V
REGULATORS
ADCVDD
DVDD
ADAS1000
10kΩ
COMMONMODE AMP
10kΩ
IOVDD
+
–
RESPIRATION
DAC
AC
LEAD-OFF
DAC
AVDD
VCM
AC
LEAD-OFF
DETECTION
VREF
PACE
DETECTION
DC LEADOFF/MUXES
ECG PATH
ECG1_LA
AMP
ADC
CS
ECG2_LL
AMP
SCLK
ADC
SDI
ECG3_RA
AMP
ECG4_V1
ADC
AMP
ADC
AMP
ADC
ECG5_V2
FILTERS,
CONTROL,
AND
INTERFACE
LOGIC
PD
RESET
SYNC_GANG
GPIO0/MCS
GPIO1/MSCLK
GPIO2/MSDO
GPIO3
EXT_RESP_LA
EXT_RESP_LL
EXT_RESP_RA
SDO
DRDY
AMP
ADC
MUX
RESPDAC_LA
RESPIRATION PATH
RESPDAC_LL
CLOCK GEN/OSC/
EXTERNAL CLK
SOURCE
CLK_IO
REFGND
AGND
DGND
Figure 53. ADAS1000 Simplified Block Diagram
Rev. A | Page 25 of 80
XTAL1
XTAL2
09660-011
RESPDAC_RA
ADAS1000/ADAS1000-1/ADAS1000-2
CAL_DAC_IO
RLD_SJ
VREF
CALIBRATION
DAC
CM_IN
CM_OUT/WCT
SHIELD
SHIELD
DRIVE
AMP
AVDD
IOVDD
ADCVDD, DVDD
1.8V
REGULATORS
VCM_REF
(1.3V)
ADCVDD
DVDD
ADAS1000-1
COMMONMODE AMP
10kΩ
AC
LEAD-OFF
DAC
RLD_OUT
DRIVEN
LEAD
AMP
VCM
10kΩ
REFIN REFOUT
Data Sheet
VREF
AC
LEAD-OFF
DETECTION
ECG PATH
ECG1_LA
AMP
ADC
CS
ECG2_LL
AMP
SCLK
ADC
SDI
ECG3_RA
DC LEADOFF/MUXES
AMP
ADC
FILTERS,
CONTROL,
AND
INTERFACE
LOGIC
DRDY
PD
RESET
SYNC_GANG
ECG4_V1
AMP
SDO
ADC
GPIO0/MCS
GPIO1/MSCLK
ECG5_V2
GPIO2/MSDO
GPIO3
ADC
CLOCK GEN/OSC/
EXTERNAL CLK
SOURCE
REFGND
AGND
DGND
Figure 54. ADAS1000-1 Simplified Block Diagram
Rev. A | Page 26 of 80
XTAL1 XTAL2
CLK_IO
09660-012
AMP
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
REFIN
REFOUT CAL_DAC_IN
RLD_SJ
AVDD
CM_IN
VREF
10kΩ
COMMONMODE AMP
10kΩ
AC
LEAD-OFF
DAC
ADCVDD, DVDD
1.8V
REGULATORS
IOVDD
ADCVDD
DVDD
ADAS1000-2
VREF
AC
LEAD-OFF
DETECTION
ECG PATH
ECG1
AMP
ADC
CS
ECG2
AMP
SCLK
ADC
SDI
ECG3
DC LEADOFF/MUXES
AMP
ADC
FILTERS,
CONTROL,
AND
INTERFACE
LOGIC
SDO
DRDY
PD
AMP
ADC
RESET
SYNC_GANG
AMP
ADC
GPIO0
GPIO1
ECG4
ECG5
REFGND
AGND
09660-013
GPIO2
GPIO3
DGND
Figure 55. ADAS1000-2 Slave Device Simplified Block Diagram
Rev. A | Page 27 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
ECG INPUTS—ELECTRODES/LEADS
The ADAS1000/ADAS1000-1/ADAS1000-2 ECG product
consists of 5 ECG inputs and a reference drive, RLD (right leg
drive). In a typical 5-lead/vector application, four of the ECG
inputs (ECG3_RA, ECG1_LA, ECG2_LL, ECG4_V1) would
be used in addition to the RLD path. This leaves one spare
ECG path (which could be used for other purposes, such as
calibration or temperature measurement). Both V1 and V2
input channels can be used for alternative measurements if
desired. When used in this way, the negative terminal of the
input stage can be switched to the fixed internal VCM_REF =
1.3 V; see details in Table 48.
In a 5-lead system, the ADAS1000/ADAS1000-1/ADAS1000-2
can provide Lead I, Lead II, and Lead III data or electrode data
directly via the serial interface at all frame rates. The other ECG
leads can be calculated by the user’s software from either the
lead data or the electrode data provided by the ADAS1000/
ADAS1000-1/ADAS1000-2. Note that in 128 kHz data rate, lead
data is only available when configured in analog lead mode as
shown in Figure 57. Digital lead mode is not available for this
data rate.
A 12-lead (10-electrode) system can be achieved using one
ADAS1000 or ADAS1000-1 device ganged together with one
ADAS1000-2 slave device as described in the Gang Mode
Operation section. Here, 9 ECG electrodes and one RLD
electrode achieve the 10 electrode system, again leaving one
spare ECG channel that could be used for alternate purposes as
suggested previously. In such a system, having nine dedicated
electrodes benefits the user by delivering lead information
based on electrode measurements and calculations rather
than deriving leads from other lead measurements.
Table 10 outlines the calculation of the leads (vector) from the
individual electrode measurements.
Table 10. Lead Composition 1
ADAS1000 or ADAS1000-1
12 Leads Achieved by Adding ADAS1000-2 Slave
Lead Name
I
II
III
aVR 2
aVL2
aVF2
V1’
V2’
V3’
V4’
V5’
V6’
Composition
LA – RA
LL – RA
LL – LA
RA – 0.5 × (LA + LL)
LA – 0.5 × (LL + RA)
LL – 0.5 × (LA + RA)
V1 – 0.333 × (LA + RA + LL)
V2 – 0.333 × (LA + RA + LL)
V3 – 0.333 × (LA + RA + LL)
V4 – 0.333 × (LA + RA + LL)
V5 – 0.333 × (LA + RA + LL)
V6 – 0.333 × (LA + RA + LL)
Equivalent
−0.5 × (I + II)
0.5 × (I − III)
0.5 × (II + III)
These lead compositions apply when the master ADAS1000 device is configured into lead mode (analog lead mode or digital lead mode) with VCM = WCT = (RA + LA
+ LL)/3. When configured for 12-lead operation with a master and slave device, the VCM signal derived on the master device (CM_OUT) is applied to the CM_IN of the
slave device. For correct operation of the slave device, the device must be configured in electrode mode (see the FRMCTL register in Table 35).
2
These augmented leads are not calculated within the ADAS1000, but can be derived in the host DSP/microcontroller/FPGA.
1
Rev. A | Page 28 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
The ADAS1000/ADAS1000-1/ADAS1000-2 implementation
uses a dc-coupled approach, which requires that the front end
be biased to operate within the limited dynamic range imposed
by the relatively low supply voltage. The right leg drive loop
performs this function by forcing the electrical average of all
selected electrodes to the internal 1.3 V level, VCM_REF,
maximizing each channel’s available signal range.
ECG CHANNEL
The ECG channel consists of a programmable gain, low noise,
differential preamplifier; a fixed gain anti-aliasing filter; buffers;
and an ADC (see Figure 56). Each electrode input is routed to
its PGA noninverting input. Internal switches allow the PGA’s
inverting inputs to be connected to other electrodes and/or
the Wilson central terminal to provide differential analog
processing (analog lead mode), to a computed average of some
or all electrodes, or the internal 1.3 V common-mode reference
(VCM_REF). The latter two modes support digital lead mode
(leads computed on-chip) and electrode mode (leads calculated
off-chip). In all cases, the internal reference level is removed
from the final lead data.
All ECG channel amplifiers use chopping to minimize 1/f noise
contributions in the ECG band. The chopping frequency of
~250 kHz is well above the bandwidth of any signals of interest.
The 2-pole anti-aliasing filter has ~65 kHz bandwidth to support
digital pace detection while still providing greater than 80 dB of
attenuation at the ADC’s sample rate. The ADC itself is a 14-bit,
2 MHz SAR converter; 1024 × oversampling helps achieve the
required system performance. The ADC’s full-scale input range
is 2 × VREF, or 3.6 V, although the analog portion of the ECG
channel limits the useful signal swing to about 2.8 V.
ADAS1000
AVDD
ELECTRODE
ELECTRODE
EXTERNAL
RFI AND DEFIB
PROTECTION
EXTERNAL
RFI AND DEFIB
PROTECTION
PREAMP
G = 1, 1.5, 2, 3
+
DIFF AMP
BUFFER
FILTER G = 1.4
–
fS VREF
ADC
14
ELECTRODE
VCM
SHIELD DRIVER
Figure 56. Simplified Schematic of a Single ECG Channel
Rev. A | Page 29 of 80
09660-014
PATIENT
CABLE
TO COMMON-MODE AMPLIFIER
FOR DRIVEN LEG AND
SHIELD DRIVER
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Digital Lead Configuration and Calculation
ELECTRODE/LEAD FORMATION AND INPUT
STAGE CONFIGURATION
When the ADAS1000/ADAS1000-1/ADAS1000-2 are
configured for digital lead mode (see the FRMCTL register,
0x0A[4], Table 35), the digital core will calculate each lead
from the electrode signals. This is straightforward for Lead I/
Lead II/Lead III. Calculating V1’ and V2’ requires WCT, which
is also computed internally for this purpose. This mode ignores
the common-mode configuration specified in the CMREFCTL
register (0x05). Digital lead calculation is only available in
2 kHz and 16k Hz data rates.
The input stage of the ADAS1000/ADAS1000-1/ADAS1000-2
can be arranged in several different manners. The input amplifiers are differential amplifiers and can be configured to generate
the leads in the analog domain, before the ADCs. In addition
to this, the digital data can be configured to provide either
electrode or lead format under user control as described in
Table 35. This allows maximum flexibility of the input stage
for a variety of applications.
Single-Ended Input Electrode
Analog Lead Configuration and Calculation
In this mode, the electrode data are digitized relative to the
common-mode signal, VCM, which can be arranged to be any
combination of the contributing ECG electrodes. Commonmode generation is controlled by the CMREFCTL register as
described in Table 30.
Leads are configured in the analog input stage when CHCONFIG
= 1, as shown in Figure 58. This uses a traditional in-amp
structure where lead formation is performed prior to digitization, with WCT created using the common-mode amplifier.
While this results in the inversion of Lead II in the analog
domain, this is digitally corrected so output data have the
proper polarity.
0
0
1
2
3
4
5
6
7
0
0
0
1
MODE
WORD1
WORD2
In this mode, all electrodes are digitized relative to a common
electrode, for example, RA. Standard leads must be calculated by
post processing the output data of the ADAS1000/ADAS1000-1/
ADAS1000-2. See Figure 60.
WORD3
WORD5
LEAD I
SINGLE-ENDED
INPUT, DIGITALLY
CALCULATED LEADS (LA − RA)
LEAD II
V1’
V2’
(LL − RA) (LL − LA)
(V1 − WCT4)
(V2 − WCT4)
LEAD I
COMMON
ELECTRODE (CE)
(LA − RA)
LEADS (HERE RA
ELECTRODE IS
CONNECTED TO THE
CE ELECTRODE
(CM_IN) AND V3 IS ON
ECG3 INPUT)5
LEAD II
V1’
V2’
0
1
0
ANALOG LEADS6
LEAD I
(LA − RA)
1
0
0
SINGLE-ENDED
INPUT ELECTRODE
RELATIVE TO VCM7
1
0
1
LEADS FORMED
RELATIVE TO A
COMMON
ELECTRODE (CE) 5
LEAD III
WORD4
V3’
(LL − RA) (V3 – RA) − (LA − RA) − (LL − RA)
(V1 − RA) − (LA − RA) + (LL − RA)
(V2 − RA) − (LA − RA) + (LL − RA)
3
3
3
LEAD II
LEAD III
(LL − RA) (LL − LA)
V1’
(V1 − VCM)
V2’
(V2 − VCM)
LA − VCM LL − VCM RA − VCM
V1 − VCM
V2 − VCM
LA − CE
V2 − CE
V3 − CE
LL − CE
V1 − CE
REGISTER FRMCTL, BIT DATAFMT: 0 = LEAD/VECTOR MODE; 1 = ELECTRODE MODE.
REGISTER ECGCTL, BIT CHCONFIG: 0 = SINGLE ENDED INPUT (DIGITAL LEAD MODE OR ELECTRODE MODE); 1 = DIFFERENTIAL INPUT (ANALOG LEAD MODE).
REGISTER CMREFCTL, BIT CEREFEN: 0 = CE DISABLED; 1 = CE ENABLED.
WILSON CENTRAL TERMINAL (WCT) = (RA + LA + LL)/3, THIS IS A DIGITALLY CALCULATED WCT BASED ON THE RA, LA, LL MEASUREMENTS.
COMMON ELECTRODE MODE AS SHOWN IN FIGURE 60.
ANALOG LEAD MODE AS SHOWN IN FIGURE 58.
SINGLE-ENDED INPUT ELECTRODE MODE AS SHOWN IN FIGURE 59.
Figure 57. Electrode and Lead Configurations
Rev. A | Page 30 of 80
09660-061
0x0A 0x01 0x05
[4]1 [10]2 [8]3
Common Electrode Configuration
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
CM_OUT/WCT
VCM = WCT = (LA + LL + RA)/3
COMMONMODE AMP
ECGCTL 0x01[10] CHCONFIG = 1
FRMCTL 0x0A[4] DATAFMT = 0
DIFFERENTIAL INPUT – LEAD FORMAT
LEAD I
+
(LA – RA)
ADC
AMP
ECG1_LA
–
LEAD III
(LL – LA)
+
AMP
ECG2_LL
ADC
–
LEAD II
(LL – RA)*
+
AMP
ECG3_RA
ADC
–
*GETS MULITPLED
BY –1 IN DIGITAL
+
AMP
ECG4_V1
ADC
V1’ = V1 – WCT
–
WCT = (LA + LL + RA)/3
+
AMP
ECG5_V2
ADC
V2’ = V2 – WCT
–
WCT = (LA + LL + RA)/3
09660-015
CM_IN
for example RA
COMMON
ELECTRODE CE IN
Figure 58. Flexible Front-End Configuration—Showing Analog Lead Mode Configuration Relative to Wilson Central Terminal (WCT)
VCM
CM_OUT/WCT
COMMONMODE AMP
COMMON MODE
CAN BE ANY
COMBINATION
OF ELECTRODES
ECGCTL 0x01[10] CHCONFIG = 0
FRMCTL 0x0A[4] DATAFMT = 1
SINGLE ENDED ELECTRODE FORMAT
ECG1_LA
+
AMP
ADC
+
AMP
ADC
+
AMP
ADC
+
AMP
ADC
+
AMP
ADC
LA – VCM
–
ECG 2_LL
LL – VCM
–
ECG3_RA
RA – VCM
–
ECG4_V1
V1 – VCM
–
ECG5_V2
V2 – VCM
CM_IN
for example, RA
COMMON
ELECTRODE CE IN
FRMCTL 0x0A[4] = 0 CONFIGURES
SINGLE-ENDED LEAD FORMAT
WHERE LEADS ARE CALCULATED
DIGITALLY AFTER ADC PROCESSING.
Figure 59. Flexible Front-End Configuration—Showing Single-Ended Electrode Configuration
Rev. A | Page 31 of 80
09660-016
–
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
VCM = RA
CM_OUT/WCT
COMMONMODE AMP
ECG1_LA
ECGCTL 0x01[10] CHCONFIG = 0
FRMCTL 0x0A[4] DATAFMT = 1
CMREFCTL 0x05[8] CEREFEN = 1
SINGLE ENDED ELECTRODE FORMAT
+
AMP
ADC
+
AMP
ADC
+
AMP
ADC
+
AMP
ADC
+
AMP
ADC
LA – RA
–
ECG2_LL
–
ECG3_RA = V3
–
ECG4_V1
LL – RA
V3 – RA
V1 – RA
–
–
CM_IN = RA
COMMON
ELECTRODE CE IN
V2 – RA
FRMCTL 0x0A[4] = 0 CONFIGURES
SINGLE-ENDED ELECTRODE FORMAT
WHERE LEADS ARE CALCULATED
DIGITALLY AFTER ADC PROCESSING.
Figure 60. Flexible Front-End Configuration—Showing Common Electrode Configuration
Rev. A | Page 32 of 80
09660-017
ECG5_V2
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
DEFIBRILLATOR PROTECTION
ESIS FILTERING
The ADAS1000/ADAS1000-1/ADAS1000-2 do not include
defibrillation protection on chip. Any defibrillation protection
required by the application requires external components.
Figure 61 and Figure 62 show examples of external defibrillator
protection, which is required on each ECG channel, in
the RLD path and in the CM_IN path if using the CE input
mode. Note that, in both cases, the total ECG path resistance is
assumed to be 5 kΩ. The 22 MΩ resistors shown connected to
RLD are optional and used to provide a safe termination voltage
for an open ECG electrode; they may be larger in value. Note
that, if using these resistors, the dc lead-off feature works best
with the highest current setting.
The ADAS1000/ADAS1000-1/ADAS1000-2 do not include
electrosurgical interference suppression (ESIS) protection on
chip. Any ESIS protection required by the application requires
external components.
As shown in Figure 63, signal paths for numerous functions are
provided on each ECG channel (except respiration, which only
connect to the ECG1_LA, ECG2_LL, and ECG3_RA pins).
Note that the channel enable switch occurs after the RLD
amplifier connection, thus allowing the RLD to be connected
(re-directed into any one of the ECG paths). The CM_IN path
is treated the same as the ECG signals.
500Ω
4kΩ
ELECTRODE
ARGON/NEON
BULB
PATIENT
CABLE
RLD
ECG1
ADAS1000/
ADAS1000-1/
ADAS1000-2
SP724
22MΩ1
500Ω
4kΩ
ELECTRODE
500Ω
AVDD
22MΩ1
500Ω
ECG2
AVDD
ARGON/NEON
BULB
09660-018
PATIENT
CABLE
ECG PATH INPUT MULTIPLEXING
SP724
1OPTIONAL.
Figure 61. Possible Defibrillation Protection on ECG Paths Using Neon Bulbs
PATIENT
CABLE
ELECTRODE
500Ω
4.5kΩ
AVDD
ECG1
22MΩ1
SP7242
ADAS1000/
ADAS1000-1/
ADAS1000-2
RLD
PATIENT
CABLE
22MΩ1
4.5kΩ
500Ω
ELECTRODE
ECG2
AVDD
09660-019
SP7242
1OPTIONAL.
2TWO SP724 CHANNELS PER ELECTRODE MAY PROVIDE BEST PROTECTION.
Figure 62. Possible Defibrillation Protection on ECG Paths Using Diode Protection
RLD AMP
RESPIRATION
INPUT
DCLO
CURRENT
ACLO
CURRENT
CALDAC
INPUT AMPLIFIER
ECG PIN
+
–
CHANNEL
ENABLE
TO
FILTERING
MUX FOR LEAD CONFIG,
COMMON ELECTRODE
+
–
TO CM
AVERAGING
ADAS1000
Figure 63. Typical ECG Channel Input Multiplexing
Rev. A | Page 33 of 80
VCM
FROM CM
AVERAGING
09660-020
1.3V VCM_REF
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
common-mode block. If the physical connection to each
electrode is buffered, these buffers are omitted for clarity.
COMMON-MODE SELECTION AND AVERAGING
The common-mode signal can be derived from any combination of one or more electrode channel inputs, the fixed internal
common-mode voltage reference, VCM_REF, or an external
source connected to the CM_IN pin. One use of the latter
arrangement is in gang mode where the master device creates
the Wilson central terminal for the slave device(s). The fixed
reference option is useful when measuring the calibration DAC
test tone signals or while attaching electrodes to the patient,
where it allows a usable signal to be obtained from just two
electrodes.
There are several restrictions on the use of the switches:
•
•
•
If SW1 is closed, SW7 must be open.
If SW1 is open, at least one electrode switch (SW2 to SW7)
must be closed.
SW7 can be closed only when SW2 to SW6 are open, so
that the 1.3 V VCM_REF gets summed in only when all
ECG channels are disconnected.
The CM_OUT output is not intended to supply current or
drive resistive loads, and its accuracy is degraded if it
is used to drive anything other than the slave ADAS1000-2
devices. An external buffer is required if there is any loading
on the CM_OUT pin.
The flexible common-mode generation allows complete
user control over the contributing channels. It is similar to,
but independent of, circuitry that creates the right leg drive
(RLD) signal. Figure 64 shows a simplified version of the
ADAS1000
CM_IN
SW1
SW2
+
–
ECG1_LA
VCM
CM_OUT
SW3
ECG2_LL
SW4
ECG3_RA
SW5
ECG4_V1
SW6
ECG5_V2
(WHEN SELECTED, IT GETS
SUMMED IN ON EACH ECG CHANNEL)
09660-021
SW7
VCM_REF = 1.3V
Figure 64. Common-Mode Generation Block
Table 11. Truth Table for Common-Mode Selection
ECGCTL
Address
0x01 1
PWREN
0
1
1
1
1
DRVCM
X
X
0
0
0
EXTCM
X
0
0
0
0
LACM
X
0
1
1
1
LLCM
X
0
0
1
1
RACM
X
0
0
0
1
.
1
.
X
.
1
.
X
.
X
.
X
1
2
CMREFCTL Address 0x05 2
V1CM V2CM On Switch
X
X
0
0
SW7
0
0
SW2
0
0
SW2, SW3
0
0
SW2, SW3,
SW4
.
.
.
X
X
SW1
See Table 26.
See Table 30.
Rev. A | Page 34 of 80
Description
Powered down, paths disconnected
Internal VCM_REF = 1.3 V is selected
Internal CM selection: LA contributes to VCM
Internal CM selection: LA and LL contribute to VCM
Internal CM selection: LA, LL, and RA contribute to
VCM (WCT)
.
External VCM selected
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
In some cases, adding lead compensation will prove necessary,
while in others lag compensation may be more appropriate. The
RLD amplifier’s summing junction is brought out to a package
pin (RLD_SJ) to facilitate compensation.
WILSON CENTRAL TERMINAL (WCT)
The flexibility of the common-mode selection averaging allows
the user to achieve a Wilson central terminal voltage from the
ECG1_LA, ECG2_LL, ECG3_RA electrodes.
The RLD amplifier’s short circuit current capability exceeds
regulatory limits. A patient protection resistor is required to
achieve compliance.
RIGHT LEG DRIVE/REFERENCE DRIVE
The right leg drive amplifier or reference amplifier is used as
part of a feedback loop to force the patient’s common-mode
voltage close to the internal 1.3 V reference level (VCM_REF)
of the ADAS1000/ADAS1000-1/ADAS1000-2. This centers all
the electrode inputs relative to the input span, providing
maximum input dynamic range. It also helps to reject noise
and interference from external sources such as fluorescent
lights or other patient-connected instruments, and absorbs
the dc or ac lead-off currents injected on the ECG electrodes.
Within the RLD block, there is lead-off comparator circuitry
that monitors the RLD amplifier output to determine whether
the patient feedback loop is closed. An open-loop condition,
typically the result of the right leg electrode (RLD_OUT)
becoming detached, tends to drive the amplifier’s output
low. This type of fault is flagged in the header word (see
Table 52), allowing the system software to take action by
notifying the user, redirecting the reference drive to another
electrode via the internal switches of the ADAS1000/
ADAS1000-1/ADAS1000-2, or both. The detection circuitry
is local to the RLD amplifier and remains functional with
a redirected reference drive. Table 30 provides details on
reference drive redirection.
The RLD amplifier can be used in a variety of ways as shown in
Figure 65. Its input can be taken from the CM_OUT signal
using an external resistor. Alternatively, some or all of the
electrode signals can be combined using the internal switches.
The DC gain of the RLD amplifier is set by the ratio of the
external feedback resistor (RFB) to the effective input resistor,
which can be set by an external resistor, or alternatively, a
function of the number of selected electrodes as configured in
the CMREFCTL register (see Table 30). In a typical case, using
the internal resistors for RIN, all active electrodes would be used
to derive the right leg drive, resutling in a 2 kΩ effective input
resistor. Achieving a typical dc gain of 40 dB would thus require
a 200 kΩ feedback resistor.
While reference drive redirection may be useful in the event
that the right leg electrode cannot be reattached, some precautions must be observed. Most important is the need for a
patient protection resistor. Because this is an external resistor, it
does not follow the redirected reference drive; some provision
for continued patient protection is needed external to the
ADAS1000/ADAS1000-1/ADAS1000-2. Any additional
resistance in the ECG paths will certainly interfere with
respiration measurement and may also result in an increase in
noise and decrease in CMRR.
The dynamics and stability of the RLD loop depend on the
chosen dc gain and the resistance and capacitance of the
patient cabling. In general, loop compensation using
external components is required, and must be determined
experimentally for any given instrument design and cable set.
The RLD amplifier is designed to stably drive a maximum
capacitance of 5 nF based on the gain configuration (see
Figure 65) and assuming a 330 kΩ patient protection resistor.
EXTERNALLY SUPPLIED COMPONENTS
CZ
TO SET RLD LOOP GAIN
2nF
40kΩ
RIN*
RLD_SJ
100kΩ
RZ
4MΩ
RFB*
RLD_OUT
CM_OUT/WCT
10kΩ
SW2
10kΩ
SW3
10kΩ
SW4
10kΩ
SW5
10kΩ
ELECTRODE LL
ELECTRODE RA
ELECTRODE V1
ELECTRODE V2
SW6
CM_IN OR
CM BUFFER OUT
VCM_REF
(1.3V)
–
SW1
ELECTRODE LA
+
10kΩ
RLD_INT_REDIRECT
*EXTERNAL RESISTOR RIN IS OPTIONAL. IF DRIVING RLD FROM
THE ELECTRODE PATHS, THEN THE SERIES RESISTANCE WILL
CONTRIBUTE TO THE RIN IMPEDANCE. WHERE SW1 TO SW5
ARE CLOSED, RIN = 2kΩ. RFB SHOULD BE CHOSEN
ACCORDINGLY FOR DESIRED RLD LOOP GAIN.
09660-022
ADAS1000
Figure 65. Right Leg Drive—Possible External Component Configuration
Rev. A | Page 35 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
either case, all active channels use the same detection
thresholds.
CALIBRATION DAC
Within the ADAS1000/ADAS1000-1, there are a number of
calibration features.
The 10-bit calibration DAC can be used to correct channel gain
errors (to ensure channel matching) or to provide several test
tones. The options are as follows:
•
•
•
DC voltage output (range: 0.3 V to 2.7 V). The DAC
transfer function for dc voltage output is

code 

0.3 V +  2.4 V× 10
(2 − 1) 

1 mV p-p sine wave of 10 Hz or 150 Hz
1 mV 1 Hz square wave
A lead-off event sets a flag in the frame header word (see
Table 52). Identification of which electrode is off is available
as part of the data frame or as a register read from the leadoff status register (Register LOFF, see Table 45). In the case of
ac lead-off, information about the amplitude of the lead-off
signal(s) can be read back through the serial interface (see
Table 50).
DC Lead-Off Detection
Internal switching allows the calibration DAC signals to be
routed to the input of each ECG channel (see Figure 63).
Alternatively, it can be driven out from the CAL_DAC_IO
pin, enabling measurement and correction for external error
sources in the entire ECG signal chain and/or for use as an
input to the ADAS1000-2 companion chip calibration input.
To ensure a successful update of the calibration DAC (see
Table 34), the host controller must issue four additional SCLK
cycles after writing the new calibration DAC register word.
GAIN CALIBRATION
The gain for each ECG channel can be adjusted to correct for
gain mismatches between channels. Factory trimmed gain
correction coefficients are stored in nonvolatile memory
on-chip for GAIN 0, GAIN 1, and GAIN 2; there is no factory
calibration for GAIN 3. The default gain values can be
overwritten by user gain correction coefficients, which are
stored in volatile memory and available by addressing the
appropriate gain control registers (see Table 49). The gain
calibration applies to the ECG data available on the standard
interface and applies to all data rates.
LEAD-OFF DETECTION
This method injects a small programmable dc current into each
input electrode. When an electrode is properly connected, the
current flows into the right leg (RLD_OUT) and produces a
minimal voltage shift. If an electrode is off, the current charges
that pin’s capacitance, causing the voltage at the pin to float
positive and create a large voltage change that is detected by the
comparators in each channel.
The dc lead-off detection current can be programmed via the
serial interface. Typical currents range from 10 nA to 70 nA in
10 nA steps.
The propagation delay for detecting a dc lead-off event depends
on the cable capacitance and the programmed current. It is
approximately
Delay = Voltage × Cable Capacitance/Programmed Current
For example:
Delay = 1.2 V × (200 pF/70 nA) = 3.43 ms
AC Lead-Off Detection
The alternative method of sensing if the electrodes are
connected to the patient is based on injecting ac currents into
each channel and measuring the amplitudes of the resulting
voltages. The system uses a fixed carrier frequency slightly
above 2 kHz, high enough to be removed by the ADAS1000/
ADAS1000-1/ ADAS1000-2 on-chip digital filters without
introducing phase or amplitude artifacts into the ECG signal.
An ECG system must be able to detect if an electrode is no
longer connected to the patient. The ADAS1000/ADAS1000-1/
ADAS1000-2 support two methods of lead-off detection: either
ac or dc lead-off detection. The two systems are independent
and can be used singly or together under the control of the
serial interface (see Table 27).
The polarity of the ac lead-off signal can be configured on a
per-electrode basis. All electrodes can be driven in phase, or
some can be driven with reversed polarity to minimize the total
injected ac current. Drive amplitude is also programmable. The
propagation delay for detecting an ac lead-off event is <10 ms.
For both ac and dc lead-off detection, the upper and lower
threshold voltages are programmable via Table 37 and Table 38.
Note that these programmed thresholds voltage vary with the
ECG channel gain. The threshold voltages are not affected by
the current level that is programmed.
SHIELD DRIVER
DC lead-off detection uses fixed gain-independent upper
and lower threshold voltages. AC lead-off detection offers
user-programmable thresholds; because the detection is
performed digitally, it may be necessary to adjust the
thresholds, depending on the selected ECG channel gain. In
Note that the ac lead-off function is disabled when the
calibration DAC is enabled.
The shield drive amplifier is a unity gain amplifier. Its purpose
is to drive the shield of the ECG cables. For power consumption
purposes, it can be disabled if not in use. Note that, the SHIELD
pin is shared with the respiration pin function, where it can be
muxed to be one of the pins for external capacitor connection.
If the pin is being used for the respiration feature, the shield
function is not available. In this case, if the application requires
Rev. A | Page 36 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Only one lead measurement can be made at one time. The
respiration measurement path is not suited for use as additional
ECG measurements because the internal configuration and
demodulation do not align with an ECG measurement; however,
the EXT_RESP_LA, EXT_RESP_RA, or EXT_RESP_LL paths
can be multiplexed into one of the ECG ADC paths, if required,
as discussed in the Extend Switch On Respiration Paths section.
a shield drive, an external amplifier connected to the CM_OUT
pin can be used.
RESPIRATION (ADAS1000 MODEL ONLY)
The respiration measurement is performed by driving a high
frequency (programmable from 46.5 kHz to 64 kHz) differential
current into two electrodes; the resulting impedance variation
caused by breathing causes the differential voltage to vary at the
respiration rate. The signal is ac-coupled onto the patient. The
acquired signal is AM, with a carrier at the driving frequency
and a shallow modulation envelope at the respiration frequency.
The modulation depth is greatly reduced by the resistance of the
customer-supplied RFI and ESIS protection filters, in addition
to the impedance of the cable and the electrode to skin interface
(see Table 12). The goal is to measure small ohm variation to sub
ohm resolution in the presence of large series resistance. The
circuit itself consists of a respiration DAC that drives the accoupled current at a programmable frequency onto the chosen
pair of electrodes. The resulting variation in voltage is amplified,
filtered, and synchronously demodulated in the digital domain;
what results is a digital signal that represents the total thoracic
or respiration impedance, including cable and electrode contributions. While it is heavily low-pass filtered on-chip, the user is
required to further process it to extract the envelope and perform
the peak detection needed to establish breathing (or lack thereof).
Internal Respiration Capacitors
The internal respiration function uses an internal RC network
(5 kΩ/100 pF), and this circuit is capable of 200 mΩ resolution
(with up to 5 kΩ total path and cable impedance). The current
is ac-coupled onto the same pins that the measurement is sensed
back on. Figure 66 shows the measurement on Lead I, but,
similarly, the measurement can be configured to measure on
either Lead II or Lead III. The internal capacitor mode requires
no external capacitors and produces currents of ~64 µA p-p
amplitude when configured for maximum amplitude setting
(±1V) through the RESPCTRL register (see Table 28).
Table 12. Maximum Allowable Cable and Thoracic Loading
Cable Resistance
R < 1 kΩ
1 kΩ < R < 2.5 kΩ
2.5 kΩ < R < 5 kΩ
Cable Capacitance
C < 1200 pF
C < 400 pF
C < 200 pF
Rthoracic < 2 kΩ
Respiration measurement is available on one of the leads (Lead I,
Lead II, or Lead III) or on an external path via a pair of dedicated
pins (EXT_RESP_LA, EXT_RESP_RA, or EXT_RESP_LL).
±1V
RESPIRATION DAC
DRIVE +
46.5kHz TO
64kHz
ADAS1000
5kΩ
100pF
CABLE AND ELECTRODE
IMPEDANCE < 5kΩ
FILTER
IN-AMP AND
ANTI-ALIASING
EXT_RESP_LA
LL CABLE
ECG2_LL
OVERSAMPLED
HPF
FILTER
EXT_RESP_LL
RA CABLE
SAR
ADC
LPF
MAGNITUDE
AND
PHASE
ECG3_RA
FILTER
EXT_RESP_RA
100pF
5kΩ
RESPIRATION DAC
DRIVE–
46.5kHz TO
64kHz
±1V
Figure 66. Simplified Respiration Block Diagram
Rev. A | Page 37 of 80
09660-023
LA CABLE
RESPIRATION
MEASURE
ECG1_LA
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
External Respiration Path
External Respiration Capacitors
The EXT_RESP_xx pins are provided for use either with the
ECG electrode cables or, alternatively, with a dedicated external
sensor independent of the ECG electrode path. Additionally,
the EXT_RESP_xx pins are provided such that the user can
measure the respiration signal at the patient side of the RFI/
ESIS protection filters. In this case, the user must take
precautions to protect the EXT_RESP_xx pins from any signals
applied that are in excess of the operating voltage range.
If necessary, the ADAS1000 allows the user to connect external
capacitors into the respiration circuit to achieve higher
resolution (<200 mΩ). This level of resolution requires that the
cable impedance be ≤1 kΩ. The diagram in Figure 67 shows the
connections at RESPDAC_xx paths for the extended respiration
configuration. Again, the EXT_RESP_xx paths can be
connected at the patient side of any filtering circuit; however,
the user must provide protection for these pins. While this
external capacitor mode requires external components, it can
deliver a larger signal-to-noise ratio. Note again that respiration
can be measured on only one lead (at one time); therefore, only
one pair of external respiration paths (and external capacitors)
may be required.
±1V
46.5kHz TO
64kHz
1nF TO 10nF
RESPDAC_LA
RESPIRATION DAC
DRIVE +
ADAS1000
1kΩ
RESPDAC_LL
1kΩ
5kΩ
100pF
MUTUALLY
EXCLUSIVE
CABLE AND ELECTRODE
IMPEDANCE < 1kΩ
RESPIRATION
MEASURE
ECG1_LA
FILTER
IN-AMP AND
ANTI-ALIASING
EXT_RESP_LA
LL CABLE
RA CABLE
ECG2_LL
FILTER
OVERSAMPLED
HPF
EXT_RESP_LL
SAR
ADC
LPF
MAGNITUDE
AND
PHASE
ECG3_RA
FILTER
EXT_RESP_RA
MUTUALLY
EXCLUSIVE
100pF
5kΩ
1nF TO 10nF
RESPDAC_RA
1kΩ
46.5kHz TO
64kHz
RESPIRATION DAC
DRIVE –
±1V
Figure 67. Respiration Measurement Using External Capacitor
Rev. A | Page 38 of 80
09660-024
LA CABLE
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
1nF TO 10nF
RESPDAC_LA
50kHz TO
56kHz
1kΩ 100Ω
RESPIRATION DAC
DRIVE + ve
CABLE AND ELECTRODE
IMPEDANCE < 1kΩ
RESPIRATION
MEASURE
LA CABLE
EXT_RESP_LA
IN-AMP AND
ANTI-ALIASING
10kΩ
RA CABLE
ADAS1000
±1V
OVERSAMPLED
HPF
SAR
ADC
GAIN
10kΩ
LPF
MAGNITUDE
AND
PHASE
EXT_RESP_RA
1/2 OF AD8606
10kΩ
0.9V
1nF TO 10nF RESPDAC_RA
1kΩ 100Ω
46.5kHz TO
64kHz
±1V
09660-025
REFOUT = 1.8V
10kΩ
RESPIRATION DAC
DRIVE – ve
1/2 OF AD8606
Figure 68. Respiration Using External Capacitor and External Amplifiers
If required, further improvements in respiration performance
may be possible with the use of an instrumentation amplifier
and op amp external to the ADAS1000. The instrumentation
amplifier must have sufficiently low noise performance to meet
the target performance levels. This mode uses the external
capacitor mode configuration and is shown in Figure 68. Bit 14
of the RESPCTL register (Table 28) allows the user to bypass
the on-chip amplifier when using an external instrumentation
amplifier.
Respiration Carrier
In applications where an external signal generator is used to
develop a respiration carrier signal, that external signal source
can be synchronized to the internal carrier using the signal
available on GPIO3 when Bit 7, RESPEXTSEL, is enabled in
the respiration control register (see Table 28).
EXTEND SWITCH ON RESPIRATION PATHS
There is additional multiplexing on the external respiration
inputs to allow them to serve as additional electrode inputs
to the existing five ECG ADC channels. This approach allows
a user to configure eight electrode inputs; however, it is not
intended as a true 8-channel/12-lead solution. Time overheads
will be required to reconfigure the multiplexer arrangement
using the serial interface in addition to filter the latency as
described in Table 14.
The user has full control over the SW1/SW2/SW3 configuration
as outlined in Table 48.
SW1a
SW2a
SW3a
SW1b
SW2b
SW3b
EVALUATING RESPIRATION PERFORMANCE
ECG simulators offer a convenient means of studying the
ADAS1000’s performance. While many simulators offer a
variable-resistance respiration capability, care must be taken
when using this feature.
SW1c
SW2c
TO ECG1_LA CHANNEL
TO ECG2_LL CHANNEL
TO ECG3_RA CHANNEL
SW3c
TO ECG4_V1 CHANNEL
SW1e
SW2e
TO ECG5_V2 CHANNEL
SW3e
TO RESPIRATION CIRCUITRY
EXT_RESP_LA
EXT_RESP_RA
EXT_RESP_LL
Rev. A | Page 39 of 80
Figure 69. Alternative Use of the Respiration Paths
09660-032
Some simulators use electrically programmable resistors, often
referred to as digiPOTs, to create the time-varying resistance to
be measured by the respiration function. The capacitances at the
digitPOT's terminals are often unequal and code-dependent, and
these unbalanced capacitances can give rise to unexpectedly
large or small results on different leads for the same programmed
resistance variation. Best results are obtained with a purposebuilt fixture that carefully balances the capacitance presented to
each ECG electrode.
SW1d
SW2d
SW3d
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
PACING ARTIFACT DETECTION FUNCTION
(ADAS1000 ONLY)
The pacing artifact validation function qualifies potential
pacing artifacts and measures the width and amplitude of valid
pulses. These parameters are stored in and available from any of
the pace dataregisters (Address 0x1A, Address 0x3A to Address
0x3C). This function runs in parallel with the ECG channels.
Digital detection is performed using a state machine operating
on the 128 kHz 16-bit data from the ECG decimation chain.
The main ECG signals are further decimated before appearing
in the 2 kHz output stream so that detected pace signals are not
perfectly time-aligned with fully-filtered ECG data. This time
difference is deterministic and may be compensated for.
The pacing artifact validation function can detect and
measure pacing artifacts with widths from 100 μs to 2 ms
and with amplitudes of <400 μV to >1000 mV. Its filters are
designed to reject heartbeat, noise, and minute ventilation
pulses. The flowchart for the pace detection algorithm is
shown in Figure 70.
The ADAS1000 pace algorithm can operate with the ac lead-off
and respiration impedance measurement circuitry enabled.
Once a valid pace has been detected in the assigned leads, the
pace-detected flags appear in the header word (see Table 52) at
the start of the packet of ECG words. These bits indicate that a
pace was qualified. Further information on height and width of
pace is available by reading the contents of Address 0x1A (Register
PACEDATA, see Table 42). This word can be included in the
ECG data packet/frame as dictated by the frame control register
(see Table 35). The data available in the PACEDATA register
is limited to seven bits total for width and height information;
therefore, if more resolution is required on the pace height
and width, this is available by issuing read commands of the
PACExDATA registers (Address 0x3A to Address 0x3C) as
shown in Table 51.
Some users may not wish to use three pace leads for detection.
In this case, Lead II would be the vector of choice
because this lead is likely to display the best pacing artifact. The
other two pace instances can be disabled if not in use.
The on-chip filtering contributes some delay to the pace signal
(see the Pace Latency section).
Choice of Leads
Three identical state machines are available and can be
configured to run on up to three of four possible leads (Lead I,
Lead II, Lead III, and aVF) for pacing artifact detection. All
necessary lead calculations are performed internally and are
independent of EGG channel settings for output data rate, lowpass filter cutoff, and mode (electrode, analog lead, common
electrode). These calculations take into account the available
front-end configurations as detailed in Table 13.
The pace detection algorithm searches for pulses by analyzing
samples in the 128 kHz ECG data stream. The algorithm
searches for an edge, a peak, and a falling edge as defined by
values in the PACEEDGETH, PACEAMPTH, and PACELVLTH
registers, along with fixed width qualifiers. The post-reset
default register values can be overwritten via the SPI bus, and
different values can be used for each of the three pace detection
state machines.
The first step in pace detection is to search the data stream for
a valid leading edge. Once a candidate edge has been detected,
the algorithm begins searching for a second, opposite-polarity
edge that meets with pulse width criteria and passes the
(optional) noise filters. Only those pulses meeting all the
criteria are flagged as valid pace pulses. Detection of a
valid pace pulse sets the flag(s) in the frame header register
and stores amplitude and width information in the PACEDATA
register (Address 0x1A; see Table 42). The pace algorithm looks
for a negative or positive pulse.
Table 13. Pace Lead Calculation
0x01 [10] 1
0
0x05 [8] 2
0
Configuration
Digital leads
0
1
Common
electrode leads
1
X
Analog leads
00
Lead I
(LA − RA)
LA − RA
CH1 − CH3
Lead I
CH1
Lead I
CH1
01
Lead II
(LL − RA)
LL − RA
CH2 − CH3
Lead II
CH2
Lead II
−CH3
0x04 [8:3] 3
10
11
Lead III
aVF
(LL − LA)
(Lead II + Lead III)/2
LL − LA
LL − (LA + RA)/2
CH2 − CH1
CH2 − (CH1 + CH3)/2
lead
II − 0.5 × Lead I
Lead II – Lead I
CH2 − 0.5 × CH1
CH2 − CH1
Lead III
Lead II − 0.5 × Lead I − CH3 − 0.5 × CH1
CH2
Register ECGCTL, Bit CHCONFIG, see Table 26.
Register CMREFCTL, Bit CEREFEN, see Table 30.
3
Register PACECTL, Bit PACExSEL [1:0], see Table 29.
1
2
Rev. A | Page 40 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
START
ENABLE PACE DETECTION
SELECT LEADS
START PACE DETECTION
ALGORITHM
START PULSE
WIDTH TIMER
LOOK FOR
TRAILING EDGE
START NOISE
FILTERS (if enabled)
TRAILING
EDGE
DETECTED?
NO
YES
NOISE FILTER
PASSED?
NO
YES
2ms > PULSE WIDTH
> 100µs
NO
YES
09660-026
FLAG PACE DETECTED
UPDATE REGISTERS WITH
WIDTH AND HEIGHT
Figure 70. Overview of Pace Algorithm
PACE
PULSE
PACELVLTH
LEADING EDGE
LEADING EDGE STOP
PACEAMPTH
PACEEDGETH
RECHARGE
PULSE
Figure 71. Typical Pace Signal
Rev. A | Page 41 of 80
09660-027
PACE WIDTH
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Pace Amplitude Threshold
Pace Validation Filter 2
This register (Address 0x07, see Table 32) can be used to set the
minimum valid pace pulse amplitude:
This filter is also used to reject sub threshold pulses such as
MV pulses and inductive implantable telemetry systems. It is
normally enabled and is controlled via the PACECTL register,
Bit 10 (see Table 29). Filter 2 applies to all leads enabled for pace
detection.
N × VREF
, (corresponds to a
GAIN × 216
range of 20 µV to 5 mV in the ×1.4 gain setting (GAIN 0))
PACEAMPTH setting =
where:
N = 0 to 255 (8 bits), register default N = 0x24, (PACEAMPTH
= 706 μV in the ×1.4 gain setting)
GAIN = 1.4, 2.1, 2.8, or 4.2 (programmable).
VREF = 1.8 V.
This level would typically be set to be the minimum expected
pace amplitude.
For best results under most operating conditions for both
biventricular and unipolar pacing, it is suggested to set the
pace amplitude threshold to a value of approximately 700 μV
to 1 mV. It is not recommended to set the threshold below
250 μV to avoid ambient noise from the patient. The amplitude
may need to be adjusted much higher than 1 mV when other
medical devices are connected to the patient.
Pace Edge Threshold
This programmable level (Address 0x0E, see Table 39) is used to
find a leading edge, signifying the start of a pace pulse:
N × VREF
(corresponds to a
GAIN × 216
range of 20 µV to 5 mV in the ×1.4 gain setting)
PACEEDGETH setting =
where:
If N = 0, PACEEDGETH = PACEAMPTH/2, then N = 0 to 255
(8 bits).
GAIN = 1.4, 2.1, 2.8 or 4.2 (programmable).
VREF = 1.8 V.
Pace Level Threshold
This programmable level (Address 0x0F, see Table 40) is used to
find the leading edge peak:
N × VREF
PACELVLTH setting =
, signed (FF = −1,
GAIN × 216
01 = +1), default = 0
Pace Width Filter
When enabled, this filter searches for an edge of opposite polarity
to the leading edge with a magnitude of at least half the original
trigger. The second edge must be between 100 µs to 2 ms from the
original edge. When a valid pace width is detected, the width is
stored. When disabled only the minimum pulse width of 100 μs
is disabled. This filter is controlled by the PACECTL register, Bit 11
(see Table 29).
BIVENTRICULAR PACERS
As described previously, the pace algorithm expects the pace
pulse to be less than 2 ms wide. In a pacer where both ventricles
are paced, they can be paced simultaneously. Where they fall
within the width and height limits programmed into the
algorithm, a valid pace will be flagged, but only one pace pulse
may be visible.
With the pace width filter enabled, the pace algorithm seeks
pace pulse widths within a 100 μs to 2 ms window. Assuming
that this filter is enabled and in a scenario where two ventricle
pacer pulses fire at slightly different times, resulting in the pulse
showing in the lead as one large, wider pulse, a valid pace is
flagged so long as the total width does not exceed 2 ms.
PACE DETECTION MEASUREMENTS
Design verification of the ADAS1000 digital pace algorithm
includes detection of a range of simulated pace signals in
addition to using the ADAS1000 and evaluation board with
one pacemaker device connected to various simulated loads
(approximately 200 Ω to over 2 kΩ) and covering the following
4 waveform corners.
•
•
Minimum pulse width (100 μs), minimum height (to
<300 μV)
Minimum pulse width (100 μs), maximum height (up to
1.0 V)
Maximum pulse width (2 ms), minimum height (to
<300 μV)
Maximum pulse width (2 ms), maximum height (up to
1.0 V)
where:
N = 0 to 255 (8 bits).
GAIN = 1.4, 2.1, 2.8 or 4.2 (programmable).
VREF = 1.8 V.
•
Pace Validation Filter 1
These scenarios passed with acceptable results. The use of the
ac lead-off function had no obvious impact on the recorded
pace height, width, or the ability of the pace detection algorithm
to identify a pace pulse. The pace algorithm was also evaluated
with the respiration carrier enabled; again, no differences in the
threshold or pacer detect were noted from the carrier.
•
This filter is used to reject sub threshold pulses such as minute
ventilation (MV) pulses and inductive coupled implantable
telemetry systems. It is typically enabled and is controlled via
the PACECTL register, Bit 9 (see Table 29) register. Filter 1
applies to all leads enabled for pace detection.
Rev. A | Page 42 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
delays are summarized in Table 14 and must be taken into
account to enable correct positioning of the pace event relative
to the ECG data.
While these experiments validate the pace algorithm over a
confined set of circumstances and conditions, they do not
replace end system verification of the pacer algorithm. This
can be performed in only the end system, using the system
manufacturer’s specified cables and validation data set.
EVALUATING PACE DETECTION PERFORMANCE
ECG simulators offer a convenient means of studying the
performance and ability of the ADAS1000 to capture pace
signals over the range of widths and heights defined by the
various regulatory standards. While the pace detection algorithm
of the ADAS1000 is designed to conform to medical instrument
standards, some simulators put out signals wider (or narrower)
than called for in the standards, and these will be rejected as
invalid by the algorithm of the ADAS1000.
The pace width acceptance window of the ADAS1000 is the
tightest at the 2 ms limit. If this proves problematic, margin can
be obtained by reducing the master clock frequency. As an
example, using an 8.000 MHz crystal in place of the
recommended 8.192 MHz crystal increases the high limit of the
pace acceptance window from 2.000 ms to 2.048 ms. The low
limit also increases; however, this does not impair the ability of
the algorithm to detect 100 µs pace pulses.
Changing the clock frequency affects all of the other ADAS1000
frequency-related functions. Continuing with the 8.000 MHz
example, the −3 dB frequencies for ECG scales by a factor of
8000/8192, with 40 Hz becoming 39.06 Hz and 150 Hz becoming
146.5 Hz, which are both still well within regulatory requirements.
The respiration and ac leads-off frequencies, as well as the
output data rates, also scale by the same 8000/8192 fraction.
PACE LATENCY
The pace algorithm always examines 128 kHz, 16-bit ECG data,
regardless of the selected frame rate and ECG filter setting. A
pace pulse is qualified when a valid trailing edge is detected and
is flagged in the next available frame header. Pace and ECG data
is always correctly time-aligned at the 128 kHz frame rate, but
the additional filtering inherent in the slower frame rates delays
the frame's ECG data relative to the pace pulse flag. These
There is an inherent one-frame-period uncertainty in the exact
location of the pace trailing edge.
PACE DETECTION VIA SECONDARY SERIAL
INTERFACE (ADAS1000 AND ADAS1000-1 ONLY)
The ADAS1000/ADAS1000-1 provide a second serial interface
for users who wish to implement their own pace detection
schemes. This interface is configured as a master interface. It
provides ECG data at the 128 kHz data rate only. The purpose
of this interface is to allow the user to access the ECG data at a
rate sufficient to allow them to run their own pace algorithm,
while maintaining all the filtering and decimation of the ECG
data that the ADAS1000/ADAS1000-1 offer on the standard
serial interface (2 kHz and 16 kHz data rates). This dedicated
pace interface uses three of the four GPIO pins, leaving one
GPIO pin available even when the secondary serial interface is
enabled. Note that the on-chip digital calibration to ensure
channel gain matching does not apply to data that is available
on this interface. This interface is discussed in more detail in
the Secondary Serial Interface section.
FILTERING
Figure 72 shows the ECG digital signal processing. The ADC
sample rate is programmable. In high performance mode, it
is 2.048 MHz; in low power mode, the sampling rate is reduced
to 1.024 MHz. The user can tap off framing data at one of three
data rates, 128 kHz, 16 kHz, or 2 kHz. Note that although the
data-word width is 24 bits for the 2 kHz and 16 kHz data rate,
the usable bits are 19 and 18, respectively.
The amount of decimation depends on the selected data rate,
with more decimation for the lower data rates.
Four selectable low-pass filter corners are available at the 2 kHz
data rate.
Filters are cleared by a reset. Table 14 shows the filter latencies
at the different data rates.
AC LEAD-OFF
DETECTION
2.048MHz
PACE
DETECTION
128kHz
–3dB AT 13kHz
ACLO
CARRIER
NOTCH
2kHz
AVAILABLE DATA RATE
CHOICE OF 1:
128kHz DATA RATE
16-BITS WIDE
128kHz
16kHz DATA RATE
24-BITS WIDE
18 USABLE BITS
16kHz
–3dB AT 3.5kHz
16kHz
2kHz DATA RATE
24-BITS WIDE
19 USABLE BITS
2kHz
–3dB AT 450Hz
40Hz
150Hz
250Hz
(PROGRAMMABLE BESSEL )
~7Hz
Figure 72. ECG Channel Filter Signal Flow
Rev. A | Page 43 of 80
CALIBRATION
31.25Hz DATA RATE
24-BITS WIDE
~22 USABLE BITS
09660-028
ADC DATA
14-BITS
2.048MHz
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Table 14. Relationship of ECG Waveform to Pace Indication 1, 2, 3
Data Rate
2 kHz
16 kHz
128 kHz
Conditions
450 Hz ECG bandwidth
250 Hz ECG bandwidth
150 Hz ECG bandwidth
40 Hz ECG bandwidth
Apparent Delay of ECG Data Relative to Pace Event 4
0.984 ms
1.915 ms
2.695 ms
7.641 ms
109 μs
0
ECG waveform delay is the time required to reach 50% of final value following a step input.
Guaranteed by design, not subject to production test.
There is an unavoidable residual uncertainty of 8 μs in determining the pace pulse trailing edge.
4
Add 38 μs to obtain the absolute delay for any setting.
1
2
3
Rev. A | Page 44 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
The ADAS1000/ADAS1000-1/ADAS1000-2 have a high
performance, low noise, on-chip 1.8 V reference for use in
the ADC and DAC circuits. The REFOUT of one device is
intended to drive the REFIN of the same device. The internal
reference is not intended to drive significant external current;
for optimum performance in gang operation with multiple
devices, each device should use its own internal reference.
An external 1.8 V reference can be used to provide the
required VREF. In such cases, there is an internal buffer provided for use with external reference. The REFIN pin is a
dynamic load with an average input current of approximately
100 μA per enabled channel, including respiration. When
the internal reference is used, the REFOUT pin requires
decoupling with a10 μF capacitor with low ESR (0.2 Ω
maximum) in parallel with 0.01 μF capacitor to REFGND,
these capacitors should be placed as close to the device pins
as possible and on the same side of the PCB as the device.
GANG MODE OPERATION
While a single ADAS1000 or ADAS1000-1 provides the ECG
channels to support a five-electrode and one-RLD electrode
(or up to 8-lead) system, the device has also been designed so
that it can easily extend to larger systems by paralleling up
multiple devices. In this mode of operation, an ADAS1000 or
ADAS1000-1 master device can easily be operated with one
or more ADAS1000-2 slave devices. In such a configuration,
one of the devices (ADAS1000 or ADAS1000-1) is designated
as master, and any others are designated as slaves. It is
important that the multiple devices operate well together;
with this in mind, the pertinent inputs/outputs to interface
between master and slave devices have been made available.
Note that when using multiple devices, the user must collect
the ECG data directly from each device. If using a traditional
12-lead arrangement where the Vx leads are measured
relative to WCT, the user should configure the ADAS1000
or ADAS1000-1 master device in lead mode with the slave
ADAS1000-2 device configured for electrode mode. The
LSB size for electrode and lead data differs (see Table 41 for
details).
In gang mode, all devices must be operated in the same
power mode (either high performance or low power) and
the same data rate.
Master/Slave
When a device is configured as a slave (ADAS1000-2), the
SYNC_GANG and CLK_IO pins are set as inputs.
Synchronizing Devices
The ganged devices need to share a common clock to ensure
that conversions are synchronized. One approach is to drive
the slave CLK_IO pins from the master CLK_IO pin. Alternatively, an external 8.192 MHz clock can be used to drive
the CLK_IO pins of all devices. The CLK_IO powers up high
impedance until configured in gang mode.
In addition, the SYNC_GANG pin is used to synchronize the
start of the ADC conversion across multiple devices. The
SYNC_GANG pin is automatically driven by the master and
is an input to all the slaves. SYNC_GANG is in high
impedance until enabled via gang mode.
When connecting devices in gang mode, the SYNC_GANG
output is triggered once when the master device starts to
convert. Therefore, to ensure that the slave device(s) receive
this synchronization signal, configure the slave device first for
operation and enable conversions, followed by issuing the
conversion signal to the ECGCTL register in the master
device.
MASTER
SLAVE 0
CLK_OI
SYNC_GANG
CM_OUT
CAL_DAC_IO
CLK_IO
SYNC_GANG
CM_IN
CAL_DAC_IO
SLAVE 1
CLK_IO
SYNC_GANG
CM_IN
CAL_DAC_IO
09660-029
VOLTAGE REFERENCE
Figure 73. Master/Slave Connections in Gang Mode, Using Multiple
ADAS1000/ADAS1000-1/ADAS1000-2 Devices
Calibration
The calibration DAC signal from one device (master) can be
output on the CAL_DAC_IO pin and used as the calibration
input for other devices (slaves) when used in the gang mode
of operation. This ensures that they are all being calibrated
using the same signal which results in better matching across
channels. This does not happen automatically in gang mode
but, rather, must be configured via Table 34.
The ADAS1000 or ADAS1000-1 can be configured as a
master or slave, while the ADAS1000-2 can only be configured as a slave. A device is selected as a master or slave using
Bit 5, master, in the ECGCTL register (see Table 26). Gang
mode is enabled by setting Bit 4, gang, in the same register.
When a device is configured as a master, the SYNC_GANG
pin is automatically set as an output.
Rev. A | Page 45 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Right Leg Drive
The ADAS1000/ADAS1000-1 have a dedicated CM_OUT
pin serving as an output and a CM_IN pin as an input. In
gang mode, the master device determines the common-mode
voltage based on the selected input electrodes. This commonmode signal (on CM_OUT) can then be used by subsequent
slave devices (applied to CM_IN) as the common-mode
reference. All electrodes within the slave device are then
measured with respect to the CM_IN signal from the master
device. See the CMREFCTL register in Table 30 for more
details on the control via the serial interface. Figure 74 shows
the connections between a master and slave device using
multiple ADAS1000/ADAS1000-2 devices.
The right leg drive comes from the master device. If the
internal RLD resistors of the slave device are to contribute
to the RLD loop, tie the RLD_SJ pins of master and slave
together.
REFOUT CAL_DAC_IO
REFIN
RLD_SJ
Common Mode
Sequencing Devices into Gang Mode
When entering gang mode with multiple devices, both
devices can be configured for operation, but the conversion enable bit (ECGCTL register, Bit 2, Table 26) of the
master device should be set after the conversion enable bit
of the slave device. When the master device conversion
signal is set, the master device generates one edge on its
SYNC_GANG pin. This applies to any slave SYNC_GANG
inputs, allowing the devices to synchronize ADC conversions.
CM_IN
CM_OUT/
RLD_OUT
WCT
SHIELD
AVDD
IOVDD
DRIVEN
LEAD AMP
VREF
CALIBRATION
DAC
SHIELD
DRIVE
AMP
VCM_REF
(1.3V)
RESPIRATION
DAC
ADCVDD, DVDD
1.8V
REGULATORS
ADAS1000
LEAD-OFF
DETECTION
10kΩ
PACE
DETECTION
MUXES
CS
SCLK
SDI
SDO
DRDY
SYNC_GANG
5 × ECG PATH
ELECTRODES
×5
AMP
ADC
EXT RESP_LA
AMP
ADC
EXT RESP LL
EXT RESP_RA
RESPIRATION PATH
FILTERS,
CONTROL,
AND
INTERFACE
LOGIC
CLOCK GEN/OSC/
EXTERNAL CLK
SOURCE
XTAL1
REFOUT
CAL_DAC_IN
RLD_SJ
ADCVDD, DVDD
1.8V
REGULATORS
VCM_REF
(1.3V)
TAKE LEAD
DATA
CLK_IO
XTAL2
AVDD
CM_IN
VREF
AC
LEAD-OFF
DAC
DVDD
(optional)
COMMONMODE AMP
AC
LEAD-OFF
DAC
REFIN
ADCVDD
(optional)
IOVDD
ADCVDD
(optional)
DVDD
(optional)
ADAS1000-2
SLAVE
COMMONMODE AMP
LEAD-OFF
DETECTION
5 × ECG PATH
ELECTRODES
×5
AMP
ADC
FILTERS,
CONTROL,
AND
INTERFACE
LOGIC
CLOCK GEN/OSC/
EXTERNAL CLK
SOURCE
CS
SCLK
SDI
SDO
DRDY
SYNC_GANG
TAKE
ELECTRODE
DATA
CLK_IO
Figure 74. Configuring Multiple Devices to Extend Number of Electrodes/Leads
(This Example Uses ADAS1000 as Master and ADAS1000-2 as Slave. Similarly the ADAS1000-1 Could be Use as Master.)
Rev. A | Page 46 of 80
09660-030
PACE
DETECTION
MUXES
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Table 15. Some Possible Arrangements for Gang Operation
Slave 1
ADAS1000-2
ADAS1000-2
ADAS1000-3
ADAS1000-2
ADAS1000-2
ADAS1000-2
Slave 2
ADAS1000-2
Features
ECG, respiration, pace
ECG, respiration, pace
ECG, respiration, pace
ECG
ECG
ECG, respiration, pace
INTERFACING IN GANG MODE
Number of Leads
12-lead + spare ADC channel
15-lead + 3 spare ADC channels
12-lead (derived leads)
12-lead + spare ADC channel
12-lead (derived leads)
12-lead (derived leads)
have the relevant synchronized data. Alternative methods
might use individual controllers for each device or separate
SDO paths.
As shown in Figure 74, when using multiple devices, the
user must collect the ECG data directly from each device.
The example shown in Figure 75 illustrates one possibility
of how to approach interfacing to a master and slave device.
For some applications, digital isolation is required between
the host and the ADAS1000. The example shown illustrates a
means to ensure that the number of lines requiring isolation
is minimized.
Note that SCLK, SDO, and SDI are shared here with
individual CS lines. This requires the user to read the data
on both devices twice as fast to ensure that they can capture
all the data to maintain the chosen data rate and ensure they
MICROCRONTROLLER/
DSP
Number of Electrodes
10 ECG, CM_IN, RLD
15 ECG, CM_IN, RLD
8 ECG, CM_IN, RLD
10 ECG, CM_IN, RLD
8 ECG, CM_IN, RLD
8 ECG, CM_IN, RLD
SCLK
SDI
CS1
CS2
SDO
MASTER
SLAVE
SCLK
SCLK
SDI
SDI
CS
CS
DRDY (OPTIONAL)
DRDY (OPTIONAL)
SDO
SDO
Figure 75. One Method of Interfacing to Multiple Devices
Rev. A | Page 47 of 80
09660-031
Master
ADAS1000
ADAS1000
ADAS1000
ADAS1000-1
ADAS1000-3
ADAS1000-4
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
SERIAL INTERFACES
The ADAS1000/ADAS1000-1 also provide an optional
secondary serial interface that is capable of providing ECG data
at the 128 kHz data rate for users wishing to apply their own
digital pace detection algorithm. This is a master interface that
operates with an SCLK of 20.48 MHz.
STANDARD SERIAL INTERFACE
The standard serial interface is LVTTL-compatible when
operating from a 2.3 V to 3.6 V IOVDD supply. This is the
primary interface for controlling the ADAS1000/ADAS1000-1/
ADAS1000-2, reading and writing registers, and reading frame
data containing all the ECG data-words and other status
functions within the device.
The SPI is controlled by the following five pins:
•
•
•
•
•
CS (frame synchronization input). Asserting CS low
selects the device. When CS is high, data on the SDI
pin is ignored. If CS is inactive, the SDO output driver is
disabled, so that multiple SPI devices can share a common
SDO pin. The CS pin can be tied low to reduce the number
of isolated paths required. When CS is tied low, there is no
frame around the data-words; therefore, the user must be
aware of where they are within the frame. All data-words
with 2 kHz and 16 kHz data rates contain register addresses
at the start of each word within the frame. Users can
resynchronize the interface by holding SDI high for 64
SCLK cycles, followed by a read of any register so that SDI
is brought low for the first bit of the following word.
SDI (serial data input pin): Data on SDI is clocked into the
device on the rising edges of SCLK.
SCLK (clocks data in and out of the device). SCLK should
idle high when CS is high.
SDO (serial data output pin for data readback). Data is
shifted out on SDO on the falling edges of SCLK. The
SDO output driver is high-Z when CS is high.
DRDY (data ready, optional). Data ready when low,
busy when high. Indicates the internal status of the
ADAS1000/ADAS1000-1/ADAS1000-2 digital logic. It is
driven high/busy during reset. If data frames are enabled
and the frame buffer is empty, this pin is driven busy/high.
If the frame buffer is full, this pin is driven low/ready. If
data frames are not enabled, this pin is driven low to
indicate that the device is ready to accept register
read/write commands. When reading packet data, the
entire packet must be read to allow the DRDY return
back high.
MICROCONTROLLER/
DSP
ADAS1000
SCLK
CS
SDI
SDO
DRDY
SCLK
CS
MOSI
MISO
GPIO
09660-033
The ADAS1000/ADAS1000-1/ADAS1000-2 are controlled via
a standard serial interface allowing configuration of registers
and readback of ECG data. This is an SPI-compatible interface
that can operate at SCLK frequencies up to 40 MHz.
Figure 76. Serial Interface
Write Mode
The serial word for a write is 32 bits long, MSB first. The
serial interface works with both a continuous and a burst
(gated) serial clock. The falling edge of CS starts the write
cycle. Serial data applied to SDI is clocked into the ADAS1000/
ADAS1000-1/ADAS1000-2 on rising SCLK edges. At least 32
rising clock edges must be applied to SCLK to clock in 32 bits of
data before CS is taken high again. The addressed input register
is updated on the rising edge of CS. For another serial transfer
to take place, CS must be taken low again. Register writes are
used to configure the device. Once the device is configured and
enabled for conversions, frame data can be initiated to start
clocking out ECG data on SDO at the programmed data rate.
Normal operation for the device is to send out frames of ECG
data. Typically, register reads and writes should be needed only
during start-up configuration. However, it is possible to write
new configuration data to the device while in framing mode.
A new write command is accepted within the frame and,
depending on the nature of the command, there may be a need
to flush out the internal filters (wait periods) before seeing
usable framing data again.
Write/Read Data Format
Address, data, and the read/write bits are all in the same word.
Data is updated on the rising edge of CS or the first cycle of the
following word. For all write commands to the ADAS1000/
ADAS1000-1/ADAS1000-2, the data-word is 32 bits, as shown
in Table 16. Similarly, when using data rates of 2 kHz and
16 kHz, each word is 32 bits (address bits and data bits).
Table 16. Serial Bit Assignment (Applies to All Register
Writes, 2 kHz and 16 kHz Reads)
B31
R/W
[B30:B24]
Address bits[6:0]
[B23:B0]
Data bits [23:0] (MSB first)
For register reads, data is shifted out during the next word, as
shown in Table 17.
Table 17. Read/Write Data Stream
Digital
Pin
SDI
SDO
Rev. A | Page 48 of 80
Command 1
Read Address 1
Command 2
Read Address 2
Address 1
Read Data 1
Command 2
Write Address 3
Address 2
Read Data 2
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Read Mode
In the 128 kHz data rate, all write words are still 32-bit writes
but the read words in the data packet are now 16 bits (upper
16 bits of register). There are no address bits, only data bits.
Register space that is larger than 16 bits spans across 2 ×
16-bit words (for example, pace and respiration).
Although the primary reading function within the ADAS1000/
ADAS1000-1/ADAS1000-2 is the output of the ECG frame
data, the devices also allow reading of all configuration registers. To read a register, the user must first address the device
with a read command containing the particular register address.
If the device is already in data framing mode, the read register
command can be interleaved between the frames by issuing a
read register command during the last word of frame data.
Data shifted out during the next word is the register read data.
To return to framing mode, the user must re-enable framing
by issuing a read of the frame header register (Address 0x40)
(see Table 52). This register write can be used to flush out the
register contents from the previous read command.
Data Frames/Packets
The general data packet structure is shown in Table 18. Data
can be received in two different frame formats. For the 2 kHz
and 16 kHz data rates, a 32-bit data format is used (where the
register address is encapsulated in the upper byte, identifying
the word within the frame) (see Table 20). For the 128 kHz data
rate, words are provided in 16-bit data format (see Table 21).
When the configuration is complete, the user can begin reading
frames by issuing a read command to the frame header register
(see Table 52). The ADAS1000/ADAS1000-1/ADAS1000-2
continue to make frames available until another register address
is written (read or write command). To continue reading frame
data, continue to write all zeros on SDI, which is a write of the
NOP register (Address 0x00). A frame is interrupted only when
another read or write command is issued.
Table 18. Example of Reading Registers and Frames
SDI
…..
NOP
SDO
…..
Frame
data
Read
Address
N
Frame
CRC
Read
frames
NOP
NOP
…..
Register
Data N
Frame
header
Frame
data
…..
Regular register reads are always 32 bits long and MSB first.
Each frame can be a large amount of data plus status words. CS
can toggle between each word of data within a frame, or it can
be held constantly low during the entire frame.
Serial Clock Rate
The SCLK can be up to 40 MHz, depending on the IOVDD
voltage level as shown in Table 5. The minimum SCLK
frequency is set by the requirement that all frame data be
clocked out before the next frame becomes available.
By default, a frame contains 11 × 32 bit words when reading at
2 kHz or 16 kHz data rates; similarly, a frame contains 13 × 16bit words when reading at 128 kHz. The default frame
configuration does not include the optional respiration phase
word; however, this word can be included as needed.
Additionally any words not required can be excluded from the
frame. To arrange the frame with the words of interest,
configure the appropriate bits in the frame control register (see
Table 35). The complete set of words per frame are 12 × 32-bit
words for the 2 kHz or 16 kHz data rates, or 15 × 16-bit words
at 128 kHz.
SCLK (min) = frame_rate × words_per_frame × bits_per_word
The minimum SCLK for the various frame rates is shown in
Table 19.
Table 19. SCLK Clock Frequency vs. Packet Data/Frame Rates
Frame
Rate
128 kHz
16 kHz
2 kHz
Any data not available within the frame can be read between
frames. Reading a register interrupts the frame and requires the
user to issue a new read command of Address 0x40 (see
Table 52) to start framing again.
1
Word
Size
16 bits
32 bits
32 bits
Maximum
Words/Frame 1
15 words
12 words
12 words
Minimum
SCLK
30.72 MHz
6.14 MHz
768 kHz
This is the full set of words that a frame contains. It is programmable and can
be configured to provide only the words of interest. See Table 35.
Table 20. Default 2 kHz and 16 kHz Data Rate: 32-Bit Frame Word Format
Register
Address
Header
0x40
Lead I/LA
0x11
Lead II/LL
0x12
Lead III/RA
0x13
V1’/V1
0x14
V2’/V2
0x15
PACE
0x1A
RESPM
0x1B
RESPPH
0x1C
LOFF
0x1D
GPIO
0x06
CRC
0x41
Table 21. Default 128 kHz Data Rate: 16-Bit Frame Word Format 1
Register
Address
1
Header
0x40
Lead I/LA
0x11
Lead II/LL
0x12
Lead III/RA
0x13
V1’/V1
0x14
V2’/V2
0x15
PACE1 PACE2
0x1A
Respiration phase words (2x 16-bit words) are not shown in this frame, but could be included.
Rev. A | Page 49 of 80
RESPM1 RESPM2
0x1B
LOFF
0x1D
GPIO
0x06
CRC
0x41
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Internal operations are synchronized to the internal master
clock at either 2.048 MHz or 1.024 MHz (ECGCTL[3]: HP = 1
and HP = 0, respectively, see Table 26). Because there is no
guaranteed relationship between the internal clock and the SPI's
SCLK signal, an internal handshaking scheme is used to ensure
safe data transfer between the two clock domains. A full
handshake requires three internal clock cycles and imposes an
upper speed limit on the SCLK frequency when reading frames
with small word counts. This is true for all data frame rates.
SCLK (max) = (1.024 MHz × (1 + HP) × words_per_frame ×
bits_per_word)/3; or 40 MHz, whichever is lower.
Exceeding the maximum SCLK frequency for a particular
operating mode causes erratic behavior in the DRDY signal
and results in the loss of data.
Data Rate and Skip Mode
When reading packets of data, the entire data packet must be
read; otherwise, DRDY stays low.
There are three methods of detecting DRDY status.
•
•
•
Although the standard frame rates available are 2 kHz, 16 kHz,
and 128 kHz, there is also a provision to skip frames to further
reduce the data rate. This can be configured in the frame
control register (see Table 35).
Data Ready (DRDY)
The DRDY pin is used to indicate that a frame composed of
decimated data at the selected data rate is available to read. It
is high when busy and low when ready. Send commands only
when the status of DRDY is low or ready. During power-on, the
status of DRDY is high (busy) while the device initializes itself.
When initialization is complete, DRDY goes low and the user
can start configuring the device for operation. When the device
is configured and enabled for conversions by writing to the
conversion bit (CNVEN) in the ECGCTL register, the ADCs
start to convert and the digital interface starts to make data
available, loading them into the buffer when ready. If
conversions are enabled and the buffer is empty, the device is
not ready and DRDY goes high. Once the buffer is full, DRDY
goes low to indicate that data is ready to be read out of the
device. If the device is not enabled for conversions, the DRDY
ignores the state of the buffer full status.
DRDY pin. This is an output pin from the ADAS1000/
ADAS1000-1/ADAS1000-2 that indicates the device read
or busy status. No data is valid while this pin is high.
The DRDY signals that data is ready to be read by driving
low and remaining low until the entire frame has been
read. It is cleared when the last bit of the last word in the
frame is clocked onto SDO. The use of this pin is optional.
SDO pin. The user can monitor the voltage level of the
SDO pin by bringing CS low. If SDO is low, data is ready;
if high, busy. This does not require clocking the SCLK
input. (CPHA = CPOL = 1 only).
One of the first bits of valid data in the header word
available on SDO is a data ready status bit (see Table 41).
Within the configuration of the ADAS1000/ADAS1000-1/
ADAS1000-2, the user can set the header to repeat until
the data is ready. See Bit 6 (RDYRPT) in the frame control
register in Table 35.
Detecting Missed Conversion Data
To ensure that the current data is valid, the entire frame must
be read at the selected data rate. If a read of the entire frame
takes longer than the selected data rate allows, the internal
buffer is not loaded with the latest conversion data. The frame
header register (see Table 52) provides four settings to indicate
an overflow of frame data. The settings of Bits[29:28] report
how many frames have been missed since the last valid frame
read. A missed frame may occur as a result of the last read
taking too long. The data in the current frame is valid data, but
it is not the current data. It is the calculation made directly after
the last valid read.
To clear such an overflow, the user must read the entire frame.
Rev. A | Page 50 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
CRC Word
ADAS1000
Framed data integrity is provided by CRCs. For the 128 kHz
frame rates, the 16-bit CRC-CCITT polynomial is used. For the
2 kHz and 16 kHz frame rates, the 24-bit CRC polynomial used.
XTAL1
CLK_IO
09660-034
In both cases, the CRC residue is preset to all 1s and inverted
before being transmitted. The CRC parameters are summarized
in Table 22. To verify that data was correctly received, software
should compute a CRC on both the data and the received
checksum. If data and checksum are received correctly, the
resulting CRC residue should equal the check constant shown
in Table 22. Note that data is shifted through the generator
polynomial MSB first, the same order that it is shifted out
serially. The bit and byte order of the CRC that is appended to
the frame is such that the MSB of the CRC is shifted through
the generator polynomial first in the same order as the data so
that the CRC residue XOR’d with the inverted CRC at the end
of the frame is all 1s (which is why the check constant is
identical for all messages). The CRC is based only on the data
that is sent out.
XTAL2
Figure 77. Input Clock
Clocks
The ADAS1000/ADAS1000-1/ADAS1000-2 run from an
external crystal or clock input frequency of 8.192 MHz.
The external clock input is provided for use in gang mode
so conversions between the two devices are synchronized.
In this mode, the CLK_IO pin is an output from the master
and an input from the slave. To reduce power, the CLK_IO is
disabled when not in gang mode. All features within the
ADAS1000 are a function of the frequency of the externally
applied clock. Using a frequency other than the 8.192 MHz
previously noted causes scaling of the data rates, filter corners,
ac lead off frequency, respiration frequency, and pace algorithm
corners accordingly.
Table 22. CRC Polynomials
Frame Rate
2 kHz, 16 kHz
128 kHz
CRC Size
24 bits
16 bits
Polynomial
x24 + x22 + x20 + x19 + x18 + x16 + x14 + x13 + x11 + x10 + x8 + x7 + x6 + x3 + x1 + x0
x16 + x12 + x5 + x0
Rev. A | Page 51 of 80
Polynomial
in Hex
0x15D6DCB
0x11021
Check Constant
0x15A0BA
0x1D0F
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
SECONDARY SERIAL INTERFACE
This second serial interface is an optional interface that can
be used for the user’s own pace detection purposes. This
interface contains ECG data at 128 kHz data rate only. If using
this interface, the ECG data is still available on the standard
interface discussed previously at lower rates with all the
decimation and filtering applied. If this interface is inactive,
it draws no power.
Data is available in 16-bit words, MSB first.
This interface is a master interface, with the ADAS1000/
ADAS1000-1 providing the SCLK, CS, SDO. Is it shared
across some of the existing GPIO pins as follows:
•
•
•
This interface can be enabled via the GPIO register (see
Table 31).
CS
MISO/GPIO
ADAS1000
MASTER SPI
MSCLK/GPIO1
MCS/GPIO0
MSDO/GPIO2
09660-035
SCLK
Table 23. Master SPI Frame Format; All Words are 16 Bits
Header
Lead
I/LA
Lead
II/LL
Lead
III/RA
V1’/V1
V2’/V2
CRC
The header word consists of four bits of all 1s followed by a 12bit sequence counter. This sequence counter increments after
every frame is sent, thereby allowing the user to tell if any
frames have been missed and how many.
RESET
GPIO1/MSCLK
GPIO0/MCS
GPIO2/MSDO
MICROCONTROLLER/
DSP
on the falling edge of MSCLK. MSCLK idles high when MCS is
deasserted.
Figure 78. Master SPI Interface for External Pace Detection Purposes
The data format of the frame starts with a header word and five
ECG data-words, and completes with the same CRC word as
documented in Table 22 for the 128 kHz rate. All words are
16 bits. MSCLK runs at approximately 20 MHz and MCS is
asserted for the entire frame with the data available on MSDO
There are two methods of resetting the ADAS1000/ADAS1000-1/
ADAS1000-2 to power-on default. Bringing the RESET line low
or setting the SWRST bit in the ECGCTL register (Table 26)
resets the contents of all internal registers to their power-on
reset state. The falling edge of the RESET pin initiates the reset
process; DRDY goes high for the duration, returning low when
the RESET process is complete. This sequence takes 1.5 ms
maximum. Do not write to the serial interface while DRDY is
high handling a RESET command. When DRDY returns low,
normal operation resumes and the status of the RESET pin is
ignored until it goes low again. Software reset using the SWRST
bit (see Table 26) requires that a NOP (no operation) command
be issued to complete the reset cycle.
PD FUNCTION
The PD pin powers down all functions in low power mode.
The digital registers maintain their contents. The power-down
function is also available via the serial interface (ECG control
register, see Table 26).
Rev. A | Page 52 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
SPI OUTPUT FRAME STRUCTURE (ECG AND STATUS DATA)
Three data rates are offered for reading ECG data: low speed 2 kHz/16 kHz rates for electrode/lead data (32-bit words) and a high speed
128 kHz for electrode/lead data (16-bit words).
DRDY
CS1
EACH SCLK WORD IS 32 CLOCK CYCLES
1
2
3
4
5
6
7
8
9
10
11
SCLK
DRIVEN OUTPUT DATA STREAM
ORD
GPIO
ANOTHER FRAME OF DATA
CRC
W
LEA
D-OF
F
RES
P
MAG IRATIO
NITU N
DE
2
PAC
E DE
TEC
TION
1
V1’/V
V2’/V
RA
LEA
D III/
L
A
LEA
D II/L
LEA
D I/L
HEA
DER
SDO2
32-BIT
DATA WORDS
1 CS MAY BE USED IN ONE OF THE FOLLOWING WAYS:
09660-036
A) HELD LOW ALL THE TIME.
B) USED TO FRAME THE ENTIRE PACKET OF DATA.
C) USED TO FRAME EACH INDIVIDUAL 32-BIT WORD.
2 SUPER SET OF FRAME DATA, WORDS MAY BE EXCLUDED.
Figure 79. Output Frame Structure for 2 kHz and 16 kHz Data Rates with SDO Data Configured for Electrode or Lead Data
DRDY
CS1
EACH SCLK WORD IS 16 CLOCK CYCLES
1
2
3
4
5
6
8
7
9
10
11
12
13
SCLK
DRIVEN OUTPUT DATA STREAM
ORD
ANOTHER FRAME
CRC
W
GPIO
LEA
D-OF
F
RES
P
MAG IRATIO
NITU N
DE
PAC
E
V2
V1
RA
LL
LA
HEA
DER
SDO2
16-BIT
DATA WORDS
1 CS
09660-037
MAY BE USED IN ONE OF THE FOLLOWING WAYS:
A) HELD LOW ALL THE TIME.
B) USED TO FRAME THE ENTIRE PACKET OF DATA.
C) USED TO FRAME EACH INDIVIDUAL 16-BIT WORD.
2 SUPER SET OF FRAME DATA, WORDS MAY BE EXCLUDED.
Figure 80. Output Frame Structure for 128 kHz Data Rate with SDO Data Configured for Electrode Data
(The 128 kHz Data Rate Can Provide Single-Ended Electrode Data or Analog Lead Mode Data Only. Digital Lead Mode Is Not Available at 128 kHz Data Rate.)
Rev. A | Page 53 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
SPI REGISTER DEFINITIONS AND MEMORY MAP
In 2 kHz and 16 kHz data rates, data takes the form of 32-bit words. Bit A6 to Bit A0 serve as word identifiers. Each 32-bit word has 24
bits of data. A third high speed data rate is also offered: 128 kHz with data in the form of 16-bit words (all 16 bits as data).
Table 24. SPI Register Memory Map
R/W121
R
R/W
R/W
R/W
R/W
R/W
R/W
R/W
A[6:0]
0x00
0x01
0x02
0x03
0x04
0x05
0x06
0x07
D[23:0]
XXXXXX
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
Register Name
NOP
ECGCTL
LOFFCTL
RESPCTL
PACECTL
CMREFCTL
GPIOCTL
PACEAMPTH
Table
Table 26
Table 27
Table 28
Table 29
Table 30
Table 31
Table 32
R/W
R/W
R/W
R/W
R/W
R/W
R/W
0x08
0x09
0x0A
0x0B
0x0C
0x0D
0x0E
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
TESTTONE
CALDAC
FRMCTL
FILTCTL
LOFFUTH
LOFFLTH
PACEEDGETH
Table 33
Table 34
Table 35
Table 36
Table 37
Table 38
Table 39
R/W
0x0F
dddddd
PACELVLTH
Table 40
R
R
R
R
R
R
0x11
0x12
0x13
0x14
0x15
0x1A
XXXXXX
XXXXXX
XXXXXX
XXXXXX
XXXXXX
XXXXXX
LADATA
LLDATA
RADATA
V1DATA
V2DATA
PACEDATA
Table 41
Table 41
Table 41
Table 41
Table 41
Table 42
R
0x1B
XXXXXX
RESPMAG
Table 43
Register Description
NOP (no operation)
ECG control
Lead-off control
Respiration control2
Pace detection control
Common-mode, reference, and shield drive control
GPIO control
Pace amplitude threshold2
Test tone
Calibration DAC
Frame control
Filter control
AC lead-off upper threshold
AC lead-off lower threshold
Reset Value
0x000000
0x000000
0x000000
0x000000
0x000F88
0xE00000
0x000000
0x242424
0x000000
0x002000
0x079000
0x000000
0x00FFFF
0x000000
0x000000
Pace edge threshold2
Pace level threshold2
LA or Lead I data
LL or Lead II data
RA or Lead III data
V1 or V1’ data
V2 or V2’ data
0x000000
Read pace detection data/status2
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
Read respiration data—magnitude2
0x000000
2
0x000000
R
0x1C
XXXXXX
RESPPH
Table 44
R
R
R
R/W
R/W
R/W
R/W
R/W
R/W
R
R
R
R
R
R
0x1D
0x1E
0x1F
0x20
0x21
0x22
0x23
0x24
0x25
0x31
0x32
0x33
0x34
0x35
0x3A
XXXXXX
XXXXXX
XXXXXX
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
dddddd
LOFF
DCLEAD-OFF
OPSTAT
EXTENDSW
CALLA
CALLL
CALRA
CALV1
CALV2
LOAMLA
LOAMLL
LOAMRA
LOAMV1
LOAMV2
PACE1DATA
Table 45
Table 46
Table 47
Table 48
Table 49
Table 49
Table 49
Table 49
Table 49
Table 50
Table 50
Table 50
Table 50
Table 50
Table 51
Pace1 width and amplitude2
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
0x000000
R
0x3B
dddddd
PACE2DATA
Table 51
Pace2 width and amplitude2
0x000000
R
0x3C
dddddd
PACE3DATA
Table 51
2
0x000000
R
R
x
0x40
0x41
Other
dddddd
XXXXXX
XXXXXX
FRAMES
CRC
Reserved13
Table 52
Table 53
Read respiration data—phase
Lead-off status
DC lead-off
Operating state
Extended switch for respiration inputs
User gain calibration LA
User gain calibration LL
User gain calibration RA
User gain calibration V1
User gain calibration V2
Lead-off amplitude for LA
Lead-off amplitude for LL
Lead-off amplitude for RA
Lead-off amplitude for V1
Lead-off amplitude for V2
Pace3 width and amplitude
Frame header
Frame CRC
Reserved
1
0x800000
0xFFFFFF
XXXXXX
R/W = register both readable and writable; R = read only.
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
3
Reserved bits in any register are undefined. In some cases a physical (but unused) memory bit may be present—in other cases not. Do not issue commands to
reserved registers/space. Read operations of unassigned bits are undefined.
2
Rev. A | Page 54 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
CONTROL REGISTERS DETAILS
For each register address, the default setting is noted in a default column in addition to being noted in the function column by “(default)”;
this format applies throughout the register map.
Table 25. Serial Bit Assignment
B31
R/W
[B30:B24]
Address bits
[B23:B0]
Data bits (MSB first)
Table 26. ECG Control Register (ECGCTL) Address 0x01, Reset Value = 0x000000
R/W
R/W
R/W
R/W
R/W
R/W
R
R/W
Default
0
0
0
0
0
0
0
Bit
23
22
21
20
19
[18:11]
10
Name
LAEN
LLEN
RAEN
V1EN
V2EN
Reserved
CHCONFIG
R/W
00
[9:8]
GAIN [1:0]
R/W
0
7
VREFBUF
R/W
0
6
CLKEXT
R/W
0
5
Master
R/W
0
4
Gang
R/W
0
3
HP
R/W
0
2
CNVEN
R/W
0
1
PWREN
R/W
0
0
SWRST
Function
ECG channel enable; shuts down power to the channel; the input becomes high-Z.
0 (default) = disables ECG channel. When disabled, the entire ECG channel is shut down and
dissipating minimal power.
1 = enables ECG channel.
Reserved, set to 0.
Setting this bit selects the differential analog front end (AFE) input. See Figure 57.
0 (default) = single-ended input (digital lead mode or electrode mode).
1 = differential input (analog lead mode).
Preamplifier and anti-aliasing filter overall gain.
00 (default) = GAIN 0 = ×1.4.
01 = GAIN 1 = ×2.1.
10 = GAIN 2 = ×2.8.
11 = GAIN 3 = ×4.2 (user gain calibration is required for this gain setting).
VREF buffer enable.
0 (default) = disabled.
1 = enabled (when using the internal VREF, VREFBUF must be enabled).
Use external clock instead of crystal oscillator. The crystal oscillator is automatically disabled if
configured as a slave in gang mode and the slave device should receive the clock from the master
device.
0 (default) = XTAL is clock source.
1 = CLK_IO is clock source.
In gang mode, this bit selects the master (SYNC_GANG pin is configured as an output). When in single
channel mode (gang = 0), this bit is ignored. ADAS1000-2 cannot be configured as a master device.
0 (default) = slave.
1 = master.
Enable gang mode. Setting this bit causes CLK_IO and SYNC_GANG to be activated.
0 (default) = single channel mode.
1 = gang mode.
Selects the noise/power performance. This bit controls the ADC sampling frequency. See the
Specifications section for further details.
0 (default) = 1 MSPS, low power.
1 = 2 MSPS, high performance/low noise.
Conversion enable. Setting this bit enables the ADC conversion and filters.
0 (default) = idle.
1 = conversion enable.
Power enable. Clearing this bit powers down the device. All analog blocks are powered down and the
external crystal is disabled. The register contents are retained during power down as long as DVDD is
not removed.
0 (default) = power down.
1 = power enable.
Software reset. Setting this bit clears all registers to their reset value. This bit automatically clears itself.
The software reset requires a NOP command to complete the reset.
0 (default) = NOP.
1 = reset.
Rev. A | Page 55 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Table 27. Lead-Off Control Register (LOFFCTL) Address 0x02, Reset Value = 0x000000
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R
R/W
Default
0
0
0
0
0
0
0
0
0
0
0
0
0
00
Bit
23
22
21
20
19
18
17
16
15
14
13
12
[11:9]
[8:7]
Name
LAPH
LLPH
RAPH
V1PH
V2PH
CEPH
LAACLOEN
LLACLOEN
RAACLOEN
V1ACLOEN
V2ACLOEN
CEACLOEN
Reserved
ACCURRENT
R/W
00
000
[6:5]
[4:2]
Reserved
DCCURRENT
R/W
0
1
ACSEL
R/W
0
0
LOFFEN
Function
AC lead-off phase.
0 (default) = in phase.
1 = 180° out of phase.
Individual electrode ac lead-off enable. AC lead-off enables are the OR of ACSEL and the
individual ac lead-off channel enables.
0 (default) = ac lead-off disabled.
1 = ac lead-off enabled.
Reserved, set to 0.
Set current level for ac lead-off.
00 (default) = 12.5 nA rms.
01 = 25 nA rms.
10 = 50 nA rms.
11 = 100 nA rms.
Reserved, set to 0.
Set current level for dc lead-off (active only for ACSEL = 0).
000 (default) = 0 nA.
001 = 10 nA.
010 = 20 nA.
011 = 30 nA.
100 = 40 nA.
101 = 50 nA.
110 = 60 nA.
111 = 70 nA.
DC or AC (out-of-band) lead-off detection.
ACSEL acts as a global ac lead-off enable for RA, LL, LA, V1, V2 electrodes (CE ac lead-off is not
enabled using ACSEL). AC lead-off enables are the OR of ACSEL and the individual ac lead-off
channel enables.
If LOFFEN = 0, this bit is don’t care.
If LOFFEN = 1,
0 (default) = dc lead-off detection enabled. (Individual ac lead-off can be enabled through
Bits[17:12].)
1 = dc lead-off detection disabled. AC lead-off detection enabled (all electrodes except CE
electrode).
When the calibration DAC is enabled, ac lead-off is disabled.
Enable lead-off detection.
0 (default) = lead-off disabled.
1 = lead-off enabled.
Rev. A | Page 56 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Table 28. Respiration Control Register (RESPCTL) Address 0x03, Reset Value = 0x000000 1
R/W
Default
R/W
0
Bit
[23:17]
16
Name
Reserved
RESPALTFREQ
R/W
0
15
RESPEXTSYNC
R/W
0
14
RESPEXTAMP
R/W
0
13
RESPOUT
R/W
0
12
RESPCAP
R/W
0000
[11:8]
RESPGAIN [3:0]
R/W
0
7
RESPEXTSEL
R/W
00
[6:5]
RESPSEL [1:0]
R/W
00
[4:3]
RESPAMP
R/W
00
[2:1]
RESPFREQ
R/W
0
0
RESPEN
1
Function
Reserved, set to 0.
Setting this bit to 1 makes the respiration waveform on the GPIO3 pin periodic every cycle. Use in
conjunction with RESFREQ to select drive frequency.
0 (default) = periodic every N cycles (default).
1 = periodic every cycle.
Set this bit to 1 to drive the MSB of the respiration DAC out onto the GPIO3 pin. This signal can be
used to synchronize an external generator to the respiration carrier. It is a constant period only
when RESPALTFREQ = 1.
0 (default) = normal GPIO3 function.
1 = MSB of RESPDAC driven onto GPIO3 pin.
For use with an external instrumentation amplifier with respiration circuit. Bypasses the on-chip
amplifier stage and input directly to the ADC. See Figure 68.
0 (default) = disabled
1 = enabled
Selects external respiration drive output. RESPDAC_RA is automatically selected when RESPCAP = 1
0 (default) = RESPDAC_LL and RESPDAC_RA.
1 = RESPDAC_LA and RESPDAC_RA.
Selects source of respiration capacitors.
0 (default) = use internal capacitors.
1 = use external capacitors.
Respiration in amp gain (saturates at 10).
0000 (default) = ×1 gain.
0001 = ×2 gain.
0010 = ×3 gain.
…
1000 = ×9 gain.
1001 = ×10 gain.
11xx = ×10 gain.
Selects between EXT_RESP _LA or EXT_RESP_LL paths. Applies only if the external respiration is
selected in RESPSEL. EXT_RESP_RA is automatically enabled.
0 (default) = EXT_RESP_LL.
1 = EXT_RESP_LA.
Set leads for respiration measurement.
00 (default) = Lead I.
01 = Lead II.
10 = Lead III.
11 = external respiration path.
Set the test tone amplitude for respiration drive signal.
00 (default) = amplitude/8.
01 = amplitude/4.
10 = amplitude/2.
11 = amplitude.
Set frequency for respiration.
RESPFREQ
RESPALTFREQ = 0
RESPALTFREQ = 1 (periodic)
00 (default)
56 kHz
64 kHz
01
54 kHz
56.9 kHz
10
52 kHz
51.2 kHz
11
50 kHz
46.5 kHz
Enable respiration.
0 (default) = respiration disabled.
1 = respiration enabled.
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
Rev. A | Page 57 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Table 29. Pace Detection Control Register (PACECTL) Address 0x04, Reset Value = 0x000F88 1
R/W
Default
R/W
1
Bit
[23:12]
11
Name
Reserved
PACEFILTW
R/W
1
10
PACETFILT2
R/W
1
9
PACETFILT1
R/W
R/W
R/W
11
00
01
[8:7]
[6:5]
[4:3]
PACE3SEL [1:0]
PACE2SEL [1:0]
PACE1SEL [1:0]
R/W
R/W
R/W
0
0
0
2
1
0
PACE3EN
PACE2EN
PACE1EN
1
Function
Reserved, set to 0
Pace width filter
0 = filter disabled
1 (default) = filter enabled
Pace Validation Filter 2
0 = filter disabled
1 (default) = filter enabled
Pace Validation Filter 1
0 = filter disabled
1 (default) = filter enabled
Set lead for pace detection measurement
00 = Lead I
01 = Lead II
10 = Lead III
11 = Lead aVF
Enable pace detection algorithm
0 (default) = pace detection disabled
1 = pace detection enabled
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
Rev. A | Page 58 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Table 30. Common-Mode, Reference, and Shield Drive Control Register (CMREFCTL) Address 0x05, Reset Value = 0xE00000
R/W
R/W
R/W
R/W
R/W
R/W
Default
1
1
1
0
0
Bit
23
22
21
20
19
Name
LACM
LLCM
RACM
V1CM
V2CM
R/W
R/W
R/W
R/W
R/W
R/W
R/W
0
0
0
0
0
0
0
0
[18:15]
14
13
12
11
10
9
8
Reserved
LARLD
LLRLD
RARLD
V1RLD
V2RLD
CERLD
CEREFEN
R/W
0000
[7:4]
RLDSEL [3:0]1
R/W
0
3
DRVCM
R/W
0
2
EXTCM
R/W
0
1
RLDSEL1
R/W
0
0
SHLDEN 1
1
Function
Common-mode electrode select.
Any combination of the five input electrodes can be used to create the common-mode
signal, VCM. Bits[23:19] are ignored when Bit 2 is selected. Common mode is the average of
the selected electrodes. When a single electrode is selected, common mode is the signal
level of that electrode alone. The common-mode signal can be driven from the internal
VCM_REF (1.3 V) when Bits [23:19] = 0.
0 = does not contribute to the common mode.
1 = contributes to the common mode.
Reserved, set to 0.
RLD summing junction. Note that if the RLD amplifier is disabled (using RLDSEL), these
switches are not automatically forced open, and the user must disable them using Bits[9:14].
0 (default) = does not contribute to RLD input.
1 = contributes to RLD input.
Common electrode (CE) reference, see Figure 57.
0 (default) = common electrode disabled.
1 = common electrode enabled.
Select electrode for reference drive.
0000 (default) = RLD_OUT.
0001 = LA.
0010 = LL.
0011 = RA.
0100 = V1.
0101 = V2.
0110 to 1111 = reserved.
Common-mode output. When set, the internally derived common-mode signal is driven
out of the common-mode pin. This bit has no effect if an external common mode is
selected.
0 (default) = common mode is not driven out.
1 = common mode is driven out of the external common-mode pin.
Select the source of common mode (use when operating multiple devices together).
0 (default) = internal common mode selected.
1 = external common mode selected (all the internal common-mode switches are off ).
Enable right leg drive reference electrode.
0 (default) = disabled.
1 = enabled.
Enable shield drive.
0 (default) = shield drive disabled.
1 = shield drive enabled.
ADAS1000 and ADAS1000-1 models only, ADAS1000-2 models does not contain these features.
Rev. A | Page 59 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Table 31. GPIO Control Register (GPIOCTL) Address 0x06, Reset Value = 0x000000
R/W
R/W
Default
0
0
Bit
[23:19]
18
Name
Reserved
SPIFW
Function
Reserved, set to 0
Frame secondary SPI words with chip select
0 (default) = MCS asserted for entire frame
1 = MCS asserted for individual word
R/W
R/W
0
0
17
16
Reserved
SPIEN
R/W
00
[15:14]
G3CTL [1:0]
R/W
0
13
G3OUT
R
0
12
G3IN
R/W
00
[11:10]
G2CTL [1:0]
R/W
0
9
G2OUT
R
0
8
G2IN
R/W
00
[7:6]
G1CTL [1:0]
R/W
0
5
G1OUT
R
0
4
G1IN
R/W
00
[3:2]
G0CTL [1:0]
R/W
0
1
G0OUT
R
0
0
G0IN
Reserved, set to 0
Secondary SPI enable (ADAS1000 and ADAS1000-1 only); SPI interface providing ECG data at
128 kHz data rate for external digital pace algorithm detection, uses GPIO0, GPIO1, GPIO2 pins
0 (default) = disabled
1 = enabled; he individual control bits for GPIO0, GPIO1, GPIO2 are ignored; GPIO3 is not affected by
SPIEN
State of GPIO3 pin
00 (default) = high impedance
01 = input
10 = output
11 = open drain
Output value to be written to GPIO3 when the pin is configured as an output or open drain
0 (default) = low value
1 = high value
Read only; input value read from GPIO3 when the pin is configured as an input
0 (default) = low value
1 = high value
State of GPIO2 pin
00 (default) = high impedance
01 = input
10 = output
11 = open drain
Output value to be written to GPIO2 when the pin is configured as an output or open drain
0 (default) = low value
1 = high value
Read only Input value read from GPIO2 when the pin is configured as an input
0 (default) = low value
1 = high value
State of GPIO1 pin
00 (default) = high impedance
01 = input
10 = output
11 = open drain
Output value to be written to GPIO1 when the pin is configured as an output or open drain
0 (default) = low value
1 = high value
Read only; input value read from GPIO1 when the pin is configured as an input
0 (default) = low value
1 = high value
State of the GPIO0 pin
00 (default) = high impedance
01 = input
10 = output
11 = open drain
Output value to be written to GPIO0 when pin is configured as an output or open drain
0 (default) = low value
1 = high value
(Read only) input value read from GPIO0 when pin is configured as an input
0 (default) = low value
1 = high value
Rev. A | Page 60 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Table 32. Pace Amplitude Threshold Register (PACEAMPTH) Address 0x07, Reset Value = 0x242424 1
R/W
R/W
R/W
R/W
1
Default
0010 0100
0010 0100
0010 0100
Bit
[23:16]
[15:8]
[7:0]
Name
PACE3AMPTH
PACE2AMPTH
PACE1AMPTH
Function
Pace amplitude threshold
Threshold = N × 2 × VREF/GAIN/216
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
Table 33. Test Tone Register (TESTTONE) Address 0x08, Reset Value = 0x000000
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
Default
0
0
0
0
0
0
00
Bit
23
22
21
20
19
[18:5]
[4:3]
Name
TONLA
TONLL
TONRA
TONV1
TONV2
Reserved
TONTYPE
R/W
0
2
TONINT
R/W
0
1
TONOUT
R/W
0
0
TONEN
Function
Tone select
0 (default) = 1.3 V VCM_REF
1 = 1 mV sine wave or square wave for TONINT = 1, no connect for TONINT=0
Reserved, set to 0
00 (default) = 10 Hz sine wave
01 = 150 Hz sine wave
1x = 1 Hz, 1 mV square wave
Test tone internal or external
0 (default) = external test tone; test tone to be sent out through CAL_DAC_IO and
applied externally to enabled channels
1 = internal test tone; disconnects external switches for all ECG channels and
connects the calibration DAC test tone internally to all ECG channels; in gang
mode, the CAL_DAC_IO is connected, and the slave disables the calibration DAC
Test tone out enable
0 (default) = disconnects test tone from CAL_DAC_IO during internal mode only
1 = connects CAL_DAC_IO to test tone during internal mode
Enables an internal test tone to drive entire signal chain, from preamplifier to SPI
interface; this tone comes from the calibration DAC and goes to the preamplifier
through the internal mux; when TONEN (calibration DAC) is enabled, ac lead-off is
disabled
0 (default) = disable the test tone
1 = enable the 1 mV sine wave test tone (calibration mode has priority)
Rev. A | Page 61 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Table 34. Calibration DAC Register (CALDAC) Address 0x09, Reset Value = 0x002000 1
R/W
R/W
Default
0
1
Bit
[23:14]
13
Name
Reserved
CALCHPEN
R/W
0
12
CALMODEEN
R/W
0
11
CALINT
R/W
0
10
CALDACEN
R/W
0000000000
[9:0]
CALDATA[9:0]
1
Function
Reserved, set to 0.
Calibration chop clock enable. The calibration DAC output (CAL_DAC_IO) can be
chopped to lower 1/f noise. Chopping is performed at 256 kHz.
0 = disabled.
1 (default) = enabled.
Calibration mode enable.
0 (default) = disable calibration mode.
1 = enable calibration mode; connect CAL DAC_IO, begin data acquisition on ECG
channels.
Calibration internal or external.
0 (default) = external calibration to be performed externally by looping CAL_DAC_IO
around into ECG channels.
1 = internal calibration; disconnects external switches for all ECG channels and
connects calibration DAC signal internally to all ECG channels.
Enable 10-bit calibration DAC for calibration mode or external use.
0 (default) = disable calibration DAC.
1 = enable calibration DAC. If a master device and not in calibration mode, also
connects the calibration DAC signal out to the CAL_DAC_IO pin for external use. If in
slave mode, the calibration DAC is disabled to allow master to drive the slave
CAL_DAC_IO pin. When the calibration DAC is enabled, ac lead-off is disabled.
Set the calibration DAC value.
To ensure successful update of the calibration DAC, the serial interface must issue four additional SCLK cycles after writing the new calibration DAC register word.
Rev. A | Page 62 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Table 35. Frame Control Register (FRMCTL) Address 0x0A, Reset Value = 0x079000
R/W
R/W
R/W
R/W
R/W
R/W
R/W
R/W
Default
0
0
0
0
0
1111
0
Bit
23
22
21
20
19
[18:15]
14
Name
LEAD I/LADIS
LEADII/LLDIS
LEADIII/RADIS
V1DIS
V2DIS
Reserved
PACEDIS1
R/W
0
13
RESPMDIS 1
R/W
1
12
RESPPHDIS1
R/W
0
11
LOFFDIS
R/W
0
10
GPIODIS
R/W
0
9
CRCDIS
R/W
R/W
0
0
8
7
RESERVED
ADIS
R/W
0
6
RDYRPT
R/W
0
5
Reserved
R/W
0
4
DATAFMT
R/W
00
[3:2]
SKIP[1:0]
R/W
00
[1:0]
FRMRATE[1:0]
1
Function
Include/exclude word from ECG data frame. If the electrode/lead is included in the dataword and the electrode falls off, the data-word is undefined.
0 (default) = included in frame.
1 = exclude from frame.
Reserved, set to 1111.
Pace detection.
0 (default) = included in frame.
1 = exclude from frame.
Respiration magnitude.
0 (default) = included in frame.
1 = exclude from frame.
Respiration phase.
0 = included in frame.
1 (default) = exclude from frame.
Lead-off status.
0 (default) = included in frame.
1 = exclude from frame.
GPIO word disable.
0 (default) = included in frame.
1 = exclude from frame.
CRC word disable.
0 (default) = included in frame.
1 = exclude from frame.
Reserved, set to 0.
Automatically excludes PACEDIS[14], RESPMDIS[13], LOFFDIS[11] words if their flags are
not set in the header.
0 (default) = fixed frame format.
1 = autodisable words (words per frame changes).
Ready repeat. If this bit is set and the frame header indicates data is not ready, the frame
header is continuously sent until data is ready.
0 (default) = always send entire frame.
1 = repeat frame header until ready.
Reserved, set to 0 .
Sets the output data format, see Figure 57.
0 (default) = digital lead/vector format (available only in 2 kHz and 16 kHz data rates).
1 = electrode format.
Skip interval. This field provides a way to decimate the data.
00 (default) = output every frame.
01 = output every other frame
1× = output every 4th frame.
Sets the output data rate.
00 (default) = 2 kHz output data rate.
01 = 16 kHz output data rate.
10 = 128 kHz output data rate (DATAFMT must be set to 1).
11 = 31.25 Hz.
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
Rev. A | Page 63 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Table 36. Filter Control Register (FILTCTL) Address 0x0B, Reset Value = 0x000000
R/W
R/W
R/W
Default
0
0
Bit
[23:6]
5
Name
Reserved
MN2K
R/W
0
4
N2KBP
R/W
00
[3:2]
LPF[1:0]
R/W
00
[1:0]
Reserved
Function
Reserved, set to 0
2 kHz notch bypass for SPI master
0 (default) = notch filter bypassed
1 = notch filter present
2 kHz notch bypass
0 (default) = notch filter present
1 = notch filter bypassed
00 (default) = 40 Hz
01 = 150 Hz
10 = 250 Hz
11 = 450 Hz
Reserved, set to 0
Table 37. AC Lead-Off Upper Threshold Register (LOFFUTH) Address 0x0C, Reset Value = 0x00FFFF
R/W
Default
0
Bit
[23:20]
Name
Reserved
Function
Reserved, set to 0
R/W
0
[19:16]
ADCOVER[3:0]
ADC overrange threshold
An ADC out-of-range error is flagged if the ADC output is greater than the overrange
threshold; the overrange threshold is offset from the maximum value
Threshold = max_value – ADCOVER × 26
R/W
0xFFFF
[15:0]
LOFFUTH[15:0]
0000 = maximum value (disabled)
0001 = max_value − 64
0010 = max_value − 128
…
1111 = max_value − 960
Applies to ac lead-off upper threshold only; lead-off is detected if the output is ≥ N ×
2 × VREF/GAIN/216
0=0V
Table 38. AC Lead-Off Lower Threshold Register (LOFFLTH) Address 0x0D, Reset Value = 0x000000
R/W
Default
0
Bit
[23:20]
Name
Reserved
Function
Reserved, set to 0
R/W
0
[19:16]
ADCUNDR[3:0]
ADC underrange threshold
An ADC out-of-range error is flagged if the ADC output is less than the underrange
threshold
Threshold = min_value + ADCUNDR × 26
R/W
0
[15:0]
LOFFLTH[15:0]
0000 = minimum value (disabled)
0001 = min_value + 64
0010 = min_value + 128
…
1111 = min_value + 960
Applies to ac lead-off lower threshold only; lead-off is detected if the output is ≤ N ×
2 × VREF/GAIN/216
0=0V
Rev. A | Page 64 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Table 39. Pace Edge Threshold Register (PACEEDGETH) Address 0x0E, Reset Value = 0x000000 1
R/W
R/W
R/W
Default
0
0
Bit
[23:16]
[15:8]
Name
PACE3EDGTH
PACE2EDGTH
R/W
0
[7:0]
PACE1EDGTH
1
Function
Pace edge trigger threshold
0 = PACEAMPTH/2
1 = VREF/GAIN/216
N = N × VREF/GAIN/216
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
Table 40. Pace Level Threshold Register (PACELVLTH) Address 0x0F, Reset Value = 0x000000 1
R/W
R/W
R/W
Default
0
0
Bit
[23:16]
[15:8]
Name
PACE3LVLTH[7:0]
PACE2LVLTH[7:0]
R/W
0
[7:0]
PACE1LVLTH[7:0]
1
Function
Pace level threshold; This is a signed value
−1 = 0xFF = −VREF/GAIN/216
0 = 0x00 = 0 V
+1 = 0x01 = +VREF/GAIN/216
N = N × VREF/GAIN/216
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
Table 41. Read Electrode/Lead Data Registers (Electrode/Lead) Address 0x11 to 0x15, Reset Value = 0x000000 1
R/W
Default
Bit
[31:24]
Name
Address [7:0]
R
0
[23:0]
ECG data
Function
0x11: LA or Lead I.
0x12: LL or Lead II.
0x13: RA or Lead III.
0x14: V1 or V1’.
0x15: V2 or V2’.
Channel data value. Data left justified (MSB) irrespective of data rate.
In electrode format, the value is an unsigned integer.
In vector format, the value is a signed two’s complement integer format.
Lead/Vector format has a 2× range compared to electrode format because it can swing
from +VREF to –VREF; therefore, the LSB size is doubled.
Electrode format and Analog Lead Format:
Minimum value (000…) = 0 V
Maximum value (1111….) = VREF/GAIN
LSB = (2 × VREF/GAIN)/(2N– 1)
Digital Lead format:
Minimum value (1000…) = −(VREF/GAIN)
Maximum value (0111….) = +VREF/GAIN
LSB = 4 × (VREF/GAIN)/(2N – 1)
where N = number of data bits: 16 for 128 kHz data rate or 24 for 2 kHz/16 kHz data rate.
1
If using 128 kHz data rate in frame mode, only the upper 16 bits are sent. If using the 128 kHz data rate in regular read/write mode, all 32 bits are sent.
Rev. A | Page 65 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Table 42. Read Pace Detection Data/Status Register (PACEDATA) Address 0x1A, Reset Value = 0x000000 1, 2, 3
R/W
R
Default
0
Bit
23
Name
Pace 3 detected
R
000
[22:20]
Pace Channel 3 width
R
0000
[19:16]
Pace Channel 3 height
R
0
15
Pace 2 detected
R
000
[14:12]
Pace Channel 2 width
R
0000
[11:8]
Pace Channel 2 height
R
0
7
Pace 1 detected
R
000
[6:4]
Pace Channel 1 width
R
0000
[3:0]
Pace Channel 1 height
Function
Pace 3 detected. This bit is set once a pace pulse is detected. This bit is set on the
trailing edge of the pace pulse.
0 = pace pulse not detected in current frame.
1 = pace pulse detected in this frame.
This bit is log2 (width) − 1 of the pace pulse.
Width = 2N + 1/128 kHz.
This bit is the log2 (height) of the pace pulse.
Height = 2N × VREF/GAIN/216.
Pace 2 detected. This bit is set once a pace pulse is detected. This bit is set on the
trailing edge of the pace pulse.
0 = pace pulse not detected in current frame.
1 = pace pulse detected in this frame.
This bit is log2 (width) − 1 of the pace pulse.
Width = 2N + 1/128 kHz.
This bit is the log2 (height) of the pace pulse.
Height = 2N × VREF/GAIN/216.
Pace 1 detected. This bit is set once a pace pulse is detected. This bit is set on the
trailing edge of the pace pulse.
0 = pace pulse not detected in current frame.
1 = pace pulse detected in this frame.
This bit is log2 (width) − 1 of the pace pulse.
Width = 2N + 1/128 kHz.
This bit is the log2 (height) of the pace pulse.
Height = 2N × VREF/GAIN/216.
If using 128 kHz data rate in frame mode, this word is stretched over two 16-bit words. If using the 128 kHz data rate in regular read/write mode, all 32 bits are sent.
Log data for width and height is provided here to ensure that it fits in one full 32-bit data-word. As a result there may be some amount of error in the resulting value.
For more accurate reading, read the 0x3A, 0x3B, 0x3C registers (see Table 51).
3
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
1
2
Table 43. Read Respiration Data—Magnitude Register (RESPMAG) Address 0x1B, Reset Value = 0x000000 1, 2
R/W
R
1
2
Default
0
Bit
[23:0]
Name
Respiration magnitude[23:0]
Function
Magnitude of respiration signal. This is an unsigned value.
2 × VREF × (224 − 1)/(1.6468 × respiration gain).
If using 128 kHz data rate in frame mode, this word is stretched over two 16-bit words. If using the 128 kHz data rate in regular read/write mode, all 32 bits are sent.
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
Table 44. Read Respiration Data—Phase Register (RESPPH) Address 0x1C, Reset Value = 0x000000 1, 2
R/W
R
1
2
Default
0
Bit
[23:0]
Name
Respiration
phase[23:0]
Function
Phase of respiration signal. Can be interpreted as either signed or unsigned value. If
unsigned, the range is from 0 to 2π. If signed, the range is from – π to +π.
0x000000 = 0.
0x000001 = 2π/224.
0x400000 = π/2.
0x800000 = +π = − π.
0xC00000 = +3π/2 = − π/2.
0xFFFFFF = +2π(1 − 2−24) = −2π/224.
This register is not part of framing data, but may be read by issuing a register read command of this address.
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
Rev. A | Page 66 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Table 45. Lead-Off Status Register (LOFF) Address 0x1D, Reset Value = 0x000000
R/W
R
Default
0
R
R
0
0
R
0
Bit
23
22
21
20
19
18
13
Name
RLD lead-off
Status
LA lead-off status
LL lead-off status
RA lead-off status
V1 lead-off status
V2 lead-off status
CELO
[17:14]
12
11
10
9
8
[7:0]
Reserved
LAADCOR
LLADCOR
RAADCOR
V1ADCOR
V2ADCOR
Reserved
Function
Electrode connection status.
If either dc or ac lead-off is enabled, these bits are the corresponding lead-off status. If
both dc and ac lead-off are enabled, these bits reflect only the ac lead-off status. DC
lead-off is available in the DCLEAD-OFF register (see Table 46).
The common electrodes have only dc lead-off detection.
An ac lead-off signal can be injected into the common electrode, but there is no ADC
input to measure its amplitude. If the common electrode is off, it affects the ac lead-off
amplitude of the other electrodes.
These bits accumulate in the frame buffer and are cleared when the frame buffer is
loaded into the SPI buffer.
0 = electrode is connected.
1 = electrode is disconnected.
Reserved.
ADC out of range error.
These status bits indicate the resulting ADC code is out of range.
These bits accumulate in the frame buffer and are cleared when the frame buffer is
loaded into the SPI buffer.
Reserved.
Table 46. DC Lead-Off Register (DCLEAD-OFF) Address 0x1E, Reset Value = 0x000000 1
R/W
R
Default
0
Bit
23
22
21
20
13
R
0
R
0
[19:14]
[8:3]
12
11
10
9
2
R
1
0
[1:0]
Name
RLD input
overrange
LA input
overrange
LL input
overrange
RA input
overrange
CE input
overrange
Reserved
Function
The dc lead-off detection is comparator based and compares to a fixed level. Individual
electrode bits flag indicate if the dc lead-off comparator threshold level has been
exceeded.
0 = electrode < overrange threshold, 2.4 V.
1 = electrode > overrange threshold, 2.4 V.
RLD input
underrange
LA input
underrange
LL input
underrange
RA input
underrange
CE input
underrange
Reserved
The dc lead-off detection is comparator based and compares to a fixed level. Individual
electrode bits indicate if the dc lead-off comparator threshold level has been exceeded.
0 = electrode > underrange threshold, 0.2 V.
1 = electrode < underrange threshold, 0.2 V.
Reserved.
This register is not part of framing data, but can be read by issuing a register read command of this address.
Rev. A | Page 67 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Table 47. Operating State Register (OPSTAT) Address 0x1F, Reset Value = 0x000000 1
R/W
R
R
R
Default
0
0
0
Bit
[23:4]
3
2
Name
Reserved
Internal error
Configuration status
R
0
1
PLL lock
R
0
0
PLL locked status
1
Function
Reserved.
Internal digital failure. This is set if an error is detected in the digital core.
This bit is set after a reset indicating that the configuration has not been read yet.
Once the configuration is set, this bit is ready.
0 = ready.
1 = busy.
PLL lock lost. This bit is set if the internal PLL loses lock after it is enabled and
locked. This bit is cleared once this register is read or the PWREN bit (Address
0x01[1]) is cleared.
0 = PLL locked.
1 = PLL lost lock.
This bit indicates the current state of the PLL locked status.
0 = PLL not locked.
1 = PLL locked.
This register is not part of framing data, but can be read by issuing a register read command of this address. This register assists support efforts giving insight into
potential areas of malfunction within a failing device.
Table 48. Extended Switch for Respiration Inputs Register (EXTENDSW) Address 0x20, Reset Value = 0x000000
R/W
R/W
Default
0
R/W
R/W
R/W
R/W
1
0
Bit
23
22
21
20
19
18
17
16
15
14
13
12
11
10
9
8
Name
EXT_RESP_RA to ECG1_LA
EXT_RESP_RA to ECG2_LL
EXT_RESP_RA to ECG3_RA
EXT_RESP_RA to ECG4_V1
EXT_RESP_RA to ECG5_V2
EXT_RESP_LL to ECG1_LA
EXT_RESP_LL to ECG2_LL
EXT_RESP_LL to ECG3_RA
EXT_RESP_LL to ECG4_V1
EXT_RESP_LL to ECG5_V2
EXT_RESP_LA to ECG1_LA
EXT_RESP_LA to ECG2_LL
EXT_RESP_LA to ECG3_RA
EXT_RESP_LA to ECG4_V1
EXT_RESP_LA to ECG5_V2
AUX_V1
0
7
AUX_V2
0
6
NOMATH_V1
0
5
NOMATH_V2
0
[4:0]
Reserved
Switch
SW1a
SW1b
SW1c
SW1d
SW1e
SW2a
SW2b
SW2c
SW2d
SW2e
SW3a
SW3b
SW3c
SW3d
SW3e
Function
External respiration electrode input switch to channel electrode input (see
Figure 69). 1
0 = switch open.
1 = switch closed.
V1 and V2 electrodes can be used for measurement purposes other than ECG.
To achieve this, they must be disconnected from the patient VCM voltage
provided from the internal common-mode buffer and, instead, connected to
the internal VCM_REF level of 1.3 V.
Setting the AUX_Vx bits high connects the negative input of the V1 channel
amplifier to internal VCM_REF level. This allows the user to make alternative
measurements on V1 relative to the VCM_REF level.
If using digital lead mode, use these bits in conjunction with Bits[6:5]
(NO_MATH_Vx).
In digital lead mode, the digital core always calculates V1’ and V2’ with respect
to WCT (LA + LL + RA)/3.
Where V1 or V2 are used for alternative measurement purposes other than
ECG, this math calculation must be disabled. These bits are most likely used in
conjunction with Bits[8:7] (AUX_Vx).
Set NOMATH_Vx bits high to disable the math calculation in V1 and V2,
respectively.
Reserved, set to 0.
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these EXT_RESP_xx pins.
Rev. A | Page 68 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Table 49. User Gain Calibration Registers (CALxx) Address 0x21 to Address 0x25, Reset Value = 0x000000
R/W
Default
Bit
Name
Function
[31:24]
Address [7:0]
0x21: calibration LA.
0x22: calibration LL.
0x23: calibration RA.
0x24: calibration V1.
0x25: calibration V2.
User can choose between default calibration values or user calibration values for GAIN 0,
GAIN 1, GAIN 2.
Note that for GAIN 3, there is no factory calibration.
0 = default calibration values (factory calibration).
1 = user calibration values.
Reserved, set to 0
Gain calibration value.
Result = data × (1 + GAIN × 2−17).
The value read from this register is the current gain calibration value. If the USRCAL bit is set
to 0, this register returns the default value for the current gain setting.
0x7FF (+2047) = ×1.00000011111111111b.
0x001 (+1) = ×1.00000000000000001b.
0x000 (0) = ×1.00000000000000000b.
0xFFF (−1) = ×0.11111111111111111b.
0x800 (−2048) = ×0.11111100000000000b.
R/W
0
23
USRCAL
R/W
R/W
0
0
[22:12]
[11:0]
Reserved
CALVALUE
Table 50. Read AC Lead-Off Amplitude Registers (LOAMxx) Address 0x31 to Address 35, Reset Value = 0x000000 1
R/W
Default
Bit
[31:24]
Name
Address [7:0]
R/W
R
0
0
[23:16]
[15:0]
Reserved
LOFFAM
1
Function
0x31: LA ac lead-off amplitude.
0x32: LL ac lead-off amplitude.
0x33: RA ac lead-off amplitude.
0x34: V1 ac lead-off amplitude.
0x35: V2 ac lead-off amplitude.
Reserved.
Measured amplitude.
When ac lead-off is selected, the data is the average of the rectified 2 kHz band-pass filter
with an update rate of 8 Hz and cutoff frequency at 2 Hz. The output is the amplitude of the
2 kHz signal scaled by 2/π approximately = 0.6 (average of rectified sine wave). To convert to
RMS, scale the output by π/(2√2).
Lead-off (unsigned).
Minimum 0x0000 = 0 V.
LSB 0x0001= VREF/GAIN/216.
Maximum 0xFFFF = VREF/GAIN.
This register is not part of framing data, but can be read by issuing a register read command of this address.
Rev. A | Page 69 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Table 51. Pace Width and Amplitude Registers (PACExDATA) Address 0x3A to Address 0x3C, Reset Value = 0x000000 1, 2
R/W
Default
Bit
[31:24]
Name
Address [7:0]
R
0
[23:8]
Pace height
R
0
[7:0]
Pace width
1
2
Function
0x3A: PACE1DATA
0x3B: PACE2DATA
0x3C: PACE3DATA
Measured pace height in signed two’s complement value
0=0
1 = VREF/GAIN/216
N = N × VREF/GAIN/216
Measured pace width in 128 kHz samples
N: N/128 kHz = width
12: 12/128 kHz = 93 µs
255:255/128 kHz = 2.0 ms
These registers are not part of framing data but can be read by issuing a register read command of these addresses.
ADAS1000 model only, ADAS1000-1/ADAS1000-2 models do not contain these features.
Rev. A | Page 70 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Table 52. Frame Header (FRAMES) Address 0x40, Reset Value = 0x800000 1
R/W
R
R
Default
1
0
Bit
31
30
Name
Marker
Ready bit
R
0
[29:28]
Overflow [1:0]
R
0
27
Fault
R
0
26
Pace 3 detected
R
0
25
Pace 2 detected
R
0
24
Pace 1 detected
R
0
23
Respiration
R
0
22
Lead-off detected
R
0
21
DC lead-off detected
R
0
20
ADC out of range
0
[19:0]
Reserved
1
Function
Header marker, set to 1 for the header.
Ready bit indicates if ECG frame data is calculated and ready for reading.
0 = ready, data frame follows.
1 = busy.
Overflow bits indicate that since the last frame read, a number of frames have
been missed. This field saturates at the maximum count. The data in the frame
including this header word is valid but old if the overflow bits are >0.
When using skip mode (FRMCTL register (0x0A), Bits[3:2]), the overflow bit acts as a
flag, where a nonzero value indicates an overflow.
00 = 0 missed.
01 = 1 frame missed.
10 = 2 frames missed.
11 = 3 or more frames missed.
Internal device error detected.
0 = normal operation.
1 = error condition.
Pace 3 indicates pacing artifact was qualified at most recent point.
0 = no pacing artifact.
1 = pacing artifact present.
Pace 2 indicates pacing artifact was qualified at most recent point.
0 = no pacing artifact.
1 = pacing artifact present.
Pace 1 indicates pacing artifact was qualified at most recent point.
0 = no pacing artifact.
1 = pacing artifact present.
0 = no new respiration data.
1 = respiration data updated.
If both dc and ac lead-off are enabled, this bit is the OR of all the ac lead-off detect
flags. If only ac or dc lead-off is enabled, this bit reflects the OR of all dc and ac
lead-off flags.
0 = all leads connected.
1 = one or more lead-off detected.
0 = all leads connected.
1 = one or more lead-off detected.
0 = ADC within range.
1 = ADC out of range.
Reserved
If using 128 kHz data rate in frame mode, only the upper 16 bits are sent. If using the 128 kHz data rate in regular read/write mode, all 32 bits are sent.
Table 53. Frame CRC Register (CRC) Address 0x41, Reset Value = 0xFFFFFF 1
R/W
R
1
Bit
[23:0]
Name
CRC
Function
Cyclic redundancy check
The CRC register is a 32-bit word for 2 kHz and 16 kHz data rate and a 16-bit word for 128 kHz rate. See Table 22 for more details.
Rev. A | Page 71 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
EXAMPLES OF INTERFACING TO THE ADAS1000
The following examples shows register commands required to
configure the ADAS1000 device into particular modes of
operation and to start framing ECG data.
Example 2: Enable Respiration and Stream Conversion
Data
1.
Example 1: Initialize the ADAS1000 for ECG Capture and
Start Streaming Data
1.
2.
3.
4.
5.
Write 1 configures the CMREFCTL register for CM =
WCT = (LA + LL + RA)/3; RLD is enabled onto the
RLD_OUT electrode. The shield amplifier is enabled.
Write 2 configures the FRMCTL register to output nine
words per frame/packet. The frame/packet of words
consist of the header, five ECG words, pace, respiration
magnitude, and lead-off. The frame is configured to
always send, irrespective of ready status. The ADAS1000
is in vector format mode with a data rate of 2 kHz.
Write 3 addresses the ECGCTL register, enabling all
channels into a gain of 1.4, low noise mode, and differential input, which configures the device for analog lead
mode. This register also configures the device as a master,
using the external crystal as the input source to the XTALx
pins. The ADAS1000 is also put into conversion mode in
this write.
Write 4 issues the read command to start putting the
converted data out on the SDO pin.
Continue to issue SCLK cycles to read the converted data
at the configured packet data rate (2 kHz). The SDI input
should be held low when reading back the conversion
data because any commands issued to the interface during
read of frame/packet are understood to be a change of
configuration data and will stop the ADC conversions
to allow the interface to process the new command.
2.
3.
4.
Write 1 configures the RESPCTL register with a 56 kHz
respiration drive signal, gain = 1, driving out through
the respiration capacitors and measuring on Lead I.
Write 2 issues the read command to start putting the
converted data out on the SDO pin.
Continue to issue SCLK cycles to read the converted data
at the configured packet data rate.
Note that this example assumes that the FRMCTL register
has already been configured such that the respiration
magnitude is available in the data frame, as arranged in
Write 2 of Example 1.
Example 3: DC Lead-Off and Stream Conversion Data
1.
2.
3.
4.
Write 1 configures the LOFFCTL register with a dc lead-off
enabled for a lead-off current of 50 nA.
Write 2 issues the read command to start putting the
converted data out on the SDO pin.
Continue to issue SCLK cycles to read the converted
data at the configured packet data rate.
Note that this example assumes that the FRMCTL register
has already been configured such that the dc lead-off word
is available in the data frame, as arranged in Write 2 of
Example 1.
Table 54. Example 1: Initialize the ADAS1000 for ECG Capture and Start Streaming Data
Write Command
Write 1
Write 2
Write 3
Write 4
Register Addressed
CMREFCTL
FRMCTL
ECGCTL
FRAMES
Read/Write Bit
1
1
1
0
Register Address
000 0101
000 1010
000 0001
100 0000
Data
1110 0000 0000 0000 0000 1011
0000 0111 1001 0110 0000 0000
1111 1000 0000 0100 1010 1110
0000 0000 0000 0000 0000 0000
32-Bit Write Command
0x85E0000B
0x8A079600
0x81F804AE
0x40000000
Data
0000 0000 0010 0000 1001 1001
0000 0000 0000 0000 0000 0000
32-Bit Write Command
0x83002099
0x40000000
Data
0000 0000 0000 0000 0001 0101
0000 0000 0000 0000 0000 0000
32-Bit Write Command
0x82000015
0x40000000
Table 55. Example 2: Enable Respiration and Stream Conversion Data
Write Command
Write 1
Write 2
Register Addressed
RESPCTL
FRAMES
Read/Write Bit
1
0
Register Address
000 0011
100 0000
Table 56. Example 3: Enable DC Lead-Off and Stream Conversion Data
Write Command
Write 1
Write 2
Register addressed
LOFFCTL
FRAMES
Read/Write Bit
1
0
Register Address
000 0010
100 0000
Rev. A | Page 72 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
Example 4: Configure 150 Hz Test Tone Sine Wave on
Each ECG Channel and Stream Conversion Data
Example 5: Enable Pace Detection and Stream
Conversion Data
1.
1.
2.
3.
4.
5.
6.
7.
Write 1 configures the CMREFCTL register to VCM_REF
= 1.3 V (no electrodes contribute to VCM). RLD is enabled
to RLD_OUT, and the shield amplifier enabled.
Write 2 addresses the TESTTONE register to enable the
150 Hz sine wave onto all electrode channels.
Write 3 addresses the FILTCTL register to change the
internal low-pass filter to 250 Hz to ensure that the 150 Hz
sine wave can pass through.
Write 4 configures the FRMCTL register to output nine
words per frame/packet. The frame/packet of words
consists of the header and five ECG words, pace,
respiration magnitude, and lead-off. The frame is
configured to always send, irrespective of ready status. The
ADAS1000 is in electrode format mode with a data rate of
2 kHz. Electrode format is required to see the test tone
signal correctly on each electrode channel.
Write 5 addresses the ECGCTL register, enabling all
channels into a gain of 1.4, low noise mode. It configures
the device as a master and driven from the XTAL input
source. The ADAS1000 is also put into conversion mode
in this write.
Write 6 issues the read command to start putting the
converted data out on the SDO pin.
Continue to issue SCLK cycles to read the converted data
at the configured packet data rate.
2.
3.
4.
5.
Write 1 configures the PACECTL register with all three
pace detection instances enabled, PACE1EN detecting on
Lead II, PACE2EN detecting on Lead I, and PACE3EN
detecting on Lead aVF. The pace width filter and validation
filters are also enabled.
Write 2 issues the read command to start putting the
converted data out on the SDO pin.
Continue to issue SCLK cycles to read the converted data
at the configured packet data rate. When a valid pace is
detected, the detection flags are confirmed in the header
word and the PACEDATA register contains information
on the width and height of the measured pulse from each
measured lead.
Note that the PACEAMPTH register default setting is
0x242424, setting the amplitude of each of the pace
instances to 1.98 mV/gain.
Note that this example assumes that the FRMCTL register
has already been configured such that the PACEDATA
word is available in the data frame, as arranged in Write 2
of Example 1.
Table 57. Example 4: Configure 150 Hz Test Tone Sine Wave on Each ECG Channel and Stream Conversion Data
Write Command
Write 1
Write 2
Write 3
Write 4
Write 5
Write 6
Register Addressed
CMREFCTL
TESTTONE
FILTCTL
FRMCTL
ECGCTL
FRAMES
Read/Write Bit
1
1
1
1
1
0
Register Address
000 0101
000 1000
000 1011
000 1010
000 0001
100 0000
Data
0000 0000 0000 0000 0000 1011
1111 1000 0000 0000 0000 1101
0000 0000 0000 0000 0000 1000
0000 0111 1001 0110 0001 0000
1111 1000 0000 0000 1010 1110
0000 0000 0000 0000 0000 0000
32-Bit Write Command
0x8500000B
0x88F8000D
0x8B000008
0x8A079610
0x81F800AE
0x40000000
Data
0000 0000 0000 1111 1000 1111
0000 0000 0000 0000 0000 0000
32-Bit Write Command
0x84000F8F
0x40000000
Table 58. Example 5: Enable Pace Detection and Stream Conversion Data
Write Command
Write 1
Write 2
Register Addressed
PACECTL
FRAMES
Read/Write Bit
1
0
Register Address
000 0100
100 0000
Rev. A | Page 73 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
Example 6: Writing to Master and Slave Devices and Streaming Conversion Data
Master Configuration
Slave Configuration
1. Write 4 configures the FRMCTL register to output nine
1. Write 1 configures the FRMCTL register to output seven
words per frame/packet (note that this differs from the
words per frame/packet. The frame/packet of words consist
number of words in a frame available from the slave
of the header, five ECG words, and lead-off. The frame is
device). The frame/packet of words consists of the header,
configured to always send, irrespective of ready status The
five ECG words, pace, respiration magnitude, and lead-off.
slave ADAS1000-2 is in electrode mode format with a data
In this example, the frame is configured to always send
rate of 2 kHz.
irrespective of ready status. The master, ADAS1000, is in
2. Write 2 configures the CMREFCTL register to receive an
vector mode format with a data rate of 2 kHz. Similar
external common mode from the master.
to the slave device, the master could be configured for
3. Write 3 addresses the ECGCTL register, enabling all
electrode mode; the host controller would then be required
channels into a gain of 1.4, low noise mode. It configures
to make the lead calculations.
the device as a slave, in gang mode and driven from the
2. Write 5 configures the CMREFCTL register for CM =
CLK_IN input source (derived from master ADAS1000).
WCT = (LA + LL + RA)/3; RLD is enabled onto
The ADAS1000-2 slave is also put into conversion mode
RLD_OUT electrode. The shield amplifier is enabled.
in this write, but waits for the SYNC_GANG signal from
The CM = WCT signal is driven out of the master device
the master device before it starts converting.
(CM_OUT) into the slave device (CM_IN).
3. Write 6 addresses the ECGCTL register, enabling all
channels into a gain of 1.4, low noise mode. It configures
the device as a master in gang mode and driven from
the XTAL input source. The ADAS1000 master is set to
differential input, which places it in analog lead mode.
This ECGCTL register write puts the master into
conversion mode, where the device sends an edge on
the SYNC_GANG pin to the slave device to trigger the
simultaneous conversions of both devices.
4. Write 7 issues the read command to start putting the
converted and decimated data out on the SDO pin.
5. Continue to issue SCLK cycles to read the converted data
at the configured packet data rate.
Table 59. Example 6: Writing to Master and Slave Devices and Streaming Conversion Data
Device
Slave
Master
Write Command
Write 1
Write 2
Write 3
Write 4
Write 5
Write 6
Write 7
Register Addressed
FRMCTL
CMREFCTL
ECGCTL
FRMCTL
CMREFCTL
ECGCTL
FRAMES
R/W
1
1
1
1
1
1
0
Register Address
000 1010
000 0101
000 0001
000 1010
000 0101
000 0001
100 0000
Rev. A | Page 74 of 80
Data
0000 0111 1111 0110 0001 0000
0000 0000 0000 0000 0000 0100
1111 1000 0000 0000 1101 1110
0000 0111 1001 0110 0000 0000
1110 0000 0000 0000 0000 1011
1111 1000 0000 0100 1011 1110
0000 0000 0000 0000 0000 0000
32-Bit Write Command
0x8A07F610
0x85000004
0x81F800DE
0x8A079600
0x85E0000B
0x81F804BE
0x40000000
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
SOFTWARE FLOWCHART
Figure 81 shows a suggested sequence of steps to be taken to interface to multiple ADAS1000/ADAS1000-1/ADAS1000-2 devices.
POWER UP ADAS1000
DEVICES
WAIT FOR POR ROUTINE
TO COMPLETE, 1.5ms
INITIALIZE SLAVE
DEVICES
INITIALIZE MASTER DEVICE
ENABLING CONVERSION
ISSUE READ FRAME
COMMAND (WRITE TO 0x40)
NO
DRDY
LOW?
YES
ISSUE SCLK CYCLES (SDI = 0)
TO CLOCK FRAME DATA OUT
AT PROGRAMMED DATA RATE
IS CRC
CORRECT?
NO
DISCARD
FRAME DATA
YES
NO
ACTIVITY
ON
SDI?
YES
ADAS1000 STOPS CONVERTING,
SDI WORD USED TO
RECONFIGURE DEVICE
RETURN
TO ECG
CAPTURE?
NO
YES
ISSUE READ FRAME
COMMAND (WRITE TO 0x40)
NO
POWER-DOWN?
YES
ADAS1000 GOES INTO
POWER-DOWN MODE
09660-038
ECG CAPTURE COMPLETE
POWER-DOWN ADAS1000
ECGCTL = 0x0
Figure 81. Suggested Software Flowchart for Interfacing to Multiple ADAS1000/ADAS1000-1/ADAS1000-2 Devices
Rev. A | Page 75 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
POWER SUPPLY, GROUNDING, AND DECOUPLING
STRATEGY
The ADAS1000/ADAS1000-1/ADAS1000-2 should have ample
supply decoupling of 0.01 μF on each supply pin located as close
to the device pin as possible, ideally right up against the device.
In addition, there should be one 4.7 μF capacitor for each of the
power domains, AVDD and IOVDD, again located as close to
the device as possible. IOVDD is best split from AVDD due to
its noisy nature.
Similarly, the ADCVDD and DVDD power domains each
require one 2.2 μF capacitor with ESR in the range of 0.5 Ω to
2 Ω. The ideal location for each 2.2 μF capacitor is dependent
on package type. For the LQFP package and DVDD decoupling,
the 2.2 μF capacitor is best placed between Pin 30 and Pin 31,
while for ADCVDD, the 2.2 μF capacitor should be placed
between Pin 55 and Pin 56. Similarly for the LFCSP package,
the DVDD 2.2 μF capacitor is ideal between Pin 43 and Pin 44,
and between Pin 22 and Pin 23 for ADCVDD. A 0.01 μF
capacitor is recommended for high frequency decoupling at
each pin. The 0.01 μF capacitors should have low effective series
resistance (ESR) and effective series inductance (ESL), such as
the common ceramic capacitors that provide a low impedance
path to ground at high frequencies to handle transient currents
due to internal logic switching.
Digital lines running under the device should be avoided
because these couple noise onto the device. The analog ground
plane should be allowed to run under the device to avoid noise
coupling. The power supply lines should use as large a trace as
possible to provide low impedance paths and reduce the effects
of glitches on the power supply line. Fast switching digital
signals should be shielded with digital ground to avoid radiating
noise to other parts of the board and should never be run near
the reference inputs. It is essential to minimize noise on VREF
lines. Avoid crossover of digital and analog signals. Traces on
opposite sides of the board should run at right angles to each
other. This reduces the effects of feedthrough throughout the
board. As is the case for all thin packages, take care to avoid
flexing the package and to avoid a point load on the surface of
this package during the assembly process.
During layout of board, ensure that bypass capacitors are placed
as close to the relevant pin as possible, with short, wide traces
ideally on the topside.
ADCVDD AND DVDD SUPPLIES
The AVDD supply rail powers the analog blocks in addition to
the internal 1.8 V regulators for the ADC and the digital core.
If using the internal regulators, connect the VREG_EN pin to
AVDD and then use the ADCVDD and DVDD pins for
decoupling purposes.
The DVDD regulator can be used to drive other external digital
circuitry as required; however the ADCVDD pin is purely
provided for bypassing purposes and does not have available
current for other components.
Where overall power consumption must be minimized, using
external 1.8 V supply rails for both ADCVDD and DVDD
would provide a more efficient solution. The ADCVDD and
DVDD inputs have been designed to be driven externally and
the internal regulators may be disabled by tying VREG_EN pin
directly to ground.
UNUSED PINS/PATHS
In applications where not all ECG paths or functions might be
used, the preferred method of biasing the different functions is
as follows:
•
•
•
•
Unused ECG paths power up disabled. For low power
operation, they should be kept disabled throughout
operation. Ideally, these pins should be connected to
RLD_OUT if not being used.
Unused external respiration inputs can be tied to ground if
not in use.
If unused, the shield driver can be disabled and output left
to float.
CM_OUT, CAL_DAC_IO, DRDY, GPIOx, CLK_IO,
SYNC_GANG can be left open.
LAYOUT RECOMMENDATIONS
To maximize CMRR performance, pay careful attention to
the ECG path layout for each channel. All channels should be
identical to minimize difference in capacitance across the paths.
Place all decoupling as close to the ADAS1000/ADAS1000-1/
ADAS1000-2 devices as possible, with an emphasis on ensuring
that the VREF decoupling be prioritized, with VREF decoupling
on the same side as the ADAS1000/ADAS1000-1/ADAS1000-2
devices, where possible.
AVDD
While the ADAS1000/ADAS1000-1/ADAS1000-2 are
designed to operate from a wide supply rail, 3.15 V to 5.5 V,
the performance is similar over the full range, but overall
power increases with increasing voltage.
Rev. A | Page 76 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
OUTLINE DIMENSIONS
9.10
9.00 SQ
8.90
0.60
0.42
0.24
0.60
0.42
0.24
0.275
43
PIN 1
INDICATOR
8.75
BSC SQ
6.05
5.95 SQ
5.85
*EXPOSED PAD
1
0.50
BSC
0.75
0.65
0.55
29
14
15
28
TOP VIEW
BOTTOM VIEW
6.50 REF
0.70 MAX
0.65 NOM
12° MAX
0.05 MAX
0.01 NOM
0.30
0.23
0.18
0.20 REF
06-20-2012-A
SEATING
PLANE
*FOR PROPER CONNECTION OF THE EXPOSED PAD, REFER TO
THE PIN CONFIGURATIONS AND FUNCTION DESCRIPTIONS
SECTION OF THIS DATA SHEET.
Figure 82. 56-Lead, Lead Frame Chip Scale Package [LFCSP_VQ]
9 mm × 9 mm Body, Very Thin Quad
(CP-56-7)
Dimensions shown in millimeters
0.75
0.60
0.45
12.20
12.00 SQ
11.80
1.60
MAX
64
49
1
48
PIN 1
10.20
10.00 SQ
9.80
TOP VIEW
(PINS DOWN)
1.45
1.40
1.35
0.15
0.05
SEATING
PLANE
VIEW A
0.20
0.09
7°
3.5°
0°
0.08
COPLANARITY
16
33
32
17
VIEW A
0.50
BSC
LEAD PITCH
0.27
0.22
0.17
ROTATED 90° CCW
COMPLIANT TO JEDEC STANDARDS MS-026-BCD
Figure 83. 64-Lead Low Profile Quad Flat Package [LQFP]
(ST-64-2)
Dimensions shown in millimeters
Rev. A | Page 77 of 80
051706-A
0.90
0.85
0.80
0.150
56
42
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
ORDERING GUIDE
Model 1
ADAS1000BSTZ
ADAS1000BSTZ-RL
ADAS1000BCPZ
ADAS1000BCPZ-RL
ADAS1000-1BCPZ
ADAS1000-1BCPZ-RL
ADAS1000-2BSTZ
ADAS1000-2BSTZ-RL
ADAS1000-2BCPZ
ADAS1000-2BCPZ-RL
EVAL-ADAS1000SDZ
EVAL-SDP-CB1Z
Description
5 ECG Channels, Pace Algorithm, Respiration Circuit
5 ECG Channels, Pace Algorithm, Respiration Circuit
5 ECG Channels, Pace Algorithm, Respiration Circuit
5 ECG Channels, Pace Algorithm, Respiration Circuit
5 ECG Channels
5 ECG Channels
Companion for Gang Mode
Companion for Gang Mode
Companion for Gang Mode
Companion for Gang Mode
ADAS1000 Evaluation Board
System Demonstration Board (SDP), used as a controller
board for data transfer via USB interface to PC
Temperature
Range
−40°C to +85°C
−40°C to +85°C
−40°C to +85°C
−40°C to +85°C
−40°C to +85°C
−40°C to +85°C
−40°C to +85°C
−40°C to +85°C
−40°C to +85°C
−40°C to +85°C
Package Description
64-Lead LQFP
64-Lead LQFP
56-Lead LFCSP_VQ
56-Lead LFCSP_VQ
56-Lead LFCSP_VQ
56-Lead LFCSP_VQ
64-Lead LQFP
64-Lead LQFP
56-Lead LFCSP_VQ
56-Lead LFCSP_VQ
Evaluation Kit 2
Controller Board 3
Package
Option
ST-64-2
ST-64-2
CP-56-7
CP-56-7
CP-56-7
CP-56-7
ST-64-2
ST-64-2
CP-56-7
CP-56-7
1
Z = RoHS Compliant Part.
2
This evaluation kit consists of ADAS1000BSTZ × 2 for up to 12-lead configuration. Because the ADAS1000 contains all features, it is the evaluation vehicle for
all ADAS1000 variants.
3
This board allows a PC to control and communicate with all Analog Devices evaluation boards ending in the SD designator.
Rev. A | Page 78 of 80
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
NOTES
Rev. A | Page 79 of 80
ADAS1000/ADAS1000-1/ADAS1000-2
Data Sheet
NOTES
©2012–2013 Analog Devices, Inc. All rights reserved. Trademarks and
registered trademarks are the property of their respective owners.
D09660-0-1/13(A)
Rev. A | Page 80 of 80