GSM90B series

90W AC-DC Single Output Medical Adaptor
GSM90B
series
EN 60601-1
IEC 60601-1
8
■ Features
Universal AC input / Full range
2 pole AC inlet IEC320-C8
Built-in active PFC function, PF>0.91
High efficiency up to 91%
Low leakage current <100 A
Protections: Short circuit / Overload / Over voltage/
■ Applications
Mobile clinical workstation
Oral irrigator
Portable hemodialysis machine
Breath Machine
Medical computer monitor
Over temperature
Fully enclosed plastic case
Medical safety approved (2xMOPP between primary
to secondary)
Class Ⅱ power (without earth pin)
LED indicator for power on
No load power consumption<0.15W
ErP step2 compliant (level V)
Meet EISA 2007 (Energy Independence and Security Act)
100% full load burn-in test
Optional lock type DC plug
Suitable for BF application with appropriate system consideration
3 years warranty
■ Description
GSM90B is a highly reliable, 90W single-output green medical adaptor series. This product is equipped
with a 2-pin (no FG) standard IEC320-C8 power plug, adopting the input range from 80VAC to 264VAC.
The entire series supplies different output voltages between 12VDC and 48VDC that can satisfy the demands
for various kinds of medical electrical devices. The circuitry design meets the international medical standards
(2*MOPP), having an ultra low leakage current (<100 A), fitting the medical devices in direct electrical
contact with the patients.
With the efficiency up to 91% and the extremely low no-load power consumption below 0.15W, the design
of GSM90B observes the latest energy regulation (Level V); the supreme feature allows the adaptor to
save the energy when it is either under the operating mode or the standby mode. The entire series utilizes
the 94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical
shock. GSM90B is approved with the international medical safety certificates.
■ Model Encoding
GSM 90 B 12 - P1M
DC plug type
Output voltage
IEC320-C8 AC inlet
Output wattage
Series name
File Name:GSM90B-SPEC 2015-02-03
GSM90B
90W AC-DC Single Output Medical Adaptor
series
SPECIFICATION
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE
OUTPUT
48V
19V
24V
6A
4.74A
3.75A
1.87A
CURRENT RANGE
0 ~ 6.67A
0 ~ 6A
0 ~ 4.74A
0 ~ 3.75A
0 ~ 1.87A
RATED POWER (max.)
80W
90W
90W
90W
90W
RIPPLE & NOISE (max.) Note.3 120mVp-p
150mVp-p
180mVp-p
200mVp-p
240mVp-p
VOLTAGE TOLERANCE Note.4 ±5.0%
±5.0%
±4.0%
±3.0%
±2.5%
Note.5 ±1.0%
±1.0%
±1.0%
±1.0%
±1.0%
±5.0%
±5.0%
±4.0%
±3.0%
±2.5%
90%
91%
Note.6 1000ms, 50ms / 230VAC
1500ms, 50ms / 115VAC at full load
HOLD UP TIME (Typ.)
VOLTAGE RANGE
20ms / 230VAC
20ms / 115VAC at full load
80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE
47 ~ 63Hz
POWER FACTOR (Typ.)
PF>0.91 / 230VAC
EFFICIENCY (Typ.)
88%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.3A / 115VAC
OVERLOAD
OVER VOLTAGE
PF>0.95 / 115VAC at full load
89%
30A / 115VAC
65A / 230VAC
Touch current < 100 A/264VAC
110 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
105 ~ 135% rated output voltage
Protection type : Shut down o/p voltage, re-power on to recover
Shut down o/p voltage, re-power on to recover
WORKING TEMP.
WORKING HUMIDITY
-30 ~ +60℃ (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
SAFETY STANDARDS
SAFETY & ISOLATION LEVEL
WITHSTAND VOLTAGE
EMC
(Note. 7)
ISOLATION RESISTANCE
89%
0.6A / 230VAC
OVER TEMPERATURE
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
NOTE
GSM90B48-P1M
GSM90B48
15V
LEAKAGE CURRENT(max.)
CONNECTOR
GSM90B24-P1M
GSM90B24
6.67A
SETUP, RISE TIME
OTHERS
GSM90B19-P1M
GSM90B19
Note.2 12V
LOAD REGULATION
PROTECTION
GSM90B15-P1M
GSM90B15
RATED CURRENT
LINE REGULATION
INPUT
GSM90B12-P1M
GSM90B12
-40 ~ +85℃, 10 ~ 95% RH
±0.03% / ℃ (0 ~ 40℃)
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1 / ES60601-1-11, TUV EN60601-1 / EN60601-1-11 approved
Primary-Secondary: 2xMOPP
I/P-O/P: 4KVAC
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
EMC IMMUNITY
Compliance to EN55011(CISPR11) class B, EN61000-3-2,3, FCC PART 15 class B
MTBF
DIMENSION
PACKING
PLUG
405.6K hrs min. MIL-HDBK-217F(25℃)
145*60*32mm (L*W*H)
0.45Kg; 30pcs/14.5Kg/1CUFT
See page 2 ; Other type available by customer requested
CABLE
See page 2 ; Other type available by customer requested
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3 medical level, criteria A
1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1uf & 47uf capacitor.
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
File Name:GSM90B-SPEC 2015-02-03
GSM90B
90W AC-DC Single Output Medical Adaptor
Derating Curve
series
Static Characteristics
100
100
90
80
60
70
LOAD (%)
LOAD (%)
50
20
-30
0
10
20
30
40
50
60
70
(HORIZONTAL)
60
50
40
80
AMBIENT TEMPERATURE ( )
90
100
110
120
140
160
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
Mechanical Specification
Case No. GS90A
Unit:mm
UL1185 14AWG 1000±50mm for 12 ~ 15V
UL1185 16AWG 1200±50mm for 19 ~ 48V
145
32
180
70 10mm
60
C"+"
11 0.5mm
POWER LED
ID 2.5 x OD 5.5
Outside
Inside
Plug Assignment
Standard plug: P1M
P1M
P/N
OUTPUT
CENTER
+
Optional lock type plug: P2S
SWITCHCRAFT S761K plug equivalent
Installation Manual
Please refer to : http://www.meanwell.com/webnet/search/InstallationSearch.html
File Name:GSM90B-SPEC 2015-02-03