NI PXI-4132 Specifications

NI PXI-4132 Specifications
Precision Source-Measure Unit
This document provides the specifications for the NI PXI-4132 precision source-measure unit (SMU). Specifications are subject
to change without notice. For the most recent NI PXI-4132 specifications, visit ni.com/manuals.
National Instruments defines the capabilities and performance of its Test & Measurement instruments as Specifications, Typical
Specifications, and Characteristic or Supplemental Specifications. Data provided in this document are Specifications unless
otherwise noted.
Specifications characterize the warranted performance of the instrument within the recommended calibration interval and under
the stated operating conditions.
Typical Specifications are specifications met by the majority of the instrument within the recommended calibration interval and
under the stated operating conditions. The performance of the instrument is not warranted.
Characteristic or Supplemental Specifications describe basic functions and attributes of the instrument established by design or
during development and not evaluated during Verification or Adjustment. They provide information that is relevant for the
adequate use of the instrument that is not included in the previous definitions.
Unless otherwise noted, specifications are valid under the following conditions:
•
Ambient temperature 23 °C ± 5 °C
•
After 30 minute warm-up time
•
niDCPower Auto Zero property/attribute set to On
To access the NI PXI-4132 documentation, navigate to Start»Programs»National Instruments»NI-DCPower»Documentation.
Device Capabilities
Table 1 and Figure 1 illustrate the voltage and the current
source and sink ranges of the NI PXI-4132.
I+
100mA
Table 1. NI PXI-4132 Current Source and Sink Ranges
DC Voltage
Ranges (CAT I)
±10 V
±100 V
DC Current Source
and Sink Ranges
10 μA
100 μA
1 mA
10 mA
100 mA
Caution Shock hazards exist when voltage levels
are greater than 30 VRMS, 42.4 V peak, or 60 VDC.
Use extreme caution when a shock hazard is present.
Always ensure the test system is de-energized
before connecting or disconnecting the backshell
assembly or cables from the NI PXI-4132.
Quadrant II
Sink
Quadrant I
Source
50mA
20mA
V–
–100V
20V 40V
–20mA
–40V –20V
–50mA
Quadrant III
Source
V+
100V
Quadrant IV
Sink
–100mA
I–
Figure 1. NI PXI-4132 Quadrant Diagram, characteristic
SMU Specifications1
Voltage Programming Accuracy/Resolution
Accuracy ± (% of output + offset)
Range
1 Year
23 °C ± 5 °C
Resolution
Peak to Peak Noise, typical
(0.1 Hz to 10 Hz)
±10 V
50 μV
0.025% + 3.0 mV
70 μV
±100 V
500 μV
0.025% + 10 mV
300 μV
Current Programming Accuracy/Resolution2
Accuracy ± (% of output + offset)
Range
1 Year
23 °C ± 5 °C
Resolution
Peak to Peak Noise, typical
(0.1 Hz to 10 Hz)
10 μA
500 pA
0.034% + 2.0 nA
90 pA
100 μA
5 nA
0.034% + 20 nA
900 pA
1 mA
50 nA
0.034% + 0.2 μA
9 nA
10 mA
500 nA
0.034% + 2.0 μA
90 nA
100 mA
5 μA
0.034% + 20 μA
900 nA
Voltage Measurement Accuracy/Resolution
Accuracy ± (% of reading + offset)
Range
1 Year
23 °C ± 5 °C
Resolution
±10 V
10 μV
0.02% + 2.0 mV
±100 V
100 μV
0.02% + 5.0 mV
Current Measurement Accuracy/Resolution
Accuracy ± (% of reading + offset)
Range
1
2
1 Year
23 °C ± 5 °C
Resolution
10 μA
10 pA
0.028% + 1.0 nA
100 μA
100 pA
0.028% + 10 nA
1 mA
1 nA
0.028% + 0.1 μA
10 mA
10 nA
0.028% + 1.0 μA
100 mA
100 nA
0.020% + 10 μA
Temperature coefficient (Tempco) is 15% of accuracy specification per °C.
Minimum programmable current limit/level is 2% of range.
NI PXI-4132 Specifications
2
ni.com
Additional Specifications
Figures 2 through 4 illustrate the step response of the
NI PXI-4132 for different loads.
Settling Time, typical...................... <300 μs
Settled to 0.1% of final
value (1 V step at 50%
load of current range)
1.2
1
Amplitude (Volts)
Transient Response, typical ............ Recovers to <0.1% of
voltage range within
100 μs after a change in
load current from 10% to
90% of current range
Wideband Source Noise,
typical.............................................. 8 mVp-p into resistive
load <1 mV RMS (20 Hz
to 20 MHz bandwidth)
0.8
0.6
0.4
0.2
<50 pF
3.3 μF
10 μF
0
–0.2
0
Remote Sense.................................. Add 0.5% of HI lead drop
to voltage accuracy
specification
Maximum Lead Drop ................. Up to 1 V drop per lead
0.5
1.0
1.5
2.0
Time (milliseconds)
Figure 2. 100 mA Range Step Response, typical
Load Regulation
Voltage ....................................... 0.5 mV per mA of output
load using Local Sense
Current........................................ 0.01% of range per volt
of output change
1.2
Amplitude (Volts)
1
Guard Offset Voltage, typical ......... <4 mV
(Current ≤ 10 mA)
Isolation Voltage (continuous), characteristic
Channel-to-earth Ground............ 150 VDC, CAT I,
verified by dielectric
withstand test, 5 s
0.8
0.6
0.4
0.2
<50 pF
330 nF
1 μF
0
–0.2
0
0.1
0.2
0.3
0.4
0.5
Time (milliseconds)
Caution Do not connect to MAINs. Do not
connect to signals or use for measurements within
CAT II, III, or IV.
Figure 3. 10 mA Range Step Response, typical
1.2
Amplitude (Volts)
1
0.8
0.6
0.4
10 μA
0.2
100 μA
1 mA
<50 pF <50 pF <50 pF
330 pF 3.3 nF 33 nF
10 nF 100 nF
1 nF
0
–0.2
0
0.1
0.2
0.3
0.4
0.5
Time (milliseconds)
Figure 4. 1 mA, 100 μA, and 10 μA Range Step
Response, typical
© National Instruments Corporation
3
NI PXI-4132 Specifications
RMS (ppm of range)
600
Voltage
Current
10
60
1
6
0.1
0.01
0.6
0.1
1
Peak-to-Peak (ppm of range)
Output Triggers (Events)
100
Types .......................................... Source Complete,
Sequence Iteration
Complete, Sequence
Engine Done, Measure
Complete
Destinations ................................ PXI Trigger lines 0–73
Polarity................................... Configurable
Pulse Width............................ Configurable between
150 ns and 1.6 μs
10
Figure 6 illustrates the programming flow in NI-DCPower
using Sequence source mode with automatic measurements.
For more information about programming the NI PXI-4132,
refer to the NI DC Power Supplies and SMUs Help.
NPLC
Figure 5. Measurement Noise vs. Aperture Time, typical
Supplemental Specifications
Sequence
Loop
Count
Measurement Speed1
Num
Steps
Table 2. Maximum Operation Rates Per Second for
60 Hz (50 Hz)
ADC Aperture
Time
Measure
to Host
3490 (2900)
1900 (1700)
1/8 PLC
470 (390)
425 (360)
1 PLC
59.9 (49.9)
59.0 (49.3)
“Step”
Source
Trigger
Start Sequence
Trigger Advance
Trigger
Source-measure
to Host2
1/64 PLC
“Sequence”
Source
Source
Measure
Delay
Source
Complete
Event
Measure
Complete
Event
Figure 6. NI-DCPower Programming Flow
Maximum Source Update Rate2 ......4,200 updates/s
Additional Information
Trigger In to Source Delay ..............500 ns
Recommended Calibration
Interval ............................................ One year
Triggers
Physical Characteristics
Input Triggers
Dimensions............................................ 3U, one-slot,
PXI/cPCI module;
2.0 cm × 13.0 cm × 21.6 cm
(0.8 in. × 5.1 in. × 8.5 in.)
Types...........................................Start, Source, Sequence
Advance, Measure
Sources........................................PXI Trigger lines 0–73
Polarity ...................................Configurable
Weight............................................. 295 g (10.4 oz)
Minimum Pulse Width ...........100 ns
Destinations4 ...............................PXI Trigger lines 0–73
Front panel connectors.................... COMBICON, 5.08 mm
(8 position)
Polarity ...................................Active high
(unconfigurable)
Note Front panel connectors can accept wire
gauges from 12 AWG to 28 AWG.
Pulse Width ............................150 ns
1
2
3
4
Sequence Sequence
Iteration
Engine
Complete
Done
Event
Event
Does not include load dependent settling time; niDCPower Auto Zero property/attribute set to Off.
Source-measure to Host and Maximum Source Update Rate are performed with the source delay set to 200 μs. This is
sufficient for the output to settle within 1% of the requested level with a simple resistive load. As you adjust the source delay
for your application’s requirements, maximum rates vary.
Pulse widths and logic levels compliant with PXI specification.
Input triggers can be reexported.
NI PXI-4132 Specifications
4
ni.com
Power Requirements
PXI power requirement ...................10 W at 5 V,
1 W at 3.3 V,
2 W at 12 V
Environment
NI PXI-4132
Precision SMU
A
Maximum altitude ...........................2,000 m (at 25 °C
ambient temperature)
B
OUTPUT
STATUS
REMOTE
SENSE
Pollution Degree..............................2
100 VDC MAX
100 mA MAX
Indoor use only.
C
Operating Environment
LO
Ambient temperature range .............0 °C to 55 °C
(Tested in accordance
with IEC 60068-2-1
and IEC 60068-2-2.)
SENSE
LO
D
E
GUARD
F
HI
G
H
Relative humidity range ..................10% to 70%,
noncondensing;
derate 1.3% per °C above
40 °C
(Tested in accordance
with IEC 60068-2-56.)
GUARD
I
J
SENSE
HI
OUTPUT CAT I
Storage Environment
Ambient temperature range .............–40 °C to 70 °C
(Tested in accordance
with IEC 60068-2-1
and IEC 60068-2-2.)
K
Relative humidity range ..................5% to 95%,
noncondensing
(Tested in accordance
with IEC 60068-2-56.)
Shock and Vibration
A
B
C
D
E
F
Status Indicator
Sense Status Indicator
Output Low Terminal
Sense Low Terminal
Guard Terminal
Output High Terminal
G
H
I
J
K
Operational shock............................30 g peak, half-sine,
11 ms pulse
(Tested in accordance
with IEC 60068-2-27.
Test profile developed
in accordance with
MIL-PRF-28800F.)
Guard Terminal
Guard Terminal
Guard Terminal
Sense High Terminal
Chassis Ground
Random vibration
Figure 7. NI PXI-4132 Front Panel
Operating ....................................5 Hz to 500 Hz, 0.3 grms
Nonoperating ..............................5 Hz to 500 Hz, 2.4 grms
(Tested in accordance
with IEC 60068-2-64.
Nonoperating test profile
exceeds the requirements
of MIL-PRF-28800F,
Class 3.)
© National Instruments Corporation
5
NI PXI-4132 Specifications
Accessories
Compliance and Certifications
Visit ni.com for more information about the following
accessory.
Safety
This product meets the requirements of the following
standards of safety for electrical equipment for measurement,
control, and laboratory use:
Caution For safety, always operate the
NI PXI-4132 with suitably rated cables and the
backshell kit provided in the shipping kit. Operating
the device without the provided backshell may
expose users to high voltage.
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications,
refer to the product label or the Online Product
Certification section.
Caution The protection provided by this
equipment may be impaired if it is used in a manner
not described in the product documentation.
100V
CAT I
Electromagnetic Compatibility
This product meets the requirements of the following EMC
standards for electrical equipment for measurement, control,
and laboratory use:
Figure 8. Assembled High Voltage Backshell for the NI PXI-4132
Table 3. NI Accessory for the NI PXI-4132
Accessory
Additional Connector and
Backshell Kit for the
NI PXI-4132
Part Number (P/N)
781175-01
COMBICON,
5.08 mm
(8 position)
Description
Mating
connector
for I/O
NI PXI-4132 Specifications
Phoenix
Contact
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note For EMC compliance, operate this device
with shielded cables and the ferrite included in the
NI PXI-4132 shipping kit.
CE Compliance
This product meets the essential requirements of applicable
European Directives as follows:
Table 4. Third-Party Accessory for the NI PXI-4132
Accessory
EN 61326 (IEC 61326): Class A emissions; Basic
immunity
Note For the standards applied to assess the EMC
of this product, refer to the Online Product
Certification section.
Caution You must install mating connectors
according to local safety codes and standards
and according to the specifications provided by
the manufacturer. You are responsible for verifying
the safety compliance of third-party connectors and
their usage according to the relevant standard(s),
including UL and CSA in North America and IEC
and VDE in Europe.
Manufacturer
•
P/N
•
2006/95/EC; Low-Voltage Directive (safety)
•
2004/108/EC; Electromagnetic Compatibility Directive
(EMC)
1825378
6
ni.com
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for
additional regulatory compliance information. To obtain
product certifications and the DoC for this product, visit
ni.com/certification, search by model number or
product line, and click the appropriate link in the Certification
column.
Environmental Management
National Instruments is committed to designing and
manufacturing products in an environmentally responsible
manner. NI recognizes that eliminating certain hazardous
substances from our products is beneficial not only to the
environment but also to NI customers.
For additional environmental information, refer to the NI and
the Environment Web page at ni.com/environment. This
page contains the environmental regulations and directives
with which NI complies, as well as other environmental
information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle,
all products must be sent to a WEEE recycling
center. For more information about WEEE
recycling centers, National Instruments WEEE
initiatives, and compliance with WEEE Directive
2002/96/EC on Waste Electrical and Electronic
Equipment, visit ni.com/environment/weee.
⬉ᄤֵᙃѻક∵ᶧ᥻ࠊㅵ⧚ࡲ⊩ ˄Ё೑ RoHS˅
Ё೑ᅶ᠋ National Instruments ヺড়Ё೑⬉ᄤֵ
ᙃѻકЁ䰤ࠊՓ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ (RoHS)DŽ݇Ѣ
National Instruments Ё೑ RoHS ড়㾘ᗻֵᙃˈ
䇋ⱏᔩ ni.com/environment/rohs_chinaDŽ
(For information about China RoHS compliance,
go to ni.com/environment/rohs_china.)
© National Instruments Corporation
7
NI PXI-4132 Specifications
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on
ni.com/legal for more information about National Instruments trademarks. Other product and company names mentioned herein are
trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate
location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents.
© 2009 National Instruments Corporation. All rights reserved.
375128B
Nov09
NI PXI-4132 仕様
高精度ソースメジャーユニット
このドキュメントには、NI PXI-4132 高精度ソースメジャーユニット(SMU)の仕様が記載されています。仕様は
事前の通知なしに変更されることがあります。最新の NI PXI-4132 の仕様については、ni.com/manuals を参照し
てください。
ナショナルインスツルメンツでは、テスト・計測用機器の機能および性能を「仕様」、「標準仕様」
、および「特性
または補足仕様」として定義しています。このドキュメントに記載されたデータは注釈がない限り「仕様」です。
「仕様」は推奨キャリブレーション間隔内において、記載された動作条件下で保証される計測器の性能を示しま
す。
「標準仕様」は推奨キャリブレーション間隔内において、記載された動作条件下で大多数の計測器が満たす仕様を
示します。計測器の性能は保証されません。
「特性または補足仕様」は、設計または開発中に特定された計測器の基本的機能および属性を示し、検証または調
整中に評価されたものではありません。これには、前述の定義に含まれていない、計測器の標準的な使用に関す
る情報が記載されています。
特に注釈がない限り、これらの仕様は以下の条件に対して有効です。
•
•
•
周囲温度 : 23 ℃ ± 5 ℃
ウォームアップ時間 : 30 分
niDCPower オートゼロプロパティ / 属性を ON に設定
NI PXI-4132 のドキュメントにアクセスするには、スタート→すべてのプログラム→ National Instruments →
NI-DCPower →ドキュメントを選択します。
デバイス機能
表 1 および図 1 は、NI PXI-4132 の電圧および電流ソー
スとシンクレンジを示します。
I+
100mA
表 1 NI PXI-4132 の電流ソースおよびシンクレンジ
DC 電圧レンジ(CAT I)
±10 V
±100 V
DC 電流ソース /
╙ੑ⽎㒢
䉲䊮䉪
シンクレンジ
10 μA
100 μA
1 mA
10 mA
100 mA
注意 電圧レベルが 30 VRMS、42.4 V peak、
または 60 VDC を超える場合、感電する危険
性があります。感電の危険性がある場合は特に
注意してください。NI PXI-4132 をバックシェ
ルアセンブリまたはケーブルに接続 / 接続解
除する前に、必ずテストシステムの電源が落
ちていることを確認してください。
50mA
╙৻⽎㒢
䉸䊷䉴
20mA
V–
–100V
V+
20V 40V
–20mA
–40V –20V
–50mA
╙ਃ⽎㒢
䉸䊷䉴
100V
╙྾⽎㒢
䉲䊮䉪
–100mA
I–
図 1 NI PXI-4132 の象限図、特性
SMU 仕様1
電圧プログラミング確度 / 分解能
確度 ±(出力値の % + オフセット)
レンジ
1年
23 ℃ ± 5 ℃
分解能
ピーク間ノイズ、標準
(0.1 Hz ~ 10 Hz)
±10 V
50 μV
0.025% + 3.0 mV
70 μV
±100 V
500 μV
0.025% + 10 mV
300 μV
電流プログラミング確度 / 分解能2
確度 ±(出力値の % + オフセット)
レンジ
1年
23 ℃ ± 5 ℃
分解能
ピーク間ノイズ、標準
(0.1 Hz ~ 10 Hz)
10 μA
500 pA
0.034% + 2.0 nA
90 pA
100 μA
5 nA
0.034% + 20 nA
900 pA
1 mA
50 nA
0.034% + 0.2 μA
9 nA
10 mA
500 nA
0.034% + 2.0 μA
90 nA
100 mA
5 μA
0.034% + 20 μA
900 nA
電圧測定確度 / 分解能
確度 ±(読み取り値の % + オフセット)
レンジ
1年
23 ℃ ± 5 ℃
分解能
±10 V
10 μV
0.02% + 2.0 mV
±100 V
100 μV
0.02% + 5.0 mV
電流測定確度 / 分解能
確度 ±(読み取り値の % + オフセット)
レンジ
1年
23 ℃ ± 5 ℃
分解能
10 μA
10 pA
0.028% + 1.0 nA
100 μA
100 pA
0.028% + 10 nA
1 mA
1 nA
0.028% + 0.1 μA
10 mA
10 nA
0.028% + 1.0 μA
100 mA
100 nA
0.020% + 10 μA
1
温度係数(Tempco)は、1 ℃あたりの確度仕様の 15%。
2
プログラム可能な最小電流制限 / レベルはレンジの 2%。
NI PXI-4132 仕様
2
ni.com/jp
図 2 ~ 4 は、異なる負荷に対する NI PXI-4132 のス
テップ応答を示します。
追加仕様
整定時間、標準 ................................<300 μs 。最終値の
0.1% に整定(電流レ
ンジの 50% 負荷で 1
V ステップ)
1.2
1
過渡応答、標準 ................................ 電流レンジで 10% ~
90% の負荷電流変化
後、100 μs 内で電圧レ
ンジの <0.1% に回復
ᝄ᏷䋨V䋩
0.8
広帯域ソースノイズ、標準...........8 mVp-p (負荷抵抗)
<1 mV RMS(20 Hz ~
20 MHz 帯域幅)
0.6
0.4
0.2
<50 pF
3.3 µF
10 µF
0
リモートセンス ................................ 電圧確度仕様に 0.5%
の HI リード降下を追
加
–0.2
0
0.5
1.0
1.5
2.0
ᤨ㑆䋨䊚䊥⑽䋩
最大リード降下 ...........................1 リードにつき最大
1 V の降下
図 2 100 mA レンジステップ応答、標準
負荷変動
1.2
電圧 .................................................0.5 mV(ローカルセン
ス使用時の出力負荷の
1 mA あたり)
1
0.8
ᝄ᏷䋨V䋩
電流 ................................................. レンジの 0.01%
(出力変更の 1 ボルト
あたり)
ガードオフセット電圧、標準 .....<4 mV
(電流 ≤ 10 mA)
0.6
0.4
0.2
<50 pF
330 nF
1 µF
0
絶縁電圧(連続)、特性
–0.2
チャンネル / アース間 ..............150 VDC、CAT I、
5 秒間の耐電圧試験で
確認済み
0
0.1
0.2
0.3
0.4
0.5
ᤨ㑆䋨䊚䊥⑽䋩
図 3 10 mA レンジステップ応答、標準
注意 電源に接続しないでください。CAT II、
III、または IV で、信号を接続したり測定用に
使用したりしないでください。
1.2
1
ᝄ᏷䋨V䋩
0.8
0.6
0.4
10 µA
0.2
100 µA
1 mA
<50 pF <50 pF <50 pF
330 pF 3.3 nF 33 nF
10 nF 100 nF
1 nF
0
–0.2
0
0.1
0.2
0.3
0.4
0.5
ᤨ㑆䋨䊚䊥⑽䋩
図 4 1 mA、100 μA、および 10 μA レンジステップ
応答、標準
© National Instruments Corporation
3
NI PXI-4132 仕様
トリガ
入力トリガ
600
㔚࿶
㔚ᵹ
10
60
1
6
0.1
0.01
タイプ ............................................ 開始、ソース、シーケ
ンスアドバンス、測定
入力元 ............................................ PXI トリガライン
0 ~ 73
䊏䊷䉪㑆 䊧䊮䉳䈱ppm
RMS 䊧䊮䉳䈱ppm
100
極性 ........................................... 構成可能
最小パルス幅 .......................... 100 ns
出力先4 .......................................... PXI トリガライン
0 ~ 73
0.6
0.1
1
極性 ........................................... アクティブ HIGH
(構成不可)
10
NPLC
パルス幅 ................................... 150 ns
図 5 測定ノイズ対アパーチャ遅延、標準
出力トリガ(イベント)
タイプ ............................................ ソース完了、
シーケンス反復完了、
シーケンスエンジン完
了、測定完了
補足仕様
測定速度1
表 2 60 Hz における 1 秒あたりの最大動作レート
(括弧内は 50 Hz でのレート)
ADC アパー
チャ遅延
測定から
ホストへ
出力先 ............................................ PXI トリガライン
0 ~ 73
極性 ........................................... 構成可能
ソース測定から
ホスト 2 へ
1/64 PLC
3490 (2900)
1900 (1700)
1/8 PLC
470 (390)
425 (360)
1 PLC
59.9 (49.9)
59.0 (49.3)
パルス幅 ................................... 150 ns ~ 1.6 μs
に構成可能
最大ソースアップデート
レート2................................................ 4,200 アップデート /s
トリガ入力から
ソースまでの遅延 ........................... 500 ns
1
負荷に依存する整定時間は含まれません。niDCPower オートゼロプロパティ / 属性は OFF に設定されていま
す。
2 ソース測定からホスト、および最大ソースアップデートレートは、ソース遅延が 200 μs に設定された状態で実
行します。簡単な負荷抵抗で、要求されたレベルの 1% 以内に出力を整定するにはこれで十分です。アプリ
ケーションの要求に合わせてソース遅延を調整する場合、最大レートは一定ではありません。
3 パルス幅および論理レベルは PXI 仕様に準拠。
4
入力トリガは再エクスポート可能。
NI PXI-4132 仕様
4
ni.com/jp
図 6 は、シーケンスソースモードで自動測定を使用し
た場合の NI-DCPower のプログラミングフローを示し
ます。NI PXI-4132 でのプログラミングについては、
『NI DC 電源および SMU ヘルプ』を参照してください。
䉲䊷䉬䊮䉴
䉲䊷䉬䊮䉴
䊦䊷䊒
䉦䉡䊮䊃
㐿ᆎ
䊃䊥䉧
䉲䊷䉬䊮䉴
䉝䊄䊋䊮䉴
䊃䊥䉧
NI PXI-4132
Precision SMU
䉴䊁䉾䊒㩷
䉴䊁䉾䊒
䉴䊁䉾䊒
ᢙ
A
B
OUTPUT
STATUS
䉸䊷䉴
䊃䊥䉧
REMOTE
SENSE
100 VDC MAX
100 mA MAX
䉸䊷䉴
䉸䊷䉴
ㆃᑧ
᷹ቯ
䉸䊷䉴 ᷹ቯቢੌ
ቢੌ 䉟䊔䊮䊃
䉟䊔䊮䊃
C
䉲䊷䉬䊮䉴 䉲䊷䉬䊮䉴
෻ᓳቢੌ 䉣䊮䉳䊮
䉟䊔䊮䊃
ቢੌ
䉟䊔䊮䊃
LO
SENSE
LO
D
E
GUARD
F
図 6 NI-DCPower プログラミングフロー
HI
G
H
追加情報
GUARD
I
推奨キャリブレーション
間隔 ......................................................1 年
J
SENSE
HI
物理特性
OUTPUT CAT I
外形寸法 .............................................3U、1 スロット、
PXI/cPCI モジュール、
2.0 cm × 13.0 cm ×
K
21.6 cm
(0.8 in. × 5.1 in. ×
8.5 in.)
重量 ......................................................295 g (10.4 oz)
フロントパネルコネクタ ..............COMBICON、
5.08 mm(8 ピン)
メモ フロントパネルコネクタは、12 AWG
から 28 AWG までのワイヤゲージに対応しま
す。
A ステータスインジケータ F 出力 HIGH 端子
B SENSE ステータスインジ G ガード端子
ケータ
H ガード端子
C 出力 LOW 端子
I ガード端子
D SENSE LOW 端子
J SENSE HIGH 端子
E ガード端子
K シャーシグランド
図 7 NI PXI-4132 フロントパネル
© National Instruments Corporation
5
NI PXI-4132 仕様
所要電力
アクセサリ
PXI 所要電力 ..................................... 10 W(5 V 時)、
1 W(3.3 V 時)、
2 W(12 V 時)
以下のアクセサリの詳細については、ni.com/jp を参
照してください。
注意 安全対策として、NI PXI-4132 は出荷
キットに含まれている適切な定格ケーブルお
よびバックシェルキットを使用して操作して
ください。付属のバックシェルを取り付けず
にデバイスを操作すると、ユーザが高電圧に
さらされる恐れがあります。
環境
最大使用高度 .................................... 2,000 m
(周囲温度 25 ℃時)
汚染度 ................................................. 2
室内使用のみ。
動作環境
周囲温度範囲 .................................... 0 ~ 55 ℃
(IEC 60068-2-1/
IEC 60068-2-2 に準拠
して試験済み)
100V
CAT I
相対湿度範囲 .................................... 10 ~ 70%、結露なき
こと。40 ℃以上では
1 ℃あたり 1.3% の
ディレーティング
(IEC 60068-2-56 に準
拠して試験済み)
図 8 組み立てられた NI PXI-4132 の高電圧バックシェル
保管環境
表 3 NI PXI-4132 対応の NI アクセサリ
周囲温度範囲 ................................ –40 ℃~ 70 ℃
(IEC 60068-2-1/
IEC 60068-2-2 に準拠
して試験済み)
相対湿度範囲 .................................... 5 ~ 95%、
結露なきこと
(IEC 60068-2-56 に準
拠して試験済み)
耐衝撃 / 振動
動作時衝撃 .............................................. 最大 30 g(半正弦波)
、
11 ms パルス
(IEC 60068-2-27 に準拠
して試験済み。
MIL-PRF-28800F に準拠
してテストプロファイ
ルを確立。
)
ランダム振動
動作時 ............................................ 5 ~ 500 Hz、0.3 grms
非動作時 ........................................5 ~ 500 Hz、2.4 grms
(IEC 60068-2-64 に準
拠して試験済み。非動
作時テストプロファイ
ルは MIL-PRF-28800F、
Class 3 の要件を上回
る。)
NI PXI-4132 仕様
アクセサリ
製品番号(P/N)
NI PXI-4132 用追加コネ
クタおよびバックシェ
ルキット
781175-01
注意 地域の安全コードと基準、および製造
元によって提供された規格に従ってメイトコ
ネクタを取り付ける必要があります。他社製
コネクタの安全適合指令、また該当する基準
(北米の UL および CSA、ヨーロッパの IEC お
よび VDE を含む)に従った使用方法を確認し
てください。
表 4 NI PXI-4132 対応の他社製アクセサリ
アクセサリ
説明
製造元
P/N
COMBICON、
I/O 用メイ
Phoenix
Contact
1825378
5.08 mm
(8 ピン)
6
トコネクタ
ni.com/jp
認可および準拠
環境管理
安全性
ナショナルインスツルメンツは、環境に優しい製品の
設計および製造に努めています。NI は、製品から特定
の有害物質を除外することが、環境のみならず NI のお
客様にとって有益であると考えています。
この製品は、計測、制御、実験に使用される電気装置
に関する以下の規格および安全性の必要条件を満たし
ます。
•
•
IEC 61010-1、EN 61010-1
UL 61010-1、CSA 61010-1
環境の詳細な情報については、ni.com/environment
(英語)の NI and the Environment を参照してくださ
い。このページには、ナショナルインスツルメンツが準
拠する環境規制および指令、およびこのドキュメント
メモ UL およびその他の安全保証について
に含まれていないその他の環境に関する情報が記載さ
は、製品ラベルまたは「オンライン製品認証」
れています。
セクションを参照してください。
廃電気電子機器(WEEE)
注意 製品ドキュメントに記載されている手
順以外の方法で使用した場合、この装置に装
備されている保護機能が正常に動作しない場
合があります。
欧州のお客様へ 製品寿命を過ぎたすべての
製品は、必ず WEEE リサイクルセンターへ送
付してください。WEEE リサイクルセンター、
ナショナルインスツルメンツの WEEE への取
り組み、および廃電気電子機器に関する
WEEE 指令 2002/96/EC との準拠について
は、ni.com/environment/weee(英語)を
参照してください。
電磁両立性
この製品は、計測、制御、実験に使用される電気装置
に関する以下の EMC 規格の必要条件を満たします。
•
EN 61326 (IEC 61326): Class A エミッション、
⬉ᄤֵᙃѻક∵ᶧ᥻ࠊㅵ⧚ࡲ⊩ ˄Ё೑ RoHS˅
基本イミュニティ
•
EN 55011 (CISPR 11): Group 1、Class A エミッ
National Instruments ヺড়Ё೑⬉ᄤֵ
ᙃѻકЁ䰤ࠊՓ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ (RoHS)DŽ݇Ѣ
National Instruments Ё೑ RoHS ড়㾘ᗻֵᙃˈ
䇋ⱏᔩ ni.com/environment/rohs_chinaDŽ
Ё೑ᅶ᠋
ション
•
AS/NZS CISPR 11: Group 1、Class A エミッショ
•
•
FCC 47 CFR Part 15B: Class A エミッション
ICES-001: Class A エミッション
ン
(For information about China RoHS compliance,
go to ni.com/environment/rohs_china.)
メモ 製品の EMC 決定に適用する基準に関
しては、「オンライン製品認証」セクションを
参照してください。
メモ EMC に適合させるには、このデバイ
スを NI PXI-4132 キットに同梱されるシールド
ケーブルおよびフェライトと併用してくださ
い。
CE マーク準拠
この製品は、該当する EC 理事会指令による基本的要
件に適合しています。
•
•
2006/95/EC、低電圧指令(安全性)
2004/108/EC、電磁両立性指令(EMC)
オンライン製品認証
その他の適合規格については、適合宣言(DoC)を参
照してください。この製品の製品認証および適合宣言を
入手するには、ni.com/certification(英語)にア
クセスして型番または製品ラインで検索し、保証の欄
の該当するリンクをクリックしてください。
© National Instruments Corporation
7
NI PXI-4132 仕様
National Instruments、NI、ni.com、および LabVIEW は National Instruments Corporation
(米国ナショナルインスツルメンツ社)の商標です。National Instruments の商標の詳細については、
ni.com/legal の「Terms of Use」セクションを参照してください。本文書中に記載されたその他の
製品名および企業名は、それぞれの企業の商標または商号です。National Instruments の製品 / 技術を保
護する特許については、ソフトウェアで参照できる特許情報 ( ヘルプ→特許情報 )、メディアに含まれてい
る patents.txt ファイル、または「National Instruments Patent Notice」(ni.com/patents)のうち、
該当するリソースから参照してください。
© 2009 National Instruments Corporation. All rights reserved.
375128B-01
2009 年 11 月